Extech Extech RHT510 Manual

Hygro-Thermometer

Mo
del RHT510

USER MANUAL

99 Washington Street Melrose, MA 02176
Phone 781-665-1400 Toll Free 1-800-517-8431
2
Introduction
Thank you for selecting the Extech Model RHT510. This hand-held device measures Relative Humidity, Air temperature, Dew point temperature, Wet bulb temperature, and Type K temperature (using external probe). This device is shipped fully tested and calibrated and, with proper use, will provide years of reliable service.
Features
Backlit LCD with Display Hold
Light weight, ergonomic design with lanyard
Displays Relative Humidity and Temperature (Air, Type-K, Dew point, and Wet bulb)
Selectable temperature units C/F
MAX-MIN Recording
Tripod Mounting
Battery operated
Safety
Please read the entire User Manual and Quick Start before operating this device. Use the meter only as specified and do not attempt to service or open the meter housing. Do not allow children to handle the meter. Please dispose of batteries and meter responsibly and in accordance with
all applicable laws and regulations.
RHT510-en-GB_V1.0 5/17
3
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com
Meter Description
1. Air Temperature/Humidity Sensor
2. Type K Temperature Probe Input
3. LCD Display
4. ON-OFF button
5. button
6. Mode button
7. Backlight button
8. SET and button
9. Record button (MAX-MIN)
10. Display Hold button
Note: Battery compartment and tripod mount on back of meter
LCD Description
1. Type K Temperature Icon
2. Function Icons
3. MAX-MIN Record Icon
4. MAX Display Icon
5. MIN Display Icon
6. Display Hold Icon
7. Battery Status Indicator
8. Auto Power OFF (APO)
9. Primary Display
10. Secondary Display
RHT510-en-GB_V1.0 5/17
4
Operation
Basic Use
1. Short press the power button to power the meter. If the LCD does not switch on, check the batteries located in the rear battery compartment.
2. Place the sensor in the testing environment.
3. View the measurement on the meter’s display.
4. To power OFF the meter, short press the
Mode Selection
Press the M button to select the mode of operation. The meter has four (4) modes: Relative Humidity %, Dew point temperature, Wet bulb temperature and Type K temperature (external probe required). Release the M button to make a selection.
Display Hold
Short press the H button to freeze or unfreeze a reading on the display. The Display Hold and APO functions are not operational in record mode.
LCD Backlight
The LCD is equipped with backlighting for easier viewing, especially in dimly lit areas. Press the
backlight button to turn the backlight on. The backlight will automatically turn off after 10
seconds.
Type K Measurement
Insert the optional Type K temperature probe into the Type K temperature probe socket (left side of meter). Use the mode (M) button to set the mode of operation to Type K temperature and view the readings on the display.
button.
MAX-MIN Recording
The meter records the maximum and minimum values over time. Note: Display Hold, Auto Power OFF setup, and the Power button do not function in Record
mode.
1. Press the R button to enter the Record mode. The display. The maximum and minimum values will be recorded and updated during the measurement period.
2. Press the R button to display the maximum value recorded over time. The maximum value will appear on the display.
3. Press the R button to display minimum value recorded over time. The icon and minimum value will appear on the display.
4. Press the H button to reset the recorded data. The previously recorded MAX-MIN readings will be cleared and the meter will start recording new MAX-MIN readings.
5. Long press the R button to exit the record mode.
record icon will appear on the
icon and
RHT510-en-GB_V1.0 5/17
5
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com
Auto Power OFF Control
In order to conserve battery life, the meter will automatically shut off after approximately 10 minutes of inactivity. The APO icon appears in the display when APO is selected.
To set APO ON or OFF as the default condition:
1. Long press SET button to enter APO mode. The bottom display will show ‘PoFF’.
2. Press the or button to turn the APO ON (YES) or APO OFF (NO).
3. Press the R button to save the selection.
4. Wait approximately 10 seconds for the meter to return to normal operation.
Temperature Units Selection (C/F)
1. Long press SET button. The bottom display will show ‘PoFF’.
2. Press the R button. The bottom display will show t-CF’.
3. Press the or button to select C or F.
4. Press the R button to save the selection.
5. Wait approximately 10 seconds for the meter to return to normal operation.
Maintenance
Battery Replacement
1. Power OFF the meter.
2. Remove the flat head screw that secures the battery compartment at the back of the meter.
3. Open the battery compartment and replace the 3 AAA batteries observing correct polarity. Re-assemble the meter before use.
Safety: Please dispose of batteries responsibly; never dispose of batteries in a fire, batteries may explode or leak. If the meter is not to be used for 60 days or more, remove the batteries and store separately.
Never dispose of used batteries or rechargeable batteries in household waste. As consumers, users are legally required to take used batteries to appropriate collection sites, the retail store where the batteries were purchased, or wherever batteries are sold.
Disposal: Do not dispose of this instrument in household waste. The user is obligated to take end-of-life devices to a designated collection point for the disposal of electrical and electronic equipment.
Cleaning and Storage
Periodically wipe the case with a damp cloth and mild detergent; do not use abrasives or solvents.
RHT510-en-GB_V1.0 5/17
6
Specifications
Electrical Specifications (23±5°C)
Humidity/Air Temperature
Unit Range Resolution Accuracy
%RH 10 to 95 0.1 <70%: ±(4%RH)
≥70%: ±(4%rdg + 1.2%RH)
Temp oC 0 to 50 0.1 ±1.2oC
Temp oF 32 to 122 0.1 ±2.5oF
Type K Thermocouple (external probe)
Unit Range Resolution Accuracy*
Type K Temp oC
Type K Temp oF
*Accuracy spec. does not include additional error introduced by the external probe
Dew Point Temperature
o
C -25.3 to 49.0 0.1 Calculation based on
o
F -13.5 to 120.0 0.1
-50.0 to 1300.0 0.1 ±(0.4%rdg + 0.5 oC)
-50.1 to -100.0 0.1 ±(0.4%rdg + 1.0 oC)
-58.0 to 2372.0 0.1 ±(0.4%rdg + 1.0
-58.1 to -148.0 0.1 ±(0.4%rdg + 1.8 oF)
humidity/temp values.
o
F)
Wet Bulb Temperature
o
C -5.4 to 49.0 0.1 Calculation based on
o
F 22.2 to 120.0 0.1
humidity/temp values.
RHT510-en-GB_V1.0 5/17
7
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com
Ge
neral Specifications
Display Backlit LCD 35 x 30mm (1.38 x 1.18”)
Measurements Humidity/Temp; Dew Point; Wet Bulb; Type K Temp.
Input Overload Display “----“
Tripod Mount On rear panel
Power 1.5V AAA batteries (3)
Power Consumption Approx. 5mA DC
Automatic Power Off After approx. 10 minutes of inactivity
Operating Temperature 0 to 50oC (32 to 122oF)
Operating Humidity 95% RH max
Storage Temperature -10 to 60oC (14 to 140oF)
Storage Humidity 10 to 75% RH
Dimensions/Weight 153 x 45 x 20mm (6.0 x 1.8 x 1.2”) /160g (5.64oz)
Copyright © 2017 FLIR Systems, Inc.
All rights reserved including the right of reproduct
ISO 9001 Certified
whole or in part in any form
ion in
RHT510-en-GB_V1.0 5/17