HJW
© Excell Precision Limited 2014. All rights reserved Worldwide.
The information contained herein is the property of Excell Precision Limited and is supplied without liability for errors or omissions. No part may be reproduced or used except as authorised by contract or other written permission. The copyright and the foregoing restriction on reproduction and use extend to all media in which the information may be embodied.
EXCELL PRECISION CO., LTD
02022010 ZSME300000197
1
Table of Contents
Before Using the Scale ..............................................................................................................................2
Instructions for Use .....................................................................................................................................2
Preparing to Use the Scale........................................................................................................................2
Chapter 1 Introduction ................................................................................................................................4
1-1 Features.............................................................................................................................................4
1-2 Display Description ..........................................................................................................................4
1-3 Keypad Function...............................................................................................................................5
1-4 Power Supply ....................................................................................................................................7
1-5 Error Messages ................................................................................................................................8
1-6 Weight Units ......................................................................................................................................8
1-7 Self-Test Mode ..................................................................................................................................9
Chapter 2 Configurations ........................................................................................................................ 10
2-1 General Function Setting ................................................................................11
2-1-1 Auto Backlight Function Setting ............................................................ 12
2-1-2 Auto Power-off Timer Setting ................................................................. 13
2-1-3 HI/LO/OK Settings ................................................................................... 14
2-1-4 Restore to the Default Settings ................................................................ 15
2-1-5 Noise Filter Settings ................................................................................ 16
2-1-6 Hold Function Settings ............................................................................ 17
2-1-7 Auto Unit Weight Averaging Setting ...................................................... 19
2-1-8 Record Last Zero .......................................................................................... 19
2-2 External Weight Calibration .............................................................................. 20
Appendix 1 ASCII Code Table................................................................................................................ 21
Appendix 2 Wiring Instructions .............................................................................................................. 22
Appendix 3 7-Segment Display Characters......................................................................................... 23
EXCELL PRECISION CO., LTD
02022010 ZSME300000197
2
Before Using the Scale
In order to use this scale correctly, we suggest you read this manual carefully.
Instructions for Use
1. Please keep the scale in a cool and dry place. Do not store at high temperatures or damp
places.
2. Please keep the scale clean and free from insect infestation.
3. To avoid impact and it is to be used under designated pressure (the weight put on the platter
can not exceed the maximum capacity of the scale).
4. If the scale is not going to be used for some time, please clean and store it in a plastic bag
with desiccants in a dry condition. The scale should be charged every 3 months to prevent failure of the internal rechargeable battery.
5. Please don’t use different types of dry battery or mix new batteries with used ones.
6. Please operate or charge the scale in an open area. Do not squeeze the power cord to avoid
wire on fire.
7. Operating temperature: -10℃~ +40℃
8. Any suggestion for product is welcome.
Preparing to Use the Scale
1. Avoid operating the scale in direct sunlight or damp environment.
2. If possible avoid connecting the scale to ac power outlet sockets which are adjacent to other
appliances to minimise the possibility of interference affecting the performance of the scale.
3. Remove any weight that might be on the weigh pan before the scale is switched on and
avoid leaving weight on the pan for long periods of time.
4. All goods weighed should be placed in the centre of the weigh pan for accurate weighing.
The overall dimensions of the goods being weighed should not exceed the dimension of the
weigh pan.
5. Once the scale has been powered on, it will go through an LCD display test and it is ready
for use when the display shows zero.
6. The scale requires 15~20 minutes to warm up before operation to ensure the best accuracy.
7. Please note when the symbol keeps flashing on the screen, the scale needs to be
charged.
8. Introduction of Storage Battery
Thank you for choosing EXCELL Weighing Hanging Scale. To guide you to use our product correctly, please read this User Manual carefully to extend the life of machine and to avoid maloperation.
EXCELL PRECISION CO., LTD
02022010 ZSME300000197
3
No burning
Caution Corrosion
Due to the storage battery adopt the advanced free-maintaining technique, customers need not to replenish electrolyte. The scale should be recharged every 3 months to prevent failure of the internal rechargeable battery.
1. The battery should be charged for 8~10 hours.
2. The temperature of battery should below 45.
Maintaining
1. Please do not discharge with over-current when using the battery. Please charge the battery
after discharging current.
2. Please take down the battery when the scale is not used for a long time or break the
connection of cathode.
