Excell Counting Scale User Manual

EXCELL PRECISION CO., LTD.
Counting ZSME200000003
USER MANUAL
© Excell Precision Limited 2006. All rights reserved Worldwide.
The information contained herein is the property of Excell Precision Limited and is supplied without liability for errors or omissions. No part may be reproduced or used except as authorised by contract or other written permission. The copyright and the foregoing restriction on reproduction and use extend to all media in which the information may be embodied.
EXCELL PRECISION CO., LTD.
Counting ZSME200000003
1
TABLE OF CONTENTS
INSTRUCTIONS FOR USE.............................................................................................3
PREPARING TO USE THE SCALE................................................................................3
CHAPTER 1 PRODUCT INTRODUCTION.....................................................................4
1-1 DISPLAYS.........................................................................................................4
1-2 KEYBOARD FUNCTION...................................................................................6
1-3 POWER SOURCES...........................................................................................9
CHAPTER 2 OPERATION............................................................................................10
2-1 SWITCH SCALE ON........................................................................................10
2-2 SWITCH SCALE OFF......................................................................................10
2-3 DISPLAY BACKLIGHT....................................................................................10
2-4 SAMPLING SPEED & BACK TO ZERO SETTING.........................................11
2-5 AUTO CALIBRATION......................................................................................12
2-6 ZERO FUNCTION............................................................................................12
2-7 UNIT WEIGHT CALIBRATION........................................................................13
2-8 COUNTING......................................................................................................14
2-8-1 The Unit Weight of Object is Unknown..............................................14
2-8-2 The Unit Weight of Object is Known..................................................14
2-9 DEDUCTION OF CONTAINER WEIGHT........................................................15
2-9-1 The Weight of Container is Unknown ( Tare )...................................15
2-9-2 Clear Tare Value...................................................................................15
2-9-3 The Weight Container is Known (Pre-set Tare).................................16
2-9-4 Clear The Pretare Value.......................................................................17
2-10 TOTALISING.................................................................................................18
2-10-1 Quantity totalising.............................................................................18
2-10-2 Recall Quantity Totalised..................................................................19
2-10-3 Weight Totalising...............................................................................19
2-10-4 Recall Weight Total............................................................................19
2-10-5 Clear all Totalised Values.................................................................20
CHAPTER 3 USER INTERFACE SETTING.................................................................21
3-1 M+ KEY ACTION........................................................................................21
3-2 PRE-SET TARE...............................................................................................21
3-3 PRE-SET QUANTITY CONDITIONS...............................................................22
3-4 M+ OPERATION.........................................................................................22
3-5 M+ CONTINUOUS OPERATION (Only After Return to Zero)..................23
3-6 MULTIPLE UAGE KEY U.W.PST / (kg/lb) FUNCTION SELECTION...........23
3-7 BAUD RATE SETTING (for only with the serial interface)..........................25
EXCELL PRECISION CO., LTD.
Counting ZSME200000003
2
CHAPTER 4 KEYBOARD TYPES................................................................................26
4-1 STANDARD TYPE...........................................................................................26
4-2 TYPE I DUAL WEIGHING UNIT..................................................................27
4-2-1 Unit Selection.......................................................................................27
4-2-2 Quantity Preset....................................................................................27
4-2-3 Weight Preset.......................................................................................28
4-3 TYPE II UNIT WEIGHT PRESET................................................................29
4-3-1 Unit Weight Preset Stores...................................................................29
4-3-2 Quantity Preset....................................................................................30
4-3-3 Weight Preset.......................................................................................31
4-4 COMBINATION TYPE DUAL WEIGHING UNIT + UNIT WEIGHT PRESET
...............................................................................................................................32
4-4-1 Unit Selection.......................................................................................32
4-4-2 Unit Weight Preset Stores...................................................................33
4-4-3 Quantity Preset....................................................................................34
4-4-4 Weight Preset.......................................................................................35
CHAPTER 5 INTERFACE.............................................................................................36
5-1 TOTALISING...................................................................................................36
5-2 PRESET TARE OPERATION CONDITIONS...................................................37
5-3 QUANTITY PRESET OPERATION CONDITIONS..........................................38
5-4 TOTALISING OPERATION CONDITIONS......................................................39
5-5 THE CONDITION TO ALLOW THE NEXT TOTALISING OPERATION.........40
5-6 COMBINATION KEY OPERATION MODE......................................................41
5-7 BAUD RATE SETTING....................................................................................42
5-8 TRANSMIT MODE SELECTION......................................................................43
EXCELL PRECISION CO., LTD.
Counting ZSME200000003
3
INSTRUCTIONS FOR USE
1) Please keep the scale in a cool and dry place. Do not store at high temperatures.
2) Please keep the scale clean and free from insect infestation.
3) Avoid objects impacting with the scale.
4) The load placed on the weigh pan must not exceed the maximum weighing capacity of the scale.
