MANUFACTURER: THERMO FISHER SCIENTIFIC
Manufacturer's PHONE: (508) 482-7000 FAX: (508) 634-2110
AAC 'Y' ITEM NSN 6640-01-348-2460
For use in preparing large or small, thin, unwrinkled sections of single or multiple pieces of
fresh frozen tissue in military surgical pathology laboratories. The Cryostome FSE Cryostat has the microtome
located outside the cryo-chamber for quick temperature adjustment and safer, easier cryo-chamber decontamination
and cleaning. A simple, intuitive multi-lingual touch screen display provides easy instrument setup, operation and
control of the cryo-chamber and microtome. Peace of mind operation is ensured with a dual compressor system, a
remote alarm to signal system failure, and multi-level security code protection to prevent unauthorized access to
protocols. The 6 position Cryobar has a dedicated refrigeration system, cover, rapid cooling boost to -55 Celcius,
and 3 heat sinks. The Specimen head has X- and Y- orientation with user defined temperature control. The 3 level
foot rest is adjustable for better ergonomics and comfort. A lockable heated window eliminates frost buildup. The
hand wheel may be locked to prevent injury when working around the blade. CFE-Free refrigeration and foam
insulation. An Automated Fumigation Cycle to disinfect the cryochamber and programmable auto-defrost cycle are
standard. "Quick heat" specimen release system is also included.
• Rotary
• Refrigerated
• Mounted on casters, two of which lock
• Has a compressor capable of maintaining selected knife temperature of 0 to -20°C 24 hours a
day for an indefinite period of time
• Automatic defrosting
• Quick freeze timer
• Interior cabinet temperature readout w/low of -28°C
• Automatic feed w/min range of 2-4 microns
USAMMA POC: Customer Relations Management 301-619-4301/4316 - DSN 343-4301/4316
Manufacturer Model# CRYOSTOME FSE
<http://www.thermo.com/com/cda/product/detail/0,1055,10121433,00.html>

MICROTOME CRYOSTAT
PART # CRYOSTOME FSE
NSN 6640013482460

PHYSICAL & ENVIRONMENTAL SPECIFICATIONS
-13°F to 131°F (-25°C to 55°C)
41°F to 104°F (5°C to 40°C)
80% max < 88°F (31°C) 50% max 88°F to 104°F (31°C to 40°C)
80% max < 88°F (31°C) 50% max 88°F to 104°F (31°C to 40°C)
115 VAC 60Hz, 6,147 BTU/Hr Max Current @ compressor start: 38A
Nominal Voltage @ max current: 103VAC Max Voltage @ no load: 126VAC
Chamber Access Width: 19.7 inches minimum
Microtome Section Thickness: 1-20 µm in 1 µm steps, 23-60 µm in 5 µm steps
Total Horizontal Specimen Advance: 25 mm
Cutting Stroke: 57 mm
Trim Wheel: 8 µm indents
Motorized Coarse Advance: 1 mm/ sec
Specimen Retraction: Standard
Refrigeration-Cryochamber: 32°F to -31°F (0 to -35°C)
Refrigeration-Cryobar: 32°F to -67°F (0 to -55°C) or colder
Refrigeration-Specimen Head: 32°F to -40°F (0 to -40°C)
Agency Approval: Meets IEC61010, CSA, UL and CE Mark (IVDD compliant)

MICROTOME CRYOSTAT
WEB LINKS
ACCESSORIES & CONSUMABLES - STARTUP
MFR PN
Thermo
Fisher
Scientific
Knife Microtome Plain Wedge
185mm
<http://www.thermo.com/eThermo/CMA/PDFs/Product/productPDF_4865.pdf>
Click Below for Disposable Blade Specifications
<http://www.thermo.com/eThermo/CMA/PDFs/Product/productPDF_7345.pdf>
Click Below for Product Brochure