Page 1
EXCELL PRECISION CO., LTD.
DIGITAL HIGH PRECISION
WEIGHING SCALES
USER INSTRUCTIONS
© Excell Precision Limited 2006. All rights reserved Worldwide.
The information contained herein is the property of Excell Precision Limited and is
supplied without liability for errors or omissions. No part may be reproduced or used
except as authorised by contract or other written permission. The copyright and the
foregoing restriction on reproduction and use extend to all media in which the
information may be embodied.
V1.1 ZSME300000011
Page 2
Page 3
EXCELL PRECISION CO., LTD.
TABLE OF CONTENTS
BEFORE USING THE SCALE.....................................................................................................3
PREPARING TO USE THE SCALE.............................................................................................3
INTRODUCTION..........................................................................................................................4
1. FEATURES...........................................................................................................................4
2. SPECIFICATIONS................................................................................................................5
3. POWER CONSUMPTION.....................................................................................................5
4. LOW BATTERY WARNING...................................................................................................5
LCD DISPLAY SYMBOLS...........................................................................................................6
KEYBOARD FUNCTION.............................................................................................................7
OPERATION................................................................................................................................8
1. DISPLAY BACKLIGHT..........................................................................................................8
2. ZERO....................................................................................................................................8
3. WEIGHING MODE................................................................................................................8
(1). Units Selection................................................................................................................8
(2). Totalising.........................................................................................................................8
I. Weight Totalising........................................................................................................9
II. Recall Totalised Weight Values...................................................................................9
III. Clear Totalised Weight Values....................................................................................9
(3). Deduction of the Container Weight...............................................................................10
I. The weight of the container is unknown ( Tare ).......................................................10
II. Recall the semi-auto tare value................................................................................10
III. Clear the semi-auto tare...........................................................................................10
IV. The weight of the container is known (Pre-set tare).................................................11
V. Recall the pre-set tare value.....................................................................................11
VI. Clear the pre-set tare value......................................................................................11
(4). Check Weighing............................................................................................................12
I. Pre-set the “ Low limit value” , “ High limit value” , and “ the buzzer” operation.............12
II. Pre-set the Low limit value only................................................................................13
III. High/OK/Low indications..........................................................................................13
IV. Recall the check weighing setting............................................................................13
V. Clear the check weighing settings............................................................................13
4. COUNTING FUNCTION.....................................................................................................14
(1). Sampling.......................................................................................................................14
(2). Totalising.......................................................................................................................14
(3). Check Weighing............................................................................................................14
CONFIGURATION SETTINGS..................................................................................................15
1. CHECK-WEIGHING CONFIGURATION.............................................................................16
2. RS-232 SETTING...............................................................................................................17
3. EXIT CONFIGURATION SETTINGS...................................................................................18
RS-232 AND RELAY OUTPUT..................................................................................................19
1. RS-232 OR SERIAL PRINTER OUTPUT............................................................................19
(1) RS-232 (25 Pin ‘ D’ type) Pin Description.......................................................................19
(2) RS-232 Interface Format...............................................................................................19
(3) Data Format...................................................................................................................20
V1.1 ZSME300000011
1
Page 4
EXCELL PRECISION CO., LTD.
2. RELAY OUTPUT.................................................................................................................23
(1) Relay Output..................................................................................................................23
(2) Relay Pin Description....................................................................................................23
(3) Connection Illustration...................................................................................................23
(4) Power Source Illustration...............................................................................................24
(5) Connection Setting........................................................................................................24
V1.1 ZSME300000011
2
Page 5
EXCELL PRECISION CO., LTD.
BEFORE USING THE SCALE
To enable you to use this scale correctly, we suggest that you read this manual carefully.
1. Do not use scale in areas with excessive water and don’ t spray the scale or indicator with
water when cleaning. Remove all water from the scale and indicator with a clean dry cloth.
2. The load placed on the platform must not exceed the maximum weighing capacity of the
scale.
3. Keep the scale away from high temperatures and damp conditions.
4. If the scale is not going to be used for some time, please clean it and store it in a plastic bag
in dry conditions. A desiccant sachet may be included to prevent moisture build up.
5. If the scale is not going to be used for some time, the internal rechargeable battery should
be recharged every three months.
6. Before using the scale after a long period of storage, please ensure that the internal battery
is fully charged. NOTE: Care should be taken not to leave the internal battery on charge
for too long, as this may decrease the life of battery.
