13
CHECKING PUMP ACCURACY
Fluid head height and viscosity are two variables that can affect flow rate. To check
flow rate, use the following procedure.
Select a pump set (or fully close clamp on flush bag,
if using a feeding pump set with flush bag). Be sure to use a graduated
cylinder for volume measurement—not an enteral container. Using Ensure
®
Complete, Balanced Nutrition™or Osmolite®Isotonic Liquid Nutrition, fill the
feeding container only and set up pump according to instructions. Suspend bag
to side or behind and 20 inches above pump (see illustration). With a test pump
connected to AC power, start the pump and let it run for 15 minutes. Then run
product into a graduated cylinder for 30 minutes and compare to rate set on pump:
Amount Expected in Graduated Cylinder
Rate Set @ 30 minutes
76 mL/hr = 34.2 mL to 41.8 mL
150 mL/hr = 67.5 mL to 82.5 mL
If flow rate is incorrect, try a new pump set and repeat procedure.
12
INSTRUCTIONS FOR USE (continued)
NOTE: Automatic priming expels air
from the set, which results in a small
amount of fluid being expelled
beyond the adapter.
9. Confirm proper placement and
function of enteral feeding tube.
10. Attach adapter to enteral feeding
tube.
11. Select flow rate, from 1 to 300 mL/hr,
by pressing individual arrows. Never
use sharp objects such as pens, pencils, fingernails or paper clips to
depress the keypad arrows.
12. If desired, turn dial to SET DOSE
and select dose by pressing individual
arrows.
NOTE: If zero dose is entered (0000),
this feature is deactivated.
13. Turn dial to RUN to start feeding.
14. To see feeding volume delivered (and
flush volume delivered, if this feature
is used), turn dial to VOL FED.
15. To clear the feeding and flush volume
delivered, turn dial to CLEAR VOL.
(This occurs after four beeps.)
NOTE: If alarm sounds, turn pump
dial to HOLD. Correct alarm condition indicated by visual display, then
turn pump dial to RUN to restart
feeding. (For explanation of alarms,
see pages 14-15.)
16. When desired, disconnect pump-set
tubing from patient’s enteral feeding
tube. Close clamp on feeding bag
(and flush bag, if used). Lift blue
lever on face of pump, remove cassette and discard set and container.
17. When feeding is complete, turn
pump dial to OFF.
NOTE: Pump stops and sounds an
alarm automatically when the container is empty. If the DOSE function
is in use, the pump will stop, sound
the alarm and display the DOSE
COMPLETE message when the
selected dose is reached.
PRECAUTION: The pump set and
container should be replaced as
needed, or at least every 24 hours, to
avoid product-contamination problems. For single-patient use only.
TO GRAVITY FEED OR
MANUALLY PRIME:
1. Completely close CAIR
®
clamp (and
slide clamp, if flush set is used).
2. Remove cassette from pump.
3. Pull out orange plunger knob and
regulate flow with CAIR clamp.
* CAIR®clamp manufactured under license from Adelberg
Laboratories, Inc. Covered by one or more of the following
U.S. Patent Nos: 3,685,787; 3,893,468; 4,013,263; 4,047,694;
4,238,108; 4,181,476.
†
If desired, CAIR clamp may be used to regulate flow rate by
gravity, but all tubing must be completely disconnected
from pump. This set delivers approximately 14 drops
per mL.
without a flush bag