Flexim PIOX R400, PIOX R400B, PIOX R400C, PIOX R400A2 User Manual

Page 1
User Manual UMPIOXR400V2-7-2EN
Sensor at riser pipe
Sensor
Process Refractometer
PIOX® R400
PIOX® R400B
PIOX® R400C
PIOX® R400A2
Page 2
PIOX® is a protected trademark of FLEXIM GmbH.
FLEXIM GmbH
Wolfener Straße 36
12681 Berlin
Germany
Tel.: +49 (30) 936 67 660
Fax: +49 (30) 936 67 680
E-mail: flexim@flexim.de
www.flexim.com
User manual for PIOX R400
UMPIOXR400V2-7-1EN, 2012-07-01
Copyright (©) FLEXIM GmbH 2012
Subject to change without notification.
Page 3
UMPIOXR400V2-7-1EN, 2012-07-01
Table of Contents
1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
1.1 Regarding this Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.2 Safety Instructions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.3 Suitability and Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2 Measuring Principle of PIOX R400. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7
2.1 Determining the Concentration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2.2 Temperature Compensation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
3 Description of PIOX R400 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
3.1 Transmitter Design . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
3.2 Sensor Design. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
4Handling. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11
4.1 First Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
4.2 Type and Serial Number . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
4.3 Firmware Version . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
4.4 General Precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
4.5 Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
4.6 Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
4.7 Calibration and Correction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
5 Selecting the Measuring Point . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13
6 Installation of PIOX R400, R400B, R400C, R400A2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15
6.1 Measuring Setups . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
6.2 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
6.3 Connection of Sensor PIOX R400 with Transmitter PIOX TR374. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
6.4 Connection of Sensor PIOX R400 with Transmitter PIOX TR379. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
6.5 Connection of Sensor PIOX R400B, R400C, R400A2 with Transmitter PIOX TR374B, TR374C,
TR374A2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .22
6.6 Connection of the Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
6.7 Connection of the Outputs. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
6.8 Connection of the Serial Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
7 Start-Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .30
7.1 Keyboard . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
7.2 Displays. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
8 Configuration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .33
8.1 Selection of the Physical Quantity and of the Unit of Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
8.2 Input of Offsets . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
8.3 Additional Input Values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
9Outputs. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .36
9.1 Installation of an Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
9.2 Output Range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
9.3 Error Output. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
9.4 Function Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
10 Serial Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .41
10.1 Definition of the Storage Rate and of the Physical Quantities to be Output . . . . . . . . . . . . . . . . . . . . . . . 41
10.2 Format of the Serial Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
10.3 Transmission Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
10.4 Format Settings of the Serial Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
Page 4
UMPIOXR400V2-7-1EN, 2012-07-01
11 Output Options. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
11.1 Damping Factor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
11.2 Activation of the Serial Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
11.3 Activation of the Outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
11.4 Status Output (Alarms). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
11.5 Activation of the Inputs. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
12 Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
12.1 Explanation of the Status Line . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
12.2 Service Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
12.3 Reset of the Status Outputs. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
12.4 Protection against Interruption . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
13 Program RMKoeff. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
13.1 Create a New Data Record . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
13.2 Calculation of Characteristics and Coefficients . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
13.3 Transmission of a Data Record . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
13.4 Displaying the Fluid on the Transmitter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
14 Inputs (Option) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
14.1 Configuring the Inputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
14.2 Assignments of Inputs to Measuring Channels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
14.3 Activation of the Inputs. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
14.4 Functional Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
A Technical Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
B Start-Up and Maintenance Protocols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
C Certificates . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
Page 5
UMPIOXR400V2-7-1EN, 2012-07-01 5
1 Introduction
1 Introduction
1.1 Regarding this Manual
This user manual has been written for the personnel operating the process refractometer PIOX R400. It contains important information about the instrument, how to handle it correctly, how to avoid damages.
Make sure you have read and understood the safety instructions (see document SIPIOX_R400) and the user man­ual before using the instrument.
All reasonable effort has been made to ensure the correctness of the content of this manual. Should you however find some erroneous information, please inform us. We will be grateful for any suggestions and comments regard­ing the concept and your experience working with PIOX R400. This will ensure that we can further develop our products for the benefit of our customers and in the interest of technological progress. If you have any suggestions about improving the documentation and particularly this user manual, please let us know so that we can consider your comments for future reprints.
We also provide special customer solutions and will be pleased to advise you in using PIOX R400 for specific appli­cations and finding the most appropriate solution for your measurement problem.
The content of this manual may be changed without prior notice. All rights reserved. No part of this manual may be reproduced in any form without FLEXIM's written permission.
1.2 Safety Instructions
In this user manual, important safety instructions are given. They are marked as follows:
Always observe the safety instructions!
1.3 Suitability and Warranty
We take no responsibility whatsoever concerning the suitability of the PIOX R400 for particular measuring tasks. It is the duty of the user to ensure that the refractometer is used only for the purpose for which it has been designed ("correct use"). The correct use of the instrument includes the respect of the mounting, commissioning and mainte­nance instructions.
The material and workmanship of PIOX R400 is guaranteed for the term specified in the sales contract provided the equipment has been used for the purpose for which it has been designed and operated according to the instructions given in this user manual.
Misuse of PIOX R400 will immediately revoke any warranty given or implied. This includes:
replacement of a component of the instrument by a component that was not authorized by FLEXIM
unsuitable or insufficient maintenance
repair of the instrument by unauthorized personnel
Attention!
Always observe the "Safety Instructions according to Directive 94/9/EC" (see document SIPIOX_R400).
Note!
The notes contain important information which help you use PIOX R400 optimally.
Attention!
This text contains important instructions which should be observed to avoid damage or destruction of PIOX R400.
This text denotes instructions according to directive 94/9/EC.
Attention!
The user manual and any complementary instructions you have received with the instrument have to be read before mounting, commissioning, operating or servicing the instrument.
Attention!
Improper installation and operation of the instrument can lead to the revocation of the guaran­tee.
Page 6
6 UMPIOXR400V2-7-1EN, 2012-07-01
1 Introduction
The manufacturer assumes no responsibility for damages resulting from incorrect use of the instrument.
FLEXIM assumes no responsibility for injury to the customer or third persons proximately caused by the material owing to defects in the product which were not predictable or for any indirect damages.
PIOX R400 is a very reliable instrument. It is manufactured under strict quality control, using modern production techniques. If installed as recommended in an appropriate location, used cautiously and taken care of conscien­tiously, no troubles should appear.
If any problem appears which can not be solved with the help of this user manual, contact our sales office giving a precise description of the problem. Specify the type, serial number and firmware version of the instrument (see section 4.2).
Page 7
UMPIOXR400V2-7-1EN, 2012-07-01 7
2 Measuring Principle of PIOX R400
2 Measuring Principle of PIOX R400
The refractive index n of a solution will be determined by means of transmitted light refractometry. A light beam propagates through the solution and will be refracted at the interface to a prism. The angle of refraction will be measured by a detector. The refractive index n of the solution can be calculated from Snell’s law of refraction:
n
i
. sin i= nt . sin
t

with n
i
refractive index of fluid
i
angle of incidence
n
t
refractive index of prism
t
angle of refraction
Fig. 2.1: Path of light beam through the sensor
The light source is a special LED with a wave length = 590 nm (sodium D-line) (see Fig. 2.1). The light passes an aperture, will be parallelized by a lens and reversed by a Dove prism. Then it enters the fluid through a window in the sensor head. When the light beam re-enters the sensor, it will be split at the vertex of a biprism and refracted at its inclined edges.
The two resulting measuring beams will be focussed by a lens projecting sharp images of the aperture on the detector, a CCD sensor with 2048 pixels.
The angle of refraction will be determined from the difference between the two measuring beam projections. The zero point will be calculated continuously, thus eliminating influences of process pressure and temperature.
Dove prism
pipe
biprism
fluid
window
temperature probe
LED
CCD image CCD sensor
sensor
Page 8
8 UMPIOXR400V2-7-1EN, 2012-07-01
2 Measuring Principle of PIOX R400
2.1 Determining the Concentration
The concentration can be calculated using the refractive index. In solutions, the refractive index varies with con­centration and temperature (see Fig. 2.2).
If this relationship is known, one-to-one throughout the working area and external effects can be excluded, the measured refractive index and temperature can be used to determine the concentration according to the character­istics (see Fig. 2.2).
The characteristics can be determined at the FLEXIM laboratory or directly in the process. Characteristics for vari­ous sugar solutions have been measured in accordance with ICUMSA
1
and have been stored in the internal data-
base.
2.2 Temperature Compensation
The refractive index varies with temperature (see Fig. 2.2). In industrial applications, it is often not necessary to determine the refractive index at the actual process temperature. Instead, an equivalent value, measured at a ref­erence temperature, is used. The measured refractive index will be converted by the transmitter accordingly.
The refractive index will be corrected automatically if temperature coefficients are stored for the medium. The refer­ence temperature is used to calculate the temperature coefficients. This is an integral part of the set of coefficients for the medium.
1
ICUMSA (International Commission for Uniform Methods of Sugar Analysis)
Note!
If only the refractive index and temperature are measured, the concentration can only be determined for two-component liquid systems.
1 0 2 0 3 0 4 0 5 0 6 0 7 0 8 0
1 , 3 5 0 0
1 , 4 0 0 0
1 , 4 5 0 0
1 , 5 0 0 0
T e m p e r a t u r :
2 0 ° C 5 0 ° C 8 0 ° C
K o n z e n t r a t i o n ( ° B r i x )
B r e c h u n g s i n d e x n
refractive index n
temperature [°C]
concentration [°Brix]
20
50
80
Fig. 2.2: Relationship between refractive index, concentration and temperature
in a saccharose solution
Page 9
UMPIOXR400V2-7-1EN, 2012-07-01 9
3 Description of PIOX R400
3 Description of PIOX R400
The design of PIOX R400 is shown in Fig. 3.1. Transmitter and sensor are the main components. A junction box can be mounted between transmitter and sensor.
sensor cable
extension cable
direct connection
connection via junction box
sensor cable
Fig. 3.1: Measurement setup for PIOX R400
Page 10
10 UMPIOXR400V2-7-1EN, 2012-07-01
3 Description of PIOX R400
3.1 Transmitter Design
The command panel and connections for the transmitter PIOX TR374 are shown in Fig. 3.2.
3.2 Sensor Design
For the components of sensor PIOX R400 see Fig. 3.3. The sensor is equipped with a PT1000 temperature probe. Window and prism are the optical components. LED and CCD are located in the sensor housing.
serial interface RS232
terminal strip KL3: connection of the power supply
terminal strip KL4: connection of the outputs
terminal strip KL1: connection of the sensor
terminal strip KL2: connection of the inputs
2x16-digit LCD display
keyboard
Fig. 3.2: Design of transmitter PIOX TR374
window
prism
temperature probe
Attention!
Always handle the sensor carefully since it contains fragile optical components.
Fig. 3.3: Design of sensor PIOX R400
Page 11
UMPIOXR400V2-7-1EN, 2012-07-01 11
4 Handling
4 Handling
4.1 First Inspection
Upon delivery, check if all parts have been delivered. Please make sure that the specifications of the delivered sen­sor and transmitter correspond to the specifications given on the purchase order.
4.2 Type and Serial Number
Type and serial number are on the nameplates of the transmitter and of the sensor (see Fig. 4.1 and Fig. 4.2).
4.3 Firmware Version
Select SPECIAL FUNCTION\INSTRUM. INFORM. to obtain informa­tion about PIOX R400. Press ENTER.
Type and serial number of PIOX R400 will be displayed in the upper line. Firmware version and its data are given in the lower line. Press ENTER to return to the program branch SPECIAL FUNCTION.
R E F R A C T O M E T E R T R A N S M I T T E R
I I 3 G E x n A I I T 4 T
a :
- 4 0 . . . + 6 0 ° C
T y p e : P I O X®T R 3 7 4 A 2
F l e x i m G m b H D - 1 2 6 8 1 B e r l i n
S e r . N o :
0 7 4 3 0 1 7 0
B u i l t : 2 0 0 8
M a d e i n G e r m a n y
I P 6 5
Fig. 4.1: Nameplate of the transmitter PIOX TR374A2 (example)
P R O C E S S R E F R A C T O M E T E R
I I 3 G E x n A I I T 4
T
a :
- 4 0 . . . + 6 0 ° C
T y p e : P I O X
®
R 4 0 0 A 2
F l e x i m G m b H D - 1 2 6 8 1 B e r l i n
S e r . N o :
2 2 2 0 0 1 0 9
B u i l t : 2 0 0 8
M a d e i n G e r m a n y
T
m :
- 2 0 . . . + 1 3 0 ° C
I P 6 7 P N 1 6
M o d e l : M S 4 F K A 2 F
P o w e r :
1 2 V / 1 0 0 m A
D O N O T O P E N !
Fig. 4.2: Nameplate of the sensor PIOX R400A2 (example)
S p e c i a l F u n c t . I n s t r u m . I n f o r m .
P X T R 3 7 4 - 0 7 4 3 0 1 4 7 V 1 . 1 2 3 1 . 0 3 . 0 6
Page 12
12 UMPIOXR400V2-7-1EN, 2012-07-01
4 Handling
4.4 General Precautions
PIOX R400 is a precision measuring instrument and must be handled with care. To obtain good measurement results and not to damage the instrument, it is important to pay attention to the instructions given in this user man­ual, and particularly to the following points:
protect the instrument from excessive shock
manipulate the cables cautiously and avoid excessive cable bend
connect the transmitter correctly to the power supply (voltage, frequency, connection to ground)
4.5 Service
Regular maintenance is normally not required. Always observe the handling precautions and the instructions given in this manual. If the instrument is installed correctly, in an appropriate location and as recommended, used cau­tiously and taken care of conscientiously, no troubles should appear.
4.6 Cleaning
Generally, the regular process cleaning and rinsing cycles are sufficient to keep the optical surfaces (prism and window) clean. However, depending on the liquid to be measured, it might be necessary to clean the measuring prism regularly to prevent the formation of deposits.
