22. Return of Items ........................................................................................................... 42
23. Notice of Compliance ................................................................................................ 43
24. Declaration of Conformity ......................................................................................... 44
2
1. Introduction
Thank you for selecting the Fisher Scientific accumet AB series benchtop meter.
All models include an electrode arm with metal bracket (13-637-671), 110/220
universal power supply (13-636-104), RS-232, and USB cable. The electrode arm
can be attached to the left, right, or center of the meter according to your
preference using a Phillips screwdriver.
The AB150 replaces the AB15 and AB15Plus series while the AB200 and AB250
are new introductions to the product line.
Please note that this manual serves three models, so not all sections of the
manual are relevant to each model. The chart below lists the parameters served
for the various models:
13-637-671 13-636-104
3
2.Keypad & Display
FIXED KEY DESCRIPTION
to initialize and complete standardization (calibration) of
STD
displayed parameter
MODE
SETUP
PRINT / ENTER
SAVE / ▲
VIEW / ▼
STIRRER
POWER / BACKLIGHT
SOFT KEY DESCRIPTION
HELP
DISPLAY
REPORT
TEMP STD
ACCEPT
CLEAR
BACK
to toggle between available measurement modes
to enter Setup mode; use to change system settings or
measurement parameters
to print, also to confirm selection during setup
to record into memory without printing, also increment
to examine saved data, also decrement
to turn 13-620-BSP stirrer (optional separately) on / off
to adjust backlight brightness, also hold >3 seconds to power off
to view screen specific assistance
to toggle between available display types
to examine current standardization data such as slope
to initialize temperature standardization (calibration)
to accept existing values
to erase existing values
to return to previous screen
ESC
to escape or cancel from the current display
4
Stable
(ATC)
(MTC)
USB port is active.
RS-232 port is active.
Based on the stability speed selected, instrument has detected a
stable reading.
Data was sent to the printer; either manually by pressing PRINT key or
automatically at timed intervals as entered thru the System Setup mode.
“High Alarm”, “Low Alarm”, or “Std Due” alarm limit had been reached.
Note: High/Low alarms also use an audible notification.
Standardization detail.
Measuring Range # of displayed value. Applicable to Conductivity, TDS,
Salinity, or Resistivity measurements only.
Automatic Temperature Compensation is active; temperature is being actively
acquired with attached electrode. ATC is recommended.
Manual Temperature Compensation is active. This occurs when a temperature
sensor may not be attached and the default temperature is used.
See Section 9 – Temperature Standardization to calibrate the ATC or MTC
temperature values.
5
Connections:
Power
RS-232
MINI-B
USB
2/4 Cell
TEMP
GROUND
BNC
Power supply, 110/230 VAC, 9 VDC, 18 W, center positive.
Use only model 13-636-104 (included).
RS-232 serial communication output, 2.5 mm jack.
For use with RS-232 cable 13-637-681 (included).
USB communication for field software upgrades (cable included)
and data output.
8-pin DIN connection for 2-cell or 4-cell
Con/TDS/Resistivity/Salinity/Temperature electrodes.
Automatic Temperature Compensation probe (13-620-19) and
pH electrodes that also include ATC.
For electrical grounding / electrical noise reduction.
For combination pH, mV (ORP, Redox), or
Ion Selective Electrodes (ISE). Half-cell indicating electrodes
require 13-620-489 adapter.
6
3. System Setup & Configuration
Use the System Setup to customize operation of your AB series meter.
Press SETUP from the measurement screen and then press PRINT/ENTER
when SYSTEM is highlighted to access these settings.
The following settings can be customized for each model:
STABILITYSTIRRER
EXPORT DATASAMPLE ID
COMM SETUPLANGUAGE
DATE & TIMEBEEP
PASSWORD
FACT RESET
CONTRAST
STABILITY
Adjust how quickly and frequently the ‘Stable’ indicator appears. There are four
levels of adjustment; SLOW, MEDIUM, FAST, or OFF (prevents the Stable indicator from appearing in measurement mode). To display ‘Stable’ more
quickly and more often, use “FAST” setting. MEDIUM or SLOW is recommended
for most applications.
SLOW: The ‘Stable’ indicator will take longer to appear and will appear less
frequently. Use this setting if you want the best results and don’t mind waiting
longer to get them.
MEDIUM: This is the factory default setting. It provides a balanced response
which works best for most applications.
FAST: The ‘Stable’ indicator will appear quickest of any setting. If you find that
the indicator appears and disappears too frequently, you may want to select a
slower setting. This setting is often not suitable for use with most non-refillable
(gel filled) electrodes which are generally slower to respond.
Tip: Select a lower pH resolution X.XX instead of X.XXX to decrease
stabilization time.
EXPORT DATA
Printing: Select ON or OFF.
Datalog: Select ON or OFF.
