Excell ESW-3, ESW-6, ESW-30, ESW-15 User Manual

User Manual
Waterproof Weighing scale
ESW
© Excell Precision Limited 2015. All rights reserved Worldwide. The information contained herein is the property of Excell Precision Limited and is supplied without liability for errors or omissions. No part may be reproduced or used except as authorized by contract or other written permission. The copyright and the foregoing restriction on reproduction and use extend to all media in which the information may be embodied.
02005316 ZSME3000000306
1
EXCELL PRECISION CO., LTD.
Table of Content
INSTRUCTION FOR USE ...................................................................................................2
PREPARING TO USE THE SCALE.....................................................................................2
CHAPTER 1 INTRODUCTION............................................................................................4
1-1 FEATURES AND SPECIFICATIONS.....................................................................................4
1-2 POWER SUPPLY............................................................................................................4
1-3 PANEL AND KEYBOARD INTRODUCTION ...........................................................................5
1-4 OPERATING THE SCALE.................................................................................................5
1-5 SELF-TEST MODE.........................................................................................................7
1-6
ERROR MESSAGES .......................................................................................................8
1-7
WEIGHT UNIT ...............................................................................................................8
CHAPTER 2 ADVANCED FUNCTIONS..............................................................................9
2-1 ADVANCED FUNCTION SETTING TABLE............................................................................9
2-2 ADVANCED FUNCTION SETTING WORKFLOW .................................................................10
2-3 GENERAL FUNCTION SETTING ..............................................................11
2-3-1 Automatic Backlight Function Setting .......................................12
2-3-2 Automatic Power-off Timer Setting ...........................................13
2-3-3 Hi/Lo/OK Function Setting ........................................................14
2-3-4 Restore to the Default Setting ...................................................15
2-3-5 Noise Filter Setting ...................................................................16
2-3-6 Hold Function Setting ..............................................................17
2-3-7 Auto Unit Weight Averaging Setting ........................................18
2-4 WEIGHT CALIBRATION ..............................................................................19
APPENDIX 1 7-SEGMENT DISPLAY CHARACTERS ......................................................20
02005316 ZSME3000000306
2
EXCELL PRECISION CO., LTD.
Instruction for Use
1. Please keep the scale in a cool dry place. Do not store it at high temperature.
2. Avoid objects impacting with the scale. Do not drop loads onto the scale or subject the weigh pan to any strong shock loads.
3. The load placed on the weigh pan must not exceed the maximum weighing capacity of the scale.
4. If the scale is not going to be used for some time, please clean it and store it in a plastic bag in dry conditions. A desiccant sachet may be included to prevent any moisture build up.
5. Please operate or charge the scale in an open area. Do not squeeze the power cord to avoid wire on fire.
Preparing To Use the Scale
1. Adjust the four levelling feet (if fitted) to set the scale pan level using the spirit level
bubble located at the front of the scale.
2. Avoid operating the scale in direct sunlight or drafts of any kind.
3. If possible avoid connecting the scale to ac power outlet sockets which are adjacent to
other appliances to minimise the possibility of interference affecting the performance of the scale.
4. Remove any weight that might be on the weigh pan before the scale is switched on and
avoid leaving weight on the pan for long periods of time
5. All goods weighed should be placed in the centre of the weigh pan for accurate
weighing. The overall dimensions of the goods being weighed should not exceed the dimension of the weigh pan.
6. We suggest to warm up the scale for 15~20 minutes before operation to ensure best
accuracy.
7. When the symbol keeps flashing on the display, please charge the batteries soon.
8. Introduction of Storage Battery
Thank you for purchasing EXCELL waterproof weighing scale. In order to use the scale properly, please read this instruction carefully before use. If you have a problem concerning the scale, please contact your supplier.
02005316 ZSME3000000306
3
EXCELL PRECISION CO., LTD.
No burning
Caution Corrosion
Due to the storage battery adopt the advanced free-maintaining technique, customers need not to replenish electrolyte. The scale should be recharged every 3 months to prevent failure of the internal rechargeable battery.
