Fisher ControlWave MRMS-IC (D5140) Configuration Manual

s
Instruction Manual
Doc Number D5140
Part Number D301719X012
May 2012
ControlWave® MRMS-IC Configuration Manual
Remote Automa ti on Solution
www.EmersonProcess.com/Remote
This page is intentionally left blank
ControlWave MRMS-IC Configuration Manual (D5140)
Contents
Chapter 1 – Getting Started 1-1
1.1What is MRMS-IC? ................................................................................................................... 1-1
1.2Before You Begin ...................................................................................................................... 1-1
1.3Installing MRMS-IC Software .................................................................................................... 1-2
1.4Starting MRMS-IC Software ...................................................................................................... 1-6
1.4.1MRMS_IC IP Startup ..................................................................................................... 1-6
1.4.2MRMS_IC Serial Startup ............................................................................................... 1-7
1.4.3Logging Onto the ControlWave Micro ........................................................................... 1-7
1.5Accessing Pages of the MRMS_IC Application ........................................................................ 1-8
1.6Entering Data in Fields of the MRMS_IC Application ............................................................... 1-9
Chapter 2 – Configuring Inputs and Outputs (I/O Tab) 2-1
2.1I/O Tab ...................................................................................................................................... 2-1
2.2I/O Usage .................................................................................................................................. 2-2
2.2.1Discrete Inputs (DI) ....................................................................................................... 2-5
2.2.2Discrete Outputs (DO) ................................................................................................... 2-5
2.2.3Analog Inputs (AI) ......................................................................................................... 2-6
2.2.4Analog Outputs (AO) ..................................................................................................... 2-7
2.2.5High Speed Counters (HSC) ......................................................................................... 2-8
2.2.6Ultrasonic Flow Meters (UFM) ...................................................................................... 2-9
2.2.7Multi-variable Transmitters (Transducers) .................................................................. 2-10
2.3Local DLM ............................................................................................................................... 2-12
2.4Customer Modbus Slave ......................................................................................................... 2-14
2.4.1Signal List Grid ............................................................................................................ 2-17
2.4.2Floating Point Format .................................................................................................. 2-19
2.5Generic Modbus Master .......................................................................................................... 2-20
2.6Load/Save Configuration ......................................................................................................... 2-25
2.6.1Performing I/O Array Operations. ............................................................................... 2-26
2.6.2Creating a Recipe ....................................................................................................... 2-27
2.6.3Saving the Recipe ....................................................................................................... 2-28
2.6.4Recalling a Saved Recipe, and Sending Its Values to the Controller ......................... 2-29
2.7Time Set/Daylight Saving Time ............................................................................................... 2-29
Chapter 3 – Configuring Stations and Runs (Measurement Tab) 3-1
3.1Measurement Tab ..................................................................................................................... 3-2
3.2Status/Configuration .................................................................................................................. 3-3
3.2.1Site Configuration Data Tab (Site Configuration) ......................................................... 3-4
3.2.2MVT Common Settings Tab (Site Configuration).......................................................... 3-7
3.2.3Station Summaries Tab (Site Configuration) ................................................................ 3-8
3.2.4Station Configuration Tab (Station Configuration) ...................................................... 3-10
3.2.5Station Data Tab (Station Configuration) .................................................................... 3-15
3.2.6Run Configuration Tab ................................................................................................ 3-16
3.2.7Alarm Configuration Tab (Run Configuration) ............................................................. 3-21
3.2.8Linearization Config Tab (Run Configuration) ............................................................. 3-23
3.2.9PV/GQ Averages Tab (Run Configuration) ................................................................. 3-24
3.2.10Orifice Tab (Run Configuration) .................................................................................. 3-25
3.2.11Turbine Tab (Run Configuration) ................................................................................ 3-29
3.2.12Auto-Adjust Tab (Run Configuration) .......................................................................... 3-32
3.2.13Ultrasonic Tab (Run Configuration) ............................................................................ 3-34