3. Do not short the battery terminals to check whether there is current. Please check whether
the connection point is firm to guarantee good connection.
4. The battery should be replaced by specialized person. No reverse-battery or the product
will be damaged.
a) Anode of battery should be connected with Anode of product batteryusually red
cable
b) Cathode of battery should be connected with Cathode of product battery (usually brown
cable or black cable)
c) See the picture following
Safety warnings
1. The electrolyte of battery is caustic which causes metal, cotton, etc to corrode.
2. The hydrogen will be resolved when using or charging the battery and it will cause explosion
when approaches fire.
Brown cable(or black cable) connected with Anode of battery
Red cable connected with Cathode of battery
Warning explosion
Children faraway
EXCELL PRECISION CO., LTD
02022010 ZSME300000197
4
Pcs
“” indications description: 1 : (STABLE) stable indication 2 : (GROSS) gross weight indication 3 : ( ) unit weight insufficient indication 4 : Samples insufficient indication 5 : ( PT ) pre-tare indication 6 : (Hold) weight hold indication 7 : M+ indication
Chapter 1 Introduction
1-1 Features
Up to 1/6,000 display resolution Hold function, Check weighing with LO / HI / OK Simple counting function Full range tare, Pre-tare, Auto zero tracking, Gross/Net Indication Selectable unit mode Low power indication and auto power off LCD display with LED backlight
1-2 Display Description
When the scale with 2-segment specification, the indications description will be:
3 : Range 1
4 : Range 2
1 2 3 4 5 6 7
HI : Upper limit OK : Value between HI and LO LO : Lower limit Kg : Unit “kilogram” lb : Unit “pound” t : hk tael Pcs : Counting mode indication
0: Zero point indication
Net : Net weight indication
: Low Power indication
EXCELL PRECISION CO., LTD
02022010 ZSME300000197
5
1-3 Keypad Function
ON/ZERO
KEY
When the power is off, press
ON/ZERO
key to switch the indicator on.
When the power is on, press
ON/ZERO
key to return to zero.
Zero Range for non-approval models is ±2% Full Scale.
OFF
KEY
Press the
OFF
key to switch the indicator off.
TARE/PT
KEY: Tare / Pre-tare key
To TARE:
Place the container onto the scale, wait until the weight value is stable, press TARE
/PT key for zero return and the NET indication is shown on the display. Place the object onto the container and the display shows the net weight value of the object. Remove both object and container, and negative value of the container will show on the display. Press TARE
/PT again
to clear “tare value”. The scale returns to zero and NET indication
goes off.
TARE function can remain effective until tare value=full load capacity Continuous Tare  Press TARE
/PT key
for continuous weight increase/decrease on
platter.
Pre-tare cannot be done in tare function. However, if a pre-tare had been preset, and
the tare weight is more than that pre-tare weight, then the Tare can be done.
No Tare can be done under gross weight display mode.
To PRE-TARE:
Press TARE / PT key, use keypad to input pre-tare weight. When the cursor flashes on the rightmost digit, press TARE / PT key again to complete.
Pre-tare mode, keypad function as followed:
MC => upward key(0~9 digit entry) NET/GROSS => move cursor leftward UNIT => downward key(0~9 digit entry) TARE/PT => move cursor rightward
UNIT
KEY
Press the
UNIT
key to switch weight units; the display icons will indicate the active units.
EXCELL PRECISION CO., LTD
02022010 ZSME300000197
6
Pcs
NET/GROSS
KEY
In the Tare mode, the screen displays the “
TARE
” icon; press the
NET/GROSS
key to switch between the “Net value” and the “Gross value”. When the ”GROSS” arrow
shows up, the Gross Weight in the screen = Tare weight +
Net weight, meanwhile, only
NET/GROSS
key is available.
M+
KEY
Sum-up function. This key is a composite key, while the item total is shown and weight returns to net zero,
press M+ key to erase memory.
When a is new weight added on platter, the sum-up function can be effective. If this
weight had not been removed, no new item can be added. Display will show the item total for one second, and show the net weight for one second, then the scale returns to the current weight, and print out the last total.
To clear the Sum-up info, press M+ key to bring up the item total show up, then press
M+ again, to clear the Sum-up value data.
P.s. To execute the clear function, the weight must return to zero.
MC
KEY
Memory clear for accumulation data.
Simple Counting Mode
Use UNIT key to switch unit to Pcs, to enter simple counting mode.
