Programming Menu
The meter can be configured and calibrated through simple button presses in the
programming menu. Press the MENU button to access the menu and refer to the menu
‘tree’ below. The menu uses UP/DOWN, SELECT, BACK, & ESC presses for navigation
and selection. In the table below the factory default settings are in bold with an asterisk.
Each parameter is explained in detail in the subsequent sections.
Top level Sub level 1 Sub level 2 Notes
AVG* Average of a series of readings
MIN Lowest of a series of readings
MAX Highest of a series of readings
NO Number of sampled readings
STATISTICAL VIEW
SDEV Standard deviation of a series
Single* One reading at a time
Measure mode
Continuous Continuous measurements
Direct* Readings are not stored in groups
Working mode
Group 1…4 Store readings in groups
Auto* Meter automatically selects mode
Fe Ferrous measurement mode
Probe used
No Fe Non-ferrous measurement mode
μm*
Micrometers
mils Mils = mm * 2.54 / 100
Unit settings
mm Millimeters
ON* Enables backlight operation
Backlight
OFF Disables backlight operation
AVG* Average of a series of readings
MAX Lowest of a series of readings
MIN Highest of a series of readings
LCD Statistics
SDEV Standard deviation of a series
Enable* Allows auto power off to operate
OPTIONS
Auto Power OFF
Disable Defeats the auto power off feature
High Limit High Alarm alerts user when reached
Limit settings
Low Limit Low Alarm alerts user when reached
LIMIT
Delete Limits Clear the alarm limit values
Current Data Delete current data
All Data Delete all stored data
DELETE
Group Data
Delete data stored data plus alarm
and calibration data
MEASUREMENT
VIEW
View stored data in all groups
Enable Allow calibration access
Disable Lock out the calibration mode
Delete Zero N Clear the zero cal. data (non-ferrous)
CALIBRATION
Delete Zero F Clear the zero cal. data (ferrous)
NOTE: Disable the Auto Power OFF feature before lengthy programming to avoid
inconvenient automatic power down while programming.