Fairchild T5220, T5220 Series, TXPD5220 Installation, Operation And Maintenance Instructions

FAIRCHILD T5220 SERIES
ELECTRO-PNEUMATIC TRANSDUCER
Installation, Operation and Maintenance Instructions
Figure 1. Model T5220 Outline Dimensions
Figure 2. Mounting Kit 15268. (Sold Separately)
INST ALLATION
The Model T5220 can be mounted directly onto a flat surface using the two 10-32 tapped mounting holes in the base of the housing. For more information, see Figure 1. “T5220 Outline Dimensions” above.
The Model T5220 can be Panel Mounted using the mounting bracket supplied in Mounting Kit 15268. For more information, see Figure 2. above.
1
Installation (Continued)
NOTE:
The Model T5420 can be mounted on a 2" pipe using the Optional Mounting Kit 14596. For more information, see Figure 3. below.
For outdoor use, mount the T5220 in an upright position. This will help prevent rain water from entering into the unit. For more information, see Figure 3. “Mounting Kit 14596” below.
Figure 3. Optional Mounting Kit 14596. (Sold Separately)
2
INSTALLATION
The Model TXPD5220 can be mounted directly onto a flat surface using the four 1/4-20 x 7/16 tapped mounting holes in the base of the housing . For more information, see Figure
4. “TXPD5220 Outline Dimensions” below. 14140 is available for Panel or 2" Pipe Mounting. For more information, see Figure 5 below.
Mounting Kit
Figure 4. TXPD5220 Outline Dimensions.
Figure 5. Mounting Kit 14140. (Sold Separately)
3
Pneumatic Connections
Electric Connection
Clean all pipelines to remove dirt and scale before instal­lation.
Apply a minimum amount of pipe compound to the male threads of the fitting only. Do Not use teflon tape as a sealant. Start with the third thread back and work away from the end of the fitting to avoid the possibility of contaminating the transducer. Install the transducer in the air line.
NOTES:
Instrument quality air, per ISA Standards D7.3-1981, is required. Use a filter to re­move dirt and liquid in the air line ahead of the transducer for correct performance. If an air line lubricator is used, it MUST be located downstream, beyond the trans­ducer.
The user is responsible for insuring that the environment in which the unit will be installed, and the operating gas, are com­patible with the materials in the trans-
WARNING:
To prevent possible ignition of hazardous atmosphere, DO NOT REMOVE COVER from the TXPD5220 Transducer while current is on.
To maintain Explosion-Proof capability, DO NOT DAMAGE mating surfaces between cover and base. NEVER ADD GASKETS.
For the T5220 Transducer, make connections to the Termi­nal Block through a 1/2 NPT Conduit Connector (not supplied) and a plastic sleeve as shown below in Figure 7. “Electrical Connections”.
For the TXPD5220 Transducer, make connections from the Sealed Connector Wires to an external Terminal Block. The TXPD5220 Transducer has an additional ground wire that is used. For more information, see Figure 7. “Electrical Connections” below.
NOTE: The T5220 & TXPD5220 should be wired in
accordance with the wiring diagram inside of cover.
Wiring in Hazardous Areas
Wiring in hazardous areas should be performed in accor­dance with the Table 1. and any local codes that apply.
Table 1. Hazardous Location Wiring Practices.
Country
U.S. Canada
Agency
FM CSA
Code
ANSI/ISA RP 12.6 & ANSI/NFPA 70 CEC Part 1
Intrinsically Safe Connections
Refer to the latest revision of the indicated drawing.
Table 2. Intrinsically Safe Installation Drawings.
Underwriting Group Drawing Number
FM (Factory Mutual) CSA (Canadian Standards)
EC-14007 EC-16097
Figure 6. Electrical Connections.
4
CALIBRATIONS / ADJUSTMENTS
Equipment Required for Calibration:
Pneumatic Supply capable of delivering up to 150 psig.
Current Supply capable of delivering up to 50 mA.
Voltage Supply capable of delivering up to 10 VDC.
Pressure Gage capable of a digital readout up to 50 psig
with an accuracy of .1%.
Digital Volt Meter capable of a readout up to 30 mA or 10 VDC with an accuracy of .02%.
The following adjustments are provided:
Full Range Operation Forward/Reverse Mode Calibration - Zero and Span
FULL RANGE OPERATION Forward Acting Mode Adjustment
NOTE:
1.
Forward Acting Calibration-Zero
Air Supply must be on before adjustments are made.
Connect the input signal to the transducer as shown in Figure 6. “Electrical Connections.”
2.
Apply the minimum input signal and adjust the Zero screw for minimum output pressure. Turn screw counterclockwise to increase pressure and clock­wise to decrease pressure.
Forward Acting Calibration-Span
3.
Apply the maximum input signal and adjust the Span screw on the potentiometer for maximum output pressure. Turn screw clockwise to increase pres­sure and counterclockwise to decrease pressure.
Repeat steps 2-3 until the desired output range is
4.
obtained. For more information, see Figure 7. “T5220/ TXPD5220 Calibration Configuration.”