5) If the scale is not going to be used for some time, please clean it and store it in a plastic bag in dry conditions. A desiccant sachet may be included to prevent moisture build up.
6) If the scale is not going to be used for some time, the internal rechargeable battery should be recharged every three months. (If using dry batteries, take the dry batteries out before storing)
7) Remove the plastic drip cover from the scale for best accuracy.
PREPARING TO USE THE SCALE
1. Locate the scale on a firm level surface free from vibrations for accurate weight readings.
2. Adjust the four levelling feet to set the scale pan level using the spirit level bubble located at the front of the scale.
3. Avoid operating the scale in direct sunlight or drafts of any kind.
4. If possible avoid connecting the scale to ac power outlet sockets which are adjacent to other appliances to minimise the possibility of interference affecting the performance of the scale.
5. Remove any weight that might be on the weigh pan before the scale is switched on and avoid leaving weight on the pan for long periods of time.
6. Once the scale has been switched on, it will go through a LCD display test and then re-zero to be ready for use.
7. The scale requires 15~20 minutes warm up before operation to ensure best accuracy.
8. Please note when symbol appears on the screen, the internal battery needs to be recharged.
9. All goods weighed should be placed in the centre of the weigh pan for accurate weighing. The overall dimensions of the goods being weighed should not exceed the dimension of the weigh pan.
EXCELL PRECISION CO., LTD.
Counting ZSME200000003
4
T
.
. .
. . .
. . . .
Pcs
T
Pcs
.
. .
. . .
. . . .
CHAPTER 1 PRODUCT INTRODUCTION
1-1 DISPLAYS
WEIGHT
UNIT WEIGHT
QUANTITY
0
3 3
. . . . .
3 3
. . . . .
+
stable
3 3
. . . . . .
DIGITAL DISPLAYS
1. WEIGHT: (5 digits)
Weight of objects on the weigh pan or totalised weight.
2. UNIT WEIGHT: (5 digits) Unit weight of objects on the weigh pan or number of weighing in the totaliser.
3. QUANTITY: (6 digits) Quantity (count) of objects on the weigh pan or total quantity in the totaliser.
SYMBOL ICONS 3
1. or
Net
or
Tare
: Tare displays the deduction of the container weight
2. à0ß or
Zero
: Zero indication
3. + or M+ : M+ Totaliser
4. Stable or ~ : Stable indication
5. : If the piece weight is smaller than Minimum Sampling
Weight, 3 will be displayed. In order to clear the symbol 3, increase the size of the sample.
6. : If the piece weight is smaller than “Minimum Unit Weight”,
3 will be displayed. The scale still weighs even though
the piece weight is too small, this may affect the count accuracy.
7. : “Low Power warning.
PERFORMANCE:
High Precision Counting Scale (1/15,000 and 1/ 30,000 Display Resolution)
Minimum Sampling Weight = at least 20d. Minimum Unit Weight = at least 0.2d Standard Counting Scale (1/6,000 Display Resolution) Minimum Sampling Weight = at least 40d. Minimum Unit Weight = at least 0.8d
EXCELL PRECISION CO., LTD.
Counting ZSME200000003
5
(d=division)
EXCELL PRECISION CO., LTD.
Counting ZSME200000003
6
T
0
.
. .
. . .
. . . .
.
. .
. . .
. . . .
R
R
+
1-2 KEYBOARD FUNCTION
STANDARD TYPE
7 8 9
SMPL
QTY
PST
4 5 6
UNIT
W.T.
PST
CE
1 2 3
Z
M+
0
.
CE
T
MC
ADDITIONAL FUNCTIONS TYPE I Dual Weighing Unit
7 8 9
SMPL
QTY
PST
4 5 6
UNIT
W.T.
lb/kg
1 2 3
Z M+
0
.
CE
T MC
7 8 9 4 5 6
lb/kg
1 2 3 0
.
CE
This is a CLEAR key to delete all the High Limit values. Please refer to page 21.
This Key is to select weight units in lbor kg. Please refer to page 21.
This Key is to select weight units of lbor kg. Please refer to page 21.
EXCELL PRECISION CO., LTD.
Counting ZSME200000003
7
T
0
R
R
.
. .
. . .
. . . .
.
. .
. . .
. . . .
+
ADDITIONAL FUNCTIONS TYPE II Unit Weight Preset
7 8 9
SMPL
QTY
PST
4 5 6
UNIT
W.T.
U.W.
PST
1 2 3
Z M+
0
.
CE
T MC
7 8 9 4 5 6
U.W.
PST
1 2 3 0
.
CE
This key selects 10 pre-set stores of piece weight. Please refer to page 24.
This key selects 10 pre-set stores of piece weight. Please refer to page 24.
EXCELL PRECISION CO., LTD.
Counting ZSME200000003
8
T
0
R
R
.
. .
. . .
. . . .
.
. .
. . .
. . . .