1. Use
NET/GROSS
key to select a sample number from “10 , 20, 50, 100, or 200”. Display will show in an sequential order by pressing
NET/GROSS key.
2. Select a sampling number, and put appropriate weight on platter, and press UNIT key. Display will show “ ”. The scale will enter counting mode after weight is stable, and display will show the sample number.
Unit weight insufficient ( ) : Sample unit weight < 0.2d (d=division) Samples insufficient ( ) : total sample < 20d (d=division) While sampling, if there is insufficient sample or unit weight insufficient indication “”, the
scale is still usable, but there may be slight inaccuracy.
After Power-off, the scale automatically memorizes the sampling number, and it is available
when “Pcs” unit is selected next time.
If the setting is “automatic average unit weight “, if the object on platter > the previous
sampling number more than 5 pcs, and also < less than 100% the previous sampling number, the scale will execute unit weight calibration automatically.
.
. .
. . .
. . . .
EXCELL PRECISION CO., LTD
02022010 ZSME300000197
7
1-4 Power Supply
Power Supply Selection
1. 6 V / 7 Ah Rechargeable battery
2. 110 / 220 V ±15 % AC Main power
Power Consumption
40 mA (Indicator + Load Cell) About 180 hours
70 mA (Indicator + Load Cell + display backlight) About 100 hours
Charging Voltage
DC 10V/1A
Low Battery Warning
Please note when the (
) symbol keeps flashing on the display, the internal battery
should be recharged.
The scale will turn off automatically after a few hours when the low battery warning symbol
shows up. The scale must be fully charged before operating again.
When the battery status indication is full, the power is about 6.1V; each block is about 0.2V;
When the battery warning symbol shows up, the power is about 5.4V.
Safety Reminder:
Please make sure “+” , “- “ poles are placed in the correct direction Please don’t’ place the battery in hot areas, or try to dissemble the battery, to avoid
electricity leakage.
Please don’t mix or use batteries of different types. Please don’t keep the used-up battery in the battery case to avoid breakdown.
(dry cell models)
The battery inside the indicator was used for testing in factory.
Therefore, the life of the battery cannot be count from the day you purchase it.
EXCELL PRECISION CO., LTD
02022010 ZSME300000197
8
1-5 Error Messages
Zero is higher than the zero range when switching the scale on. Zero is lower than the zero range when switching the scale on. Internal value is unstable.
Continuous unstable occurred over 10 sec when switching the scale on, or after pressing ON/ZERO or TARE/PT key.
The weight of the object is over 9 divisions of the maximum capacity.
1/6 full scale lower than zero.
1-6 Weight Units
(kg)
1 g = 0.001 kg
(g)
1 g = 1 g
(lb)
1 g = 0.002204623 lb
(lb/oz)
1 g = 0.03527396 oz
(oz)
1 g = 0.03527396 oz
(GN)
1 g = 15.432358 GN
(dwt)
1 g = 0.6430149 dwt
(ct)
1 g = 5 ct
(hk.tael)
1 g = 0.02645546 Hk.catty
viss
1kg = 0.612245 viss
EXCELL PRECISION CO., LTD
02022010 ZSME300000197
9
Press
UNIT
key
Press
UNIT
key
1-7 Self-Test Mode
When power is off, hold
NET/GROSS
and press
ON/ZERO
key, Wait till display shows
to enter “Self-Test Mode”.
INTERNAL VALUE MODE
(must hook up full-bridge Load Cell to test)
Press
TARE / PT
to enter, and the display shows internal value
Put on the weight and check whether the internal value has obvious change. Press
ON/ZERO
key to back to the last screen , the display shows
KEYPAD TEST MODE
Press
TARE/ PT
to enter, display shows
Keypad’s internal code:
TARE/ PT
key= 03 ,
UNIT
key= 05,
NET/GROSS
key=06,
MC key = 04 , M+ key=02
Press
ON/ZERO
key to back to the last screen , the display shows
FIRMWARE VERSION DISPLAY MODE
Press
TARE/ PT
to enter , display shows the firmware version ,
Press
TARE/ PT
key again, the display shows maintenance number for 1
seconds
Press
ON/ZERO
key to back to the last screen, display shows
BACK TO THE LAST SCREEN
Press
TARE/ PT
key to exit self-test mode, the scale will re-power on automatically.
