The information contained herein is the property of Excell Precision Limited and is supplied without liability for errors or omissions. No part may be reproduced or used except as authorised by contract or other written permission. The copyright and the foregoing restriction on reproduction and use extend to all media in which the information may be embodied.
EXCELL PRECISION CO., LTD.
FM User Manual SME300000119
1
TABLE OF CONTENT
SAFETY PRECAUTION.......................................................................3
<CHAPTER 1> FEATURES AND SPECIFICATION..........................4
1-1 FEATURES............................................................................................................................4
1-2 SPECIFICATION.....................................................................................................................5
<CHAPTER 2> KEYPAD OPERATION INSTRUCTION....................6
<CHAPTER 3> FRONT & REAR PANELS........................................7
3-1 FRONT PANEL......................................................................................................................7
3-2 REAR PANEL ........................................................................................................................8
<CHAPTER 4> INSTALLATION........................................................9
4-1 LOAD CELL..........................................................................................................................9
4-2 DIMENSION.........................................................................................................................10
4-3 BATTERY ASSEMBLY ..........................................................................................................11
<CHAPTER 5> INTERNAL CALIBRATION....................................12
5-1 SPECIFICATION SETTING ......................................................................13
2 Specification Parameter Description .........................................................................................................14
2 Parameter Display Description ...................................................................................................................15
5-2 INTERNAL WEIGHT CALIBRATION ..........................................................16
2 Password Setting..........................................................................................................................................17
5-3 INTERNAL FUNCTION SETTING ..............................................................18
2 Internal Function Parameter Description..................................................................................................19
2 Error Message................................................................................................................................................20
<CHAPTER 6> SPECIAL FUNCTION.............................................21
6-1 ANIMAL SCALE FUNCTION SETTING ...................................................................................21
6-2 DUAL RANGE RESOLUTION SWITCH FUNCTION...................................................................22
6-3 PRE-TARE FUNCTION ........................................................................................................23
6-4 RESOLUTION SWITCH FUNCTION........................................................................................23
EXCELL PRECISION CO., LTD.
FM User Manual SME300000119
2
<CHAPTER 7> TRANSMISSION INTERFACE...............................24
7-1 OP-01 RS232 / RS485 SERIAL OUTPUT (WITH RTC)........................................................24
2 Pin Allocation of Rear Panel........................................................................................................................24
2 Connection Description (RS485) ................................................................................................................25
<CHAPTER 8> FUNCTION PARAMETER SETTING......................26
8-1 FUNCTION SETTING
...........................................................................27
2 Function Parameter Setting.........................................................................................................................28
8-2 RS232 SETTING
...............................................................................29
2 OP-01 RS232 / RS485 Interface Function..................................................................................................30
2 Transmission Format....................................................................................................................................32
2 Command Mode .............................................................................................................................................33
<CHAPTER 9> MAINTENANCE.....................................................34
9-1 DEFAULT RECOVERY FOR ALL PARAMETERS........................................................................34
9-2 DEFAULT RECOVERY FOR GENERAL FUNCTION PARAMETERS ...............................................34
9-3 SELF-DIAGNOSIS MODE.......................................................................................................34
9-3-1 Program Version Number.........................................................................................................................35
9-3-2 7-segment Display Testing.......................................................................................................................35
9-3-3 Keypad & Calibration Switch Testing.....................................................................................................35
9-3-4 AD Conversion Value................................................................................................................................35
9-3-5 EEPROM Testing........................................................................................................................................35
9-3-6 RTC Time & Date Testing..........................................................................................................................35
9-3-7 RS-232 Serial Output Interface Testing ( OP-01 ).................................................................................35
EXCELL PRECISION CO., LTD.
FM User Manual SME300000119
3
Safety Precaution
& Turn off power before installing or disassembling. & Keep the product away from sunshine. The temperature range for operation
is -10 ~ +40℃℃.
& To connect the ground is a must for this equipment. The ground impedance
is less than 100.
& Never connect the ground with other equipments which are huge in power
consumption.
& No ground or incorrect ground connecting might cause the electric shocks or
breakdowns.
BEFORE USING THE INDICATOR
Thanks for purchasing EXCELL FM Mini-indicator. In order to operate smoothly, to last the durability, and to reduce the chance of breakdown for this product, please read the following instructions carefully.