PREPARING TO USE THE SCALE
1. Locate the scale on a firm level surface free from vibrations for accurate weight readings.
2. Adjust the four levelling feet to set the scale platform level using the spirit level bubble
located on the scale platform frame.
3. Avoid operating the scale in direct sunlight or drafts of any kind.
4. Remove any weight that might be on the scale platform before the scale is switched on.
5. Once the scale has been switched on, it will go through a LCD display test and then re-zero
to be ready for use.
6. The scale requires 15~20 minutes warm up before operation to ensure best accuracy.
7. Please note when the symbol is shown on the display, the internal battery needs to
be recharged.
8. All goods weighed should be placed in the centre of the platform for accurate weighing.
The footprint of the goods being weighed should not overhang the edges of the platform.
V1.1 ZSME300000011
3
Page 6
EXCELL PRECISION CO., LTD.
INTRODUCTION
1. FEATURES
1. Dual-weighing units: Kilogram (kg) and pound (lb).
2. Multi-function operation: Full range tare; Pre-tare; Hold function; Net / Gross weight
display; Simple counting; Preset weight value and quantity value; Adjustable noise filter
setting from 01 ~ 15 (³ 15,000 division configuration only); Check-weighing configuration.
3. User-friendly design:
¨ Auto calibration
¨ AC / DC power supply
¨ Large LCD display with built-in backlight
¨ Auto power-off design to ensure the performance stability
¨ Double over-load protection
4. Variable calibration settings depending on the different calibration division:
¨ Standard division (under 10,000 internal resolution): Capacity and weight calibrations
are available for accurate weighing.
¨ High precision division (over 10,000 to 150,000 internal resolution): Linearity, capacity,
and weight calibrations are available for accurate weighing.
5. Options
¨ RS-232 interface
¨ Relay output
6. High performance in A/D converter (weighing indicators)
¨ Input sensitivity: 0.3 µ V/D
¨ Sample speed up to 16 times / second
¨ Non-linearity: 0.01 % of full scale
¨ Input zero range: -1 mv ~ +5 mv
¨ Input signal range: –1 mv ~ +14 mv
¨ Load cell excitation: 5V DC ± 5% 100mA
¨ Load cell drive capacity: up to 4 350 Ω load cells
V1.1 ZSME300000011
4
Page 7
EXCELL PRECISION CO., LTD.
10° C ~ 40° C (14° F ~ 104° F)
2. SPECIFICATIONS
The available capacities for selection are different and depend on various models:
Capacity Division
3 kg (6 lb) 0.2 g (0.0005 lb)
6 kg (12 lb) 0.2 g (0.0005 lb)
7.5 kg (15 lb) 0.5 g (0.001 lb)
15 kg (30 lb) 1 g (0.002 lb)
30 kg (60 lb) 2 g (0.005 lb)
75 kg (60 lb) 5 g (0.01 lb)
150 kg (150 lb) 10 g (0.02 lb)
300 kg (600 lb) 20 g (0.05 lb)
600 kg (1200 lb) 50 g (0.1 lb)
Operating Temperature: 0° C ~ 40° C (32° F ~ 104° F) OIML: -
Power Source: AC 110V/220V (± 10%) + DC 6V/4.5AH Rechargeable Battery
Display: LCD, 6 digits, 25 mm (Height), EL backlight
3. POWER CONSUMPTION
1. The power consumption is around DC 20 mA (system + load cell), the rechargeable battery
can be used up to 200 hours.
2. The power consumption is around DC 67 mA (system + load cell + backlight), the
rechargeable battery can be used for 60 hours approximately.
3. The power consumption is around DC 71 mA (system + load cell + backlight + RS-232), the
rechargeable battery can only be used for 56 hours approximately.
4. LOW BATTERY WARNING
Please note when the symbol is shown on the display, the internal battery needs to be
recharged.
v The scale will power off automatically without recharging after the low battery symbol
shows upon for 20 to 30 hours (3 to 5 hours if the backlight is active) on the display. As a
recommendation, the scale must be fully recharged before operating the scale again.
V1.1 ZSME300000011
5
Page 8
EXCELL PRECISION CO., LTD.
