Extech Instruments EX900 User manual

User's Guide
Extech 800 Series 1000 Amp Clamp Meters with IR Thermometer
EX810 AC Clamp meter
EX820 True RMS AC Clamp meter
EX830 AC/DC True RMS Clamp meter
Patented
Find Quality Products Online at: sales@GlobalTestSupply.com
Introduction
Congratulations on your purchase of the EX800 Series Clamp device. The series consists of the following models:
EX810 1000A AC Current Clamp meter
EX820 1000A True RMS AC Current Clamp
EX830 1000A True RMS AC/DC Current Clamp
All models measure:
Infrared Temperature AC/DC Voltage Resistance Capacitance Frequency Continuity Diode
Model specific functions:
EX830: DC Current, True RMS, Type K thermometer, DC Zero EX820: True RMS, Type K thermometer EX810: Average responding
The EX800 Series features:
Auto Power OFF Max/Min recording Data Hold Peak Hold Backlit LCD display
Careful use of this meter will provide many years of reliable service.
EX800 Series Version 3.2 8/07 2
Find Quality Products Online at: sales@GlobalTestSupply.com
Safety
International Safety Symbols
This symbol, adjacent to another symbol or terminal, indicates the user must refer to the manual for further information.
This symbol, adjacent to a terminal, indicates that, under normal use, hazardous voltages may be present
Double insulation
SAFETY NOTES
Do not exceed the maximum allowable input range of any function.
Do not apply voltage to meter when resistance function is selected.
Set the function switch OFF when the meter is not in use.
Remove the battery if meter is to be stored for longer than 60 days.
WARNINGS
Set function switch to the appropriate position before measuring.
When measuring volts do not switch to current/resistance modes.
Do not measure current on a circuit whose voltage exceeds 600V.
When changing ranges always disconnect the test leads from the circuit under test.
UL Note
A UL mark does not indicate that this product has been evaluated for accuracy.
CAUTIONS
Improper use of this meter can cause damage, shock, injury or death. Read and
understand this user manual before operating the meter.
Always remove the test leads before replacing the battery or fuses.
Inspect the condition of the test leads and the meter itself for any damage before
operating the meter. Repair or replace any damage before use.
Use great care when making measurements if the voltages are greater than 25VAC
rms or 35VDC. These voltages are considered a shock hazard.
Always discharge capacitors and remove power from the device under test before
performing Diode, Resistance or Continuity tests.
Voltage checks on electrical outlets can be difficult and misleading because of the
uncertainty of connection to the recessed electrical contacts. Other means should be used to ensure that the terminals are not "live".
If the equipment is used in a manner not specified by the manufacturer, the protection
provided by the equipment may be impaired.
Function Maximum Input
A AC, A DC (A DC on Model EX830 only) 1000A DC/AC
V DC, V AC 600V DC/AC
Resistance, Capacitance, Frequency, Diode Test 250V DC/AC
Type K Temperature (EX820 & EX830 only) 60V DC, 24V AC
EX800 Series Version 3.2 8/07 3
Find Quality Products Online at: sales@GlobalTestSupply.com
Description
Meter Description (EX830 pictured)
1. Current clamp
2. Clamp opening trigger
3. Data Hold Button
4. Mode
5. Peak
6. Range
7. DCA Zero (EX830 only)
8. MIN/MAX
9. Backlit LCD Display