1. The battery should be charged for 8~10 hours.
2. The temperature of battery should below 45℃.
Maintaining
1. Please do not discharge with over-current when using the battery. Please charge the
battery after discharging current.
2. Please take down the battery when the scale is not used for a long time or break the
connection of cathode.
3. Do not short the battery terminals to check whether there is current. Please check
whether the connection point is firm to guarantee good connection.
4. The battery should be replaced by specialized person. No reverse-battery or the
product will be damaged.
a) Anode of battery should be connected with Anode of product batteryusually red
cable
b) Cathode of battery should be connected with Cathode of product battery (usually
brown cable or black cable)
c) See the picture following
Safety warnings
1. The electrolyte of battery is caustic which causes metal, cotton, etc to corrode.
2. The hydrogen will be resolved when using or charging the battery and it will cause
explosion when approaches fire.
Brown cable(or black cable) connected with Anode of battery
Red cable connected with Cathode of battery
Warning explosion
Children faraway
02005316 ZSME3000000306
4
EXCELL PRECISION CO., LTD.
Chapter 1 Introduction
1-1 Features and Specifications
Features:
Sealed waterproof silica gel strip blocks water from infiltrating into the scale. Surrounded by waterproof grade sheeting to ensure the water free. 1/3,000~1/6,000 display resolution available. Aluminum frames coated with anti-corrosion treatment for longer lifespan. Stainless steel made feet and platter for anti-rust. High speed of 24bits AD fast reacts and shortens the weighing operation duration. Selectable units: Kilogram (kg), gram (g), ounce (oz), and pound (lb) weighing units
available.
Built-in rechargeable battery can be easily replaced. Low power indication and auto power off. Well-designed protection point for transportation. Options: Double-sided display.
Specifications:
1-2 Power Supply
Power Supply and Power Consumption
DC 6V / 4Ah or 6V / 7Ah rechargeable battery AC 110V~240V (±10) adapter
14 mA (system no backlightSingle display) about 275 hours 28 mA (system no backlightdouble display) about 132 hours 19 mA (system backlightsingle display) about 200 hours 45 mA (system backlightdouble display) about 78 hours
Low Battery Warning
When the symbol keeps flashing on the display; please recharge the batteries soon.
Model Capacity Division Resolution
ESW-3 3kg
0.5g 1g
1/6,000 1/3,000
ESW-6 6kg
1g 2g
1/6,000 1/3,000
ESW-15 15kg
2g 5g
1/7,500 1/3,000
ESW-30 30Kg
5g
10g
1/6,000 1/3,000
Operating Temperature: -10°C ~ 40°C (14°F ~ 104°F) Weight of the scale: approximate 3.5kg
02005316 ZSME3000000306
5
EXCELL PRECISION CO., LTD.
.
.
.
.
.
.
.
. .
.
Pcs
1-3 Panel and Keyboard Introduction
Icon Introduction
STABLE : The weight is stable. GROSS : The scale is in the gross mode. The display shows the goods and any
container weight. This Gross status indication is on when the TARE function is used.
: The unit weight is not sufficient. When the icon is on, the counting
function is operational but the count may contain errors.
: The sampling size is not sufficient When the icon is on, the counting
function is operational but the count may contain errors.
Keyboard Function
ON/ZERO KEY
This key possesses two functions: Power On and Zero function.
OFF KEY
When the scale is switched on, press the OFF key, the scale will switch off.
TARE KEY
The tare function will not operate during the following conditions:
1. When the scale powers on if the weight is negative and after a container is placed on the weigh pan if the weight is still below zero.
2. The tare value is over the full scale capacity.
UNITS KEY
Press the UNITS key to switch weight units; the icons will indicate the active units.
NET/GROSS KEY
In the Tare mode, the screen displays the “TARE” icon; press the NET/GROSS key to switch between the “Net value” and the “Gross value”.
1-4 Operating the Scale
POWER ON
When the scale is off, press the ON/ZERO key, the scale will switch on.
POWER OFF
When the scale is on, press the OFF key, the scale will switch off.
02005316 ZSME3000000306
6
EXCELL PRECISION CO., LTD.
Pcs
.
.
.
.
. .
.
. .
.
Pcs
.
