eXact Well EZ Quick Start Manual

Page 1
ADVANCED PHOTOMETER SYSTEM
Well
DIGITAL ACCURACY FOR RELIABLE WATER CARE
QUICK START GUIDE
Page 2
2
©2017, Industrial Test Systems, Inc. Printed in USA
Welcome. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Well eXact® EZ Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Install “AAA” Batteries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Fill, Dip, Read
pH . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5
Total Iron . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-7
Nitrate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-9
Total Alkalinity
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-11
Total Hardness High. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12-13
For Best Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Technical Support . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Well eXact® EZ Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Method Verifi cation Ready Snap® 3. . . . . . . . . . . . . . . . . . . . . 15
About
Built-in Sample Cell . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Warranty (2 years) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Carrying Case . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
MSDS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .17-19
eXact® Micro Tests & Reagents . . . . . . . . . . . . . . . . . . . . . . . . 20
Patent Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
WELCOME TO YOUR NEW WELL EXACT® EZ PHOTOMETER.
Thank you for your Well eXact® EZ purchase! This guide will quickly walk you through the technical details of your new photometer. After initial set-up, test procedures, and tips, you will be on your way to digital water testing! Each test will require the use of one or more of the testing methods outlined in this manual.
YOUR WELL EXACT® EZ COMES WITH:
• Cleaning Brush
• Quick Start Guide (this booklet)
• 30 foil packet strips (6 of each): pH, Total Iron, Nitrate, Total Alkalinity, and Total Hardness, and 6 powder pillows of Iron Reducer.
WHAT YOU WILL NEED TO GET STARTED:
• Four (4) AAA batteries
• #4 Phillips head screwdriver
Note:
This system has been calibrated for use with only our
eXact® Micro reagents.
Page 3
3
Sensafe.com
WELL EXACT® EZ OVERVIEW
SCREW
Unscrew to remove base
INSTALL “AAA” BATTERIES (NOT INCLUDED)
1. Use a #4 Phillips head screwdriver to remove the screw from the base of your Well eXact
®
EZ .
2. Remove the base.
3. Install Four (4) new AAA batteries as illustrated inside your photometer’s battery compartment. We recommend using high quality batteries.
4. Replace the base fi rmly with pressure while tightening the screw. The meter will turn on automatically.
5. Tighten the screw with #4 Phillips head screwdriver. Be sure not to over tighten.
SAMPLE CELL - 4 ML
Built-in plastic 4mL cell for collecting your water sample
LCD DISPLAY
Displays results and test name
LIGHT SOURCE
Uses 525nm wavelength and 11mm path-length for testing
READ BUTTON
Starts test timer
MENU BUTTON
Cycles through available tests
BASE
Install/replace batteries here (IP67 rated waterproof)
RESULT
UNITS
LOW
BATTERY
LCD DISPLAY
ZERO/ON BUTTON
Turns the meter on and creates a baseline for your water testing
CELL COVER
Covers the cell for mixing and bright light situations
Protected against the effects of immersion in water to depth between 15cm and 1m
Part no. 486203
YOUR NEW WELL EXACT® EZ PHOTOMETER IS IDEAL FOR TESTING AND MAINTAINING YOUR WELL WATER.
Page 4
4
©2017, Industrial Test Systems, Inc. Printed in USA
Use the included cleaning brush to clean the cell after each test to achieve the most accurate result.
TIP
FILL CELL
Rinse the cell 3 times with the water sample to be tested and
FILL to the top to begin test.
2
In bright light situations use cell cover for zeroing and reading.
TIP
FILL, DIP, READ
3
SELECT TEST
Press and re-press the button until the display shows PH1.
ZERO METER
Press and the LCD Display reads 0.0PH, indicating the meter is ready for testing.
4
1
POWER ON PHOTOMETER
Press the button to power on the Well eXact® EZ.
PH
Page 5
5
Sensafe.com
REMOVE STRIP
Remove one eXact® Strip Micro PH and set in a dry, convenient place. Replace cap on bottle.
5
FILL, DIP, READ
READ RESULTS
READ result displayed as pH. This result is
automatically stored in the PH1 menu. After testing is complete, rinse the sample cell immediately.
