Fisher 444 Alphaline User Manual

00809-0100-4263
English
Rev. AA
Model 444 Alphaline® Temperature Transmitters
Product Manual
1
Model 444 Alphaline
®
Temperature Transmitters
NOTICE
Read this manual before working with the product. For personal and system safety, and for optimum product performance, make sure you thoroughly understand the contents before installing, using, or maintaining this product.
Within the United States, Rosemount Inc. has two toll-free assistance numbers. Customer Central: 1-800-999-9307 (
Technical support, quoting, and order-related questions.
North American 1-800-654-7768 ( Response Center: Equipment service needs.
For equipment service or support needs outside the United States, contact your local Rosemount representative.
7:00 a.m. to 7:00 p.m. CST)
24 hours a day – Includes Canada)
Rosemount Inc.
8200 Market Boulevard Chanhassen, MN 55317 USA Tel 1-800-999-9307 Telex 4310012 Fax (612) 949-7001
00809-0100-4263 © Rosemount Inc. 1998.
http://www.rosemount.com
N
I
R
P
IN
U.
S.
The products described in this document are NOT designed for nuclear­qualified applications.
Using non-nuclear qualified products in applications that require nuclear­qualified hardware or products may cause inaccurate readings.
For information on Rosemount nuclear-qualified products, contact your local Rosemount Sales Representative.
SNF-0004
Ros
emount, the Rosemount logotype, and Alphaline are registered trademarks of Rosemount Inc.
Chromel and Alumel are trademarks of Hoskins Mfg. Co.
Cover Photo: 444-005AC
Fisher-Rosemount satisfies all obligations coming from legislation to harmonize product requirements in the European Union.
T
E
D
A.
Table of Contents
SECTION 1 Introduction
SECTION 2 Installation
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
General Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
Mechanical Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Mounting Stability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Access Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Housing Rotation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Terminal Side of Electronics Housing . . . . . . . . . . . . . . . . . 2-2
Circuit Side of Electronics Housing . . . . . . . . . . . . . . . . . . 2-2
Electrical Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Field Wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
Sensor Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4
RTD Inputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4
Thermocouple or Millivolt Inputs . . . . . . . . . . . . . . . . . . . . 2-6
Grounding . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6
Multi-Channel Installations . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6
Surges/Transients . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6
Environmental Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-7
Temperature Environment . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-7
Moist or Corrosive Environments . . . . . . . . . . . . . . . . . . . . . . . 2-9
Hazardous Location Installation . . . . . . . . . . . . . . . . . . . . . . . . 2-9
Intrinsically Safe Installation . . . . . . . . . . . . . . . . . . . . . . . . . . 2-9
Installation Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-10
Mechanical . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-10
Electrical . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-11
SECTION 3 Calibration
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Safety Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Calibrating a RTD Transmitter . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
Calibrating a Thermocouple Transmitter
Using a Compensated Thermocouple Simulator . . . . . . . . . . . 3-5
Using an Ice Bath . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7
Calibrating a Low-Power Transmitter . . . . . . . . . . . . . . . . . . . . . . 3-8
Calibrating a Millivolt Transmitter . . . . . . . . . . . . . . . . . . . . . . . . . 3-10
i
SECTION 4
Maintenance and
Troubleshooting
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
Hardware Diagnostics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
Repair . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
Disassembly Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
Reassembly Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5
Interchangeability of Parts
Mechanical Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-6
Electrical Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-6
Burnout Protection Adjustments . . . . . . . . . . . . . . . . . . . . . . . . 4-6
Repair and Warranty Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-7
Return of Materials . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-7
SECTION 5 Specifications and Reference Data
SECTION 6 Options
APPENDIX A Approval Drawings
APPENDIX B T emperature Sensor Reference Information
Functional Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
Performance Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
Physical Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
LCD Meter Specifications
Functional Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-7
Performance Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-7
Physical Specification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-7
Analog Meter Specifications
Functional Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-8
Performance Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-8
Physical Specification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-8
Mounting Bracket . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
LCD / Analog Meter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-2
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-1
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-1
ii
Section
1 Introduction
OVERVIEW
This manual is designed to assist in installing, operating, and maintaining Rosemount Transmitters.
Section 2 Installation
provides mechanical, electrical, and environmental considerations to guide you through a safe and effective transmitter installation.
Section 3 Calibration
provides different Model 444 calibration procedures.
Section 4 Maintenance and Troubleshooting
provides hardware diagnostics, maintenance tasks, basic hardware troubleshooting techniques, and considerations for returning materials.
Section 5 Specifications and Reference Data
provides functional, performance, and physical transmitter specifications; also includes transmitter dimensional drawings, ordering information, and spare parts.
Section 6 Options
provides a listing of transmitter options and a description of each.
Appendix A Approval Drawings
contains approval drawings for Canadian Standards Association (CSA) and Factory Mutual (FM) instrinsic safety installation.
Appendix B Temperature Sensor Reference Information
provides reference information regarding the application of various Rosemount temperature sensors.
®
Model 444 Alphaline® Temperature
1-1
Rosemount Model 444 Alphaline Temperature Transmitters
1-2
Section
2 Installation
OVERVIEW
This section includes the following transmitter installation information:
• General Considerations
• Mechanical Considerations Mounting Stability
Access Requirements
• Electrical Considerations Power Supply
Field Wiring Sensor Connections Grounding Multi-Channel Installations Surges/Transients
• Environmental Considerations Temperature Environment
Moist or Corrosive Environments Hazardous Location Installation Intrinsically Safe Installation
• Installation Procedure Mechanical
Electrical
GENERAL CONSIDERATIONS
Failure to follow these installation guidelines may result in death or serious injury. Make sure only qualifiedpersonnel perform the installation.
Explosions can cause death or serious injury. Verify that the operating atmosphere of the transmitter is consistent with the appropriate hazardous locationscertifications.
Use the Rosemount Model 444 Alphaline Temperature Transmitter when the temperature measurement point is remote from the control, readout, or recording point, or where the measurement point is exposed to environmental conditions that would be harmful to unprotected signal conditioning equipment.
Electrical temperature sensors such as RTDs and thermocouples
2-1
Rosemount Model 444 Alphaline Temperature Transmitters
produce low-level signals proportional to their sensed temperature. Model 444 temperature transmitters convert the low-level sensor signal to a standard 4–20 mA dc signal that is relatively insensitive to lead length and electrical noise. This current signal is then transmitted to the control room via two wires.
Figures 2-1, and 2-2 show recommended mounting configurations for transmitter and sensor assemblies. See Section 6 Options for information regarding Model 444 transmitter accessories.
MECHANICAL CONSIDERATIONS
Mounting Stability
Access Requirements
Housing Rotation You may rotate the transmitter in 90-degree increments to improve field access
Terminal Side of Electronics Housing
Circuit Side of Electronics Housing
You can attach the transmitter directly to the sensor assembly as shown in Figures 2-1 and 2-2. An optional mounting bracket permits the transmitter to be mounted remotely from the sensor(s), either on a flat surface or attached to a two-inch pipe (See Figure 2-11 on page 2-13). The choice of mounting method must take into account a number of factors:
Mounting stability is an important consideration. The transmitter, though rugged, may require supplementary support under high-vibration conditions, particularly if extensive thermowell lagging or long extension fittings are used. In such instances, the pipestand mounting technique shown in Figure 2-11 on page 2-13 is preferable.
When choosing an installation location and position, take into account the need for access to the transmitter.
to both compartments. Make wiring connections through the conduit openings on the terminal side of
the electronics housing. Mount the transmitter so the terminal side is accessible, and be sure to provide adequate clearance for cover removal.
The transmitter electronics are installed in the circuit side of the transmitter housing. In case of electronic malfunction, provide adequate clearance for circuit-side cover removal. Also, be sure to account for additional clearance if a meter is to be installed. For more information regarding the meter option, refer to Section 6 Options.
2-2
FIGURE2-1. Recommended Process Mounting.
Installation
Thermowell
NOTE Dimensionsare in inches (millimeters).
FIGURE2-2. Recommended Process Mounting with Drain Seal.
Thermowell Hex
Extension Nipple
Extension
Length
Union or Coupling
Sensor Hex
Transmitter
Conduit for
Field Wiring
(dc Power)
3.2
(81.3)
444-0200C
Extension Nipple
Union or Coupling
Sensor Hex
Drain Seal
Conduit for
Field Wiring
(dc Power)
Coupling
Close Nipple
Street Ell
Transmitter Terminal Side
444-0200F
2-3
Rosemount Model 444 Alphaline Temperature Transmitters
ELECTRICAL CONSIDERATIONS
Po wer Supply
FIGURE 2-3. Model 444 Load Limits.
This section contains information that you should consider when preparing to install Model 444 transmitters. Read this section carefully before going on to the installation procedures. Metal conduit should be used to enclose cabling for best results in electrically noisy environments.
The dc power supply should provide power with less than 2% ripple. The input voltage versus load limitation relationship for 4–20 mA transmitters is shown in Figure 2-3. Figure 2-4 shows field wiring for a standard Model 444 transmitter. Figure 2-5 shows field wiring for Models 444LL and 444LM low-power voltage output units, which require 100K ohms minimum load. The total R-load is the sum of the resistance of the signal leads and the load resistance of the controller, indicators, and related devices. Note that the resistance of intrinsic safety barriers, if used, must be included in the total load.
Power Supply Load Limitations
R
LOAD(MAX.)
1650 1500
1000
500
Load (Ohms)
0
Voltage
Too Low
12 20 30 40
=503 (V
Power Supply (V dc)
(MIN.)
Operating
Region
– 12)
Field Wiring
Explosions mayresult in death or serious injury. Do not remove the instrument cover in explosive atmospheres when the circuit is alive.
High voltage that may be present on leads can cause electrical shock.Avoid contact with leads and terminals.
Do not apply high voltage (e.g. ac line voltage) to the transmitter terminals. Abnormally high voltage can damage the unit.
All power to the transmitter is supplied over the signal wiring. Signal wiring need not be shielded, but use twisted pairs for best results. Do not run unshielded signal wiring in conduit or open trays with power wiring, or near heavy electrical equipment. To power the transmitter, connect the positive power lead to the terminal marked “+” and the negative power lead to the terminal marked “–” (see Figures 2-4 and 2-5). Tighten the terminal screws to ensure adequate contact. No additional power wiring is required.
2-4
FIGURE 2-4. Field Wiring for the StandardModel 444 Transmitters.
Meter Connections
and S ignal
Test Points
(+)
{
(–) (–)
(+)
{
dc Power
Installation
FIGURE 2-5. Field Wiring for Low-Power Model 444 Transmitters (444LL and LM).
RTD Input
(typical)
Zero Adjust
RTD Input
Zero Adjust
Output Voltage (+)
Optional
Ground
Span Adjust
(+) dc Power
(–) Common
Shield
Span Adjust
+–+
Power
Supply
+
++–
444-0000C02A
Output Load Limitation
Minimum Load = 100K
AtoD
Converter
Optional
Ground
444-0000A02A
Sensor Connections
RTD Inputs
Explosion mayresult in death or serious injury. Do not remove the instrument cover in explosive atmos pheres when the circuit is alive.
Highvoltagethatmaybe present on leads can causeelectrical shock. Avoid contact with the leads and the terminals.
Various RTD configurations are used in industry; each configuration offers a specific solution for compensating the effects of lead wire resistance. They include 3- and 4-wire designs. The correct installation for each of these RTDs is shown in Figures 2-6a and b on page 2-6.
If the transmitter is mounted remotely from the RTD, operation will be satisfactory, without recalibration, for lead wire resistances of up to 2 ohms per lead (equivalent to 200 feet of 20 AWG wire). In this case, the leads between the RTD and transmitter should be shielded.
2-5
Rosemount Model 444 Alphaline Temperature Transmitters
The correct connections for a compensation loop RTD and a 2-wire RTD are shown in Figure 2-6c and Figure 2-6d, respectively. In a 2-wire RTD, however, both leads are in series with the sensor element, so significant errors (0.1 °C) could occur if the lead lengths are greater than one foot. For longer runs when using a 2-wire RTD, attach a third lead and connect as shown in Figure 2-6a.
FIGURE 2-6.Sensor Wiring Diagrams.
Signal
+
Red
Red
White
Model 444 with 3-Wire RTD
Figure 2-6a
Signal
+ Output Common –
Red
Model 444LL or 444LM
with 3-Wire RTD
Figure 2-6e
Signal
+
Red
Red
White
Model 444 with 4-Wire RTD
Figure 2-6b
Red
White
Model 444MV with Millivolt
Input or Model444T Series
with Grounded or
Ungrounded Thermocouple
White
Signal
+
Figure 2-6f
Model 444 with
Comp. Loop RTD
Figure 2-6c
+
High T/C
+
Signal
+
Red
Jumper
White
Model 444 with 2-Wire RTD
Black
Black
Signal
++ ––
Model 444MV used as
DifferentialMillivolt
Transmitter (T/C Junctions
must be ungrounded)
Figure 2-6g
Figure 2-6d
+
Low T/C
Signal
White
Red
+
444-0203A;B;C;D
FIGURE 2-7.Characteristics of Thermocouple and RTD Input Wires.
Thermocouple Type Positive Lead Negative Lead
J Iron (Magnetic) Constantan (Non-magnetic)
K Chromel (Non-magnetic) Alumel (Magnetic)
T Copper (Yellow color) Constantan (Silver color) E Chromel (Shiny metal) Constantan (Dull metal) R Platinum 1 3% Rhodium Platinum S Platinum 10% Rhodium Platinum
Single Element RTD Compensation Loop RTD Dual E lement RTD
Red
Red
White
White
2-6
Red
White
Black
Black
Red
White White
Black
Green Green
444-0207A
Installation
Thermocouple or Millivolt Inputs
Grounding
In the case of thermocouples, make connections between the sensor and the transmitter with thermocouple wire. For process mounting applications, connect the thermocouple directly to the transmitter. For installations where the transmitter is mounted remotely from the sensor, use appropriate thermocouple extension wire. As with all low-level signal wiring, shielding is recommended for long runs. Make input connections for the Model 444MV Millivolt Transmitter using copper wires. The correct connections for thermocouple and millivolt inputs are shown in Figures 2-6f and g.