LCD DISPLAY SYMBOLS
HIGH : Preset High Limit Weight Value
OK : The range between Low & High Limit Weight Value
LOW : Preset Low Limit Weight Value
kg : “kg” unit
3
lb : “lb” unit
Pcs : “COUNTING” mode indicating the number of pieces
à0ß : “ZERO” indication and platform stable confirmation
Net : “Net Weight ” indication
: “ Low Battery” indication
61 : (STABLE) “ Stable” indication
6 2 : ( PT) “ Pre-set Tare” mode
6 3 : (M+) “ Accumulation” mode
.
. .
6 4 : ( ) “Sample Too Small” indication
6 5 : ( ) “ Piece Weight Too Small” indication
. . .
. . . .
Pcs
V1.1 ZSME300000011
6
Page 9
EXCELL PRECISION CO., LTD.
KEYBOARD FUNCTION
UNITS
1. : Units Key to choose the desired unit weight (kg or lb).
2. : Preset High & Low Limit value.
^ /
3. : Ù Þ Data Entry Key, Þ Backlight Key.
4. : Þ Print Key to print the total data ; ¿ Þ Confirmation Key.
5.
Pcs
6. : Accumulation Key to add the displayed value into memory.
:
Sample Key to set the sample quantity.
7. : Re-call accumulated or preset high & low limit values or preset tare value.
8. : Clear Key to delete the existing accumulated values.
9. : Press this key to enter the counting mode.
10. : Preset Tare Key.
R
CE
.
. .
. . .
. . . .
PT
11. : Zero Key to re-zero the scale.
12. : Tare Key to deduct the container weight.
T
KEYBOARD SYMBOLS:
COUNT SMPL UNITS
PRE-TATR
M+ PST
.
. .
. . .
. . . .
M+
PT
Pcs
UNITS
Þ
ZERO RE-CALL
Ù
→0←
R
Ù
BL
TARE CE
PRINT
¿
V1.1 ZSME300000011
7
T
CE
Page 10
EXCELL PRECISION CO., LTD.
OPERATION
1. DISPLAY BACKLIGHT
Press the key to choose the desired display backlight mode:
. Þ “Auto Backlight” mode. When the weight is over 10 divisions or any
. Þ Display backlight is on all the time.
^ /
key is pressed, the display backlight will be switched on. When the
weight returns to zero or the weight on platform is less than 10
divisions, the display backlight will switch off after 5 seconds.
. Þ Display backlight is off.
2. ZERO
Press the key to re-zero the display with no load on the weigh pan. When zero is set,
the (à0 ß ) symbol will be displayed.
3. WEIGHING MODE
(1). Units Selection
1. Press the key in the weighing mode to choose the desired weighing unit and the
display will show the “ kg” or “ lb” symbol on the top right of the LCD display.
2. The unit status will be memorized when turning the machine off.
(2). Totalising
1. The scale allows the next totalising operation, even when the weight value does not return
back to zero. The
than 10 d. The scale will store the totalised weight value after the weight is stable.
2. The scale can totalise positive or negative weight but not both at the same time. The
totalised weight store must be reset to zero before it is possible to select positive or
negative totalising mode.
V1.1 ZSME300000011
key is functional, when the weight value changes by more
8
Page 11
EXCELL PRECISION CO., LTD.
I. Weight Totalising
ACTION
DISPLAY
Place the object onto the
scale.
The display shows
the weight value
Press the M+ key
.
STABLE
.
STABLE
kg
kg
The display shows the
total number of additions
The total weight totalised
shows and the 3
rd
“6 ”
kg
.
flashes on the display
STABLE
M+
v The totalising function can be used up to a maximum of 9999 times before it must be reset.
The totalising display is limited to 6 digits maximum.
II. Recall Totalised Weight Values
1. Press the key to display the total number of additions and the totalised weight
value. The 3
R
rd
“6 ” symbol located above (M+) icon will flash on the display. The scale will
return to the weighing mode after 3 seconds.
v The scale will not display the negative sign “ -” for negative totalised weight values when
recalling an a totalised weight value, but the negative sign “ -“ will be printed out (transmitted
serially) for each negative weight and negative totalised weight.
III. Clear Totalised Weight Values
1. Press the key followed by pressing the key to clear the totalised weight
R
values.
2. All totals will be lost in the following three circumstances:
¨ The mode is changed from weighing to counting or vice versa.
¨ The scale is switched off.
¨ The weight unit is changed.
V1.1 ZSME300000011
9
Page 12
EXCELL PRECISION CO., LTD.
