HC360 HC360
myAIRVO™ 2 Quick Start Guide myAIRVO™ 2 Quick Start Guide
Consumable parts
Please refer to Section 4, “Cleaning and maintenance”, of your myAIRVO 2 User Manual for information
on consumable parts. myAIRVO 2 has four parts which you will need to replace regularly to keep your
device as clean as possible. The times recommended below assume your myAIRVO 2 is only being used
by one person, and rely on the correct cleaning procedures being followed.
Optifl ow™+ nasal cannula
Replace every 30 days.
Product code: OPT942E (Small)
OPT944E (Medium)
OPT946E (Large)
Reorder code:
Heated breathing tube
Replace every 60 days.
Product code: 900PT560E (top right)
900PT500E (bottom right)
Reorder code:
Air fi lter
Replace every 90 days or 1000 hours.
Product code: 900PT913 (2-pack)
Reorder code:
Important note: This is a Quick Start Guide only and should not be considered in any way a
replacement for your myAIRVO 2 User Manual. Before using your myAIRVO 2 device you must read
your myAIRVO 2 User Manual to understand the device’s intended use, warnings and cautions.
myAIRVO 2 is not intended for life support.
Setup steps for your myAIRVO 2 with the HC360 reusable water chamber
Please refer to Section 2, “Setting up myAIRVO 2”, and Section 3, “Using myAIRVO 2”, of your
myAIRVO 2 User Manual for information on setup.
1 2
3 4
Place your myAIRVO 2 on
a low shelf or near the fl oor
beside your bed. Make sure
it is placed below your head
height and fl at.
Fit the water chamber onto
your device by depressing
the fi nger guard, pushing
forward until the fi nger
guard clicks into place.
With the aid of the supplied
funnel, fi ll the chamber
with enough distilled water
for the period of use, but
never above the 560 mL
maximum water level line.
Attach the heated breathing
tube to your myAIRVO 2.
HC360 reusable water chamber
N/A
Product code: HC360
Reorder code:
Some products may not be available in your country. Please contact your local Fisher and Paykel Healthcare
representative.
Contact details for support and ordering
www.fphcare.com/myairvo
REF 614104 REV A November 2017 © Fisher & Paykel Healthcare Limited
5 6
7 8
Plug myAIRVO 2’s power
cord into the mains/utility
power socket and into the
rear of the unit.
When myAIRVO 2 is ready
for you to use, a tick symbol
will be displayed.
Switch on myAIRVO 2
by pressing and holding
the On/Off button for 5
seconds.
myAIRVO 2 will begin
to warm up, displaying a
warm-up symbol on the
screen.
Connect your interface
and begin therapy. During
use, if excess condensate
forms in the tube, refer to
Condensate Management
in Section 2, “Setting up
myAIRVO 2”.
myAIRVO™ 2 Quick Start Guide
HC360
Understanding myAIRVO 2 buttons
Please refer to Section 3, “Using myAIRVO 2”, of your myAIRVO 2 User Manual for information on how to
operate your myAIRVO 2.
On/Off button
The On/Off button turns
myAIRVO 2 on when held for
5 seconds. After use, press
for 3 seconds to begin Drying
Mode. Your device will turn off
automatically after 99 minutes.
Mode button
The Mode button allows you to
cycle between modes:
• Temperature
• Flow Rate
• Day/Night
• Compliance
Audio Pause button
The Audio Pause button allows
you to temporarily silence audio
alarms.
Up and Down buttons
The Up and Down buttons allow
you to change settings within the
following modes:
• Temperature
• Flow Rate
• Day/Night
Understanding myAIRVO 2 modes and settings in default mode
Please refer to Section 3, “Using myAIRVO 2”, of your myAIRVO 2 User Manual for information on how to
operate your myAIRVO 2. Your myAIRVO 2 will remember your settings when switched off.
Temperature Mode
Use the Up and Down buttons to
cycle between temperatures of
31 °C, 34 °C and 37 °C.
Flow Rate Mode
Use the Up and Down buttons
to select your fl ow rate from
10-60 L/min. Please see your
myAIRVO 2 User Manual for
information about fl ow ranges for
your interface.
Cleaning myAIRVO 2
Please refer to Section 3, “Using myAIRVO 2”, and Section 4, “Cleaning and maintenance”, of your
myAIRVO 2 User Manual for information on how to clean myAIRVO 2.
Daily cleaning
1
After use, remove your interface. To begin
the 99-minute Drying Mode, press and hold
the On/Off button for 3 seconds. To avoid
burns, do not wear the interface during
Drying Mode and do not remove the water
chamber until Drying Mode is complete.
Weekly cleaning
1
2
Rinse the interface in drinking-quality water
then reconnect to the heated breathing
tube whilst still in Drying Mode to dry the
interface.
2 3
3
After Drying Mode is complete, remove the
water chamber by pushing down the fi nger
guard and pulling out the chamber. Wash
and rinse the chamber and then refi ll as per
set up steps on previous page.
Day/Night Mode
Use the Up and Down buttons
to cycle between Day Mode and
Night Mode. In Night Mode some
audio signals will become quieter
and your display screen will dim.
Compliance Mode
Compliance Mode automatically
cycles between device usage
statistics:
• Total hours used
• Hours per day
• Checksum (for your clinician)
myAIRVO 2 visual and auditory alarms
Please refer to Section 3, “Using myAIRVO 2”, of your myAIRVO 2 User Manual for information on alarms.
Your myAIRVO 2 has alarms to warn you about interruptions to your treatment. You can use the Audio
Pause button to silence the alarm for 115 seconds.
Audio Pause button
Switch off the unit and unplug from the
mains/utility power socket.
4
Remove water chamber. Pour out and discard the remaining
water. Remove chamber base. Wash the chamber top and base
in mild dishwashing detergent then rinse. Soak the chamber in
a solution of vinegar (1 part) and water (2 parts) for 10 minutes.
Rinse and dry.
Remove the heated breathing tube and
drain any excess condensate.
5
Thoroughly wipe the inside of the heated breathing tube
connection port with a clean, low-lint cloth dipped in warm
water with mild dishwashing detergent added. Wipe the exterior
of the unit with a clean damp (not wet) cloth dipped in warm
water with mild dishwashing detergent added.
Remove the interface from the heated
breathing tube, wash it in warm water
with mild dishwashing detergent added,
rinse it in drinking-quality water, then
reconnect it to the heated breathing tube.
When cleaning is complete, set up your myAIRVO 2 using the setup steps previously described.