Reverse Acting Mode Adjustment
NOTE:
5.
Reverse Acting Calibration-Zero
Reverse Acting Calibration-Span
Intrinsically safe units cannot be set for re­verse acting mode in field.
Connect the input signal to the transducer as shown in Figure 6. “Electrical Connections.”
6.
Apply the minimum input signal and adjust the Zero screw for maximum output pressure. Turn screw counterclockwise to increase pressure and clock­wise to decrease pressure.
7.
Apply the maximum input signal and adjust the Span screw on the potentiometer for minimum output pres­sure. Turn screw counterclockwise to increase pres­sure and clockwise to decrease pressure.
Repeat steps 6-7 until the desired output range is
8.
obtained. For more information, see Figure 8. “T5200/ TXPD5200 Calibration Configuration.”
Figure 7. T5220 / TXPD5220 Calibration Configuration.
5
MAINTENANCE
To clean the Orifice, use the following procedure:
Turn off the air supply to the transducer. If the
1.
transducer is a common supply unit, remove the Pipe Plug from the In Port located on the rectangular base. If the transducer is a separate supply unit, remove the Air Line Fitting from the In Port.
2.
Remove the Orifice Assembly (9) from the unit by threading a 1/4-20" Screw into the orifice housing and pulling it out through the inlet port. For more informa-
tion, see Figure 8. “Exploded Drawing” below.
Clean with alcohol and dry with compressed air.
3.
NOTE:
To clean the Inner Valve Assembly, use the following procedure:
1.
2.
3.
4.
NOTE:
Parts must be completely dry before reas­sembling.
Shut off the valve that is supplying air to transducer. It is not necessary to remove the Transducer from air line.
Remove Screws (7) holding the Retainer Cap (6) and Plate (5) to Booster (1).
Pull out Inner Valve Assembly (4) and wash the seat carefully.
Wipe off any particles that may be attached to rubber Seat Ring Assembly (2). For more information, see Detail A, below.
Avoid such solvents as acetone, carbon tetra­chloride and trichlorethylene.
To clear probblems caused by Magnet malfunction, use the following procedure:
1.
Shut off the valve that is supplying air to transducer. It is not necessary to remove the Transducer from air
line.
2.
Remove three Screws (12) holding Magnet (11) to Transducer (8).
3. Remove Magnet (11) and clean gap of any foreign material with masking tape. For more information, see
View B, below.
4.
Place Magnet (11) over Coil (10) and tighten with three Screws (12).
5.
Using a non-matalic rod, Raise up on the arm to make sure Coil (10) is not sticking in Magnet (11).
NOTE:
To replace PC Board, use the following procedure:
1.
2.
3.
4.
If Coil is sticking, loosen Screws and repo­sition Magnet. Tighten Screws and repeat step 5.
Disconnect wires from Terminal Block on PC Board (14).
Unsolder Red (+) and Black (-) Leads on PC Board (14).
Remove four Screws (13) holding PC Board (14) to Transducer (8). For more information, see Figure 8. “Exploded Drawing” below.
Replace PC Board (14) with the appropriate PC Board from Table 3. on page 7.
Figure 8. Exploded Drawing.
6
Maintenance (continued)
TROUBLE-SHOOTING
Table 3. T5220 Series Components.
Item
Qty.
1 2 3 4 5 6 7 8 9 10 11 12 13 14 14 14 14 14 14 14 14
1
2
1 1 1 1 1 1 2 1 1 4 1 1 1 1 1 1 1 1 3 1 1
For T5220, TFI5220, TFN5220, TFNI5220, TXPD5220, & TXPDI5220 Units Only.
For TCI5220 Only.
Part No.
NS NS NS NS NS NS NS NS NS NS NS NS
NS 13744-2 15638-2 13745-2 15483-2 14639-2 15640-2 13746-2 15639-2
Description
Transducer Orifice Coil Magnet Screw Screw Booster Seat Ring Assembly Screen Inner Valve Assembly Retainer Plate Retainer Cap Screw
1
1-5 mA PC Board
2
1-5 mA PC Board
1
4-20,10-50 mA PC Board
2
4-20,10-50 mA PC Board
1
1-5 VDC PC Board
2
1-5 VDC PC Board
1
1-9 VDC PC Board
2
1-9 VDC PC Board
NS: (Not Serviceable) In the Part No. column, NS indi­cates that the part is not available for field replacement.
Table 4. Trouble-Shooting.
Problem
No Output
Solution (check)
Supply Pressure Clogged Orifice Power Supply
Leakage
Low or Improper Span Adjust
Pneumatic Connections
Zero and SpanAdjust Supply Pressure Low Output Leakage
Erratic Operation
DC Signal Loose Wires or Connections Liquid in Air Supply Dirt in Magnet Gap
Unit Does Not
Replace PC Board
Operate
NOTE: Booster Repair Kits are available for the T5220.
Ratio 1:1 1:2 1:3 1:4 1:5 1:6
Kit Number 15288-1 15288-2 15288-3 15656-4 15656-5 15656-6
7
IS-500T5220
Litho in USA
Rev. M 06/07
Loading...