Fisher Scientific XL150, XL200, XL250, XL500, XL600 Instruction Sheet

Fisher Scientific accumet® XL
XL150
XL200
XL250 XL500 XL600
Thank you for selecting the Fisher Scientific accumet XL series benchtop meter. All models include an electrode arm with metal bracket (13-637-671), 100/240 universal power supply (13-636-104), RS-232, and USB cable. This instrument incorporates a full color touch screen which can be controlled by a finger or an optional USB mouse. Avoid damage to the display by keeping display from contact with chemicals, hot liquids, and all sharp or pointed objects. This instruction sheet serves five
models, so not all information presented here is relevant to each model.
Get It Together
2
MINI-B USB:
USB communication
for field software
upgrades (cable
included) and
data output
STIRRER A: For use with optional stirring probe (13-620-BSP)
Right
RS-232:
RS-232 serial communication output, 2.5 mm
jack. For use with
RS-232 cable
13-637-681 (included)
GROUND: For electrical grounding/electrical noise reduction
Rear View of XL600
TEMP:
Automatic Temperature Compensation probe (13-620-19) and pH electrodes that also include ATC. Note:
Models XL250, XL500, & XL600 have Dual BNC Channels, each with Temperature
DO: 8-pin DIN connection for 13-620-SSP Self­Stirring Dissolved Oxygen/Temperature electrode. Note:
Model XL600 only
2/4 CELL:
8-pin DIN connection for 2-cell or 4-cell Cond/TDS/ Resistivity/Salinity/ Temperature electrodes. Note:
Models XL200, XL500, XL600 only
BNC: For combination pH, mV (ORP, Redox), or Ion Selective Electrodes (ISE).
Note: Models XL250, XL500, & XL600 have Dual BNC Channels, each with Temperature
POWER:
Power supply, 100/240 VAC, 9 VDC, 18W, center positive. Use only model 13-636-104 (included)
Left
STIRRER B:
For use with optional stirring probe (13-620-BSP)
RJ45:
Ethernet port for network/ internet connectivity
1
In The Box
Electrode Arm With Metal Bracket (13-637-671)
Navigation
3
100/240 Universal Power Supply (13-636-104)
Administration: Use to create and edit user login ID
profiles and passwords.
System Setup: Use to configure and view system
settings. These include stirrer speed, printer, touch screen calibration, date & time and brightness settings.
Multi Channel: Use to display measurement mode of
multiple channels simultaneously. Not available with XL150.
Single Channel: Use to display one measurement mode.
Note: Single Channel is required to access the Standardi­zation mode, Printing function, and on-screen graphing.
RS-232 USB Cable
Setup: Use to customize changes to the
measurement parameter.
Print: Use to print data manually from measurement
mode. If you selected the “LOG DATA” option in the “Print Criteria” setup, the Print icon in the measurement mode will send data to the printer. Available from
single-channel measurement display only.
Log Data: Use to save the current measurement data
manually to memory. Sample ID entry is required before data can be saved (this is available from Setup).
Connect the metal electrode bracket to the base of the meter using a Phillips screwdriver. The bracket may be attached to the left, right, or center according to your preference.
Attach the electrode arm to the bracket & tighten the wing nut. Purchase an additional arm/bracket (13-637-671) to utilize two arms in left & right positions simultaneously.
If the instrument will be wall mounted using the wall mount holes, the electrode arm should be removed from the bracket.
Attach electrodes and output cables as required into the corresponding ports of the meter.
Prior to attaching pH, ORP, or Ion electrodes, the BNC shorting cap (13-620-99) should be removed. Do not discard the shorting cap. This cap is useful for simulating a 0 mV signal and also as a protective covering when the BNC connection is not in use. A strap will retain the cap in place when an electrode is connected.
Display Setup: Use to configure the display options and
select input parameters.
Standardize: Use to standardize (calibrate) the
desired parameter. Note: Available from single-channel measurement display only and not used for mV / relative mV offset adjustment.
Measure: Use to resume live measurement after
“Auto Read” function has been activated and the measurement is frozen. Note: Not available with mV,
BOD, OUR, or SOUR display.
Mode: Use to change operation modes. Also use to
gain access to Administration, System Setup, About, and Internet Explorer icons.
Log View: View saved data in memory on the display.
Standardization View: View standardization
(calibration) data on the display.
Stirrer A/B: On/Off control of optional stirring probe
(13-620-BSP) connected to corresponding Stirrer port A or Stirrer port B.
IExplorer: Internet Explorer mobile browser.