Place the container onto the
(3). Deduction of the Container Weight
I. The weight of the container is unknown ( Tare )
ACTION
scale
Press the
The weight value becomes
zero and the
displayed
Place the object into the
container and the display
shows the net weight value
of the object.
T
key when the
stable.
.
STABLE
Net
.
STABLE
Net
.
STABLE
II. Recall the semi-auto tare value
DISPLAY
kg
kg
kg
R
Press the key followed by pressing the key Þ The display shows
T
the tare weight value.
III. Clear the semi-auto tare
When the container is removed from the scale, the display shows the container weight
value with a negative sign. Press the key to reset the scale to zero, and the (Net)
symbol will switch off.
NOTE:
v Multiple tare operation Þ Users can continuously increase or decrease the tare value by
pressing the key.
T
v The total tare value (tare value + pre-set tare value) can equal the full capacity of the scale.
V1.1 ZSME300000011
10
Page 13
EXCELL PRECISION CO., LTD.
weight value of the container
IV. The weight of the container is known (Pre-set tare)
Press the key
Use the Ù key to enter the
followed by pressing the ¿
key
Place the object with the
container onto the weight pan.
The display shows the net
value of the object
ACTION
PT
.
STABLE
Net
STABLE
Net
.
STABLE
DISPLAY
.
PT
PT
V. Recall the pre-set tare value
kg
kg
DESCRIPTION
The display shows the weight value
of the container with the negative
sign and the (Net) symbol. The
second “6 ” above the (PT) icon
indicates the tare is active
kg
R
Press the key followed by pressing the key Þ The display shows the
PT
pre-set tare value.
VI. Clear the pre-set tare value
PT
Press the key followed by pressing the key, then press the
R
key to clear the pre-set tare value. The scale resets back to zero, and the (Net) symbol
and the second “6 ” icon (PT) will switch off.
NOTE:
v In Tare mode, the Preset tare function is disabled.
v The scales with two weighing ranges can NOT pre-set the tare value larger than the first
weighing range. For example: a 30kg scale is set by two weighing ranges. The first range is
0 to 15kg, and the second range is 15 to 30kg. The pre-set tare value can not be larger
than 15kg.
V1.1 ZSME300000011
11
Page 14
EXCELL PRECISION CO., LTD.
weight of the object is lower or
(4). Check Weighing
Press the key to enter the check weighing function. Press the key to
return back to the weighing mode.
I. Pre-set the “ Low limit value” , “ High limit value” , and “ the buzzer” operation
Press the key to enter the value required and step to the next digit by
^ /
pressing the key.
ACTION
DISPLAY
DESCRIPTION
Press the
the check weighing function
Set the “ Low limit value”
Press the Ù key to enter the
Low limit value. Press the ¿
key to confirm the setting
Set the “ High limit value”
Press the Ù key to enter the
High limit value. Press the ¿
key to confirm the setting
Set the “ buzzer” operation
Press the Ù key to set the
buzzer operation. Press the ¿
key to confirm the setting
The scale returns to the
︿ key = Upward key (0~9 digit entry)
key to select
.
STABLE
.
STABLE
.
STABLE
.
STABLE
STABLE
STABLE
.
STABLE
kg
kg
kg
kg
kg
kg
kg
Buzzer Setting
C C
A B
Þ
= When the weight is stable the
buzzer sounds and the High, OK or
Low icon is displayed.
1 =. The High, OK or Low icon is
displayed irrespecti
of the weight. The buzzer sounds
only when the weight is stable.
2 = The buzzer sounds and the
High, OK or Low icon is displayed
irrespective of the stability of the
.
B Þ
0 = The buzzer does not sound.
1 = The buzzer sounds at the OK
status.
¿ key = Confirmation key
2 = The buzzer sounds when the
(move the cursor one step to the right)
equal to the lower limit value, or
higher than the upper limit value.
V1.1 ZSME300000011
12
Page 15
EXCELL PRECISION CO., LTD.
II. Pre-set the Low limit value only
After completing the Low limit value setting, the display shows . . Press the
key to complete the setting, and the display shows
STABLE
kg
v When users only pre-set the Low limit value, the buzzer setting is fixed at:
III. High/OK/Low indications
LOW OK HIGH
½ппппппп½ пппппррррр½ рррррр
10d
Low limit
value
High limit
value
IV. Recall the check weighing setting
1. Press the key followed by pressing the key to recall the Low limit
R
value.