7
Timed: When either printing or datalog is ON, the option to select Timed
readings ON or OFF appears. Select “ON” to automatically print or save data
at selectable intervals. Select “OFF” to save data manually using the
PRINT/ENTER key.
Time Interval: (MM:SS) (Minutes:Seconds). When timed reading is ON, you
must select an interval at which the data will be sent to the Printer and/or Datalog
memory. Choose from 3 seconds to 60 minutes. Data will be collected until the
data limit is reached or the option is turned “OFF”. This feature is useful for
gathering data from a single sample over time. Note that “SAVE” is not available
during TIMED setting. To view the stored data, use “VIEW” key.
COMM SETUP (Communication Setup)
Format: To send data as a comma separated value, choose CSV – best for
exporting data into spreadsheet software. Choose PRINTER to send the data in
an easily viewable format – best for printing.
Communication Setup: Choose RS-232 (9600 or 19200 baud rate) or USB.
Baud Rate: When RS-232 is selected, choose (9600 or 19200)
DATE & TIME
Setting the correct date and time is required for GLP and will apply to power on,
measurement, data log, and print functions. Factory reset will not apply to date
and time setting once it has been set. Changes related to daylight savings time
must be manually entered.
Date Format: Select (MM DD YY) or (DD MM YY)
Time Format: Select 12Hrs (AM/PM) or 24Hrs
PASSWORD
Select “ENABLE” to restrict access to Calibration and Setup modes. When
password protection is enabled, password entry is required before performing any
calibration, or making changes to the setup mode. Setup parameters can be
viewed, but not changed without correct password entry. The password is a user
selectable number from 1 to 99999.
Select “DISABLE” if password protection is not desired.
The meter does not allow you to edit setup parameters or perform a new
calibration unless you enter the correct password. If an incorrect password is
entered 3 consecutive times, the meter returns to measurement mode.
If a password has been ENABLED but cannot be recovered, a password can be
provided via a written request. The instrument serial number and your contact
information are required.
8
FACT RESET
Select “YES” to reset the to the factory default settings except; Date & Time
setting, and data stored into memory.
CONTRAST
Optimize the contrast setting of your display for best visibility in your surrounding
lighting conditions. Test various contrast settings for best results. This setting will
be applied to both backlight and non-backlight conditions.
STIRRER
Set the stirring speed of the 13-620-BSP Benchtop Stirring Probe (optional
accessory) when stirrer is on. Choose from 5 different speeds.
SAMPLE ID
This is a user selectable number from 1 to 99999. Incorporating a sample ID to
identify one or more data points is useful to distinguish data that is saved into
memory or sent to a PC or printer. Use the ▲/▼ key to adjust the values and
ENTER to move the cursor of a previous Sample ID.
LANGUAGE
Select the preferred language that is displayed on the instrument. Choose from
English, Deutsch, 中文, Français, Italiano, & Español.
Tip: Check www.fishersci.com/accumet
updates which may include additional languages.
for free instrument software
BEEP
Enable or disable an audible tone with each key press.
9
4. Setup pH & mV
Use Setup pH or mV mode to customize these parameters. Note: mV setup offers
Alarm setting only.
Press SETUP from the measurement screen and then press ENTER when pH or
mV is selected.
BUFFER (pH Only)
Choose one of (5) pH buffer groups for calibration; The pH values in the selected
buffer group are used for auto-recognition during pH calibration. Choose the
appropriate buffer group based on the pH calibration standards you intend to use.
Buffer Group pH Values
USA 2.000, 4.010, 6.997, 10.013, 12.000
NIST 1.678, 4.010, 6.865, 9.184, 12.460
DIN 19267 1.090, 3.060, 4.650, 6.790, 9.230, 12.750
FSCI 1.000, 3.000, 6.000, 8.000, 10.000, 13.000
CUSTOM Any 2-5 values, ≥ 1.0 pH unit apart
RESOLUTION (pH Only)
Select your desired pH resolution: X.X, X.XX, X.XXX
ALARM
Use a visual and audible alarm to alert you when High or Low values that have
been set from this menu are exceeded. “High Alarm” or “Low Alarm” will blink on
the display if the values are exceeded while simultaneously, a loud, intermittent
beeping sound is heard. The alarms will continue until the conditions are no
longer met, and will only be active during measurement mode.
STD DUE (pH Only)
When enabled, the “Cal Due” indicator blinks if the number of days since the last
calibration has been exceeded. Set the time from 8 hours, 16 hours, or from 1 to
31 days.
10
5. pH Standardization (With Preset Buffer Group)
For best results, periodic standardization (calibration) with known accurate
standards is recommended prior to measurement. Use standards that bracket
your intended measuring range while including a neutral point (7.00, 6.86, or
6.79). For example, if you expect to measure samples from pH 6.2 to 9.5,
calibration with 4.01, 7.00, and 10.01 will work well.