3.2.14PD Tab (Run Configuration) ........................................................................................ 3-37
Issued: May-2012 Contents iii
ControlWave MRMS-IC Configuration Manual (D5140)
3.3View Local Archives ................................................................................................................ 3-39
3.3.1Selecting Logs to View ................................................................................................ 3-39
3.4Collect Local Logs ................................................................................................................... 3-40
3.4.1Selecting Archives for Collection ................................................................................ 3-40
3.4.2Collecting a Single Archive ......................................................................................... 3-40
3.4.3Collecting Multiple Archives ........................................................................................ 3-40
3.4.4Log Collection Parameters .......................................................................................... 3-40
3.5View Audit Log ........................................................................................................................ 3-43
3.5.1Data Storage Parameters dialog box .......................................................................... 3-44
3.5.2Search Data Collection Criteria dialog box ................................................................. 3-45
3.6Maintenance Mode .................................................................................................................. 3-46
3.6.1Site Tab ....................................................................................................................... 3-46
3.6.2Station Tab .................................................................................................................. 3-49
3.6.3Run Tab ...................................................................................................................... 3-51
3.6.4PVs Tab ...................................................................................................................... 3-53
3.6.5AI Maintenance Tab .................................................................................................... 3-54
3.7Gas Chromatograph Configuration ......................................................................................... 3-56
3.7.1General ....................................................................................................................... 3-56
3.7.2Current Tab (Gas Chromatograph Configuration) ...................................................... 3-61
3.7.3Component Tab (Gas Chromatograph Configuration) ................................................ 3-65
3.7.4Delta Limit Tab (Gas Chromatograph Configuration) ................................................. 3-67
3.7.5Normalization Tab (Gas Chromatograph Configuration) ............................................ 3-68
3.7.6Custom Tab (Gas Chromatograph Configuration) ...................................................... 3-69
3.8Summary Pages ...................................................................................................................... 3-70
3.8.1Measurement Tab ....................................................................................................... 3-70
3.8.2Alarm Tab .................................................................................................................... 3-71
3.9Limits Page (Gas Composition Allowable Ranges) ................................................................ 3-72
3.10Daily Run Corrected and Uncorrected Volume ....................................................................... 3-73
Chapter 4 – Sampler (Control Tab) 4-1
4.1Sampler ..................................................................................................................................... 4-1
4.1.1List 29 ............................................................................................................................ 4-4
Appendix E - Troubleshooting E-1
Index IND-1
iv Contents Issued: May-2012
ControlWave MRMS-IC Configuration Manual (D5140)
Chapter 1 – Getting Started
This chapter discusses how to install the MRMS-IC application and provides some general information about how to use it.
In This Chapter
1.1What is MRMS-IC? .......................................................................... 1-1
1.2Before You Begin ............................................................................. 1-1
1.3Installing MRMS-IC Software .......................................................... 1-2
1.4Starting MRMS-IC Software ............................................................ 1-6
1.4.1MRMS_IC IP Startup ............................................................ 1-6
1.4.2MRMS_IC Serial Startup ...................................................... 1-7
1.4.3Logging Onto the ControlWave Micro .................................. 1-7
1.5Accessing Pages of the MRMS_IC Application ............................... 1-8
1.6Entering Data in Fields of the MRMS_IC Application ...................... 1-9
1.1 What is MRMS-IC?
1.2 Before You Begin
Multi-Run Multi-Station (MRMS) software with Industry Canada (IC) / Measurement Canada approvals is a software application that allows the ControlWave Micro controller to manage up to six (6) natural gas measurement stations.
The MRMS-IC application consists of: A ControlWave project file (*.PRO) pre-programmed for multi-run
multi-station natural gas measurement.
A customized flash configuration profile (*.FCP) file that configures
the ports, audit, and archive parameters of the ControlWave Micro for the MRMS_IC.
A TechView session. This includes the TechView session file
(*.TVS), associated *.INI files, and a set of HTM menus customized for the MRMS-IC application. You use these menus to configure the application.
You must install the ControlWave Micro controller on site and
connect field devices to its I/O modules. For information on ControlWave Micro hardware, see document CI-ControlWave Micro.
