EZ Digital FC-3000, FC-7015, FC-7150, FC-7150U, FC-7015U Operation Manual

Page 1
WARRANTY
Warranty service covers a period of one year from the date of original purchase. In case of technical failure within one year, our service center or sales outlet free of
charge will provide repair service. We charge customers for repair after the one-year warranty period has been expired.
Provided that against any failure resulted from the user’s negligence, natural disaster or accident, we charge you for repairs regardless of the warranty period.
For more professional repair service, be sure to contact our service center or sales outlet.
Page 2
Introduction
Thank you for purchasing a EZ product. Electronic measuring instruments produced by EZ Digital are high technology products made under strict quality control. We guarantee their exceptional precision and utmost reliability. For proper use of the product, please read this operation manual carefully. EZ Digital Co., Ltd.
Note
1. To fully maintain the precision and reliability of the product use it within the range of standard setting( temperature 10 °C~35 °C, humidity 45%~85%)
2. After turning of power, please allow a pre-heating period of as long as some 30 minutes before use.
3. This equipment should be used with a triple line power cord for safety.
4. For quality improvement the exterior design and specification of the product can be changed without prior notice.
5. If you have further questions concerning use, please contact the EZ Digital service center or sales outlet
Safety Summary
Please take a moment to read these operating instructions thoroughly and completely before operating this instrument. Pay particular attention to WARNINGS used for conditions and actions that pose hazard to the user and CAUTIONS used for conditions and actions that may damage the instrument.
Always to inspect the instrument and other accessories for any sign of damage or
abnormality before every use. Never ground yourself and keep your body isolated from ground.
Never touch exposed wiring, connections or any live circuit conductors.
Do not install substitute parts or perform any unauthorized modification to the
instrument. Use caution when working above 60V DC or 30V AC rms. Such voltages pose a
shock hazard.
Remember that line voltage is present on some power input circuit points such as
on-off switches, fuse, power transformers, etc., even when the equipment is turn off.
Also, remember that high voltage may appear at unexpected points in defective
equipment.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
FAX 781.665.0780 - TestEquipmentDepot.com
3
Page 3
CONTENT
1. PRODUCT DESCRIPTION
1-1. Introduction ------------------------------------------------------ 5 1-2. Technical Specifications -------------------------------------- 6 1-3. Equipment Ratings --------------------------------------------- 9
1-4. Supplied Accessories ------------------------------------------ 10
2. INSTALLATION
2-1. Initial Inspection ------------------------------------------------- 10
2-2. Connecting AC Power ----------------------------------------- 10 2-3. Cooling and Ventilation ---------------------------------------- 10 2-4. Position ------------------------------------------------------------ 10 2-5. Warming-Up ------------------------------------------------------ 10
3. OPERATION
3-1. Controls, indicators and connectors ------------------------ 11 3-2. Operating instruction ------------------------------------------- 17
3-3. Frequency Measurements ------------------------------------- 18 3-4. Period Measurements ------------------------------------------ 19 3-5. Total Measurements -------------------------------------------- 19 3-6. RPM measurement ---------------------------------------------- 19 3-7. Time interval measurement(A-B)----------------------------- 19 3-8. Ratio Measurement (A/B)-------------------------------------- 20
4. MAINTENANCE
4-1. Fuse Replacement ---------------------------------------------- 21 4-2. Adjustment and Calibration ----------------------------------- 21
4-3. Cleaning and decontamination ------------------------------- 21
5. OTHERS
5-1. BNC Cable Considerations ----------------------------------- 22 5-2. Use of Attenuator probes ------------------------------------- 22 5-3. Line Frequency Measurements ------------------------------ 23
1. PRODUCT DESCRIPTION
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
FAX 781.665.0780 - TestEquipmentDepot.com
4
Page 4
1-1. Introduction
This reciprocal UNIVERSAL(FREQUENCY) COUNTER series are microprocessor controlled instrument for frequency measurement at high resolution within a short period of 7 digit display with one second gate time due to uniquely developed LSI as well as the expanding/reciprocal system. It covers a frequency range from 0.1Hz to 3.3 GHz (FC-3000: 3.3GHz, FC-7150U, FC-7150: 1.5GHz, FC-7015U,FC-7015: 100MHz) based on 10MHz time base T.C.O (temperature controlled oscillator) and also featuring,
FC-7150U/FC-7015U Universal Counter
Trigger Function Time Interval Measurement Function Frequency Ratio Measurement Function Common or Separate Input Selection External Frequency Standard Input with 9 Digits LED Display Attenuator Check Period Total Low Pass Filter Line Filter
FC-3000/FC-7150/FC-7015 Frequency Counter
RPM(Rotation Per Minute) Measuring Function External Frequency Standard Input with 9 Digits LED Display Attenuator Check Period Total Low Pass Filter Line Filter
A self-test mode is also provided for a quick check of several facts of operation. Each operating mode can be selected by front panel push button switches with automatic decimal points and indicators. The high accuracy, sensitivity and versatility of this counter make it an extremely valuable instrument to the scientist, engineer, experimenter and communications technician. Light weight and compact size make it practical for use by the hobbyist or field technicians.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
FAX 781.665.0780 - TestEquipmentDepot.com
5
Page 5
1-2. Technical Specifications
INPUT A CHARACTERISTICS
- FREQUENCY RANGE : 0.1Hz to 100MHz (DC Coupled) 30Hz to 100MHz (AC Coupled)
- SENSITIVITY : 0.1Hz to 100MHz: 30mV
- COUPLING : AC or DC Selectable
- IMPEDANCE : 1 M Resistance, Shunted by < 40pF
- ATTENUATOR : x 1 or x 10 Switch Selectable
- LOW PASS FILTER : -3dB Point of 100KHz, Switch Selectable
- TRIGGER LEVEL : + 350mV to - 350mV (PRESET 0V)
(FC-7150U, FC-7015U only)
- SLOPE : Positive or Negative Slope Switch Selectable
(FC-7150U, FC-7015U only)
* NOTE: Trigger error is typically ±0.3% of reading by the number of cycles averaged for input signals having better than 40dB S/N ratio and greater than 100mV amplitude.
