Excell ESW-E User Manual

User Manual
Waterproof Weighing Scale
ESW-E
1
EXCELL PRECISION CO., LTD.
ZSME300000276
02005324
Content of Table
Instruction for Use................................................................................................................ 2
Preparing to Use the Scale..................................................................................................2
Chapter 1 Introduction .........................................................................................................4
1-1 Features.................................................................................................................4
1-2 Specifications .........................................................................................................4
1-3 Power Supply .........................................................................................................5
1-4 Display and Keyboard Introduction.........................................................................5
1-5 Self-Test Mode........................................................................................................6
1-6 Error Messages......................................................................................................7
1-7 Weight Unit.............................................................................................................7
Chapter 2 External Function Setting....................................................................................8
2-1 General Function Setting ..............................................................9
2-1-1 Power Saving Function Setting .........................................10
2-1-2 Automatic Power-off Timer Setting ....................................11
2-1-3 Hi/Lo/OK Function Setting .................................................12
2-1-4 Restore to the Default Setting ...........................................13
2-1-5 Noise Filter Setting ............................................................14
2-2 Weight Calibration
...........................................................................15
Appendix1 7-Segment Display Characters........................................................................16
2
EXCELL PRECISION CO., LTD.
ZSME300000276
02005324
Thank for your purchasing of our EXCELL Weighing Scale. To guide you to use our product correctly, please read this Manual carefully to extend the life of machine and to avoid error.
Instruction for Use
1) Please keep the scale in a cool dry place. Do not store it at high temperature.
2) Avoid objects impacting with the scale. Do not drop loads onto the scale or subject the weigh pan to any strong shock loads.
3) The load placed on the weigh pan must not exceed the maximum weighing capacity
of the scale.
4) If the scale is not going to be used for some time, please clean it and store it in a plastic bag in dry conditions. A desiccant sachet may be included to prevent any moisture build up.
Preparing to Use the Scale
1. Adjust the four levelling feet (if fitted) to set the scale pan level using the spirit level bubble located at the front of the scale.
2. Avoid operating the scale in direct sunlight or drafts of any kind.
3. If possible avoid connecting the scale to ac power outlet sockets which are adjacent to other appliances to minimise the possibility of interference affecting the performance of the scale.
4. Remove any weight that might be on the weigh pan before the scale is switched on and avoid leaving weight on the pan for long periods of time
5. All goods weighed should be placed in the centre of the weigh pan for accurate weighing. The overall dimensions of the goods being weighed should not exceed the dimension of the weigh pan.
6. We suggest to warm up the scale for 15~20 minutes before operation to ensure best accuracy.
7. Please note when the symbol keeps flashing on the screen, the batteries need to be replaced.
8. Introduction of Storage Battery
3
EXCELL PRECISION CO., LTD.
ZSME300000276
02005324
No burning
Caution Corrosion
Due to the storage battery adopt the advanced free-maintaining technique, customers need not to replenish electrolyte. The scale should be recharged every 3 months to prevent failure of the internal rechargeable battery.
1. The battery should be charged for 8~10 hours.
2. The temperature of battery should below 45℃.
Maintaining
1. Please do not discharge with ov er-current when using the battery. Please charge the battery after discharging current.
2. Please take down the battery when the scale is not used for a long time or break the connection of cathode.
3. Do not short the battery terminals to check whether there is current. Please check whether the connection point is firm to guarantee good connection.
4. The battery should be replaced by specialized person. No reverse-battery or the product will be damaged.
a) Anode of battery should be connected with Anode of product batteryusually red
cable
b) Cathode of battery should be connected with Cathode of product battery (usually
brown cable or black cable)
c) See the picture following
Safety warnings
1. The electrolyte of battery is caustic which causes metal, cotton, etc to corrode.
2. The hydrogen will be resolved when using or charging the battery and it will cause explosion when approaches fire.
Brown cable(or black cable) connected with Anode of battery
Red cable connected with Cathode of battery
Warning explosion
Children faraway
4
EXCELL PRECISION CO., LTD.