7
DIP STRIP AND PRESS READ
Press to initiate a 20 second countdown and simultaneously DIP the eXact
®
Strip by submerging all pads in the sample then use a gentle, constant, back and forth motion (2 strokes per second) until the timer displays
‘1’
. Remove and discard the strip.
6
PH
Page 6
6
©2017, Industrial Test Systems, Inc. Printed in USA
ADD REDUCER
Tear open one eXact® Reagent EZ Open Reducer powder pillow. Add contents of powder pillow to the cell. Make sure entire powder pillow is emptied.
3
4
CAP & PRESS READ
Place the cell cover onto the sample cell. Press to initiate a 20 second countdown. Place thumb or fi nger over cap and turn meter upside-down repetitively during the 20 second countdown. When timer displays 1, place the meter on a fl at surface. The display will automatically start to count up for 40 seconds. Ignore result displayed.
Clean cell with Distilled Vinegar (5%), or 0.1N HCl before fi lling the meter with the sample to be tested. If running multiple tests in a row, using the same water sample, the cell does not have to be rinsed or cleaned with acid between each test. Rinse the sample cell 3 times with the water sample to be tested and FILL to the top to begin test.
2
RINSE AND FILL CELL
Use the included cleaning brush to clean the cell after each test to achieve the most accurate result.
TIP
In bright light situations use cell cover for zeroing and reading.
TIP
FILL, DIP, READ
TOTAL IRON
1
POWER ON PHOTOMETER
Press the button to power on the Well eXact® EZ.
Press and re-press the button until the
display shows FE2.
Page 7
7
Sensafe.com
FILL, DIP, READ
TOTAL IRON
REMOVE REAGENT STRIP
Remove one eXact® Strip Micro FE TPTZ and set in a dry, convenient place. Replace cap on bottle.
5
ZERO METER
Press and the LCD display reads 0.00PPM, indicating the meter is ready for testing.
6
DIP STRIP AND PRESS READ
Press to initiate a 20 second countdown and simultaneously DIP the eXact® Strip by submerging all pads in the sample then use a gentle, constant, back and forth motion (2 strokes per second) until the timer displays '1'. Remove and discard the strip. The display will automatically start to count up for 40 seconds.
7
READ RESULTS
READ result displayed as Total Iron. The result is
automatically stored in the FE3 menu. After testing is complete, rinse the sample cell immediately.
8
Page 8
8
©2017, Industrial Test Systems, Inc. Printed in USA
Use the included cleaning brush to clean the cell after each test to achieve the most accurate result.
TIP
FILL CELL
Rinse the cell 3 times with the water sample to be tested and
FILL to the top to begin test.
2
3
SELECT TEST
Press and re-press the button until the display shows No3.
ZERO METER
Press and the LCD Display reads 0.00
PPM
,
indicating the meter is ready for testing.
4
In bright light situations use cell cover for zeroing and reading.
TIP
FILL, DIP, READ
1
POWER ON PHOTOMETER
Press the button to power on the Well eXact® EZ.
NITRATE
Page 9
9
Sensafe.com
FILL, DIP, READ
NITRATE
REMOVE STRIP
Remove one eXact® Strip Micro NO3 Nitrate and set in a dry, convenient place. Replace cap on bottle.
5
READ RESULTS
READ result displayed as Total Nitrate (as NO
3
). This result is automatically stored in the No3 menu. After testing is complete, rinse the sample cell immediately and clean with brush.
7
DIP STRIP AND PRESS READ
Press to initiate a 20 second countdown and simultaneously DIP the eXact® Strip by submerging all pads in the sample then use a gentle, constant, back and forth motion (2 strokes per second) until the timer displays
‘1’
. Remove and discard the strip. The display
will automatically start to count up for 600 seconds.
6
Page 10
10
©2017, Industrial Test Systems, Inc. Printed in USA
3
SELECT TEST
Press and re-press the button until the display shows AL4.
Use the included cleaning brush to clean the cell after each test to achieve the most accurate result.
TIP
FILL CELL
Rinse the cell 3 times with the water sample to be tested and
FILL to the top to begin test.
2
ZERO METER
Press and the LCD Display reads 0
PPM
,
indicating the meter is ready for testing.
4
In bright light situations use cell cover for zeroing and reading.