The following pH calibration buffers in the groups listed below are automatically
recognized:
Buffer Group pH Values
USA 2.000, 4.010, 6.997, 10.013, 12.000
NIST 1.678, 4.010, 6.865, 9.184, 12.460
DIN 19267 1.090, 3.060, 4.650, 6.790, 9.230, 12.750
FSCI 1.000, 3.000, 6.000, 8.000, 10.000, 13.000
To eliminate temperature errors associated with the pH electrode, attach the
automatic temperature compensation (ATC) probe for best accuracy. Without
temperature compensation, pH accuracy will decrease as samples deviate from
25 ºC and pH 7.
If the pH electrode has been stored dry, soak in storage solution for 10 minutes
before standardization to saturate the pH electrode surface and minimize drift. If
storage solution is not available, use a neutral pH buffer. Do not reuse buffer
solutions after calibration.
1) Turn meter on, press MODE if needed for pH measurement mode.
2) Rinse the pH and ATC electrodes with clean water then submerse in your first
pH standard buffer.
3) Press STD to enter standardization mode.
If the meter is password protected, you will be prompted to
enter a password.
4) The AB meter automatically searches for the appropriate standard from
the chosen buffer group in the secondary display. This value will blink
when the ‘Stable’ indicator appears. Selecting CLEAR will clear existing
calibration values.
11
5) Press STD to accept the value of the measured buffer. If successful, the
Standardization Report will show the accepted value along with other data.
The percentage slope is visible after two or more calibration points have
been completed.
6) Repeat Steps 4 & 5 with additional pH buffers.
6.pH Standardization (With Custom Buffers)
Buffer Group pH Values
CUSTOM Any 2-5 values, ≥ 1.0 pH unit apart
1) Turn meter on, press MODE if needed for pH measurement mode.
2) Rinse the pH and ATC electrodes then submerse in your custom pH buffer –
any pH value can be used.
3) Press STD to enter standardization mode. When stable, use ▲/▼ keys to
adjust the custom pH buffer value.
For best results, enter the value of the custom buffer that
corresponds to the measured temperature. For example, pH
10 buffer is actually 10.06 at 20°C, so the value should be
adjusted to 10.06 instead of 10.00 if the temperature is 20°C.
4) Press STD to accept the value.
5) Rinse the pH and ATC electrodes then submerse in next custom pH buffer.
Use any pH value that is at least 1.0 pH unit from custom pH standards that
have already been calibrated. Press CLEAR to clear existing values – this will
be necessary when repeating a standardization using the same buffer value.
6) When stable, use ▲/▼ keys to adjust the custom pH buffer value. Previously
accepted values will be visible inside of a beaker icon. Press STD to accept
the new value. Percentage slope is visible after two or more values have
been completed.
7) Repeat Steps 2 thru 4 with additional pH buffers (up to 5 custom buffers) in
any order.
12
7. mV Standardization (Offset Adjustment)
Oxidization Reduction Potential (ORP or Redox) as measured by an ORP
electrode in milliVolt (mV) units is not a precise measurement, but is useful as a
relative indicator. As such, mV offset adjustment is not meant to enhance
accuracy, but rather to make readings comparable to a reference.
Commercial ORP solutions are commonly used as a check standard – a
meter/electrode system is verified to be close to a given value although
adjustments are not made. These solutions can be used as a calibration standard
in which adjustments are made in an attempt to match the ORP value, however
results are often difficult to reproduce.
1) Turn meter on, connect a combination ORP electrode (such as 13-620-81),
and press MODE if needed for mV (or R.mV) measurement.
2) Dip the ORP electrode into a solution with a known mV value (i.e. Zobel,
Light’s, quinhydrone, or iodidetriiodide) with brief or slow stirring.
3) Press STD to begin standardization.
4) Adjust the value using ▲/▼ keys, press STD when the reading is stable to
accept the calibration offset. The maximum adjustable value is ±150 mV from
the factory default mV value.
5) To view the Ion Standardization Report at any time from the measure screen,
press REPORT.
When an offset has been stored successfully, R.mV replaces
mV in measurement mode.
8. Temperature Setup
Use Setup Temperature mode to desired temperature units; ºC or ºF.
Press SETUP from the measurement screen and then press ENTER when
TEMPERATURE is selected.
Choose ºC or ºF units of measure.
Automatic Temperature Compensation (ATC) is automatically applied when a
temperature sensor is connected.
Manual Temperature Compensation (MTC) of 25 ºC is used if the temperature
sensor is not connected. See Section 9 – Temperature Standardization
the MTC temperature value.
13
to adjust
9. Temperature Standardization
The thermistor sensor used for automatic temperature compensation and
measurement is both accurate and stable, so require frequent calibration isn’t
required. Temperature calibration is recommended upon electrode replacement,
whenever the temperature reading is suspect, or if matching against a certified
thermometer is desired.