ZSME300000276
02005324
Chapter 1 Introduction
1-1 Features
Sealed waterproof silica gel strip blocks water from infiltrating into the scale. Surrounded by waterproof grade sheeting to ensure the water free. 1/3,000~1/6,000 display resolution available. Aluminum frames coated with anti-corrosion treatment for longer lifespan. Stainless steel made feet and platter for anti-rust. High speed of 24bits AD fast reacts and shortens the weighing operation duration. Selectable units: Kilogram (kg), gram (g), ounce (oz), and pound (lb) weighing units
available.
Built-in rechargeable battery can be easily replaced. Low power indication and auto power off. Well-designed protection point for transportation. Concave handle for easy carrying.
Options:
Double-sided display
1-2 Specifications
Model Capacity Division Resolution
ESW-3 3kg
0.5g 1g
1/6,000 1/3,000
ESW-6 6kg
1g 2g
1/6,000 1/3,000
ESW-15 15kg
2g 5g
1/7,500 1/3,000
ESW-30 30Kg
5g
10g
1/6,000
1/3,000
Operating Temperature: Non-approval model: 0°C ~ 40°C (32°F ~ 104°F) Approval model: -10°C ~ 40°C (14°F ~ 104°F)
Dimensions: 240 x 120 x 280 mm ( W x H x D) Weight of the scale: Around 3.5kg
240mm
280mm
120mm
5
EXCELL PRECISION CO., LTD.
ZSME300000276
02005324
1-3 Power Supply
Power Supply Selection
1. DC 6V / 4Ah rechargeable battery
2. AC 110V~240V (±10) adapter
Power Consumption
25 mA (system + L/C + front display;Power saving mode) 35 mA (system + L/C + front and back display;Power saving mode) 55 mA (system + L/C + front display)
70 mA (system + L/C + double display)
Low Battery Warning
Please note when the ( ) symbol keeps flashing on the display, the batteries should be charged.
The scale will turn off automatically after 1~2 hours, when the low battery warning
symbol shows up. The scale must be fully charged, before operating again.
Low battery shift voltage ≒5.3V DC
1-4 Display and Keyboard Introduction
Icon Introduction
Æ0Å Zero point indication
STABLE : stable indication
GROSS
gross weight indication
Low battery indication. When this symbol is flashing, the batteries should be charged.
L1,L2 : Units indication
HI : The weight on weigh pan is greater than the high limit
LO : The weight on weigh pan is lower than the check value
Keyboard Function
OFF: power off ON/ZERO: This key possesses two functions: Power On and Zero function.
n When the scale is off, press the ON/ZERO key, the scale will switch on.
o When the scale is on, the ON/ZERO key has zero function.
When the weigh pan is empty (free of load) and the display is not showing zero,
press the ON/ZERO key to zero the scale. At zero, the “Æ0Å” indication is on.
Units indication
L1 L2 Units
kg
g
lb
HI
LO
L2
ST ABLE
GROSS Æ0Å
L1
6
EXCELL PRECISION CO., LTD.
ZSME300000276
02005324
UNITS:
switching units
Press the UNITS key to switch weight units, the icons or arrows will indicate the active units as appropriate. The units available are dependent on the exact scale model.
After power off, the scale will memorize the active units. When the scale is powered
on again, it displays the previously active units.
TARE: tare key
The tare function will not operate during the following conditions: When the scale powers on, the weight is still below zero after a container is placed on the weigh pan. Or the tare value is over the full scale capacity.
Tare function
(1) Put a container on the weigh pan and after the weight is stable, press the TARE key
to zero the weight of the container. The screen displays the “Æ0Å” “STABLE” icon.
(2) Put the goods in the container, the screen displays the net weight value of the goods.
(3) Remove the full container; the screen displays the negative weight value of the
container. At this time pressing the TARE key again will cancel the tare and the scale reverts back to zero. The screen displays the “Æ0Å” “STABLE” icon.