.
.
.
. .
.
. .
.
ZERO
When the weigh pan is empty (free of load) and the display is not showing zero, press the ON/ZERO key to zero the scale. At zero, the “Æ0Å” indication is on.
When the weight value is within the zero range, the zero function operates to zero the
scale or cancel the tare function.
Zero range: The OIML & NTEP models have a zero range of ± 2% of Full Scale. The
Sri Lanka model has a zero range of ± 4% of Full Scale.
SWITCHING UNITS
Press the UNITS key to switch weight units, the icons or arrows will indicate the active units as appropriate. The units available are dependent on the exact scale model.
After power off, the scale will memorize the active units. When the scale is powered on
again, it displays the previously active units.
TARE FUNCTION
(1) Put a container on the weigh pan and after the weight is stable, press the TARE
key to zero the weight of the container. The screen displays the “Net” icon.
(2) Put the goods in the container, the screen displays the net weight value of the
goods.
(3) Remove the full container; the screen displays the negative weight value of the
container. At this time pressing the TARE key again will cancel the tare and the scale reverts back to zero. The “Net” icon is switched off.
The tare function can be operated continually to the full weighing capacity of the scale. Continual tare operation is adding or removing tare objects on weigh pan and pressing
the TARE key each time.
NET/GROSS FUNCTION
In the Tare mode, the screen displays the “Net” icon, press the NET/GROSS key to switch between the “Net value” and the “Gross value”.
When the GROSS icon is on, the weight value on the display is the total amount of
the tare value and net value.
At the Gross status, only OFF and NET/GROSS keys are functional. NET/GROSS key is only used in Tare mode.
SIMPLE COUNTING FUNCTION
(1) Use the UNITS key to enter into the “PCS” mode (2) Press the NET/GROSS key to select the counting sample size
(S = 10, S = 20, S = 50, S = 100, S=200). The LCD shows , , , , in order.
(3) Put the samples on the weigh pan and press the UNITS key, the screen displays
“------”. After the sampling process is complete, put the goods on the weigh pan and the screen shows the quantity of the items.
The sample weight should be heavier then the minimum capacity of the scale (20d), If
not the arrow pointing to the icon will be activated.
The weight of a sample should be heavier than the 0.2d (d=division), or the arrow
pointing to the icon will be on.
When the or are indicated, the scale is still operational but the count may
contain errors.
02005316 ZSME3000000306
7
EXCELL PRECISION CO., LTD.
To power off in this mode, the scale will memorize the “Pcs” unit. When the scale is
powered on again, it directly enters the simple counting mode.
While the “Auto unit weight average” function is available in the Advanced Function,
the goods on the weigh pan are 5pcs greater than the sample size and less than double the sample size, the scale will automatically re-sample the unit weight.
1-5 Self-Test Mode
Set the switch SWA1 on the bottom of machine to the LOCK position.When power is off,
hold
NET/GROSS
, and press
ON/ZERO
key, Wait till display shows to
enter “Self-Test Mode”.
Press
UNITS
key
Press
ON/ZERO
key
Press
ON/ZERO
key
Press
ON/ZERO
key
Press
ON/ZERO
key
Press
UNITS
key
Press
UNITS
key
Press
UNITS
key
INTERNAL VALUE MODE (
must hook up Load Cell to test)
c Press
T ARE
to enter, and the display shows internal value
d Please check whether the internal value has changed obviously with weight
changing.
e Please check the backlight. f Press
ON/ZERO
key to back to the last screen , the display shows
KEYPAD TEST MODE
c Press
TARE
to enter, display shows
Keypad’s internal code:
T ARE
key= 07
UNITS
key= 06
NET/GROSS
key=05
d Press
ON/ZERO
key to back to the last screen , the display shows
FIRMWARE VERSION DISPLAY MODE
c Press
T ARE
to enter , display shows the firmware version
d Press
T ARE
key again, the display shows maintenance number for 2 secs
e Press
ON/ZERO
key to back to the last screen, display shows
BACK TO THE LAST SCREEN
Press
T ARE
key to exit self-test mode, the scale will restart automatically.
