FISCHER DA01 VUW, DA01 VUW ATEX, DA03 0A Series, DA03 1C Series, DA03 1B Series Operating Manual

...
Operating manual
DA01 VUW
Differential pressure measuring device
Pressure levels PN250/PN400
Standard version
09015042 • BA_EN_DA01_VUW • Rev. ST4-A • 02/18
*09015042*
| Masthead FISCHER Mess- und Regeltechnik GmbH

Masthead

Manufacturer:
Technical editorial team:
FISCHER Mess- und Regeltechnik GmbH
Bielefelderstr. 37a D-32107 Bad Salzuflen
Telefon: +49 5222 974 0 Telefax: +49 5222 7170
eMail: info@fischermesstechnik.de web: www.fischermesstechnik.de
Documentation representative: S. Richter Technical editor: R.Kleemann
All rights, also those to the translation, reserved. No part of these instructions may be reproduced or processed, duplicated or distributed using electronic sys­tems or any other form (print, photocopy, microfilm or another process) without the written consent of the company FISCHER Mess- und Regeltechnik GmbH, Bad Salzuflen.
Reproduction for internal use is expressly allowed.
Brand names and procedures are used for information purposes only and do not take the respective patent situation into account. Great care was taken when compiling the texts and illustrations; nevertheless, errors cannot be ruled out. The company FISCHER Mess- und Regeltechnik GmbH will not accept any legal responsibility or liability for this.
Subject to technical amendments.
© FISCHER Mess- und Regeltechnik 2018

Version history

Rev. ST4-A 11/15 Version 1 (first edition)
FISCHER Mess- und Regeltechnik GmbH Table of Contents

Table of Contents

1 Sicherheitshinweise ....................................................................................................................................... 4
1.1 General .....................................................................................................................................................4
1.2 Personnel Qualification.............................................................................................................................4
1.3 Risks due to Non-Observance of Safety Instructions ...............................................................................4
1.4 Safety Instructions for the Operating Company and the Operator............................................................4
1.5 Unauthorised Modification ........................................................................................................................4
1.6 Inadmissible Modes of Operation .............................................................................................................4
1.7 Safe working practices for maintenance and installation work .................................................................5
1.8 Pictogram explanation ..............................................................................................................................5
2 Product and functional description ..............................................................................................................6
2.1 Delivery scope ..........................................................................................................................................6
2.2 Equipment versions ..................................................................................................................................6
2.3 Function diagram ....................................................................................................................................11
2.4 Design and mode of operation................................................................................................................11
3 Assembly.......................................................................................................................................................12
3.1 General information ................................................................................................................................12
3.2 Process connection ................................................................................................................................12
3.3 Electrical connections .............................................................................................................................13
4 Commissioning.............................................................................................................................................15
4.1 General ...................................................................................................................................................15
4.2 Venting of the pressure lines ..................................................................................................................15
4.3 Zero point correction...............................................................................................................................16
4.4 Switch point setting.................................................................................................................................16
5 Servicing .......................................................................................................................................................17
5.1 Maintenance ...........................................................................................................................................17
5.2 Transport ................................................................................................................................................17
5.3 Service....................................................................................................................................................17
5.4 Disposal ..................................................................................................................................................17
6 Technical Data .............................................................................................................................................. 18
6.1 Allgemeines ............................................................................................................................................18
6.2 Input variables ........................................................................................................................................19
6.3 Operating conditions...............................................................................................................................19
6.4 Construction design ................................................................................................................................20
7 Order Codes..................................................................................................................................................27
7.1 accessories.............................................................................................................................................29
8 EU Declarations of conformity ....................................................................................................................30
9 EAC Declaration ...........................................................................................................................................32
BA_EN_DA01_VUW 3 / 32
1 | Sicherheitshinweise FISCHER Mess- und Regeltechnik GmbH

1 Sicherheitshinweise

1.1 General

WARNING
This operating manual contains instructions fundamental to the installation, op­eration and maintenance of the device that must be observed unconditionally. It must be read by the assembler, operator and the specialized personnel in charge of the instrument before it is installed and put into operation.
This operating manual is an integral part of the product and therefore needs to be kept close to the instrument in a place that is accessible at all times to the re­sponsible personnel.
The following sections, in particular instructions about the assembly, commis­sioning and maintenance, contain important information, non-observance of which could pose a threat to humans, animals, the environment and property.
The instrument described in these operating instructions is designed and manu­factured in line with the state of the art and good engineering practice.