Press
MC
key
Press
MC
key
Press
MC
key
Press
UNIT
key
Press
UNIT
key
EXCELL PRECISION CO., LTD
02022010 ZSME300000197
10
Chapter 2 Configurations
Configuration Workflow
In the weighing mode, press the
NET/GROSS
and
ZERO
keys at the same time to enter the
configuration mode. The LCD shows .
Press
UNIT
key
Press
UNIT
key
Press
UNIT
key
Press
MC
key
Press
MC
key
Press
MC
key
General Function Setting Mode External Weight Calibration Exit the Advanced Function Setting Mode
EXCELL PRECISION CO., LTD
02022010 ZSME300000197
11
2-1 General Function Setting
ON/ZERO
key Upward key (0~9 digit entry)
UNIT
key Downward key (0,9~1 digit entry)
TARE/PT
key Move cursor rightward
NET/GROSS
key Move cursor leftward
Press
TARE/PT
key
key in the parameter
“00 ~ 12
Key in “00” and
press
TARE/PT
key to
return to the Advanced function setting
Return to the Advanced Function Setting Mode Menu Automatic Backlight Function Settings
Automatic Power-off Timer Settings Hi/Lo/OK Settings Restore the Default Settings Noise Filter Settings Hold Function Settings Auto Unit Weight Averaging Setting Reserved Reserved Record Last Zero
EXCELL PRECISION CO., LTD
02022010 ZSME300000197
12
2-1-1 Auto Backlight Function Setting
Select in the general function setting mode to change the backlight function setting.
LCD displays the last status Default setting = off
Auto backlight function
When the weight is over 10d, the display backlight will be on. After the weight is stable for 10 seconds or when the scale returns to zero, the display backlight switches off. (d=division)
Press
TARE /PT
key
Use
ON/ZERO
or
UNIT
key
to select function “on” or “off
Press
TARE/PT
key to go back to the advanced function setting menu
On backlight function is on Off backlight function is off
EXCELL PRECISION CO., LTD
02022010 ZSME300000197
13
2-1-2 Auto Power-off Timer Setting
Select in the general function setting mode to change the automatic power-off timer setting.
ON/ZERO
key Upward key (0~9 digit entry)
UNIT
key Downward key (0,9~1 digit entry)
TARE /PT
key Move cursor rightward
NET/GROSS
key Move cursor leftward
LCD displays the last status Default setting = 0 (No auto power-off)
Auto power-off function
When the weight on the platter is less than 10d or keeps idle for the set time, the scale will automatically switch off. (d=division)
Press
TARE/PT
key
Use
MC
or
UNIT
key to
key in parameter
Press
TARE/PT
key to go back to the
Advanced function setting menu
Automatic power-off timer setting Use
MC
or
UNIT
key to key in
parameter 0  No auto power-off
1  When the scale is idle for 1
minute, the scale automatically switches off
2  When the scale is idle for 2
minutes, the scale automatically switches off
9  When the scale is idle for 9
minutes, the scale automatically switches off
EXCELL PRECISION CO., LTD
02022010 ZSME300000197
14
lb
lb
lb
lb
2-1-3 HI/LO/OK Settings
Select in the general function setting mode to set the HI/LO/OK function.
When the high limit and low limit are both set as “0”, the Hi/Lo/OK function is disabled.
MC
key Upward key (0~9 digit entry)
NET/GROSS
key Move cursor leftward
UNIT
key Downward key (0,9~1 digit entry)
TARE/PT
key Move cursor rightward
.
.
.
.
LCD displays the last status
Press
TARE/PT
key
Key in the desired high limit value
Press
TARE/PT
key
Key in the desired low limit value
Press
TARE/PT
key
Key in the parameter setting for alarm
Press
TARE/PT
key
LCD displays the last status HI indicator lights.
LCD displays the last status LO indicator lights.
Alarm setting
(a)(b)(c) The definition of a, b, c positions:
(a) 1 = beep sound on
0 = beep sound off (b) 1 = beep sound on when stable
0 = beep sound on when unstable (c) 1 = beep sound on when the weight is
between high and low limits
0 = beep sound on when the weight is
not between high and low limits and higher than 10d (d=division)
EXCELL PRECISION CO., LTD
02022010 ZSME300000197
15
2-1-4 Restore to the Default Settings
Select in the general function setting mode to restore to the default settings.
