Congratulations on your purchase of the Extech EX810 Clamp Meter. This meter measures AC/DC
Voltage, AC Current, Resistance, Capacitance, Frequency, Diode Test, Continuity, plus NonContact IR Temperature. This device is shipped fully tested and calibrated and, with proper use, will
provide years of reliable service.
Safety
International Safety Symbols
This symbol, adjacent to another symbol or terminal, indicates the user must refer to the
manual for further information.
This symbol, adjacent to a terminal, indicates that, under normal use, hazardous voltages
may be present
Double insulation
SAFETY NOTES
• Do not exceed the maximum allowable input range of any function.
• Do not apply voltage to meter when resistance function is selected.
• Set the function switch OFF when the meter is not in use.
• Remove the battery if meter is to be stored for longer than 60 days.
WARNINGS
• Set function switch to the appropriate position before measuring.
• When measuring volts do not switch to current/resistance modes.
• Do not measure current on a circuit whose voltage exceeds 600V.
• When changing ranges always disconnect the test leads from the circuit under test.
Function Maximum Input
A AC 1000A DC/AC
V DC, V AC 600V DC/AC
Resistance, Capacitance, Frequency, Diode Test 250V DC/AC
• Improper use of this meter can cause damage, shock, injury or death. Read and understand
this user manual before operating the meter.
• For all measurements, always use approved measurement category (CAT) voltage and current
rated test leads, test probes, and adapters. Do not exceed the CAT rating of the lowest rated
test lead, probe, adaptor, or other product accessory.
• Always remove the test leads before replacing the battery or fuses.
• Inspect the condition of the test leads and the meter itself for any damage before operating the
meter. Repair or replace any damage before use.
• Use great care when making measurements if the voltages are greater than 25VAC rms or
35VDC. These voltages are considered a shock hazard.
• Always discharge capacitors and remove power from the device under test before performing
Diode, Resistance or Continuity tests.
• Voltage checks on electrical outlets can be difficult and misleading because of the uncertainty
of connection to the recessed electrical contacts. Other means should be used to ensure that
the terminals are not "live".
• If the equipment is used in a manner not specified by the manufacturer, the protection provided
by the equipment may be impaired.
• This device is not a toy and must not reach children’s hands. It contains hazardous objects as
well as small parts that the children could swallow. In case a child swallows any of them,
please contact a physician immediately
• Do not leave batteries and packing material lying around unattended; they can be dangerous
for children if they use them as toys
• In case the device is going to be unused for an extended period of time, remove the batteries
to prevent them from draining
• Expired or damaged batteries can cause cauterization on contact with the skin. Always,
therefore, use suitable hand gloves in such cases
• See that the batteries are not short-circuited. Do not throw batteries into the fire.
• Do not directly view or direct the laser pointer at an eye. Low power visible lasers do not
normally present a hazard, but may present some potential for hazard if viewed directly for
extended periods of time
NOTES: Read and understand all Warning and Caution statements in this operation manual prior
to using this meter. Set the function select switch to the OFF position when the meter is not in use.
AC Current Measurements
WARNING: Ensure that the test leads are
disconnected from the meter before making current
clamp measurements.
1. Set the Function switch to the A
2. Press the trigger to open the jaw. Fully
enclose only one conductor. For optimum
results, center the conductor in the jaw.
3. The clamp meter LCD will display the reading.
AC/DC Voltage Measurements
1. Insert the black test lead into the negative COM terminal and the red test lead into the positive
V terminal.
2. Set the function switch to the V
3. Connect the test leads in parallel to the circuit under test.
4. Read the voltage measurement on the LCD display.
To freeze the LCD reading, press the HOLD button. While data hold is active, the HOLD icon
appears on the LCD. Press the HOLD button again to return to normal operation.
Peak Hold
The Peak Hold function captures the peak AC or DC voltage or current. The meter can capture
negative or positive peaks as fast as 1 millisecond in duration.
1. Turn the function switch to the A or V position.
2. Allow time for the display to stabilize.
3. Press and Hold the PEAK button until “CAL” appears in the display. This procedure will zero
the range selected.
4. Press the PEAK button, Pmax will display.
5. The display will update each time a higher positive peak occurs.
6. Press the PEAK button again, Pmin will display. The display will now update and indicate the
lowest negative peak.