The transmitter will operate with the current signal loop either floating or grounded. However, many types of readout devices are affected by the extra noise in floating systems. If operation appears noisy or erratic, grounding the current signal loop at a single point may solve the problem. The negative terminal of the power supply is the best place to ground the loop. Alternately, either side of the readout device could be grounded. Do not ground the current signal loop at more than one point.
Thermocouple and millivolt transmitters are isolated, so the input circuit also may be grounded at any single point (when a grounded thermocouple is used, this point is the grounded junction), and the signal loop may be grounded at any point.
The 444RL transmitter is not isolated, so there can be no grounds in the RTD circuit. Since RTDs must be well-insulated from ground in order to give correct temperature readings, this is not normally an installation limitation. The positive side of the power supply should not be grounded for use with RTD input transmitters. The 444RI9 should be used with grounded RTDs.
Multi-Channel Installations
FIGURE 2-8. Multi-Channel Installation.
Transmitter
No. 1
Transmitter
No. 2
If using shielded wire, connect the shield of the sensor-to-transmitter cable to the shield of the transmitters-to-receiver cable. Ground the shielding only at the signal loop ground.
Figure 2-8 illustrates how several transmitters can be connected to a single master power supply. In this instance, the system can be grounded only at the negative power supply terminal. Since several channels are dependent on one supply, an uninterruptible power supply or backup battery should be considered if loss of all channels would pose operational problems. The diodes shown in Figure 2-8 prevent unwanted charging or discharging of the battery.
ToAdditional
Transmitters
Readout or
Controller No. 1
Readout or
Controller No. 2
+
Backup Battery
dc
Power
Supply
+
444-0202A
2-7
Rosemount Model 444 Alphaline Temperature Transmitters
Surges/Transients
ENVIRONMENTAL CONSIDERATIONS
Temperature Environment
The transmitter will withstand electrical transients of the energy level usually encountered in static discharges or induced switching transients. However, high­energy transients, such as those induced in wiring from nearby lightning strikes, can damage both the transmitter and the sensor.
To protect against high-energy transients, install Model 444 transmitters in conjunction with the Rosemount Model 470 Transient Protector. The Model 470 prevents damage from transients induced by lightning, welding, heavy electrical equipment, or switch gears. Refer to the Model 470 Transient Protector product data sheet, pub. no. 00813-0100-4191 for more information.
The transmitter will operate within specifications for ambient temperatures between –25 and 85 °C. It will function, but not necessarily within specifications, in ambient temperatures between –40 and 100 °C.
Aside from ambient temperature variations, heat from the process is transferred from the thermowell to the transmitter housing. If the process temperature is near or beyond specification limits, use excess thermowell lagging or an extension nipple to protect the transmitter from the high temperature condition. See Figure 2-10.
EXAMPLE:
Suppose the maximum ambient temperature is 40 °C and the temperature to be measured is 540 °C. The maximum allowable housing temperature rise is the rated temperature specification limit minus the existing ambient temperature (85 – 40), or 45 °C. As shown in Figure 2-9, an “E” dimension of
3.6 inches will result in a housing temperature rise of 22 °C. An “E” dimension of 4 inches would therefore be the minimum recommended length, and would provide a safety factor of about about 25 °C. A longer “E” dimension, such as 6 inches, would be desirable in order to reduce errors caused by transmitter temperature effect, although in that case the transmitter would probably require extra support. If a thermowell with lagging is used, the “E” dimension may be reduced by the length of the lagging.
2-8
Installation
FIGURE 2-9. Model 444 Transmitter Housing Temperature Rise.
FIGURE 2-10. Sensor Assembly Dimensional Drawings.
HOUSING
RISE
ABOVE
AMBIENT
°C (°F)
60
(108)
50
(90)
40
(72)
30
(54)
22
20
(36)
10
(18)
Connection Head with Extended Cover
25
0
3 4 5 6 7 8 9
Bayonet Spring Loaded Sensor
Transmitter Housing Temperature Rise
E
vs. Length for a Test Installation
8
1
5
°
C
(
1
540
°
C
(
1000
0
3.6
°
C
(
4
8
°
F)
O
v
e
nT
E
5
0
0
°
F
)
O
v
e
n
T
e
F
)
O
v
e
n
T
e
m
pe
e
mp
e
r
a
t
u
r
e
Length in Inches
½–14 NPT Thread
m
p
e
r
a
tu
r
e
r
a
t
ur
e
3044-0123A
NOTE Dimensions are in inches (millimeters).
5.5
(140)
3.5
(89)
Chain
Sensor Installed in
Connection Head (Extended Cover)
with Coupling and
Nipple Extension
1.0
(25)
2.2
(56)
Spring Loaded Sensor
Connection Head with
Flat Cover
¾–14 NPT
Nominal Fitting
Sensor Installed in ConnectionHead(Flat Cover)withUnionand
Nipple Extension and
Thermowell
Sensor Length (Ref.)
Length
X
E
V
(1)
+ 1.75 (44)
T
(1)
T
= 0.0 on Standard Assembly Thermowells
L
0.53 (13) Max. Thread
Engagement
¾–14 NPT
on Thermowell
U
0.25 (6)
Sensors-0000A06A
2-9
Rosemount Model 444 Alphaline Temperature Transmitters
Moist or Corrosive Environments
HazardousLocation Installation
The transmitter is designed to resist attack by moisture and other corrosives. The coated circuit boards are mounted in a compartment completely sealed from the conduit entrances. O-ring seals protect the interior when the covers are installed. In humid environments, however, it is still possible for moisture “breathing” to occur in conduit lines. If the transmitter is mounted at a low point in the conduit run, the terminal compartment could fill with water, causing electrical shorting. The transmitter should be mounted so moisture from the conduit will not drain into the housing. In some instances a drain seal, installed as shown in Figure 2-2 on page 2-3, is advisable.
Explosions mayresult in death or serious injury.Verify that the operating atmosphere of the transmitter is consistent with the appropriate hazardous locations certifications.
Explosions mayresult in death or serious injury.Both transmitter covers must be fully engaged to meet explosion­proof requirements.
Intrinsically Safe Installation
The Model 444 is designed with an explosion-proof housing and circuitry suitable for intrinsically safe and non-incendive operation. Individual transmitters are clearly marked with a tag indicating the approvals they carry. The various approvals are available as options. Refer to Section 5 Specifications and Reference Data for a complete listing of available approvals.
To maintain certified ratings for installed transmitters, install in accordance with applicable installation codes and approval drawings. Refer to Appendix A Approval Drawings for Model 444 installation drawings. For future orders, refer to the current product price list for the most up-to-date information on these approvals.
For explosion-proof installations, installation location must be m ade in accordance with Rosemount drawing 00444­0261, Rev. E.
For intrinsically safe installations, installation location must be m ade in accordance with Rosemount drawing 00444­0034, Rev. C (CSA) or00444-0264, Rev. B (SAA).
You can use Intrinsically safe installations instead of explosion-proof installations in hazardous areas. In such configurations, the transmitter and sensor are located in a hazardous area, and the current signal leads are connected to equipment in a non-hazardous area through intrinsic safety barriers that limit the voltage and current fed into the hazardous area. Install in accordance with the barrier manufacturer’s instructions for the specific barrier used. For approval information, refer to Table 5-1 on page 5-4, and Table 5-2 on page 5-5. For installation information, refer to the intrinsically safe barrier systems reference drawings in Appendix A Approval Drawings.
2-10
Installation
INSTALLATION PROCEDURE
Installation consists of mounting the transmitter and sensor assembly and making electrical connections. If mounting the transmitter directly to the sensor assembly, use the process mounting technique shown in Figure 2-1 or Figure 2-2, on page 2-3. For transmitter locations remote from the sensor, use conduit between the sensor and transmitter. Transmitter hubs will accept male conduit fittings with ½–14 NPT; ½–14 NPSM; or ½–14 taper thread per ANSIC 80.4.
Explosion mayresult in death or serious injury. Do not remove the instrument cover in explosive atmos pheres when the circuit is alive.
High voltage that may be present on leads may cause electrical shock.Avoid contact with the leads and the terminals.
Process leaks may result in death or serious injury.Install and tighten thermowells or sensors before applying pressure, or process leakage may result. Removing the thermowell or sensor while in operation may cause process fluid leaks.
Mechanical
Electrical
1. Mount the thermowell to the pipe or process container wall.
2. Attach any necessary extension nipples and adapters. Seal the nipple and adapter threads with silicone or tape.
3. Screw the sensor into the thermowell.
4. Install drain seals if required for severe environments or to satisfy code requirements (See Figure 2-2 on page 2-3).
5. Attach the transmitter to the thermowell assembly. Seal the adapter threads with silicone or tape.
6. Install conduit for field wiring to the remaining conduit entry of the transmitter. Seal conduit threads with silicone or tape.
7. Pull field wiring leads through the conduit into the terminal side of the transmitter housing.
For explosion-proof installations, wiring connections must be m ade in accordance with Rosemount drawing 00444­0261, Rev. E.
For intrinsically safe installations, wiring connections must be m ade in accordance with ANSI/ISA-RP12.6, and Rosemount drawing 00444-0034, Rev.C (CSA) or 00444­0264, Rev. B (SAA).
For all installations, wiring connections must follow the National Electric Code.
2-11
Rosemount Model 444 Alphaline Temperature Transmitters
Preliminary Checkout
1. For any Model 444 unit, first verify that the transmitter is calibrated to the required range. Calibration is usually performed by substituting an input in place of the sensor, and this is most conveniently accomplished prior to sensor connection. Refer to the calibration procedures in Section 4 Maintenance and Troubleshooting.
Input Connections
2. Model 444RL: Connect the RTD leads as shown in Figure 2-6a, b, c, d, or e depending upon the lead compensation method used.
Model 444T series: Connect the thermocouple leads as shown in Figure 2-6f. Polarity is important; be sure to identify the leads accurately. The negative lead is usually red; if there is no color coding, the characteristics provided in Figure 2-7 may be helpful.
Model 444MV: If using the transmitter as a millivolt-to-milliampere converter, use ordinary copper leads for input connections as shown in Figure 2-6f. If using the transmitter with two thermocouples to measure differential millivolt, connect the thermocouples as shown in Figure 2-6g. The “high” thermocouple causes the transmitter output to increase when its temperature increases relative to the “low” thermocouple. Grounded thermocouples cannot be used for differential measurements.
Models 444LL and LM: In these low-power option packages, the RTD leads are connected the same as in the conventional RTD arrangements shown in Figure 2-6a, b, c, and d.
Output Connections
3. For all 4–20 mA models, use ordinary copper wire of sufficient size to assure that the voltage across the transmitter power terminals does not go below 12 V dc (See Figure 2-3). For multi-channel or intrinsically safe installations, see applicable paragraphs in this section.
Model 444RL: Connect current signal leads as shown in Figure 2-6a, b, c, or d.
Model 444T series: Connect current signal leads as shown in Figure 2-6f. Model 444MV: Connect current signal leads as shown in Figure 2-6f or g. Models 444LL and LM: Connect current signal leads as shown in Figure
2-6e.
Final Checkout
4. For all models, recheck the polarity and correctness of connections; then turn the power on.
2-12
FIGURE 2-11. Model 444 with Optional Mounting Bracket.
Installation
PIPESTAND MOUNTING
Clearance Hole
for ¼-inch Bolt
(eight places)
NOTE Dimensions are in inches (millimeters).
FIGURE 2-12. Model 444 Dimensional Drawings
4.5 Max. (114)
Transmitter can be Rotated 90°
Mounting Bracket
¼–20 3½-inch
5
/16 –18 U-bolt for
2-inch Pipe (2)
Permanent Tag
(Optional)
Bolt (4)
PANEL OR SURFACE MOUNTING
5.00 (127)
7.5 (191) Max. with Optional Meter
Meter Housing
5
/16 -inch Bolts
(four required,
not furnished)
2.81 (81)
4.5 Max. (114)
Hole for
5
/16 -inch
Bolts
(four
places)
0.75 (19)
Clearance for
Cover Removal
(Typical)
444-1151G, 1151F04A
4.5 Max.
Nameplate
Explosion Proof or
Intrinsic Safety L abel
(Optional)
0.375 (10) Min. Dp.
NOTE Dimensions are in inches (millimeters).
(114)
Mounting Holes
¼–20 UNC–2B
(four places)
½–14 NPT per
ANSI C80.4 for
Conduit or Sensor
Connection
(two places)
0.87 (22)
1.7
(44)
(117)
Terminal
Connections
this Side
0.36 (9)
0.72 (18)
4.2
Terminal Circuitry this Side
444-51LTE 05A, 51LTG05A, 51LTF 05A
2-13
Rosemount Model 444 Alphaline Temperature Transmitters
2-14
Section
3 Calibration
OVERVIEW
SAFETY MESSAGES
Each transmitter is factory calibrated to the temperature range shown on the nameplate. If calibration to a specific range is not specified on the purchase order, the transmitter is calibrated to maximum span with a base temperature of 0 °C, and the “Calibration” entry on the transmitter nameplate is left blank. For more specific calibration information and a complete breakdown of transmitter parts, refer to Section 5 Specifications and Reference Data.
Only a few calibration laboratories have the kind of precision temperature baths necessary for accurate direct calibration of a temperature sensor or sensor/transmitter system. As a result, the transmitter is normally calibrated by substituting a resistance decade box for an RTD or a compensated thermocouple simulator for a thermocouple.
This section contains the following transmitter calibration information:
• Calibrating a RTD Transmitter
• Calibrating a Thermocouple Transmitter
• Calibrating a Low-Power Transmitter
• Calibrating a Millivolt Transmitter
This section contains procedures that require removing the transmitter covers and making electrical connections. The following safety messages apply to all such procedures.
Explosion mayresult in death or serious injury. Do not remove the instrument cover in explosive atmos pheres when the circuit is alive.