1.2 Personnel Qualification

The instrument may only be installed and commissioned by specialized person­nel familiar with the installation, commissioning and operation of this product.
Specialized personnel are persons who can assess the work they have been assigned and recognize potential dangers by virtue of their specialized training, their skills and experience and their knowledge of the pertinent standards.

1.3 Risks due to Non-Observance of Safety Instructions

Non-observance of these safety instructions, the intended use of the device or the limit values given in the technical specifications can be hazardous or cause harm to persons, the environment or the plant itself.
The supplier of the equipment will not be liable for damage claims if this should happen.

1.4 Safety Instructions for the Operating Company and the Operator

The safety instructions governing correct operation of theinstrument must be observed. The operating company must make them available to the installation, maintenance, inspection and operating personnel.
Dangers arising from electrical components, energy discharged by the medium, escaping medium and incorrect installation of the device must be eliminated. See the information in the applicable national and international regulations.
Please observe the information about certification and approvals in the Tech­nical Data section.

1.5 Unauthorised Modification

Modifications of or other technical alterations to the instrument by the customer are not permitted. This also applies to replacement parts. Only the manufacturer is authorised to make any modifications or changes.

1.6 Inadmissible Modes of Operation

The operational safety of this instrument can only be guaranteed if it is used as intended. The instrument model must be suitable for the medium used in the system. The limit values given in the technical data may not be exceeded.
The manufacturer is not liable for damage resulting from improper or incorrect use.
FISCHER Mess- und Regeltechnik GmbH Sicherheitshinweise | 1

1.7 Safe working practices for maintenance and installation work

The safety instructions given in this operating manual, any nationally applicable regulations on accident prevention and any of the operating company's internal work, operating and safety guidelines must be observed.
The operating company is responsible for ensuring that all required mainten­ance, inspection and installation work is carried out by qualified specialized per­sonnel.

1.8 Pictogram explanation

DANGER
Type and source of danger
This indicates a direct dangerous situation that could lead to death or serious injury (highest danger level).
a) Avoid danger by observing the valid safety regulations.
WARNING
Type and source of danger
This indicates a potentially dangerous situation that could lead to death or ser­ious injury (medium danger level).
a) Avoid danger by observing the valid safety regulations.
CAUTION
Type and source of danger
This indicates a potentially dangerous situation that could lead to slight or seri­ous injury, damage or environmental pollution (low danger level).
a) Avoid danger by observing the valid safety regulations.
NOTICE
Note / advice
This indicates useful information of advice for efficient and smooth operation.
BA_EN_DA01_VUW 5 / 32
2 | Product and functional description FISCHER Mess- und Regeltechnik GmbH
Large measuring cell Ø130 Small measuring cell Ø75
NG100
NG160
(mbar ranges) (bar ranges)
1/4-18 NPT 1/2-14 NPT
1/4-18 NPT 1/2-14 NPT
Inner thread Outer thread Cutting ring-
screw connection
12 mm tube
Flange connection
based on DIN 615018
with internal thread G½

2 Product and functional description

2.1 Delivery scope

• Differential pressure measuring device DA01
• Operating Manual

2.2 Equipment versions

The following illustrations depict typical combinations of the measuring cell, measured value display and contact elements. However, these can be freely combined according to the order code. Wherever this is not possible, this is clearly stated.
For instance, a small measuring cell with an NG160 display and a contact ele­ment is also available.
Fig.1: Device overview

2.2.1 Process connection

Fig.2: Options for the process connection
FISCHER Mess- und Regeltechnik GmbH Product and functional description | 2
Low-action contacts Snap-actiocontacts Inductive contacts
KINAX 3W2 708-226D0 KINAX 3W2 708-226E0
for standard devices
Snap-action contacts Inductive contacts
for ATEX devices
KINAX 3W2 708-226D0 KINAX 3W2 708-226E0
Rotation angle encoder in accordance with data sheet KE09
Limit switch in accordance with data sheet KE##
Marker needle Trailing needle
Fluid fillings
Unit without contacts
Paraffin oil, glycerine, silicon oil
Unit with inductive contacts
Paraffin oil, silicon oil
Unit with low-action contacts
Paraffin oil, silicon oil
Units with magnetic spring contacts
Silicon oil

2.2.2 Nameplate

2.2.3 Contact elements

Fig.3: Contact elements

2.2.4 Special functions

Fig.4: Special functions
BA_EN_DA01_VUW 7 / 32
2 | Product and functional description FISCHER Mess- und Regeltechnik GmbH
Wall mounting Pipe mounting
Panel mounting set type 1 Panel mounting set type 2
with panel mounting set with front ring
available on request

2.2.5 Assembly

Fig.5: Assembly
The panel installation fittings can only be used in devices with a small measur­ing cell (Ø75) and a display in the NG100 bayonet ring casing.
WARNING
Panel mounting set
Due to the heavy weight, the operator needs to install a support construction for installation of the front panel.