02005316 ZSME3000000306
8
EXCELL PRECISION CO., LTD.
1-6 Error Messages
Initial zero is higher than the zero range when switching the indicator on.
(Approval models)
Initial zero is lower than the zero range when switching the indicator on.
(Approval models)
Internal value is below Zero.
The weight of the object is over 9 divisions of the maximum capacity. A/D IC value is over the maximum range.
1-7 Weight Unit
kg 1 g = 0.001 kg
g 1 g = 1 g
lb 1 g = 0.002204623 lb
lb/oz 1 g = 0.03527396 oz
oz 1 g = 0.03527396 oz GN 1 g = 15.432358 GN dwt 1 g = 0.6430149 dwt
ct 1 g = 5 ct
02005316 ZSME3000000306
9
EXCELL PRECISION CO., LTD.
Chapter 2 Advanced Functions
2-1 Advanced Function Setting Table
Below is an overview of the advanced functions. For detailed settings refer to the following sections:
DISPLAY LEVEL 1 FUNCTION DISPLAY LEVEL 2 FUNCTIONS
Exit the ADVANCED
FUNCTION setting mode
--- ---
Return to theADVANCED
FUNCTION setting menu
Automatic backlight
function setting
Automatic power-off
timer setting
Hi/Lo/OK function setting
Restore the default
settings
Noise filter setting
Hold function setting
General Function setting
mode
Auto unit weight
averaging setting
External Weight
Calibration
--- ---
02005316 ZSME3000000306
10
EXCELL PRECISION CO., LTD.
2-2 Advanced Function Setting Workflow
In the weighing mode, press the NET/GROSS and ON/ZERO keys at the same time to enter the Advanced Function setting mode. The LCD shows .
Overall workflow of the Advanced Function setting mode:
General Function setting mode External Weight Calibration Exit the Advanced Function setting mode
Refer to the following sections for the detailed operation procedures of each function setting.
Press UNITS key
Press ON/ZERO
key
Press UNITS key
Press UNITS keyPress ON/ZERO key
Press ON/ZERO
key
02005316 ZSME3000000306
11
EXCELL PRECISION CO., LTD.
2-3 General Function Setting
There are 7 functions in the general function setting mode from
to .
Workflow of the General Function setting:
ACTIONS
DISPLAY NOTE
Return to the Advanced Function Setting Mode Menu
Automatic Backlight Function Settings
Automatic Power-off Timer Settings
Hi/Lo/OK Settings
Restore the Default Settings
Noise Filter Settings
Hold Function Settings
Auto Unit Weight Averaging Setting
Refer to the following sections for detailed operation procedures of each setting.
Press
TARE
key
Use the keys to select the function
Key in “00” and press TARE key to return to the Advanced function setting menu
ON/ZERO
key Upward key (from 0 to 9)
UNITS
key Downward key (from 9 to 0)
TARE
key Move the cursor to right
NET/GROSS
key Move the cursor to left
02005316 ZSME3000000306
12
EXCELL PRECISION CO., LTD.
2-3-1 Automatic Backlight Function Setting
Select in the General Function setting mode to change the backlight function setting.
Automatic backlight function
When the weight is over 10d, the display backlight will be on. After the weight is stable for 10 seconds or when the scale returns to zero, the display backlight switches off.
DISPLAY
ACTIONS
NOTE
Key in the parameter “00”
Press
TARE
key
Use
ON/ZERO
or
UNITS
key to select function “on” or “off
Press
TARE
key
Press
TARE
key to complete the general setting procedures
LCD displays the last status
02005316 ZSME3000000306
13
EXCELL PRECISION CO., LTD.
2-3-2 Automatic Power-off Timer Setting
Select in the General Function setting mode to change the automatic power-off timer setting.
Automatic power-off function
When the weight on weigh pan is less than 10d or keeps idle for the set time, the scale will automatically switch off.
ACTIONS
DISPLAY NOTE
LCD displays the last status
ON/ZERO
key Upward key (from 0 to 9)
UNITS
key Downward key (from 9 to 0)
TARE
key Move the cursor to right
NET/GROSS
key Move the cursor to left.