High voltage that may be present on leads can cause electrical shock.Avoid contact with the leads and the terminals.
3-1
Rosemount Model 444 Alphaline Temperature Transmitters
CALIBRATING A RTD TRANSMITTER
Calibration Procedure To calibrate a model 444RL or 444RL___B0912, perform the following procedure:
Calibration Equipment Required:
Readout Resistor. The transmitter test terminals give a 40–200 mV signal. The Models 444RL and444RL ___B0912 have a jumper-selectable 4–20 mA test output option (2-board sets). If this is not suitable for the test equipment available, a ±0.1% tolerance, 0.5 W precision wirewound resistor is needed. Suggested values include a 100-ohm resistor to give a 0.4 to 2 volt output; or 500 ohms for 2 to 10 volts.
Voltmeter (such as a 5-digit DVM). Voltage rating is dependent upon the test signal. Accuracy is ±0.01%; resolution is 1 mV.
dc Power Supply. Power capability is 24 V dc at 35 mA. Resistance Decade Box. Precision type, 5-dial, with largest dial providing 100-
ohm steps. Accuracy is ±0.02 ohm. The decade box should be periodically calibrated against a 5-dial Wheatstone bridge.
Lead Simulation Resistors. If the transmitter is to be mounted remote from the RTD, and the lead resistance between the transmitter and the RTD is greater than 2 ohms per lead (equivalent to 200 ft of 20 AWG wire), the transmitter should be trimmed with simulated lead resistances for best accuracy. This requires wirewound resistors with resistance values equal to the nominal lead resistance of the RTD.
1. The Models 444RL and 444RL ___B0912 have a jumper-selectable 4–20 mA test output option. If a 4–20 mA test output is required, reposition the test terminal output jumper on the range board (the default setting is 40– 200 mV). Refer to steps 2 through 4 of the disassembly procedure, on page 4-4, for information on removing the circuit board assembly.
Place the jumper in the position labelled “A” for a 4–20 mA test output. See Figure 3-1. Refer to steps 5 through 12 of the reassembly procedure, on page 4-5, for information on reinstalling the circuit board assembly.
2. Remove the cover from the terminal side of the transmitter housing.
3. If an RTD is already connected, remove all RTD lead connections.
4. Attach the calibration test equipment as shown in Figure 3-2. Use miniature banana plugs to make terminal connections. Use simulated lead resistors only if necessitated by long lead wire lengths, as discussed above.
3-2
NOTE
If using RTD configurations other than the 3-wire design shown in Figure 3-2, refer to Figure 2-6 on page 2-6 for the correct wiring.
5. If trimming the transmitter to a new range, you may have to reposition the Coarse Zero Jumper on the Range Board. If so, see the disassembly procedure on page 4-4. Position the jumper in the location shown in Table 3-1. (A transmitter with a Base Temperature outside the regions shown in Table 3-1 is a special design, and does not contain a Coarse Zero Jumper.) Reassemble the circuit boards.
Calibration
6. Determine the RTD resistance at the desired base and full scale temperatures. For Calibration Code 1 (see Table 5-3, on page 5-9), these resistances are listed in Table B-1.
7. Turn the power on.
8. Set the decade box to the resistance corresponding to the desired base temperature. Adjust the zero potentiometer until the output is 4 mA. Remember that recovery time of the unit from an underscale condition is longer than from an over-scale condition. Therefore, set the box to a higher resistance than that desired, then bring it down to the correct value.
9. Set the decade box to the resistance corresponding to the desired full scale temperature. Adjust the span potentiometer until the output is 20 mA.
10. Repeat steps 8 and 9 until you obtain the 4 and 20 mA readings without readjusting the span and zero potentiometers. Complete this process more quickly by noting the full scale reading before readjusting the span pot, using the span pot to overshoot the desired reading by 20%, and then using the zero pot to readjust the full scale reading to 20 mA.
EXAMPLE:
To calibrate the Model 444RL1U1 for a range of 100 to 150 °F (38 to 66 °C), first consult Table 3-1, and plug the jumper into pins Z2. From Table B-1, trim points are 114.68 and 125.37 ohms corresponding to 100 °F and 150 °F respectively. After adjusting the base to 4 mA, and setting the decade to full scale resistance, output equals 22.5 mA, or 2.5 mA greater than desired. Set the span pot to an output lower than 20 mA by the amount equal to 20% of
2.5 equals 0.5 mA, or 19.5 mA. Reset the zero pot so the output equals 20 mA. Repeat steps 8 and 9 and this procedure until readjustments are no longer necessary.
TABLE 3-1. Coarse Jumper Location, Model 444R.
11. Disconnect the decade box and the readout resistor. Reconnect the RTD and power leads. Replace the terminal cover.
12. Mark the correct range in the “Calibration” space on the nameplate
.
Base Temperature
Region
°C °F 444RL1 444RL2 444RL3
–50 to 0
0to50
50 to 100
100 to 150
NOTE
If the base temperature is at a dividing point between regions, use the lowerjumper position optimum performance; i.e.,uselocationZ1 rather than Z2 for Model 444RL1 with a base temperature of 0 °C.
–58 to 32
32 to 122 122to212 212to302
Jumper Location
Z1 Z2 Z3 Z4
Z1 Z1 Z2 Z2
Continuously
adjustable
(no jumper)
3-3
Rosemount Model 444 Alphaline Temperature Transmitters
FIGURE 3-1.Location of Test Input and Burnout Protection Jumper on Model 444RL Range Board.
Test Output
Jumper Position
BurnoutProtection
Jumper Position
444-0002ACCA
FIGURE 3-2. RTD Transmitter Calibration Diagram.
Decade Box
DVM
Transmitter
Lead Simulator Resistors (If required)
Power
Supply
Readout Resistor
DVM
Alternate Readout
444-0215A
3-4
CALIBRATING A THERMOCOUPLE TRANSMITTER
Calibration
Using a Compensated Thermocouple Simulator
Calibration Procedure 1. Remove the cover from the terminal side of the transmitter housing.
Calibration Equipment Required
Compensated Thermocouple Simulator. Precision voltage source providing conformity to NIST Monograph 125 thermocouple curves. Reflect accuracy of simulator to desired calibration span. A simulator accuracy four times better than the transmitter is recommended (0.05% of calibrated span or 0.005mV whichever is greater). Simulator inaccuracies greater than this will degrade system accuracy and factory calibration is recommended.
Voltme t er. Such as a 5-digit DVM. Accuracy is ±0.01%; resolution is 1 mV. dc Power Supply. Power capability is 24 Vdc at 35 mA. Thermocouple Wire.Use the same type as that used in the construction of the
thermocouple. Readout Resistor. The transmitter test terminals give a 40–200 mV signal. If
this is not suitable for the test equipment available, a ±0.1% tolerance, 0.5 W precision wirewound resistor is needed. Suggested values include a 100-ohm resistor to give a 0.4 to 2 volt output; 250 ohms for 1 to 5 volts; or 500 ohms for 2 to 10 volts.
2. If a thermocouple is already connected, remove all thermocouple lead connections.
3. Connect the equipment as shown in Figure 3-4. Be sure to maintain polarity from the transmitter to the thermocouple simulator. Make terminal connections using miniature banana plugs.
4. If trimming the transmitter to a new range, you may have to reposition the Coarse Zero Jumper on the Range Board. If so, see the Disassembly Procedure on page 4-4. Position the jumper in the location shown in Table . (A transmitter with a base temperature outside the regions shown in Table is a special design and does not contain a Coarse Zero Jumper. Also, Model 444 TR and TS transmitters do not have Coarse Zero Jumpers.) Reassemble the circuit boards.
5. Determine the base and full scale temperatures.
6. Turn the power on.
7. Refer to the thermocouple simulator instructions for setting the thermocouple type and engineering units. Set the simulator to the base (zero) temperature and adjust the zero pot until the output is 4 mA (or 40 mV at the test terminals).
8. Set the simulator to the full scale temperature and adjust the span pot until the output is 20 mA (or 200 mV at the test terminals).
9. Repeat steps 7 and 8 until you obtain the 4 and 20 mA readings without readjusting the pots. Use the “overshoot” technique described in step 9 of the RTD calibration procedure, if desired.
10. Disconnect the simulator leads. Reconnect the thermocouple and power leads, if required. Replace the terminal side housing cover.
11. Mark the new range in the “Calibration” space on the nameplate.
3-5
Rosemount Model 444 Alphaline Temperature Transmitters
TABLE 3-2. Coarse Zero Jumper Location, Model 444TJ, TK, TF, TT, and MV Range Code 1.
Transmitter Base Region
°C °F mV
–50to50 50 to 100
NOTE
Range Codes 2 and 3 are continuously adjustable over the range showninTable1.(NoCoarseZeroJumper)
FIGURE 3-3.Location of Burnout Protection Jumper on Model 444T Range Board.
–58to122 122to302
-2 to 3 3to8
Coarse Zero
Jumper Location
Z1 Z2
FIGURE 3-4. Compensated Thermocouple Simulator Calibration Diagram.
Thermocouple Simulator
Burnout R2, R3
Protection Resistors
.
Thermocouple
Wire
+ –
444-0010ACCA
DVM
+
+
Readout Resistor
+
Power
Supply
–+
Transmitter
(Alternate Readout)
DVM
444-0216A
3-6
Calibration
UsinganIceBath
Calibration Procedure 1. Remove the cover from the terminal side of the transmitter housing.
Calibration Equipment Required
Millivolt Source. Precision voltage source providing outputs from –10 to 100 mV. Reflect accuracy of four times better that the 444 transmitter is recommended (0.05% of calibrated span or 0.005 mV which ever is greater).
Voltme t er. Such as a 5-digit DVM. Accuracy is ±0.01%; resolution is 1 mV. dc Power Supply. Power capability is 24 V dc at
35 mA. Thermocouple Wire.Use the same type as that used in the construction of the
thermocouple. Readout Resistor. The transmitter test terminals give a 40–200 mV signal. If
this is not suitable for the test equipment available, a ±0.1% tolerance, 0.5 W precision wirewound resistor is needed. Suggested values include a 100-ohm resistor to give a 0.4 to 2 volt output; 250 ohms for 1 to 5 volts; or 500 ohms for 2 to 10 volts.
Ice Bath. For highest accuracy, a stirred ice bath (such as a Rosemount 911A) should be used, as well as ice made from deionized or distilled water.
Input Monitor Voltmeter. Use to monitor source when required. Resolution of ±0.001 mV for ranges up to 100 mV. This can be the same as voltmeter used to measure transmitter output if rangeability and resolution are sufficient for both levels.
2. Install the equipment as shown in Figure 3-5 and allow the thermocouple junctions to stabilize at the ice point. Make terminal connections with miniature banana plugs.
3. If trimming the transmitter to a new range, you may have to reposition the Coarse Zero Jumper on the Range Board. If so, see the disassembly procedure on page 4-4. Position the jumper in the location shown in Table . (A transmitter with a base temperature outside the regions shown in Table is a special design and does not contain a Coarse Zero Jumper. Also, Model 444 TR and TS transmitters do not have Coarse Zero Jumpers.)
4. Determine the thermocouple millivolt levels at the desired base and full scale temperatures. See NIST Monograph 125 or Table B-1.
5. Turn the power on.
6. Set the millivolt source until the monitoring voltmeter reads the emf corresponding to the desired base temperature. Adjust the zero pot until the output is 4 mA.
7. Set the millivolt source until the monitoring voltmeter reads the emf corresponding to the desired full scale temperature. Adjust the span pot until the output is 20 mA.
8. Repeat steps 6 and 7 until you obtain the 4 and 20 mA readings without readjusting the pots. Use the “overshoot” technique described in step 9 of the RTD calibration procedure, if desired.
9. Mark the correct range in the “calibration” space on the nameplate.
3-7
Rosemount Model 444 Alphaline Temperature Transmitters
FIGURE 3-5. Ice Bath Calibration Diagram.
Input
Monitor
DVM
Millivolt
Source
+
Thermocouple
Wire
DVM
+
+
Power
Supply
+ –
+
Ice B ath
+
DVM
(Alternate Readout)
444-0217A
CALIBRATING A LOW-POWER TRANSMITTER
Calibration Equipment Required
Voltme t er. Such as a 5-digit DVM. Accuracy is ±0.01%; resolution is 1 mV. dc Power Supply. Power capability is 5 V dc at 1.5 mA for Model 444LL and 8 V
dc at 2 mA for Model 444LM. Resistance Decade Box. Precision type, 5-dial, with largest dial providing 100-
ohm steps. Accuracy is ±0.02 ohm. The decade box should be periodically calibrated against a 5-dial Wheatstone bridge.
Lead Simulation Resistors. If the transmitter is to be mounted remote from the RTD, and the lead resistance between the transmitter and the RTD is greater than 2 ohms per lead (equivalent to 200 ft of 20 AWG wire), the transmitter should be trimmed with simulated lead resistances for best accuracy. This requires wirewound resistors with resistance values equal to the nominal lead resistance of the RTD.
Load Resistor. If the transmitter is to be operated with a load that is significantly different from the DVM used for calibration, a load resistor can be used for best calibration accuracy. A metal film, carbon comp, or wirewound resistor, as well as a decade box can be used to simulate the actual load.
3-8
Calibration Procedure The following steps describe the procedure for calibrating a low-power
transmitter, as shown in Figure 3-6:
1. Remove the cover from the terminal side of the transmitter housing.
2. If an RTD is already connected, remove all RTD lead connections.
3. Attach the calibration test equipment as shown in Figure 3-6. Make terminal connections using miniature banana plugs. Use simulated lead resistors only if necessitated by long lead wire lengths, as discussed above.
NOTE
If using RTD configurations other than the 3-wire design shown in Figure 3-2, refer to Figure 2-6 on page 2-6 for the correct wiring.
4. Determine the RTD resistance at the desired base- and full-scale temperatures. For Calibration Code 1 (see Table 5-5), obtain these resistances from Table B-1.