R
2. Then press the key again to recall the High limit value.
3. Then press the key again to recall the buzzer setting.
4. Then press the key again to return the weighing mode.
R
R
V. Clear the check weighing settings
1. Press the key followed by pressing the key, and then press the
R
key to clear the Low limit value.
2. Press the key again to clear the High limit value and the buzzer setting.
V1.1 ZSME300000011
13
Page 16
EXCELL PRECISION CO., LTD.
The display shows the number
4. COUNTING FUNCTION
.
. .
. . .
Press the key to enter the counting function. Press the key to return back
to the weighing mode.
(1). Sampling
Press the key
counting function
Press
number of desired items in the
sample (10, 20, 50 or 100)
Place the sample on the scale
and press the key
of items in the sample
. . . .
ACTION
.
. .
. . .
. . . .
Pcs
to choose the
.
. .
. . .
. . . .
the weight is stable
DISPLAY
STABLE
STABLE
STABLE
Pcs
Pcs
Pcs
Pcs
NOTE:
v The larger the sample size, the more accurate the unit weight. (The minimum sample
weight = 20d)
v A sample size of 10 pieces gives a typical count accuracy of 95%.
v A sample size of 50 pieces gives a typical count accuracy of 98%.
.
. .
SAMPLE TOO SMALL ( ) Þ Sample is less than 20 divisions.
UNIT WEIGHT TOO SMALL ( ) Þ Unit weight is less than 1/5 of a division.
. . .
. . . .
Pcs
Under such conditions, the scale can still work, but may result in lower count
accuracy.
(2). Totalising
Refer to the operation of totalising in the weighing function on page 9.
(3). Check Weighing
Refer to the operation of check weighing function on page 13.
V1.1 ZSME300000011
14
Page 17
EXCELL PRECISION CO., LTD.
CONFIGURATION SETTINGS
Switch on the scale. While the scale is counting backward to zero, press and hold the
key until the display shows the software program version number: “ ” .
Release the key, the scale enters the configuration setting mode. is displayed.
Ù
Press the
Press the
Ù
Ù
key
key
key
Press the
Ù
key
Press the
Press the
Press the
TARE
TARE
TARE
key
key Press the
key
Press the
Press the Ù key
Press the
Press the
Press the
Press the
Press the
Press the
Ù
Ù
Ù
Ù
Ù
Ù
Ù
key
key
key
key
key
key
key
Press the
Press the
Press the
Press the
Press the
Press the
Press the
TARE
TARE
TARE
TARE
TARE
TARE
TARE
Þ INTERNAL CALIBRATION (LOCKED)
key
Þ INTERNAL CALIBRATION (LOCKED)
Þ INTERNAL CALIBRATION (LOCKED)
key
Þ INTERNAL CALIBRATION (LOCKED)
Þ CHECK-WEIGHING CONFIGURATION
key
Þ RS-232 SETTING
Þ EXIT FROM CONFIGURATION
key
Þ INTERNAL CALIBRATION (LOCKED)
SETTINGS
key
Þ INTERNAL CALIBRATION (LOCKED)
Þ INTERNAL CALIBRATION (LOCKED)
key
key
V1.1 ZSME300000011
15
Page 18
EXCELL PRECISION CO., LTD.
1. CHECK-WEIGHING CONFIGURATION
¨ F8 (Hold) affects access to F4. If F8 is set, access to F4 is denied.
Press the or key to select the F4 functionÞ the display shows
^ /
T
Press the ¿ key to enter the
check weighing configuration
Press the Ù key to enter the
low limit value. Press the ¿
key to confirm the setting
Set the “ High limit value”
Press Ù key to enter the high
limit value. Press the ¿ key
confirm the setting
Set the “ buzzer” operation
Press the Ù key to set the
buzzer. Press the ¿ key to
confirm the setting
The scale returns to the
ACTION
.
STABLE
.
STABLE
.
STABLE
.
STABLE
STABLE
STABLE
ode.
︿ key = Upward key (0~9 digit entry)
¿ key = Confirmation key
(move the cursor one step to the right)
DISPLAY
Press the ¿ key
kg
for 5 times to clear
the configuration
settings
kg
Press the ¿ key
kg
Buzzer Setting
kg
C C
A B
Þ 0 = When the weight is stable
kg
the buzzer sounds and the Hi
OK or Low icon is displayed.