7. To return to normal operation, press and hold the PEAK button until the Pmin or Pmax
indicator switches off.
Note: If the Function switch position is changed after a calibration the Peak Hold calibration must
be repeated for the new function selected.
MAX/MIN
1. Press the MAX/MIN key to activate the MAX/MIN recording mode. The display icon "MAX" will
appear. The meter will display and hold the maximum reading and will update only when a new
“max” occurs.
2. Press the MAX/MIN key and “MIN” will appear The display icon "MIN" will appear. The meter
will display and hold the minimum reading and will update only when a new “min” occurs
3. Press the MAX/MIN key and a blinking “MAXMIN” will appear. The meter will display the
present reading, but will continue to update and store the max and min readings.
4. To exit MAX/MIN mode press and hold the MAX/MIN key for 2 seconds.
Temperature Units (˚F / ˚C)
The temperature units selection switch is located in the battery
compartment. To change the units, remove the battery door, lift out the
battery and set the switch for the desired units.
LCD Backlight Button
The LCD is equipped with backlighting for easier viewing, especially in
dimly lit areas. Press the backlight button to turn the backlight on. Press
again to turn the backlight off.
Automatic Power OFF
In order to conserve battery life, the meter will automatically turn off after approximately 25 minutes.
To turn the meter on again, turn the function switch to the OFF position and then to the desired
function position.
WARNING: To avoid electrical shock, disconnect the meter from any circuit, remove the test leads
from the input terminals, and turn OFF the meter before opening the case. Do not operate the meter
with an open case.
Cleaning and Storage
Periodically wipe the case with a damp cloth and mild detergent; do not use abrasives or
solvents. If the meter is not to be used for 60 days or more, remove the battery and store it
separately.
Battery Replacement
1. Remove the Phillips head screw that secures the rear battery door
2. Open the battery compartment
3. Replace the 9V battery
4. Secure the battery compartment
Never dispose of used batteries or rechargeable batteries in household waste.
As consumers, users are legally required to take used batteries to appropriate
collection sites, the retail store where the batteries were purchased, or wherever
batteries are sold.
Disposal: Do not dispose of this instrument in household waste. The user is obligated to take end-oflife devices to a designated collection point for the disposal of electrical and electronic equipment.
Other Battery Safety Reminders
o Never dispose of batteries in a fire. Batteries may explode or leak.
o Never mix battery types. Always install new batteries of the same type.
Function Range and Resolution Accuracy (% of reading + digits)
Temp
(IR)
Clamp jaw opening 43mm (1.7") approx.
Display 3-3/4 digits (4000 counts) backlit LCD
Continuity check Threshold 40Ω; Test current < 0.5mA
Diode test Test current of 0.3mA typical;
Open circuit voltage < 3VDC typical
Low Battery indication Battery symbol is displayed
Over-range indication ‘OL’ display
Measurement rate 2 readings per second, nominal
PEAK Captures peaks >1ms
IR Spectral response 6 to 16µm
IR Emissivity 0.95 fixed
IR distance ratio 8:1
Input Impedance 10MΩ (VDC and VAC)
AC bandwidth 50 to 400Hz (AAC and VAC)
Crest Factor 3.0 in 40A and 400A ranges, 1.4 in 1000A range (50/60Hz and 5% to
Operating Temperature 5°C to 40°C (41°F to 104°F)
Storage Temperature -20°C to 60°C (-4°F to 140°F)
Operating Humidity Max 80% up to 31°C (87°F) decreasing linearly to 50% at 40°C
Storage Humidity <80%
Operating Altitude 2000meters (7000ft) maximum.
Battery One (1) 9V Battery (NEDA 1604)
Auto power OFF After approx. 25 minutes
Dimensions & Weight 270x110x50mm (10.6x4.3x2”) 386g (13.6 oz)
Safety For indoor use and in accordance with the requirements for double
Patent notice U.S. Patent 7163336
-58 to -4°F ± 9 °F
-4 to 518°F
-50 to -20°C ±5°C
-20 to 270°C
100% of range)
(104°F)
insulation to IEC1010-1 (2001): EN61010-1 (2001) Overvoltage
Category III 600V, Pollution Degree 2.