2.2.6 Equipment features (overview)

The following shows the configuration possibilities of the DA01 depending on the measuring cell used and the pressure level.
Key
FISCHER Mess- und Regeltechnik GmbH Product and functional description | 2
Measuring range
Measured value display
Ø100
Inductive contacts
Rotation angle transducer
Trailing needle
Marker needle
Remote seal
1 2 3 1 2
0 ... 0.6 bar
0 ... 1 bar
0 ... 1.6 bar
0 ... 2.5 bar
0 ... 4.0 bar
0 ... 6bar
0 ... 10 bar
0 ... 16 bar
0 ... 25 bar
-1 ... 0.6 bar
-1 ... 1.5 bar
-1 ... 3 bar
Snap-action contacts
-1 ... 5 bar
1 2 3
Low-action contacts
Pressure levelPN250/PN400
Measuring range
Measured value display
Ø160
Inductive contacts
Rotation angle transducer
Trailing needle
Marker needle
Remote seal
1 2 3 1 2
0 ... 0.6 bar
0 ... 1 bar
0 ... 1.6 bar
0 ... 2.5 bar
0 ... 4.0 bar
0 ... 6bar
0 ... 10 bar
0 ... 16 bar
0 ... 25 bar
-1 ... 0.6 bar
-1 ... 1.5 bar
-1 ... 3 bar
Snap-action contacts
-1 ... 5 bar
1 2 3
Low-action contacts
Pressure levelPN250/PN400
Small measuring cell Ø75
BA_EN_DA01_VUW 9 / 32
Fig.6: Small measuring cell Ø75 Measured value display Ø100
Fig.7: Small measuring cell Ø75 Measured value display Ø160
2 | Product and functional description FISCHER Mess- und Regeltechnik GmbH
-100 ... +250 mbar
Measuring range
Inductive contacts
Rotation angle transducer
Trailing needle
Marker needle
Remote seal
1 2 3 1 2
0 ... 40 mbar
0 ... 60 mbar
0 ... 100 mbar
0 ... 160 mbar
0 ... 250 mbar
0 ... 400 mbar
-40 ... +60 mbar
-60 ... +100 mbar
-100 ... +150 mbar
Snap-action contacts
1 2 3
Low-action contacts
Pressure level
PN250
Measured value display
Ø100
-100 ... +250 mbar
Measuring range
Inductive contacts
Rotation angle transducer
Trailing needle
Marker needle
Remote seal
1 2 3 1 2
0 ... 40 mbar
0 ... 60 mbar
0 ... 100 mbar
0 ... 160 mbar
0 ... 250 mbar
0 ... 400 mbar
-40 ... +60 mbar
-60 ... +100 mbar
-100 ... +150 mbar
Snap-action contacts
1 2 3
Low-action contacts
Pressure level
PN250
Measured value display
Ø160
Large measuring cell Ø130
Fig.8: Large measuring cell Ø130 Measured value display Ø100
Fig.9: Large measuring cell Ø130 Measured value display Ø160
FISCHER Mess- und Regeltechnik GmbH Product and functional description | 2
1
2
3
4
5
6
7
8
8
7

2.3 Function diagram

Fig.10: Function diagram
1 Motion train 2 Transfer lever 3 Measuring shaft 4 Pressure transfer fluid 5 Separating membrane 6 Connecting rod 7 Measuring diaphragm 8 Pressure chamber

2.4 Design and mode of operation

The pressures in the pressure chambers that are to be compared are each ex­erted onto a measuring membrane that can be rigidly connected using a con­nection rod. To compensate the static pressure, the space between the separat­ing and measuring membrane is filled with a pressure transfer fluid.
During pressure equalisation, the two measuring membranes are in an idle pos­ition. In case of pressure difference, the force acting on the membranes causes it to be moved towards the side of the lower pressure.
The connecting rod transfers the deflection of the measuring membranes onto the transfer lever mounted to the measuring shaft. Proportional to the current differential pressure, the measurement shaft makes a rotational movement that the indicator translates into a rotation angle between 0 and 270°.
In the case of one-sided pressure by the measuring system above and beyond the measuring range, the over-pressure guard will be activated which supports the overlosaded membrane.
BA_EN_DA01_VUW 11 / 32
3 | Assembly FISCHER Mess- und Regeltechnik GmbH

