Clamp jaw opening 1.7" (43mm) approx.
Display 3-3/4 digits (4000 counts) backlit LCD
Continuity check Threshold 40 ; Test current < 0.5mA
Diode test Test current of 0.3mA typical;
Low Battery indication Battery symbol is displayed
Over-range indication ‘OL’ display
Measurement rate 2 readings per second, nominal
PEAK Captures peaks >1ms
Thermocouple sensor Type K thermocouple required
IR Spectral response 6 to 16µm
IR Emissivity 0.95 fixed
IR distance ratio 8:1
Input Impedance 10M (VDC and VAC)
AC bandwidth 50 to 400Hz (AAC and VAC)
AC response True rms (AAC and VAC) on EX820 & EX830
Crest Factor 3.0 in 40A and 400A ranges, 1.4 in 1000A range (50/60Hz
Operating Temperature 41 F to 104 F (5 C to 40 C)
Storage Temperature -4 F to 140 F (-20 C to 60 C)
Operating Humidity Max 80% up to 87 F (31 C) decreasing linearly to 50% at
Storage Humidity <80%
Operating Altitude 7000ft. (2000meters) maximum.
Battery One (1) 9V Battery (NEDA 1604)
Auto power OFF After approx. 25 minutes
Dimensions & Weight 10.6x4.3x2” (270x110x50mm); 13.6 oz. (386g)
Safety For indoor use and in accordance with the requirements for
Patent notice U.S. Patent 7163336
Range and
Resolution
-4 to 1400oF NA ± (3%rdg + 9oF) ± (3%rdg + 9oF)
-20 to 760
-58 to -4 F 9 F 9 F 9 F
-4 to 518 F
-50 to -20 C
-20 to 270 C
o
C NA ± (3%rdg + 5oC) ± (3%rdg + 5oC)
2.0% reading or
whichever is >
2.0% reading or
whichever is >
Open circuit voltage < 3VDC typical
and 5% to 100% of range)
104 F(40 C)
double insulation to IEC1010-1 (2001): EN61010-1 (2001)
Overvoltage Category III 600V and Category II 1000V,
Pollution Degree 2.
NOTES: Read and understand all Warning and Caution statements in this operation
manual prior to using this meter. Set the function select switch to the OFF position when
the meter is not in use.
AC/DC Current Measurements (DC Current on Model EX830 only)
WARNING: Ensure that the test leads are disconnected from the meter before making
current clamp measurements.
1. Set the Function switch to the A
or ADC range
2. Press the trigger to open jaw.
Fully enclose only one conductor.
For optimum results, center the
conductor in the jaw.
3. The clamp meter LCD will display
the reading.
DCA ZERO (EX830 only)
The DC Zero feature removes offset
values and improves accuracy for DC
current measurements. To perform a
zero, select ADC and with no
conductor in the jaw:
1. Press the DC ZERO button to
zero the display. “ZERO” will appear in the display. The offset value is now stored and
removed from all measurements.
2. To view the stored value, press the DC ZERO button. “ZERO” will flash and the
stored value will be displayed.
3. To exit this mode, press and Hold the ZERO button until “ZERO” is no longer in the
display.
AC
AC/DC Voltage Measurements
1. Insert the black test lead into the negative COM terminal and the
red test lead into the positive V terminal.
2. Set the function switch to the VAC or VDC position.
3. Use the MODE button to select AC or DC Voltage (EX830 only).
4. Connect the test leads in parallel to the circuit under test.
5. Read the voltage measurement on the LCD display.
1. Set the function switch to the K Temp position.
2. Insert the Temperature Probe into the negative COM and the
positive TEMP jacks, observing polarity.
3. Touch the Temperature Probe tip to the device under test.
Continue to touch the part under test with the probe until the
reading stabilizes.
4. Read the temperature on the display. The digital reading will
indicate the proper decimal point and value.
WARNING: To avoid electric shock, be sure the thermocouple probe
has been removed before changing to another measurement function.
Note: An open input or a temperature overrange the meter will display
“OL” and beep.
Note: See the “Temperature Units” paragraph to select °F or °C
Continuity Measurements
1. Insert the black test lead into the negative COM terminal and the red test lead into the
positive terminal.
