Extech Instruments 380950-NIST User manual

Page 1
Model 380950 80A Mini AC/DC Clamp Meter
User's Guide
www.GlobalTestSupply.com
www. .com
information@itm.com1.800.561.8187
Page 2
Introduction
Congratulations on your purchase of the Extech 80A Mini AC/DC Clamp Meter. The Model 380950 measures AC/DC Current, AC/DC Voltage, Resistance, Frequency, Capacitance, Duty Cycle, Diode Test, and Continuity. This clamp meter is shipped fully tested and calibrated and, with proper use, will provide years of reliable service.
Safety
International Safety Symbols
This symbol, adjacent to another symbol or terminal, indicates the user must refer to the manual for further information.
This symbol, adjacent to a terminal, indicates that, under normal use, hazardous voltages may be present
Double insulation
SAFETY NOTES
Do not exceed the maximum allowable input range of any function.
Do not apply voltage to meter when resistance function is selected.
Set the function switch OFF when the meter is not in use.
Remove the battery if meter is to be stored for longer than 60 days.
WARNINGS
Set function switch to the appropriate position before measuring.
When measuring volts do not switch to current/resistance modes.
Do not measure current on a circuit whose voltage exceeds 240V.
When changing ranges always disconnect the test leads from the circuit under test.
CAUTIONS
Improper use of this meter can cause damage, shock, injury or death. Read and understand
this user manual before operating the meter.
Always remove the test leads before replacing the battery or fuses.
Inspect the condition of the test leads and the meter itself for any damage before operating the
meter. Repair or replace any damage before use.
Use great care when making measurements if the voltages are greater than 25VAC rms or
35VDC. These voltages are considered a shock hazard.
Always discharge capacitors and remove power from the device under test before performing
Diode, Resistance or Continuity tests.
Voltage checks on electrical outlets can be difficult and misleading because of the uncertainty
of connection to the recessed electrical contacts. Other means should be used to ensure that the terminals are not "live".
If the equipment is used in a manner not specified by the manufacturer, the protection provided
by the equipment may be impaired.
2
www.GlobalTestSupply.com
www. .com
information@itm.com1.800.561.8187
380950-en-GB_V1.2 8/15
Page 3
Meter Description
1. Conductor jaws
2. Jaw opening trigger
3. Function select switch
4. LCD Display
5. ZERO button
6. Data Hold and Backlight Button
7. Mode select button
8. Range select button
9. Hz/%/Duty Cycle button
10. COM input jack
11. V/Hz jack
12. Battery cover (rear)
AC AC (alternating current)
DC DC (direct currrent)
Minus sign
AUTO AutoRange mode ZERO ZERO mode
Audible Continuity
•))) HOLD Data Hold mode
Diode test mode
m milli V Volts A Amps K kilo M Mega
Ohms
Low Battery icon
3
www.GlobalTestSupply.com
www. .com
information@itm.com1.800.561.8187
380950-en-GB_V1.2 8/15
Page 4
Operation
Notice: Read and understand all WARNING and CAUTION statements listed in the safety section of this operation manual prior to using this meter. Set the function select switch to the OFF position when the meter is not in use.
DC/AC Current Measurements
Warning: Disconnect the test leads from the meter before making current
clamp measurements.
1. Set the Function switch to the 80ADC, 4ADC, 80AAC or 4AAC range. If the range of the measured is not known, select the higher range first then move to the lower range if necessary.
2. For DC current measurement, press the ZERO key to null the meter display.
3. Press the trigger to open jaw. Fully enclose one conductor to be measured.
4. The clamp meter LCD will display the reading.
DC/AC Voltage Measurements
1. Set the rotary function switch to the V position.
2. Insert the black test lead banana plug into the negative (COM) jack Insert the red test lead banana plug into the positive (V/Hz) jack
3. Select AC or DC with the MODE button
4. Connect the test leads to the circuit under test
5. Read the voltage on the display. The display will indicate the proper decimal point and value.
Resistance Measurements
1. Set the function switch to the
2. Insert the black test lead banana plug into the negative (COM) jack Insert the red test lead banana plug into the positive (V Hz jack.
3. Touch the test probe tips across the circuit or part under test. It is best to disconnect one side of the part under test so the rest of the circuit will not interfere with the resistance reading.