3 Assembly

3.1 General information

The device can be mounted in one of the following ways (see Assembly):
1. Wall mounting
The device is designed for installation onto flat assembly plates. The unit is equipped with a wall mounting plate for this mounting type.
2. Pipe assembly
The device is equipped with a special pipe assembly set and is suitable for mounting to a 2" pipe (DN50).
3. Panel mounting
Panel mounting is available in two installation models.
Panel mounting fittings (Type 1)
This variant is only suitable for models with a small measuring system (Ø75) and a bayonet ring casing NG100.
Panel installation with front ring (Type 2)
All models are suitable for this variant. A suitable steel construction must be used to ensure that the front plate can bear the weight of the DA03. This is possible e.g. by means of a combination with the wall or pipe assembly.
At the factory, the device is calibrated for vertical installation, but the installation position is arbitrary. For any installation positions that are not vertical, the zero­point signal can be corrected via the installed offset correction.
To ensure safety during installation and maintenance, we recommend installing a suitable shut-off valve on the system (see accessories).

3.2 Process connection

• By authorized and qualified specialized personnel only.
• The pipes need to be depressurized when the instrument is being connec­ted.
• Appropriate steps must be taken to protect the device from pressure surges.
• Check that the device is suitable for the medium being measured.
• Maximum pressures must be observed (cf. Tech. data)
All supply lines are arranged so that there are no mechanical forces acting on the device.
The pressure lines must be kept as short as possible and installed without any tight bends to avoid delays.
The pressure lines must be installed at an inclination so that when fluids are measured no air pockets are created or when measuring gases, no water pock­ets are created. If the required inclination is not reached, water or air filters must be installed at suitable places.
The pressure lines need to be vented for fluid measuring media. [}15].
If water is used as a measuring medium, the unit must be protected against frost.
If the pressure sensing lines are already pressurised at the time of commission­ing, zero-point control and adjustment cannot be performed. In such cases, the device should be only connected to the mains without the pressure sensing lines.
FISCHER Mess- und Regeltechnik GmbH Assembly | 3
+
Higher pressure
lower pressure
1
2
3
4
5
6
M20 x 1.5
1
2
5
6
3
4
4
The process connections are marked with (+) and (-) symbols on the device. The pressure lines must be mounted according to these symbols.
Differential pressure
Fig.11: Process connection

3.3 Electrical connections

• By authorized and qualified specialized personnel only.
• When connecting the unit, the national and international electro-technical regulations must be observed.
• Disconnect the system from the mains, before electrically connecting the device.
• Install the consumer-adapted fuses.
• Do not connect the connector if strained.
Only devices with contact elements are connected to the electrical supply. To this end there is a cable socket on the side of the device, or a HAN plug in the power plant version.

3.3.1 Cable socket / HAN plug

BA_EN_DA01_VUW 13 / 32
Fig.12: Cable socket
1 Lower part 2 Sealing ring EPDM 3 Middle part 4 Sealing ring EPDM 5 Lid 6 Lid screw
Fig.13: HAN connector
1 Cable screw connection M20 x 1.5 2 Sleeve housing Han 3A 3 Socket insert Han 7D 4 Pin insert Han 7D 5 Safety clip 6 Attachment casing Han 3D
3 | Assembly FISCHER Mess- und Regeltechnik GmbH
1 2 3 4
1 2
1 2 3 4
1 2 3
1 2 3 4
-+-
+
1 2
Eb= 5 ... 25V
1 2 3
R
L
+
-
Eb= 12 ... 30V
1 2 3
R
L
+
-
~
+
-
~
+
-
2-Wire 3-Wire

3.3.2 Contact elements

Contact elements are supplied in accordance with data sheet KE. This illus­trates all variants, their pin assignment and the technical data.
Creep and magnetic spring contacts
The terminal numbers always correspond to the contact number an are as­signed to the target indicators from left to right. Up to three contacts can be used. There are assigned to the target indicators as follows:
1. For 2 contacts Contact 1: left target indicator Contact 2:right target indicator
2. For 3 contacts Contact 1: left target
Fig.14: Contacts
Contact 2: middle target Contact 3: right target
Inductive contacts
In the case of inductive contacts, the switch function is not only determined by the slot type initiator but also by the switch amplifier. Up to max two contacts can be used. There are assigned to the target indicators as follows:
Contact 1: left target indicator
Fig.15: Inductive contacts
Contact 2:right target indicator.