When cleaning with a caustic soda lye or any other aggressive chemical, first make sure that all parts in contact with the medium are resistant. If a manual cleaning of the prism or of the window is necessary, this can be made with a small pipe cleaner or with a soft cloth.
4.7 Calibration and Correction
Transmitter and sensor were calibrated by the factory. If the measured values need to be corrected, see section 8.2.
Attention!
Always observe the "Safety Instructions according to Directive 94/9/EC" (see document SIPIOX_R400).
Attention!
The degree of protection IP 65 of the transmitter is only guaranteed if the cables fit firmly and tightly in the cable glands, the cable glands are firmly tightened and the housing has been tightly screwed.
Attention!
Always handle the sensor carefully since it contains fragile optical components.
Note!
Only clean the prism with non-abrasive material!
Page 13
UMPIOXR400V2-7-1EN, 2012-07-01 13
5 Selecting the Measuring Point
5 Selecting the Measuring Point
The sensor PIOX R400 is designed to be used directly in the process flow. It should be installed at a measuring point that is representative for the process and relevant to the given application.
If the measurement is integrated in a process control system, the sensor should be installed at a location where it can react quickly to process changes.
Fasten the sensor to a vertical pipe (rising pipe) that allows the fluid to flow upwards. The pipe is always completely full and the measurement prism is constantly surrounded by flowing fluid (see Fig. 5.1).
Fig. 5.1: Installation of the sensor with a welding adapter (full section view)
Attention!
Always observe the "Safety Instructions according to Directive 94/9/EC" (see document SIPIOX_R400).
Attention!
Make sure that the degree of protection of the transmitter, sensor and junction box is observed.
Attention!
If it is not possible to maintain the operational temperature transmitter, sensor and junction box have to be cooled.
Attention!
If the fluid temperature is > 100 °C and PIOX R400 is used in explosive atmosphere, the sen­sor must be shielded against heat radiation from the pipe. This ensures that the ambient, fluid and explosion protection temperatures remain within the specification of the sensor.
Attention!
Transmitter, sensor, junction box must not be thermally insulated (i.e., "not wrapped"). Other­wise, the permissible ambient temperature might be exceeded.
flow direction
sensor
Page 14
14 UMPIOXR400V2-7-1EN, 2012-07-01
5 Selecting the Measuring Point
Install the sensor head so that the direction of flow is parallel to the prism edge (see Fig. 5.2).
Fig. 5.2: Orientation of measuring prism in pipe
To prevent deposits from forming on the measuring prism, ensure the following conditions:
permanent minimum flow in the pipe
the pipe diameter is relatively small at measurement point, i. e., the smaller the pipe diameter, the larger the pressure and the larger the self-cleaning effect
the fluid temperature is relatively high but within the specifications of the sensor, i. e., if the temperature increases, the fluid solubility increases but the viscosity decreases
the fluid pressure is relatively high but within the specification of the sensor, i. e., as the pressure increases, the fluid solubility increases and fewer gas bubbles form
If necessary, clean the measurement prism manually. Therefore, choose a measuring point where the flow can be interrupted and the sensor can be installed/removed without difficulty.
The selected measuring point should be easily accessible for service work.
Make sure that the transmitter is installed within the range of the cable. The ambient temperature at the installation site must be within the operating temperature range of the transmitter.
Note!
Protect the sensor against direct sunlight, rain and vibrations (pumps).
Note!
If the fluid is highly viscous, abrasive or contains solids in suspension, a bypass is recom­mended.
Attention!
Never wrap the cables (extension and sensor cables) around hot pipes.
flow direction
Page 15
UMPIOXR400V2-7-1EN, 2012-07-01 15
6 Installation of PIOX R400, R400B, R400C, R400A2
6 Installation of PIOX R400, R400B, R400C, R400A2
6.1 Measuring Setups
Direct Connection
The sensor will be connected directly to the transmitter (cable length = 10 m).
Connection via Junction Box
If the distance of sensor and transmitter is > 10 m, the sensor will be connected via a junction box and an extension cable to the transmitter.
6.1.1 PIOX R400
Fig. 6.1: PIOX R400
6.1.2 PIOX R400B
Fig. 6.2: PIOX R400B
Attention!
Always observe the "Safety Instructions according to Directive 94/9/EC" (see document SIPIOX_R400).
P IO X
transmitter PIOX TR374 or PIOX TR379
sensor PIOX R400
junction box JB93
extension cable, max. 200 m
connection via junction box
OR
direct connection
10 m
10 m
fluid
outside of explosive
atmosphere
P IO X
10 m
10 m
outside of
explosive
atmosphere
zone 1
zone 0
transmitter PIOX TR374B
sensor PIOX R400B
junction box JB91
connection via junction box
OR
direct connection
fluid
extension cable, max. 100 m
Page 16
16 UMPIOXR400V2-7-1EN, 2012-07-01
6 Installation of PIOX R400, R400B, R400C, R400A2
6.1.3 PIOX R400C
Fig. 6.3: PIOX R400C
6.1.4 PIOX R400A2
Fig. 6.4: PIOX R400A2
P IO X
zone 1
10 m
10 m
junction box JB91
transmitter PIOX TR374C
sensor PIOX R400C
extension cable, max. 200 m
connection via junction box
OR
direct connection
outside of explosive
atmosphere
fluid
P IO X
10 m
10 m
transmitter PIOX TR374A2
zone 2
junction box JB92
sensor PIOX R400A2
extension cable, max. 200 m
connection via junction box
OR
direct connection
fluid
Page 17
UMPIOXR400V2-7-1EN, 2012-07-01 17
6 Installation of PIOX R400, R400B, R400C, R400A2
6.2 Installation
6.2.1 Installation of Transmitter PIOX TR374, TR374B, TR374C, TR374A2
Wall Installation
The transmitter can be fixed to the wall.
Remove the cover of the housing.
At the chosen location, drill 4 holes in the wall (see Fig. 6.5).
Insert dowels in the holes.
Fix the housing with screws to the wall.
Fig. 6.5: Assembly dimensions (in mm)
Pipe Installation
The transmitter can be fixed to the pipe. For the assembly, use the supplied pipe mounting kit consisting of a shackle (1), pipe plate (2) and instrument mounting plate (3) (see Fig. 6.6).
Fig. 6.6: Pipe mounting kit
2 5
3 5
ø 8
3 . 5
1 6 3
2 6 5
3 0
6
ø
ø
4 x
M 4
M o n t a g e m i t H o l z s c h r a u b e n u n d D ü b e l
M o n t a g e m i t L i n s e n k o p f s c h r a u b e n
2 8 0
2 0 0
163
265
200
280
2
3
1
Page 18
18 UMPIOXR400V2-7-1EN, 2012-07-01
6 Installation of PIOX R400, R400B, R400C, R400A2
6.2.2 Installation of Sensor PIOX R400, R400B, R400C, R400A2
The sensor PIOX R400 with FLEXIM flange is fastened to the process connection using a slit ring. The gasket (O­ring) which is in contact with the medium is placed correctly by a thrust ring (see Fig. 6.7).
Fig. 6.7: Installation of sensor PIOX R400 with FLEXIM flange
6.3 Connection of Sensor PIOX R400 with Transmitter PIOX TR374
Mount the sensor to the pipe at the selected measuring point (see chapter 5). The cable glands for connecting to the junction box and transmitter are included with the PIOX R400.
6.3.1 Direct Connection
Remove the blind plug for the connection of the sensor (see Fig. 6.9).
Open the cable gland of the sensor cable. The compression part remains in the cap nut (see Fig. 6.8).
Fig. 6.8: Cable gland
Attention!
The degree of protection of PIOX TR374 is only guaranteed if the cables fit firmly and tightly in the cable glands, the cable glands are firmly tightened and the housings have been tightly screwed.
Note!
It is recommended to run the cables from the measuring point to the transmitter before the connection of the sensor, to avoid load on the connectors.
B
2
1 slit ring
2 thrust ring
3 O-ring
3
B
B
2
1
3
cap nut
compression part
basic part
Page 19
UMPIOXR400V2-7-1EN, 2012-07-01 19
6 Installation of PIOX R400, R400B, R400C, R400A2
Push the sensor cable through the cap nut and the compression part.
Prepare the sensor cable with the cable gland (see Fig. 6.9).
Cut the outer shield and brush it back over the compression part.
Screw the gasket ring side of the basic part in the housing.
Insert the sensor cable in the housing.
Fix the cable gland by screwing the cap nut on the basic part.
Connect the leads to the terminals of the transmitter (see Fig. 6.9).
Fig. 6.9: Direct connection of sensor PIOX R400 with transmitter PIOX TR374
terminal sensor cable
1
measuring
channel A
V+ yellow
V- green
A+ brown
B- white
3a gray
5a pink
measuring
channel B
2V+ yellow
2V- green
2A+ brown
2B- white
6b gray
9b pink
70 mm
10 mm
outer shield
terminal strip KL1
terminal strip KL1:
sensor cable
Preparation of extension cable and sensor cable:
transmitter
1
No terminals are provided for
connecting the red and blue cables.
Important! To ensure good high frequency shielding, it is important to provide good electrical contact between the outer shield and the cable gland and hence to the housing of the transmitter. Cut the outer shield and brush it back as shown in the figure.
Attention! To connect the sensor cable to the transmitter, use only metallic cable glands.
outer shield
Page 20
20 UMPIOXR400V2-7-1EN, 2012-07-01
6 Installation of PIOX R400, R400B, R400C, R400A2
6.3.2 Connection via Junction Box
Connection of the Extension Cable with the Transmitter
Remove the blind plug for the connection of the sensor (see Fig. 6.10).
Open the cable gland of the extension cable. The compression part remains in the cap nut (see Fig. 6.8).
Push the extension cable through the cap nut and the compression part.
Prepare the extension cable with the cable gland (see Fig. 6.10).
Cut the outer shield and brush it back over the compression part.
Screw the gasket ring side of the basic part in the housing.
Insert the extension cable in the housing.
Fix the cable gland by screwing the cap nut on the basic part.
Connect the leads to the terminals of the transmitter (see Fig. 6.10).
Connection of the Extension Cable with the Junction Box
Remove the blind plug from the junction box (see Fig. 6.10).
Open the cable gland of the extension cable. The compression part remains in the cap nut (see Fig. 6.8).
Push the extension cable through the cap nut, the compression part and the basic part.
Insert the extension cable in the junction box.
Prepare the extension cable with the cable gland.
Cut the outer shield and brush it back.
Pull the extension cable back until the brushed back outer shield is below the shield terminal of the junction box (see Fig. 6.10).
Screw the gasket ring side of the basic part in the junction box.
Fix the cable gland by screwing the cap nut on the basic part.
Fix the extension cable and the outer shield to the shield terminal of the junction box (see Fig. 6.10).
Connect the leads to the terminals of the junction box (see Fig. 6.10).
Connection of the Sensor Cable with the Junction Box
Remove the blind plug from the junction box (see Fig. 6.10).
Open the cable gland of the sensor cable (see Fig. 6.8). The compression part remains in the cap nut.
Push the sensor cable through the cap nut, the compression part and the basic part.
Prepare the sensor cable with the cable gland.
Insert the sensor cable in the junction box.
Screw the gasket ring side of the basic part in the junction box.
Fix the cable gland by screwing the cap nut on the basic part.
Connect the leads to the terminals of the junction box (see Fig. 6.10).
Attention!
The outer shield of the extension cable must not have electrical contact to the junction box. The extension cable has to remain insulated completely up to the shield terminal.
Attention!
The outer shield of the sensor cable must not have electrical contact to the junction box.
Page 21
UMPIOXR400V2-7-1EN, 2012-07-01 21
6 Installation of PIOX R400, R400B, R400C, R400A2
terminal strip KL1:
KL2
KL1
KL4
extension cable
shield terminal
Extension cable and outer shield must be fixed to the shield terminal.
sensor cable
outer shield
terminal sensor cable
1
V+ yellow
V - green
A+ brown
B- white
terminal extension cable sensor
cable
1
V+ yellow -
V - green -
S - -
S - outer
shield
2
A+ brown -
B - white -
terminal strip KL1:
terminal strip KL1
Note:
Prepare the sensor cable and the extension cable as shown in Fig. 6.9.
terminal extension cable
measuring
channel A
V+ yellow
V - green
A+ brown
B - white
measuring
channel B
2V+ yellow
2V - green
2A+ brown
2B - white
transmitter
junction box
Important!
To ensure good high frequency shielding, it is important to pro­vide good electrical contact between the outer shield and the cable gland and hence to the housing of the transmitter. Cut the outer shield and brush it back as shown in the figure.
terminal strip KL2:
Attention! To connect the sensor cable and the extension cable to the transmitter, use only metal cable glands.
terminal sensor cable
1
1a gray
2a pink
terminal strip KL4:
2
The outer shield of the sensor cable must be connected to terminal
strip KL2.
1
No terminals are provided for connecting the red and blue cables.
Fig. 6.10: Connection of sensor PIOX R400 with transmitter PIOX TR374 via junction box
Page 22
22 UMPIOXR400V2-7-1EN, 2012-07-01
6 Installation of PIOX R400, R400B, R400C, R400A2
6.4 Connection of Sensor PIOX R400 with Transmitter PIOX TR379
Mount the sensor to the pipe at the selected measuring point (see chapter 5). The cable glands for connecting to the junction box and transmitter are included with the PIOX R400.
6.4.1 Direct Connection
Prepare the sensor cable (see Fig. 6.9) and connect the leads to the terminals of the transmitter (see Tab. 6.1).
6.4.2 Connection via Junction Box
If using a junction box, connect the cable leads of the extension cable to the terminals of the transmitter (see Tab. 6.1).
Connect the cable leads of the extension cable and of the senor cable to the terminals of the junction box (see Fig. 6.10).