10. Test lead input jacks
11. IR thermometer and laser pointer (rear)
12. Backlight Button
13. Laser pointer button
14. Function switch
Display icons Description
HOLD Data Hold
Minus sign Negative reading display
0 to 3999 Measurement display digits
ZERO DCA Zero (Model EX830 only)
P Peak value
AUTO Auto Range mode
DC/AC Direct Current / Alternating
MAX Max reading
MIN Min reading
Low battery
mV or V Milli-volts or Volts (Voltage)
Ohms (Resistance)
A Amperes (Current)
F Farad (Capacitance)
Hz Hertz (Frequency)
o
F and oC Fahrenheit and Celsius units (Temperature)
n, m, , M, k Unit of measure prefixes: nano, milli, micro, mega, and kilo
)
)
) Continuity test
Current
Diode test
Laser pointer
EX800 Series Version 3.2 8/07 4
Find Quality Products Online at: sales@GlobalTestSupply.com
Specifications
Function
50/60 Hz True RMS on EX820 & EX830
DC Current
EX830 only
AC Voltage
50/60Hz
True RMS on EX820 & EX830
Resistance
Capacitance
Frequency
Range & Resolution
400.0 AAC ± (2.8% + 8d) ± (2.5% + 8d) ± (2.5% + 8d) AC Current
1000 AAC ± (3.0% + 8d) ± (2.8% + 5d) ± (2.8% + 5d)
400.0 ADC ----- ----- ± (2.5% + 5d)
1000 ADC ----- ----- ± (2.8% + 5d)
400.0 mVAC ± (1.5% + 10d) ± (1.0% + 10d) ± (1.0% + 10d)
4.000 VAC
40.00 VAC
400.0 VAC
600 VAC ± (2.5% + 8d) ± (2.0% + 5d) ± (2.0% + 5d)
400.0 mVDC ± (0.8% + 2d) ± (0.8% + 2d) ± (0.8% + 2d)
4.000 VDC
40.00 VDC
400.0 VDC
600 VDC ± (2.0% + 2d) ± (2.0% + 2d) ± (2.0% + 2d)
400.0
4.000k
40.000k
400.0k
4.000M
40.00M
4.000nF ± (5.0% + 30d) ± (5.0% + 30d) ± (5.0% + 30d)
40.00nF ± (5.0% + 20d) ± (5.0% + 20d) ± (5.0% + 20d)
400.0nF
4.000µF
40.00µF
400.0µF ± (4.0% + 10d) ± (4.0% + 10d) ± (4.0% + 10d)
4.000mF ± (10% + 10d) ± (10% + 10d) ± (10% + 10d)
40.00mF unspecified unspecified unspecified
4.000kHz ± (1.5% + 2d) ± (1.5% + 2d) ± (1.5% + 2d)
Sensitivity: 100V (<50Hz); 50V (50 to 400Hz); 5V (401Hz to 4000Hz)
Accuracy
(% of reading + digits)
EX810 EX820 EX830
± (1.8% + 8d) ± (1.5% + 5d) ± (1.5% + 5d)
± (1.5% + 2d) ± (1.5% + 2d) ± (1.5% + 2d) DC Voltage
± (1.0% + 4d)
± (1.0% + 4d) ± (1.0% + 4d)
± (1.5% + 2d) ± (1.5% + 2d) ± (1.5% + 2d)
± (2.5% + 3d) ± (2.5% + 3d) ± (2.5% + 3d)
± (3.5% + 5d) ± (3.5% + 5d) ± (3.5% + 5d)
± (3.0% + 5d) ± (3.0% + 5d) ± (3.0% + 5d)
EX800 Series Version 3.2 8/07 5
Find Quality Products Online at: sales@GlobalTestSupply.com
Function
Temperature (type-K)
Temp (IR)
General Specifications
Clamp jaw opening 1.7" (43mm) approx. Display 3-3/4 digits (4000 counts) backlit LCD Continuity check Threshold 40 ; Test current < 0.5mA Diode test Test current of 0.3mA typical;
Low Battery indication Battery symbol is displayed Over-range indication ‘OL’ display Measurement rate 2 readings per second, nominal PEAK Captures peaks >1ms Thermocouple sensor Type K thermocouple required
IR Spectral response 6 to 16µm IR Emissivity 0.95 fixed
IR distance ratio 8:1 Input Impedance 10M (VDC and VAC) AC bandwidth 50 to 400Hz (AAC and VAC) AC response True rms (AAC and VAC) on EX820 & EX830 Crest Factor 3.0 in 40A and 400A ranges, 1.4 in 1000A range (50/60Hz
Operating Temperature 41 F to 104 F (5 C to 40 C) Storage Temperature -4 F to 140 F (-20 C to 60 C) Operating Humidity Max 80% up to 87 F (31 C) decreasing linearly to 50% at
Storage Humidity <80% Operating Altitude 7000ft. (2000meters) maximum. Battery One (1) 9V Battery (NEDA 1604) Auto power OFF After approx. 25 minutes Dimensions & Weight 10.6x4.3x2” (270x110x50mm); 13.6 oz. (386g) Safety For indoor use and in accordance with the requirements for