- RESOLUTION AND NUMBER OF DISPLAYED DIGIT
Time Base Selector INT EXT INT EXT INT EXT INT EXT Gate Time 0.01S 0.1S 1S 10S Number Of Displayed Digit 5 6 6 7 7 8 8 9 Frequency (Input A,B) RESOLUTION
0.1 Hz-0.99 Hz 1 Hz-9.9 Hz 10 Hz-99 Hz 100 Hz-999 Hz 1 KHz-9.9 KHz 10 KHz-99 KHz 100 KHz-999 KHz 1 MHz-9.9 MHz 10 MHz-99 MHz 100 MHz
10 uHz 1 uHz 1 uHz 0.1 uHz 0.1 uHz 10nHz 10nHz 1nHz
0.1 mHz 10 uHz 10 uHz 1 uHz 1 uHz 0.1 uHz 0.1 uHz 10nHz 1 mHZ 0.1 mHz 0.1 mHz 10 uHz 10 uHz 1 uHz 1 uHz 0.1 uHz 10 mHz 1 mHz 1 mHz 0.1 mHz 0.1 mHz 10 uHz 10 uHz 1 uHz
0.1 Hz 10 mHz 10 mHz 1 mHz 1 mHz 0.1 mHz 0.1 mHz 10 uHz 1 Hz 0.1 Hz 0.1 Hz 10 mHz 10 mHz 1 mHz 1 mHZ 0.1 mHz 10 Hz 1 Hz 1 Hz 0.1 Hz 0.1 Hz 10 mHz 10 mHz 1 mHZ 100 Hz 10 Hz 10 Hz 1 Hz Hz 0.1 Hz 0.1 Hz 10 mHz 1 KHz 100 Hz 100 Hz 10 Hz 10 Hz 1 Hz 1 Hz 0.1 Hz 10 KHz 1 KHz 1 KHz 100 Hz 100 Hz 10 Hz 10 Hz 1 Hz
- ACCURACY : ± Time base Error ± Resolution(Table 1)
- PERIOD RANGE : 10 nS to 10 S
- DISPLAY : n. u. m., Sec with decimal point
- TOTAL RANGE : DC to 30 MHz
- CAPACITY : 0 to 999 999 999
- OVER FLOW : “OF”
7
- RPM RANGE : 6 to 600 x 10
RPM , OVER FLOW : “OF”
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
FAX 781.665.0780 - TestEquipmentDepot.com
6
Page 6
<FC-3000, FC-7150, FC-7015 only>
- MAX. INPUT VOLTAGE LEVEL
FIG 1. MAX. Input Level. (input A,B)
INPUT B. CHARACTERISTICS <FC-7150U, FC-7015U>
- FREQUENCY RANGE : 0.1Hz to 100MHz (DC Coupled) 30Hz to 100MHz (AC Coupled)
- SENSITIVITY : 0.1Hz to 100MHz: 30mV
- COUPLING : AC or DC Selectable
- IMPEDANCE : 1 MΩ Resistance, Shunted by < 40pF
- ATTENUATOR : x 1 or x 10 Switch Selectable
- LOW PASS FILTER : -3dB Point of 100KHz, Switch Selectable
- SLOPE : Positive or Negative Switch Selectable
* Resolution and Number of displayed digit (Table 1): Same as INPUT A
* Max. INPUT VOLTAGE LEVEL (FIG 1): Same as INPUT A
TIME INTERVAL(AB) <FC-7150U, FC-7015U>
- RANGE : 0.1 uSec-10Sec(0.1Hz-10MHz)
- LSD : 100nSec
- RESOLUTION : ± LSD ± Trigger Error*
- ACCURACY : ± LSD ± Trigger Error ±Time base error x T.l
- MULTIPLIER : 1,10,100,1000 (Gate Time: 10S,1S,0.1S,0.01S)
RATIO(A/B) <FC-7150U, FC-7015U>
- RANGE : 0.1Hz to 10MHz(input A)
: 0.1Hz to 10MHz(input B)
- RESOLUTION : ± LSD ±(B TrIg. ERROR x FREQ. A)/N
- ACCURACY : ±1 COUNT of A ± B TRIG. ERROR x FREQ. A.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
FAX 781.665.0780 - TestEquipmentDepot.com