6.5 Connection of Sensor PIOX R400B, R400C, R400A2 with Transmitter PIOX
TR374B, TR374C, TR374A2
Mount the sensor to the pipe at the selected measuring point (see chapter 5). The cable glands for connecting to the junction box and transmitter are included with the PIOX R400.
6.5.1 Direct Connection
Remove the blind plug for the connection of the sensor (see Fig. 6.12).
Open the cable gland of the sensor cable. The compression part remains in the cap nut (see Fig. 6.11).
Push the sensor cable through the cap nut and the compression part.
Note!
It is recommended to run the cables from the measuring point to the transmitter before the connection of the sensor, to avoid load on the connectors.
Note!
The terminals for connecting the sensor to the transmitter PIOX TR379 are located on termi­nal strip KL6.
Tab. 6.1: Connection of sensor PIOX R400 to transmitter PIOX TR379
measuring
channel A
terminal sensor cable
measuring
channel B
terminal sensor cable
V+ yellow 2V+ yellow
V- green 2V- green
A+ brown 2A+ brown
B- white 2B- white
Attention!
The degree of protection IP 65 of the transmitter is only guaranteed if the cables fit firmly and tightly in the cable glands, the cable glands are firmly tightened and the housing has been tightly screwed.
Note!
It is recommended to run the cables from the measuring point to the transmitter before the connection of the sensor, to avoid load on the connectors.
Attention!
Always observe the "Safety Instructions according to Directive 94/9/EC" (see document SIPIOX_R400).
Page 23
UMPIOXR400V2-7-1EN, 2012-07-01 23
6 Installation of PIOX R400, R400B, R400C, R400A2
Prepare the sensor cable with the cable gland.
Cut the outer shield and brush it back over the compression part.
Fig. 6.11: Cable gland
Screw the gasket ring side of the basic part in the housing.
Insert the sensor cable in the housing.
Fix the cable gland by screwing the cap nut on the basic part.
Connect the leads to the terminals of the transmitter (see Fig. 6.12).
Fig. 6.12: Direct connection of sensor PIOX R400B, R400C, R400A2 with transmitter PIOX TR374B, TR374C, TR374A2
cap nut
compression part
basic part
equipotential
bonding terminal
terminal strip KL1
70 mm
10 mm
outer shield
sensor
cable
terminal sensor cable
measuring
channel A
V+ yellow
V- green
A+ brown
B- white
measuring
channel B
2V+ yellow
2V- green
2A+ brown
2B- white
terminal strip KL1:
transmitter
Important!
To ensure good high frequency shielding, it is important to provide good electrical contact between the outer shield and the cable gland and hence to the housing of the transmitter. Cut the outer shield and brush it back as shown in the figure.
Attention! To connect the sensor cable to the transmitter, use only metallic cable glands.
outer
shield
Preparation of extension cable and sensor cable:
equipotential bonding terminal
Page 24
24 UMPIOXR400V2-7-1EN, 2012-07-01
6 Installation of PIOX R400, R400B, R400C, R400A2
6.5.2 Connection via Junction Box
Connection of the Extension Cable with the Transmitter
Remove the blind plug for the connection of the sensor (see Fig. 6.13).
Open the cable gland of the extension cable. The compression part remains in the cap nut (see Fig. 6.11).
Push the extension cable through the cap nut and the compression part.
Prepare the extension cable with the cable gland (see Fig. 6.13).
Cut the outer shield and brush it back over the compression part.
Screw the gasket ring side of the basic part in the housing.
Insert the extension cable in the housing.
Fix the cable gland by screwing the cap nut on the basic part.
Connect the leads to the terminals of the transmitter (see Fig. 6.13).
Connection of the Extension Cable with the Junction Box
Remove the blind plug from the junction box (see Fig. 6.13).
Open the cable gland of the extension cable. The compression part remains in the cap nut (see Fig. 6.11).
Push the extension cable through the cap nut, the compression part and the basic part.
Insert the extension cable in the junction box.
Prepare the extension cable with the cable gland (see Fig. 6.13).
Cut the outer shield and brush it back.
Pull the extension cable back until the brushed back outer shield is below the shield terminal of the junction box (see Fig. 6.13).
Screw the gasket ring side of the basic part in the junction box.
Fix the cable gland by screwing the cap nut on the basic part.
Fix the extension cable and the outer shield to the shield terminal of the junction box (see Fig. 6.10).
Connect the leads to the terminals of the junction box (see Fig. 6.13).
Connection of the Sensor Cable with the Junction Box
Remove the blind plug from the junction box (see Fig. 6.13).
Open the cable gland of the sensor cable. The compression part remains in the cap nut (see Fig. 6.11).
Push the sensor cable through the cap nut, the compression part and the basic part.
Prepare the sensor cable with the cable gland (see Fig. 6.13).
Insert the sensor cable in the junction box.
Screw the gasket ring side of the basic part in the junction box.
Fix the cable gland by screwing the cap nut on the basic part.
Connect the leads to the terminals of the junction box (see Fig. 6.13)
Attention!
Always observe the "Safety Instructions according to Directive 94/9/EC" (see document SIPIOX_R400).
Attention!
The outer shield of the extension cable must not have electrical contact to the junction box. The extension cable has to remain insulated completely up to the shield terminal.
Attention!
The outer shield of the sensor cable must not have electrical contact to the junction box.
Page 25
UMPIOXR400V2-7-1EN, 2012-07-01 25
6 Installation of PIOX R400, R400B, R400C, R400A2
Fig. 6.13: Connection of the sensor PIOX R400B, R400C, R400A2 to the transmitter
PIOX TR374B, TR374C, TR374A2 via junction box
KL1KL2
shield terminal
terminal strip KL1
terminal sensor cable
V+ yellow
V- green
A+ brown
B- white
terminal strip KL1:
terminal strip KL2:
transmitter
junction box
extension
cable
sensor cable
Note:
Prepare the sensor cable and the extension cable as shown in Fig. 6.12.
terminal extension cable
measuring
channel A
V+ yellow
V- green
A+ brown
B- white
measuring
channel B
2V+ yellow
2V- green
2A+ brown
2B- white
terminal strip KL1:
terminal extension cable sensor cable
V+ yellow -
V- green -
S - -
S - outer
shield
1
A+ brown -
B- white -
1
The outer shield of the sensor cable must be
connected to terminal strip KL2.
Important! To ensure good high frequency shielding, it is important to provide good electrical contact between the outer shield and the cable gland and hence to the housing of the transmitter. Cut the outer shield and brush it back as shown in the figure.
Attention! To connect the sensor cable and the extension cable to the transmitter, use only metal cable glands.
outer
shield
Extension cable and outer shield must be fixed to the shield terminal.
equipotential bonding terminal
equipotential bonding terminal
equipotential bonding terminal
Page 26
26 UMPIOXR400V2-7-1EN, 2012-07-01
6 Installation of PIOX R400, R400B, R400C, R400A2
6.6 Connection of the Power Supply
6.6.1 Transmitter PIOX TR374, TR374B, TR374C, TR374A2
Prepare the power cable and connect the leads to the terminals of the transmitter (see Fig. 6.14, Tab. 6.2 and Tab. 6.3).
Fig. 6.14: Connection of the transmitter to the power supply
6.6.2 Transmitter PIOX TR379
Prepare the power cable and connect the leads to the terminals of the transmitter (see Tab. 6.4 and Tab. 6.5).
Attention!
A switch according to IEC 61010-1:2001 has to be provided in the building installation which must be near the instrument, easily accessible for the user and marked as disconnection device of the instrument.
If the instrument is used in explosive atmosphere, the switch should be installed outside the explosive atmosphere. If this is not possible, the switch should be installed in the area with the least risk.
Attention!
Always observe the "Safety Instructions according to Directive 94/9/EC" (see document SIPIOX_R400).
Tab. 6.2: AC connection of transmitter PIOX TR374, TR374B, TR374C, TR374A2
terminal connection
PE earth
N(-) neutral
L(+) phase 100...240 V AC, 50/60 Hz
Tab. 6.3: DC connection of transmitter PIOX TR374, TR374B, TR374C, TR374A2
terminal connection
PE earth
N(-) - DC
L(+) + DC
Tab. 6.4: AC connection of transmitter PIOX TR379
terminal connection
PE earth
N neutral
L1 phase 100...240 V AC, 50/60 Hz
Tab. 6.5: DC connection of transmitter PIOX TR379
terminal connection
PE earth
L- - DC
L+ + DC
F 1
K L 3
N ( - )
P E
L ( + )
Page 27
UMPIOXR400V2-7-1EN, 2012-07-01 27
6 Installation of PIOX R400, R400B, R400C, R400A2
6.7 Connection of the Outputs
6.7.1 Transmitter PIOX TR374, TR374B, TR374C, TR374A2
Prepare the output cable and connect the leads to the terminals of the transmitter (see Fig. 6.15).
Fig. 6.15: Terminals for the connection of the outputs of the transmitter
Terminals P1 to P7 are used for connecting the outputs of the transmitter (see Tab. 6.6).
Attention!
Always observe the "Safety Instructions according to Directive 94/9/EC" (see document SIPIOX_R400).
Tab. 6.6: Circuits of the outputs
output PIOX R400 terminal circuit
active current loop
Px+
Px-
R
ext
< 500
semi-passive current loop, use as active current loop
Px+
Px-
R
ext
< 50
semi-passive current loop, use as passive current loop
Px+
Px-
U
H
= 0...24 V
U
H
> 0.021A * R
ext
[]
passive current loop
Px+
Px-
U
H
= 4...24 V
U
H
> 0.021A * R
ext
[] + 4V
Example:
If U
H
= 12V,
R
ext
must be between
0...380 !
F 1
K L 3
P 4 -
P 3 -
P 6 b
P 5 b
1 0 1
P 1 -
1 0 3
P 2 -
P 4 +
P 3 +
P 6 a
P 5 a
P 7 a
A +
P 1 +
B -
P 2 +
P 7 b
+
-
mA
+-
+
-
mA
+-
+
-
mA
+-
+
-
U
H
mA
+-
+
-
U
H
Page 28
28 UMPIOXR400V2-7-1EN, 2012-07-01
6 Installation of PIOX R400, R400B, R400C, R400A2
R
ext
is the sum of all ohmic resistances in the circuit (resistance of the conductors, resistance of the amperemeter/voltmeter, etc.)
6.7.2 Transmitter PIOX TR379
Terminals P1 to P7 are used for connecting the outputs of the transmitter (see Tab. 6.6).
voltage output
Px+
Px-
R
ext
> 2 M
Note: For smaller values of R
ext
, the precision is
smaller than the specified value.
binary output optorelay
Px+
Px-
U
H
< 32 V
I
C
100 mA
binary output open collector
Px+
Px-
U
H
= 5...24 V
R
C
[k] = UH / IC [mA]
I
C
= 1...4 mA
binary output Reed relay
Px+
Pxa
Pxb
Px-
U
MAX
= 48 V
I
MAX
= 0.25 A
Note!
The terminals required for connecting the outputs are not defined until the outputs are config­ured (see section 9.1).
Note!
The terminals for the outputs of transmitter PIOX TR379 are located on terminal strip KL1.
Tab. 6.6: Circuits of the outputs
output PIOX R400 terminal circuit
+
Ri=500
Ri=500
V
+
-
+
-
U
H
V
+
-
R
C
+
-
U
H
V
+
-
R
C
a
b
Page 29
UMPIOXR400V2-7-1EN, 2012-07-01 29
6 Installation of PIOX R400, R400B, R400C, R400A2
6.8 Connection of the Serial Interface
6.8.1 Transmitter PIOX TR374, TR374B, TR374C, TR374A2
RS232 interface
The RS232 interface is located on the front plate of transmitter (see Fig. 6.16). The RS232 interface can be connected only outside of an explosive atmosphere as the housing has to be opened.
RS485 interface (optional)
Connect the RS485 cable to the terminals A+ and B- and the shield to terminal 101 of terminal strip KL4 (see Fig. 6.16).
Fig. 6.16: Connections of the serial interfaces of the transmitter
6.8.2 Transmitter PIOX TR379
RS485 interface (optional)
Connect the RS485 cable to the terminals 4A+ and 4B- of terminal strip KL 4 of transmitter PIOX TR379.
Attention!
Always observe the "Safety Instructions according to Directive 94/9/EC" (see document SIPIOX_R400).
Note!
If the transmitter has an RS485 interface, no data can be sent via RS232.
F 1
K L 3
1 0 1
A +
B -
connection of RS232 interface
connection of RS485 interface:
Page 30
30 UMPIOXR400V2-7-1EN, 2012-07-01
7 Start-Up
7 Start-Up
The command panel of PIOX R400 consists of the keyboard and a 2 x 16-digit LCD display.
7.1 Keyboard
The keyboard has 3 function keys and 12 keys for numerical data input. Several keys have double functions. They can be used for data entry and for navigation through the scroll list.
In selection mode, for example, the arrow-shaped numerical keys are used as cursor keys. In input mode, they are used for the input of numbers and characters.
The key functions are explained in Tab. 7.1 to Tab. 7.4.
PIOX R400 will be switched on by connecting the power supply. The main menu is displayed after the initialization.
Note!
No data can be entered while the serial number is displayed.
Tab. 7.1: General functions
INIT (cold start): Most parameters and settings are reset to the factory default values. The data memory will not be cleared.
Press these two keys simultaneously while switching on the instrument until the main menu is displayed.
During the measurement a cold start can be performed:
Press the keys BRK, C and ENTER simultaneously.
Release only key ENTER. The transmitter will be restarted.
Keep the keys BRK and C pressed until the main menu is displayed.
RESET: Press these three keys simultaneously to recover from an error. The reset has the same effect as restarting the transmitter. Stored data will not be affected.