Patent notice U.S. Patent 7163336
Range and Resolution
-4 to 1400oF NA ± (3%rdg + 9oF) ± (3%rdg + 9oF)
-20 to 760
-58 to -4 F 9 F 9 F 9 F
-4 to 518 F
-50 to -20 C
-20 to 270 C
o
C NA ± (3%rdg + 5oC) ± (3%rdg + 5oC)
2.0% reading or
whichever is >
2.0% reading or
whichever is >
Open circuit voltage < 3VDC typical
and 5% to 100% of range)
104 F(40 C)
double insulation to IEC1010-1 (2001): EN61010-1 (2001) Overvoltage Category III 600V and Category II 1000V, Pollution Degree 2.
Accuracy (% of reading + digits)
EX810 EX820 EX830
4 F
5 C
2 C
2.0% reading or 4 F
whichever is >
5 C 5 C
2.0% reading or 2 C
whichever is >
2.0% reading or
whichever is >
2.0% reading or
whichever is >
4 F
2 C
EX800 Series Version 3.2 8/07 6
Find Quality Products Online at: sales@GlobalTestSupply.com
Operation
NOTES: Read and understand all Warning and Caution statements in this operation manual prior to using this meter. Set the function select switch to the OFF position when the meter is not in use.
AC/DC Current Measurements (DC Current on Model EX830 only)
WARNING: Ensure that the test leads are disconnected from the meter before making
current clamp measurements.
1. Set the Function switch to the A or ADC range
2. Press the trigger to open jaw. Fully enclose only one conductor. For optimum results, center the conductor in the jaw.
3. The clamp meter LCD will display the reading.
DCA ZERO (EX830 only)
The DC Zero feature removes offset values and improves accuracy for DC current measurements. To perform a zero, select ADC and with no conductor in the jaw:
1. Press the DC ZERO button to zero the display. “ZERO” will appear in the display. The offset value is now stored and removed from all measurements.
2. To view the stored value, press the DC ZERO button. “ZERO” will flash and the stored value will be displayed.
3. To exit this mode, press and Hold the ZERO button until “ZERO” is no longer in the display.
AC
AC/DC Voltage Measurements
1. Insert the black test lead into the negative COM terminal and the red test lead into the positive V terminal.
2. Set the function switch to the VAC or VDC position.
3. Use the MODE button to select AC or DC Voltage (EX830 only).
4. Connect the test leads in parallel to the circuit under test.
5. Read the voltage measurement on the LCD display.
EX800 Series Version 3.2 8/07 7
Find Quality Products Online at: sales@GlobalTestSupply.com
Resistance Measurements
Note: Remove power before making resistance measurements
1. Insert the black test lead into the negative COM terminal and the red test lead into the positive terminal.
2. Set the function switch to the position.
3. Touch the test probe tips across the circuit or component under test.
4. Read the resistance on the LCD display.
Capacitance Measurements
WARNING: To avoid electric shock, discharge the capacitor under test
1. Set the function switch to the capacitance position.
2. Insert the black test lead banana plug into the negative COM jack
3. Press MODE to zero any stray capacitance.
4. Touch the test probe tips across the part under test.
5. Read the capacitance value in the display.
6. The display will indicate the proper decimal point and value.
before measuring. If “dISC” appears in the display, remove and discharge the capacitor.
and the red test lead banana plug into the CAP positive jack.