You must install OpenBSI 5.7 software including TechView on your
PC workstation. You must also install the latest service pack and patches (Service Pack 2, Patch D). See the OpenBSI Utilities
Manual (D5081), the BSI_Config User’s Manual (D5128), and the TechView User Manual (D5131) for details on installation
requirements.
Issued: May-2012 Getting Started 1-1
ControlWave MRMS-IC Configuration Manual (D5140)
You must connect the PC workstation to the ControlWave Micro
controller. You can communicate using a serial cable or an Ethernet cable. Cable diagrams are included in CI-ControlWave Micro.
The ControlWave Micro must be running a flash configuration
profile file (*.FCP) compatible with MRMS-IC software. For information on updating FCP files, see Chapter 5 of the OpenBSI Utilities Manual (D5081).
The ControlWave Micro must be running the ControlWave project
(*.PRO) file configured for the MRMS-IC. See Chapter 7 of the OpenBSI Utilities Manual (D5081) for information on downloading a ControlWave project (*.PRO) file.
Note: If you ordered your ControlWave Micro with MRMS-IC
software pre-installed, the FCP and PRO files are already loaded when the unit ships from the factory.
1.3 Installing MRMS-IC Software
Note: MRMS_IC runs on the following Windows operating systems:
Windows XP
1. Double-click on the MRMS-IC application icon.
2. Click Next on the welcome screen of the installer.
Click “Next”
Figure 1-1. MRMS_IC Installer – Welcome Screen
1-2 Getting Started Issued: May-2012
ControlWave MRMS-IC Configuration Manual (D5140)
3. The next screen includes a “Read Me” file which describes the
changes since the last release of MRMS IC. Click Next after you review the changes.
Click I Agree
Figure 1-2. Read Me Page
4. Review the license agreement and click the I Agree button to
proceed with the installation or Cancel to abort the installation process.
Figure 1-3. License Agreement page
Issued: May-2012 Getting Started 1-3
ControlWave MRMS-IC Configuration Manual (D5140)
Figure 1-4. Installation in Progress
5. The installation begins. If you see the following message box,
click OK to confirm that you have the proper OpenBSI version installed.
Click OK
Figure 1-5. Confirm OpenBSI Version
6. At the completion of the installation, click Next.
1-4 Getting Started Issued: May-2012
Click Next
ControlWave MRMS-IC Configuration Manual (D5140)
Figure 1-6. Installation Completion
7. Now click Finish to exit the installer.
Click Finish
Figure 1-7. Exit the Installer
Issued: May-2012 Getting Started 1-5
ControlWave MRMS-IC Configuration Manual (D5140)
1.4 Starting MRMS-IC Software
You start the MRMS_IC software by invoking the proper TechView file. There are two ways to do this:
Starting MRMS_IC from
the Start Programs
Menu
Starting MRMS_IC from
an icon
1.4.1 MRMS_IC IP Startup
For an IP connection, click: Start > Programs > MRMS_IC > MRMS_IC_IP_tvs
For a serial connection, click: Start > Programs > MRMS_IC > MRMS_IC_Serial_tvs
From a desktop icon, similar to those below, or from the \MRMS_IC\SUPPORT folder, double-click the IP or serial TVS file, depending upon your type of connection.
Figure 1-8. MRMS_IC TVS file icons
For IP communication, see Section 1.4.1 MRMS_IC IP Startup. For serial communication see Section1.4.2 MRMS_IC Serial Startup.
Note: Although you can view data through an IP connection, the
MRMS-IC application only allows configuration changes when you establish a physical serial connection to serial communication port 1 on the ControlWave Micro.
Once you start the TVS file for IP operation, TechView opens the Runtime Configuration Parameters dialog box:
Figure 1-9. IP Runtime Parameters
1. Ignore the number of runs; this parameter does not apply for
1-6 Getting Started Issued: May-2012
MRMS-IC.
2. Enter the IP address of the ControlWave Micro IP port to
which you are connected.
3. Click OK.
4. Log onto the ControlWave Micro as described in Section 1.4.3.
1.4.2 MRMS_IC Serial Startup
Once you start the TVS file for serial operation, TechView opens the Runtime Configuration Parameters dialog box:
ControlWave MRMS-IC Configuration Manual (D5140)
Figure 1-10. Serial Runtime Parameters
1. Ignore the number of runs; this parameter does not apply to
MRMS-IC.
