Flextest FL2005.1 Instruction Manual

FLEXTEST
Multifunctional Tester
FL2005.1
INSTRUCTION MANUAL
CONTENTS
1. SAFETY INFORMATION..................................................1
2. SPECIFICATIONS …........................................................2
3. PARTS & CONTROLS......................................................4
4. BEFORE OPERATION
......................................................7
5.
V VOLTAGE MEASUREMENT........................…........9
6. F
LOOR PROBE RESISTANCE MEASUREMENT...…..…....9
7. HEATING CABLE RESISTANCE <1000
MEASUREMENT
.....………..………………….…………...9
8. MΩ INSULATION RESISTANCE MEASUREMENT......10
9. BATTERY REPLACEMENT..................................…......11
10. MAINTENANCE & CLEANING.....................................11
1. SAFETY INFORMATION
The circuit under test must be de-energized and isolated before
connections are made except for voltage measurement.
Verify operation prior to measuring hazardous voltages (voltage
above 30V AC rms, 42V AC peak and 60V DC). Do not touch the circuit connections during a test. Disconnect the live test lead before disconnecting the neutral
test lead.
After insulation tests, to protect from electric shock, capacitive
circuits must to be discharged.
Do not use the meter if the low battery indicator (
BT
) is
displayed.
Test leads (including crocodile clips) must be in good order,
clean and not broken nor cracked.
Do not use the meter if it looks damaged.
Do not use the meter around explosive gas, vapor or dust. Do not push test button before all connection and preparation is
done. The instrument must only be used by suitably trained and
competent persons. Do not use the meter with any parts or cover removed. Do not use the meter in a wet environment.
U.S. PAT. NO. 478,017 JAPAN PAT. NO. 1180870 CHINA PAT. NO. ZL02367250.1
Warnings and Safety symbols:
Caution refer to this manual before using the meter.
Dangerous voltages.
Meter is protected throughout by double insulation or reinforced insulation.
Comply with IEC1010-1
When servicing, use only specified replacement parts.
2. SPECIFICATIONS
2-1 General Information
Environment conditions :
c Installation Categories III 1000V d
Pollution Degree 2
e Altitude up to 2000 meters f Indoor use only
Safety Meets of IEC61010-1 and IEC61557
Display : Dual display, 3-3/4 Digital readout with analog bar
indication.
Sampling Rate : 1 sample/sec.
Manual data Memory and Read : Memory capacity 9 set. Over Range Indicator :
” will be displayed.
Low Battery Indication :
The (
BT
) will be displayed when the battery voltage drops
below the operating voltage.
Operating Temperature and Humidity:
0℃ to 50℃ (32℉ to 122℉) below 80% RH (noncondensing).
Temperature Coefficient : 0.10 x (specified accuracy)/ ℃
Storage Temperature and Humidity :
-10℃ to 60℃ (14℉ to 140℉) below 70% RH (noncondensing)
Battery : 6 x 1.5V Size “AA” battery.
Dimensions : 235 (L) x 116 (W) x 54(H) mm ,
(9.3”L x 4.6”W x 2.1”H)
Weight : Approx. 520g (1.15 LB), including battery
Accessories : Test leads, 6pcs batteries, carrying case,
instruction manual, reference card
2
2-2 Electrical Specifications
Accuracies are specified as:
±(...% of reading + ...digits) at 23℃ ±5℃,below 80% RH.
Insulation Resistance (MΩ)
Range Resolution Accuracy Test Voltages
250V 2MΩ/20MΩ/ 200MΩ/ 2GΩ/4GΩ
2MΩ: 1KΩ 20MΩ: 10 KΩ 200MΩ:100 KΩ
2GΩ:1MΩ
10GΩ:10MΩ
250V+30% ~ -0%
500V 4MΩ/40MΩ/ 400MΩ/2GΩ/ 4GΩ
4MΩ: 1KΩ
40MΩ: 10KΩ
400MΩ:100 KΩ
2GΩ:1MΩ
10GΩ:10MΩ
500V+30% ~ -0%
1000V 4GΩ
1MΩ
3%+5
(<1000M)
5%+5
(>1000M)
1000V+30% ~ -0%
Analog Bar Graph
0 to 10GΩ
Nominal Current
1mA
H
EATING CABLE
Resistance <1000
Range Resolution Accuracy
Max. open Circuit
Voltage
Overload
Protection
999.9Ω0.1Ω
1%+10
3V
600Vrms
F
LOOR PROBE
Resistance
Range Resolution Accuracy
Max. open
Circuit Voltage
Overload
Protection
99.99KΩ 0.01KΩ
1%+5
3V
600Vrms
3
~V AC Voltage (40Hz~500Hz)
Range Resolution Accuracy Input Impedance
Overload
Protection
999V 1V 2%+3
9MΩ
1000Vrms
Sensitivity : 2V
V DC Voltage
Range Resolution Accuracy Input Impedance
Overload
Protection
999V 1V 2%+3
9MΩ
1000Vrms
Sensitivity : 2V
3. PARTS & CONTROLS
1
2
3
6
4
10
5
7
8
9
4
Loading...
+ 4 hidden pages