Extech Extech EA10 Manual

User's Guide
EasyView Dual K Thermometer
Model EA10
T1
T2
hour: min
EasyView 10
Dual K Thermometer
C
T1
T2
MX
F K
AVGMN
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
FAX 781.665.0780 - TestEquipmentDepot.com
whether written or oral, is expressed or implied.
Introduction
Congratulations on your purchase of the Extech EasyView Dual K Thermometer. This device offers dual thermocouple inputs with multifunction LCD display. Careful use of this meter will provide years of reliable service.
Specifications
Measurement Range -200 to 1999oF (-200 to 1360oC) Resolution 0.1oC/ oF <200o , 1oC/ oF >200 Accuracy (@23 ± 5
o
C)
±(0.3% reading + 2oF) @ -200oF to 1832oF ±(0.5% reading + 2oF) @ 1832oF to 1999oF ±(0.3% reading + 5oC) @ -200oC to -93oC ±(0.3% reading + 1oC) @ -93oC to 1000oC ±(0.5% reading + 1oC) @ 1000oC to 1360oC
Temperature Coefficient 0.1 times the applicable accuracy specifications per oC from 0oC
o
to 18
C and 28oC to 50oC (32oF to 64oF and 82oF to 122oF) Note: Temperature accuracy does not include the accuracy of the type K probe. Note: The temperature scale is based on the international temperature scale of 1990 (ITS90).
General Specifications
Display Dual Display Multi-function LCD with Backlight Input Protection 60VDC; 24VAC rms Display Update Rate 1 per second Over range indication "OL" appears on the LCD Open input indication "OL" appears on the LCD Low battery indication “BT” appears on the LCD Power supply 6 AAA Batteries Battery life Approximately 200 hours with carbon zinc batteries Operating Temperature 32 to 122 Operating Humidity < 80% RH Storage Temperature 14 to 140
o
F (0 to 50oC)
o
F (-10 to 60oC) Storage Humidity 10 to 80% RH Dimensions 5.91x2.8x1.4" (150x72x35mm) Weight Approx. 8.29 oz. (235g) with battery
o
Meter Description
1. LCD display
2. Data HOLD button
3. Power button
4. Max/Min button
5. OFFSET button
6. Backlite button
7. C/F/K units select button
8. Elapsed TIME button
9. T1 or T1-T2 or T2 display select button
10. T1 Thermocouple input socket
11. T2 Thermocouple input socket
12. Battery compartment (rear)
10
1
2
3
4
5
12
T1
T2
min: s ec
hour:min
11
C
T1
T2
F K
MX
AVGMN
6
7
8
9
Operation
Connecting thermocouples
1. This meter accepts two type K thermocouples with spade plugs (sub-miniature type
with one spade wider than the other).
2. Plug the thermocouple(s) into the meter's T1 and/or T2 thermocouple input jack(s).
Turn Power ON
1. Press the button to turn power on. The meter will perform a short self-test.
2. If a probe is not inserted to the meter, an “OL” indication will appear on the display.
Selecting °C, °F or °K units of measure
Press the C/F/K button to select the desired temperature unit of measure.
Selecting T1, T2, and T1-T2 Displays
Press the T1 T2 T1-T2 button to step through and select a display of:
a. T1 (thermocouple 1) displays in the upper large display and T2 (thermocouple 2)
temp. displays in the lower small display. b. T1 plus T2 c. T1-T2 plus T1 d. T1-T2 plus T2
MIN, MAX, AVG Recording Function with Timer
Press the MX/MN button to begin capturing the Minimum (MN), Maximum (MX), and Average (AVG) temperature values. The Elapsed Time clock will appear on the lower left of the display and the lower display will indicate the MAXIMUM (MX) value captured since the MX/MN button was pressed.
Press the MX/MN button to change the lower display to reflect the stored MX to MN and then AVG value of the input selected in the upper display (T1, T2 or T1-T2). Each reading will be accompanied by a Timer value indicating the time when each reading was recorded.
Press the Time button to change the elapsed time format from min:sec to hour:min. Press and Hold the MX/MN button until the display returns to standard operation to exit the
MX/MN function.
3
EA10 V1.0 4/02
Data Hold
Press the HOLD button to freeze the reading in the display. The 'H' hold icon will appear on the upper left-hand side of the display. Press the HOLD button again to return to normal operation (the 'H' hold icon will disappear).
Offset
The Offset Mode permits the user to store the temperature reading as a reference value and compare this reference value to subsequent temperature readings. In Offset mode, the meter displays the actual measured temperature minus the stored reference value.
Press the OFFSET button to store the currently displayed temperature readings and enter the Offset mode. “OFFSET” will appear in the display.
Press the OFFSET button to exit the mode.
Backlight
Press the backlight button to turn on the LCD display backlight. The backlight will automatically turn off after one minute or press the button again to turn the light off.
Auto Power OFF
The meter will automatically turn off after 30 minutes if no keys are pressed during this period. Auto power off is disabled in the max/min recording mode.
Turning Power OFF
Press the button to turn power off.
Calibration and Repair Services
Support Hotline (781) 890-7440
Tech support: Ext. 200; Email: support@extech.com Repair/Returns: Ext. 210; Email: repair@extech.com