FL700-en-US_V1.4 09/14
4
Operation
Preparation for use
1. Remove the FL700, electrode module and sample cup from its box. Remove caps from
module
2. If not already done, fit the electrode module onto the end of the meter body, making sure
that the slots line up correctly, and turn the module retaining ring to secure the assembly.
Powering the FL700
Press the ON/OFF button to turn the meter on or off. If the batteries are weak, the ‘BAT’
indicator appears on the LCD.
Power-On Diagnostics
1. When the meter is switched ON the LCD displays “SELF” and “CAL” while the meter runs a
diagnostic routine
2. During this time the meter is recalling the user calibration data, performing self diagnostics &
initializing the circuitry
3. When completed, the meter proceeds to the normal measurement mode
Calibration
The FL700 can be calibrated between 1.0 ppm and 10 ppm or between 0.5 and 5.0 ppm Fluoride
ion. Set the Mode to ppm. The following calibration procedure assumes the normal 1.0 to 10 ppm
range has been chosen.
1. Prepare a 1 ppm fluoride standard solution by placing one TISAB tablet into a sample cup
and add 20 mls of the 1ppm Fluoride standard into the sample cup.
Note: if your 1ppm Fluoride standard already contains TISAB do not use a TISAB tablet.
2. Create a rinse solution that can be used between sample measurements buy dissolving
1 TISAB tablet in 20ml of Distilled water. The rinse solution promotes faster response times.
3. Rinse the end of the FL700 module in this Rinse solution or in Distilled water and then wipe
thoroughly with a paper tissue.
4. Place the FL700 into the prepared 1 ppm standard and switch the instrument ON using the
ON/OFF button. The instrument will now go through its self-calibration.
5. The instrument will enter HOLD mode in about 35 seconds when stabilized in the 1.0 ppm
standard solution.
6. Press and HOLD the CAL key, CAL will appear in the display followed by 0.5ppm and
5.0ppm. Continue holding the CAL button until 1.0 ppm and CAL appear in the display.
Release the CAL key.
7. Wait until the display stops blinking; the instrument will enter the normal measurement
mode.
8. The instrument is now calibrated and ready for use.
9. The circled L and H icons on the display indicate that the low range (L) and high range (H)
calibrations have been completed.
Calibration frequency
Calibrate the FL700 prior to each new measurement batch or if more than 12 hours has elapsed
since the last calibration.