5. Turn the power on.
6. Set the decade box to the resistance corresponding to the desired base temperature. Adjust the zero potentiometer until the output is 0.8 V for Model 444LL or 1.0 V for Model 444LM.
7. Set the decade box to the resistance corresponding to the desired full-scale temperature. Adjust the span potentiometer until the output is 3.2 V for Model 444LL or 5.0 V for Model 444LM.
8. Repeat steps 6 and 7 until you obtain both the zero- and full-scale readings without adjusting the span and zero potentiometers. Complete this process more quickly by noting the full-scale reading before readjusting the span pot, using the span pot to overshoot the desired reading by 20%, and then using the zero pot to readjust the full scale reading.
9. Disconnect the decade box and the readout. Reconnect the RTD and power leads. Replace the terminal cover.
10. Mark the correct range in the “Calibration” space on the nameplate.
Calibration
FIGURE 3-6. Low-Power Transmitter Calibration Diagram.
Load Re sistors
Decade Box
Transmitters are calibrated at the factory with a 220 K V load.
(If required)
DVM
444LL and LM Transmitter
+
+
dcPower
Source
Lead S imulator Resistors (If required)
444-0218A
3-9
Rosemount Model 444 Alphaline Temperature Transmitters
CALIBRATING A MILLIVOLT TRANSMITTER
Calibration is identical to the thermocouple type (see Figure 3-5 on page 3-8), except that a reference junction and ice bath are not used. The millivolt source is connected directly to the transmitter input terminals with copper wire, and the desired millivolt levels are entered directly. See Table 3-2 for Coarse Zero Jumper locations.
3-10
Section
4 Maintenance and
Troubleshooting
OVERVIEW
This section contains the following transmitter maintenance and troubleshooting information:
Hardware Diagnostics
• Troubleshooting
• Repair Disassembly Procedure
Reassembly Procedure Interchangeability of Parts Burnout Protection Adjustments
• Repair and Warranty Service
• Return of Materials
Use onlythe procedures and new parts specifically referenced in this manual. Unauthorized procedures or parts can affect product performance and the output signal used to control a process, and may render the instrument dangerous. Direct any questions concerning these procedures or parts to Rosemount Inc.
HARDWARE DIAGNOSTICS
TROUBLESHOOTING
If you suspect a malfunction, refer to Table 4-1 to verify that transmitter hardware and process connections are in good working order. Under each of the seven major symptoms, you will find specific suggestions for solving the problem. Always deal with the most likely and easiest-to-check conditions first.
This section offers tips for troubleshooting several kinds of potential malfunctions. To determine a malfunction, use pin-like probes to break through the protective coating to make measurements on a circuit board.
4-1
Rosemount Model 444 Alphaline Temperature Transmitters
TABLE 4-1. Transmitter TroubleshootingSymptoms andCorrectiveActions.
Symptom Potential Source Corrective Action
High Output Sensor Check for a sensor or thermocouple opencircuit. (RL, MV, T-Series with upscale burnout protection only)
Loop Wiring Check for dirty or defective terminals, interconnecting pins, or receptacles. Electronics Assembly Check for dirty or defective interconnectingpins.
Erratic Output Loop Wiring Check foradequate voltage to the transmitter.
Electronics Assembly Check for dirty or defective interconnectingpins.
Low Output or No Output
Excessive Current (over 30 mA)
Excessive Output Shift with Ambient Temperature
UnitCannotbe Trimmed to Desired Base Temperature
UnitCannotbe Trimmed to Desired Span.
Sensor Check RTD leadsto ensure that theyare not shortingtogether or to ground.(RL only)
Loop Wiring Check for adequate voltage tothe transmitter.(RL only)
Loop Wiring Check for short between current signal leads.
Electronics Assembly Check for defective components in amplifier or current control section. Sensor Check for incorrect thermocouple typeor Incorrect thermocouplepolarity connection (T series only) Electronics Assembly Check to ensure that the burnout -protection jumpers positioned correctly (MV, T series only).
Transmitter Check to ensure that unit iscapable of desired range. Electronics Assembly Check to ensure that the range boardjumper is in the correct position.
Transmitter Check to ensure that unit iscapable of desired range. Loop Wiring Check for adequate voltage tothe transmitter. Electronics Assembly Check for defective components in amplifier or current control section.
Check for intermittent shorts, open circuits, and multiple grounds. Check for dirty or defective terminals orinterconnectionpins.
Check for correct RTD lead connection. (RL only) Check for open RTDlead on double-lead side. (RL only)
Check for intermittent shorts, open circuits, and multiple grounds. (MV, T series only) Check for proper polarity at the signal terminal. (MV,T series only) Check for dirty or defective terminals orinterconnectionpins.
Check to ensure that current signal leadsARE NOT connected to sensor terminals. Check that sensor leads ARE NOT groundedwhen positive side of power supply is grounded (RL, RL___B0912, LL, and LM)
Check for defective components in voltage regulator or dc-to-ac converter section (MV, T series only). Check for defective components in amplifieror current control section(all models). May require replacement electronicsassembly.
Check to ensure that the burnout -protectionjumpers positioned correctly.(MV, Tseries only). Check for defective zero pot.
Check for defective components in voltage regulator section. Check for defective span pot.
4-2
REPAIR
Maintenance and Troubleshooting
Exposure to hazardous substances can cause death or serious injury. If a hazardous substance is identified, a Material Safety Data Sheet (MSDS), required by law to be availableto people exposed to specific hazardous substances, must be included with the returned materials.
In case of a failure, particularly one in which the transmitter’s output goes to one extreme and stays there, the first step is to determine whether the fault lies with the sensor(s) or the transmitter. Although only a careful calibration can determine sensor shifts, catastrophic failures (such as an open or shorted sensor element) can be checked with an ohmmeter at the time the transmitter is disconnected from the sensor(s).
NOTE
The Resistance vs. Temperature and Millivolt vs. Temperature tables for the standard Model 444 sensor input types are presented in Table B-1 in Appendix
B Temperature Sensor Reference Information.
RTD Test: A platinum RTD with an ice-point (°C) resistance of 100 ohms should
read approximately as shown in Table B-1 at other temperatures. The resistance between the two leads on the same side of the sensing element should be low, a few ohms at most. Resistance between any of the RTD leads and the sensor sheath should be high (1 megohm or greater).
Thermocouple Test: Thermocouple resistance should be low (10 ohms or less for short runs of heavy wire). For longer runs of extension wire, resistance will be roughly ten times the resistance of copper wire of the same diameter. If the sensor and receiving equipment are functioning properly, the transmitter will probably require repair.
The transmitter is designed for easy replacement of its plug-in, modular circuit boards. A malfunction can be most easily isolated by substituting boards one at a time until the unit functions properly.
It is recommended that customers return defective circuit boards to Rosemount Inc. for repair (see “RETURN OF MATERIALS” on page 4-6). This ensures that replacement parts meet the design criteria for the board and that the malfunctioning board is completely checked and repaired.
Rosemount Inc. offers a circuit board repair/replacement program through its many service centers. Please contact your Rosemount field sales office for price and delivery information.
4-3
Rosemount Model 444 Alphaline Temperature Transmitters
Disassembly Procedure
Explosion can cause death or serious injury. Do not remove the instrument cover in explosive atmospheres when the circuit is alive.
High voltage that may be present on leads can cause electrical shock. Makesure all powerto the transmitter is off before wiring.
NOTE
The numbers in parentheses refer to parts shown in the Illustrated Parts List, Table 5-7, in Section 5 Specifications and Reference Data.
1. Terminal blocks for making all field wiring electrical connections are located in a compartment identified as “Terminal Side” on the nameplate. The sensor terminals, power supply and signal-test terminals, as well as the zero and span adjustments are accessible by removing the Electronics Housing Cover (2) from the terminal side. The terminals are permanently attached to the housing and must not be removed.
2. The transmitter electronics Circuit Board Assembly (5, 6, 7) is located in a separate compartment, identified as “Circuit Side” on the nameplate. Make sure power is off. Then remove Circuit Side Cover (2).
NOTE
On the standard RTD input (Model 444RL), fast turn-on (Model 444RL ___B0912), and low-power (Models 444LL and LM) units, boards 5 and 6 are integrated into one board.
3. Remove the three Circuit Board Assembly Screws (4).
4. Push equally on the zero and span adjustment pot shafts from the terminal side. This will allow you to grasp and remove the Circuit Board Assembly.
5. If troubleshooting is required, it is best to keep the Circuit Board Assembly together for initial evaluation. Otherwise, the board assembly may be disassembled by grasping the Output Board (5) around its circumference and pulling it gently and evenly away from the other two boards. Remove the Amplifier Board (6) in the same manner. Take care not to bend the interconnection pins.
6. The adjustment pot shafts are sealed by two small O-rings (1D). Remove, if necessary, by taking off the O-ring Retainer Plate (1B), which is held in place by two small screws (1C).
7. The Nameplate (1E), Hazardous Service Certification Label (9), and Instrument Tag (8) are held in place with Drive Screws (1F). Remove any of these by gripping the head of the Drive Screw with pliers and carefully turning counter-clockwise.
4-4
Reassembly Procedure
Maintenance and Troubleshooting
Explosions can cause death or serious injury. Both transmitter covers must be fully engaged to meet explosion­proof requirements.
1. Inspect all O-rings (1D,3) and replace if necessary. Lightly grease new O­rings with silicone grease to guarantee an adequate seal.
2. If the O-ring Retainer Plate (1B) has been removed, be sure the correct side is facing outward. The resistor symbol should be visible on RTD Transmitters, while a thermocouple symbol should be visible on Thermocouple or Millivolt transmitters.
3. Inspect threaded connections on the housing and covers to make sure a minimum of five undamaged threads will be fully engaged. If the threads are shiny, apply a thin layer of molybdenum disulphide thread coating (such as Moly-Kote) to prevent galling of the aluminum threads.
4. If the Range Board (7A) requires a Coarse Zero Jumper (7B) check to make sure it is in the correct location for the desired temperature range. See Table 3-1, on page 3-3 or Table 3-2, on page 3-6.
5. Orient the Range Board (7A) and Amplifier Board (6) as shown in the Illustrated Parts List. Taking care not to bend the pins, plug the Amplifier Board into the Range Board. Press together until all three standoffs on the Amplifier Board rest against the Range Board.
6. Orient the Output Board (5) so its standoffs line up with the standoffs of the Amplifier/Range Board combination. Carefully and evenly, plug the output board into the Amplifier/Range Board. Take care not to bend the pins. Press together until all three standoffs on the Output Board rest against the Amplifier Board.
7. The circuit board assembly may be bench-tested, or calibrated outside the housing through the use of Test Terminal Strips (11 and 12). See the discussion of Calibration and Troubleshooting in this section.
8. If the circuit board assembly has been calibrated outside the housing, be very careful to ensure the zero and span adjustment pots are not moved while inserting the circuit board assembly into the housing.
9. Orient the circuit board assembly so the pot shafts line up with the pot holes in the housing.
10. Insert the circuit board assembly firmly into the housing.
11. Replace the three Circuit Board Assembly Screws (4).
12. Replace the transmitter covers (2). Tighten the covers hand-tight.
Interchangeability of Parts
Use onlythe procedures and new parts specifically referenced in this manual. Unauthorized procedures or parts can affect product performance and the output signal used to control a process, and may render the instrument dangerous. Direct any questions concerning these procedures or parts to Rosemount Inc.
4-5
Rosemount Model 444 Alphaline Temperature Transmitters
Mechanical Parts • All mechanical hardware is interchangeable among units without regard
to model numbers.
• Nameplates are interchangeable only among units that share the same input types (i.e. RTD, Thermocouple, or Millivolt).
Electrical Parts • Amplifier Board: Interchangeable among Models 444T and 444M.
• Output Board: 444T series (TE, TJ, TK, TT, TR, TS) and 444MV share a common output board.
• Range Board: Interchangeable among units of the same input code (e.g., 444RL1).
Burnout Protection Adjustments
REPAIR AND WARRANTY SERVICE
The Model 444T series (TE, TJ, TK, TT, TR, TS) and Model 444MV have a resistor network that drives the output either upscale or downscale if an open occurs in the input circuit. This option is identified in the model number. To convert from upscale to downscale, disassemble the circuit board assembly and remove R3 (22 meg, ¼WCC resistor) from the range board. To convert from downscale to upscale, replace R3. If no burnout protection is desired (as in some instances where the source has a high input impedance), remove both R2 and R3. See Figure 3-3, on page 3-6.
Model 444RL has a jumper on the range board to select burnout protection. Placing the jumper in the “U” position will cause the output to be driven upscale if the RTD opens. If the jumper is in the “D” position the output will be driven downscale. See Figure 3-1, on page 3-3. Models 444LL and 444LM have inherent upscale burnout protection that cannot be changed. Model 444RL also has a jumper on the range board to specify mA output at the transmitter test terminals. Setting the jumper at the “V” position produces a 40–200 mV output at the test terminals. Setting the jumper at the “A” position produces a 4–20 mA output at the test terminals. See Figure 3-1, on page 3-4.
Repair and warranty service is available through the Rosemount Regional Service Centers. Submit damage claims directly to the carrier.
RETURN OF MATERIALS
4-6
Exposure to hazardous substances can cause death or serious injury. If a hazardous substance is identified, a Material Safety Data Sheet (MSDS), required by law to be availableto people exposed to specific hazardous substances, must be included with the returned materials.
To expedite the return process, call the Rosemount North American Response Center toll-free at 800-654-RSMT (7768). This center, available 24 hours a day, will assist you with any needed information or materials.
The center will ask for product model and serial numbers, and will provide a Return Material Authorization (RMA) number. The center will also ask for the name of the process material the product was last exposed to.
The Rosemount North American Response Center will detail the additional information and procedures necessary to return goods exposed to hazardous substances.
Section
5 Specifications
and Reference Data
FUNCTIONAL SPECIFICATIONS
Inputs
Models 444RL, LL, and LM
100 V R
Model 444T
Thermocouple types E,J,K,T,R, and S per NIST (grounded or ungrounded).