TIP
FILL, DIP, READ
1
POWER ON PHOTOMETER
Press the button to power on the Well eXact® EZ.
TOTAL ALKALINITY
Page 11
11
Sensafe.com
FILL, DIP, READ
REMOVE STRIP
Remove one eXact® Strip Micro AL Total Alkalinity and set in a dry, convenient place. Replace cap on bottle.
5
READ RESULTS
READ result displayed as Total Alkalinity. This result is
automatically stored in the AL4 menu. After testing is complete, rinse the sample cell immediately.
7
DIP STRIP AND PRESS READ
Press to initiate a 20 second countdown and simultaneously DIP the eXact® Strip by submerging all pads in the sample then use a gentle, constant, back and forth motion (2 strokes per second) until the timer displays
‘1’
. Remove and discard the strip.
6
TOTAL ALKALINITY
If result reads "HI", repeat steps 5-7 with a Total Alkalinity Range Extender Strip (part no. 486665) until a numerical value is displayed; each strip adds 130 ppm to the fi nal result.
TIP
Page 12
12
©2017, Industrial Test Systems, Inc. Printed in USA
Use the included cleaning brush to clean the cell after each test to achieve the most accurate result.
TIP
FILL CELL
Rinse the cell 3 times with the water sample to be tested and FILL to the top to begin test.
2
1
POWER ON PHOTOMETER
Press the button to power on the Well eXact® EZ.
3
SELECT TEST
Press and re-press the button until the display shows TH5.
ZERO METER
Press and the LCD Display reads 0
PPM
,
indicating the meter is ready for testing.
4
In bright light situations use cell cover for zeroing and reading.
TIP
FILL, DIP, READ
TOTAL HARDNESS HIGH
Page 13
13
Sensafe.com
REMOVE STRIP
Remove one eXact® Strip Micro THH Total Hardness High and set in a dry, convenient place. Replace cap on bottle.
5
DIP STRIP AND PRESS READ
Press to initiate a 20 second countdown and simultaneously DIP the eXact
®
Strip by submerging all pads in the sample then use a gentle, constant, back and forth motion (2 strokes per second) until the timer displays
‘1’
. Remove and discard the strip.
6
READ RESULTS
READ result displayed as Total Hardness. This result
is automatically stored in the TH5 menu. After testing is complete, rinse the sample cell immediately and clean with brush.
7
FILL, DIP, READ
TOTAL HARDNESS HIGH
PH: The pH of the sample should be between 6.9 and 7.2. If the pH is not in this range, adjust the sample to pH 7 prior to testing THH.
AL: Total Alkalinity must be in the range of 65-220ppm. If Total Alkalinity is >220ppm, it must be conditioned with eXact
®
Strip
Micro THH Conditioner (part #486666) prior to testing THH.
TIP
Page 14
14
©2017, Industrial Test Systems, Inc. Printed in USA
FOR BEST ACCURACY
The Well eXact® EZ photometer has a 5 minute auto-shutoff timer.
Each test menu can store 20 results. To retrieve the stored results, go to the desired test using the MENU key. When the desired test is displayed, press and hold down the MENU key. Continue holding down the MENU key to scroll the stored results for that test, starting with the most recent result. The meter will display, from memory, the last 20 readings in sequence beginning with -20, which is the latest result, followed by -19, which is the 2nd latest result, etc; and fi nally -01, which is the oldest result retained. Only the last 20 readings are stored in each menu. This meter is able to store 100 results in memory (20 in each menu).
Before testing, rinse the meter sample cell with the sample water 3 times.
Always fi ll the cell to capacity (4mL); be careful not to splash liquid over the side.
Test immediately after fi lling the cell with the water sample.
To obtain optimal accuracy when testing outdoors (sunlight), use the cell cover when zeroing and reading the sample.
Due to the strip slitting process, you may fi nd one or two strips that are noticeably smaller or larger in width than the normal strips in the bottle. These should be discarded. Using these strips may give unreliable results.
Meter is not compatible for use with powder pillows, tablets, or liquids from other manufacturers.
Dip strip for entire countdown.
Each eXact® Strip Micro is valid for ONLY one test. Discard strip after use.
Dry the outside of the meter before storage.
Remove batteries before storing for prolonged periods.
Store the meter and test materials out of direct sunlight and away from chemical storage areas.