1 =. The High, OK or Low icon
is displayed irrespective of the
kg
stability of the weight. The
buzzer sounds only when the
weight is stable.
2 = The buzzer sounds and the
High, OK or Low icon is
displayed irrespective of the
stability of the weight.
B Þ 0 = The buzzer does not sound.
1 = The buzzer sounds at the
OK status.
2 = The buzzer sounds when
the weight of the object is lower
or equal to the low limit value,
or higher than the high limit
value.
DESCRIPTION
.
STABLE
V1.1 ZSME300000011
16
Page 19
EXCELL PRECISION CO., LTD.
2. RS-232 SETTING
¨ J1 and J3 on the RS-232 interface are connected together (short), when the RS-232
interface is connected to a computer.
¨ J2 and J4 on the RS-232 interface arte connected together (short), when the RS-232
interface is connected to a printer.
Press the or keys to select the F5 function Þ the display shows
^ /
T
ACTION
DISPLAY
DESCRIPTION
Press the ¿ key
Transmission mode
Press the Ù key to set
the transmission mode
Then press the ¿ key
to confirm the setting
Baud rate default setting
Press the Ù key to set
the baud rate (1200, 2400,
4800 or 9600)
Then press the ¿ key
to confirm the setting
Configuration complete
0 Þ No transmission
1 Þ Stable transmission
2 Þ Continuous transmission
3 Þ Press the ¿ key to
transmit (simple mode)
4 Þ Press the ¿ key to
transmit (complete mode)
5 Þ Stable transmission
(totalising mode). The
format is as same as
3.
6 Þ EZ-2 printer mode
The format is as same as
4.
7 Þ EZ-2 printer mode.
Press the ¿ key to
transmit
8 Þ Press the ¿ key to
transmit. The format is the
same as 1 & 2.
v When choosing “ EZ-2 printer mode “ , the baud rate should be set to 9600 Baud
V1.1 ZSME300000011
17
Page 20
EXCELL PRECISION CO., LTD.
3. EXIT CONFIGURATION SETTINGS
T
Press the or key to select the F6 functionÞ the display shows
^ /
ACTION
Press the ¿ key
DISPLAY
Scale reverts back to the
weighing mode
V1.1 ZSME300000011
18
Page 21
EXCELL PRECISION CO., LTD.
RS-232 AND RELAY OUTPUT
1. RS-232 OR SERIAL PRINTER OUTPUT
(1) RS-232 (25 Pin ‘ D’ type) Pin Description
˙ J1 - J3 SHORT ; J2 - J4 OPEN (default setting)
Pin 2 Þ RXD
Pin 3 Þ TXD
Pin 7 Þ GND
˙ J2 - J4 SHORT ; J1 - J2 OPEN
v Please refer to F5 function settings for transmission mode, baud rate setting and data
Pin 2 Þ TXD
Pin 3 Þ RXD
Pin 7 Þ GND
format.
(2) RS-232 Interface Format
I . Mode : EIA-RS232 C’ s
II . Format :
1. Baud rate: 1200, 2400, 4800 or 9600
2. Data bits: 8 BITS
3. Parity bit: None
4. Stop bits: 1 BIT
5. Code: ASCII
Start bit Data bits Stop bit
V1.1 ZSME300000011
19
Page 22
EXCELL PRECISION CO., LTD.
(3) Data Format
Stable transmission ( ) / Continuous transmission ( )
Press the ¿ key to transmit ( )
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18
HEAD 1 , HEAD 2 , DATA UNIT CR LF
HEAD 1 ( 2 BYTES ) HEAD 2 ( 2 BYTES )
OL - Overload , Under load TR - TARE Mode
ST - Display is Stable NT - NET Mode
US - Display is Unstable GS - GROSS Mode
DATA (8 or 9 BYTES)
2D ( HEX ) = “ – ” ( MINUS )
2E ( HEX ) = “ . ” ( DECIMAL POINT )
UNIT ( 2, 3 or 4 BYTES )
kg = 6B ( HEX ) ; 67 ( HEX )
lb = 6C ( HEX ) ; 62 ( HEX )
tl.T = 74 ( HEX ) ; 6C ( HEX ) ; 2E ( HEX ) ; 54 ( HEX )
hkg = 68 ( HEX ) ; 6B ( HEX ) ; 67 ( HEX )
Transmission example:
1. The gross weight (+0.876kg) shows as below, after stable: (in the tare mode)
S T , G S , + 0 . 8 7 6 k G 0D 0A
HEAD 1 , HEAD 2 , DATA UNIT CR LF
2. The net weight (-1.568lb) shows as below without weight stability: (Not in the tare mode)
U S , N T , - 1 . 5 6 8 l b 0D 0A
HEAD 1 , HEAD 2, DATA UNIT CR LF
3. The net weight (+0.876kg) shows as below, after stable: (in the tare mode)
S T , T R , + 0 . 8 7 6 k G 0D 0A
HEAD , HEAD , DATA UNIT C LF
kg/lb
V1.1 ZSME300000011
20
Page 23
EXCELL PRECISION CO., LTD.