7
Page 7
INPUT C. CHARACTERISTICS
<FC-7150U, FC-7150>
- FREQUENCY RANGE : 80 MHz to 1.5 GHz
- SENSITIVITY : 80 MHz to 1.1 GHz: 35 mV
1.1 GHz to 1.5 GHz: 70 mV
- COUPLING : AC only
- IMPEDANCE : 50 ± 5%
- MAX. INPUT LEVEL : 3 Vrms sine wave
<FC-3000>
- FREQUENCY RANGE : 80 MHz to 3.3 GHz
- SENSITIVITY : 80 MHz to 2.0 GHz: 10 mV
2.0 GHz to 3.0 GHz: 20 mV
3.0 GHz to 3.3 GHz : 35mV
- COUPLING : AC only
- IMPEDANCE : 50 ± 5%
- MAX. INPUT LEVEL : 3 Vrms sine wave
- RESOLUTION AND NUMBER OF DISPLAYED DIGIT Time Base Selector INT EXT INT EXT INT EXT INT EXT
Gate Time 0.01S 0.1S 1S 10S Number Of Displayed Digit 5 6 6 7 7 8 8 9 Frequency (Input B) RESOLUTION 80 MHz-99 MHz 100 MHz-999 MHz 1 GHz-3.3 GHz
1 KHz 100 Hz 100 Hz 10 Hz 10 Hz 1 Hz 1 Hz 0.1 Hz 10 KHz 1 KHz 1 KHz 100 Hz 100 Hz 10 Hz 10 Hz 1 Hz 100 KHz 10 KHz 10 KHz 1 KHz 1 KHz 100 Hz 100 Hz 10 Hz
(Table 2) TIME BASE CHARACTERISTICS
- TYPE : TCO (Temperature controlled oscillator)
- FREQUENCY : 10.000000 MHz
- STABILITY : ±1 PPM(±1 count)
- LINE VOLTAGE STABILITY : Less than ± 1 PPM with ± 10% line voltage
variation.
-TEMPERATURE STABILITY : ±5 PPM from 0° C to 50° C
- MAX. AGING RATE : ±5 PPM/year
- INT. STD. OUT : 10 MHz(internal standard frequency output)
- LEVEL : 1 Vpp or more
- IMPEDANCE : Approx. 600
- EXT. STD. IN : 10 MHz(external standard frequency input)
- LEVEL : 1.5 Vrms to 5Vrms
- IMPEDANCE : Approx. 600
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
FAX 781.665.0780 - TestEquipmentDepot.com
8
Page 8
DISPLAY CHARACTERISTICS
- DISPLAY : Nine Digit 0.56" LED with M/n, K/u, Hz, m, Sec, G.T, Hold, and “OF” indicators. Function and Gate time: User Selected. “OF” Display Shown When Count exceeds 999 999 999
- HOLD : In Frequency and Period, TOTAL. Mode measurement in progress is stopped, and the last complete measurement is displayed. When Hold is released, a new measurement begins.
- GATE TIME : Dipending on input frequency < 10mS------------Somewhere between 0.9 and 9mS < 0.1S--------------Somewhere between 9 and 90mS < 1S---------------- Somewhere between 90 and 900mS < 10S---------------Somewhere between 0.9 and 9S
NOTE: LAST MEASUREMENT DISPLAY WILL REMAIN FOR 10 SECONDS AFTER SIGNAL OFF.
DIMENSION AND WEIGHT
- Dimensions : 255(W) x 90(H) x 255(D)mm
- Weight : Approx. 2.0 kg.
1-3. Equipment Ratings
Power & Fuse Ratings
Power Max.