EXCELL PRECISION CO., LTD.
FM User Manual SME300000119
4
<Chapter 1> Features and Specification
1-1 Features
l Large and clear LCD (height 25.4mm × 10mm) l LED Backlight, more duration, more energy saving l Auto zero tracking l Weighing calibration in both kilogram and pound l High resolution switch function to display 10 times divisions l Weighing unit shift function & data accumulation function l Adjustable digital filter l Available in connecting with up to four 350-ohm load cells l Capable of connecting with 6-wire load cells
l Convenient to operate power switch in front panel
EXCELL PRECISION CO., LTD.
FM User Manual SME300000119
5
1-2 Specification
Analog Specification
♦ Load Cell Current: DC 5V ±5% 60mA (Up to Four 350Ω Load Cells) ♦ Max. Load Cell Input Voltage: 16 mV ♦ Input Sensitivity: 0.12µV/D or more ♦ Conversion Rate: Approximately 100 times/sec. (max.) ♦ Resolution: 19 bits
Digital Specification
♦ Display: LCD, 6 digits, height 25.4x10mm, LED backlight ♦ Display Frequency: 50 times/sec. (max.) ♦ Display Range: - 999999 ~ 999999 ♦ Min. Division: 1, 2, 5, 10, 20, 50 ♦ Decimal Point: 0, 0.0, 0.00, 0.000, 0.0000 ♦ Memory: Calibration parameter and function setting are all stored in EEPROM.
Optional Interface
♦ OP-01 RS-232 / RS-485 (Includes RTC Function)
Power Requirement
♦ Adaptor Spec.: Input 120/230VAC 50~60Hz, output 9V/1000mA 6V/5pcs Hi-MH rechargeable battery kit (1800mA/1hr) Max. Power Consumption (by rechargeable battery):
120mA (with 4 Load Cells + backlight + RS-232 interface) 15 hours
50mA (with 1 Load Cell + no backlight + no RS-232 interface) 36 hours
Others
Operation Temperature: -10 ~ 40℃℃ Operation Humidity: < 85% R.H. Dimension: W 49.5 × L193 × H134 (mm)
Weight: 700 (g)
EXCELL PRECISION CO., LTD.
FM User Manual SME300000119
6
F1
<Chapter 2> Keypad Operation Instruction
Function Operation Description
General Function
Setting
Press and hold
,
and then press
Refer to <Chapter 8> Function
Parameter Setting for details
Weighing
Parameter Setting
Adjust calibration
switch to ON
Setting for decimal point, capacity,
division, zero tracking, and unstable
detecting, etc. Refer to 5-1
Specification Setting for details.
Calibration
Adjust calibration
switch to ON
Refer to 5-2 Internal Weight
Calibration for operation.
Self-diagnosis
Mode
While turning on with countdown, press and hold .
Refer to 9-3 Self-diagnosis Mode
for details.
Default Recover
for All Parameters
Adjust calibration switch to ON, and then press and hold .
Refer to 9-1 for details.
Default Recovery
for General
Function
Parameters
While turning on with countdown, press and hold .
Refer to 9-2 for details.
4 During the operation, use the following keys to complete all the
works.
To add the value flashing To move the cursor rightward
To reduce the value flashing Storage setting
To move the cursor leftward To abort setting/to escape
GROSS
NET
3
ESC
ENTER
ZERO
F2
4
TARE
ENTER
F1
ENTER
F1
ENTER
ZERO
EXCELL PRECISION CO., LTD.
FM User Manual SME300000119
7
ESC
ZEROTARE
F2
+
ENTER
F1
R
GROSS
NET
+-
TAREMOTIONM+GROSS PT
<Chapter 3> Front & Rear Panels
3-1 Front Panel
Indication:
: Battery charged status : Battery charging status TARE : Tare status MOTION : Unstable weighing indication M+ : Accumulation status indication GROSS : Gross weight PT : Pre-tare
Keypad:
1) Power ON / OFF. Press and hold this key for 3 seconds to shut down.
2) To abort or escape when setting.
1) Weight re-zero.
2) To add the value when setting.
1) To eliminate the gross weight.
2) To reduce the value when setting.
1) To switch Gross / Net weight shown on display.
2) To move the cursor leftward when setting.