1) Connect the temperature probe and place into a solution with a known
accurate temperature such as a constant temperature bath. The temperature
probe may be separate or as part of a pH or Conductivity electrode. *The
AB200 has two temperature inputs – one built into 2/4 conductivity cell, and
another for use with the BNC connection.
2) When the reading is stable, press STD from any measurement mode, then
press TEMP STD.
3) The upper display shows the active temperature while the lower display
shows the factory default temperature without calibration.
4) Adjust the upper display using ▲/▼ keys, press ENTER to accept the
calibration temperature. The maximum adjustable value is ±5 ºC (or ±9 ºF)
from the factory default temperature value.
Additional Temperature Calibration Notes:
For Manual Temperature Compensation (MTC) the procedure is the same
although the maximum adjustable value is the full temperature range of the
parameter, and connection of a separate ATC probe is not necessary. ATC is
recommended for most applications.
If temperature standardization is performed, be sure that the
thermometry source being used as a reference is accurate!
Use Setup mode to customize Conductivity, TDS, Salinity, & Resistivity
parameters.
Press SETUP from the measurement screen and then press ENTER when
Conductivity, TDS, Salinity, or Resistivity is selected.
STD METHOD (Conductivity Mode Only)
Choose automatic or manual standardization for conductivity, and manual
standardization for TDS, Salinity, and Resistivity. This option allows you to select
AUTO (automatic) or MANUAL conductivity calibration.
In the automatic calibration mode, the meter will automatically select one of (4)
conductivity calibration standard values depending on the range and
normalization temperature being used (see table below).
Range
#
r 1 0.00 – 19.99 µS None None
r 2 20.00 – 199.9 µS 84 µS 76 µS
r 3 200.0 – 1.999 mS 1413 µS 1278 µS
r 4 2.000 – 19.99 mS 12.88 mS 11.67 mS
r 5 20.00 – 500.0 mS 111.8 mS 102.1 mS
If you will only use one or more of the (4) calibration standards listed here,
automatic calibration is recommended. Otherwise, manual calibration should
be selected.
The factory default is automatic conductivity calibration. Note that automatic
calibration is not available for conductivity range 1.
Conductivity
Range
Automatic Calibration Values
Normalization Temperature
25 ºC 20 ºC
STD POINTS
Use SINGLE for Single-Point Calibration (SPC) to apply a single calibration value
across all ranges.
Use MULTI for Multi-Point Calibration (MPC) to calibrate each range individually.
This will restrict an individual calibration so that it is applied to one range only.
When using multi-point calibration, perform a calibration in each range that you
expect to use for best results.
The factory default is SPC. SPC or MPC will be shown with the calibration details
in measurement mode.
15
PURE WAT COEF (Pure Water Coefficient)
Choose ENABLE to automatically apply pure water temperature correction for
measurements below 2 µS/cm (ultra pure water).
Note: this option does not apply to Salinity mode.
TEMP COEFFICIENT (Linear)
The temperature coefficient is the amount of change in conductivity per degree
temperature (% per ºC). The factory default setting is a temperature coefficient of
2.1 % per ºC. For most applications this will provide good results. The meter
allows adjustment from 0.0 to 10.0.
Note: this option does not apply to Salinity mode.
Select 0.0 % for uncompensated measurements.
The temperature will be measured by the electrode and
displayed in measurement mode – but without compensation.
NORMALIZATION TEMPERATURE (ºC)
When Automatic Temperature Compensation is used, measurements are
adjusted by the temperature coefficient %, to the normalization temperature. This
value is adjustable from 15.0 to 35.0 ºC.
Use the normalization temperature that is referenced on your calibration
standard(s) – usually 25 ºC.
TDS FACTOR (TDS Mode Only)
The TDS conversion factor is a multiplication factor used to convert from
conductivity to TDS. The TDS conversion factor automatically adjusts the reading.
Select the desired TDS factor from 0.400 to 1.000.
CELL CONST (Cell Constant)
The AB200 kit includes a probe with a nominal cell constant (k) of 1.0.
Use probes with k = 0.1 and 10 (sold separately) for improved performance in
extreme sample ranges. The factory default is 1.0 to match the included probe.
Cell constant can be adjusted from 0.010 to 10.000.
k = 0.1 ideal for low measurements <20 µS (<10 ppm).
k = 1.0 ideal for mid-range measurements
k = 10 ideal for high measurements >20 mS (>10 ppt).
16
ALARM
Use a visual and audible alarm to alert you when High or Low values that have
been set from this menu are exceeded. “High” or “Low” will blink on the display if
the values are exceeded while simultaneously, a loud, intermittent beeping sound
is heard. The alarms will continue until the conditions are no longer met, and will
only be active during measurement mode.
STD DUE
When enabled, the “Std Due” indicator blinks if the number of days since the last
calibration has been exceeded. Set the time from 8 hours, 16 hours, or from 1 to
31 days.
