High-Level Disinfection Guide
This guide is intended for multi-patient use of a Fisher & Paykel Healthcare Mask in the hospital or other clinical setting where proper disinfection of the device must
occur between patient uses. If you use the mask as a single-patient user in the home, please refer to your mask Instructions For Use for cleaning instructions.
F&P Zest™ Q
F&P Lady Zest
Nasal Mask
(excluding Foam Cushion and Diffuser)
™
Zest
Nasal Mask
(excluding Foam Cushion)
™
Q
FlexiFit™ 405/406/407
Nasal Mask
(excluding Foam Cushion and Diffuser)
Aclaim®2
Nasal Mask
(excluding Foam Cushion and Diffuser)
™
Forma
Full Face Mask
(excluding Foam Cushion)
Thermal
CIDEX® OPA Autoclave Sterrad®
Disinfection
Soak parts in
hot-water bath.
10 minutes at 80 °C (176 °F)
30 minutes at 70 °C (158 °F)Soak for 30 minutes
30 minutes at 70 °C (158 °F)Soak for 12 minutes
30 minutes at 70 °C (158 °F)Soak for 30 minutes
30 minutes at 70 °C (158 °F)Soak for 30 minutes
Follow manufacturer’s
instructions for high-level
disinfection. Immerse at 20 °C
(68 °F). Rinse thoroughly in
potable water.
Gravity displacement-type
autoclave at 121 °C (250 °F).
Enclose components in a
sterilization peel pouch.
See overleaf for more information.
15-minute exposure
30-minute exposure. Note:
Do not
autoclave the FlexiTube
manufacturer’s instructions for
100S & 100NX
Dry thoroughly before
disinfecting. Follow
®
100S or 100NX
the Sterrad
Sterilization System.
100S
100S
100S
100S
100S & 100NX
Sporox® II
Follow manufacturer’s
instructions for high-level
disinfection. Immerse at 20 °C
(68 °F). Rinse thoroughly in
potable water.
Soak for 6 hours
Soak for 6 hours
FlexiFit 432
Full Face Mask
(excluding Foam Cushion)
FlexiFit 431
Full Face Mask
360
™
Opus
Nasal Pillows Mask
(excluding Contoured Mask Frame)
™
Oracle
Oral Mask
General Warnings and Cautions
• Do not iron the headgear as the material is heat sensitive and will be
damaged.
• Do not clean the mask with products containing alcohol, anti-bacterial
agents, antiseptic, bleach, chlorine or moisturizer.
• Do not dry or store the mask in direct sunlight.
• Do not wash the mask in a dishwasher.
• The above actions may deteriorate or damage the mask and shorten its life.
If your Fisher & Paykel Healthcare Mask weakens or cracks, discontinue use
and seek replacement immediately.
Disclaimer: It is the user’s responsibility to qualify any deviations from the
recommended methods of reprocessing.
Contact your local homecare provider for additional information.
30 minutes at 75 °C (167 °F)Soak for 60 minutes
30 minutes at 75 °C (167 °F)
30 minutes at 75 °C (167 °F)Soak for 30 minutes
10 minutes at 80 °C (176 °F)Soak for 30 minutes
Soak for 3 hours
30-minute exposure
Cleaning the headgear
The headgear does not require disinfection. Thoroughly washing the headgear
between patients is acceptable.
1. Hand-wash the headgear in soap. Do not soak for longer than 10 minutes.
2. Rinse well in fresh water and allow it to air-dry out of direct sunlight.
Components that cannot be disinfected
The following components cannot be disinfected and should be replaced
between patients:
• Foam Cushion
• Diffusers
• Contoured Mask Frame
30-minute exposure
100S
100S
100S
Soak for 30 minutes
www.fphcare.com

Validated High-Level Disinfection Procedures
Your Fisher & Paykel Healthcare Mask is provided patient-ready. To reuse between patients please select one of the procedures below.
All masks have been validated for 20 cycles.
PROCEDURES
Thermal
Disinfection
®
CIDEX
at 20 °C (68 °F)
OPA
Autoclave
Gravity displacement-type
autoclave
at 121 ° C (250 °F)
Sterrad® 100S &
100NX
Sporox® II
at 20° C (68 °F)
Disassembly
Cleaning and
Drying
High-Level
Disinfection and
Drying
Inspection
Disassemble the mask according to the instructions shown in the relevant mask’s Instructions For Use.
Clean all surfaces of the mask and components with a soft-bristled brush and soap. Ensure all visible soil is removed.
Rinse thoroughly under running potable water to remove soap.
Allow the mask components to thoroughly air-dry out of direct sunlight before processing with High-Level Disinfection.
FAILURE TO PROPERLY CLEAN THE MASK MAY RESULT IN INADEQUATE DISINFECTION.
1. Using a certified Hot Water
Disinfection System, soak
those components of the
mask that can be disinfected.
See overleaf for appropriate
mask validated temperatures
and times.
2. On completion, remove the
mask components from
the Hot Water Disinfection
System.
3. Allow the mask components
to air-dry out of direct
sunlight.
4. Store in clean conditions.
1. Immerse those components
of the mask that can be
disinfected in the CIDEX
Solution according to the
manufacturer’s instructions.
See overleaf for appropriate
mask validated times.
2. Rinse the mask components
thoroughly in potable water.
3. Allow the mask components
to air-dry out of direct
sunlight.
4. Store in clean conditions.
®
OPA
1. Enclose those components
of the mask that can be
disinfected in a sterilization
peel pouch as described in the
instructions for the autoclave.
2. Place in the autoclave for the
required time.
See overleaf for appropriate
mask validated times.
3. Remove the components from
the autoclave.
4. Store in clean conditions.
1. Package the mask
components prior to
sterilization as described
in the manufacturer’s
instructions for the Sterrad
100S or 100NX Sterilization
System.
2. Disinfect the mask following
the manufacturer’s
instructions for the Sterrad
100S or 100NX Sterilization
System.
See overleaf for appropriate
mask Sterrad
System.
3. Remove the components from
the Sterrad
Sterilization System.
4. Store in clean conditions.
®
Sterilization
®
100S or 100NX
®
®
1. Immerse those components
of the mask that can be
disinfected in the Sporox
II Solution according to the
manufacturer’s instructions.
See overleaf for appropriate
mask validated times.
2. Rinse the mask components
thoroughly in potable water.
3. Allow the mask components
to air-dry out of direct
sunlight.
4. Store in clean conditions.
Before each use: Inspect the mask for deterioration. If any visible deterioration of a mask component is apparent, the mask component should be
discarded and replaced. Some discoloration may occur after reprocessing and is acceptable.
®
Reassembly
Packaging and
Storage
Reassemble the mask according to the instructions shown in the relevant mask Instructions For Use.
Do not store the mask in direct sunlight. Store in clean conditions.
Storage Temperature: -20 °C to 60 °C (-4 °F to 140 °F).
CIDEX and STERRAD are trademarks of Johnson & Johnson.
REF 185045078 REV K 2014-04 © 2014 Fisher & Paykel Healthcare Limited