1) Keypad function (FNC-02 & FNC-03)
2) To move the cursor rightward when setting. Keypad function (FNC-02 & FNC-03)
Confirmation key.
ESC
ZERO
TARE
ENTER
F1
F2
4
GROSS
NET
3
EXCELL PRECISION CO., LTD.
FM User Manual SME300000119
8
3-2 Rear Panel
1. Battery Case
2. RS232/485 Input/Output
3. DC 9V Power Input
4. Calibration Switch
5. Load Cell Connect Socket
5
1
2
3
4
EXCELL PRECISION CO., LTD.
FM User Manual SME300000119
9
12345
6789
Load cell cable
Shield
Load cell
EXC+ SEN+
SIG+SIG-
EXC-
SEN-
<Chapter 4> Installation
4-1 Load Cell
4-wired (5-wired) Load Cell
Short Pin4&5 to connect with EXC+ Short Pin2&3 to connect with EXC­ Pin1 connects with SIG+ Pin9 connects with SIG­ Pin6, 7, 8 connect with Shield
6-wired (7-wired) Load Cell
Pin5 connects with EXC+ Pin4 connects with SEN+ Pin3 connects with EXC­ Pin2 connects with SEN­ Pin1 connects with SIG+ Pin9 connects with SIG­ Pin6, 7, 8 connect with Shield
EXCELL PRECISION CO., LTD.
FM User Manual SME300000119
10
4-2 Dimension
EXCELL PRECISION CO., LTD.
FM User Manual SME300000119
11
餘的電塞入電池孔中
Stuff the surplus wires into the battery case.
4-3 Battery Assembly
EXCELL PRECISION CO., LTD.
FM User Manual SME300000119
12
<Chapter 5> Internal Calibration
Adjust calibration switch to ON, and the screen displays:
After 2 seconds
Press Press
Press Press
Adjust calibration switch back to OFF.
⇒ Specification Setting
⇒ Internal Weight Calibration
⇒ Internal Function Setting
ZERO
TARE
ZERO
TARE
ENTER
ENTER
ENTER
EXCELL PRECISION CO., LTD.
FM User Manual SME300000119
13
5-1 Specification Setting
Press
Input the *Parameter Code intended
Press
Continue other functions setting or press to escape.
The screen displays the parameters set previously. Input the parameters intended and then press .
*Parameter Code
Decimal Point
Capacity
Division 1
Division 2
Zero Tracking Setting
Unstable Detecting Setting
ESC
ENTER
ENTER
ZERO
TARE
GROSS
NET
3
F2
4
ENTER
ESC
⇒ To add the value flashing
⇒ To reduce the value flashing
⇒ To move the cursor leftward
⇒ To move the cursor rightward
⇒ Storage setting
⇒ To abort setting / escape
ENTER
EXCELL PRECISION CO., LTD.
FM User Manual SME300000119
14
2 Specification Parameter Description
Setting
Parameter
Code
Function
Parameter
Description
Default Setting
CSP-00
Decimal
Point
Refer to the description on next page.
CSP-01 Capacity
999999
000000
Max. value for weight display 999999
1 2
5 10 20
CSP-02 Division 1
50
Min. value for weight display 1
1
2
5 10 20
CSP-03 Division 2
50
Min. value for Weight display 1
CSP-04
Zero
Tracking
Setting
Refer to the description on next page.
CSP-05
Unstable
Detecting
Setting
Refer to the description on next page.
EXCELL PRECISION CO., LTD.
FM User Manual SME300000119
15
2 Parameter Display Description
CSP-00 Decimal Point
Display Decimal Point Digit
None
.
1 Digit
.
2 Digits
.
3 Digits
.
4 Digits
CSP-04 Zero Tracking Setting
Display Division/Period
.
0.25 D/1 sec
.
0.5 D/1 sec
.
0.75 D/1 sec 1D/1 sec
.
1.25 D/2 sec
.
1.5 D/2 sec
.
1.75 D/2 sec
2 D/2 sec
No Zero Tracking
CSP-05 Unstable Detecting Setting
Display Division / Period
.
0.25 D/1 sec
.
0.5 D/1 sec
.
0.75 D/1 sec
1D/1 sec
.
1.25 D/2 sec
.
1.5 D/2 sec
.
1.75 D/2 sec
2 D/2 sec
No Unstable Detecting
EXCELL PRECISION CO., LTD.