Note: sound and video are not enabled.
Help: Provides screen specific help from various screens.
Setup Your Parameter
4
When a user presses ‘Setup’ located on the display, one of the four windows shown at the right will open.
For All Models For XL200, XL500, XL600 For XL250, XL500, XL600 For XL600 Only
Conductivity / TDS / Resistivity / Salinity Setup
Ion Selective Setup
Dissolved Oxygen (DO) SetuppH / mV Setup
5
pH Buffer Group pH Values
USA 2.000, 4.010, 6.997, 10.013, 12.000
NIST 1.678, 4.010, 6.865, 9.184, 12.460
DIN (19267) 1.090, 3.060, 4.650, 6.790, 9.230, 12.750
FSCI
(Fisher Scientific)
PW (Pure Water) 4.10, 6.97, 9.15
CUSTOM Any 2 - 5 values, ≥ 1.0 pH unit apart
1.000, 3.000, 6.000, 8.000, 10.000, 13.000
Standardization (Calibration)
For best results, Standardize with one or more calibration standards prior to measurement. From single-channel display mode, press “Standardize” to begin standardization. Rinse your electrode with clean water and immerse in your first calibration standard. When the reading is stable, press “Confirm” to accept the value, and repeat with additional standards. Press “Clear” to erase all values.
Range#Conductivity
Range
r 1 0.00 – 19.99 µS None None
r 2 20.00 – 199.9 µS 84 µS 76 µS
r 3 200.0 – 1.999 mS 1413 µS 1278 μS
r 4 2.000 – 19.99 mS 12.88 mS 11.67 mS
r 5 20.00 – 500.0 mS 111.8 mS 102.1 mS
Standardization Icons
Use Confirm to accept the standard value.
Use Clear to erase previous standardization.
Use Cancel to returns to measurement without accepting the standard value.
Use Temp Std to standardize (calibrate) the temperature.
Automatic Calibration Values
Normalization Temperature
25 ºC 20 ºC
Setup other parameters by selecting the appropriate tab.
You Are Ready To Begin Measurement!
6
Visit www.fishersci.com/accumet for product updates and more!
For Customer Service:
1-800-766-7000 (USA) 1-800-234-7437 (Canada) exports@fishersci.com www.fishersci.com
For Technical Support:
1-888-358-4706 (North America) accumet@fishersci.com
Thermo Fisher Scientific (Australia) +1300-735-292 thermofisher.com.au Fisher Scientific GmbH (Austria) 0043-800-208840 at.fishersci.com Fisher Scientific (Belgium) 056-260-260 be.fishersci.com Fisher Scientific (Canada) 1-800-234-7437 fishersci.ca Fisher Scientific (China) +400-881-5117 fishersci.com.cn Fisher Scientific SPOL. S.R.O. (Czech Republic) +420-466-798-230 thermofisher.cz Fisher Scientific (Denmark) +45-70-27-99-20 fishersci.dk Fisher Scientific (Finland) +358-9-8027-6280 fishersci.fi Fisher Scientific (France) 03-88-67-53-20 fr.fishersci.com Fisher Scientific GmbH (Germany) 0049-2304-932-5 de.fishersci.com Fisher Scientific (India) +1-800-209-7001 fishersci.in Fisher Scientific (Ireland) +35-301-899-5854 ie.fishersci.com Fisher Scientific (Italy) 02-953-28-258 it.fishersci.com Fisher Scientific (Korea) +02-527-0300 fishersci.co.kr
© 2012 Thermo Fisher Scientific Inc. All rights reserved.
68X613701 Rev 0 July 2012
Fisher Scientific (Malaysia) +603-5122-8888 fishersci.com.my Fisher Scientific (Netherlands) 020-487-70-00 nl.fishersci.com Thermo Fisher Scientific (New Zealand) +0800-933-966 thermofisher.co.nz Fisher Scientific (Norway) +47-22-95-59-59 fishersci.no Fisher Scientific (Portugal) 21-425-33-50 pt.fishersci.com Fisher Scientific (Singapore) +65-6873-6006 fishersci.com.sg Fisher Scientific (South East Asia) +65-6873-6006 fishersci.com.sg Fisher Scientific (Spain) 902-239-303 es.fishersci.com Fisher Scientific (Sweden) +46-31-352-32-00 fishersci.se Fisher Scientific (Switzerland) 056-618-41-11 ch.fishersci.com Fisher Scientific (United Kingdom) +01-509-231166 fisher.co.uk Fisher Scientific (United States) 1-800-766-7000 fishersci.com
Loading...