FLIR SC325
Speed up your design cycle with infrared
The SC325 is designed from the ground-up to deliver
the accurate thermographic imaging and repeatable
temperature measurement necessary in demanding
science and R&D applications. Each crisp thermal image
is built from over 76,000 individual picture elements
that are sampled by the camera’s on-board electronics
and firmware. SC325 features include:
Key features and Benefits:
• Uncooled Microbolometer Detector – Maintenance-free and provides excellent
long wave imaging performance
FLIR SC325
• Optics and Focus – standard built-in 25° lens with optional 6°, 15°, 45°, and 90°
lenses available to achieve other elds of view; manual and auto focus standard
• Microscopy and Close-up Measurement – Optional 100 µm, 50 µm, and 25 µm
microscope optics and microscope stand are available for small target imaging
and measurement
• Precise Timing and Control – Optically-isolated digital I/O connections eliminate
communication latencies with digital in and out for shutter disable and program
control along with a V-sync connector for triggering and synchronization
• Plug-and-Play Compatibility – Ideal system integration solution with universal plug
and play and GigE Vision Control protocols. The camera can be fully congured
from the PC, allowing command, control, and collection of full frame data from the
camera in real time.
• Versatility – Compact, rugged, and lightweight with straightforward 3-sided
mounting feature that permits quick installation and easy movement for new
application requirements
• Fast Data Transfer – Equipped with an RJ-45 gigabit Ethernet connection that
supplies a 16-bit 320 x 240 images at rates as high as 60 Hz along with linear
temperature data; GenICam and GigE Vision Compliant
• Tailored to Your Application – FLIR Systems offers a complete line of accessories
including optics, enclosures, data systems, and software tools to suit the most
demanding applications.
320 x 240 pixels IR resolution and thermal sensitivity of
50mK provides optimized image details and temperate
difference information.
Infrared measurement allows you to see a thermal
problem and measure temperature over surfaces
accurately.
Typical applications:
The FLIR SC325 camera is an excellent choice for those who want to work in R&D
and need high frame rates but for whom 320 x 240 pixels resolution is sufcient. For
those who need to use camera in R&D, it is highly recommended to use the FLIR
ResearchIR software.
FLIR SC325 Technical Specifications
Imaging and optical data
Field of view (FOV)
Minimum focus distance
Focal length
Spatial resolution (IFOV)
Lens identication
F-number
Thermal sensitivity / NETD
Image frequency
Focus
Detector data
Detector type
Spectral range
IR resolution
Detector pitch
Detector time constant
Measurement
Object temperature range
Accuracy
Measurement analysis
Atmosphere transmission correction
Optics transmission correction
Emissivity correction
Reected apparent temperature
correction
External optics / windows correction
Measurement corrections
Ethernet
Ethernet
Ethernet, type
Ethernet, standard
Ethernet, connector type
Ethernet, communication
Ethernet, image streaming
Ethernet, protocols
Digital input / output
Digital input, purpose
Digital input
Digital output, purpose
Digital output
Digital I/O, isolation voltage
Digital I/O, supply voltage
Digital I/O, connector type
Power system
External power operation
External power, connector type
Voltage
25° x 18.8°
0.4 m
18 mm
1.36 mrad
Automatic
1.3
<0.05°C @ + 30°C / 50 mK
60 Hz
Automatic or manual (built in motor)
Focal Plane Array (FPA), uncooled
microbolometer
7.5-13 µm
320 x 240 pixels
25 µm
Typical 12 ms
-20°C to +120°C
0 °C to +350°C
±2°C or ±2% of reading
Automatic, based on inputs for distance,
atmospheric temperature and relative
humidity
Automatic, based on signals from
internal sensors
Variable from 0.01 to 1.0
Automatic, based on input of reected
temperature
Automatic, based on input of optics /
windowtransmission and temperature
Global object parameters
Control and image
Gigabit Ethernet
IEEE 802.3
RJ-45
TCP/IP socket-based FLIR proprietary
and GenICam protocol
16-bit 320 x 240 pixels @ 60 Hz
- Signal linear
- Temperature linear
- Radiometric
GigE Vision and GenICam compatible
TCP, UDP, SNTP, RTSP, RTP, HTTP, ICMP,
IGMP, ftp, SMTP, SMB (CIFS), DHCP,
MDNS (Bonjour), uPnP
Image tag (start, stop, general), Image
ow ctrl. (Stream on/off), Input ext.