FM User Manual SME300000119
16
5-2 Internal Weight Calibration
Turn on and warm up the machine for 15 to 30 minutes before calibration. Adjust calibration switch to ON, and the screen will display . Press or to select .
Procedure
Zero
Calibration
Weight Calibration
Zero Calibration
a ) Ensure nothing on the platter; after being stable, press , and the screen will display
……. . The zero calibration will be complete 5 seconds later.
b ) To abort zero calibration, just press .
Weight Calibration
a ) Place an object, whose weight is known, on the platter, and input the weight value from
front panel. After being stable, press and the screen will display ……. . The weight calibration will be complete 5 seconds later.
b ) To abort weight calibration, just press .
Zero calibration complete
Confirmed
Calibration
Switch
OFF
Place the
weight mass
and set the
weight
ENTER
ENTER
ENTER
ENTER
ESC
ENTER
ENTER
ESC
TARE
ZERO
ESC
ESC
EXCELL PRECISION CO., LTD.
FM User Manual SME300000119
17
2 Password Setting
Adjust calibration switch to ON After 2 seconds
4 After complete password setting, when entering calibration mode or function
setting mode, the screen displays
for 1 second, and then
.
Its necessary to input the correct password to continue each setting.
If the password inputted is not correct, the screen displays
.
Input new password
4 “0000 means no password.
Confirm new password (re-input again)
ENTER
ENTER
F1
ZERO
EXCELL PRECISION CO., LTD.
FM User Manual SME300000119
18
⇒ To add the value flashing
⇒ To reduce the value flashing
⇒ To move the cursor leftward
⇒ To move the cursor rightward
⇒ Storage setting
⇒ To abort setting / escape
5-3 Internal Function Setting
Press
Input the *Parameter code intended
Press
The screen displays the parameters set previously. Input the parameters intended and then press
Continue other functions setting or press to escape.
ESC
ENTER
ENTER
ENTER
*Parameter Code
⇒ Tare or Zero Function under
Unstable Status
⇒ Turning on Re-zero
⇒ Re-zero Range
⇒ Filter Strength
⇒ AD Sampling Rate
⇒ Animal Scale Mode
⇒ Animal Scale Stable Range
⇒ Animal Scale Sampling
Frequency
⇒ Dual Range Resolution
Setting
⇒ Dual Range Resolution
Middle Point Setting
ZERO
TARE
GROSS
NET
3
F2
4
ENTER
ESC
EXCELL PRECISION CO., LTD.
FM User Manual SME300000119
19
2 Internal Function Parameter Description
Setting
Parameter
Code
Function
Parameter
Description
Default Setting
0 ON
CFN-00
Tare or Zero
Function under
Unstable Status
1 OFF
0
0 OFF
CFN-01
Re-zeroing after
Turning on
1 ON
0
CFN-02 Re-zero Range 0% ~ 30%
0%: Full range re-zero 1% ~ 30%: Capacity × ± setting value%
0
CFN-03 Filter Strength 0 ~ 5 Strength increases by number 2
0 No limit 1 20 times/sec. 2 10 times/sec.
CFN-04
AD Sampling
Rate
3 5 times/sec.
2
0 OFF 1
Mode 1: No weight display under unstable status
CFN-05
Animal Scale
Mode
2
Mode 2: Weight display whether under stable or unstable status
0
CFN-06
Animal Scale
Stable Range
0 ~ 100 Mode 2: Stable Range Setting 30
0 8 times 1 16 times 2 32 times 3 64 times
CFN-07
Animal Scale
Sampling
Frequency
4 128 times
2
0 Multi-interval
CFN-08
Dual Range
Resolution
Setting
1 Multi-range
0
CFN-09
Dual Range
Resolution
Middle Point
Setting
0 ~ 65535 5000
EXCELL PRECISION CO., LTD.
FM User Manual SME300000119
20
2 Error Message
(1) Load Cell or A/D circuit extraordinary
(2) Real weighing value zero value
(3) Internal resolution 0.12μV/D range
(4) . Incorrect password
(5) Turning on zero ≦zero range
(6) Turning on zero ≧zero range
EXCELL PRECISION CO., LTD.