11. Conductivity Standardization (Automatic)
For best results, periodic standardization (calibration) with known accurate
standards is recommended prior to measurement. Use standards that are close in
value to your intended sample(s).
When using multi-point calibration, perform a calibration in each range that you
expect to use for best results.
If the conductivity electrode has been stored dry for some time, soaking in alcohol
or clean water for a few minutes can help performance.
1) Turn meter on, press MODE if needed for conductivity measurement mode.
2) Rinse the electrode with clean water then submerse in your first your
conductivity standard – for automatic mode, this is either 84 µS, 1413 µS,
12.88 mS, or 111.8 mS. Provide stirring for best results.
3) Press STD to enter standardization mode.
4) When ‘Stable’ appears, use the ▲/▼ keys to adjust the cell constant
manually or press STD without adjustment to keep the nominal cell constant
value displayed and proceed with automatic calibration.
5) Wait for the meter to lock on the standard and press STD after ‘Stable’
appears to confirm the value. The calibrated value, Range (R) that was
calibrated and the calculated cell constant are now shown.
6) For multi-point calibration repeat steps 2 & 3 with additional standards.
Calibrate one point per range, up to 4.
7) To view the Conductivity Standardization Report at any time from the
measure screen, press REPORT.
17
Additional Notes On Automatic Conductivity Calibration:
A maximum of one calibration point per range can be performed. If multiple
calibration points are used in the same range, the most recent one will replace
the previous one.
If the electrode is replaced, it is best to clear the calibration to the factory
default values.
Rinse or immerse the probe before standardization and between samples with
clean water (deionized water is ideal).
For best results always begin with your lowest calibration standard value,
followed by the next lowest, and so on.
Low conductivity standard solutions (less than 20 µS) are unstable and are very
temperature dependent. As a result, reproducible calibration results are
challenging in lowest measurement range (0.00 to 20.0 µS).
“2 Cell’ or “4 Cell’ electrode is automatically detected and
displayed on the measurement screen when connected.
12. Cond/TDS/Salinity/Resistivity Standardization
(Manual Adjustment)
For best results, periodic standardization (calibration) with known accurate
standards is recommended prior to measurement. Use standards that are close in
value to your intended sample(s).
When using multi-point calibration (MPC), perform a calibration in each range that
you expect to use for best results.
If the conductivity electrode has been stored dry for some time, soaking in alcohol
or clean water for a few minutes can help performance.
1) Turn meter on, press MODE if needed for the appropriate mode.
2) Rinse the electrode with clean water then submerse in your first your conductivity standard. Press STD to enter standardization mode.
3) When ‘Stable’ appears, use the ▲/▼ keys to adjust the cell constant
manually or press STD without adjustment to keep the nominal cell constant
value displayed and proceed with automatic calibration.
18
4) The upper display is the active reading while the lower display is the factory
default value. Use the ▲/▼ keys to adjust the upper display to match the
desired value.
5) Press STD after ‘Stable’ appears to confirm the value. The calibrated
value, Range (R) that was calibrated and the calculated cell constant are
now shown.
6) For multi-point calibration repeat steps 2 thru 5 with additional standards.
Calibrate one point per range, up to 5.
7) To view the Conductivity Standardization Report at any time from the
measure screen, press REPORT.
Additional Manual Calibration Notes:
A maximum of one calibration point per range can be performed. If multiple
calibration points are used in the same range, the most recent one will replace
the previous one.
When the electrode is replaced, it is best to clear the calibration to the factory
default values.
Rinse or immerse the probe before calibration and between samples with clean
water (deionized water is ideal).
For best results always begin with your lowest calibration standard value,
followed by the next lowest, and so on.
Low conductivity standard solutions (less than 20 µS) are unstable and are very
temperature dependent. As a result, reproducible calibration results are
challenging in lowest measurement range #1 (0.00 to 20.0 µS).
19
13. Ion Setup
Use Setup ISE mode to customize this parameter. Press SETUP from the
measurement screen and then press ENTER when ISE is selected.
MEASURE UNIT
Select ppm, molar, or mg/L ion concentration units.
ALARM
The AB250 offers a visual and audible alarm to alert you when the High or Low
values that have been set from this menu are exceeded. “High Alarm” or “Low
Alarm” will blink on the display if the values are exceeded while simultaneously, a
loud, intermittent beeping sound is heard. The alarms will continue until the
conditions are no longer met, and will only be active during measurement mode.
STD DUE
When enabled, the “Std Due” indicator blinks if the number of days since the last
calibration has been exceeded. Set the time from 8 hours, 16 hours, or from 1 to
31 days.
STANDARDS
Choose one of (5) ISE buffer groups available for standardization, based on the
standard calibration values you intend to use.
Group ISE Values
1 0.001, 0.01, 0.1, 1, 10, 100
2 0.01, 0.1, 1, 10, 100, 1000
3 0.1, 1, 10, 100, 1000, 10000
4 0.02, 0.2, 2, 20, 200, 2000
5 0.05, 0.5, 5, 50, 500, 5000
20
14. Ion Standardization
The AB250 can measure ion concentration such as ammonia or fluoride when
using an ion selective electrode (ISE) for the specific ion of interest.