Automatic power-off timer setting
Use
ON/ZERO
or
UNITS
key to key in
parameter 0 No auto power-off
1 When the scale is idle for 1
minute, the scale automatically switches off
2 When the scale is idle for 2
minutes, the scale automatically switches off
9 When the scale is idle for 9
minutes, the scale automatically switches off
Press
TARE key
Use
ON/ZERO
or UNITS
key
to key in parameter
Key in the parameter “00”
Press
TARE
key to complete
the general setting procedures
Press TARE key
02005316 ZSME3000000306
14
EXCELL PRECISION CO., LTD.
2-3-3 Hi/Lo/OK Function Setting
Select in the General Function setting mode to set the Hi/Lo/OK function. This function is available in all unit modes. In one specific unit mode,
enter to set the Hi/Lo/OK values. When the high limit and low limit are both set as “0”, the Hi/Lo/OK function is disabled.
DISPLAY
ACTIONS
NOTE
LCD displays the last status
LCD displays the last status
.
lb
LCD displays the last status
.
lb
.
lb
.
lb
Alarm setting
(a) (b) (c) The definition of a, b, c positions: (a) 1 = beep sound on
0 = beep sound off (b) 1 = beep sound on when stable
0 = beep sound on when unstable (c) 1 = beep sound on when the weight is
between high and low limits
0 = beep sound on when the weight is
not between high and low limits and higher than 10d
ON/ZERO
key Upward key (from 0 to 9)
UNITS
key Downward key (from 9 to 0)
TARE
key Move the cursor to right
NET/GROSS
key Move the cursor to left.
Key in the desired high limit value
Key in the parameter setting for alarm
Press
TARE
key
Press
TARE
key
Key in the desired low limit value
Press
TARE
key
Press
TARE
key
Key in the parameter “00”
Press
TARE
key to complete
the general setting procedures
02005316 ZSME3000000306
15
EXCELL PRECISION CO., LTD.
2-3-4 Restore to the Default Setting
Select in the General Function setting mode to restore to the default setting.
The default setting includes the following:
1) External weight calibration
2) HI/LO/OK setting value
3) Noise filter setting (External)
4) Sampling setting for the counting function
Setting is only available for non-approval models.
ACTIONS
DISPLAY NOTE
Return (Cancel the
restoration)
⇒ Restore to the default
setting
ON/ZERO
key Upward key (from 0 to 9)
UNITS
key Downward key (from 9 to 0)
TARE
key Move the cursor to right
NET/GROSS
key Move the cursor to left.
Press
TARE
key
Use
ON/ZERO
or
UNITS
keys
to select “return” or “format”
Press
TARE
key to complete the
general setting procedures
Key in the parameter “00”
Press
TARE
key
02005316 ZSME3000000306
16
EXCELL PRECISION CO., LTD.
2-3-5 Noise Filter Setting
Select in the General Function setting mode to set the noise filter setting.
When modifying , the parameters of remain un-altered.
DISPLAY
ACTIONS
NOTE
Returning to the zero point setting
Use ON/ZERO or UNITS key to key in the parameters or zero point Default setting = 0 0 ⇒ No skip 5 skip 5d 1 skip 1d 6 skip 6d 2 skip 2d 7 skip 7d 3 skip 3d 8 skip 8d 4 skip 4d 9 skip 9d When the weight on the scale is over
1/3 full capacity, the function is on.
Digital switch & Stabilization range setting
LCD displays the last parameter setting
Filter parameter setting
LCD displays the last parameter setting
Filter parameter setting Use
ON/ZERO
or
UNITS
keys to key
in the parameters. Default setting = 5 Parameter 0 ~ 9, the larger the number, the faster the filter response. Fast response can lead to weight instability.
Digital switch & Stabilization range setting
Use
ON/ZERO or UNITS keys to key in
the parameters. Default setting = 0 Parameter 0 ~ 9, the larger the number the more stable the weight.
Returning to zero point setting
LCD displays the last status
Use
ON/ZERO
or
UNITS
key to key in the parameters
Press
TARE
key to complete
general setting procedures
Press
TARE
key
Press
TARE
key
Use
ON/ZERO
or
UNITS
key to key in the parameters
Press
TARE
key
Use
ON/ZERO
or
UNITS
key to key in the parameters
Press
TARE
key
Key in the parameter “00”
Setting is only
available for non-approval models.