2. Enter the BSAP local address of the ControlWave Micro to
which you are connected.
3. Select the serial communication port on the PC which you are
using to communicate with the ControlWave Micro.
4. Select the baud rate on the serial communication line.
5. Click OK.
6. Log onto the ControlWave Micro as described in Section 1.4.3.
1.4.3 Logging Onto the ControlWave Micro
In the SignOn to RTU dialog box, enter a Username / Password combination that allows full access to the ControlWave Micro, then click the SignOn button.
Issued: May-2012 Getting Started 1-7
ControlWave MRMS-IC Configuration Manual (D5140)
y
Figure 1-11. Logging onto the ControlWave Micro
1.5 Accessing Pages of the MRMS_IC Application
To access various pages of the MRMS_IC application, click on the tab for the function you want to configure, then click on the buttons which appear on that tab. By default, the I/O tab appears first.
Click on the tab you want to access, then click on an
of the buttons for that tab
Tabs
Buttons
Figure 1-12. Tabs and Buttons in MRMS_IC
The next several chapters describe the functions available on each tab of the application.
1-8 Getting Started Issued: May-2012
ControlWave MRMS-IC Configuration Manual (D5140)
You need not configure all the features of the application; only those that you need for your particular purpose and measurement needs.
1.6 Entering Data in Fields of the MRMS_IC Application
Whenever you select a field and enter data, or select from a drop-down menu, you must press the [Enter] key to confirm and save your choice.
To exit a field without entering data, press the [Esc] key.
Issued: May-2012 Getting Started 1-9
This page is intentionally left blank
ControlWave MRMS-IC Configuration Manual (D5140)
Chapter 2 – Configuring Inputs and Outputs (I/O Tab)
This chapter discusses configuring the MRMS-IC application to accept field inputs and outputs (I/O). This is accomplished from the MRMS­IC’s I/O tab.
Note: Although you can view data through an IP connection, the
MRMS-IC application only allows configuration changes when you establish a physical serial connection to serial communication port 1 on the ControlWave Micro.
In This Chapter
2.1 I/O Tab ............................................................................................. 2-1
2.2 I/O Usage ......................................................................................... 2-2
2.2.1 Discrete Inputs (DI) .............................................................. 2-5
2.2.2 Discrete Outputs (DO) .......................................................... 2-5
2.2.3 Analog Inputs (AI) ................................................................. 2-6
2.2.4 Analog Outputs (AO) ............................................................ 2-7
2.2.5 High Speed Counters (HSC) ................................................ 2-8
2.2.6 Ultrasonic Flow Meters (UFM) ............................................. 2-9
2.2.7 Multi-variable Transmitters (Transducers) ......................... 2-10
2.3 Local DLM ...................................................................................... 2-12
2.4 Customer Modbus Slave ............................................................... 2-14
2.4.1 Signal List Grid ................................................................... 2-17
2.4.2 Floating Point Format ......................................................... 2-19
2.5 Ultrasonic Data .............................................................................. 2-20
2.6 Load/Save Configuration ............................................................... 2-25
2.6.1 Performing Array Read or Write Operations. ..................... 2-26
2.6.2 Creating a Recipe ............................................................... 2-27
2.6.3 Changing Floating Point Format of Values in the Recipe .. 2-28
2.6.4 Saving the Recipe .............................................................. 2-28
2.6.5 Recalling a Saved Recipe, and Sending Its Values to the
Controller .......................................................................... 2-29
2.7 Generic Modbus Master ................................................................ 2-29
2.8 Time Set/Daylight Saving Time ..................................................... 2-29
2.1 I/O Tab
Click the I/O tab to display the various I/O options you can configure. We’ll discuss each of these in the sections that follow.
To configure an I/O function or feature, click on its button.