2. Set the function switch to the
3. Use the MODE button to select continuity ))). The display icons will change when the
MODE button is pressed.
4. Touch the test probe tips across the circuit or component under test.
5. If the resistance is < 40 , a tone will sound.
))) position.
Diode Test
1. Insert the black test lead banana plug into the negative COM jack and the red test
lead banana plug into the positive jack
2. Turn the function switch to position. Use the MODE button to select the diode
function if necessary (diode symbol will appear on the LCD when in Diode test mode)
3. Touch the test probe tips to the diode or semiconductor junction under test. Note the
meter reading
4. Reverse the test lead polarity by reversing the red and black leads. Note this reading
5. The diode or junction can be evaluated as follows:
If one reading displays a value (typically 0.400V to 0.900V) and the other
reading displays OL, the diode is good.
If both readings display OL the device is open.
If both readings are very small or ‘0’, the device is shorted.
To freeze the LCD reading, press the HOLD button. While data hold is active, the HOLD
icon appears on the LCD. Press the HOLD button again to return to normal operation.
Peak Hold
The Peak Hold function captures the peak AC or DC voltage or current. The meter can
capture negative or positive peaks as fast as 1 millisecond in duration.
1. Turn the function switch to the A or V position.
2. Use the MODE button to select AC or DC (ADC EX830 only).
3. Allow time for the display to stabilize.
4. Press and Hold the PEAK button until “CAL” appears in the display. This procedure
will zero the range selected.
5. Press the PEAK button, Pmax will display.
6. The display will update each time a higher positive peak occurs.
7. Press the PEAK button again, Pmin will display. The display will now update and
indicate the lowest negative peak.
8. To return to normal operation, press and hold the PEAK button until the Pmin or
Pmax indicator switches off.
Note: If the Function switch position is changed after a calibration the Peak Hold
calibration must be repeated for the new function selected.
MAX/MIN
1. Press the MAX/MIN key to activate the MAX/MIN recording mode. The display icon
"MAX" will appear. The meter will display and hold the maximum reading and will
update only when a new “max” occurs.
2. Press the MAX/MIN key and “MIN” will appear The display icon "MIN" will appear.
The meter will display and hold the minimum reading and will update only when a
new “min” occurs
3. Press the MAX/MIN key and a blinking “MAX MIN” will appear. The meter will display
the present reading, but will continue to update and store the max and min readings.
4. To exit MAX/MIN mode press and hold the MAX/MIN key for 2 seconds.
Temperature Units (˚F / ˚C)
The temperature units selection switch is located in the battery
compartment. To change the units, remove the battery door, lift
out the battery and set the switch for the desired units.
LCD Backlight Button
The LCD is equipped with backlighting for easier viewing,
especially in dimly lit areas. Press the backlight button to turn the
backlight on. Press again to turn the backlight off.
Automatic Power OFF
In order to conserve battery life, the meter will automatically turn off after approximately 25
minutes. To turn the meter on again, turn the function switch to the OFF position and then
to the desired function position.
WARNING: To avoid electrical shock, disconnect the meter from any circuit, remove the
test leads from the input terminals, and turn OFF the meter before opening the case. Do
not operate the meter with an open case.
Cleaning and Storage
Periodically wipe the case with a damp cloth and mild detergent; do not use abrasives or
solvents. If the meter is not to be used for 60 days or more, remove the battery and store it
separately.
Battery Replacement
1. Remove the Phillips head screw that secures the rear battery door
2. Open the battery compartment
3. Replace the 9V battery
4. Secure the battery compartment
Temperature Probe Replacement
The replacement bead wire probe (with banana plug connectors) is Part Number TP873.
Note: To use a Type K thermocouple probe that is terminated by a subminiature (flat
blade) connector, a subminiature-to-banana plug adaptor (Part Number TP879) is
required.
Extech offers repair and calibration services for the products we sell. Extech also
provides NIST certification for most products. Call the Customer Service Department for
information on calibration services available for this product. Extech recommends that
annual calibrations be performed to verify meter performance and accuracy.