Input Voltage Fuse
FC-7150U,FC-7015U FC-3000, FC-7150,
FC-7015
103 126V AC(50/60Hz) F 0.5A 250V 15W 10W 206 252V AC(50/60Hz) F 0.25A 250V 15W 10W
Operating Environment TEMPERATURE : 0 ° C to + 40 ° C (Accuracy Specified at 25° C ± 5 ° C)
HUMIDITY : Up to 85% to 40°C without temperature extremes causing condensation within the instrument.
Storage Environment TEMPERATURE : -20° C to +70° C HUMIDITY : below 85% RH
Insulation Category II : Portable equipment of local level. Pollution Degree : 2 Protection to IEC 529 : Ordinary
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
FAX 781.665.0780 - TestEquipmentDepot.com
9
Page 9
1-4. Supplied Accessories
Operation Manual ----------------------------------------------------------------------------- 1 BNC cable -------------------------------------------------------------------------------------- 1 Power cord-------------------------------------------------------------------------------------- 1 Spare Fuse-------------------------------------------------------------------------------------- 1
Specifications are subject to change without notice.
2. INSTALLATION
2-1. Initial Inspection This instrument was carefully inspected both mechanically and electrically before
shipment. It should be physically free of damage. To confirm this, the instrument should be inspected for physical damage in transit. Also, check for supplied accessories.
2-2. Connecting AC Power This instrument requires ac 230V 50-60 Hz power through 3-conductor ac power
cable to be fit into three-contact electrical outlet to secure grounding. If forced to use 2­conductor cable, use ground terminal in rear panel for grounding instrument.
CAUTION
THIS INSTRUMENT IS SET TO AC 230V. BEFORE POWERING ON THIS INSTRUMENT, MAKE SURE THE VOLTAGE OF THE POWER SOURCE IS AC 230V.
2-3. Cooling and Ventilation No special cooling and ventilation is required. However, the instrument should be
operated where the ambient temperature is maintained. 2-4. Position This instrument is built as a bench-type instrument with rubber feet and tilt stand in
place. Stand-up angle can be adjusted by rotating angle of carrying handle. 2-5. Warming-Up
Allow more than 20 minutes for the unit to warm up so that it is stabilized and ready for use.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
FAX 781.665.0780 - TestEquipmentDepot.com
10
Page 10
3. OPERATION 3-1. Controls, indicators and connectors 3-1-1. FC-7150U, FC-7015U Universal Counter
FIG 2. FC-7150U FRONT PANEL
GATE INDICATOR : The gate light, when lit, indicates the main gate is open
and measurement in progress.
DISPLAY : 9 digit(O.56") green LED display used for all read
readings.
NOTE: LAST MEASUREMENT DISPLAY WILL REMAIN FOR 10 SECONDS AFTER SIGNAL OFF.
UNIT INDICATOR : When lit, indicates that the frequency displayed is in KHz
or MHz, and period is in n, u or mSec.
HOLD INDICATOR : When lit, engaged the hold function.
HOLD SWITCH : In hold function the display held but the counters
continue to increment. When the hold is released, the display is updated and resumes counting.
RS-232C INDICATOR : TX(transmitting), RX(receiving) blinking(OPTION)
TRIG. LEVEL VR : Adjusts trigger threshold level on the input A input signal.
Pushing the knob in (preset) sets this level at the mid­ point of a symmetrical sinewave input. Pulling the knob out and rotating it varies the level from negative to positive around the mid point.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
FAX 781.665.0780 - TestEquipmentDepot.com
11
Page 11
INPUT A, BNC : Input for frequency measurements below 100MHz.
Female BNC connector terminated in 1 M input resistance, Shunted by < 40pF capacitance.
INPUT C, BNC : Input for frequency measurements above 80MHz female
<FC-7150U Only> BNC connector terminated in 50Ω.
INPUT B, BNC : Input for frequency measurements below 100MHz
Female BNC connector terminated in 1 M input
resistance, shunted by < 40pF capacitance.
COUPLE. SWITCH : The switch is used to select the input-coupling mode AC or
DC.
ATT. SWITCH : When this switch is set to x 10 (pushed in) the input B is
attenuated 10 :1 before application to the counter with the switch set to x 1 (pushed out), the input B signal is applied. The attenuator has no effect on the input C.
LOW PASS FILTER : With this switch pushed in, the input B is routed through
a SWITCH(LPF) low-pass filter with-3dB point of approximately 100KHz. When it is released, the input B signal is applied to the counter. Same as input B function for input A.
OVER FLOW INDICATOR : OF is displayed when overflow. COM/SEP. : COM. or SEPARATION for INPUT A,B SLOPE SWITCH : Selects positive going or negative going edge of input A,
B signal for triggering. When pushed in, negative going edge is selected; when pushed out, positive going edge is selected.
TILT STAND : Pull out to adjust Tilt. GATE TIME SWITCH : This switch select the degree of resolution of the display
in all modes except TOTAL.