The defaults are included the following:
1) External weight calibration
2) HI/LO/OK setting values
3) Noise filter setting (External)
4) Sampling settings for the counting function This function is not available for approval models.
MC
key Upward key (0~9 digit entry)
UNIT
key Downward key (0~9 digit entry)
TARE
key Move cursor rightward
NET/GROSS
or key Move cursor leftward
Return (Cancel the
restoration)
Restore the default
setting
Press
TARE/PT
key
Use
MC
or
UNIT
keys to
select “return” or “format”
Press
TARE/PT
key
EXCELL PRECISION CO., LTD
02022010 ZSME300000197
16
2-1-5 Noise Filter Settings
Select in the general function setting mode to set the noise filter settings.
This function is not available for approval models.
Returning to the zero point setting
Use
MC
or
UNIT
key to key in the
parameters or zero point
Default setting = 0
0  No skip 5 skip 5d 1  skip 1d 6  skip 6d 2  skip 2d 7  skip 7d 3  skip 3d 8  skip 8d 4  skip 4d 9  skip 9d
When the weight on the scale is over 1/3
full capacity, the function is on.
Digital switch & Stabilization range setting LCD displays the last parameter setting
Filter parameter setting
LCD displays the last parameter setting.
Returning to zero point setting
LCD displays the last status
Filter parameter setting
Use
MC
or
UNIT
keys to key in the
parameters. Default setting = 5
Parameter 0 ~ 9. The larger the number is, the faster the filter responds. Fast response can lead to weighing instability. Parameter 9 the AD value is not filtered.
Input AD value = Output AD value
Digital switch & Stabilization range setting Use
MC
or
UNIT
keys to key in the
parameters. Default setting = 0 Parameter 0 ~ 9, the larger the number the more stable the weight.
Press
TARE/PT
key
Use
MC
or
UNIT
key to key in the parameters
Press
TARE/PT
key
Use
MC
or
UNIT
key to key in the parameters
Press
TARE/PT
key
Use
MC
or
UNIT
key to key in the parameters
Press
TARE/PT
key
EXCELL PRECISION CO., LTD
02022010 ZSME300000197
17
2-1-6 Hold Function Settings
This function is not available for OIML or NTEP approval models.
MC
key Upward key (0~9 digit entry)
TARE/PT
key Move cursor rightward
UNIT
key Downward key (0,9~1 digit entry)
NET/GROSS
key Move cursor leftward
Hold function setting Use
MC
or
UNIT
keys to key in the
parameters
Default setting = 0 0  Hold function disabled 1 “Peak hold” mode 2 “Stable hold 1” mode 3 “Stable hold 2” mode
4 “Animal scale hold” mode
Hold function setting
LCD displays the last parameter setting
Press
TARE/PT
key
Use
MC
or
UNIT
key to
key in the parameters
Press
TARE/PT
key
= Hold function disabled = “Peak hold” mode
The scale keeps displaying the maximum weight when the weight is continually changing. Press any key to exit this mode.
= “Stable hold 1” mode
When the weight is stable, the LCD shows the current weight value. Press any key to exit this mode.
= “Stable hold 2” mode
When the weight is stable, the LCD shows the current weight value. When the weight returns to zero (<10d), the hold mode is cancelled automatically.
= “Animal scale hold” mode
When the scale returns back to zero, the display shows “ - - - - - - ”. After the animal or object is on the platter and the scale becomes stable, the display shows the weight value and hold. Then whe n the animal or object is off the platter, the display shows “ - - - - - - (or the weight is less than 10e) and the hold function is off. If the scale is hardly stable when the animal is on
the platter, the scale shows the average weight in 10 seconds and holds the status.
EXCELL PRECISION CO., LTD
02022010 ZSME300000197
18
Use
MC
or
UNIT
key to key
in the parameters
Use
MC
or
UNIT
key to key
in the parameters
Use
MC
or
UNIT
key to key
in the parameters
Hold Function Setting (Animal scale)
%
%
MC
key Upward key (0~9 digit entry)
UNIT
key Downward key (0,9~1 digit entry)
TARE/PT
key Move cursor rightward
NET/GROSS
key Move cursor leftward
Press
TARE/PT
key
Press
TARE/PT
key
Press
TARE/PT
key
Press
TARE/PT
key
Hold function setting
Use
MC
or
UNIT
keys to key in 4 and
press
TARE/PT
key to enter the hold function
setting mode.
