Congratulations
intended for scanning groups of individuals or monitoring an individual for elevated temperatures. It
is not a substitute for a clinical thermometer. Always use a clinical thermometer when high accuracy
body temperature measurements are required.
on your purchase of the Model IR200 IR Thermometer. This thermometer is
Considerations and Warnings
Read all instructions before using this product.
For the most accurate results, make reading with an ambient (room) temperature of 73 to
82°F (23 to 27°C)
Make sure the skin is dry and that no hair interferes with the measurement.
Do not immerse the meter in water.
Avoid touching and/or scratching the infrared sensor lens.
Always use a clinical thermometer to verify any abnormal temperature measurements.
Clean the lens area by gently blowing with compressed air and use a damp swab to wipe
the lens. Do not use any solvents to clean the lens.
Measurement Notes
1.If the meter has been stored in a cold or hot environment, allow it at least 20 minutes to
acclimate to room temperatu
To prevent transmission of disease, avoid direct contact with th
2.
After heavy exercise always wait at least 10 minutes before taking forehead temperat
3.
readings.
re before making measurements.
e skin.
ure
2
IR200-en-GB_V1.6 10/16
Page 3
Meter Description
1. IR
sensor
2.
Body – Surface switch
3.▼ button
4.▲ button
5. MODE
6. Measurement
7. LCD
8. Batter
DISPL
1. Surface
2. Bod
3. Temper
4. Memor
5. Alarm
6. Temper
7.
8.
button
displa
y compartm
AY
mode
y mode
ature displa
y
location
active
ature un
Low battery icon
Memory temperature displa
y
trigger
ent
its
1
2
3
4
5
6
7
8
y
y
1
2
3
4
5
6
7
8
3
IR200-en-GB_V1.6 10/16
Page 4
Operating Instructions
Body Temperature Measurements
1.
Set the Body-Surface- switch for “Body” for accurate measurements in the range of 89
°
108.5
F (32.0 to 42.5°C).
2.Hold the meter by its handle and point it toward the surface to be measured. Meas
distance should be 2 to 6”
Press the trigger to turn the meter on and take a temperature reading. The temperatur
3.
reading appe
smaller
Release the Trigger and the reading will hold for approximately 7 seconds after whic
4.
meter
quickly log samples.
If the temperature goes above 109°F or 43°C, Hi will appear in the display.
5.
If the temperature goes below 86°F or 30°C, Lo will appear in the display.
Body Temperature Mode settings
1.
With the meter OFF, press the MODE button once to set the C/F temperatur
temper
Press the MODE button a second time to set the alarm temperature limit. Press the ▲ or ▼
2.
buttons to chang
Press the MODE button a third time to enter the long term calibration drift correction mode.
3.
entering the mo
make a correction, measure a known, fixed temperature source. Enter the corr
press the ▲ or ▼ buttons to cha
readings. Repeat
IR200 matches t
4.
Press the MODE button a fourth time to set the alarm buzzer status. Press the ▲ or ▼
to sw
Surface Temperature Measurements
1.
Set the Body-Surface- switch for “Surface” for wide range surface measurements of
materials.
2.
Hold the meter by its handle and point it toward the surface to be
3.
Press the trigger to turn the meter on and take a temperature reading. The temperatur
reading appe
smaller
4.
Release the Trigger and the reading will hold for approximately 7 seconds after whic
meter
quickly log samples.
If the temperature goes above 140°F or 60°C, “Hi” will appear in the display.
5.
If the temperature goes below 32°F or 0°C, “Lo” will appear in the display.
ars in the large display and the logged data number and value appear in
display.
will automatically shut off. Trigger presses can occur up to a rate of one a se
ature units will flash. Press the ▲ or ▼ buttons to change th
e the value.
de, the previous temperature correction factor will appear on the display. To
and adjust the correction value as needed until the measurement
he known temperature.
itch from ON to OFF.
ars in the large display and the logged data number and value appear in
display.
will automatically shut off. Trigger presses can occur up to a rate of one a se
(5 to 15
cm).
e units.
nge the correction value and minimize the differenc
measured.
.6.0 to
urement
e
the
h the
cond to
e units. The
On
ection mode and
e in
on the
buttons
most
e
the
h the
cond to
4
IR200-en-GB_V1.6 10/16
Page 5
Surface Temperature Mode settings
1.