FUNCTION SWITCH : Select the desired operating mode.
a. FREQ. A. When this mode is selected, the counter reads the frequency of the
input A. Resolution is selected using the GATE TIME.
b. FREQ. B. When this mode is selected, the counter reads the frequency of the
input B. Resolution is selected using the GATE TIME.
c. FREQ. C<FC-7150U only>: When this mode is selected, the counter reads the
frequency of the input C.All readings are in MHz.
d. PERIOD A. When this mode is selected, the counter read the period of the input
A. Resolution is selected using the GATE TIME.
e. TOTAL A. When this mode is selected, the unit counts cycles of the input A
signal and continuously displays that count.
f. T.l (A→B). When this mode is selected, the unit measures the time interval from
an edge of the input A signal to an edge of the input B signal. Positive going or
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
FAX 781.665.0780 - TestEquipmentDepot.com
12
Page 12
Negative going edge of each signal is selected By the slope switches.
g. RATIO (A/B). When this mode is selected, the unit measures the ratio of input A
frequency to the input B frequency.
POWER SWITCH : Push in the unit ON and push out the unit power OFF.
3
VOLTAGE POWER MAX
FUSE
115V
0.5AF
0.25AF
10W 10W
115V 230V
230V
SN :
2
54
16
FIG 3.
VOLTAGE SELECTOR : Select ac power, 115V AC or 230V AC.
FC-7150U/FC-7015U REAR PANEL
AC INLET :For connecting of the supplied ac power INT/EXT TIME BASE SELECTOR : Select the time base, switch position
EXT. STD.IN provides a nominal 600 input impedance
path for an external 10 MHz time base. switch position INT. STD.OUT monitors the internal time base signal.
GROUND TERMINAL
FUSE HOLDER :Replacing fuse with unscrewing INT/EXT TIME BASE BNC: Serves as a monitoring point for the internal time
base signal, or provides an input path for an external
time base signal depending on the INT/EXT switch setting. The external signal should have a voltage range of
1.5V~5 Vrms.
RS-232C CONNECTOR : Connector for serial interfacing with computer.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
FAX 781.665.0780 - TestEquipmentDepot.com
13
Page 13
3-1-2. FC-3000, FC-7150, FC-7015 Frequency Counter
FIG 4. FC-7150 FRONT PANEL
1
62
14
3GHz FREQUENCY COUNTER
FC−3000
0.FTX
RX
RMT
G.T
A
AC
TOTAL
PRIOD RPM
FREQ.
POWER
OFFON
FUNCTION
20
AA
HOLD
519 18
GATE TIME
MULTIPLIER
TOTAL RESET
3
M/n
Hz
INPUT C
SEC
3V MAX.
H
INPUT A
ATT 1/1
1/10
50
80MHz
3GHz
INPUT A
LPF OFF
ON
100MHz
250V MAX.
12 13 17
10S1SS1.0 0 0.1S
COUPLE
k/u
m
AC
DC
11
948
1M
FIG 5. FC-3000 FRONT PANEL
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
FAX 781.665.0780 - TestEquipmentDepot.com
14
Page 14
GATE INDICATOR : The gate light, when lit, indicates the main gate is open
and measurement in progress.
DISPLAY : 9 digit(O.56") green LED display used for all read
readings.
NOTE: LAST MEASUREMENT DISPLAY WILL REMAIN FOR 10 SECONDS AFTER SIGNAL OFF.
UNIT INDICATOR : When lit, indicates that the frequency displayed is in KHz
or MHz, and period is in n, u or mSec.
HOLD INDICATOR : When lit, engaged the hold function.
HOLD SWITCH : In hold function the display held but the counters
continue to increment. When the hold is released, the display is updated and resumes counting.
RS-232C INDICATOR : TX(transmitting), RX(receiving) blinking(OPTION)
INPUT A, BNC : Input for frequency measurements below 100MHz.
Female BNC connector terminated in 1 MΩ input
resistance, Shunted by < 40pF capacitance. INPUT C, BNC : Input for frequency measurements above 80MHz female BNC connector terminated in 50Ω. COUPLE. SWITCH : The switch is used to select the input-coupling mode AC
or DC.
ATT. SWITCH : When this switch is set to x 10 (pushed in) the input B is
attenuated 10 :1 before application to the counter with
the switch set to x 1 (pushed out), the input B signal is
applied. The attenuator has no effect on the input C. LOW PASS FILTER : With this switch pushed in, the input B is routed through
a SWITCH(LPF) low-pass filter with-3dB point of
approximately 100KHz. When it is released, the input B
signal is applied to the counter. Same as input B
function for input A.
OVER FLOW INDICATOR : OF is displayed when overflow. TILT STAND : Pull out to adjust Tilt.