> P A R < m e a o p t s f P a r a m e t e r
2x16-digit LCD display
8
2
0
5
7
N E X T
1
M U X
9
D I S P
3
D I S P
L F
E N T E R
L I G H T
R E S E T
I N I T
B R K
3 x O F F
C
O N
6
4
keyboard
Fig. 7.1: Command panel of transmitter PIOX TR374
I N I T
B R K
C
E N T E R
R E S T A R T
I N I T
B R K
C
Page 31
UMPIOXR400V2-7-1EN, 2012-07-01 31
7 Start-Up
7.2 Displays
7.2.1 Main Menu
For an overview of the menu structure see section 7.2.3.
Attention!
Be careful not to interrupt a current measurement by inadvertently pressing key BRK!
Tab. 7.2: Navigation
selection of the main menu
scroll to the left or right through a scroll list
scroll upwards and downwards through a scroll list
confirmation of the selected menu item
Tab. 7.3: Iinput of numbers
...
input of the numerical value shown on the key
sign for the input of negative values
decimal point
delete values.
After the value has been deleted, the previous value will be displayed.
confirmation of input
Tab. 7.4: Input of text
selection of the position of the character to be input
changes the currently selected character to an "A"
changes the currently selected character to a "Z"
changes between small and capital letters
selection of the next/previous ASCII character
deletes the character currently shown and inserts a blank space
...
scrolls upwards/downwards through the restricted ASCII character set. The char­acter changes every second. The scrolling can be interrupted by pressing any other key.
finishes editing
After switching on and initialization, the main menu will be displayed in the upper line. The main menu has the program branches:
• PAR (PARAMETER)
• MEA (MEASURING)
• OPT (OUTPUT OPTIONS)
• SF (SPECIAL FUNCTION)
The actually selected program branch is displayed in capital letters between arrows. The full name of the selected program branch is dis­played in the lower line.
Use the arrow keys and to select a program branch. Press ENTER.
B R K
6
Q +
Q
-
4
8
3
O N
2
3
O F F
E N T E R
0
9
D I S P
L F
L I G H T
+
E N T E R
6
Q +
Q
-
4
9
D I S P
3
D I S P
5
8
3
O N
2
3
O F F
0
7
N E X T
1
M U X
E N T E R
> P A R < m e a o p t s f P a r a m e t e r
4
6
Page 32
32 UMPIOXR400V2-7-1EN, 2012-07-01
7 Start-Up
7.2.2 Navigation
7.2.3 Menu Structure
If a vertical arrow () is displayed beside a menu item, this menu item contains a scroll list. This list is displayed in the lower line.
Use the arrow keys and , to scroll through the list. Press ENTER.
Sometimes a horizontal selection is requested in the lower line. The selected list item is displayed in capital letters between arrows.
Use the arrow keys and , to scroll through the list. Press ENTER.
Note!
Press key BRK to return to the main menu.
Note!
In this user manual, all program entries and keys are indicated in capital letters. Program entries are indicated in typewriter characters (PARAMETER). Submenus are separated from the main menu entry by a backslash.
P a r a m e t e r f o r C h a n n e l A :
8
3
O N
2
3
O F F
I 1 e n a b l e n o > Y E S <
4
6
>PAR<
>MEA<
>OPT<
>SF<
Parameter
Measuring
Output Options
Special Function
Instrument information
Select measuring
channel
Select measuring
channel
Select measuring
channel
Select fluid
Enter measuring point
Enter damping
Select physical
quantity
Configure serial output
Activate inputs
Select unit of
measurement
Activate program
code
Offset of
physical quantity
Activate serial
output
Activate service
mode
Offset of
sensor value
Activate alarm
System settings
Substitute values of
input values
Scale outputs
will be displayed
only if serial output is
activated
activate
deactivate
Start Measurement
will be displayed only if inputs are
activated
storage rate
output values
will be displayed
during measurement
after pressing
ENTER
serial number
firmware version
time, date
dialogs and menus
input, measuring, outputs
serial transmission
miscellaneous
Page 33
UMPIOXR400V2-7-1EN, 2012-07-01 33
8 Configuration
8 Configuration
The measurement has now to be configured, i.e. the physical quantity, the unit of measurement and the offsets have to be defined. The configuration will be made separately for every available measuring channel. It can be modified at any time later by selecting the program branch PARAMETER again.
8.1 Selection of the Physical Quantity and of the Unit of Measurement
The following physical quantities can be measured with PIOX R400:
refractive index
normalized refractive index (a corrected refractive index based on a reference temperature)
concentration in mass percent
concentration in volume percent
other customer specific physical quantities
The selected physical quantity is displayed by PIOX R400 during measurement. The normalized refractive index can be displayed simultaneously.
Note!
It is recommended to record the settings and measured values during the start-up of the PIOX R400 (see annex B "Start-Up and Maintenance Protocols").
In the main menu, select the program branch PARAMETER. Press ENTER.
Select the fluid to be measured in the scroll list. The scroll list contains glucose, saccharose, invert sugar, fructose, etc.
These fluids and their data sets are programmed in the internal data­base.
Three additional fluids each with two target values can be defined by means of the program RMKoeff and added to the scroll list (see chap­ter 13).
If a fluid is to be measured which is not in the fluid scroll list, select OTHER MEDIUM.
Note: Only Brix scales will be displayed as coefficients are not avail­able for this fluid.
Press ENTER.
If data sets for two target values have been defined for a fluid, the parameters of the first target value (main phys­ical quantity) will be requested first and then the parameters of the second target value (auxiliary physical quan­tity).
Note!
These displays is indicated only, if the menu item 1xMain+1xAUX has been selected in SPE­CIAL FUNCTION\SYSTEM SETTINGS\MEASURING\CONFIGURATION.
Note!
The correction of the refractive index takes place automatically if temperature coefficients are stored for the selected fluid.
> P A R < m e a o p t s f P a r a m e t e r
F l u i d G l u c o s e
A : M a i n m e a s u r e P l e a s e s e t u p . . .
A : A u x i l a r y m e a s . P l e a s e s e t u p . . .
Page 34
34 UMPIOXR400V2-7-1EN, 2012-07-01
8 Configuration
8.2 Input of Offsets
Corrections can be done as follows:
input of offsets for the values calculated by the transmitter (e.g. concentration in mass percent)
input of offsets for the measured primary physical quantities (e.g. refractive index, temperature)
8.2.1 Input of Offsets for the Values Calculated by the Transmitter
It is possible to correct the values calculated by the transmitter, i.e. by addition of a constant value to the selected physical quantity or the normalized refractive index. The corrections can be defined in the following displays:
.
The physical quantities and scales available for the selected fluid are displayed. Possible physical quantities are e.g. volume concentration or mass concentration. A typical scale is for example the Brix value for the concentration of a sugar solution.
Select the physical quantity. Press ENTER.
If several units of measurement are available for the selected physical quantity, they are now displayed.
The display will not be indicated if a scale is selected.
Select the unit of measurement to be used. Press ENTER.
If no offset is to be defined for the selected physical quantity or for the normalized refractive index, select NO. Otherwise select YES. Press ENTER.
The selected physical quantity is displayed in the upper line. Select YES to define an offset for this quantity. Press ENTER.
Enter the offset. The previously selected unit of measurement is dis­played. The maximum possible offset is 10 % of the valid value range of the selected physical quantity.
Example: The glucose concentration is measured with the Brix scale selected as unit of measurement. The
Brix scale ranges from 0...100 °. The maximum offset is 10 °. Press ENTER.
If no offset for the normalized refractive index is to be defined, select NO. Otherwise select YES. Press ENTER.
Enter the offset for the normalized refractive index. The offset has to be within the range ± 0.100000. Press ENTER.
M e a s u r i n g v a l u e M a s s - C o n c e n t r .
U n i t O f M e a s u r e % m a s
M e a s u r e a d j u s t s ? n o > Y e s <
M a s s - C o n c e n t r . n o > Y e s <
O F S : M a s s - C o n c e n t 0 . 7 5 % m a s
n o r m a l i z e d n D n o > Y E S <
O F S : n o r m a l i z e d n 0 . 0 0 1 2 3 4
Page 35
UMPIOXR400V2-7-1EN, 2012-07-01 35
8 Configuration
8.2.2 Input of Offsets for the Measured Primary Physical Quantities
It is possible to correct the primary physical quantities (refractive index and temperature) measured by PIOX R400. These values will be corrected before they are used by the transmitter for the calculation of the selected physical quantities. The correction can be defined in the following displays:
8.3 Additional Input Values
If no supplementary values are needed for the measurement and if no auxiliary physical quantity needs to be defined, the configuration of the measurement is finished. Proceed with the configuration of the outputs (see chap­ter 9).
Supplementary input values are quantities that are measured by another instrument and transmitted to PIOX R400 via the inputs. Approximated values for these quantities have to be entered in the case that no input values are available during the measurement (e.g. failure of the corresponding instrument, defective cable).
Supplementary input values can only be taken into account when the corresponding coefficients for the selected fluid are available.
Supplementary input values of the fluid are: temperature, pressure, density, kinematic viscosity, dynamic viscosity, conductibility, pH value, sound velocity, volume flow.
The inputs are described in chapter 14.
If the sensor values are not to be corrected, select NO, press ENTER and continue with section 8.3. Otherwise select YES. Press ENTER.
Select YES if a correction for the measured refractive index is to be defined. Press ENTER.
The offset has to be within the range ± 0.0500. Enter the offset. Press ENTER.
Select YES if an offset has to defined for the measured temperature. Press ENTER.
The offset has to be within the range ± 10.00 °C. Enter the offset. Press ENTER.
S e n s o r a d j u s t s ? n o > Y E S <
R e f r a c t i v e i n d e x n o > Y E S <
R e f r a c t i v e i n d e x O f f s e t : 0 . 0 0 4 3 2 1
F l u i d t e m p . n o > Y E S <
F l u i d t e m p . O f f s e t : 0 . 5 6 C
Page 36
36 UMPIOXR400V2-7-1EN, 2012-07-01
9 Outputs
9 Outputs
If the measured values are to be transmitted to another instrument, the outputs have to be configured.
The outputs have to be installed in the program branch SPECIAL FUNCTION (see section 9.1) and to be activated in the program branch OUTPUT OPTIONS (see chapter 11). The installation of an output consists of three steps:
assigning a measuring channel (source channel) to the output
defining the physical quantity (source item) to be transmitted to the output by the source channel and the prop­erties of the signal
defining the behavior of the output in case no valid measured values are available
9.1 Installation of an Output
Note!
The settings will be stored at the end of the installation dialog. If the installation dialog is inter­rupted by key BRK, the setting will not be stored.
Select SPECIAL FUNCTION\SYSTEM SETTINGS\PROC. OUTPUTS. Press ENTER.
Select the output to be installed. The scroll list contains all actually available outputs.
A tick after a list item indicates that this output has already been installed. Press ENTER.
This display is indicated if the output has not been installed yet. Select YES. Press ENTER.
If the output is already installed, select NO to reconfigure it or YES to return to the previous menu item and to select another output. Press ENTER.
Select in the scroll list the measuring channel to be assigned as source channel to the output. Press ENTER.
This display will not be indicated if the transmitter has only one mea­suring channel.
Select the source item type to be transmitted to the output. The differ­ent source item types and source items are given in Tab. 9.1. Press ENTER.
S y s t e m s e t t i n g s P r o c . o u t p u t s
I n s t a l l O u t p u t C u r r e n t I 1
I 1 e n a b l e n o > Y E S <
I 1 d i s a b l e n o > Y E S <
S o u r c e c h a n . C h a n n e l A :
S o u r c e i t e m M e a s u r e d v a l u e s
Page 37
UMPIOXR400V2-7-1EN, 2012-07-01 37
9 Outputs
Tab. 9.1: Available source items for the outputs
source item type possible source items description
MEASURED VALUES MAIN MEASURE
main physical quantity selected in program branch PARAMETER
AUXILARY VALUE
auxiliary physical quantity
NORMALIZED nD
normalized refractive index (nDT)
SENSOR VALUES REFRACTIVE INDEX
measured refractive index (not normalized, nD)
FLUID TEMPERAT.
fluid temperature measured by the sensor
SENSOR TEMPERAT.
average inner temperature of the sensor
HUMIDITY
sensor humidity
SIGNAL AMPLITUDE
relative signal amplitude
SIGNAL QUALITY
relative signal quality
PIXEL POSITION
absolute pixel position
ZERO POSITION
zero point of the pixels
LINKED INPUTS FLUID TEMP.
fluid temperature
FLUID PRESSURE
fluid pressure
DENSITY
density
KINEM.VISCOSITY
kinematic viscosity
DYNAM.VISCOSITY
dynamic viscosity
CONDUCTANCE
conductance
pH VALUE
pH value
SOUND SPEED
sound velocity
VOLUME FLOW
volume flow
COMMONINP x
with x = 1...4, any output
STATUSBITS STATUSBITS R1
first status bit of the channel
STATUSBITS R2
second status bit of the channel
STATUSBITS R3
third status bit of the channel
STATUSBITS R4
fourth status bit of the channel
Select the source item and press ENTER. If e.g. an alarm output has to be installed, select STATUSBITS in the scroll list (see section 11.4).
S o u r c e i t e m
S t a t u s B i t s
Page 38
38 UMPIOXR400V2-7-1EN, 2012-07-01
9 Outputs
9.2 Output Range
9.3 Error Output
Now the value can be defined which is to be output if the source item can not be measured.
When configuring an analog output, the output range will be defined now. Select one of the ranges in the scroll list or OTHER RANGE... to enter the output range manually.
If OTHER RANGE.. is selected, enter the minimum output value (OUTPUT MIN). Press ENTER.
If OTHER RANGE.. is selected, enter the maximum output value (OUTPUT MAX). Press ENTER.
The entered output range should cover at least 10 % of the full physi­cal output range (e.g.I
MAX
- I
MIN
2 mA for a 20 mA current loop).
If this is not the case, the smallest maximum output value (OUTPUT MAX) possible for the entered minimal output value (OUTPUT MIN) will be displayed.
Tab. 9.2: Error value options
error value option result
MINIMUM
output of the lower limit of the output range
HOLD LAST VALUE
output of the last measured value
MAXIMUM
output of the upper limit of the output range
OTHER VALUE
The output has to be entered manually. It has to be within the limits of the output.