Ion standardization is required with at least two calibration standards in order for
the instrument to calculate the measurement.
The primary display will show “- - - -” when ion calibration is
required for ion measurement.
Prepare ion standards that bracket your measurement range. Follow the ion
selective electrode instruction manual for the appropriate ionic strength
adjustment and sample preparation.
For best results always begin with your lowest standard value, followed by the
next lowest, and so on.
1) Connect the ISE and press MODE as needed for concentration mode.
2) Rinse the ISE with clean water then dip into your lowest calibration standard
and stir. Observe the lower display which contains the mV reading of the ISE.
3) Press STD. Use ▼/▲ keys to match the upper display to your ISE calibration
standard value.
4) After ‘Stable’ appears, press STD to accept the standardization value.
5) Rinse the ISE with clean water and dip into your next lowest calibration
standard and stir. Press STD after “Stable” appears to accept the next
standardization value.
6) The mV/decade slope value will be displayed upon a successful calibration.
“Slope Error” indicates that the calibration for the current point was not
successful. This occurs when the slope (mV difference between two
consecutive points) is <15 mV/decade or >90 mV/decade.
7) Repeat steps 4 & 5 with additional ISE standards if desired.
8) To view the Ion Standardization Report at any time from the measure screen,
press REPORT.
Refer to your Ion Selective Electrode instruction manual for
details on conditioning, storage, maintenance, calibration
standard preparation, Ionic Strength Adjustment,
troubleshooting, etc. Each ISE is unique and requires care and
operation that is specific to the electrode and ion of interest.
21
15. Viewing, Transferring, and Printing Data
To send data directly to a printer or PC you will require the appropriate RS-232 or
USB cable.
Press VIEW key to view calibration or stored data directly on the display.
Press REPORT soft key to view calibration data directly on the display.
Press PRINT key to print/export stored data manually to a dedicated printer
or computer.
Press PRINT key to print/export calibration data manually to a dedicated
printer or computer.
Use the TIMED Datalog function to print/export live data automatically at
intervals to a dedicated printer or computer (3 second minimum rate). See
EXPORT DATA within the SYSTEM SETUP MENU to use this setting.
Choose the “PRINTER” format option to transfer the data in an easily viewable
format – best for printing.
Choose the “CSV” format option to send the data as a comma separated value –
best for exporting data into software
Press VIEW and select DELETE ALL to erase all saved data that was manually
or automatically stored into memory. This step is most commonly required when
the TIMED Datalog interval is not turned off and as a result, fills the memory to
capacity. When the memory is full, deleting the saved data is required in order to
save additional data. The instrument will never erase old data in favor of new data
when the memory is full.
If you wish to view, print or send the saved data to a PC before deleting, use the
PRINT key.
To make changes to the printer, datalog, and output settings, see Section 3 –
EXPORT DATA.
22
16. Standardization Report
The Standardization Report provides detailed information on the most recent
calibration. It includes date & time, calibration standard(s), offset, temperature,
number of days calibration is overdue, and slope information.
The Standardization Report is automatically displayed every time a standard has
been successfully accepted.
Important Note For pH Standardization:
If the calculated pH slope is not within 80.0 % - 110.0 %, the value is rejected and
returns to standardization mode after the following message is displayed;
"Slope Error! Allowable limit is 80 % to 110 %"
If the calculated pH slope is within 80.0 % - 110.0 %, but not within
90.0 % - 102.0 %, the value is accepted after the following message is displayed;
"Warning: Slope is not within 90-102 %.
Check electrode / buffer / temperature or clear values."
Important Note For ISE Standardization:
The allowable limit for ISE slope is 15 to 90 mV per decade.
To view the calibration report at any time, press REPORT from the desired
measurement screen. To send the calibration report data to a printer or PC, press
PRINT/ENTER key.
Check out www.fishersci.com/accumet for the latest
software available for your meter and other useful information.
23
17.Troubleshooting
Parameter Cause Solution
pH Cal points are
too close
pH Reading is 7.00 and
won’t change
pH or Ion Slope error /
Standardization
error
Ion Display reads “- - -“ 2-point standardization has not
Conductivity Can’t adjust /
can’t calibrate
Conductivity 2nd Calibration point
replaces the 1st
Custom pH buffer calibration values must
be at least 1.0 pH unit apart.
Electrode tip may be cracked or broken.
Replace electrode.
Calibration values are too far apart,
or too far from ideal values. Check the
electrode, use fresh buffers, and check
meter settings. Electrode may require
filling solution, conditioning, or complete
replacement. Refer to electrode manual.
been performed.
AUTO calibration values (84 µS, 1413
µS, 12.88 mS, or 111.8 mS) are not
used. Change Cal Method to MANUAL.