FM User Manual SME300000119
21
32 Times
30d
10d
WT
T
32
量保持
取消重保持
<Chapter 6> Special Function
6-1 Animal Scale Function Setting
2 CFN-05 = 1
(Animal Scale Mode1: No weight display under unstable status)
When no any object is on the platter, the screen will display:
When the object is on the platter, after weight has been measured, the screen will display:
If the display weight value keeps being lower than zero plus 10d, the screen will display:
2 CFN-05 = 2
(Animal Scale Mode 2: Weight display weather under unstable or stable status)
When the weight value reaches the range of CFN-06 and CFN-07 setting, the screen will keep displaying the weight value.
When the weight value is over the range of CFN-06 and CFN-07 setting, the screen will display the normal weight measurement.
Example: CFN-06=30 CFN-07=2
.
Weight Holding
Weight Holding Cancellation
EXCELL PRECISION CO., LTD.
FM User Manual SME300000119
22
最小刻度1
1
2
最小刻度2
WT
5000
0
1
零點恢復
2
2
2
WT
5000
0
6-2 Dual Range Resolution Switch Function
If the setting of CSP-02 is not the same as CSP-03, the dual range resolution will be activated.
2 CFN-08 = 0 Multi-interval
If CFN-09 = 5000
2 CFN-08 = 1 Multi-range
If CFN-09 = 5000
Division 2
Division 1
Recover when back to zero
EXCELL PRECISION CO., LTD.
FM User Manual SME300000119
23
6-3 Pre-tare Function
2 FNC-02 or FNC-03 setting is at parameter 6. (Pre-tare Function)
Under weight display status, press or key (according to FNC setting), the screen will display:
Input the pre-tare value intended
Press key
Back to weight display status (PT indication lights up)
Pre-tare Cancellation
When the gross weight is displayed as 0, press key to cancel the pre-tare value.
6-4 Resolution Switch Function
2 FNC-02 or FNC-03 setting is at parameter 5. (HR)
Under weight display status, press or key (according to FNC setting), the screen will display 10 times resolution, and then, recovery back to original after 5 seconds.
ENTER
F1
F2
4
F1
F2
4
TARE
EXCELL PRECISION CO., LTD.
FM User Manual SME300000119
24
<Chapter 7> Transmission Interface
7-1 OP-01 RS232/RS485 Serial Output (with RTC)
To short 1 and 2 pins is RS485 output. To short 2 and 3 pins is RS232 output.
2 Pin Allocation of Rear Panel
Pin Function
2 RXD 3
TXD 5 SG 6 DA 7 DB
4 RS485 interface is capable to connect up to 10 mini-indicators.
1
23
4
5
7689
CN3 CN
2
1 2 3 1 2 3
EXCELL PRECISION CO., LTD.
FM User Manual SME300000119
25
電腦主機
EX0419 # 1 RS - 06 = 01
EX0419 # 2 RS - 06 = 02
EX0419 # 10 RS - 06 = 10
電阻
2 Connection Description (RS485)
4 Notice
If the terminal resistant is built-in the host interface, it s not necessary to connect with
another one outside.
If the host computer is no signal ground (SG), its not necessary to connect with it.
終端電阻
Terminal Resistant
Host
EXCELL PRECISION CO., LTD.
FM User Manual SME300000119
26
<Chapter 8> Function Parameter Setting
Under general weight display status, press , and the screen will display:
Press key Press key
Press to escape. Back to general weight display status.
⇒ External function setting
⇒ RS232/RS485 interface function
ZERO
TARE
ENTER
ENTER
ESC
F1
ENTER
EXCELL PRECISION CO., LTD.
FM User Manual SME300000119
27
8-1 Function Setting
Press
Input the *parameter code intended
Press
Continue other function settings
or press to escape
*Parameter Code
Key Disable
DSP Update
F1 Key Function Setting
F2 Key Function Setting
Backlight Setting
ENTER
ENTER
The screen displays the parameters set previously. Input the parameters intended and then press .
ENTER
ZERO
TARE
GROSS
NET
3
F2
4
ENTER
ESC
⇒ To add the value flashing
⇒ To reduce the value flashing
⇒ To move the cursor leftward
⇒ To move the cursor rightward
⇒ Storage setting
⇒ To abort setting or escape
ESC
EXCELL PRECISION CO., LTD.