device (programmatically read)
2 opto-isolated, 10-30 VDC
Output to ext. device
(programmatically set)
2 opto-isolated, 10-30 VDC, max 100 mA
500 VRMS
12/24 VDC, max 200 mA
6-pole jackable screw terminal
12/24 VDC, 24W absolute max
2-pole jackable screw terminal
Allowed range 10-30 VDC
Environmental data
Operating temperature range
Storage temperature range
Humidity (operating and storage)
EMC
Encapsulation
Bump
Vibration
-15°C to +50°C
-40°C to +70°C
IEC 60068-2-30/24 h 95% relative humidity
+25°C to +40°C
• EN 61000-6-2:2001 (Immunity)
• EN 61000-6-3:2001 (Emission)
• FCC 47 CFR Part 15 Class B (Emission)
IP 40 (IEC 60529)
25 g (IEC 60068-2-29)
2 g (IEC 60068-2-6)
Physical data
Weight
Camera size (L x W x H)
Tripod mounting
Base mounting
0.7 kg
170 x 70 x 70 mm
UNC1/4”-20 (on three sides)
2 x M4 thread mounting holes
(on three sides)
Housing material
Aluminium
Scope of delivery
Hard transport case or cardboard box
Infrared camera with lens
Calibration certicate
Ethernet™ cable
Mains cable
Power cable, pig-tailed
Power supply
Printed Getting Started Guide
Printed Important Information Guide
User documentation CD-ROM
Utility CD-ROM
Warranty extension card or Registration card
Optional Accessories
IR lens f = 30 mm, 15° incl. case
IR lens f = 10 mm, 45° incl. case
Close-up 4× (100 µm) incl. case
Close-up 2× (50 µm) incl. case
Lens 76 mm (6°) with case and mounting support for A/SC3XX
Lens 4 mm (90°) with case and mounting support for A/SC3XX
Close-up 1× (25 µm) incl. case and mounting support for A/SC3XX
High temp. option +1200°C/+2192°F for FLIR T/B2XX to T/B4XX and A/SC3XX Series
Power supply for A/SC3XX and A/SC6XX
Power cord EU
Power cord US
Power cord UK
Ethernet cable CAT-6, 2m/6.6 ft.
Power cable, pig-tailed
Hard transport case for A/SC3XX and A/SC6XX series
Delivery Box for A/SC3XX
Recommended softwares for documentation and analysis:
- ThermoVision(TM) System Developers Kit
- FLIR ResearcherIR
- FLIR QuickPlot
Specications and prices subject to change without notice. Copyright © 2010 FLIR Systems. All right reserved including the right of reproduction in whole or in part in any form.
Asia Pacific Headquarter
Hong Kong
FLIR Systems Co Ltd.
Room 1613 – 16, Tower 2
Grand Central Plaza
138 Shatin Rural Committee
Road, N.T, Hong Kong
Tel: +852 2792 8955
Fax: +852 2792 8952
Email: ir@ir.com.hk
China Head Office - Shanghai
FLIR Systems (Shanghai) Co., Ltd
Tel: +86 21 5169 7628
Fax: +86 21 5466 0289
e-mail: info@ir.cn
Japan Office - Tokyo
FLIR Systems Japan K.K.
Tel: +81 3 6277 5681
Fax: +81 3 6277 5682
e-mail: info@ir.jp
Korea Office - Seoul
FLIR Systems Korea Co., Ltd
Tel: +82 2 565 2714
Fax: +82 2 565 2718
e-mail: sales@irkorea.com
Taiwan Representative Office
Tel: +886 2 27579662
Fax: +886 2 27576723
e-Mail: ir@ir.com.hk
India Representative Office
Tel: +91 11 4606 7100
Fax: +91 11 4606 7110
e-mail: ir@ir.com.hk
100930 SC325 datasheet_en
www.flir.com/thg