Only one point per range can be
calibrated. Re-consider your SINGLE or
MULTI Cal Method setting.
TDS Standard does
not match
Temperature Instrument reads
inaccurate
temperature
ANY “OR” or “UR” error
message
ANY I forgot my
password
Adjust TDS factor as needed to
correct value.
Calibrate using water bath or known
accurate thermometer. Faulty thermistor
– repair or replace.
Over Range or Under Range condition.
Check that electrode is connected.
Please send a written request with your
name, contact information, and
instrument serial number; a temporary
24
password will be provided.
ALL My problem is not
listed here.
Please contact your Fisher Scientific
representative or email technical
support directly with your feedback:
accumet@fishersci.com
18.USB Driver & Boot Loader Documentation
This document is intended to guide the reader through the process of A) installing
the USB driver for communication with a PC, and B) boot loading (upgrading) the
latest firmware for Fisher Scientific accumet
meters which may become available.
USB Driver Installation Guide For Windows XP:
To install the USB device driver under Windows XP, follow the instructions below.
This will allow your computer to recognize the instrument.
1. Obtain the USB driver file (via web download, disc, email, etc.) and save
on your computer in a location that can be easily found.
2. Connect the meter to a spare USB port on your PC using the USB cable
provided with the instrument. Once the USB is connected, Windows
Found New Wizard will launch as shown below:
®
AB150, AB200, and AB250 series
25
3. Select “Install from a list or specific location (Advanced)” as shown in
figure above and then click “Next”.
4. Select “Search for the best driver in these locations” and enter the file
path in the combo-box (“C:\Accumet_AB_USB_Driver” as in figure below)
or browse to it by clicking the browse button. Once the file path has been
entered in the box, Click “Next” to proceed.
26
5. If Windows XP is configured to warn when unsigned (non-WQHL
certified) drivers are about to be installed, the message dialogue
shown in figure below will be displayed unless installing a Microsoft
WHQL certified driver. Click on “Continue Anyway” to continue with
the installation.
27
6. The screen shown in the figure below will be displayed as Windows XP
copies the required driver files.
7. Once the installation is successful as shown in the figure below.
Click “Finish” to complete.
28
USB Driver Installation Guide For Windows 7:
To install the USB device driver under Windows 7, follow the instructions below.
This will allow your computer to recognize the instrument.
1. Obtain the USB driver file (via web download, disc, email, etc.) and save
on your computer in a location that can be easily found.
2. Connect the meter to a spare USB port on your PC using the USB cable
provided with the instrument.
3. Open the Device Manager:
In the Device Manager window there will be a device under Other
Devices with a yellow warning symbol to indicate a problem i.e. no driver
installed with text “USB VCOM Port”.
29
4. Right click on the other devices (USB VCOM Port) to bring up a menu as
shown below. From the Displayed menu, select “Update Driver
Software…”.
5. Select “Browse my computer for driver software” and enter the file
path in the combo-box or browse to it by clicking the browse button.
30
6. Once the file path has been entered in the box, check “Include
subfolders” box and click “Next” to proceed.
7. If necessary, click on “Install this driver software anyway” to continue
with the installation if you receive Windows 7 security pop up window.
31
The following window appears when installation is in progress.
When the installation has finished completion screen is displayed as
shown below.
32
8. Click “Close” and go back to the Device Manager Window.
The below figure shows the successful installation of USB Driver for
Accumet AB meter.
Note: If the USB cable is removed and connected to a new (different) USB
port, please repeat the driver installation procedure for the new USB port.
33
Boot Loading Procedure:
The software version of your instrument is displayed on the screen during power
on. Check www.fishersci.com/accumet
new software is available. As new releases become available, the boot loading
procedure described here is the process by which you can upgrade your Fisher
Scientific accumet®AB150, AB200, and AB250 series meter to have the latest
firmware. This procedure applies to both Windows XP and Windows 7. Please
read the entire procedure before starting.
1. Obtain the latest firmware via web download, disc, email, etc. and save
on your computer in a location that can be easily found or retrieved.
2. Remove the power cord from the instrument.
3. Plug in the USB cable.
4. Press and hold the left soft key (see image below) as you plug in the
power cord.
or email accumet@fishersci.com to see if
34
5. The instrument will beep twice and a pop-up window will appear showing
contents of a new drive named “ACCUMET AB” on your computer
similar to the one seen below. This folder represents the hard drive
of the instrument. The left soft key can be released as soon as this
window is visible.
35
6. Similar to the method for replacing a file that resided on a thumb drive,
you will need to delete the firmware.bin
window and replace it with the new file. Highlight “firmware.bin” and
delete this file.
file as seen in the pop-up
7. Locate the new firmware file “Accumet_ABSeries.bin
paste or drag & drop it into the pop-up window.