GATE TIME SWITCH : This switch select the degree of resolution of the display
in all modes except TOTAL. FUNCTION SWITCH : Select the desired operating mode.
a.FREQ. A. : When this mode is selected, the counter reads the frequencyof
the input A. Resolution is selected using the GATE TIME.
b. FREQ. C
<FC-3000,7150 >
: When this mode is selected, the counter reads the
frequency of the input C.All readings are in MHz.
c. PERIOD A. : When this mode is selected, the counter read the period of the
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
FAX 781.665.0780 - TestEquipmentDepot.com
15
Page 15
input A. Resolution is selected using the GATE TIME.
e. TOTAL A. : When this mode is selected, the unit counts cycles of the input A
signal and continuously displays that count.
f. RPM A. : When this mode is selected, the unit displays the RPM(rotation per
minute) of the input A signal.
POWER SWITCH : Push in the unit ON and push out the unit power OFF.
3
VOLTAGE POWER MAX
FUSE
115V
0.5AF
0.25AF
10W 10W
115V 230V
SN :
230V
2
54
FIG 6. FC-7150/FC-7015 REAR PANEL
7
RS−232C
INT STD OUT
EXT STD IN
FIG 7 . FC-3000 REAR PANEL
36
WARNING
TO PREVENT ELECTRIC SHOCK HAZARD AND FIRE, REPLACE ONLY WITH FUSE OF SAME RATING AND TYPE. DISCONNECT POWER SUPPLY BEFORE REPLACING FUSE.
DO NOT REMOVE COVER. REFER SERVICING TO QUALIFIED PERSONNEL.
16
4
230V~, 50Hz/60Hz
FUSE
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
FAX 781.665.0780 - TestEquipmentDepot.com
16
Page 16
VOLTAGE SELECTOR : Select ac power, 115V AC or 230V AC.
AC INLET :For connecting of the supplied ac power INT/EXT TIME BASE SELECTOR : Select the time base, switch position
EXT. STD.IN provides a nominal 600 input impedance
path for an external 10 MHz time base. switch position INT. STD.OUT monitors the internal time base signal.
GROUND TERMINAL
FUSE HOLDER :Replacing fuse with unscrewing INT/EXT TIME BASE BNC: Serves as a monitoring point for the internal time
base signal, or provides an input path for an external
time base signal depending on the INT/EXT switch setting. The external signal should have a voltage range of
1.5V~5 Vrms.
RS-232C CONNECTOR :
Connector for serial interfacing with computer.
3-2. Operating Introductions
Below is the basic operating information needed for frequency counter.
a. Connect the unit to AC power cord into receptacle on rear panel and
plug into AC inlet. b. To Turn on equipment, push power on-off switch on. c. set the function indicator position to FREQ A and Gate time indicator to 1 Sec position.
CAUTION
1. APPLICATION OF INPUT VOLTAGES HIGHER THAN THE LIMITS LISTED IN THE SPECIFICATIONS SECTION MAY DAMAGE THE COUNTER. BEFORE APPLYING ANY SIGNAL TO THE INPUTS, MAKE CERTAIN THAT IT DOES NOT EXCEED THESE SPECIFIED MAXIMUMS.
2. FREQUENCY COUNTER GROUND POINTS ARE CONNECTED DIRECTLY TO EARTH GROUND. ALWAYS CONNECT FREQUENCY COUNTER GROUND ONLY TO GROUND POINTS IN THE CIRCUIT UNDER TEST.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
FAX 781.665.0780 - TestEquipmentDepot.com
17
Page 17
3-3. Frequency Measurements
3-3-1 INPUT A, B (0.1 Hz to 100 MHz)
a. Apply the signal to be measured to the input A and/or B(FC-7150U, FC-
7015U only) BNC.
b. Set the FREQ. A and/or B(FC-7150U, FC-7015U only) of function switch.
c. Select the degree of resolution desired, using the gate time selector switch. d. Frequency is given by the display. The gate indicator lights while each measurement in progress, and the display is updated at the end of each measurement interval.
e. Engaging the hold switch “freezes” the display at the existing reading,
When hold is released, the display is updated and resumes counting. f. If necessary, engage the attenuator switch When set to x 10 (pushed
in), this switch attenuates the input A and/or B(FC-7150U, FC-7015U only)
signal by a factor of approximately 10 before application to the counter. This helps prevent miscounting caused by noisy or improperly terminated
high amplitude signals. g. If necessary, engage the LPF (low pass filter) switch This route the input A and/or B through a low pass filter (-3 dB point of approximately 100 KHz) before application to the frequency counter. This helps eliminate counting errors in low frequency measurements by minimizing effects of high frequency noise present on the input. h. When measurement the lower cut-off frequency (10 Hz), pushed in DC coupling position.