3.3.3 Rotation angle transducers KINAX 3W2

The rotation angle encoder serves to record angular positions, to prepare and provide the measured values as electrical output signals 0/4…20mA for the following device. Rotation angle encoder is supplied in accordance with data
sheet KE09. This contains further information about the pin assignment and the
technical data.
Fig.16: Rotation angle transducers connection
FISCHER Mess- und Regeltechnik GmbH Commissioning | 4
Venting screw
Closing screw0.8 x 4
Venting screws
5
Seal (FKM)

4 Commissioning

4.1 General

All electrical supply, operating and measuring lines, and the pressure connec­tions must have been correctly installed before commissioning. All supply lines are arranged so that there are no mechanical forces acting on the device.
Check that all pressure connections are free of leaks before commissioning.
In models filled with fluid, the venting valve on the upper side of the bayonet casing must be opened before commissioning! To do this, turn the venting screw anticlockwise as far as it will go.
Fig.17: Venting valve

4.2 Venting of the pressure lines

WARNING
Risk connected to pressure
Never remove the venting screw if the unit is still pressurised! Close the shut-off valves of the flanged fittings or depressurize the system.
The pressure lines need to be vented for before commissioning on devices that work with fluid media. Proceed as follows:
• Remove the venting screws of the two pressure chambers.
• Carefully increase the system pressure until the fluid level is 5 mm below the sealing surface of the threaded borehole.
• Close the device with the venting screws
Fig.18: Venting with fluid media
BA_EN_DA01_VUW 15 / 32
4 | Commissioning FISCHER Mess- und Regeltechnik GmbH
Without fluid filling With fluid filling
Closing screw 11
1.2 x 7
1
2
3
4
5
6

4.3 Zero point correction

The differential pressure measuring units are set in the factory before delivery so that they do not usually need to be adjusted at the assembly site. If this is still necessary, proceed as follows:
• Depressurize the measuring chamber (+) und (-) side or only exert the exist­ing static system pressure.
• Remove the closing screw The zero point correction screw is located be­hind.
• Set the measurement value pointer to the scale zero point using the zero point correction screw.
• Mount the closing screw
Fig.19: Zero point correction screw

4.4 Switch point setting

There is an adjustment lock attached to the front pane of the measuring unit on units with contact elements. This means that the contacts attached to the target indicators can be set to any point along the scale.
To facilitate switching precision and the service life of the mechanical measur­ing system, the switching points should lie between 10% and 90% of the meas­uring range.
Fig.20: Switch point setting
1 Adjustment key 2 Adjusting lock 3 Axle 4 Drive arm 5 Set-point display 6 Actual value display
Adjustment sequence:
• Press the axle inwards until the drive arm reaches behind the setting pin of the target value indicator.
• Set the target value indicator to the required switch point by turning the set­ting key.
• Relieve the axle.
• The switch-point setting can be secured against unintentional adjustment by removing the attachment screw and the adjustment key.
FISCHER Mess- und Regeltechnik GmbH Servicing | 5

5 Servicing

5.1 Maintenance

The instrument is maintenance-free. We recommend the following regular in­spection to guarantee reliable operation and a long service life:
• Check the function in combination with downstream components.
• Check the leak-tightness of the pressure connection lines.
• Check the electrical connections.
The exact test cycles need to be adapted to the operating and environmental conditions. In combination with other devices, the operating instructions for the other devices also need to be observed.

5.2 Transport

The measuring device must be protected against impacts. It should be transpor­ted in the original packaging or a suitable transport container.

5.3 Service

All defective or faulty devices should be sent directly to our repair department. Please coordinate all shipments with our sales department.
WARNING
Process media residues
Process media residues in and ondismantled devices can be a hazard to people, animals and the environment. Take adequate preventive measures. If required, the devices must be cleaned thoroughly.
Return the device in the original packaging or a suitable transport container.

5.4 Disposal

WARNING
Incorrect disposal may pose a risk to the environment.
Please help to protect the environment by always disposing of the work pieces and packaging materials in compliance with the valid national waste and recyc­ling guidelines or reuse them.
BA_EN_DA01_VUW 17 / 32
6 | Technical Data FISCHER Mess- und Regeltechnik GmbH