+
COMBINATION TYPE Dual Weighing Units + Preset Piece Weight Stores
7 8 9
SMPL
QTY
PST
4 5 6
UNIT
W.T.
U.W.
PST
1 2 3
Z M+
0
.
CE
T MC
7 8 9 4 5 6
U.W.
PST
1 2 3 0
.
CE
This is a dual function key which selects: Preset piece weight stores and weighing units selection. Please refer to page 27.
This is a dual function key which selects: Preset piece weight stores and weighing units selection. Please refer to page 27.
EXCELL PRECISION CO., LTD.
Counting ZSME200000003
9
1-3 POWER SOURCES
Applicable Power Sources
1. AC 115V ± 10%
2. AC 230V ± 10%
3. 6V/1.2Ah Rechargeable battery
4. 6V/4Ah Rechargeable battery
5. 6V/10Ah Rechargeable battery
6. 9V DC via mains adapter
Power Consumption
Approximately DC 20 mA Approximately DC 40 mA (With display Backlight)
Low Power Warning
When the symbol is displayed, it indicates that the scale needs to be
recharged or the dry batteries need to be replaced.
If the scale is not recharged after the low battery symbol is displayed, the scale will
switch itself off after 20-30 minuets.
4 Red light: Scale needs to be recharged.
Green light: Scale is recharged.
EXCELL PRECISION CO., LTD.
Counting ZSME200000003
10
CHAPTER 2 OPERATION
2-1 SWITCH SCALE ON
To switch on the scale, ensure that the weigh pan is empty and press the ON/OFF switch towards the ( | ) position and the scale will power on and set zero ready for
weighing or counting operations.
4 If the scale has a RS232 option fitted, the interface card number will be displayed
during the power up sequence.
2-2 SWITCH SCALE OFF
To switch the scale off, press the ON/OFF switch towards the ( ¡ ) position.
2-3 DISPLAY BACKLIGHT
AUTO BACKLIGHT
Press
Z
key, the unit weight displays followed by pressing
5
When the weight is over 10 divisions or any key is pressed, the display backlight
will be switched on. When the weight returns to zero or the weight on platform is less than 10 divisions, the display backlight will switch off after 5 seconds.
BACKLIGHT ON
Press
Z
key, the unit weight displays followed by pressing
4
Backlight is on all the time.
BACKLIGHT OFF
Press
Z
key, the unit weight displays followed be pressing
6
Backlight is off.
4 The backlight mode is stored in memory and will restored when the scale is
switched off and back on again.
EXCELL PRECISION CO., LTD.
Counting ZSME200000003
11
2-4 SAMPLING SPEED & BACK TO ZERO SETTING
Sampling speed: 01 ~ 15 (01 = slow, 15 = Very fast) The slower the sampling speed the more accurate the count will be. The faster the sampling speed the less accurate the count may be.
Zero display mask:
The standard auto zero tracking facility will normally ensure that the zero remains stable despite fluctuations in temperature or ac mains voltage etc. However in exceptional circumstances particularly mist conditions & other unstable environments a useful additional facility is available.
With the zero display mask it is possible to configure the scale not to show the first division of the weight. When configured the weight display will show zero as long as the weight is less than two display divisions.
1= 1st division is displayed as zero 0= normal mode (1st division is displayed)
Press
Z
key, the unit weight displays followed be pressing
8
Weight Unit Weight Quantity
X X
Sampling speed
Press 8 to increment (up) or 2
To decrement (down) the sampling speed
Zero display mask: Press 4 to select 0 (normal mode) or
6 to select 1 (1st division is displayed as zero)
Press Q’TY PST to exit and save the settings.
EXCELL PRECISION CO., LTD.
Counting ZSME200000003
12
2-5 AUTO CALIBRATION
Œ When the scale is at zero, press and hold . followed by pressing
U.W.
PST
until the weight display shows the scale capacity flashing, then release
the keys.
0
3
.
.
Stable
3
Value of full capacity of the scale
Place weights on the scale equivalent to full load capacity. When the scale
is stable, it will calibrate automatically. While the scale is counting back to zero, remove all the weights from the scale. The Calibration procedure is finished.
4 If full load is not available in weights it is possible to enter a weight value of less
than full load. Enter the weight value to be used and then place the weight on the scale, when the reading is stable the scale will automatically calibrate. While the scale is counting back to zero, remove all the weights from the scale. The Calibration procedure is finished. The new calibration will be stored and recalled when the scale is switched off and on again.
Press
CE
key to exit from the calibration mode.
4 Restore to the default setting: When the Auto calibration is miss-operated, and
the scale should be restored to the default setting, please hold
CE
key and
press
U.W. PST
key, the default setting will be restored after the scale counting
back to zero.
2-6 ZERO FUNCTION
Press the
Z
key to re-zero the display with no load on the weigh pan. When the zero
is set correctly the 3 icon will be displayed adjacent to the (à0ß) symbol on the weight display.
The zero setting range is ± 2% of the capacity of the scale.