Model 444MV
Millivolt input (grounded or ungrounded) source impedance less than 100 V.
R-numbers, specials
Special inputs other than standards, consult factory.
Spans
RTD
Platinum 45 to 135 °F (25 to75 °C).
Copper 180 to 540 °F (100 to 300 °C). Nickel 45 to 360 °F (25 to 200 °C).
Thermocouples
Type J, K, E, T 180 to 540 °F (100 to 300 °C). Type J 504 to 1458 °F (280 to 810 °C). Type K, E 504 to 1510 °F (280 to 840 °C). Type K 845 to 2540 °F (470 to 1410 °C). Type R, S 1467 to 3000 °F (815 to 1670 °C).
Millivolt
5 to 15 mV. 15 to 45 mV.
Outputs
Linear with temperature for RTD inputs. Linear with millivolt input signal for thermocouple or millivolt inputs; thermocouple and millivolt models input/output isolated to 500 V dc.
Models 444RL, T, MV
4–20 mA.
Model 444LL
0.8–3.2 V dc.
Model 444LM
1.0–5.0 V dc.
platinum RTD per IEC 751.
0
125 to 380 °F (70 to 210 °C). 360 to 1080 °F (200 to 600 °C).
5-1
Rosemount Model 444 Alphaline Temperature Transmitters
Output Limits (approximate)
Models 444RL, T, MV
Low: 3.9 mA dc. High: 30.0 mA dc.
Model 444LL
Low: 0.1 V dc. High: 4.2 V dc.
Model 444LM
Low: 0.125 V dc. High: 6.2 V dc.
Power Supply
Models 444RL, T, and MV
12 to 45 V dc at terminals of transmitter.
Model 444LL
5 to 12 V dc (overvoltage protected to 24 V dc) max current = 1.5 mA.
Model 444LM
8 to 12 V dc (overvoltage protected to 24 V dc) max current = 2.0 mA.
Load Limits
Models 444RL, T, and MV
4–20 mA.
1650 1500
1000
500
Load (Ohms)
0
12 20 30 40
Maximum Load = 50 3 (Supply Voltage – 12)
4–20 mA dc
Voltage
Too Low
Operating
Region
Power Supply (V dc)
Span and Zero
Continuously adjustable, as defined in the ordering table. Adjustments are accessible from the terminal side of the transmitter housing.
Tran smitter Temperature Limits
–13 to 185 °F (–25 °C to 85 °C), transmitter operates within specifications. –40 to 212 °F (–40 °C to 100 °C), transmitter operates without damage. –58 to 248 °F (–50 °C to 120 °C), storage. –13 to 149 °F (–25 °C to 65 °C), transmitter operates within specifications for
meter option.
Loss of Input
Upscale burnout indication standard for RTD inputs, downscale burnout indication optional. Upscale burnout indication standard for thermocouple and millivolt inputs; downscale burnout indication or no indication optional.
Turn-on Time
2 seconds. No warm-up required.
5-2
Specifications and Reference Data
PERFORMANCE SPECIFICATIONS
Accuracy
±0.2% of calibrated span (or, for thermocouple and millivolt inputs, ±0.02 millivolts, whichever is greater). ±0.5% for copper, nickel, and isolated RTD inputs, 0.1% for differential RTD inputs. Includes combined effects of transmitter repeatability, hysteresis, linearity (conformity instead of linearity for thermocouple input), and adjustment resolution. Does not include sensor error.
Stability
±0.2% of calibrated span for six months.
Ambient Temperature Effect
Errors for 50 °F (28 °C) change in ambient temperature.
RTD Inputs
Zero: ±0.17 °C,
plus
Span: ±0.22%,
plus
Elevation/Suppression: ±0.083% of base temperature in °C.
T/C Inputs (Includes Effect of Cold Junction)
Zero: ±1.38 °C,
plus
Span: ±0.28% of span,
plus
Elevation/Suppression: ±0.11% of base temperature in °C.
Millivolt Inputs
Zero: ±0.038 mV,
plus
Span: ±0.28% of span,
plus
Elevation/Suppression: ±0.11% of base input in mV.
Input Impedance (Thermocouple and mV Inputs)
More than 1 megohm—burnout resistors disconnected.
Power Supply Effect
±0.005% per volt.
Load Effect
No load effect other than the change in voltage supplied to the transmitter.
Vibration Effect
±0.05% of span per g to 200 Hz in any axis for 3 g’s up to 33 Hz, 2 g’s from 33 to 70 Hz and 1 g from 70 to 200 Hz.
Mounting Position Effect
None.
5-3
Rosemount Model 444 Alphaline Temperature Transmitters
PHYSICAL SPECIFICATIONS
Materials of Construction
Electronics Housing
Low-copper aluminum. (NEMA 4X). IP 54, IP 65, IP 66.
Housing Paint
Polyurethane.
Housing O-rings
Buna-N.
Sensor and Conduit Connections
1
/2-inch conduit on electronics housing. Screw terminals and integral test jacks
compatible with miniature banana plugs (Pomona 2944, 3690 or equal).
Weight
Transmitter: 3 lb (1.4 kg). Transmitter with mounting bracket: 4 lb (1.8 kg).
Hazardous Location Certifications
Factory Mutual (FM) Approvals E5 Explosion Proof: Class I, Division 1, Groups B, C, and D. Dust Ignition
Proof: Class II, Division 1, Groups E, F, and G; Class III, Division 1 hazardous locations. Indoor and outdoor use. NEMA Enclosure Type 4X.
Refer to Factory Mutual Explosion Proof Drawing 00444-0261. I5 Intrinsic Safety: Class I, Division 1, Groups A, B, C, and D; Class II,
Division 1, Groups E, F, and G; Class III, Division 1 hazardous locations; Intrinsically safe system only when applying Table 5-1 entity parameters. Nonincendive: Class I, Division 2, Groups A, B, C, and D; Indoor and outdoor use. NEMA Enclosure Type 4X.
Refer to Factory Mutual Intrinsic Safety Drawing 01151-0214. K5 Combination of E5 and I5.
TABLE 5-1. FM Entity Parameters.
FM Approved for
Class I, II, III,
Division 1, Groups
A B CthruG
•• •
NA NA
Model 444
Parameters
V
= 40 V
MAX
I
=165mA
MAX
Ci= 0.044µF
=0
L
i
=40V
V
MAX
I
= 225 mA
MAX
C
= 0.044µF
i
L
=0
i
Associated
Equipment
Parameters
VOCor V I
SC
C L
VOCor V I
SC
C L
40 V
T
or IT ≤ 165 mA
> 0.044µF
A
>0
A
40 V
T
or IT ≤ 225 mA
> 0.044µF
A
>0
A
Canadian Standards Association (CSA) Approvals E6 Explosion Proof: Class I, Division 1, Groups C and D; Dust-ignitionproof
Class II, Division 1, Groups E, F, and G; Class III, Division 1 hazardous locations; Suitable for Class I, Division 2, Groups A, B, C, and D; CSA Enclosure Type 4X.
I6 Intrinsic Safety: Class I, Division 1, Groups A, B, C, and D. Intrinsically
safe system only when applying Table 5-2 parametric parameters. Temperature code T2D. CSA Enclosure Type 4X.
Refer to CSA Intrinsic Safety Drawing 00444-0034. C6 Combination of E6 and I6.
5-4
TABLE 5-2. CSA P arametric Parameters.
Specifications and Reference Data
CSA Approved for
Barrier Manufacturer/Model
A B C D
Any CSA approved zenerbarrier
Class I ,
Division 1 Groups
••••
30 V,330 V or28 V,300 V or22 V,180 V
Foxboro Converters 2AI-I2V-CGB 2AI-I3V-CGB 2AS-I2I-CGB 2AS-I3I-CGB 3AD-I3IA CS-E/CGB-A 3A2-I2D CS-E/CGB-A, 3A2-I3D CS-E/CGB-A 3A4-I2DA CS-E/CGB-A 3F4-I2DA1CS-E/CGB-A
Any CSA approved zenerbarrier
NA•••
NA NA
30 V,150V
Standards Association of Australia (SAA) Certifications E7 Flameproof:
Ex d IIB+H2 T6 Class I, Zone 1.
DIP T6 Class II.
Special Conditions for Safe Use (“X”):
For transmitters having NPT or PG cable entry thread, an appropriate flameproof thread adaptor shall be used to facilitate application of certified flameproof cable glands. Only SAA-certified flameproof temperature sensors shall be used with the Model 444 Temperature Transmitter if fitted directly into the tapped entry of the enclosure.
I7 Intrinsic Safety:
Ex ia IIC T6 (T Ex ia IIC T5 (T
= 40 °C)
amb
= 70 °C)
amb
Class I, Zone 0 Entity Parameters:
= 30 V
U
i
= 200 mA
I
i
= 1.0 W
P
i
= 0.024 mF
C
i
= 0
L
i
Refer to SAA Intrinsic Safety Drawing 00444-0264. Special Conditions for Safe Use (“X”):
The equipment has been assessed to the “Entity” concept and upon installation the barrier/entity parameters must be taken into account.
N7 Type “n”:
Ex n IIC T6 (T Ex n IIC T5 (T
= 40 °C)
amb
= 70 °C)
amb
Class I, Zone 2
Special Conditions for Safe Use (“X”):
The equipment must be connected to a supply voltage which does not exceed the rated voltage. The enclosure end caps must be correctly fitted while the equipment is powered.
5-5
Rosemount Model 444 Alphaline Temperature Transmitters
Centro Elettrotecnico Sperimentale Italiano (CESI/CENELEC) Certifications E8 Flameproof:
EEx d IIC T6
I8 Intrinsic Safety:
Model 444RL EEx ia IIC T6 (T EEx ia IIC T5 (T EEx ia IIC T4 (T
= 40 °C) [Pi = 0.75 W]
amb
= 55 °C) [Pi = 1.0 W]
amb
= 80 °C) [Pi = 1.0 W]
amb
Model 444T_ & MV EEx ib IIB T6 (T EEx ib IIB T5 (T EEx ib IIB T4 (T
= 40 °C) [Pi = 0.75 W]
amb
= 55 °C) [Pi = 1.0 W]
amb
= 80 °C) [Pi = 1.0 W]
amb
Entity Parameters:
= 30 V dc
V
i
= 200 mA
I
i
= 0.75 W (T6)
P
i
= 1.0 W (T5)
P
i
= 1.0 W (T4)
P
i
= 0.024 mF (Model 444RL)
C
i
= 0.006 mF (Model 444T_ & MV)
C
i
= 0.
L
i
Special Conditions for Safe Use (“X”):
If the temperature sensor connected to the input circuit does not tolerate an alternating tension of 500V for 60 seconds, the certified transducers must be powered by galvanically-insulated equipment. Model 444RL temperature transducers must be connected to associated electronic equipment certified to EN 50.014/EN 50.020 standards.
British Approvals Service for Electrical Equipment Flammable Atmospheres (BASEEFA) Type N Certification N1 Ex N II T5
Maximum Voltage: 35 V dc Maximum Current: 30 mA dc
An appropriate stainless steel tag will be supplied according to the certification option selected.
5-6
LCD METER SPECIFICATIONS (OPTIONS D AND E)
Specifications and Reference Data
Functional Specifications
Performance Specifications
Configuration
4 mA point limits: –999 to 1000. Span limits: 0200 to 9999.
The sum of the 4 mA point and span must not exceed 9999. Adjustments are made using non-interactive zero and span buttons.
Temperature Limits
Storage: –40 to 85 °C (–40 to 185 °F). Operating: –20 to 70 °C (–4 to 158 °F). Between –40 to –20 °C (–40 to –4 °F) loop is intact and the meter is not damaged.
Humidity Limitation
0 to 95% non-condensing relative humidity.
Update Period
750 ms.
Response Time
Responds to changes in input within a maximum of two update periods. If the filter is activated, then the display responds to the change within nine update periods.
Digital Display Resolution
0.05% of calibrated range ±1 digit.
Analog Bar Graph Resolution
0.05% of calibrated range.
Indication Accuracy
0.25% of calibrated range ±1 digit.
Stability
Over Time: 0.1% of calibrated range ±1 digit per six months.
Temperature Effect
0.01% of calibrated range per °C on zero.
0.02% of calibrated range per °C on span over the operating temperature range.
Power Interrupt
All calibration constants are stored in EEPROM memory and are not affected by power loss.
Failure Mode
LCD meter failure will not affect transmitter operation.
Under/Over Range Indication
Input current < 3.5 mA: Display blank. Input current > 22.0 mA: Display flashes 112.5% of full scale value or 9999,
whichever is less.
Physical Specification
Meter Size
2¼-inch diameter face with four, ½-inch high characters.
5-7
Rosemount Model 444 Alphaline Temperature Transmitters
ANALOG METER SPECIFICATIONS (OPTIONS B AND C)
Functional Specifications
Performance Specifications
Physical Specification
Meter Indication
0 to 100% linear scale. Special optional ranges.
Temperature Limits
–40 to 65 °C (–40 to 150 °F).
Humidity Limits
0 to 100% relative humidity.
Zero Adjustment
Adjustment screw on face of meter.
Indication Accuracy
±2% of calibrated span.
Temperature Effect
Less than 2% of full scale at any point within the temperature limits.
Meter Size
2¼-inch diameter face with 2-inch long scale.
5-8
Specifications and Reference Data
TABLE 5-3. 444 Model Structure.