Note: For very large values of capacitance measurement time can be
several minutes before the final reading stabilizes.
Frequency Measurements
1. Set the function switch to the V Hz Position.
2. Press and hold the MODE button to select the Frequency (Hz) function. “k Hz” will appear in the display.
3. Insert the black test lead banana plug into the negative COM jack and the red test lead banana plug into the Hz positive jack.
4. Touch the test probe tips across the part under test.
5. Read the Frequency value on the display.
6. The display will indicate the proper decimal point and value.
7. Press and hold the MODE button again to return to the voltage mode
EX800 Series Version 3.2 8/07 8
Find Quality Products Online at: sales@GlobalTestSupply.com
Type K Temperature Measurements
(Models EX820 and EX830 only)
1. Set the function switch to the K Temp position.
2. Insert the Temperature Probe into the negative COM and the positive TEMP jacks, observing polarity.
3. Touch the Temperature Probe tip to the device under test. Continue to touch the part under test with the probe until the reading stabilizes.
4. Read the temperature on the display. The digital reading will indicate the proper decimal point and value.
WARNING: To avoid electric shock, be sure the thermocouple probe has been removed before changing to another measurement function.
Note: An open input or a temperature overrange the meter will display
OL” and beep.
Note: See the “Temperature Units” paragraph to select °F or °C
Continuity Measurements
1. Insert the black test lead into the negative COM terminal and the red test lead into the
positive terminal.
2. Set the function switch to the
3. Use the MODE button to select continuity ))). The display icons will change when the MODE button is pressed.
4. Touch the test probe tips across the circuit or component under test.
5. If the resistance is < 40 , a tone will sound.
))) position.
Diode Test
1. Insert the black test lead banana plug into the negative COM jack and the red test lead banana plug into the positive jack
2. Turn the function switch to position. Use the MODE button to select the diode function if necessary (diode symbol will appear on the LCD when in Diode test mode)
3. Touch the test probe tips to the diode or semiconductor junction under test. Note the meter reading
4. Reverse the test lead polarity by reversing the red and black leads. Note this reading
5. The diode or junction can be evaluated as follows:
If one reading displays a value (typically 0.400V to 0.900V) and the other
reading displays OL, the diode is good.
If both readings display OL the device is open.
If both readings are very small or ‘0’, the device is shorted.
EX800 Series Version 3.2 8/07 9
Find Quality Products Online at: sales@GlobalTestSupply.com
Non-Contact InfraRed Temperature Measurements
1. Set the function switch to the IR Temp
position.
2. Aim the infrared sensor (rear of the meter) at the surface to be measured.
3. Press the button in the center of the rotary function switch to turn on the laser pointer and identify the surface spot to be measured.
4. The area of the surface to be measured must be larger than the spot size as determined by the distance to spot size specification.
5. Read the temperature in the display.
o
Note: See the “Temperature Units” paragraph to select
WARNING: Do not directly view or direct the laser pointer at an eye. Low power visible lasers do not normally present a
hazard, but may present some potential for hazard if viewed directly for extended periods of time.
IR Spot to Distance Diagram
The 8:1 spot to distance ratio determines the size of the measured surface area with respect to the distance the meter is held away from the surface.
F or oC
IR Measurement Notes
1. The object under test should be larger than the spot (target) size calculated by the field of view diagram.
2. If the surface of the object under test is covered with frost, oil, grime, etc., clean before taking measurements.
3. If an object's surface is highly reflective, apply masking tape or flat black paint to the surface before measuring.
4. The meter may not make accurate measurements through transparent surfaces such as glass.
5. Steam, dust, smoke, etc. can obscure measurements.
6. To find a hot spot, aim the meter outside the area of interest then scan across (in an up and down motion) until the hot spot is located.
EX800 Series Version 3.2 8/07 10
Find Quality Products Online at: sales@GlobalTestSupply.com
Data Hold
To freeze the LCD reading, press the HOLD button. While data hold is active, the HOLD icon appears on the LCD. Press the HOLD button again to return to normal operation.