The tare function can be operated continually to the full weighing capacity of the scale. Continual tare operation is adding or removing tare objects on weigh pan and
pressing the TARE key each time.
NET | GROSS: switch NET or GROSS key
NET/GROSS key is only used in Tare mode.
In the Tare mode, the screen displays the “Net” icon,
In tare mode, when the GROSS icon is on, the weight value on the display is the total
amount of the tare value. When the GROSS icon is off, the weight value on the display is the net value. press the NET/GROSS key to switch between the “Net value” and the “Gross value”.
At the Gross status, only OFF and NET/GROSS key s are functional.
1-5 Self-Test Mode
Set the switch SWA1 on the bottom of machine to the LOCK position.When the scale
count down, press
NET/GROSS
and
ON/ZERO
key together, Wait till display shows
to enter “Self-Test Mode”.
Press
UNITS
key
Press
ON/ZERO
key
Press
ON/ZERO
key
Press
ON/ZERO
key
Press
ON/ZERO
key
Press
UNITS
key
Press
UNITS
key
Press
UNITS
key
7
EXCELL PRECISION CO., LTD.
ZSME300000276
02005324
1-6 Error Messages
Unable to read the EEPROM
The EEPROM is not mounted yet, or the circuit near EEPROM is broken.
Initial zero is higher than the zero range when switching the scale on. (approval models)
Initial zero is lower than the zero range when switching the scale on. (approval models)
Internal value is below Zero.
A/D IC malfunction (cannot read A/D value). The load cell may not be connected to the scale correctly.
The weight of the object is over 9 divisions of the maximum capacity.
A/D IC value is over the maximum range.
Weight is lower than -1/6 full scale.
1-7 Weight Unit
(kg) 1 g = 0.001 kg (g) 1 g = 1 g (lb) 1 g = 0.002204623 lb (lb/oz) 1 g = 0.03527396 oz (oz) 1 g = 0.03527396 oz
INTERNAL VALUE MODE (
must hook up Load Cell to test)
c Press
T ARE
to enter, and the display shows internal value
d Please check whether the internal value has changed obviously with weight
changing.
e Please check the backlight. f Press
ON/ZERO
key to back to the last screen , the display shows
KEYPAD TEST MODE
c Press
TARE
to enter, display shows
Keypad’s internal code:
T ARE
key= 07
UNITS
key= 06
NET/GROSS
key=05
d Press
ON/ZERO
key to back to the last screen , the display shows
FIRMWARE VERSION DISPLAY MODE
c Press
T ARE
to enter , display shows the firmware version
d Press
T ARE
key again, the display shows maintenance number for 2 secs
e Press
ON/ZERO
key to back to the last screen, display shows
BACK TO THE LAST SCREEN
Press
T ARE
key to exit self-test mode, the scale will restart automatically.
8
EXCELL PRECISION CO., LTD.
ZSME300000276
02005324
Chapter 2 External Function Setting
In the weighing mode, press the NET/GROSS and ON/ZERO keys at the same time to
enter the External Function setting mode. The LCD shows .
Press
UNITS
key
Press
ON/ZERO
key
Press
UNITS
key
Press
UNITS
key
Press
UNITS
key
Press
ON/ZERO
key
Press
ON/ZERO
key
Press
ON/ZERO
key
General Function setting mode External Weight Calibration Function Reserved Exit the Advanced Function setting mode
Refer to the following sections for the detailed operation procedures of each function setting.
9
EXCELL PRECISION CO., LTD.
ZSME300000276
02005324
2-1 General Function Setting
There are 7 functions in the general function setting mode from
to .
Workflow of the General Function setting:
Press TARE key
Use the keys to select the function
Key in “00” and press
TARE key to return to
the Advanced function setting menu
ON/ZERO
key Upward key (from 0 to 9)
UNITS
key Downward key (from 9 to 0)
TARE
key Move the cursor to right
NET/GROSS
key Move the cursor to left.