Example: The normalized refractive index was selected as source item for the current loop, the output range
was set to 4/20 mA, the error value delay t
d
> 0.
The measurement of the refractive index is impossible during the time interval t
0
...t1. The error
value will be output.
I 1
O u t p u t r a n g e
4 / 2 0 m A
I 1 O u t p u t M I N 1 0 m A
I 1 O u t p u t M A X 1 1 m A
I 1 O u t p u t M A X 1 2 m A M I N I M A L
???
v [m3/h]
t
t
0
t
1
Fig. 9.1: Error output
???
nD
Page 39
UMPIOXR400V2-7-1EN, 2012-07-01 39
9 Outputs
Select a list item for the error output in the scroll list. Press ENTER.
Tab. 9.3: Examples for the error output
list items for the error output output signal
If OTHER VALUE ... is selected, enter an error value. The value must be within the physical limits of the output. Press ENTER.
The terminals to be used for the connection of the output are now dis­played (here: P1+ and P1- for the active current loop).
Connect the leads of the output cable to the displayed terminals.
Press ENTER.
E r r o r - v a l u e M i n i m u m ( 4 . 0 0 m A )
I [mA]
20
4
t
t
d
E r r o r - v a l u e H o l d l a s t v a l u e
I [mA]
20
4
t
E r r o r - v a l u e M a x i m u m ( 2 0 . 0 m A )
I [mA]
20
4
t
t
d
E r r o r - v a l u e O t h e r v a l u e . . .
I [mA]
20
4
t
t
d
I 1 = a c t i v l o o p T e r m i n a l : P 1 + , P 1 -
I 1 = a c t i v l o o p T e r m i n a l : P 1 + , P 1 -
Page 40
40 UMPIOXR400V2-7-1EN, 2012-07-01
9 Outputs
9.4 Function Check
The function of the installed output can now be tested. Connect a multimeter to the installed output.
Test of an analog output:
Test of a binary output:
The outputs are now configured. For the use of the serial output see chapter 10. For the activation of the outputs see chapter 11.
Enter a test value (in the example, the current output is tested). It has to be within the selected output range. Press ENTER.
If the multimeter displays the entered value, the output works cor­rectly.
Select YES to repeat the test, NO to return to SYSTEM SETTINGS. Press ENTER.
Select REED-RELAIS OFF or OPEN COLLECT.OFF in the scroll list OUTPUT TEST to test the de-energized state of the output. Press ENTER.
Measure the resistance at the output. The value has to be high ohmic.
Select YES. Press ENTER.
Select REED-RELAIS ON or OPEN COLLECT.ON in the scroll list OUTPUT TEST to test the energized state of the output. Press ENTER.
Measure the resistance at the output. The value has to be low ohmic.
Select YES to repeat the test, NO to return to SYSTEM SETTINGS. Press ENTER.
I 1 : O u t p u t T e s t 4 m A
I 1 = 4 . 0 m A A g a i n ? n o > Y E S <
B 1 : O u t p u t T e s t R e e d - R e l a i s O F F
B 1 = O F F A g a i n ? n o > Y E S <
B 1 : O u t p u t T e s t R e e d - R e l a i s O N
B 1 = O N A g a i n ? n o > Y E S <
Page 41
UMPIOXR400V2-7-1EN, 2012-07-01 41
10 Serial Output
10 Serial Output
For each measuring channel PIOX R400 can output max. 15 different values in regular intervals via the interface. To enable the serial output the output rate and the physical quantities to be output have to be defined in SPECIAL
FUNCTION\SETUP SERIAL OUT and the serial output has to be activated in the program branch OUTPUT OPTIONS.
10.1 Definition of the Storage Rate and of the Physical Quantities to be Output
Select the program branch SPECIAL FUNCTION\SETUP SERIAL OUT. Press ENTER.
In the first display of the program branch, the output rate will be dis­played. The output rate is the frequency of the output of a data line. The rate selected here will be used for all measuring channels.
Select one of the output rates from the scroll list or EXTRA to define a storage rate. Press ENTER.
If EXTRA is selected, enter the storage rate. Press ENTER.
Select the channel for which the serial output has to be configured. Press ENTER.
This display will not be indicated if the transmitter has only one mea­suring channel.
The scroll list COLUMN CONTENT is displayed. It contains the actual configuration of the serial output for the previously selected channel. Each entry of the list defines a column of the output.
The first character of each entry of the list is the designation of the column. The 15 possible columns of the out­put are designated by the letters a to o.
The designation of the column is followed by the number of the column in the serial output. The 15 possible col­umns will not be always output by PIOX R400. Only the columns will be output to which a source item has been assigned.
Afterward, the type of source item contained in the column is given. The different source item types and source items are given in Tab. 10.1. If a column was not assigned a source item, the designation of the column is:
-- = >off<.
Example: The scroll list contains the following list items: A01=MAIN MEASURE, B02=LIMIT R1, all other
columns are empty. Two columns with the previously selected rate will be output. The first output contains the main measurement, the second one the status bit R1.
To change the content of a column, select this column in the scroll list. Press ENTER.
Select the source item type to be transmitted to the output.
S p e c i a l F u n c t . S e t u p s e r i a l o u t
S t o r a g e R a t e E v e r y s e c o n d
S t o r a g e R a t e 1 s
S e r i a l o u t p u t f o r C h a n n e l A :
C o l u m n c o n t e n t a 0 1 = M a i n M e a s u r e
C o l u m n c o n t e n t a 0 1 = M a i n M e a s u r e
S o u r c e I t e m M e a s u r e d v a l u e s
Page 42
42 UMPIOXR400V2-7-1EN, 2012-07-01
10 Serial Output
The different source item types are given in Tab. 10.1.
Tab. 10.1: Available source items for the serial output
source item type possible source items
MEASURED VALUES MAIN MEASURE
main physical quantity selected in program branch PARAMETER
AUXILARY VALUE
auxiliary physical quantity
NORMALIZED nD
normalized refractive index (nDT)
SENSOR VALUES REFRACTIVE INDEX
measured refractive index (not normalized, nD)
FLUID TEMPERAT.
fluid temperature measured by the sensor
SENSOR TEMPERAT.
average inner temperature of the sensor
HUMIDITY
sensor humidity
SIGNAL AMPLITUDE
relative signal amplitude
SIGNAL QUALITY
relative signal quality
PIXEL POSITION
absolute pixel position
ZERO POSITION
zero point of the pixels
LINKED INPUTS FLUID TEMP.
fluid temperature
FLUID PRESSURE
fluid pressure
DENSITY
density
KINEM. VISCOSITY
kinematic viscosity
DYNAM.VISCOSITY
dynamic viscosity
CONDUCTANCE
conductance
pH VALUE
pH value
SOUND SPEED
sound velocity
VOLUME FLOW
volume flow
COMMONINP x
with x=1...4, any output
STATUSBITS STATUSBITS R1
first status bit of the channel
STATUSBITS R2
second status bit of the channel
STATUSBITS R3
third status bit of the channel
STATUSBITS R4
fourth status bit of the channel
If the source item type MEASURED VALUES is selected, select the source item in the scroll list. Press ENTER.
If the source item type LINKED INPUTS is selected, select the source item in the scroll list. Press ENTER.
When the input value in the scroll list is preceded by a question mark, no input is configured for the measurement of this value. If an input value is selected preceded by a question mark, a warning message will be displayed.
If the source item type SENSOR VALUES is selected, select the source item in the scroll list. Press ENTER.
If the source item type STATUSBITS is selected, select one of the sta­tus bits of the channel. Press ENTER (see section 11.4).
S o u r c e I t e m M e a s u r e d v a l u e s
S o u r c e I t e m L i n k e d i n p u t s
S o u r c e I t e m S e n s o r v a l u e s
S o u r c e I t e m S t a t u s B i t s
Page 43
UMPIOXR400V2-7-1EN, 2012-07-01 43
10 Serial Output
10.2 Format of the Serial Output
At the beginning of the measurement, a header describing the format of the output table is sent. The measured val­ues are transmitted afterwards. In standard format, the columns are separated by a semicolon. Each line ends with a line feed. The first three lines of the header contain general information about the instrument and the measure­ment (see example below):
Configuration parameters output for each measuring channel in a data block are in the following lines.
The line \DATA will be transmitted next.
\DEVICE
serial number of the instrument and firmware version
\STORAGE DATE
date and time of the measurement
\STORAGE RATE
output rate (time between two data lines belonging to the same channel)
\A:-;CHANNEL_INFORMATION
information about the channel (designation of the measuring point, name of the fluid, selected physical quantity). Each line begins with the name of the channel (\A:, \B:)
\A:=;COLUMN DESCRIPTIONS
The content of each column sent for this channel is specified on one line:
- column number (COL_01, COL_02, ...)
- source item
- unit of measurement
- column title
- information
Example:
\DEVICE ;PXTR372-07200101(V1.92)
\STORAGE_DATE;12.08.2004 14:51:46
\STORAGE_RATE;3 sec
\A:-;CHANNEL_INFORMATION
\A:-;Measuring_point=;123-00-33
\A:-;Bank[0].Main.Fluid=;MyOwnSubstance;Scale:Brix
\A:-;COLUMN DESCRIPTION
\A:-;COL_01=;Mass-Concentr.;[% mas];(MME);
\A:-;COL_02=;StatusBit R1[bit];(R1S);Measured values/Main values
\A:-;COL_03=;Fluid temperat.;[C];(SFT);
\A:-;COL_04=;Fluid pressure[?];(LPR);UNLINKED!
\A:-;COL_05=;StatusBit R3[bit];(R3S);Sensor values/Refr.index
\A:-; (MME)[% mas]; (R1S)[bit]; (SFT)[C]; (LPR)[?]; (R3S)[bit];
\DATA
A:; 10.26; NOT; 22.13; ?E?; NOT;
A:; 10.26; NOT; 22.12; ?E?; NOT;
A:; 10.26; NOT; 22.12; ?E?; NOT;
A:; 10.27; TRIG; 22.11; ?E?; NOT;
Page 44
44 UMPIOXR400V2-7-1EN, 2012-07-01
10 Serial Output
10.3 Transmission Parameters
PIOX R400 sends CRLF-terminated ASCII. The max. length of the line is 255 characters. The transmission param­eters of the serial interface RS232 are listed in Tab. 10.3 and of RS485 in Tab. 10.4.
Tab. 10.2: Titles of the columns of the serial output
column title contains units of measurement
SRI sensor - refractive index (none)
SFT sensor - fluid temperature [°C] [°F]
SDT sensor - instrument temperature [°C] [°F]
SHL sensor - humidity 1 [%H]
SHR sensor - humidity 2 [%H]
SAM sensor - signal amplitude [%]
SQU sensor - signal quality [%]
SPP sensor - pixel position (none)
SGZ sensor - pixel zero point (none)
MEA measured value - main measurement (depend on the physical quantity)
NDT measured value - normalized refractive index (none)
MAU measured value - auxiliary value (depend on the physical quantity)
R1S StatusBit - R1 [bit]
R2S StatusBit - R2 [bit]
R3S StatusBit - R3 [bit]
R4S StatusBit - R4 [bit]
LTF input - fluid temperature [°C] [°F]
LPR input - fluid pressure [bar]
LDE input - fluid density [g/cm³]
LVI input - kinematic viscosity [cSt=mm²/s]
LVD input - dynamic viscosity [mPas]
LGS input - conductance [mS]
LPH input - ph value [pH]
LSS input - sound velocity [m/s]
LVF input - volume flow [m³/h]
LC1 input - general input 1 (depend on the connected input)
LC2 input - general input 2
LC3 input - general input 3
LC4 input - general input 4
Tab. 10.3: Transmission parameters of RS232 serial interface
baud rate 9600 bits per seconds
data bits 8
parity even
stop bits 2
protocol RTS/CTS
Tab. 10.4: Transmission parameters of RS485 serial interface
baud rate 9600 bits per seconds
data bits 8
parity even
stop bits 1
Page 45
UMPIOXR400V2-7-1EN, 2012-07-01 45
10 Serial Output
These settings can be changed in SPECIAL FUNCTION\SYSTEM SETTINGS\SERIAL TRANSMIS. if an RS485 interface is used.
10.4 Format Settings of the Serial Output
Some format settings of the serial output can be edited in SPECIAL FUNCTION\SYSTEM SETTINGS\
SERIAL TRANSMIS..
Press ENTER to confirm the menu item PROTOCOL FOR SERIAL OUTPUT.
Select DEFAULT to display the default transmission parameters.
Select SETUP to change the transmission parameters. Press ENTER.
The display contains three scroll lists:
BAUD for the baud rate
PARITY for the parity
ST for the number of stop bits
If ON is selected, space characters will not be transmitted. The file size will be considerably reduced (i.e. shorter transmission time).
Decimal marker to be used for floating point numbers (point or comma).
Character to be used for separating columns (semicolon or tabula­tor). The setting depends on the requirements of the PC program.
Note!
If the transmitter has an RS485 interface, no data can be sent via RS232.
P r o t o c o l f o r S e r i a l o u t p u t
S e r i a l p r o t o c o l d e f a u l t > S E T U P <
> B A U D < p a r i t y s t 2 4 0 0 E V E N 1
S E R : k i l l s p a c e s o f f > O N <
S E R : d e c i m a l p o i n t ' . ' > ' , ' <
S E R : c o l - s e p a r a t . ' ; ' > ' T A B ' <
Page 46
46 UMPIOXR400V2-7-1EN, 2012-07-01
11 Output Options
11 Output Options
If analog outputs or the serial output are configured, they have to be activated in program branch OUTPUT OPTIONS. The status outputs also will be configured in this program branch
.
In the program branch OUTPUT OPTIONS:
the damping factor of the measured values will be defined
the installed inputs and outputs will be activated
alarms will be defined and activated
the serial output of the measured values will be activated
11.1 Damping Factor
Each displayed measured value is the floating average of all measured values of the last x seconds, where x is the damping factor.