Figure 2-1. I/O Tab in MRMS-IC
Issued: May-2012 Configuring Inputs and Outputs (I/O Tab) 2-1
ControlWave MRMS-IC Configuration Manual (D5140)
2.2 I/O Usage
When you click the button on the I/O tab, the I/O Usage page displays a graphical representation of the ControlWave Micro, showing each of the I/O modules detected by the MRMS-IC. If MRMS-IC cannot detect a particular module or an I/O slot is empty, its graphic shows “Not Present.” MRMS-IC only uses Mixed I/O modules.
Click on the graphical representation of a module to configure its I/O.
I/O modules
Figure 2-2. I/O Usage Screen Showing I/O Modules Detected
When you move the cursor over the CPU module, an Expansion Communication (ECOM) module, or any I/O module, you’ll see a yellow box on the screen. To configure I/O, follow these steps:
1. From the I/O tab, click the I/O Usage button.
2. Position the cursor over the I/O module you want to configure; a
yellow box indicates the cursor position on any configurable module.
3. Click on the module you want to configure. This opens a screen
showing the possible choices for I/O. The Mixed I/O Module shows multiple types of I/O (see Figure 2-3).
2-2 Configuring Inputs and Outputs (I/O Tab) Issued: May-2012
ControlWave MRMS-IC Configuration Manual (D5140)
Figure 2-3. Mixed I/O Module
4. The I/O assignments in the MRMS-IC application are fixed
based on the I/O slot. Because you can have an expanded communication module (ECOM) in either I/O slot 1 or 2, I/O designations begin with the right-most slot and go in reverse order. In other words, the mixed I/O module in I/O slot 6 holds user DI1, whereas the mixed I/O module in I/O slot 1 holds user DI6.
5. Configuration is limited based on the I/O type. Refer to the sub-
sections that follow.
Note: You may have noticed that when the cursor is left hovered over
an IO point, the graphics to the left display the applicable connection points for direct and remote IO.
Issued: May-2012 Configuring Inputs and Outputs (I/O Tab) 2-3
ControlWave MRMS-IC Configuration Manual (D5140)
Graphic displays location of physical I/O connections for an I/O point, when you hover the cursor over the PNT.
Figure 2-4. Connection Points for Physical I/O
Notes:
The MRMS-IC application only uses Mixed I/O modules. Depending upon your particular configuration, you might not use all
the inputs or outputs in a particular meter run or station.
If you have I/O that comes from an ultrasonic flow meter or a multi-
variable transmitter that communicates with the MRMS-IC through a communication port, instead of an I/O module, you configure it from the UFM or Transducer pages, discussed later in this chapter.
To return to the I/O Usage page from any page underneath it, click
the Go Back link.
2-4 Configuring Inputs and Outputs (I/O Tab) Issued: May-2012
2.2.1 Discrete Inputs (DI)
Discrete inputs (DIs) include the following fields:
Field Description PNT
INV If you click this box for a given I/O point so that “Y” is
Live
ControlWave MRMS-IC Configuration Manual (D5140)
This read-only field displays the I/O point number.
displayed, MRMS-IC inverts the real-live field value and uses the inverted value as the process value. For example, if the Live value of discrete I/O point 5 is OFF, and INV is selected for that point, PV is set ON and that’s what MRMS-IC uses for control and processing.
This read-only field shows the actual ON/OFF status of this discrete input point.
PV
Assignment
2.2.2 Discrete Outputs (DO)
Discrete outputs (DOs) include the following fields:
Field Description PNT
Points that are ON show in red. Points that are OFF show in green.
This read-only field shows the value of the process variable (PV) used in MRMS-IC. This matches the Live value unless you invert the input using INV.
Points that are ON show in red. Points that are OFF show in green.
This read-only field shows details of the fixed I/O assigned to this point.
This read-only field displays the I/O point number.
PV
Assignment
This read-only field shows the value of the process variable (PV) MRMS-IC will output to the field device.
This read-only field shows details of the fixed I/O assigned to this point.
Issued: May-2012 Configuring Inputs and Outputs (I/O Tab) 2-5
ControlWave MRMS-IC Configuration Manual (D5140)
A
2.2.3 Analog Inputs (AI)
Analog inputs (AIs) include the following fields:
Field Description PNT
This read-only field displays the I/O point number.