02005316 ZSME3000000306
17
EXCELL PRECISION CO., LTD.
2-3-6 Hold Function Setting
Select in the General Function setting mode to set the hold function.
= Hold function disabled = “Peak hold” mode
Keep displaying the maximum weight when the weight is continually changing To exit this mode, press any key.
= “Stable hold 1” mode
When the weight is stable, the LCD shows the current weight value. To exit this mode, press any key.
= “Stable hold 2” mode
When the weight is stable, the LCD shows the current weight value. When the weight returns to zero (<10d), the hold mode is cancelled automatically.
ACTIONS
DISPLAY
NOTE
Hold function setting
Use
ON/ZERO
or
UNITS
keys to key in the parameters Default setting = 0 0 Hold function disabled 1
“Peak hold” mode
2 “Stable hold 1” mode 3 “Stable hold 2” mode
Hold function setting
LCD displays the last parameter setting
ON/ZERO
key Upward key (from 0 to 9)
UNITS
key Downward key (from 9 to 0)
TARE
key Move the cursor to right
NET/GROSS
key Move the cursor to left.
Use
ON/ZERO
or
UNITS
key
to key in the parameters
Press
TARE
key to complete
the general setting procedures
Press
TARE
key
Key in the parameter “00”
Press
TARE
key
02005316 ZSME3000000306
18
EXCELL PRECISION CO., LTD.
2-3-7 Auto Unit Weight Averaging Setting
Select in the General Function setting mode to set Auto Unit Weight Averaging.
ON/ZERO
key Upward key (from 0 to 9)
UNITS
key Downward key (from 9 to 0)
TARE
key Move the cursor to right
NET/GROSS
key Move the cursor to left.
Auto unit weight averaging setting Use
ON/ZERO
or
UNITS
keys to key in the parameters Default setting = 0 0 Auto unit weight averaging 1 Non auto unit weight averaging
Auto unit weight averaging setting
LCD displays the last parameter setting
DISPLAY
ACTIONS
NOTE
Use
ON/ZERO
or
UNITS
key
to key in the parameters
Press
TARE
key to complete
the general setting procedures
Press
TARE
key
Key in the parameter “00”
Press
TARE
key
02005316 ZSME3000000306
19
EXCELL PRECISION CO., LTD.
kg
kg
kg
2-4 Weight Calibration
In the weighing mode, press the NET/GROSS and ON/ZERO keys at the same time to enter the Advanced Function setting mode. The LCD shows and use the NET/GROSS or UNITS key to select to enter the weight calibration mode.
Unavailable for approval models Weight calibration conditions:
The calibration weight value placed on the weight pan must be over 100e, and must be within the 90% and 110% of the full weight
The LCD displays the calibration weight value (the maximum capacity of 1st unit) and the unit. The digit on the right side keeps flashing.
The scale is calculating the internal value of the zero point. After stabilization, the LCD shows the calibration weight value.
Press
TARE
key
When the digit on the right flashes, ensure that the weigh pan is empty. Then Press the
TARE
key. After stabilization
Place the displayed weight on the weigh pan. Press the
TARE
key. The scale calculates the full capacity internal value and after stabilization, the buzzer beeps 3 times. Remove the weight from the weigh pan and the procedure is completed.
.
.
ACTIONS
DISPLAY
NOTE
ON/ZERO
key Upward key (from 0 to 9)
UNITS
key Downward key (from 9 to 0)
TARE
key Move the cursor to right
NET/GROSS
key Move the cursor to left.
02005316 ZSME3000000306
20
EXCELL PRECISION CO., LTD.
Appendix 1 7-Segment Display Characters
Digit 7 segments letter Alphabet 7 segments letter Alphabet 7 segments letter
0
A
N
1
B
O
2
C
P
3
D
Q
4
E
R
5
F
S
6
G
T
7
H
U
8
I
V
9
J
W
K
X
L
Y
M
Z