Hold function setting
LCD displays the last parameter setting
Error range setting
Use
MC
or
UNIT
keys to key in the
parameters ( Among 1%~100%) Default setting = 10%
Counts averaging setting
Use
MC
or
UNIT
keys to key in the
parameters (1,2,4,8,16,32,64)
Default setting = 8
EXCELL PRECISION CO., LTD
02022010 ZSME300000197
19
Press
TARE/PT
key
Use
MC
or
UNIT
key to key
in the parameters
Press
TARE/PT
key
Use
MC
or
UNIT
key to enter
the parameter.
2-1-7 Auto Unit Weight Averaging Setting
2-1-8 Record Last Zero
This function is not available for approval
models.
MC
key Upward key (0~9 digit entry)
UNIT
key Downward key (0,9~1 digit entry)
TARE/PT
key Move cursor rightward
NET/GROSS
key Move cursor leftward
MC
key Upward key (0~9 digit entry)
UNIT
key Downward key (0,9~1 digit entry)
TARE/PT
key Move cursor rightward
NET/GROSS
key Move cursor leftward
Auto unit weight averaging setting
Use
MC
or
UNIT
keys to key in the
parameters
Default setting = 0 0  Auto unit weight averaging
1  Non auto unit weight averaging
Auto unit weight averaging setting
LCD displays the last parameter setting
Press
TARE/PT
key
Record Last Zero
Use
MC
or
UNIT
Key to select the setting.
Default setting =
Record the last zero setting
No record
LCD displays the last parameter setting Press
TARE/PT
key
EXCELL PRECISION CO., LTD
02022010 ZSME300000197
20
When the right digit is flashing, press
TARE/PT
key while
there is no load on platter
kg
Put the calibrated weight on platter and press
TARE/PT
key; After 3 beep sounds, calibration completes; Remove weight
kg
Press
NET/GROSS
key
to skip
Press
NET/GROSS
key to move the flashing digit to the left.
About 2 seconds later
Press
NET/GROSS
key to exit
About 2 seconds later
2-2 External Weight Calibration
The conditions for external weight calibration: The calibration weight value placed on the platter must be over 100e, and the standard
deviation of the weight must be within 10%.
This function is not available for approval models.
.
.
.
ON/ZERO
key Upward key (0~9 digit entry)
UNIT
key Downward key (0,9~1 digit entry)
TARE/PT
key Move cursor rightward
NET/GROSS
key Move cursor leftward
Press
TARE/PT
key, the display shows the calibration value. The right digit keeps flashing.
The calibration value could
be changed
kg
Reading Zero
kg
EXCELL PRECISION CO., LTD
02022010 ZSME300000197
21
Appendix 1 ASCII Code Table
Symbol
ASC II
Code
Symbol
ASC II
Code
Symbol
ASC II
Code
A
41H
a
61H
0
30H
B
42H
b
62H
1
31H
C
43H
c
63H
2
32H
D
44H
d
64H
3
33H
E
45H
e
65H
4
34H
F
46H
f
66H
5
35H
G
47H
g
67H
6
36H
H
48H
h
68H
7
37H
I
49H
i
69H
8
38H
J
4AH
j
6AH
9
39H
K
4BH
k
6BH
0DH
L
4CH
l
6CH M 4DH
m
6DH
N
4EH
n
6EH
O
4FH
o
6FH P 50H
p
70H
Q
51H
q
71H
R
52H
r
72H
S
53H
s
73H
T
54H
t
74H
U
55H
u
75H
V
56H
v
76H W 57H
w
77H X 58H
x
78H Y 59H
y
79H Z 5AH
z
7AH
EXCELL PRECISION CO., LTD
02022010 ZSME300000197
22
Appendix 2 Wiring Instructions
Load cell wiring method:
(1) As shown below, when a load cell is connected with a 4PIN cable, SEN+
and SEN- can be unconnected. J11 and J12 on PCB must be tin-soldered in short circuit.
(2) If Load Cell is 6PIN, please wire as the following diagram:
EXCELL PRECISION CO., LTD
02022010 ZSME300000197
23
Appendix 3 7-Segment Display Characters
Digit
7 segment letter
Alphabet
7 segment letter
Alphabet
7 segment letter
0 A N
1 B O
2 C P
3 D Q
4 E R
5 F S
6 G T
7 H U
8 I V
9 J W
K X
L Y
M Z