PV
Zero
Span Enter the value that, when added to the Zero value,
Units
Assignment
Use Stacked DP Transmitter / Use Single DP Transmitter Only button
This read-only field shows the calculated value of the analog input process variable (PV) based on the configured Zero and Span.
If the value shows in red, the value is questionable
connection, a communication problem with the field device, data timeout or some other problem that could cause the value to be invalid.
Enter the value that the process variable should read when the AI field input is 4mA. Press [Enter] to save your selection.
represents what the process variable should display when the AI field input is 20m your selection.
For example, if Zero is 5 and Span is 20, then:
If the AI field input is: 4mA 5 20mA 25 12mA 15
The engineering units for this process variable. Click in the field and select the proper units from the drop-down menu. Press [Enter] to save your selection.
This read-only field shows details of the fixed I/O assigned to this point.
This button toggles I/O assignments based on whether you use a single DP transmitter, or stacked DP transmitters.
To use stacked, click the Use Stacked DP Transmitter and assignments change to stacked; the label on the button then changes to Use Single DP Transmitter Only.
To use a single DP transmitter, click the Use Single DP Transmitter and assignments change to single; the label on the button then changes to Use Stacked DP Transmitter.
. This could indicate no
. Press [Enter] to save
PV will be:
2-6 Configuring Inputs and Outputs (I/O Tab) Issued: May-2012
2.2.4 Analog Outputs (AO)
Analog outputs (AOs) include the following fields:
Field Description PNT
ControlWave MRMS-IC Configuration Manual (D5140)
This read-only field displays the I/O point number.
PV
Zero
Span Enter the value that, when added to the Zero value,
Units
Assignment
This read-only field shows the calculated value of the analog output process variable (PV) based on the configured Zero and Span. This value will be sent to the field device.
Enter the value that the process variable should read when the AO field output is 4mA. Press [Enter] to save your selection.
represents what the process variable should display when the AO field output is 20mA. Press [Enter] to save your selection.
For example, if Zero is 5 and Span is 20, then:
If PV is: 5 4mA 25 20mA 10 8mA
The engineering units for this process variable. Click in the field and select the proper units from the drop-down menu. Press [Enter] to save your selection.
This read-only field shows details of the fixed I/O assigned to this point.
The AO field output is:
Issued: May-2012 Configuring Inputs and Outputs (I/O Tab) 2-7
ControlWave MRMS-IC Configuration Manual (D5140)
2.2.5 High Speed Counters (HSC)
High speed counters (HSC) include the following fields:
Field Description PNT
This read-only field displays the I/O point number. The number varies depending upon the type of I/O module.
Counts
Time Stamp
Assignment
Use AutoAdjust / Use Single HSC Input button
This read-only field displays the number of counts since the last power cycle.
This read-only field displays the timestamp of the last sample from the HSC module. The timestamp is the number of milliseconds since boot.
This read-only field shows details of the fixed I/O assigned to this point.
This button toggles I/O assignments based on whether you use a single HSC input, or you use Auto Adjust which requires two HSC inputs.
To use auto adjust, click the Use AutoAdjust and assignments change to show two HSC inputs; the label on the button then changes to Use Single HSC Input.
To use a single HSC input, click the Use Single HSC Input and assignments change to a single HSC; the label on the button then changes to Use AutoAdjust.
2-8 Configuring Inputs and Outputs (I/O Tab) Issued: May-2012
ControlWave MRMS-IC Configuration Manual (D5140)
2.2.6 Ultrasonic Flow Meters (UFM)
If you have one or more ultrasonic flow meters, click the UFMs link on the top of the I/O Usage page to call up the UFM page.
Figure 2-5. Ultrasonic Flow Meter (UFM) Configuration
Configure the following fields for your ultrasonic flow meter.
Field Description Enabled/Disabled
Port
Address Enter the address of the UFM here. Press [Enter] to
Type
Issued: May-2012 Configuring Inputs and Outputs (I/O Tab) 2-9
Click this button to enable communication from this UFM to the MRMS-IC.