A damping factor of 1 s means that the measured values are not averaged as the measuring rate is approx. 1/s. The default value is 10 s. This is appropriate for normal measuring conditions.
Strongly fluctuating readings caused by high dynamics require a larger damping factor.
11.2 Activation of the Serial Output
11.3 Activation of the Outputs
The activation displays of all the configured outputs will now be displayed one after another. The configuration of the outputs is described in chapter 9.
Three values relating the measured values to the output values (current, voltage, frequency) have to be entered.
Select the program branch OUTPUT OPTIONS. Press ENTER.
Select the channel for which the output options are to be set.
This display will not be indicated if the transmitter has only one mea­suring channel.
Enter the damping factor. Values between 1 s and 100 s will be accepted. Press ENTER.
If the serial output has been configured in program branch SPECIAL FUNCTION, SERIAL OUTPUT will be displayed now. The installation of the serial output is described in chapter 10.
Select YES to activate the output of the measured values to the serial interface, NO to deactivate it. Press ENTER.
Select YES to activate the specified output,
NO to deactivate it. Press
ENTER.
p a r m e a > O P T < s f O u t p u t O p t i o n s
O u t p u t O p t i o n s f o r C h a n n e l A :
D a m p i n g 1 0 s
S e r i a l O u t p u t n o > Y E S <
C u r r e n t L o o p I 1 : n o > Y E S <
Page 47
UMPIOXR400V2-7-1EN, 2012-07-01 47
11 Output Options
In the menu item MEAS.VALUES will be requested whether the sign should be considered for the output of the measured values.
The LOW-SCALE VALUE and FULL-SCALE VALUE determine the range of measured values which will be mapped on the physical range of the output. For all values greater than the FULL-SCALE VALUE, the output value will be clipped to the maximum output value (OUTPUT MAX). For all values smaller than the LOW-SCALE VALUE, the min­imal value (OUTPUT MIN) will be output. The LOW-SCALE VALUE and FULL-SCALE VALUE can be negative or positive. The range of measured values does not need to be symmetrical. In most of the cases, the LOW-SCALE VALUE will be the lowest measured value expected, the FULL-SCALE VALUE the highest value expected.
After the program branch OUTPUT OPTIONS is completed, the main menu will be displayed again. Start the mea­surement as described in chapter 12.
11.4 Status Output (Alarms)
Each measuring channel of PIOX R400 has 4 status bits (R1...R4) which can used to define a status output (alarm output). First activate the status bits and then configure the status output.
In the following displays, only the status bits which have been linked to an output or to the serial output will be shown. If the needed status bit is not displayed, it has to be linked to an output (see chapter 9).
11.4.1 Selection of the Status Bit
If no status output is to be defined or no settings are to be modified, select SKIP SETUP in the scroll list STATUS OUTPUT. The program branch OUTPUT OPTIONS will be finished. Start the measurement as described in chapter
12.
All status outputs can be switched off at once by selecting the list item DISABLE ALL. The program branch OUT- PUT OPTIONS will be finished. Start the measurement as described in chapter 12.
11.4.2 Activation of the Status Bit
Select SIGN if the sign of the measured values should be considered for the output, ABSOLUT if it should not be considered.
Enter as LOW-SCALE VALUE the lowest measured value expected. The displayed unit of measurement is the unit of the source item of the output. The LOW-SCALE VALUE is the measured value corresponding to the lower limit of the physical range of the output.
Enter as FULL-SCALE VALUE the highest measured value expected. The FULL-SCALE VALUE is the measured value corresponding to the upper limit of the physical range of the output.
Select the status bit to be used for a status output. The designation of the status bit is followed by the output to which it is linked.
Select YES to activate the selected status bit. Select NO to deactivate the status bit and return to the previous menu item. Press ENTER.
M a s s - C o n c e n t r . > A B S O L U T < s i g n
I 1 : L o w - s c a l e 0 . 0 0 % m a s
I 1 : F u l l - s c a l e 1 0 . 0 0 % m a s
S t a t u s O u t p u t R 1 - > B 1
R 1 - > B 1 n o > Y E S <
Page 48
48 UMPIOXR400V2-7-1EN, 2012-07-01
11 Output Options
11.4.3 Selection of the Source Item
The source item is the quantity triggering the status bit, i.e the quantity about which information is to be sent to the status output.
Example: PIOX R400 will be used to control the Brix value of a sugar solution. The value must not exceed
30 °. Select the Brix value as source item for the status output. PIOX R400 will monitor the Brix value and raise an alarm when the defined limit is not respected. The alarm can be sent to the pro­cess control system.
Select the source item type in the scroll list. The different source item types and source items are given in Tab. 11.1. Press ENTER.
If the source item type MEASURED VALUES is selected, select the source item in the scroll list. Press ENTER.
Tab. 11.1: Available source items for the status outputs
source item type possible source items
MEASURED VALUES MAIN MEASURE
main physical quantity selected in program branch PARAMETER
AUXILARY VALUE
auxiliary physical quantity
NORMALIZED nD
normalized refractive index
SENSOR VALUES REFRACTIVE INDEX
measured refractive index (not normalized)
FLUID TEMPERAT.
fluid temperature measured by the sensor
SENSOR TEMPERAT.
average inner temperature of the sensor
HUMIDITY
sensor humidity
SIGNAL AMPLITUDE
relative signal amplitude
SIGNAL QUALITY
relative signal quality
PIXEL POSITION
absolute pixel position
ZERO POSITION
absolute zero point of the pixels
LINKED INPUTS FLUID TEMP.
fluid temperature
FLUID PRESSURE
fluid pressure
DENSITY
density
KINEM. VISCOSITY
kinematic viscosity
DYNAM.VISCOSITY
dynamic viscosity
CONCUCTANCE
conductance
pH VALUE
pH value
SOUND SPEED
sound velocity
VOLUME FLOW
volume flow
COMMONINP x
with x = 1...4, any output
If the source item type LINKED INPUTS is selected, select the source item in the scroll list. Press ENTER.
R 1 : S o u r c e I t e m M e a s u r e d v a l u e s
M e a s u r e d v a l u e s M a s s - C o n c e n t r .
L i n k e d i n p u t s ? F l u i d t e m p .
Page 49
UMPIOXR400V2-7-1EN, 2012-07-01 49
11 Output Options
11.4.4 Definition of the Triggering Condition
When the source item is selected, the triggering condition of the status output can be defined, i.e. the condition that has to be fulfilled to trigger the output. Three alarm functions are available. They are listed in Tab. 11.2.
The hysteresis is a symmetrical range around the limit in which fluctuation is permitted. The output will be triggered when the measuring value reach the upper limit of this range and when the measuring value fall below the lower limit.
11.4.5 Setting of the Triggering Period
It is possible to trigger the status output as soon as the defined condition is met. It is also possible to trigger the out­put only when the condition has been met without interruption for a certain predefined period of time.
When the input value in the scroll list is preceded by a question mark, no input is configured for the measurement of this value. If an input value is selected preceded by a question mark, a warning message will be displayed.
If the source item type SENSOR VALUES is selected, select the source item in the scroll list. Press ENTER.
Tab. 11.2: Alarm functions
alarm function description
MAXIMUM (x>trig)
The status output will trigger when the source item exceeds a certain maximum value.
MINIMUM (x<trig)
The status output will trigger when the source item falls below a certain minimum value.
SOURCE ERROR
The status output will trigger when the source item is not valid.
Example: In the example given above, the function needed can be reached by selecting the alarm type MAXI-
MUM (x>trig) for the status output. HIGH LIMIT of the output should be set to 30 °.
Select the alarm function. Press ENTER.
If HIGH LIMIT or LOW LIMIT is selected as alarm function, enter now the maximum or minimum value for the source item. Note that the signs of the source item and of the limit will be considered. Press ENTER.
It is now possible to define an hysteresis for the status output. This function is useful to avoid a constant triggering of the output by mea­sured values fluctuating around the limit.
Example: If a variation of 0.5 °Brix about the upper limit is allowed, set the hysteresis to 0.5 °Brix for the 30
°Brix limit. The alarm will trigger at 30.5 °Brix and at 29.5 °Brix. Small fluctuations around 30° Brix do not have any effect on the alarm.
Select
YES to activate the alarm immediately. Select NO to activate
the alarm only after the switching condition has been met for a certain period of time. Press ENTER.
W A R N I N G : U N L I N K E D ? F l u i d t e m p .
S e n s o r v a l u e s R e f r a c t i v e i n d e x
R 1 : f u n c t i o n M a x i m u m ( x > t r i g )
H i g h L i m i t 2 2 . 5 % m a s
H y s t e r e s e : 0 . 5 % m a s
I n s t a n t l y t r i g . ? n o > Y E S <
Page 50
50 UMPIOXR400V2-7-1EN, 2012-07-01
11 Output Options
11.4.6 Behavior of the Status Output after Triggering
11.4.7 Idle State of the Status Output
Depending on the physical type of the output, the idle state of the output will be as follows:
11.5 Activation of the Inputs
The installed inputs will be activated now one after another.
Enter now the period of time in which the switching condition must be met without interruption before the status output is activated. Press ENTER.
If the status bit has to return to its default state when the triggering condition is not met any more, select NON-HOLD. Press ENTER.
Otherwise select HOLD. The alarm remains in the activated state until being reset manually (see section 12.3). Press ENTER.
Select the behavior when the measurement of the source item is not possible any more after the status output is triggered.
Select HOLD OUTPUT to keep the status output activated.
Select DEFAULT OUTPUT to deactivate the status output immedi­ately.
Select DELAYED DEFAULT to deactivate the status output after a certain period of time.
Press ENTER.
If DELAYED DEFAULT is selected in the previous step, enter the delay for the deactivation. Press ENTER.
The idle state of the status output has to be defined.
•If NORMALLY OPEN is selected, the idle state of the status output will be “logical zero” (StatusBit=0).
•If NORMALLY CLOSED is selected, the default state of the status output will be “logical one” (StatusBit=1).
idle state state of a relay state of an optocoupler state of a current output
STATUSBIT=0
no current (contacts are open) no current (auxiliary voltage) current has lowest value
STATUSBIT=1
current (contacts are closed) current I < UHV/2 current has highest value
Note!
These settings are not active when the status bit is sent to the serial output. In this case, the message NOT will be sent to the serial output when the status bit is in default state. When the trigger condition is met, the message
TRIG will be sent.
The type and number of the input is displayed in the upper line (here: I1 for the first current input), followed by the physical quantity linked to this input (here: DENSITY).
Select
YES to activate the input, NO to deactivate it. Press ENTER.
M i n . t r i g . p e r i o d 0 s
A f t e r e v e n t N o n - h o l d
I f s o u r c e f a i l D e l a y e d d e f a u l t
R 1 : D e f a u l t a f t e r 0 s
D e f a u l t c o n t a c t n o r m a l l y o p e n
I 1 - > D e n s i t y n o > Y E S <
Page 51
UMPIOXR400V2-7-1EN, 2012-07-01 51
12 Measurement
12 Measurement
During the measurement, more information about the measurement can be obtained by scrolling.
Scrolling in the upper line (backward/forward) with the keys and .
Scrolling in the lower line (backward/forward) with the keys and .
Select in the main menu the program branch MEASURING to start the measurement. Press ENTER.
The channels on which is to be measured have to be activated, the other channels have to be deactivated.
This display will not be indicated if the transmitter has only one mea­suring channel.
measuring channel is activated
- measuring channel is deactivated
· measuring channel can not be activated
Select a measuring channel with key and .
Press key to activate or deactivate the selected channel.
A deactivated channel will be ignored during the measurement. All parameters entered for this channel remain unchanged.
Press ENTER when finished.
Note!
A measuring channel can not be activated if no parameters have been entered.
The measurement is started. The designation of the measured quan­tity is displayed in the upper line, its value in the lower line.
Press key to display the normalized refractive index
.
Tab. 12.1: Values that can be displayed in the upper line
display explanation explanation
Sens= ***.* status line of the sensor short status
Humid= 33% relative air humidity in the sensor chamber sensor humidity
Tdev= 25 C inner temperature of the sensor sensor temp.
Tflu= 22.34 C temperature of the fluid measured by the sensor fluid temp.
nD = 1.345678 refractive index measured by the sensor refractive index
Amp= 82.34564% signal amplitude
(values > 100% are possible) signal amplitude
Quality= 91.2% signal quality
(values >100% are possible) signal quality
Tab. 12.2: Values that can be displayed in the lower line
display explanation explanation
A: 3.44 Brix physical quantity (here: Brix) (designation of the physical quantity)
A: 1.345678 nDT refractive index
(with temperature correction)
norm.nD(T=xxC) The reference temperature defined in the temperature coefficients of the fluid and relative to which the normaliza­tion took place is given in parentheses.
p a r > M E A < o p t s f M e a s u r i n g
C H A N N : A B M e a s u r -
Ö
4
6
8
3
O N
S c a l e : B r i x A : 3 . 4 4 B r i x
n o r m . n D ( T = 2 0 C ) A : 1 . 3 4 5 6 7 8 n D T
3
DISP
8
3
O N
9
DISP
2
3
O F F
3
DISP
Page 52
52 UMPIOXR400V2-7-1EN, 2012-07-01
12 Measurement
12.1 Explanation of the Status Line
Each character in the status line indicated the state of a physical quantity. For the physical quantities covered by the status line and their order see Tab. 12.3.
The following characters are used in the status line: *, ?, H and L (see Tab. 12.4).
Limits have to be defined to indicate the status (see Tab. 12.5).
For an example see Tab. 12.6.
12.2 Service Mode
If key ENTER is pressed during the measurement, the service mode will be activated.
Press ENTER again to stop the service mode.
12.3 Reset of the Status Outputs
All alarms will be switched to their idle state by pressing key three times during measurement. Alarm outputs whose switching condition is still met will be activated again after 1 s. Use this function to reset alarm outputs of type HOLD when the switching conditions is not met anymore.