3-3-2 INPUT C
FC-7150U, FC-7150: 80 MHz to 1.5 GHz
FC-3000 : 80MHz to 3.3GHz
CAUTION
THE MAXIMUM INPUT LIMIT TO THIS INPUT IS 3 Vrms MAXIMUM OVER THE INPUT FREQUENCY RANGE. THE X 10 ATTENUATOR DOES NOT APPLY.
a. Apply the signal to be measured to the input C. BNC. b. Set the function indicator to the FREQ. C position. c. Select the degree of resolution desired, using the gate time switch.
d. Frequency is given by the display. The indicator lights while each measurement in progress.
e. Engaging the hold switch “freezes” the display at the existing reading,
When hold is released, the display is updated and resumes counting.
f. The attenuator and LPF coupling switch have no effect in input C.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
FAX 781.665.0780 - TestEquipmentDepot.com
18
Page 18
3-4. Period Measurements
a. Apply the signal to be measured to the input A BNC. b. Select the degree of resolution desired, using the gate time switch.
c.Period is given by the display. The gate indicator lights while each measurement is in progress.
d. attenuator, low pass fitter, coupling switch application is same as
frequency measurements mode.
3-5. Total Measurements
The totalize mode is used to count the total number of events occurring during a specific time period. Maximum frequency is 30 MHz.
a. Set the totalize mode. Any gate and units setting is ignored. b. Apply the signal to be measured to input A , and then the counter display is the count continually. Maximum count is 999999999. if this is exceed
the overflow message display as ”OF” c. Low pass filter and attenuator, coupling switch application is same as frequency measurements mode.
NOTE THE HOLD SWITCH MAY BE USED TO BE LATCH THE DISPLAY. HOWEVER, THE COUNTER TO INCREMENT AND WHEN THE HOLD IS RELEASED, THE UPDATE COUNT IS DISPLAY.
3-6. RPM (Rotation Per Minute) Measurement <FC-3000, FC-7150, FC-7015>
a. Apply the signal to be measured to the input A BNC, and then the counter displays the RPM. Maximum count is 999999999. If this is exceed, the overflow message displays as ”OF”
b. Low pass filter and attenuator, coupling switch application is same as frequency measurements mode.
3-7. Time Interval Measurements(AB) <FC-7150U, FC-7015U>
In time interval mode, unit measures the elapsed time from a selected edge of the input A waveform to a selected edge of the input B waveform. For a stable reading, the two input signals should be related to each other such that this time interval remains reasonably constant from one measurement to the next. For example, two digital waveforms derived from the same clock would be suitable; two arbitrary frequencies from separate function generators would not.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
FAX 781.665.0780 - TestEquipmentDepot.com
19
Page 19
a. Connect the signal to be measured to the input A and input B.
b. Set the T.l(AB) of function switch. The switch label serves as a reminder that the measurement starts at the input A edge and stops at the input B edge.
c. Select desired edge of each waveform using the input A, and input B slope switches push switch in for negative-going edge, leave out for positive-going edge.
d. Set a trigger level control to preset (Push In). This ensures that both input A and input B are triggering at the same level (approximately the average) on their respective waveforms.
e. Select the degree of resolution desired, using the gate time switch
f. Time interval is given by the display, the gate indicator lights while each measurement is in progress, and the display is updated at the end of each measurement interval.
g. Press the hold switch, “freezes” the display at the existing reading when hold is released, a new measurement begins (Gate Indicator Lights), but the display continues to hold the old reading until the new measurement is completed.
Frequency Ratio Measurements (A/B) <FC-7150U, FC-7015U>
3-8.
In this mode of operation, the counter displays the ratio of the frequency applied to input A to the frequency applied to input B. The input A frequency should preferably be equal to or greater than that of input B. And both frequencies must be within the limits given in the "SPECIFICATIONS" section. Frequency ratio is determined by counting the number of input A cycles occurring during a specified number of input B cycles (1.10.100.1000) and applying the result, with a proper decimal point, to the display.
a. Connect a signals to be measured to the input A and input B
NOTE
BOTH INPUTS MAY BE CONNECTED TO THE SAME SIGNAL FOR A RATIO OF
1.000000 AT THE GATE TIME 1 SEC. THE UNIT INDICATORS ARE OFF BECAUSE THE READING DISPLAYED IS RATIO.
b. Set the ratio A/B of function switch.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
FAX 781.665.0780 - TestEquipmentDepot.com
20
Page 20
c. Select the resolution desired using the gate time switches.
d. Frequency ratio is given by the display the gate indicator lights while each measurement is in progress, and the display is updated at the end of each measurement interval. e. Press the hold switch "freezes" the display at the existing reading. When hold is released, a new measurement begins, but the display continues to hold the old reading until the new measurement is completed.