ACTIONS
DISPLAY NOTE
Return to the Advanced Function Setting Mode Menu Power Saving Function Settings Automatic Power-off Timer Settings Hi/Lo/OK Settings Restore the Default Settings Noise Filter Settings Function Reserved Function Reserved
Refer to the following sections for detailed operation procedures of each setting.
10
EXCELL PRECISION CO., LTD.
ZSME300000276
02005324
2-1-1 Power Saving Function Setting
Select in the General Function setting mode to change power saving function setting.
The displays shows the last status Default setting = off
Power saving function
When the weight is over 10d, the display shows normal. If you press the key, the display still shows normal. When the scale returns to zero or you do not pressing any key, it will enter into power saving mode after 10 seconds.
Press TARE key
Use
ON/ZERO
or UNITS key to select function “on” or “off
Press TARE
key
ACTIONS
DISPLAY NOTE
Key in the parameter “00”
Press
TARE
key to complete the general setting
p
rocedures
Power Saving Function Setting Use
ON/ZERO
or UNITS key to select
On = power saving on Off = power saving off
ON/ZERO
key Upward key (from 0 to 9)
UNITS
key Downward key (from 9 to 0)
TARE
key Move the cursor to right
NET/GROSS
key Move the cursor to left.
11
EXCELL PRECISION CO., LTD.
ZSME300000276
02005324
Automatic power-off timer setting
Use
ON/ZERO
or
UNITS
key to key in
parameter 0 No auto power-off
1 When the scale is idle for 1
minute, the scale automatically switches off
2 When the scale is idle for 2
minutes, the scale automatically switches off
9 When the scale is idle for 9
minutes, the scale automatically switches off
2-1-2 Automatic Power-off Timer Setting
Select in the General Function setting mode to change the automatic power-off timer setting.
LED displays the last status
Automatic power-off function
When the weight on weigh pan is less than 10d or keeps idle for the set time, the scale will automatically switch off.
Press
TARE
key
Use
ON/ZERO
or
UNITS
key to key in parameter
Press
TARE
key
ACTIONS
DISPLAY NOTE
ON/ZERO
key Upward key (from 0 to 9)
UNITS
key Downward key (from 9 to 0)
TARE
key Move the cursor to right
NET/GROSS
key Move the cursor to left.
Key in the parameter “00”
Press
TARE
key to complete the general setting
p
rocedures
12
EXCELL PRECISION CO., LTD.
ZSME300000276
02005324
2-1-3 Hi/Lo/OK Function Setting
Select in the General Function setting mode to set the Hi/Lo/OK function. This function is available in all unit modes. In one specific unit mode,
enter to set the Hi/Lo/OK values.
When the high limit and low limit are both set as “0”, the Hi/Lo/OK function is disabled.
ACTIONS
DISPLAY
NOTE
LED displays the last status
Press
TARE
key
Key in the desired high limit value
Press
TARE
key
Key in the desired low limit value
Press
TARE
key
Key in the parameter setting for alarm
Press
TARE
key
LED displays the last status
LED displays the last status
.
.
.
.
Alarm setting
(a) (b) (c) The definition of a, b, c positions: (a) 1 = beep sound on
0 = beep sound off (b) 1 = beep sound on when stable
0 = beep sound on when unstable (c) 1 = beep sound on when the weight is
between high and low limits
0 = beep sound on when the weight is not
between high and low limits and higher than 10d
ON/ZERO
key Upward key (from 0 to 9)
UNITS
key Downward key (from 9 to 0)
TARE
key Move the cursor to right
NET/GROSS
key Move the cursor to left.
Key in the parameter and then press
TARE
key to go back to the
Advanced function setting menu
13
EXCELL PRECISION CO., LTD.
ZSME300000276
02005324
2-1-4 Restore to the Default Setting
Select in the General Function setting mode to restore to the default setting.