6 Technical Data

6.1 Allgemeines

EXECUTION Nominal
pressure
DA01 V … PN250 Ø75 Measuring ranges: 0…0.6 bar to 0…25 bar
DA01 U … PN400 Ø75 Measuring ranges: 0…0.6 bar to 0…25 bar
DA01 W … PN250 Ø130 Measuring ranges: 0…40 mbar to 0…400 bar
Measuring cell
Application information
Remote seals: It is possible to attach remote seals for all measuring ranges. The re­mote seals need to be designed for the displacement volume, the length of the cable and the application temperature.
Remote seals: It is possible to attach remote seals for all measuring ranges. The re­mote seals need to be designed for the displacement volume, the length of the cable and the application temperature.
Limitations: Drag indicator measuring ranges ≥ 60 mbar Contacts / Transmitter measuring ranges ≥ 100 mbar
Remote seals: It is possible to attach remote seals for measuring ranges ≥ 160 mbar. The remote seals need to be designed for the displacement volume, the length of the cable and the application temperature.
FISCHER Mess- und Regeltechnik GmbH Technical Data | 6

6.2 Input variables

Measuring variable
Differential pressure in gaseous and fluid aggressive media.
General
Small measuring cell Ø75
Rated pressure of the measuring system
Durability One-sided over-pressure-proof up to the
Measurement accuracy ±1.6 % of the measuring range Temperature sensor 0.3 % / 10 °C Zero-point adjustment ±25 % of the measuring range
Max. static operating pressure
rated pressure of the measuring system resistance to under-pressure on the (+) and (-) side

Measuring ranges

Measuring range Device model
V U W
0 … 0.6 bar ● 0 … 1 bar ● 0 … 1.6 bar ● 0 … 2.5bar ● 0 … 4.0 bar ● 0 … 6 bar ● 0 … 10 bar ● 0 … 16 bar ● 0 … 25 bar
-1 … 0.6 bar
-1 … 1.5 bar
-1 … 3 bar
-1 … 5 bar
Large measuring cell Ø130
Measuring range Device model
V U W
0 … 40 mbar ● 0 … 60 mbar ● 0 … 100 mbar ● 0 … 160 mbar ● 0 … 250 mbar ● 0 … 400 mbar ● 0 … 600 mbar

6.3 Operating conditions

Permissible ambient temperature -20 … +80 °C Admissible storage temperature -40 … +80 °C Admissible media temperature Max. 100 °C Type of protection: IP 65 acc. to EN 60529
BA_EN_DA01_VUW 19 / 32
6 | Technical Data FISCHER Mess- und Regeltechnik GmbH

6.4 Construction design

Materials

Measured value display Material Material no.
EU AISI
Bayonet ring housing NG100, NG160 CrNi steel 1.4301 304 Process connection (all models) CrNi steel 1.4404 316L Intermediate plate AlMgSiPb HARD-COAT Seals (O-rings) FKM Motion train CrNi steel Dial face and needle Aluminium, painted, printed Inspection disk Safety laminated glass
®
MB = Measurement range
Materials (media-contacting)
Design of the measuring system (R) Material Material no.
EU AISI
Pressure caps CrNi steel 1.4404 316L Separation membranes CrNi steel 1.4571 361Ti
Design of the measuring system (G) Material Material no.
EU AISI
Pressure caps CrNi steel 1.4404 316L Separation membranes Hastelloy® C276
Process connection Material Material no.
EU AISI
Connecting piece and port CrNi steel 1.4404 316L Cutting ring screw connections CrNi steel 1.4571

Assembly

Wall mounting Flanged assembly plate Pipe mounting Flanged assembly plate and attachment bracket Panel mounting set
type 1 Panel mounting set
type 2
Panel installation fittings for units with a small measuring cell (Ø75) and NG100 bayonet ring casing.
Front ring and support construction
FISCHER Mess- und Regeltechnik GmbH Technical Data | 6

6.4.1 Additional Attachments

6.4.1.1 Contact elements
Limit signal transmitters (contacts) and capacitive rotation angle transducers with an output signal proportional to the angular position can be fitted into a housing augmented by a corresponding bayonet ring connector.
A certain minimum pressure level is required to operate this kind of contact ele­ment, which is why there is a lower limit for the mbar measuring ranges. This limit depends on the model type and is stated in the section ‚General‘.
The measuring deviation increases by ±0.5% per contact when the contacts are driven and switched.
For more information and the order key, please refer to the data sheet:
• for limit switch in data sheet KE
• for rotation angle converter in the data sheet KE09
6.4.1.2 Fluid charging
Under aggravated operating conditions, such as vibrations and extreme pres­sure fluctuations, or in order to avoid condensation forming if used outdoors, the casing can be filled with the following fluids depending in the type of contacts in­stalled:
without contacts Glycerine, silicon oil Low-action contacts Silicon oil Magnetic spring contacts Silicon oil Inductive contacts Paraffin oil Rotation angle transducer no filling possible
6.4.1.3 Marker needle
A settable red marker can be attached to the scale to clearly show a certain pressure (limit value).
6.4.1.4 Trailing needle
The railing needle is 'dragged' with the measured value indicator. As there is no fixed connection between the two needles, one-off maximum values are stored. The trailing needle can be reset using an adjusting dial in the window. Trailing needles cannot be used in conjunction with contacts. A certain minimum pres­sure level is required to move the drag indicator, which is why there is a lower limit for the mbar measuring ranges. This limit depends on the model type and is stated in the section ‚General‘.
6.4.1.5 Shut-off fitting
3-spindle valve block PN 100, DN 5, can be directly flanged
• Type DZ3600SV2700
• Material 1.4571
• Functions: Shut-off, pressure compensation
BA_EN_DA01_VUW 21 / 32
6 | Technical Data FISCHER Mess- und Regeltechnik GmbH
Cable socket
HAN 7D