36
” and copy &
8. Once the file is successfully transferred, close the window.
9. Remove the power cord and restart the meter as you would normally.
The new software version will be briefly visible during power on.
19. Specifications
We reserve the right to make changes, improvements and modifications to the
specifications listed here. While specification changes are quite rare, software
versions are often updated to improve the customer experience and instrument
operation. The software version is displayed during power on.
pH AB150, AB200, AB250
Range -2.000 to 20.000 pH
Resolution 0.1 / 0.01 / 0.001 pH
Accuracy ±0.002 pH + 1 LSD
Cal. Points Up to 6 preset or 5 custom
Fisher Scientific accuflex® electrode support arm and bracket 13-637-671
Replacement power supply, 100/240 V 13-636-104
Compact printer, 100-240 V. Includes 1 roll of paper 13-637-690
Replacement paper for 13-637-690 printer, pack of 2 rolls 13-637-691
41
21. Warranty
The Fisher Scientific Company (“Fisher”) warrants to the direct purchaser that the
accumet meters and accumet, accuTupH, and accupHast, electrodes will be free
from defects in material or workmanship for a specified warranty period. During
that period, Fisher will repair or replace the product or provide credit, at its sole
option, upon prompt notification and compliance with its instructions. For accumet
meter, that specified period is 36 months from delivery date. For electrodes, that
specified period is 12 months.
Unless otherwise agreed, the warranty is limited to the country in which the
product is sold.
No Fisher employee, agent or representative has the authority to bind Fisher to
any oral representation or warranty concerning any product sold. Any oral
representation or warranty made prior to purchase of any product and not set forth
in writing and signed by a duly authorized officer of Fisher shall not be enforceable
by the purchaser.
FISHER EXPRESSLY DISCLAIMS ALL OTHER WARRANTIES, EXPRESS OR
IMPLIED WARRANTY OF MERCHANTABILITY AND FITNESS FOR A
PARTICULAR PURPOSE.
Fisher’s sole responsibility and the purchaser’s exclusive remedy for any claim
arising out of the purchase or any product listed above is repair, replacement or
credit as described above, where applicable. In no event: 1) shall the cost of the
exclusive remedy exceed the purchase price: 2) shall Fisher be liable for any
special, indirect, incidental, consequential, or exemplary damages, howsoever
arising, even if Fisher has been advised of the possibility of such damages.
Each article that Fisher furnishes will conform to the written specifications given
in this manual, or those of a further improved model. Changes are made often to
the information in the manual and will be incorporated into future edition.
22. Return of Items
A “Return Goods Authorization” (RGA) must be obtained from our Customer
Service Department or before returning items for any reason. Please include data
regarding the reason the items are to be returned. For your protection, items must
be carefully packed to prevent damage in shipment and insured against possible
damage or loss. We will not be responsible for damage resulting from careless or
insufficient packing. A restocking charge will apply to all unauthorized returns.
42
23. Notice of Compliance
This equipment generates, uses, and can radiate radio frequency energy and if
not installed and used in accordance with the instruction manual, may cause
interference to radio communications. It has been tested and found to comply
with the limits for a Class A computing device pursuant to Subpart J of Part 15
of FCC Rules, which are designed to provide reasonable protection against
such interference when operated in a commercial environment. Operation of
this equipment in a residential area is likely to cause interference in which case
the user, at his own expense, will be required to take whatever measures may
be required to correct the interference.
“This digital apparatus does not exceed the (Class A) limits for radio noise
emissions from digital apparatus set out in the Radio Interference Regulations
of the Canadian Department of Communications.”
“Le present appareil numerique n’ emet pas de bruits radioelectriques
depassant les limites applicables aux appareils numeriques (de la class A)
prescrites dans le Reglement sur le brouillage radioelectrique edicte par le
ministere des Communications du Canada.”
WEEE Compliance
This product is required to comply with the European Union’s Waste
Electrical & Electronic Equipment (WEEE) Directive 2002/96/EC. It is
marked with the symbol on the left.
Thermo Fisher Scientific has contracted with one or more recycling/disposal
companies in each EU Member State and this product should be disposed of or
recycled through them. Further information on compliance with these
directives, the recyclers in your country, and product information that may
assist the detection of substances subject to the RoHS Directive are available
from www.thermofisher.com
.
43
24. Declaration of Conformity
Manufacturer: Thermo Fisher Scientific, Inc.
Address: Ayer Rajah Crescent
Blk 55 #04-16/24
Singapore 139949
Singapore
Hereby declares that the following products rated 100-240 VAC, 50/60 Hz, 2A:
Fisher Scientific accumet AB150
Fisher Scientific accumet AB200
Fisher Scientific accumet AB250
Equipment Class: Measurement, control and laboratory, EMC Class A
Conforms to the following directives and standards:
EN61326-1:2006 Electromagnetic Compatibility
(EMC Directive) Electrical equipment for
measurement, control and laboratory use
– EMC requirements