By pressing BRK measurement will be stopped and the main menu is selected. All alarms outputs will be de-ener­gized, independently of the programmed idle state. After a cold start, all alarm outputs will be re-initialized.
12.4 Protection against Interruption
The function SET PROGRAM CODE allows to input a program code that must be entered to interrupt a current mea­surement.
To interrupt a current measurement, the complete program code has to be entered (= BREAK code). To execute a command during measurement, it is sufficient to enter the first three digits of the program code (= ACCESS code).
Tab. 12.3: Order of characters in the status line
display Sens= * * * * * ...
physical quantity nD Tflu Tdev Humid
Tab. 12.4: Explanation of the characters in the status line
character explanation
* measurement is ok.
? no measured values
H (high) measured values above the upper limit
L (low) measured values below the lower limit
Tab. 12.5: Limits of the physical quantities in the status line
physical quantity nD Tflu Tdev Humid
limits 1.3...1.7 -20...+150°C -40...+60 °C 0...80 %
Tab. 12.6: Example of a status line
display Sens= * ? H * H ...
* The refractive index measured by the sensor is within the selected measuring range (ok).
? no value for the fluid temperature measured by the sensor
H the instrument is overheated (> 60 °C)
* -
H the relative air humidity in the sensor chamber is to high (> 80 %).
The set measuring and display functions will be kept during service mode (here: 60 minutes).
The last value measured before the activation will be output during service mode.
Note!
The display will be indicated only if a time > 0 has been entered for the service mode in SPE­CIAL FUNCTION\SYSTEM SETTINGS\MEASURING.
* SERVICE MODE *
* 60:00:00 *
C
Page 53
UMPIOXR400V2-7-1EN, 2012-07-01 53
12 Measurement
If a program code is active, the message PROGRAM CODE ACTIVE will be displayed when a key is pressed. The message will disappear after a few seconds.
12.4.1 Activation of a Program Code
12.4.2 Interruption of Measurement
If a program code was defined, it will be requested as soon as there is an intervention in the measurement (a com­mand or key BRK).
12.4.3 Deactivation of the Program Code
Note!
Do not forget the program code!
Select in the program branch SPECIAL FUNCTION the menu item SET PROGRAM CODE.
Select YES to enter a program code.
Enter a program code with max. 6 digits. Press ENTER.
An error message will be displayed if a reserved number has been entered (e.g. a HotCode).
A program code remains valid as long as no other valid program code is entered or the program code is not deactivated.
Enter the program code. Press ENTER.
The input of the program code will be stopped and the measurement continued by pressing key .
At first 000000 will be displayed. If the program code starts with 000, ENTER can be pressed directly if the
ACCESS code is requested.
If the entered program code is not valid, an error message will be dis­played for a few seconds. If the entered program code is valid, the command will be executed or the measurement will be interrupted.
The program code will be deleted by entering "------". Press ENTER.
Note!
If the character "-" is entered less than six times, this character string will be used as new pro­gram code.
S p e c i a l F u n c t . S e t p r o g r a m c o d e
P r o g r a m c o d e n o > Y E S <
P r o g r a m c o d e
- - - - - -
P r o g r a m c o d e I N V A L I D C O D E !
I N P U T B R E A K _ C O D E C O D E : 0 0 0 0 0 0
C
I N P U T B R E A K _ C O D E I N V A L I D C O D E
P r o g r a m c o d e
- - - - - -
Page 54
54 UMPIOXR400V2-7-1EN, 2012-07-01
13 Program RMKoeff
13 Program RMKoeff
The internal material database of the transmitter contains records for saccharose, fructose, glucose and invert sugar.
Data sets can be easily added for additional fluids.
The program RMKoeff is supplied on the delivered CD.
Insert the CD into your computer.
Start the program by double-clicking RMKoeff.exe in folder COEFFICIENT ENTRY.
The main window will be opened:
Fig. 13.1: Main window of program RMKoeff
Tab. 13.1 contains an overview of the various functions.
Tab. 13.1: Functions in window RMKoeff
menu/buttons explanation
menu DATEI open and save data records
menu PIOX receive and send data records
menu OPTIONEN configuration of the serial interface
button RECEIVE ALL COEFFICIENTS receive a data record from the transmitter
button DEFAULT ... example of data record; by changing of the default settings a new data
record can be created
button EDIT ... create a new data record
button SEND SELECTED COEFFICIENTS send a data record to the transmitter
Page 55
UMPIOXR400V2-7-1EN, 2012-07-01 55
13 Program RMKoeff
13.1 Create a New Data Record
A new data record can be created by changing of the default settings of an old data record. The new data record will be adapted to the data record of the fluid.
Click DEFAULT to create a data record for a new fluid.
In the window that will be opened, delete DEFAULT and enter the name of the fluid (see Fig. 13.2).
Fig. 13.2: Input of the fluid name
ENABLE ITEM... below ITEM 1 is activated. Click button EDIT COEFFS... below.
The window with the defaults will be opened (see Fig. 13.3). The definitions of the boxes are listed in Tab.
13.2. Change the values in the boxes
.
Note!
The name of the fluid will be displayed in the transmitter in
PARAMETER\MEDIUM.
Page 56
56 UMPIOXR400V2-7-1EN, 2012-07-01
13 Program RMKoeff
Fig. 13.3: Window with default settings
Tab. 13.2: Definition of boxes in EDIT COEFFS...
number explanation
(1) entry of refractive index of solvent for standard temperature T, e.g. for water at 20 °C: 1.33299
(2) temperature at which the compensated refractive index is output (standard temperature)
(3) temperature gradient (negative gradients will be used internally), e.g. 0.0002 K
-1
(4) data on physical quantity (name, unit of measurement, precision, etc.)
(5)
data on characteristics, entry of coefficients (can be determined by entering a test series (see section
13.2)
(6) calculation of coefficients by entering a test series (section 13.2)
(7) temperature range
(8) if the temperature exceeds the range in (7), the values in MAX and MIN will be output
(9) if the temperature exceeds the range in (7), error values will be output
(10) maximum and minimum output range for the physical quantity given in (4)
(11) if the output range exceeds the values defined in (10), the values in MAX and MIN will be output
(12) If the output range exceeds the values defined in (10), error values will be output
(13) calculation of the physical quantity using the entered values and coefficients (test)
6
7
8
9
11
12
13
10
1
2
3
4
5
Page 57
UMPIOXR400V2-7-1EN, 2012-07-01 57
13 Program RMKoeff
13.2 Calculation of Characteristics and Coefficients
The characteristics will be generated by polynomials of grade 1...3 from a data table or from a test series.
Click button GET IT FROM YOUR SAMPLES (see Fig. 13.3).
A window will be opened (see Fig. 13.4). Enter the values of the test series.
Fig. 13.4: Example of a test series
The X axis quantity is the refractive index nDT related to a standard temperature. The Y axis quantity is the required process parameter. The process parameter and its unit of measurement are user defined, e.g. concentra­tion in w%, g/l or mol/l.
In box TYPE OF FITTING, enter the type and degree of the polynomial function to be used to fit the curve to the measurement data.
The measurement range shown in the diagram can be entered manually or generated using the program RMKoeff.
Clear the two boxes FROM TABLE and AUTOMATIC if the measurement range of the diagram is to be entered manually.
Select the two boxes FROM TABLE and AUTOMATIC if the measurement range of the diagram is to be defined by the program.
Press OK.
The determined coefficients will be displayed in the boxes K0 to K3.
The created data record can be tested. Enter pairs of values for the refractive index and temperature in box 13 (see Fig. 13.3).
By pressing the right mouse button and click on the option SAVE THIS TABLE, the data table can be stored.
fit to a polynomial of degree 1
Page 58
58 UMPIOXR400V2-7-1EN, 2012-07-01
13 Program RMKoeff
13.3 Transmission of a Data Record
The data record has to be transmitted to the transmitter via the serial interface RS232. The RS232 interface can be connected only outside of an explosive atmosphere as the housing has to be opened (see Fig. 3.2).
Connect the transmitter to the computer via the serial interface.
Switch on the transmitter.
After initialization, the main menu appears.
Close all windows of the program RMKoeff except the main window.
Fig. 13.5: Main window of program RMKoeff
In the menu OPTIONS, select SERIELLE SCHNITTSTELLE.
The following window will be opened:
Fig. 13.6: Settings for serial interface
Attention!
Always observe the "Safety Instructions according to Directive 94/9/EC" (see document SIPIOX_R400).
Note!
No other settings are required on the POX R400 for reading measurement values.
Page 59
UMPIOXR400V2-7-1EN, 2012-07-01 59
13 Program RMKoeff
The transmission parameters of serial interface RS232 are given in Tab. 13.3.
Select the serial interface on the computer.
Press OK.
In the main window, click the button SEND SELECTED COEFFICIENTS to send the data record for the fluid to the transmitter.
The following window will be opened:
Fig. 13.7: Data record is sent
The data record is sent to the transmitter.
13.4 Displaying the Fluid on the Transmitter
After sending the data record to the transmitter, a new fluid can be displayed on the transmitter in PARAMETER\MEDIUM.
Scroll through the list until the new fluid appears.
The measurement can now be started with the new fluid.
Tab. 13.3: Transmission parameters of RS232 interface
baud rate 9600 bits per seconds
data bits 8
parity even
stop bits 2
protocol RTS/CTS
Page 60
60 UMPIOXR400V2-7-1EN, 2012-07-01
14 Inputs (Option)
14 Inputs (Option)
The transmitter can be equipped with max. 4 analog inputs. External transducers can be connected to the inputs and used to determine the following physical quantities:
temperature
density
pressure
kinematic viscosity
dynamic viscosity
refractive index
pH value
conductance
sound velocity
volume flow
user value
The inputs can be current, voltage or temperature inputs.
Before the inputs can be used for the measurement, proceed as follows:
configure the inputs (see section 14.1)
assign the inputs to measuring channels (see section 14.2)
activate the inputs (see section 14.3)
14.1 Configuring the Inputs
The inputs have to be scaled:
Note!
If a new input module is installed, the transmitter has to be restarted to detect the new inputs (RESET or OFF/ON).
Select SYSTEM SETTINGS\PROC. INPUTS. Press ENTER.
Select SETUP AND SCALE. Press ENTER.
Select the input to be configured (current, voltage or temperature input). Press ENTER.
Select the process quantity to be measured at the selected input (here: DENSITY). Press ENTER.
Enter the minimum value of the input signal (here: 4mA).
This value is given in the specification of the external transducer.
Press ENTER.
S Y S T E M s e t t i n g s P r o c . i n p u t s
P r o c e s s i n p u t S e t u p a n d s c a l e
P r o c e s s i n p u t I 1 : C u r r e n t 2 0 m A
I 1 : I n p u t f o r D e n s i t y
I 1 : S e n s o r M I N 4 . 0 0 m A
Page 61
UMPIOXR400V2-7-1EN, 2012-07-01 61
14 Inputs (Option)
Assign the corresponding process values to the the minimum and maximum value of the input signal.
If NO DETECTION is selected skip the following two displays.
Enter the maximum value of the input signal (here: 20 mA).
This value is given in the specification of the external transducer.
Press ENTER.
Enter the process quantity (here: DENSITY) to be assigned to the min­imum value of input signal (here
4.0 mA).
This value is given in the specification of the external transducer.
Press ENTER.
Enter the process quantity (here: DENSITY) to be assigned to the maximum value of input signal (here
20.0 mA).
This value is given in the specification of the external transducer.
Press ENTER.
Define the behavior of the external transducer in case of error. The fol­lowing items can be selected:
NO DETECTION: The external transducer does not evaluate the signal before sending it to the transmitter.
If I1 < MIN (4.0 mA): An error is sent if the value falls below the previously defined measuring range.
If I1 > MAX (20.0 mA): An error is sent if the value exceeds the previously defined measuring range.
Note!
The error behavior of the external transducer is described in the specification.
If I1 < MIN (4.0 mA) is selected:
Enter the lowest value for the measuring range before an error mes­sage is sent.
This value is given in the specification of the external transducer.
If I1 > MIN (20.0 mA) is selected:
Enter the highest value for the measuring range before an error mes­sage is sent.
This value is given in the specification of the external transducer.
The terminals to be used for the connection of the inputs are now dis­played. Press ENTER.
I 1 : S e n s o r M A X 2 0 . 0 0 m A
I 1 : M I N ( 4 . 0 m A ) = 1 . 0 0 0 0 0 g / c m 3
I 1 : M A X ( 2 0 . 0 m A ) = 1 . 2 0 0 0 0 g / c m 3
E r r o r d e t e c t i o n N o d e t e c t i o n
E r r o r i f I 1 < 3 . 0 m A
E r r o r i f I 1 > 2 1 . 0 m A
I n p u t I 1 T e r m i n a l : T 1 A , B / b
Page 62
62 UMPIOXR400V2-7-1EN, 2012-07-01
14 Inputs (Option)
14.2 Assignments of Inputs to Measuring Channels
The configured inputs have to be assigned to a measuring channel so that the measured process quantity can be considered during the measurement.
It will be requested now which other inputs are to be linked to channel A and then to the other measuring channels.
14.3 Activation of the Inputs
Inputs have to be activated if the measured process quantity has to be displayed, stored and/or output with the other measured values. The inputs will be activated in the program branch OUTPUT OPTIONS.
Press ENTER, until the input is displayed.
14.4 Functional Check
Start a measurement to test that the installed input is working.
Press key and scroll through the menu until the measured value of the process quantity is displayed.
Select SYSTEM SETTINGS\PROC. INPUTS. Press ENTER.
Select LINK OTHERS. Press ENTER.
Select the measuring channel to which the inputs are to be assigned.
Select the input to be assigned to measuring channel A (here: DEN- SITY).
Select NO MEASURING if no input is to be assigned to measuring channel A. Press ENTER.
Only the inputs that have been configured are shown in the scroll list.