Use the dropdown menu to specify the ControlWave Micro serial communication port which connects to this UFM. Press [Enter] to save the selection.
save the selection.
Use the drop-down menu to select the type of UFM. Press [Enter] to save your selection.
ControlWave MRMS-IC Configuration Manual (D5140)
2.2.7 Multi-variable Transmitters (Transducers)
If you have one or more multi-variable transmitters, click the Transducers link on the top of the I/O Usage page to call up the Transducers page.
The Transducers page shows the first three multi-variable transmitters (MVTs). If you want to view a different group of three MVTs, click the button corresponding to the range of MVTs (1-3 or 4-6) on the top of the screen.
Figure 2-6. Transducers Page (Multi-Variable Transmitters)
Each MVT includes the following fields:
Field Description Enabled/Disabled
Communication Protocol (BSAP/MODBUS)
Port
2-10 Configuring Inputs and Outputs (I/O Tab) Issued: May-2012
Click this button to enable communication from this MVT to the MRMS-IC.
Click the BSAP/MODBUS button to toggle the method used to communicate with this MVT between BSAP protocol and MODBUS protocol.
Use the dropdown menu to specify the ControlWave Micro serial communication port which connects to this MVT. Press [Enter] to save the selection.
ControlWave MRMS-IC Configuration Manual (D5140)
Address Enter the address of the MVT here. Press [Enter] to
save the selection.
Xmtr Type
Register Set
Comm Stats
% Good
Good Polls
Bad Polls
Reset
Current Status
Tag Name
Serial Number
Time Stamp
DP
SP
Use the drop-down menu to select the type of data coming from this MVT. Choose either:
: Data from this type:
Type GP/T gage pressure and temperature DP/P/T differential pressure, static pressure, and
temperature
T temperature
Press [Enter] to save your selection.
This field applies only to MODBUS communication. Click either 7xxx or 4xxxx to select the MODBUS register set used by this MVT.
This read-only field shows the percentage of successful communication transactions with this MVT.
This read-only field shows the number of good poll messages in communications with this MVT.
This read-only field shows the number of bad poll messages in communications with this MVT.
This button resets the communication statistics in the %Good, Good Polls, and Bad Polls fields.
These read-only fields display the most recent status messages from this MVT.
This read-only field shows the tag name from this MVT. (BSAP only)
This read-only field shows the serial number from this MVT. (MODBUS only)
This read-only field shows the time stamp of the most recent value received from this MVT.
This read-only field shows the most recent differential pressure reading from this MVT.
This read-only field shows the most recent static
Issued: May-2012 Configuring Inputs and Outputs (I/O Tab) 2-11
ControlWave MRMS-IC Configuration Manual (D5140)
A
FT
Units
Zero
Span
2.3 Local DLM
Notes:
The local Data Line Monitor (DLM) provides details about low-
level communication messages sent through a selected serial port used by the MRMS-IC.
pressure reading from this MVT.
This read-only field shows the most recent temperature reading from this MVT.
This read-only field shows the engineering units for this variable.
This read-only field shows the value for this variable when the MVT receives a 4mA field input.
This read-only field shows the value that, when added to the Zero value, represents what the process variable should display when the field input to the MVT is 20mA.
Typically, you would only use the local DLM if you are a very
advanced user and need to perform communication troubleshooting for a particular port.
The local DLM only displays the first 80 characters of a message. The local DLM only captures messages approximately every half
second, therefore, it can miss some messages.
Click the button on the I/O tab to activate the Data Line Monitor function. The DLM includes the following fields:
Field Description Monitor Port
TX Data
Use the dropdown menu to select the ControlWave Micro serial communication port you want the DLM to monitor. Press [Enter] to save the selection.
Note:
the monitor port, you can copy data from the window to the clipboard. You can then paste this data into another file for off-line review.
This read-only field shows the most recent message transmitted through this port.
fter you collect the data, if you select “None” for
RX Data
window The window shows successive messages detected by
2-12 Configuring Inputs and Outputs (I/O Tab) Issued: May-2012
This read-only field shows the most recent message received through this port.
the DLM. Most recent messages appear at the top; you can use the scroll bar to adjust the window to show earlier messages.