Model Product Description
444 Alphaline Temperature Transmitter
Temperature Span
Minimum Maximum
–58 °F (–50 °C) 300 °F (150 °C) –58 °F (–50 °C) 300 °F (150 °C) –58 °F (–50 °C) 300 °F (150 °C)
–58 °F (–50 °C) 300 °F (150 °C) –58 °F (–50 °C) 930 °F (500 °C) –58 °F (–50 °C) 300 °F (150 °C) –58 °F (–50 °C) 930 °F (500 °C) –58 °F (–50 °C) 930 °F (500 °C) –58 °F (–50 °C) 300 °F (150 °C) –58 °F (–50 °C) 930 °F (500 °C) –58 °F (–50 °C) 300 °F (150 °C) 0 °F (–18 °C) 1500 °F (815 °C) 0 °F (–18 °C) 1500 °F (815 °C)
–2mV 8mV –2mV 20mV
0
Millivolt Millivolt
(1)
45 °F (25 °C) 135 °F (75 °C) 125 °F (70 °C) 380 °F (210 °C) 360 °F (200 °C) 1080 °F (600 °C)
180°F (100°C) 540 °F (300 °C) 504 °F (280 °C) 1458 °F(810 °C) 180 °F (100 °C) 540 °F (300 °C) 504 °F (280 °C) 1510 °F(840 °C) 845 °F (470 °C) 2540 °F(1410°C) 180 °F (100 °C) 540 °F (300 °C) 504 °F (280 °C) 1510 °F(840 °C) 180 °F (100 °C) 540 °F (300 °C) 1467 °F (815 °C) 3000 °F (1670 °C) 1467 °F (815 °C) 3000 °F (1670 °C)
5mV 15mV 15 mV 45 mV
Code Input Type
RL1 RL2 RL3
RL9 Special Input, Range or Accuracy (minimum span 3 V)
TJ1 TJ2 TK1 TK2 TK3 TE1 TE2
TT1 TR1 TS1
T_9 Special Range or Accuracy (minimum span 3 mV, maximum span 100 mV) MV1
MV2 MV9 Special Range or Accuracy (minimum span 3 mV, maximum span 100 mV)
Platinum RTD
100 V R linearized output
Thermocouple
Type J Type J Type K Type K Type K Type E Type E Type T Type R Type S
Code Loss of Input Indication
U
Upscale (standard for all input types)
D
Downscale
N
None (not available for platinum RTD inputs)
Code Calibration
1
Trim to IEC 751 Class B (RTD) or NIST Curve (thermocouple)
2
Trim to Specific Model 68/78/88 Calibration Schedule
3
Trim to Other Nominal Curve (customer must specify separately) (Note: Millivolt input must use Code 3.)
Code Meter Options
A
None
B
Integral Analog Meter, Special Scale (same as calibrated range)
C
Integral Analog Meter, 0–100% Scale
D
Integral LCD Meter, 0–100% Scale Integral LCD Meter, Special Scale (specify range, mode, and engineering units)
E
(2)
Code Mounting Bracket
1
None
2
Mounting Bracketfor 2-inch Pipe or Surface Mounting
Code Hazardous Area Certifications
NA E5
E6
E7
N7 E8
N1
No Certification Required FM Explosion-Proof Approval
I5
FM Intrinsic Safety and Non-incendive Approval CSA Explosion-Proof Approval
I6
CSA Intrinsic Safety Approval (444RL and 444T only) SAA Explosion-Proof Certification
I7
SAA Intrinsic SafetyCertification SAA Non-incendive Certification CESI Explosion-Proof Certification (When ordering a transmitter with this option, place a W before the model number: W444.)
I8
CESI Intrinsic Safety Certification (When ordering a transmitter with this option, place a W before the model number: W444.) BASEEFANon-incendive Certification (When ordering a transmitter with thisoption, place a T before the modelnumber: T444.)
Code Options
Q4 A1 A2
2-Point Calibration Certificate One (1) ½ NPT to M20 (CM 20) SST Thread Adapter One (1) ½ NPT to M20 (CM 20) SST Thread Adapter
Code Special
RXXXX Unique Range (use with RL9, T_9, and MV9 inputs)
Typical Model N umber: 0444 RL3 U 1 A 2 E5 Q4
(1) Refer to Temperature Sensors, Assemblies , and Accessories Product Data Sheet, Rosemount pub. no. 00813-0100-2654 for information
about Rosemount sensor assemblies.
(2) LCD Meters are only available with RL1, RL2, or RL3. (may be reconfigured in the field.)
Base Temperature
Minimum Maximum
(1)
Upper Range Limit
435 °F (225 °C) 680 °F (360 °C)
1380 °F (750 °C)
840 °F (450 °C) 1400 °F (760 °C) 840 °F (450 °C) 2440 °F (1340 °C) 2500 °F (1370 °C) 840 °F (450 °C) 1830 °F (1000 °C) 750 °F (400 °C) 3200 °F (1760 °C) 3200 °F (1760 °C)
23 mV 65 mV
5-9
Rosemount Model 444 Alphaline Temperature Transmitters
TABLE 5-4. 444RL Fast Turn-on Model Structure.
Model Product Description
444 Alphaline Fast Turn-OnTemperature Transmitter
Code Input Type
RL1 RL2 RL3
Platinum RTD
100 V R linearized output linearized output
(1)
0
Code Loss of Input Indication
U Upscale (standard for all input types)
Code Calibration
1 2 3
Trim to IEC 751 Curve (RTD) Trim to Specific Model 68/78/88CalibrationSchedule Trim to Other Nominal Curve (customer must specifyseparately)
Code Meter Options
A B C
None Integral Analog Meter, Special Scale (must specify range, mode, andengineering units) Integral Analog Meter, 0–100% Scale
Code Mounting Bracket
1 2
None Mounting Bracket for 2-inch Pipe or Surface Mounting
Code Hazardous Area Certifications
NA
E5 E6
I6
Code Options
Q4
A1 A2
No Certification Required FM Explosion-Proof Approval CSA Explosion-Proof Approval CSA Intrinsic Safety Approval
2-Point Calibration Certificate One (1) ½ NPT to M20 (CM 20) SST Thread Adapter One (1) ½ NPT to M20 (CM 20) SST Thread Adapter
Code Special
B0912 Fast Turn-on Electronics
Typical Model Number: 0444 RL3 U 1 A 2 NA B0912
Temperature Span
Minimum Maximum
45 °F (25 °C) 135 °F (75 °C) 125 °F (70 °C) 380°F (210 °C) 360 °F (200 °C) 1080 °F (600 °C)
Base Temperature (4mA Point)
Minimum Maximum
–58°F(–50°C) 300°F(150°C) –58°F(–50°C) 300°F(150°C) –58°F(–50°C) 300°F(150°C)
Upper Range
Limit
435°F(225°C) 680°F(360°C) 1380 °F (750°C)
(1) Refer to Temperature Sensors, Assemblies , and Accessories Product Data Sheet, Rosemount pub. no. 00813-0100-2654 for information
about Rosemount sensor assemblies.
5-10
TABLE 5-5. 444LL, LM Model Structure.
Model Product Description
444 AlphalineLow-Power Temperature Transmitter
Code Input Type
LL1
LM1
Platinum RTD
100 V R Linearized Output
Code Loss of Input Indication
U Upscale
Code Calibration
1 2
Trim to IEC 751 Curve (RTD) Trim to Specific Model 68/78/88Calibration Schedule
Code Meter Options
ANone
Code Mounting Bracket
1 2
None Mounting Bracket for 2-inch Pipeor Surface Mounting
Code Options
NA
E5 E6
I6
No Cer tification Required FM Explosion-Proof Approval CSA Explosion-Proof Approval CSA Intrinsic Safety Approval (444RL and 444T only)
Typical Model Number: 0444 LM1 U 1 A 2 E5
(1)
Minimum Maximum
0
75 °F (42 °C) 150 °F (83 °C) 75 °F (42 °C) 150 °F (83 °C)
Temperature Span
Specifications and Reference Data
Base Temperature (4mA Point)
Minimum Maximum
–25°F(–32°C) 50°F(10°C) –25°F(–32°C) 50°F(10°C)
(1) Refer to Temperature Sensors, Assemblies , and Accessories Product Data Sheet, Rosemount pub. no. 00813-0100-2654 for information
about Rosemount sensor assemblies.
TABLE 5-6. Meter Kits.
Part Description Part Number Spares Category
(2) (3)
(4)
(2)
(2)
00444-0194-0001 00444-0194-0005 01151-0687-0003 01151-0744-0003 00444-0194-0004 01151-1300-1000 00444-0194-0002 00444-0194-0003 01151-2610-0001
scale: XX to XX
C C C C C C C C C C
40–200 mV AnalogMeter Kit 40–200 mV AnalogMeter Kit (CENELECand I.S. approved) 40–200 mV AnalogMeter 40–200 mV AnalogMeter (CENELEC approved) 4–20 mA LCD Meter Kit 4–20 mA LCD Meter Mounting Hardware and CoverKit Mounting Hardware Kit Cover Assembly Meter with SpecialScale
(1) Normally,no spare parts required for C classification. (2) Meter kit includes meter, mounting hardware, and cover assembly. (3) LCD Meters are only available with RL1, RL2, and RL3 (may be recon figured in the field). (4) For Special Scale Meters, order appropriate meter and indicate spec ial scale desired. Mounting hardware and cover
assembly must be ordered separately
(1)
5-11
Rosemount Model 444 Alphaline Temperature Transmitters
TABLE 5-7. Standard Parts List.
Item
Number
1 444R
2
j
3
j
1D
j
4
j
3
j
5
j
6
j
Model Part Description Part Number
444T 444M
ALL Housing Cover—QuantityOne 90032-0240-0003 C ALL Housing Cover O-ring 01151-0033-0003 C ALL Adjustment PotentiometerO-ring (pkg. of12) 00444-0030-0001 B ALL Electronics Assembly Screws, 6–32 317/8 (pkg. of 12) 00444-0031-0001 C ALL CoverO-rings. (pkg. of 12) 00444-0030-0001 B ALL A/O Board—RTD Input (2board set, F igure 5-1)
ALL Amplifier Board—All Inputs (3 board set, Figure5-2) 00444-0007-0003 A
Housing Kit—RTD Input Housing Kit—Thermocouple Input Housing kit—Millivolt Input Electronics housing kit consists of:
1 each electronics housing (item
1eachO-ringretainerplate
2 each 6–32 3 2 each pot O-rings (item 1 each nameplate(item 2 each #4–403
Output Board—T/C, mV Input(3 board set,Figure 5-2)
3
/16 retainerplate screws (item j1C)
3
/16 screws (item j1F)
j
1A
j
)
1B
j
)
(item
1D
j
)
1E
)
00444-0028-0001 00444-0028-0002 00444-0028-0003
00444-0228-0001 00444-0015-0001
j7A 444R Range Board—RTD Inputs
25 to 75 °C Span (InputCode RL1) 70 to 210°C Span (InputCode RL2) 200 to 600 °C Span (Input Code RL3) Special Range RL9 (requires 3 board set – c onsult factory)
00444-0221-0001 00444-0221-0002 00444-0221-0003
00444-0003-XXXX
j7A 444T Range Board—Thermocouple Inputs
Type J,100 to 300°C Span (InputCode TJ1) Type J,280 to 840°C Span (InputCode TJ2) Type K, 100 to 300 °C Span (Input Code TK1) Type K, 280 to 840 °C Span (Input Code TK2) Type K, 470 to 1410 °C Span (InputCode TK3) Type E, 100 to 300 °C Span (Input Code TE1) Type E, 280 to 840 °C Span (Input Code TE2) Type T, 100to 300 °CSpan (Input CodeTT1) Type R, 815 to 1670 °C Span(Input Code T R1) Type S,815 to 1670 °C Span (InputCode TS1) Special Range T-9 Kit (consult factory)
(2)
00444-0262-0002 00444-0262-0006 00444-0262-0003 00444-0262-0007 00444-0262-0008 00444-0262-0001 00444-0262-0005 00444-0262-0004 00444-0262-0009 00444-0262-0010
00444-0013-XXXX
j7A 444M Range Board—Millivolt Inputs
5to15mV(InputCodeMV1) 15 to 45 °C Span (InputCode MV2) Special Range MV9 (consult factory)
00444-0023-0001 00444-0023-0002
00444-0023-XXXX
j7B 444R Range Board Zero Jumpers (pkg. of 12) 00444-0036-0001 A
8
j j
10 ALL Mounting Bracket Kit
11
j
12
j
(1) Spares Categories (2) Thermocouple Range BoardKit Consists of:
Code A – Recommended 1 spare part per 25 transmitters. 1 each T/C Range Board. Code B – Recommended 1 spare part per 50 transmitters. 1 each External CJC Block. Code C – None normally required. Code D – Not used on transmitters, but recommendedfor repair actiivities.
ALL Instrument Tag (optional) 01151-0148 C
9
ALL Certification Label C
1 each mountingbracket(item ) 4 each ¼–203 ½bolts(item ) 2 each U-bolts(item ) 4each¼washers(item )
5
/16 –18 nuts(item )
4each 2 each washerplates (item )
ALL Terminal Block,Three-position
Terminal Block,Four-position
10C
10A
10B
10D
10E
10F
00444-0022-0001 C
C10448-0106 C10448-0109
Spares
Category
C C C
A A
A A A A
A A A A A A A A A A A
A A A
D D
(1)
5-12
TABLE 5-8. Parts List fo r Model W444 Transmitter (CESI/CENELEC Approval).