6.4.2 Electrical connection

In the case of devices with additional electronic equipment, the connection is realised using a cable socket attached to the side and/or with a Han 7D con­nector on th epower plant models. The pin assignment depends on the ordered mode and is stated in the data sheet KE or KE09.
Fig.21: Cable socket
Cable socket
Number of screw terminals 6 + 2PE Rated current See data sheet KE Rated voltage 250 V Cable diameter up to 1.5 mm2 with wire protection Cable screw connection M20 x 1.5, terminal range 7 … 13 mm
HAN 7D
No. of crimp contacts 7 + PE Rated current See data sheet KE Rated voltage 50 V Cable diameter 1 mm Cable screw connection M20 x 1.5, terminal range 7 … 13 mm
2
FISCHER Mess- und Regeltechnik GmbH Technical Data | 6
Small measuring system (Ø75)
Flange based on DIN EN 61518 Wall mounting plate
152
105
90
80
86.5
132
(with installed contacts)
112
213
(with installed contacts)
165
113
NG100 Ø101
NG160 Ø161
Ø9
41.3
54.4
80
132 152
R6
Ø9
Cable socket
7/16 UNF
Round steel bracket
2'' pipe (DN50)
2" pipe mounting
Pipe mounting plate
Spacer sleeve
Attachment kit for
round steel bracket
(possible for all models)
The pipe assembly set is also available as accessory.

6.4.3 Dimensional drawings

All dimensions in mm unless otherwise stated
Fig.22: Dimensional drawing (Small measuring system Ø75)
Fig.23: Pipe mounting
BA_EN_DA01_VUW 23 / 32
6 | Technical Data FISCHER Mess- und Regeltechnik GmbH
Wall mounting plate
Flange based on DIN EN 61518
Large measuring system (Ø130)
170
60
200
100
R6
Ø11
200
119
149
(with installed contacts)
170
60
120
Ø11
170
100
277.5
229.5
175
NG100 Ø101
NG160 Ø161
(with installed contacts)
7/16 UNF
54.4
41.3
Cable socket
Installation of front panel type 1
d2 d1
d3
120°
d4
d1
NG100
d2 d3 d4
101 132 116 4.8
Front ring
Bayonet ring
with connecting bracket
Holding plate
Assembly bracket
The front plate thickness must be min. 2 mm.
(only small measuring system Ø75 and NG100 display)
Fig.24: Dimensional drawing (Large measuring system Ø130)
Fig.25: Installation of front panel with panel fittings
FISCHER Mess- und Regeltechnik GmbH Technical Data | 6
Installation of front panel type 2
Front ring
Bayonet ring
with connecting bracket
The front plate thickness must be min. 2 mm.
A suitable steel construction must be used to ensure that the front plate can bear the weight of the DA03.
Examples:
d2 d1
d3
120°
d4
d1 NG100 NG160
d2 d3 d4
101 132 116 4.8 161 196 178 5.8
Assembly on a mounting plate Mounting to a 2'' pipe
Fig.26: Installation of front panel with front ring
BA_EN_DA01_VUW 25 / 32
6 | Technical Data FISCHER Mess- und Regeltechnik GmbH
Contact elements
NG100 = 90 NG160 = 120
NG100 = 138 NG160 = 167.5
100 18 118
Shut-off fitting
~147
~155
Cutting ring connection G3/8 for 12 mm pipe
DA03
Process
1
2
3
2
13
~110
Valve open
108
54
(Connection clearance)
~73
with inner spindle thread
M10
42
Fig.27: Dimensional drawing contact devices
Fig.28: Shutoff valve DZ3600SV2700
FISCHER Mess- und Regeltechnik GmbH Order Codes | 7
D A 0 1
1 2 5 6 7 8 9 10 11 12 13 14 15 163 4
0
Process connection
Output signal
Operating voltage
Measuring unit
Measured value display /
contact elements
Electrical connection
Assembly
Type
Measuring range
Code no.