The default setting includes the following:
1) External weight calibration
2) HI/LO/OK setting value
3) Noise filter setting (External)
For OIML, NTEP or Sri Lanka approved models, setting is disabled.
ACTIONS
DISPLAY NOTE
Return (Cancel the
restoration)
Restore to the default
setting
Press
TARE
key
Use
ON/ZERO
or
UNITS
keys to select “return” or “format”
Press
TARE
key
ON/ZERO
key Upward key (from 0 to 9)
UNITS
key Downward key (from 9 to 0)
TARE
key Move the cursor to right
NET/GROSS
key Move the cursor to left.
Key in the parameter “00”
Press
TARE
key to complete the general setting
p
rocedures
14
EXCELL PRECISION CO., LTD.
ZSME300000276
02005324
2-1-5 Noise Filter Setting
Select in the General Function setting mode to set the noise filter setting.
Returning to the zero point setting
Use
ON/ZERO or UNITS key to key in the
parameters or zero point Default setting = 0 0 ⇒ No skip 5 skip 5d 1 skip 1d 6 skip 6d 2 skip 2d 7 skip 7d 3 skip 3d 8 skip 8d 4 skip 4d 9 skip 9d When the weight on the scale is over
1/3 full capacity, the function is on.
Digital switch & Stabilization range setting
LED displays the last parameter setting
Filter parameter setting
LED displays the last parameter setting
Filter parameter setting Use
ON/ZERO
or
UNITS
keys to key
in the parameters. Default setting = 5 Parameter 0 ~ 9, the larger the number, the faster the filter response. Fast response can lead to weight instability.
Digital switch & Stabilization range setting
Use
ON/ZERO or
UNITS
keys to key in the parameters. Default setting = 0 Parameter 0 ~ 9, the larger the number the more stable the weight.
ACTIONS
DISPLAY
NOTE
Press
TARE
key
Use
ON/ZERO
or
UNITS
key to key in the parameters
Press
TARE
key
Use
ON/ZERO
or
UNITS
key to key in the parameters
Press
TARE
key
Use
ON/ZERO
or
UNITS
key to key in the parameters
Press
TARE
key
Returning to zero point setting
LED displays the last status
Key in the parameter and then press
TARE
key to go back to the
Advanced function setting menu
For OIML, NTEP or Sri Lanka approved models, setting is disabled.
15
EXCELL PRECISION CO., LTD.
ZSME300000276
02005324
2-2 Weight Calibration
In the weighing mode, press the NET/GROSS and ON/ZERO keys at the same time to enter the External Function setting mode. The LED shows and uses the NET/GROSS or UNITS key to select to enter the weight calibration mode.
For approved models, setting is disabled. Weight calibration conditions:
The calibration weight value placed on the weight pan must be over e100, and the standard deviation of the weight must be within 10%.
The LED displays the calibration weight value (the maximum capacity of 1
st
unit) and the unit. The digit on
the right side keeps flashing.
The scale is calculating the internal value of the zero point. After stabilization, the LED shows the calibration weight value.
Press
TARE
key
When the digit on the right flashes, ensure that the weigh pan is empty. Then Press the
TARE
key. After stabilization
Place the displayed weight on the weigh pan. Press the
TARE
key. The scale calculates the full capacity internal value and after stabilization, the buzzer beeps 3 times. Remove the weight from the weigh pan and the procedure is completed.
.
.
ACTIONS
DISPLAY
NOTE
ON/ZERO
key Upward key (from 0 to 9)
UNITS
key Downward key (from 9 to 0)
TARE
key Move the cursor to right
NET/GROSS
key Move the cursor to left.
16
EXCELL PRECISION CO., LTD.
ZSME300000276
02005324
Appendix1 7-Segment Display Characters
Digit
7-segments
letter
Alphabet
7-segments
letter
Alphabet
7-segments
letter
0
A
N
1
B
O
2
C
P
3
D
Q
4
E
R
5
F
S
6
G
T
7
H
U
8
I
V
9
J
W
K
X
L
Y
M
Z