4. MAINTENANCE
CAUTION
IT IS ESSENTIAL FOR SAFETY TO PROPERLY MAINTAIN AND SERVICE THIS INSTRUMENT
WARNING
VOLTAGES WITHIN THIS INSTRUMENT ARE SUFFICIENTLY HIGH TO ENDANGER LIFE. COVERS MUST NOT BE REMOVED EXCEPT BY PERSONS QUALIFIED AND AUTHORIZED TO DO SO AND THESE PERSONS SHOULD ALWAYS TAKE EXTREME CARE ONCE THE COVERS HAVE BEEN REMOVED.
4-1. Fuse Replacement
Disconnect and remove all connections from any live power source.
Unscrew fuse holder by screw driver.
Locate the defective fuse and remove it by gently pulling-out.
Install a new fuse of the SAME SIZE AND RATING.
Screwing fuse holder.
CAUTION
MAKE SURE THAT THE RATED AND SPECIFIED FUSES ARE USED FOR REPLACEMENT.
4-2. Adjustment And Calibration
It is recommendable to regularly adjust and calibrate this instrument. Qualified and authorized personnel only should execute performance and procedures
4-3. Cleaning and decontamination
The instrument can be cleaned with a soft clean cloth to remove any oil, grease or grime. Never use liquid solvents or detergents. If the instrument gets wet for any reason, dry the instrument using low-pressure clean air at less than 25 PSI. Use care and caution around the window cover areas where water or air could enter into the instrument while drying.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
FAX 781.665.0780 - TestEquipmentDepot.com
21
Page 21
5. OTHERS
5-1. BNC Cable Considerations Accuracy of radio frequency measurements can be affected by connections between
signal source and counter. Main considerations are standing waves and shunt cable capacitance.
Standing wave is usually present due to reflections when a transmission line is not terminated in its characteristic impedance. These standing waves may cause damage to the signal source or produce inaccurate measurements, and their effects increase as cable length reaches one-fourth of the wavelength for the frequency being measure. Standing wave can be minimized by keeping cable lengths short, and more important, providing a proper termination.
The cable's characteristic impedance and the terminating impedance should match the source impedance. For example, for a source impedance of 50, use 50 coaxial cable terminated with a 50 resistive load.
Use DC blocking capacitor in situations where bias voltage or other DC voltages could be affected by the termination resistor. Shunt cable capacitance, which can cause undesirable signal attenuation, increases with increased cable length. It is recommended that for radio frequency measurements, the cable be no longer than three feet (9O cm), to keep shunt capacitance within acceptable limit.
In 50 systems the internal 50Ω input termination of the input B. BNC minimizes reflections and the resulting standing waves. Thus, the need for an external termination is eliminated. Also, shunt capacitance has a much lesser effect at this BNC then at the input A, and the above restriction on cable length is reduced. However, prescale measurements must always be taken from a 50 point in the circuit under test.
5-2. Use of Attenuator Probes
Input A resistance (1 M ) and input capacitance (< 40 pF) are independent of the ATT switch. To decrease loading, a high impedance oscilloscope probe such as the following may be used with input A. Use the probe in the x 10 position whenever possible for less circuit loading.
NOTE
DO NOT USE A 10:1 PROBE WITH THE INPUT C, THE PROBE IS DESIGNED FOR 10:1 ATTENUATOR WITH A COUNTER INPUT RESISTANCE OF 1 M . THE 50 TERMINATION OF THE INPUT C WOULD RESULT IN UNACCEPTABLY HIGH ATTENUATION.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
FAX 781.665.0780 - TestEquipmentDepot.com
22
Page 22
5-3. Line Frequency Measurements
Use of the attenuator, low pass filter, and/or x10 probe is advisable when measuring line frequency because the high amplitude signal and noise can cause wrong counting.
WARNING
USE CAUTION IN MEASURING THE LINE FREQUENCY OF AN AC OUTLET. USING THE PROBE TIP ONLY, MEASURE BOTH SIDES OF THE LINE. THE GROUND SIDE WILL GIVE A ZERO READING AND THE HOT SIDE WILL PROVIDE THE DESIRED MEASUREMENT. DO NOT USE THE "GROUND "LEAD OF THE PROBE. REMEMBER THAT THE CHASSIS OF THE COUNTER AND THE “GROUND” LEAD OF THE PROBE ARE ALREADY AT EARTH GROUND (VIA THE 3-WIRE POWER CORD OF THE INSTRUMENT). TOUCHING THE “GROUND "LEAD TO THE 'HOT" SIDE OF THE LINE WOULD PLACE A DIRECT SHORT ON THE POWER LINE THROUGH THE PROBE CABLE, RESULTING IN POSSIBLE INJURY AND DAMAGE TO THE PROBE CABLE.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
FAX 781.665.0780 - TestEquipmentDepot.com
23