Specifications and Reference Data
Item
Number
1 Housing Kit
j j
1D
j j j
j
Part Description Part No. Spares Categor y
(2)
— RTD Input
(2)
Housing Kit Housing kit
2
Housing Cover—QuantityOne 90032-0240-0003 C
3
Housing Cover O-rings 01151-0033-0003 B Adjustment PotentiometerO-rings 00444-0030-0001 B
4
Electronics Assembly Screws 00444-0031-0001 C
5
A/O Board—RTD Input Output Board—T/C, mV Input Output Board—RD Input
6
Amplifier Board 00444-0007-0003 A
— Thermocouple Input
(2)
— Millivolt Input
00444-0028-0001 00444-0028-0002 00444-0028-0003
00444-0228-0001 00444-0015-0002 00444-0111-0001
j7A Range Board
W444 RL1 W444 RL2 W444 RL3 W444 RL9 (consult factory) W444 TJ1 W444 TJ2 W444 TK1 W444 TK2 W444 TK3 W444 TE1 W444 TE2 W444 TT1 W444 TR1 W444 TS1 W444 T_9 (consult factory) W444 MV1 W444 MV2 W444 MV9 (consult factory)
00444-0221-0001 00444-0221-0002 00444-0221-0003
00444-0003-XXXX
00444-0262-0002 00444-0262-0006 00444-0262-0003 00444-0262-0007 00444-0262-0008 00444-0262-0001 00444-0262-0005 00444-0262-0004 00444-0262-0009 00444-0262-0010
00444-0013-XXXX
00444-0023-0001 00444-0023-0002
00444-0023-XXXX
j7B Range Board Zero Jumpers 00444-0036-0001 A
10 Mounting Bracket Kit 00444-0022-0001 C
11
j
12
j
Terminal Block,Three-position Terminal Block,Four-position
C10448-0106 C10448-0109
(1)
C C C
A A A
A A A A A A A A A A A A A A A A A A
D D
(1) Spares Categories
Code A – Recommended one spare part per 25 transmitters. Code B – Recommended one spare part per 50 transmitters. Code C – None normallyrequired.Code D – Not used on transmitters, but recommended for repairactivities (seeInstruction Manual).
(2) Housing kit includes housing, adjustment potentionmeter O-rings,O-ring retainer plateand nameplate.
5-13
Rosemount Model 444 Alphaline Temperature Transmitters
TABLE 5-9. Parts List for Model T444 Transmitter (BASEEFA Approval).
Item Number
1 Housing Kit
j j
1D
j j j
j
Part Description Part No. Spares C ategory
(2)
—RTD Input
(2)
Housing Kit Housing kit
2
Housing Cover—Quantity One 90032-0240-0003 C
3
Housing Cover O-rings 01151-0033-0003 B Adjustment PotentiometerO-rings 00444-0030-0001 B
4
Electronics Assembly Screws 00444-0031-0001 C
5
A/O Board—RTD Input Output Board—T/C, mV Input
6
Amplifier Board 00444-0007-0003 A
—Thermocouple Input
(2)
—Millivolt Input
00444-0028-0001 00444-0028-0002 00444-0028-0003
00444-0228-0001 00444-0015-0001
j7A Range Boards
T444 RL1 T444 RL2 T444 RL3 T444 RL9 (consultfactory) T444 TE1 T444 TJ1 T444 TK1 T444 TT1 T444 TE2 T444 TJ2 T444 TK2 T444 TK3 T444 TR1 T444 TS1 T444 T_9 (consultfactory) T444 MV1 T444 MV2 T444 MV9 (consultfactory)
00444-0136-0001 00444-0136-0002 00444-0136-0003
00444-0136-XXXX
00444-0273-0001 00444-0273-0002 00444-0273-0003 00444-0273-0004 00444-0273-0005 00444-0273-0006 00444-0273-0007 00444-0273-0008 00444-0273-0009 00444-0273-0010
00444-0138-XXXX
00444-0139-0001 00444-0139-0002
00444-0139-XXXX
j7B Range Board ZeroJumpers 00444-0036-0001 A
10 Mounting Bracket Kit 00444-0022-0001 C
11
j
12
j
For meter information, seepage 7.
(1) Spares Categories
Code A – Recommended one spare part per 25 transmitters. Code B – Recommended one spare part per 50 transmitters. Code C – None normally required. Code D – Not used on transmitters, but recommendedfor repair activities (see Instruction Manual).
(2) Housing kit includes housing, adjustment potentiometerO-rings, O-ring retainer plate and nameplate.
Terminal Block, Three-position Terminal Block, Four-position
C10448-0106 C10448-0109
(1)
C C C
A
A A A A A A A A A A A A A A A A A A
D D
5-14
TABLE 5-10. Parts List for Model 444 RL Fast Turn On Transmitter (RTD Input).
Specifications and Reference Data
Item Number
1 Housing Kit
j j
1D
j j j
Part Description Part No. Spares Category
(2)
2
Housing Cover—Quantity One 90032-0240-0003 C
3
Housing Cover O-rings 01151-0033-0003 B Adjustment PotentiometerO-rings 00444-0030-0001 B
4
Electronics Assembly Screws 00444-0031-0001 C
5
Output Board 444 RL B0912 00444-0126-0001 A
00444-0028-0001 C
j7A Range Boards
RL1 B0912 RL2 B0912 RL3 B0912
00444-0221-0011 00444-0221-0012 00444-0221-0013
j7B Range Board Zero Jumpers 00444-0036-0001 A
10 Mounting Bracket Kit 00444-0022-0001 C
11
j
12
j
For meter information, seepage 7.
(1) Spares Categories
Code A – Recommended one spare part per 25 transmitters. Code B – Recommended one spare part per 50 transmitters. Code C – None normally required. Code D – Not used on transmitters, but recommendedfor repair activities (see Instruction Manual).
(2) Housing kit includes housing, adjustment potentiometerO-rings, O-ring retainer plate and nameplate.
TABLE 5-11. Parts List for Model 444 LL, 444 LM Low Power Transmitter (RTD Input).
Terminal Block, Three-position Terminal Block, Four-position
C104480106 C104480109
(1)
A A A
D D
Item Number
1 Housing Kit
j j
1D
j j j
Part Description Part No. Spares C ategory
(2)
2
Housing Cover—QuantityOne 90032-0240-0003 C
3
Housing Cover O-rings 01151-0033-0003 B Adjustment PotentiometerO-rings 00444-0030-0001 B
4
Electronics Assembly Screws 00444-0031-0001 C
5
Output Board 444 LL Output Board 444 LM
00444-0028-0005 C
00444-0159-0001 00444-0159-0002
j7A Range Boards
444 LL 444 LM
00444-0157-0001 00444-0157-0002
j7B Range Board Zero Jumpers 00444-0036-0001 C
10 Mounting Bracket Kit 00444-0022-0001 C
11
j
12
j
No meter available.
(1) Spares Categories
Code A – Recommended one spare part per 25 transmitters. Code B – Recommended one spare part per 50 transmitters. Code C – None normally required. Code D – Not used on transmitters, but recommendedfor repair activities (see Instruction Manual).
(2) Housing kit includes housing, adjustment potentiometerO-rings, O-ring retainer plate and nameplate.
Terminal Block,Three-position Terminal Block,Four-position
C104480106 C104480109
(1)
A A
A A
D D
5-15
Rosemount Model 444 Alphaline Temperature Transmitters
TABLE 5-12. Parts List for Model 444 Increased Accuracy or SpecialRange Transmitters.
Item Number
1 Housing Kit
j j
1D
j j j
j
Part Description Part No. Spares Category
(2)
—RTD Input
(2)
Housing Kit Housing kit
2
Housing Cover—Quantity One 90032-0240-0003 C
3
Housing Cover O-rings 01151-0033-0003 B Adjustment PotentiometerO-rings 00444-0030-0001 B
4
Electronics Assembly Screws 00444-0031-0001 C
5
Output Board RL9 Output Board T_9 Output Board MV9
6
Amplifier Board 00444-0007-0003 A
—Thermocouple Input
(2)
—Millivolt Input
00444-0028-0001 00444-0028-0002 00444-0028-0003
00444-0005-0001 00444-0015-0001 00444-0015-0001
j7A Range Boards
RL9 (consult factory) T_9 (consult factory) MV9 (consult factory)
00444-0003-XXXX 00444-0013-XXXX 00444-0023-XXXX
j7B Range Board Zero Jumpers 00444-0036-0001 C
10 Mounting Bracket Kit 00444-0022-0001 C
11
j
12
j
For meter information, seebelow.
(1) Spares Categories
Code A – Recommended one spare part per 25 transmitters. Code B – Recommended one spare part per 50 transmitters. Code C – None normally required. Code D – Not used on transmitters, but recommendedfor repair activities.
(2) Housing kit includes housing, adjustment potentiometerO-rings, O-ring retainer plate and nameplate.
Terminal Block, Three-position Terminal Block, Four-position
C10448-0106 C10448-0109
(1)
C C C
A A A
A A A
D D
5-16
FIGURE 5-1. Transmitter Exploded View1.
2
j
3
j
j
Specifications and Reference Data
View Corresponds to the Following Models:
444RL1 444LM1 444RL2 444LL1
4
5
j
8
j
j7A
1A
j
444RL3
9
j
j7B
1F
j
10E
10F
10A
1E
j
1D
j
1B
j
10D
1C
j
10B
10C
j
3
2
j
Detail o f Item 1B. O-Ring Retainer Plate Wiring Diagrams
Reversible
Thermocouple
or Millivolt
Side
Models 444T, MV
RTD Side
Models 444RL, LL, LM
Accessory Terminal Blocks For
Bench Testing
444-1151C02A, 0071A01A, 0071B01A, 1151H02A
5-17
Rosemount Model 444 Alphaline Temperature Transmitters
FIGURE 5-2. Transmitter Exploded View2.
2
j
3
j
4
j
j
5
6
j
j7A
View Corresponds to the Following Models:
444RL9 444TK3 444TR1 444TJ1 444TE1 444TS1 444TJ2 444TE2 444MV1 444TK1 444TT1 444MV2 444TK2 444T_9 444MV9
8
j
1A
j
j
9
j7B
1F
j
j
10E
10F
10A
Detail of Remote Cold Junction
Compensation Sensor. (Model 444T)
1D
j
1B
j
1C
j
1E
10D
10B
10C
j
3
2
j
5-18
3044-0X68C01A, 444-1151B02A, 0236D01A
Section
6 Options
MOUNTING BRACKET
FIGURE 6-1. Model 444 with Optional Mounting Bracket.
PIPESTAND MOUNTING
Clearance Hole
for ¼-inch Bolt
(eight places)
NOTE Dimensions are in inches (millimeters).
The mounting bracket option provides auxiliary configurations for mounting the Model 444 transmitter. With this option you can either mount the transmitter to a 2-inch pipe or a suitable flat panel. The bracket is constructed of carbon steel with carbon steel U-bolts. See Figure 6-1 for an exploded view of the Model 444 with a mounting bracket in both configurations.
Transmitter can be Rotated 90°
Mounting Bracket
¼–20 3½-inch
Bolt (4)
5
/16 –18 U-bolt for
2-inch Pipe (2)
PANEL OR SURFACE MOUNTING
5
/16 -inch Bolts
(four required,
not furnished)
2.81 (81)
5.00 (127)
Hole for
5
/16 -inch
Bolts
(four
places)
444-1151G, 1151F04A
6-1
Rosemount Model 444 Alphaline Temperature Transmitters
LCD / ANALOG METER
LCD METER
Configuration
FIGURE 6-2.LCD Meter.
The LCD and analog meters provide local indication of the transmitter output. Both meters attach easily to the terminal side of the transmitter.
The Rosemount® LCD meter plugs directly into the Model 444 to provide a highly accurate digital display of the process variable. The meter adds no voltage drop in the 4–20 mA current loop when connected directly across the transmitter test terminals.
Configure the LCD meter to meet specific requirements by using the left and right calibration buttons located on the meter face as shown in Figure 6-2. The analog bar graph is also shown in Figure 6-2. The 20-segment bar graph is factory calibrated and represents 4–20 mA directly.
.
Analog Bar Graph
LeftConfiguration Button
Right Configuration
Retaining
Ring
Button
No calibration equipment is required to configure the LCD meter, but between 4 and 20 mA must exist in the loop in order for the meter to operate. The actual value of the current is not significant. In addition, meter configuration does not affect the transmitter/loop current. Use the following meter configuration procedure to properly configure the LCD meter.
Remove the Cover 1. Unscrew the retaining ring shown in Figure 6-2 and lift the transparent
cover off of the housing.
NOTE
The LCD meter time-out is approximately 16 seconds. If keys are not pressed within this period, the indicator reverts to reading the current signal.
LCD-001AB
6-2
Options
Position the Decimal Point and Select the Meter Function
2. Press the left and right configuration buttons simultaneously and release them immediately.
3. To move the decimal point to the desired location, press the left configuration button. Note that the decimal point wraps around.
4. To scroll through the mode options, press the right configuration button repeatedly until the desired mode is displayed. See Table 6-1.
TABLE 6-1. LCD MeterModes.
Options Relationship between Input Signal and Digital Display
Lin
LinF Srt SrtF
Square root function only relatesto the digitaldisplay. The bar graphoutput remains linear with the current signal.
Square root response
The digital display willbe proportional to the square rootof the inputcurrent where 4mA=0 and 20 mA=1.0, scaled per the calibration procedure. The transition point from linear to square root is at 25% of full scale flow.
Filter response operates upon“present input” and “input received in the previous five second interval” in the followingmanner:
Display = (0.75 3 previousinput) + (0.25 3 presentinput) This relationship is maintained provided that theprevious reading minus thepresent reading is less than 25% of full scale.
Linear
Linear with five-second filter Square root Square root with five-second filter
Store the Information 5. Press both configuration buttons simultaneously for two seconds. Note
that the meter displays “----” for approximately 7.5 seconds while the information is being stored.
Set the Display Equivalent toa4mASignal
6. Press the left button for two seconds.
7. To decrement the display numbers, press the left configuration button and to increment the numbers, press the right configuration button. Set the numbers between –999 and 1000.
8. To store the information, press both configuration buttons simultaneously for two seconds.
Set the Display Equivalent to a 20 mA Signal
9. Press the right button for two seconds.
10. To decrement the display numbers, press the left configuration button on the display and to increment the numbers, press the right configuration button. Set the numbers between –999 and 9999. The sum of the 4 mA point and the span must not exceed 9999.
11. To store the information, press both configuration buttons simultaneously for two seconds. The LCD meter is now configured.
Replace the Cover 12. Make sure the rubber gasket is seated properly, replace the transparent
cover, and replace the retaining ring.
6-3
Rosemount Model 444 Alphaline Temperature Transmitters
LCD Meter Assembly
FIGURE 6-3.LCD Meter Exploded View
Figure 6-3 shows the mounting hardware required to properly install the LCD meter on a Model 444 transmitter. This mounting hardware may also be used with the Rosemount universal (analog) meter.