FM User Manual SME300000119
28
2 Function Parameter Setting
Setting
Parameter
Code
Function
Parameter
Description
Default
Setting
0 ON
FNC-00
Key Disable
0000
1111
1 OFF
0000 is corresponding to:
(from left to right)
0000
0 No Limit 1 20 times/sec. 2 10 times/sec. 3 5 times/sec.
FNC-01
DSP Update
4 1 times/sec.
1
0
Print (printing)
1 Units (units switch) 2 M+ (accumulation and printing) 3 MC (memory clearing)
4
Weight/Weight Accumulation/Times
Accumulation Display Switch
5 HR (high resolution switch)
FNC-02
F1 Key
Function
Setting
6 Pre-tare (pre-tare function)
0
0
Print (printing)
1 Units (units switch) 2 M+ (accumulation and printing) 3 MC (memory clearing)
4
Weight/Weight Accumulation/Times
Accumulation Display Switch
5 HR (high resolution switch)
FNC-03
F2 Key
Function
Setting
6 Pre-tare (pre-tare function)
1
0
Auto Backlight On
(backlight on in operation only)
1 Backlight On (backlight always on)
FNC-04
Backlight
Setting
2 Backlight Off
0
ZERO
TARE
GROSS
NET
3
F2
4
EXCELL PRECISION CO., LTD.
FM User Manual SME300000119
29
⇒ To add the value flashing
⇒ To reduce the value flashing
⇒ To move the cursor leftward
⇒ To move the cursor rightward
⇒ Storage setting
⇒ To abort setting / escape
8-2 RS232 Setting
Press
Input the *Parameter code intended
Press
The screen displays the parameters set previously. Input the parameters intended and then press
Continue other functions setting
or press to escape.
*Parameter Code
Information Pattern
Transmission Method
Transmission Rate
Parity, Bit Length, Stop Bit
Unstable or Over Load
Auto Transmission
Condition
Command Address
Output Format
Transmission Times
Date Setting
Time Setting
ESC
ENTER
ENTER
ENTER
ZERO
TARE
GROSS
NET
3
F2
4
ENTER
ESC
EXCELL PRECISION CO., LTD.
FM User Manual SME300000119
30
2 OP-01 RS232 / RS485 Interface Function
Setting
Parameter
Code
Function
Parameter
Description
Default Setting
0 Display Correspondingly 1 Gross Weight 2 Net Weight 3 Tare 4 Weight Accumulation 5 Times Accumulation
RS1-00
Information
Pattern
6 Output with Date & Time
0
0 Continuous Transmission 1 Auto Transmission
2 Press or to transmit 3 Command Mode (no address)
RS1-01
Transmission
Method
4 Command Mode (with address)
0
0 1200 1 2400 2 4800 3 9600
RS1-02
Transmission
Rate
4 19200
1
0 N, 8, 1
No Parity
8 Bits Length
1 Stop Bit
1 O, 7, 1
Odd Parity,
7 Bits Length,
1 Stop Bit
RS1-03
Parity
Bit Length
Stop Bit
2 E, 7, 1
Even Parity,
7 Bits Length,
1 Stop Bit
2
0 Continuous Output
RS1-04
Unstable or
Over Load
1 Stop Output
0
0 Positive (over + 10D)
RS1-05
Auto
Transmission
Condition
1
Positive/negative
(over + 10D, under - 10D)
0
RS1-06
Command
Address
00
99
Available only if RS1-
01 setting is 4
0
0 Standard Format
RS1-07
Output
Format
1
UMC 600
0
F2
4
F1
EXCELL PRECISION CO., LTD.
FM User Manual SME300000119
31
0 No Limit 1 1 times/sec. 2 2 times/sec 3 5 times/sec 4 10 times/sec
RS1-08
Transmission
Times
5 20 times/sec
4
RS1-09
Date Setting
RS1-10
Time Setting
EXCELL PRECISION CO., LTD.
FM User Manual SME300000119
32
2 Transmission Format
RS1-00 0 ~ 3
S T
,
G S
,
+ 1 2 3 4
.
5 6 g CR LF
Header 1 Header 2 Weight Data ( 8 digits ) Unit Terminators
Header 1
ST : Stable Weight / US : Unstable Weight / OL : Weight Over Load
Header 2
GS : Gross Weight / NT : Net Weight / TR : Tare
Weight Data ( 8 digits )
The first digit of weight data represents +/- indication for weight value. The other 7 digits, including decimal point, represent the weight value. If the weight is over load (Header 1 : OL), the screen turns into blank except +/- indication and decimal point.