Peak Hold
The Peak Hold function captures the peak AC or DC voltage or current. The meter can capture negative or positive peaks as fast as 1 millisecond in duration.
1. Turn the function switch to the A or V position.
2. Use the MODE button to select AC or DC (ADC EX830 only).
3. Allow time for the display to stabilize.
4. Press and Hold the PEAK button until “CAL” appears in the display. This procedure will zero the range selected.
5. Press the PEAK button, Pmax will display.
6. The display will update each time a higher positive peak occurs.
7. Press the PEAK button again, Pmin will display. The display will now update and indicate the lowest negative peak.
8. To return to normal operation, press and hold the PEAK button until the Pmin or
Pmax indicator switches off.
Note: If the Function switch position is changed after a calibration the Peak Hold
calibration must be repeated for the new function selected.
MAX/MIN
1. Press the MAX/MIN key to activate the MAX/MIN recording mode. The display icon "MAX" will appear. The meter will display and hold the maximum reading and will update only when a new “max” occurs.
2. Press the MAX/MIN key and “MIN” will appear The display icon "MIN" will appear. The meter will display and hold the minimum reading and will update only when a new “min” occurs
3. Press the MAX/MIN key and a blinking “MAX MIN” will appear. The meter will display the present reading, but will continue to update and store the max and min readings.
4. To exit MAX/MIN mode press and hold the MAX/MIN key for 2 seconds.
Temperature Units (˚F / ˚C)
The temperature units selection switch is located in the battery compartment. To change the units, remove the battery door, lift out the battery and set the switch for the desired units.
LCD Backlight Button
The LCD is equipped with backlighting for easier viewing, especially in dimly lit areas. Press the backlight button to turn the backlight on. Press again to turn the backlight off.
Automatic Power OFF
In order to conserve battery life, the meter will automatically turn off after approximately 25 minutes. To turn the meter on again, turn the function switch to the OFF position and then to the desired function position.
EX800 Series Version 3.2 8/07 11
Find Quality Products Online at: sales@GlobalTestSupply.com
Maintenance
WARNING: To avoid electrical shock, disconnect the meter from any circuit, remove the test leads from the input terminals, and turn OFF the meter before opening the case. Do not operate the meter with an open case.
Cleaning and Storage
Periodically wipe the case with a damp cloth and mild detergent; do not use abrasives or solvents. If the meter is not to be used for 60 days or more, remove the battery and store it separately.
Battery Replacement
1. Remove the Phillips head screw that secures the rear battery door
2. Open the battery compartment
3. Replace the 9V battery
4. Secure the battery compartment
Temperature Probe Replacement
The replacement bead wire probe (with banana plug connectors) is Part Number TP873.
Note: To use a Type K thermocouple probe that is terminated by a subminiature (flat blade) connector, a subminiature-to-banana plug adaptor (Part Number TP879) is required.
EX800 Series Version 3.2 8/07 12
Find Quality Products Online at: sales@GlobalTestSupply.com
Warranty
Calibration and Repair Services
Extech offers repair and calibration services for the products we sell. Extech also provides NIST certification for most products. Call the Customer Service Department for information on calibration services available for this product. Extech recommends that annual calibrations be performed to verify meter performance and accuracy.
Support line (781) 890-7440
Technical support: Extension 200; E-mail: support@extech.com
Repair & Returns: Extension 210; E-mail: repair@extech.com
Product specifications subject to change without notice
For the latest version of this Users Guide, Software updates, and other
up-to-the-minute product information, visit our website: www.extech.com
All rights reserved including the right of reproduction in whole or in part in any form.
Copyright © 2004 Extech Instruments Corporation
EX800 Series Version 3.2 8/07 13
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
FAX 781.665.0780 - TestEquipmentDepot.com
Find Quality Products Online at: sales@GlobalTestSupply.com
Loading...