7 Order Codes

Device model:
[1] Pressure level Measuring cell
V PN250 Ø75 U PN400 Ø75
W PN250 Ø130
Small measuring system Ø75
Measuring range:
[2.3] Measuring range Device model
V U W 01 0 … 0.6 bar 02 0 … 1 bar 03 0 … 1.6 bar 04 0 … 2.5bar 05 0 … 4.0 bar 06 0 … 6 bar 07 0 … 10 bar 08 0 … 16 bar 09 0 … 25 bar 32 -1 … 0.6 bar 33 -1 … 1.5 bar 34 -1 … 3 bar 35 -1 … 5 bar
BA_EN_DA01_VUW 27 / 32
7 | Order Codes FISCHER Mess- und Regeltechnik GmbH
Large measuring system Ø130
[2.3] Measuring range Device model
V U W 57 0 … 40 mbar 58 0 … 60 mbar 59 0 … 100 mbar 60 0 … 160 mbar 82 0 … 250 mbar 83 0 … 400 mbar
C1 0 … 600 mbar
70 -40 … +60 mbar 72 -60 … +100 mbar 74 -100 … +150 mbar 76 -100 … +250 mbar
Design of the measuring system:
[4]
R Pressure chamber CrNi steel 1.4404 (AISI 316L)
Measuring membrane standard
G Pressure chamber CrNi steel 1.4404 (AISI 316L)
Measuring membrane Hastelloy C276
Process connection:
[5.6]
03 Flange connection based on DIN EN 61518
with internal thread G½
04 Connecting piece G½ with inside thread 1/4 -18 NPT 05 Connecting piece G½ with inside thread 1/2 -14 NPT 13 Connection shanks G½ with external thread G½ 14 Connecting port G½ with outer thread 1/4-18 NPT 15 Connecting port G½ with outer thread 1/2-14 NPT 27 Cutting ring connection in brass for 12 mm pipe
Measured value display:
[7]
L Bayonet ring housing NG100
M Bayonet ring housing NG160
Assembly:
[8]
W Wall mounting
R Pipe mounting T Panel installation fittings
(only a small measuring system Ø75, NG100 measured value display without contact elements)
G Front ring for panel mounting
FISCHER Mess- und Regeltechnik GmbH Order Codes | 7
Fluid filling:
[9]
0 Without fluid filling 1 Glycerine 4 Paraffin oil 5 Silicon oil
Special functions:
[10]
0 Without special function 1 Adjustable marker needle 2 Resettable drag needle
Contacts/transmitters:
[11]
0 No contacts/transmitters 1 Built-in contacts as per data sheet KE 2 Installed capacitive rotation angle encoder in accordance with data
sheet KE09
5 Built-in contacts with plug connector (power plant model)
[12]
0 Standard version
Limitations
A minimum operating pressure, which not all measuring ranges achieve, is re­quired to activate a contact element or a drag indicator. Please also note the in­formation about the equipment features [}8].

7.1 accessories

Order no. Planned measures Material DZ3600SV2700 Triple valve block DN5 PN420
• Flange connection acc. to DIN EN 61518
• Cutting ring screw connections 12 mm pipe
• Including assembly set
Order no. Planned measures Type 05003065 Isolating unit amplifier 1-channel 24 V DC TS500Ex-ia-1R-5 05003066 Isolating unit amplifier 2-channel 24 V DC TS500Ex-ia-2R-5 05003083 Isolating unit amplifier 1-channel 230 V AC TS500Ex-ia-1R-0 05003084 Isolating unit amplifier 2-channel 230 V AC TS500Ex-ia-2R-0 05003070 Universal supplier isolator ST500Ex-10-5 05003086 Universal supplier isolator ST500Ex-10-0
1.4571
BA_EN_DA01_VUW 29 / 32
8 | EU Declarations of conformity FISCHER Mess- und Regeltechnik GmbH

8 EU Declarations of conformity

Fig.29: CE_EN_DA01_10
FISCHER Mess- und Regeltechnik GmbH EU Declarations of conformity | 8
Fig.30: CE_EN_DA01_20
BA_EN_DA01_VUW 31 / 32
9 | EAC Declaration FISCHER Mess- und Regeltechnik GmbH

9 EAC Declaration

Fig.31: ЕАЭС N RU Д-DЕ.АЛ16.В.77754
Loading...