Unit
Kg, g, t, lb or blank
Terminators
CL and LF are data termination code.
RS1-00 = 4
T N , 1 2 3 CR LF
RS1-00 = 5
RS1-00 = 6
T W , + 1 2 3 4 . 5 6 k g CR LF
D A T E
: 2 0 X X / X X / X X CR LF
T
I M E : X X : X X : X X CR LF
G R O S S : + 1 2 3 4 . 5 6 k g CR LF
N E T
: + 1 2 3 4 . 5 6 k g CR LF
T A R E
: + 1 2 3 4 . 5 6 k g CR LF
T N
: X X X CR LF
T W : + 1 2 3 4 . 5 6 k g CR LF
EXCELL PRECISION CO., LTD.
FM User Manual SME300000119
33
2 Command Mode
Command
Function Command
Function
READ, RW
Weight Reading CT Tare Clearing
ZERO, MZ
Weight Re-zeroing Rl Weight Accumulation
TARE, MT Gross Weight Reducing
Rm Times Accumulation
NTGS Gross / Net Switch Rn Date
MG Gross Weight Indicating
Ro Time
MN Net Weight Indicating AT
Weight and Times
Accumulation
DT
Weight and Times
Accumulation Clearing
Œ After setting the commands mentioned above, its necessary to add the termination
code CR(0DH) and LF(0AH).
If the command is not correct, it will reply E + Command Unidentified. Ž If setting command mode with address (RS1-06 = 4), add @ address in front of each
command.
Example: When RS1-08 = 1, for reading weight value, the whole complete command
should be @01RW(CR)(LF).
EXCELL PRECISION CO., LTD.
FM User Manual SME300000119
34
<Chapter 9> Maintenance
9-1 Default Recovery for All Parameters
(1) Adjust the calibration switch to ON, when re-zeroing after turning on, press
and hold simultaneously.
(2) The screen will display . .
(3) If decided, press and hold until displaying , and then adjust the
calibration switch to OFF.
9-2 Default Recovery for General Function Parameters
(1) When re-zeroing after turning on, press and hold simultaneously.
(2) The screen will display . . (3) If decided, press and hold until re-turning on.
9-3 Self-diagnosis Mode
(1) When re-zeroing after turning on, press and hold.
(2) The screen will display
, which means entered self-diagnosis mode.
(3) Use or keys to select item intended to test.
Press key to enter self-diagnosis, and press key to escape.
Item Display Testing Item
1
Program Version Number Displaying
2
7-segment Display Testing
3
Keypad and Calibration Switch Testing
4
AD Conversion Value Displaying
5
EEPROM Testing
6
RTC Date & Time Testing
7
OP-1 RS232 Serial Output Interface Testing
F1
ENTER
ENTER
F1
ENTER
ENTER
ZERO
ZERO
TARE
ENTER
ESC
EXCELL PRECISION CO., LTD.
FM User Manual SME300000119
35
9-3-1 Program Version Number
7-segment display reveals program version number.
9-3-2 7-segment Display Testing
7-segment display reveals ~ ” and ..
9-3-3 Keypad & Calibration Switch Testing
Adjust calibration switch to ON, and press any key, the corresponding bit will be
changed from .
9-3-4 AD Conversion Value
7-segment display reveals the internal value of the present weight.
9-3-5 EEPROM Testing
Displaying represents in normal condition.
Displaying represents in extraordinary condition.
9-3-6 RTC Time & Date Testing
Press key to enter the testing mode, and the screen will display date XX.XX.XX.
Example: 05.11.03 represents 3rd of November, 2005.
Press key again to display time XX.XX.XX.
Example: 09.45.50 represents 9 oclock, 45 minute and 50 seconds.
9-3-7 RS-232 Serial Output Interface Testing ( OP-01 )
(1) Short the 2nd pin and 3rd pin of the SER. OUT. D-SUB 9pin socket.
Displaying represents in normal condition.
Displaying represents in breakdown condition.
(2) If connected with a computer (protocol must be corresponding), the screen will
display ~ , which means RS-232 output is in normal condition.
ENTER
ENTER