ControlWave MRMS-IC Configuration Manual (D5140)
Figure 2-7. Local DLM
Issued: May-2012 Configuring Inputs and Outputs (I/O Tab) 2-13
ControlWave MRMS-IC Configuration Manual (D5140)
2.4 Customer Modbus Slave
MRMS-IC supports a single Modbus slave session you can configure to provide a Modbus slave interface to the controller.
Click the button on the I/O tab to bring up the Customer Modbus Slave page.
You can view either coils or registers in the Signal List grid.
Figure 2-8. Customer Slave Page
This page includes the following fields:
Field Description
Settings Communications Port
Serial Click the Serial selection to use serial Modbus
2-14 Configuring Inputs and Outputs (I/O Tab) Issued: May-2012
Modbus communications can use either serial or IP communications.
communication, and specify the port you want to use. (See Port).
ControlWave MRMS-IC Configuration Manual (D5140)
Field Description
Port
IP Click the IP selection to use IP Modbus (Open
Modbus Slave Address
Modbus Type
Specify the serial communication port on the ControlWave Micro you want to use for Modbus slave communication. Use the following code:
Enter this 1 COM1 2 COM2 3 COM3 4 COM4 5 COM5 6 COM6 7 COM7 8 COM8 9 COM9 10 COM10 11 COM11 Press [Enter] to save the selection.
Modbus) communication.
Enter the Modbus slave address. If the local slave address you enter has already been assigned to either the SCADA Enron Modbus slave interface, or any of the other Customer Modbus Slave sessions, you will see a Loc Addr Conflict message. Modify the Modbus Slave Address as required to resolve the conflict.
: To select this serial CW Micro port:
Enron If you want to communicate using Enron Modbus,
click this selection.
Gould If you want to communicate using Gould Modbus,
click this selection.
Data Parameters Word Order Choose the data word order to match the data word
order used by the Modbus Master that communicates with this Modbus Slave.
High Word First
Low Word First
Byte Order Choose the data byte order to match the data byte
Issued: May-2012 Configuring Inputs and Outputs (I/O Tab) 2-15
Click this to specify that the high word is first.
Click this to specify that the low word is first.
ControlWave MRMS-IC Configuration Manual (D5140)
Field Description
High Byte First
order used by the Modbus Master that communicates with this Modbus Slave.
Click this to specify that the high byte is first.
Low Byte First
Bit Order Choose the data bit order to match the data bit order
High Bit First
Low Bit First
Data Size
Click this to specify that the low byte is first.
used by the Modbus Master that communicates with this Modbus Slave.
Click this to specify that the high bit is first.
Click this to specify that the low bit is first.
Select the appropriate data format for Modbus Register data from the drop down menu. The available selections are:
Single Bit – Each Register will include a single bit Byte Data – Each Register will include a single byte 16 Bit Integer – Each Register will include a single
16-bit integer 32 Bit Int., 1 Reg., Cnt*1, Adr*1 – Each Register
will include a 32-bit double integer. 32 Bit Float, 1 Reg., Cnt*1, Adr*1 – Each Register
will include a 32-bit floating point number 32 Bit Int., 2 Reg., Cnt*2, Adr*2 – Two registers will
be used for each 32-bit double integer. The MODBUS Master must poll two registers for each 32 bit integer.
32 Bit Float, 2 Reg., Cnt*2, Adr*2 – Two registers will be used for each 32-bit floating point number. The MODBUS Master must poll two registers for each 32 bit number.
32 Bit Int., 2 Reg., Cnt*2, Adr*1 - Two registers will be used for each 32-bit double integer. The MODBUS Master must poll a single register for each 32 bit integer.
32 Bit Float, 2 Reg., Cnt*2, Adr*1 - Two registers will be used for each 32-bit floating point number. The MODBUS Master must poll a single register for each 32 bit number.
Press [Enter] to save the selection. If you don’t make a selection, the field shows Not Set.
2-16 Configuring Inputs and Outputs (I/O Tab) Issued: May-2012
Loading...
+ 110 hidden pages