4. Read the resistance on the display. The display will indicate the proper decimal point and value.
Continuity Check
1. Set the function switch to the
2. Push the mode button to indicate •))) on the display.
3. Insert the black lead banana plug into the negative (COM) jack Insert the red test lead banana plug into the positive (VHz) jack.
4. Touch the test probe tips to the circuit or wire you wish to check.
5. If the resistance is less than approximately 150, the audible signal will sound. If the circuit is open, the display will indicate “OL.”.
•))) CAP position.
•))) CAP position.
4
www.GlobalTestSupply.com
www. .com
information@itm.com1.800.561.8187
380950-en-GB_V1.2 8/15
Page 5
Diode Test
1. Turn the rotary switch to the
2. Insert the black test lead banana plug into the negative (COM) jack Insert the red test lead banana plug into the positive (VHz jack.
3. Push the mode button to indicate on the display.
4. Touch the test probes to the diode under test. Typically for a normal diode, forward voltage will indicate 0.4V to 0.7V. Reverse voltage will indicate “OL”. Shorted devices will indicate near 0V and an open device will indicate “OL” in both polarities.
Capacitance Measurements
Warning: To avoid electrical shock, disconnect power to the unit under test and discharge all
capacitors before taking any capacitance measurements. Remove the batteries and unplug the line cords.
1. Set the function switch to the •))) CAP position.
2. Push the mode button to indicate nF on the display.
3. Insert the black lead banana plug into the negative (COM) jack Insert the red test lead banana plug into the positive (VHz) jack.
4. Press the ZERO key to null the meter display.
5. Touch the test probe tips to the capacitor you wish to check.
6. Read the capacitance value on the display.
Frequency or % Duty Cycle Measurements
1. Turn the rotary switch to the Hz % position.
2. Insert the black test lead banana plug into the negative (COM) jack and the red test lead banana plug into the positive (VHz jack.
3. Select Hz or % with the HZ/% button.
4. Touch the test probe tips to the circuit under test.
5. Read the frequency on the display.
•))) CAP position.
Analog Signal Output
1. Turn the rotary switch to the DCA or ACA ranges.
2. Insert the black test lead banana plug into the negative (COM) jack and the red test lead banana plug into the positive (VHz jack.
3. Connect the test leads to a multimeter, oscilloscope or chart recorder inputs.
4. Press the trigger to open the jaw. Fully enclose the conductor to be measured.
5. The analog voltage signal is output to the measuring device.
Note: When measuring DCA, the output signal is DCV. When measuring ACA, the output signal is
both ACV and DCV.
5
www.GlobalTestSupply.com
www. .com
information@itm.com1.800.561.8187
380950-en-GB_V1.2 8/15
Page 6
Auto/Manual Ranging
The meter turns on in Autoranging mode. Press the RANGE button to enter manual ranging. Each press of the range button will step to the next range as indicated by the units and decimal point location. Press and hold the RANGE button for two seconds to return to Autoranging mode.
Note: Manual ranging does not function in AC Current or Diode and Continuity check functions. In Temperature function, it will change the resolution from 0.1° to 1°.
Data Hold
To freeze the LCD meter reading, press the HOLD button. While data hold is active, the HOLD display icon appears on the LCD. Press the HOLD button again to return to normal operation.
Backlight
Press and hold the HOLD button for >2 seconds to turn the backlight on/off. Note: The HOLD feature will activate when the backlight is turned on. Press the HOLD button again
to exit the Hold feature.
Zero Button
Zeros Capacitance and DC Current measurements. Also allows the user to offset the meter by using the displayed value as the zero reference value. Press the ZERO key momentarily to activate and to exit Zero mode.