Mounting Screw into Housing
Strap Washer
Retaining Straps
Mounting Screws
Mounting Screw into
Mounting Plate
TerminalScrews (Mount
into Transmitter “Test”
Terminal Block)
Mounting Plate
LCD Meter Specifications
Functional Specifications
Meter (Meter may be rotated in 90 degree increments)
Cover Bushing
CoverFoamSpacer
Input Signal
4–20 mA dc.
Meter Indication
4-digit LCD showing –999 to 9999. A 20-segment bar graph directly represents the 4–20 mA current.
Scaling/Calibration
4 mA Point Limits: –999 to 1000. Span limits: 0200 to 9999.
The sum of the 4 mA point and span must not exceed 9999. Adjustments are made using non-interactive zero and span buttons.
Hazardous Locations Certifications
Approved for use with Rosemount Models 444, 751, 1135, 1144, and 1151.
751-0264B
6-4
Options
Overload Limitation
666 mA.
Temperature Limits
Storage: –40 to 85 °C (–40 to 185 °F). Operating: –20 to 70 °C (–4 to 158 °F).
Between temperatures –40 to –20 °C (–40 to –4 °F), the loop is intact and the meter is not damaged.
Humidity Limitation
0 to 95% non-condensing relative humidity.
Update Period
750 ms.
Response Time
Responds to changes in input within a maximum of two update periods. If the filter is activated, then the display responds to the change within nine update periods.
Performance Specifications
Digital Display Resolution
0.05% of calibrated range ±1 digit.
Analog Bar Graph Resolution
0.05% of calibrated range.
Indication Accuracy
0.25% of calibrated range ±1 digit.
Stability
Over Time: 0.1% of calibrated range ±1 digit per six months.
Temperature Effect
0.01% of calibrated range per °C on zero.
0.02% of calibrated range per °C on span over the operating temperature range.
Power Interrupt
All calibration constants are stored in EEPROM memory and are not affected by power loss.
Failure Mode
LCD meter failure will not affect transmitter operation.
Under/Over Range Indication
Input current < 3.5 mA: Display blank. Input current > 22.0 mA: Display flashes 112.5% of full scale value or 9999, whichever is less.
Physical Specification Meter Size
2¼-inch diameter face with ½-inch high characters.
6-5
Rosemount Model 444 Alphaline Temperature Transmitters
ANALOG METER
FIGURE 6-4.Analog Meter Face.
The analog meter plugs directly into the Model 444 to provide an accurate local indication of user-specified units. It requires an analog 4–20 mA dc, 10–50 mA dc, or 40–200 mV dc transmitter output from a two-wire transmitter, and adds no voltage drop in the 4–20 mA current loop when connected directly across the transmitter test terminals.
The large 2¼-inch diameter meter face has a two-inch long scale for easy readability, as shown in Figure 6-4. A meter-zero adjustment is located on the meter faceplate. You can rotate the meter in 90-degree increments within the transmitter housing for convenient viewing.



яюэьюыъ


751-2534C01B
Analog Meter Specifications
Functional Specifications
Performance Specifications
Input Signal
4–20 mA dc.....
10–50 mA dc.. 40–200 mV
Maximum series resistance is
}
10 ohms for milliameters.
Meter Indication
0 to 100% linear scale. Special optional ranges.
Overload Limit
150% of rated end scale value for two minutes.
Temperature Limits
–40 to 65 °C (–40 to 150 °F).
Humidity Limit
0 to 100% relative humidity.
Zero Adjustment
Adjustment screw on face of meter.
Indication Accuracy
±2% of calibrated span.
Temperature Effect
Less than 2% of full scale at any point within the temperature limits.
Physical Specification Meter Size
2¼-inch diameter face with 2-inch long scale.
6-6
Appendix
A Hazardous Locations
Certifications Drawings
Rosemount Drawing 00444-0261, 2 Sheets: Model 444 Explosion-proof Installation Drawing, Factory Mutual.
Rosemount Drawing 01151-0214, 6 Sheets: Index of Intrinsically Safe Barrier Systems and Entity parameters for 444, 1135, 1144, 1151, and 2051 Transmitters and 751 Field Indicators.
Rosemount Drawing 00444-0034, 2 Sheets: CSA Intrinsic Safety Approvals for Model 444.
Rosemount Drawing 00444-0264, 1 Sheet: Model 444 SAA Intrinisc Safety Configuration
A-1
Rosemount Model 444 Alphaline Temperature Transmitters
A-2
Appendix A
A-3
Rosemount Model 444 Alphaline Temperature Transmitters
A-4
Appendix A
A-5
Rosemount Model 444 Alphaline Temperature Transmitters
A-6
Appendix
B Temperature Sensor
Reference Information
TABLE B-1. Resistance Versus Temperature Curve Ro=100.00 Ω, a=0.00385Reference Standard IEC 751.
°C
°C 0 –10 –20 –30 –40 –50 –60 –70 –80 –90
–200 18.49 –100 60.25 56.19 52.11 48.00 43.87 39.71 35.53 31.32 27.08 22.80
0 100 96.09 92.16 88.22 84.27 80.31 76.33 72.33 68.33 64.30
°C 0 10 20 30 40 50 60 70 80 90
0 100 103.90 107.79 111.67 115.54 119.40 123.24 127.07 130.89 134.70 100 138.50 142.29 146.06 149.82 153.58 157.31 161.04 164.76 168.46 172.16 200 175.84 179.51 183.17 186.82 190.45 194.07 197.69 201.29 204.88 208.45 300 212.02 215.57 219.12 222.65 226.17 229.67 233.17 236.65 240.13 243.59 400 247.04 250.48 253.90 257.32 260.72 264.11 267.49 270.86 274.22 277.56 500 280.90 284.22 287.53 290.83 294.11 297.39 300.65 303.91 307.15 310.38 600 313.59 316.80 319.99 323.18 326.35 329.51 332.66 335.79 338.92 342.03 700 345.13 348.22 351.30 354.37 357.42 360.47 363.50 366.52 369.53 372.52 800 375.51 378.48 381.45 384.40 387.33
°F
°F 0 –10 –20 –30 –40 –50 –60 –70 –80 –90
–300 25.18 22.80 20.41 18.01 –200 48.46 46.17 43.87 41.56 39.25 36.93 34.60 32.26 29.91 27.55 –100 71.00 68.77 66.54 64.30 62.05 59.80 57.55 55.28 53.02 50.74
0 93.03 90.85 88.66 86.47 84.27 82.07 79.87 77.66 75.44 73.22
°F 0 10 20 30 40 50 60 70 80 90
0 93.03 95.21 97.39 99.57 101.74 103.90 106.07 108.22 110.38 112.53 100 114.68 116.83 118.97 121.11 123.24 125.37 127.50 129.62 131.74 133.86 200 135.97 138.08 140.18 142.29 144.38 146.48 148.57 150.66 152.74 154.82 300 156.90 158.97 161.04 163.11 165.17 167.23 169.29 171.34 173.39 175.43 400 177.47 179.51 181.54 183.57 185.60 187.62 189.64 191.66 193.67 195.68 500 197.69 199.69 201.69 203.68 205.67 207.66 209.64 211.62 213.60 215.57 600 217.54 219.51 221.47 223.43 225.38 227.34 229.28 231.23 233.17 235.11 700 237.04 238.97 240.90 242.82 244.74 246.65 248.57 250.48 252.38 254.28 800 256.18 258.07 259.96 261.85 263.74 265.62 267.49 269.36 271.23 273.10 900 274.96 276.82 278.67 280.53 282.37 284.22 286.06 287.90 289.73 291.56
1000 293.38 295.21 297.03 298.84 300.65 302.46 304.27 306.07 307.87 309.66 1100 311.45 313.24 315.02 316.80 318.58 320.35 322.12 323.88 325.64 327.40 1200 329.16 330.91 332.66 334.40 336.14 337.88 339.61 341.34 343.07 344.79 1300 346.51 348.22 349.93 351.64 353.35 355.05 356.74 358.44 360.13 361.81 1400 363.50 365.18 366.85 368.52 370.19 371.86 373.52 375.18 376.18 378.48 1500 380.13 381.77 383.41 385.05 386.68 388.31 389.94
B-1
Rosemount Model 244P Head and Rail Mount Temperature Transmitters
B-2
Index
A
Accuracy 5-3 Ambient Temperature Effect 5-3 Analog Meter Specifications 5-8
B
Burnout Protection Adjustments 4-6
C
Calibration
Low-Power Transmitter 3-8 Millivolt Transmitter 3-10 RTD Transmitter 3-2 Thermocouple Transmitter 3-5
Configuration 5-7
D
Digital Display Resolution 5-7 Dimensional Drawings
Sensor Assembly 2-9 Transmitter 2-13
Disassembly Procedure 4-4
E
Electrical Installation 2-12
Final Checkout 2-12 Input Connections 2-12 Output Connections 2-12 Preliminary Checkout 2-12
Exploded View
LCD Meter 6-4 Transmitter 5-17, 5-18 Transmitter with Mounting
Bracket 2-13, 6-1
G
Grounding 2-7
Current Signal Loop 2-7 RTD Circuit 2-7 Shielded Wire 2-7 Thermocouple and Millivolt
Transmitters 2-7
H
Hazardous Location
Certifications 5-4 Hazardous Location Installation 2-10 Humidity Limitation 5-7
I
Input Impedance 5-3 Inputs 5-1 Installation 2-1
Electrical Considerations
Intrinsically Safe
Installations 2-10
Multi-Channel
Installations 2-7
Power Supply 2-4, 2-8
Mechanical Considerations 2-1
Hazardous Location
Installation 2-10
Moist or Corrosive
Atmospheres 2-10 Mounting 2-2 Temperature
Environment 2-8
Installation Procedure 2-11
Electrical 2-12
Mechanical 2-11 Interchangeability of Parts 4-6 Intrinsically Safe Installations 2-10 Introduction 1-1
L
LCD Meter
Assembly 6-4
Specifications 6-4 LCD Meter Specifications 5-7 Load Effect 5-3 Load Limitations 5-2 Loss of Input 5-2 Low-Power Transmitter 3-8
Calibration Equipment
Required 3-8
Calibration Procedure 3-9
M
Maintenance 3-1, 4-1
Burnout Protection
Adjustments 4-6 Disassembly Procedure 4-4 Interchangeability of Parts 4-6 Reassembly Procedure 4-5 Repair 4-3 Warranty Service 4-7
Materials of Construction 5-4 Mechanical Installation 2-11 Millivolt Transmitter 3-10 Moist or Corrosive Atmospheres 2-10 Mounting 2-2 Mounting Bracket 2-2, 2-13, 6-1 Mounting Position Effect 5-3 Multi-Channel Installations 2-7
O
Ordering Tables 5-9–5-11 Output Limits 5-2 Outputs 5-1
P
Power Supply 2-4, 2-8, 5-2 Power Supply Effect 5-3
R
Reassembly Procedure 4-5 Repair 4-3
Burnout Protection
Adjustments 4-6 Disassembly Procedure 4-4 Interchangeability of Parts 4-6 Reassembly Procedure 4-5 Warranty Service 4-7
Response Time 5-7 RTD Inputs 2-5
Remote RTD Mounting 2-5
RTD Transmitter 3-2
Calibration Equipment
Required 3-2 Calibration Procedure 3-2
I-1
Rosemount Model 444 Alphaline Temperature Transmitters
S
Sensor and Conduit Connections 5-4 Sensor Connections
Grounding 2-7 RTD Inputs 2-5
2-Wire RTD 2-6 3-Wire RTD 2-6 4-Wire RTD 2-6 Compensation Loop
RTD 2-6 Differential Millivolt 2-6 Differential RTD 2-5
Thermocouple or Millivolt
Inputs 2-7 Span and Zero 5-2 Spans 5-1 Spare Parts, List of 5-12–5-16 Specifications and Reference
Data 5-1
Analog Meter Specifications 5-8 LCD Meter Specifications 5-7
Configuration 5-7 Digital Display
Resolution 5-7 Humidity Limitation 5-7 Response Time 5-7 Temperature Limits 5-7 Update Period 5-7
Transmitter Specifications 5-1,
5-2
Accuracy 5-3 Ambient Temperature
Effect 5-3 Hazardous Location
Certifications 5-4 Input Impedance 5-3 Inputs 5-1 Load Effect 5-3 Materials of
Construction 5-4 Mounting Position
Effect 5-3 Power Supply Effect 5-3 Sensor and Conduit
Connections 5-4 Stability 5-3 Turn-on Time 5-2 Vibration Effect 5-3 Weight 5-4
Stability 5-3
T
Temperature Environment 2-8 Temperature Limits 5-2, 5-7 Testing
RTD Test 4-3
Thermocouple Test 4-3 Thermocouple or Millivolt Inputs 2-7 Thermocouple Transmitter 3-5
Calibration Procedure for a
Compensated Thermocouple
Calibration Equipment
Required 3-5
Calibration Procedure 3-5
Calibration Procedure
for Ice Bath
Calibration Equipment
Required 3-7
Calibration Procedure 3-7 Troubleshooting 4-2 Turn-on Time 5-2
U
Update Period 5-7
V
Vibration Effect 5-3
W
Warranty Service 4-7 Weight 5-4 Wiring 2-4
I-2
Index
I-3
Rosemount Inc.
8200 Market Boulevard Chanhassen, MN 55317 USA Tel 1-800-999-9307 Telex 4310012 Fax (612) 949-7001 © 1998 Rosemount Inc.
T
N
I
E
D
R
P
IN
U.
A.
S.
http://www.rosemount.com
¢00809-0100-4263z¤
00809-0100-4263, Rev. AA
Fisher-Rosemount Limited
Heath Place Bognor Regis West Sussex PO22 9SH England Tel 44 (1243) 863 121 Fax 44 (1243) 867 5541
Fisher-Rosemount
Singapore Pte Ltd.
1 Pandan Crescent Singapore 128461 Tel (65) 777-8211 Fax (65) 777-0947 Tlx RS 61117 FRSPL
Loading...