With the meter OFF, press the MODE button once to set the C/F temperatur
temper
ature units will flash. Press the ▲or ▼ buttons to change the
2.
Press the MODE button a second time to set the alarm temperature limit. Press the ▲or ▼
buttons to chang
Press the MODE button a third time to set the alarm buzzer status. Press the ▲or ▼
3.
Data memory
The meter can store up to 32 readings in the memory. To review the logged data points, press
both the ▲ and ▼ buttons while the unit is OFF. Use the ▲ or ▼ buttons to scroll through the
memory locations. To clear logged data, select the “0” memory location and press the MODE
button. The unit will beep twice to indicate memory has been cleared.
Battery Replacement
When the low battery symbol appears on the display, replace the meter’s battery. The battery
compartment is located on the bottom of the handle. Open the compartment by removing one screw
and sliding the cover off. Replace the (2) “AA” batteries and close the battery compartment cover.
itch from ON to OFF.
to sw
Never dispose of used batteries or rechargeable batteries in household waste.
As consumers, users are legally required to take used batteries to appropriate collection
sites, the retail store where the batteries were purchased, or wherever batteries are sold.
Disposal: Do not dispose of this instrument in household waste. The user is obligated to
take end-of-life devices to a designated collection point for the disposal of electrical and
electronic equipment
e the value.
.
e units. The
units.
buttons
IR Measure
1.
2.
3.
4.
5.
ment Notes
Before measuring, be sure to clean surfaces that are covered with fr
If an object's surface is highly reflective, apply ma
before measu
covering.
Measurements through transparent surfaces such as glass may not be accurate.
Steam, dust, smoke, etc. can obscure measuremen
The meter automatically compensates for deviations in ambient temperature. Ho
take up to 30 minutes for the meter to adjust to extremely wide changes.
ring. Allow time for the paint or tape to adjust to the temperature of the
sking tape or flat black paint to the surface
ts.
5
ost, oil, grime, etc.
surface it is
wever, it can
IR200-en-GB_V1.6 10/16
Page 6
IR200 Thermometer Cleaning Instructions
The IR200 Non-Contact Forehead IR Thermometer is a non-critical Reusable Medical Equipment
(RME) device.
Please clean and disinfect the housing and lens of this device before and after each use. To clean and
disinfect this device please read and understand these instructions.
Equipment Required
Please obtain Isopropyl Alcohol or a mild antiseptic solution (70% alcohol concentration or higher)
and low-lint soft cloths or cotton swabs.
Housing cleaning
Germicidal wipe products may be used to clean the housing if they contain at least a 55% alcohol
concentration and a 2-minute contact time rule is followed.
Lens cleaning
For lens cleaning please use 70% alcohol concentration only.
Procedure and Notes
1.
To clean the meter housing, moisten a cotton swab, cotton tissue, or cotton cloth lightly wi
alcohol and gently
‘Equipment Required’ section above
2.
Do not use an excessive amount of alcohol while cleaning; do not allow moisture to
through th
To clean the IR lens, lightly moisten a cotton swab with a 70% Isopropy
3.
solution and sw
cleaning the lens
Never submerge the device in liquid, or place it
4.
solvents or abrasives
Allow the device to dry for at least 10 to 15 minutes after cleaning before use
5.
e butto
wipe the device. Alternatively, use a germicidal wipe as described
n access areas
ab the lens. Please allow 10 to 15 minutes drying time before use af
in an autoclave device, or subject it to harsh
th
in the
seep
l alcohol based
ter
6
IR200-en-GB_V1.6 10/16
Page 7
Specifications
RangeAccuracy
Body Temp 32.0 to 42.5°C (89.6 to 108.5°F)±0.3°C/0.5°F
Surface Temp 0 to 60°C (32.0 to 140°F) ±0.8°C/1.5°F
Emissivity 0.95 fixed
Field of View D/S = Approx. 8:1 ratio (D = distance; S = spot or target)
IR Spectral response 6 to 14 m (wavelength)
Display Backlit LCD display with function indicators
Memory store up to 32 readings
Response time 500ms
Over range indication Hi or Lo
Operating Temperature 0
Relative Humidity 10% to 90%RH operating: <80%RH storage.