6
www.GlobalTestSupply.com
www. .com
information@itm.com1.800.561.8187
380950-en-GB_V1.2 8/15
Page 7
Specifications
Function Range & Resolution Accuracy (of reading)
DC Current 4.000 ADC ± (2.8% + 10 digits)
AC Current (50/60Hz)
DC Voltage
AC Voltage (50/60Hz)
Resistance
Capacitance 40.00nF ± (5% + 30 digits)
Frequency 5.000Hz ± (1.5% + 5 digits)
Duty Cycle 0.5% to 99.0% ± (1.2% + 2 digits)
Analog Output (ACA & DCA ranges)
80.0 ADC ± (3.0% + 8 digits)
4.000 AAC ± (3.0% + 10 digits)
80.0 AAC ± (3.0% + 8 digits)
400.0mV ± (1.0% + 15 digits)
4.000V ± (1.0% + 3 digits)
40.00V
400.0V 600V ± (2.0% + 3 digits)
400.0mV ± (1.0% + 30 digits)
4.000V
40.00V
400.0V 600V
400.0
4.000k
40.00k
400.0k
4.000 M
40.00M
400.0nF ± (3% + 5 digits)
4.000µF ± (3.5% + 5 digits)
40.00µF
100.0µF ± (5% + 5 digits)
50.00Hz
500.0Hz
5.000KHz
50.00KHz
500.0KHz
5.000MHz
10.00MHz
Pulse Width: 100µs-100ms, Frequency: 5Hz to 150KHz 10mV/Amp (4 Amp range), 1mV/Amp (80 Amp range) 10mV/Amp ; Accuracy: ± (5%rdg + 2mV); Output impedance: approx 3k
± (1.5% + 3 digits)
± (2.0% + 5 digits)
± (1.0% + 4 digits)
± (1.5% + 2 digits)
± (2.5% + 3 digits) ± (3.5% + 5 digits)
± (1.2% + 2 digits) Sensitivity: 10Vrms min.
7
www.GlobalTestSupply.com
www. .com
information@itm.com1.800.561.8187
380950-en-GB_V1.2 8/15
Page 8
Jaw size 12.7mm (0.5") approx.
Display 4000 count LCD
Continuity Audible tone < 150 approx.
Diode Test Open circuit voltage < 1.5VDC; Test current <1mA (typical)
AC V bandwidth 50Hz to 400Hz
AC A bandwidth 50/60Hz
Low battery indication “ ” is displayed
Overrange indication “OL” is displayed
Auto Power OFF After 25 minutes
Measurement rate 2 per second, nominal
Input Impedance 7.8M (V DC and V AC)
Operating Temperature -10ºC to 50ºC (14ºF to 122ºF)
Storage Temperature -30
Operating Humidity Max 80% up to 31ºC (87ºF) decreasing linearly to 50% at 45ºC (113ºF)
Storage Humidity <80%
Operating Altitude 2000meters (6560ft) operating
Batteries (2) 1.5V AAA batteries
Weight 200g (0.44lb)
Size 200x50x35mm (7.87” x 1.97” x 1.38”)
Safety For indoor use and in accordance with the requirements for double
PER IEC1010 OVERVOLTAGE INSTALLATION CATEGORIES
OVERVOLTAGE CATEGORY I
Equipment of OVERVOLTAGE CATEGORY I is equipment for connection to circuits in which measures are taken to limit the transient overvoltages to an appropriate low level. Note – Examples include protected electronic circuits.
OVERVOLTAGE CATEGORY II
Equipment of OVERVOLTAGE CATEGORY II is energy-consuming equipment to be supplied from the fixed installation. Note – Examples include household, office, and laboratory appliances.
OVERVOLTAGE CATEGORY III
Equipment of OVERVOLTAGE CATEGORY III is equipment in fixed installations. Note – Examples include switches in the fixed installation and some equipment for industrial use with permanent connection to the fixed installation.
OVERVOLTAGE CATEGORY IV
Equipment of OVERVOLTAGE CATEGORY IV is for use at the origin of the installation. Note – Examples include electricity meters and primary over-current protection equipment
o
C to 60oC (-22oF to 140oF)
insulation to IEC1010-1 (1995): EN61010-1 (1995) Overvoltage Category III, Pollution Degree 2.
8
www.GlobalTestSupply.com
www. .com
information@itm.com1.800.561.8187
380950-en-GB_V1.2 8/15
Page 9
Maintenance
WARNING: To avoid electrical shock, disconnect the meter from any circuit, remove the test leads from the input terminals and turn OFF the meter before opening the case. Do not operate with open case.
Cleaning and Storage
Periodically wipe the case with a damp cloth and mild detergent; do not use abrasives or solvents. If the meter is not to be used for periods of longer than 60 days, remove the batteries and store them separately
Battery Replacement
1. Remove the two rear battery cover Phillips head screws
2. Open the battery compartment
3. Replace the two 1.5V AAA batteries.
4. Re-assemble the meter
5.
You, as the end user, are legally bound (EU Battery ordinance) to return all used batteries, disposal in the household garbage is prohibited! You can hand over your used batteries / accumulators at collection points in your community or wherever batteries / accumulators are sold!
Disposal: Follow the valid legal stipulations in respect of the disposal of the device at the end of its lifecycle