Additionally, the four general inputs are displayed: CommonInp.1, CommonInp.2, CommonInp.3 and CommonInp.4.
Note!
The transmitter stores the configuration of a measurement channel before the next channel is configured. Therefore, enter all data of a channel at least once to save the changes.
Note!
The inputs will only appear if the transmitter is equipped with inputs of the corresponding type and these inputs have been configured and assigned to a measuring channel.
In OUTPUT OPTIONS, select the measuring channel whose input is to be activated. Press ENTER.
The inputs assigned to the measuring channel are displayed succes­sively under OUTPUT OPTIONS. Select YES for the inputs to be acti­vated.
Press ENTER.
S Y S T E M s e t t i n g s P r o c . i n p u t s
P r o c . i n p u t s L i n k a g e
L i n k a g e f o r C h a n n e l A :
A : D e n s i t y W i t h i n p u t I 1
O u t p u t O p t i o n s f o r C h a n n e l A :
I 1 - > D e n s i t y n o > Y E S <
9
D I S P
Page 63
UMPIOXR400V2-7-1EN, 2012-07-01 63
A Technical Data
A Technical Data
Transmitter
transmitter PIOX
TR374 TR374A2 TR374B TR374C TR379
standard field device field device for
ATEX zone 2
field device for con­nection of sensors in zone 0/1
field device for con­nection of sensors in zone 1
19 " module
power supply 100...240 V/50...60 Hz or
20...32 V DC
power consump­tion
< 15 W
measuring chan­nels
1, option: 2
signal damping 0...100 s, adjustable response time 1 s material aluminum, powder coated aluminum degree of protec-
tion according to EN 60529
IP 65 IP 20
dimensions see dimensional drawing 42TE x 3HE
(without back panel) see dimensional
drawing weight 2.8 kg 1.7 kg installation wall mounting, option: 2 " pipe mounting 19 " rack mounting operating tem-
perature
-10...+60 °C
display 2 x 16 characters, dot matrix, backlit menu language English, German data logger
(option)
SD card 2 GB---
explosion protection
zone - 2 outside of explosive
atmosphere
outside of explosive atmosphere
-
marking -
II3G Ex nA nC ic IIC T4 Gc T
a
-40...+60 °C
0637 II(1)G [Ex ia] IIB T
a
-40...+60 °C
0637
II(2)G [Ex ib] IIC T
a
-40...+60 °C
-
certification ATEX
--
IBExU06ATEX1075 X IBExU06ATEX1075 X
-
certification IECEx
- - IECEx IBE 10.0003X IECEx IBE 10.0003X -
type of protec­tion
- non sparking - - -
intrinsic safety parameters
U
m
= 250 V intrinsically safe power supply for zone 0/1 sensors: P
o
= 1.1 W
U
o
= 13.2 V
I
o
= 84 mA
C
o
= 5.1 μF
L
o/Ro
= 32.8 μH/Ω
RS485 (intrinsic safety): P
o
= 0.2 W
U
o
= 6 V
I
o
= 76 mA
C
o
= 999.3 μF
L
o/Ro
= 3.37 mH/
U
m
= 250 V intrinsically safe power supply for zone 1 sensors: P
o
= 1.1 W
U
o
= 13.2 V
I
o
= 84 mA
C
o
= 0.24 μF
L
o
= 10 mH
RS485 (intrinsic safety): P
o
= 0.2 W
U
o
= 6 V
I
o
= 76 mA
C
o
= 39.3 μF
L
o
= 10 mH
remark intrinsically safe
power supply for zone 0/1 sensors included
intrinsically safe power supply for zone 1 sensors included
Page 64
64 UMPIOXR400V2-7-1EN, 2012-07-01
A Technical Data
measuring functions
physical quanti­ties
refractive index, fluid temperature, °Brix, w% (saccharose, fructose, glucose, invert sugar), further by RMKoeff
physical quanti­ties for diagnosis
signal amplitude, sensor humidity, sensor temperature
communication
interface -
process integration:
RS485 (option)
- diagnosis: RS232
-
process integration: RS485 (option)
- diagnosis: RS232
1
-
process integration:
RS485 (option)
- diagnosis: RS232
-
process integration:
RS485 (option)
- diagnosis: RS232
-
process integration:
RS485 (option)
- diagnosis: RS232
serial data kit
software (all Win­dows
TM
ver-
sions)
RMKoeff: management of fluid data sets
cable RS232
RS232
1
RS232 RS232 RS232
adapter RS232 - USB
RS232 - USB
1
RS232 - USB RS232 - USB RS232 - USB
outputs (option)
The outputs are galvanically isolated from the transmitter.
number on request
current output
range 0/4...20 mA accuracy 0.1 % of reading ± 15 A active output R
ext
< 500
passive output U
ext
= 4...24 V, dependent on R
ext
R
ext
< 1 k
voltage output
range 0...1 V or 0...10 V accuracy 0...1 V: 0.1 % of reading ± 1 mV
0...10 V: 0.1 % of reading ± 10 mV internal re­sistance
R
i
= 500
frequency output
range 0...1 kHz or 0...10 kHz open collector 24 V/4 mA
binary output
open collector 24 V/4 mA 24 V/4 mA Reed relay 48 V/0.25 A 48 V/0.25 A optorelay 32 V/100 mA ­binary output as
alarm output
- functions limit limit
inputs (option)
The inputs are galvanically isolated from the transmitter.
number max. 4, on request
current input
range active: 0...20 mA
passive: -20...+20 mA
accuracy 0.1 % of reading ± 10 A active input U
i
= 24 V, Ri = 50 , Pi < 0.5 W,
not short-circuit proof
passive input R
i
= 50 , Pi < 0.3 W
voltage input
range 0...1 V or 0...10 V accuracy 0...1 V: 0.1 % of reading ± 1 mV
0...10 V: 0.1 % of reading ± 10mV internal re­sistance
R
i
= 1 M
1
connection of the interface RS232 outside of explosive atmosphere (housing cover open)
transmitter PIOX
TR374 TR374A2 TR374B TR374C TR379
Page 65
UMPIOXR400V2-7-1EN, 2012-07-01 65
A Technical Data
Dimensions of Transmitter (in mm)
PIOX TR374
PIOX TR379
ø 4 . 5 0
1 0
M 2 0 ( 6 x )
7 0 . 5
2 0 0
1 6 3
2 8 7
2 6 5
screw holes for wall mounting
M 2 . 5
2 1 3
1 9 8 . 1
1 2 9
1 2 2 . 5
4 3 . 7
2 0 8 . 2 5
1 2 9
1 5
1 1 0
1 7 0 . 5
2 2 2 . 2
Page 66
66 UMPIOXR400V2-7-1EN, 2012-07-01
A Technical Data
Sensor
sensor PIOX R400 R400A2 R400B R400C process parameters
medium all liquids with a turbidity < 10 000 FAU fluid temperature
(dependent on ambient tempera­ture)
-20...+150 °C (150 °C at an ambient temperature of 20 °C)
-20...+130 °C
fluid pressure max. PN 16, on request: max. PN 40
measurement
measuring principle transmitted light refractometry measuring range nD: 1.3...1.7
°Brix: 0...100 accuracy (absolute) nD: 0.0002 (corresponds to 0.1°Brix, typical 0.1 w%) repeatability nD: 0.00002 (corresponds to 0.01 °Brix, typical 0.01 w%) resolution nD: 0.000001 material housing stainless steel 304 (1.4301) wetted parts stainless steel 316Ti (1.4571), special materials on request (e.g. titanium, Hastelloy) prism sapphire, nD  1.76 gaskets EPDM, FKM (Viton) or FFKM (Kalrez) degree of protection
according to EN 60529
IP 67
flange FLEXIM, Varivent or Tri-clamp flange dimensions with
flange
see dimensional drawing
weight type of construction M: 1.7 kg
type of construction L: 2.1 kg
operating tempera­ture
-20...+60 °C
explosion protection
zone - 2 0/1 1 marking -
II3G Ex nA IIC T4 Gc T
a
-40...+60 °C
T
m
-20...+130 °C
0637; II1/2G Ex ia IIB T4 T
a
-40...+60 °C
T
m
-20...+130 °C
0637; II2G Ex ib IIC T4 T
a
-40...+60 °C
T
m
-20...+130 °C certification ATEX - - IBExU06ATEX1075 X IBExU06ATEX1075 X certification IECEx - - IECEx IBE 10.0003X IECEx IBE 10.0003X type of protection - non sparking intrinsic safety intrinsic safety intrinsic safety
parameters
- - power supply : C
i
= 141.6 nF
L
i
= 0 μH
data: C
i
= 22.7 nF
L
i
= 0 μH
power supply : C
i
= 141.6 nF
L
i
= 0 μH
data: C
i
= 22.7 nF
L
i
= 0 μH
temperature probe
type Pt1000 resolution 0.01 K accuracy at 20 °C 0.15 K
sensor cable
length 10 m cable jacket PVC operating tempera-
ture
-30...+80 °C
Page 67
UMPIOXR400V2-7-1EN, 2012-07-01 67
A Technical Data
Dimensions of Sensor (in mm)
Mounting Positions of Sensor
type of construction M type of construction L
flange type F (FLEXIM flange)
flange type V (Varivent flange N)
flange type T (Tri-clamp flange 3 ")
type of construction M type of construction L
horizontal pipe vessel
vertical pipe T-piece
241
38
46
70
150
241
46
70
28
251
140
251
241
38
150
241
Page 68
68 UMPIOXR400V2-7-1EN, 2012-07-01
A Technical Data
Extension Cable
Junction Box
Dimensions of Junction Box (in mm)
ATEX zone 1 ATEX zone 2 outside of explosive
atmosphere
type LIYCY 2x2x0.75/77 blue LIYCY 2x2x0.75/77 gray max. length m PIOX R400C: 200
PIOX R400B: 100
200
operating temperature
°C -30...+70
technical type JB91 JB92 JB93 dimensions see dimensional drawing installation wall mounting, option: 2 " pipe mounting
material
housing stainless steel 316L (1.4404) stainless steel 304 (1.4301) gasket silicone silicone degree of protection
according to EN 60529
IP 67
operating temperature
min. °C -40 max. °C +80
explosion protection
ATEX zone °C 1 2 ­marking not necessary II 3G Ex nA II
T6...T4 Ta -40...+80 °C
II 3D Ex tD A22 IP67 T 100 °C
-
type of protection °C simple apparatus according to
IEC 60079-11
non sparking, protection by
enclosure
-
remark only in combination with
intrinsic safe equipment [ia], [ib]
--
174
119
2
wall mounting holder
70
163.5
156
Ø
9
Page 69
UMPIOXR400V2-7-1EN, 2012-07-01 69
B Start-Up and Maintenance Protocols
B Start-Up and Maintenance Protocols
B.1 Start-Up
measuring point (designation):
Settings and measured values on the transmitter:
fluid characteristics was generated by:
FLEXIM customer
measured values: one representative measured value each for a typical process state at each measuring point (e.g., working point, CIP).
* measured with (e.g. laboratory refractometer, type): precision:
input of offsets:
no offset: (
MEASURE ADJUSTS? = NO, SENSOR ADJUSTS? = NO)
offset: MEASURE ADJUSTS? = offset: SENSOR ADJUSTS? =
start-up completed on: signature:
device:
sensor: R400- ser. no.:
transmitter: PXTR374 ser. no.:
process fluid
solvent
physical quantity
concentration working point:
range:
working point:
range:
other ingredients
fluid temperature
ambient temperature
pressure
selected fluid on transmitter
process
state
physical
quantity 1
(e.g., Brix,
w%)
physical
quantity 2
(e.g., Brix,
w%)
humidity
(%)
nD nDT T
fluid
T
dev
amplitude reference
value*
Page 70
70 UMPIOXR400V2-7-1EN, 2012-07-01
B Start-Up and Maintenance Protocols
B.2 Maintenance Log from
FLEXIM recommends annual maintenance.
measuring point (designation):
The fluid composition has not changed since start-up/maintenance on
A modified/new medium has been used since
type of modification:
settings and measured values on the transmitter:
fluid characteristics was generated by:
FLEXIM customer
measured values: one representative measured value each for a typical process state at each measuring point (e.g., working point, CIP).
* measured with (e.g. laboratory refractometer, type): precision:
device:
sensor: R400- ser. no.:
transmitter: PXTR374- ser. no.:
process fluid
solvent
physical quantity
concentration working point:
range:
working point:
range:
other ingredients
fluid temperature
ambient temperature
pressure
selected medium on the transmitter
process
state
physical
quantity 1
(e.g., Brix,
w%)
physical
quantity 2
(e.g., Brix,
w%)
humidity
(%)
nD nDT T
fluid
T
dev
amplitude reference
value*
Page 71
UMPIOXR400V2-7-1EN, 2012-07-01 71
B Start-Up and Maintenance Protocols
B.3 Comparison of Amplitude Values
If the amplitude value measured during maintenance is fallen significantly below the start-up value, e.g. when the medium is slightly turbid:
amplitude value
(maintenance)
1/20 amplitude value
(start-up)
,
cleaning of the sapphire optics is recommended.
Note: Make sure that the sensor material (stainless steel 1.4571) and the sensor gaskets (for material see purchase order) are resistant to the solvent. The solvent must be suitable for the fluid, too.
cleaned with .
zero point check (option), recommended after cleaning
PIOX R400 is checked with deionized water.
•Select SACCHAROSE in PARAMETER\FLUID.
target: nDT = 1.33299 ± 0.0002
input of offsets:
no offset: (
MEASURE ADJUSTS? = NO, SENSOR ADJUSTS? = NO)
offset: MEASURE ADJUSTS? = offset: SENSOR ADJUSTS? =
maintenance completed on: signature:
operating temperature nDT deviation
Page 72
72 UMPIOXR400V2-7-1EN, 2012-07-01
C Certificates
C Certificates
Page 73
Page 74
Page 75
Page 76
Page 77
Page 78
Page 79
Page 80
Page 81
Loading...