Press the ¿ key to transmit (simple mode)
S/N WT/UNIT ( kg / lb )
- - - - - - - - - - - - - -
0001 1.0000 E Press the
0002 1.0000 E Press the
0003 1.0000 E Press the
0004 1.0000 E Press the
0005 1.0000 E Press the
- - - - - - - - - - - - - -
0005 5.0000 E Press the
¿
key or the
¿
key or the
¿
key or the
¿
key or the
¿
key or the
¿
key for 2 times to printout the total
M+
M+
M+
M+
M+
key
key
key
key
key
Stable transmission (totalising mode)
S/N WT/UNIT ( kg / lb )
- - - - - - - - - - - - - -
0001 1.0000 E scale stable, transmitting
0002 1.0000 E scale stable, transmitting
0003 1.0000 E scale stable, transmitting
0004 1.0000 E scale stable, transmitting
0005 1.0000 E scale stable, transmitting
- - - - - - - - - - - - - -
0005 5.0000 E Press the ¿ key for 2 times to printout the total
EZ-2 printer mode, press the ¿ key to transmit
(Baud rate must be set at 9600)
Only prints out the “ weight value” . The proportion of the printed typeface Þ Height : Width = 3:2
+100.0 kg
V1.1 ZSME300000011
21
Page 24
EXCELL PRECISION CO., LTD.
EZ-2 printing mode (Baud rate must be set to 9600). Press the ¿ key to
transmit (complete mode)
E Press the ¿ key or the M+ key
E Press the ¿ key or the M+ key
E Press the ¿ key or the M+ key
E Press the ¿ key twice to print out the total
G 1.000kg
T 0.000kg
PT 0.000kg
N 1.000kg
(Blank line x 3)
TICKET NO .0002
G 1.000kg
T 0.000kg
PT 0.000kg
N 1.000kg
(Blank line x 3)
TICKET NO .0003
G 1.000kg
T 0.000kg
PT 0.000kg
N 1.000kg
(Blank line x 3)
TOTAL NUMBER
OF TICKETS 0003
TOTAL
(Blank line x 3)
<Remarks> : G = GROSS T = TARE PT = PRE-TARE N = NET
V1.1 ZSME300000011
22
Page 25
EXCELL PRECISION CO., LTD.
2. RELAY OUTPUT
(1) Relay Output
Relay’ s function acts as signal output in the check-weighing mode. When an object’ s
weight value reaches one of the setting points (HI, OK, LOW), the relay outputs a signal
through PIN 1, PIN2, or PIN 3. To set the check-weighing configuration by pressing the
key (in TW & AW models) or enter F4 (please refer to page 12 or 16 for operations).
(2) Relay Pin Description
PIN 1 Þ OK output
PIN 2 Þ High output
PIN 3 Þ Low output
PIN 4 Þ VDD
PIN 5 Þ GND
PIN 6 Þ COMMON
(3) Connection Illustration
V1.1 ZSME300000011
23
Page 26
EXCELL PRECISION CO., LTD.
(4) Power Source Illustration
The relay circuit board can be supplied either by the external AC power or internal DC
power.
• Relay powered by external AC power
PIN4 and PIN5 are connected with AC power source; J1 and J2 OPEN
‚ Relay powered by internal DC power
No power source input to PIN4 and PIN5; J1 and J2 SHORT
(5) Connection Setting
OK, High, Low can be set at NO (normal open status) or NC (normal close status)
respectively.
• OK、 High 、 Low at NO ( normal open status) Þ default setting
‚ OK、 High、 Low at NC ( normal close status)
Please use a knife to cut the J3, J4 and J5 not connect (open) and solder the J6, J7
and J8 connected.
V1.1 ZSME300000011
24