Minimize exposure of meter and test reagents to heat above 90°F (32°C).
When installing batteries, verify the O-ring is still attached to the screw before tightening.
Cleaning the cell with water and brush after each test is recommended for best accuracy and prior to storage of unit.
TECHNICAL SUPPORT
Please visit
sensafe.com/photometers/well-exact-ez/
for the latest technical information and how-to-videos. For additional technical support, call (803) 329-9712.
Industrial Test Systems, Inc.
1875 Langston Street, Rock Hill, SC 29730 800-861-9712 or 803-329-9712 [email protected] | sensafe.com
ITS Europe, Ltd.
The UK Centre for Homeland Security Building 7, Chilmark, Salisbury, Wiltshire SP3 5DU UK +44 1722 717911 [email protected] | itseurope.co.uk
C
E
L
E
B
R
A
T
I
N
G
E
X
C
E
L
L
E
N
C
E
years
28
1989
2017
Manufacturing quality products since 1989
Page 15
15
Sensafe.com
TROUBLESHOOTING
Listed below are possible situations that may arise while testing. Please contact one of our knowledgeable customer service representatives if you require further assistance
.
Subject Cause Solution
Dim screen or no response from meter
Low battery
Replace batteries
“LO” on LCD while zeroing
Low battery Replace batteries Dirty cell Clean cell Cloudy sample Dilute sample or use fi lter Bad LED Contact ITS
“HI” on LCD while reading
Result above detection level
Re-run test to verify result
“LO” on LCD while reading
Result below detection level
Re-run test to verify result
WELL EXACT® EZ ACCURACY
All tests have been calibrated using certifi ed reference standards and analytical spectrophotometric methods. The Well eXact
®
EZ has been factory calibrated and will stay valid because of its exceptional quality. We are so confi dent in the Well eXact
®
EZ, we
offer an industry leading 2-year warranty.
We built the Well eXact
®
EZ to be easy, accurate and environmentally friendly. We have achieved this by utilizing our patented eXact
®
Strip Micro Technology, which uses 60% less water and chemistry than alternative methods. Instead of using a 10mL water sample, eXact
®
Strip Micro uses a 4mL water sample. The accuracy of the meter is maintained by designing the sample cell with an 11mm path-length.
Ready Snap® 3 is a calibration verifi cation solution with predetermined color to verify the accuracy of your
Well
eXact® EZ photometer.
The easy 3 step procedure (snap, fi ll, and test) allows for quick verifi cation of test parameters. Each box contains 10 ampoules of 10mL solution with no dilution necessary.
METHOD VERIFICATION
WELL EXACT® EZ ASSIGNED VALUE FOR READY SNAP® 3
READY SNAP® 3
TOTAL HARDNESS
DESIRED VALUE
TOTAL HARDNESS
ACCEPTABLE RANGE
Red dye # 505
426 ppm 358 - 494
Note: Values indicated are specifi c to Well eXact® EZ readings and current
concentrations as found at time of manufacture.
Page 16
16
©2017, Industrial Test Systems, Inc. Printed in USA
BUILT IN SAMPLE CELL
The built-in sample cell is made of transparent plastic; the sturdy cell design will last for over 20,000 readings. Our studies have shown that scratches on the cell will not compromise the accuracy of your results because of the cell’s fi xed position.
WARRANTY (2 YEARS)
Registration of your eXact® photometer must be received within 30 days from date of purchase to activate the warranty. Registration is available over the phone (+1-803-329-9712 Ext. 0) or online at
sensafe.com/micro/warranty/
(Personal data is kept confi dential). The eXact
®
photometer is warranted to be free from defects in materials and workmanship for a period of two (2) years from the date of purchase by the customer. ITS will repair or replace, at its discretion, product which is deemed to be faulty due to manufacturing defect. Warranty does not cover product damage caused by abuse (such as crushing a tablet in the cell), battery corrosion damage, or improper use. If the meter is faulty or otherwise defective contact ITS by phone (+1-803-329­9712 Ext. 0) or email ([email protected]) to describe the problem and obtain a return authorization form before returning the photometer to ITS. Damage caused by improper packing of the photometer for return shipment to ITS will not be covered by the warranty. Customer is responsible for shipping charges to ITS. ITS pays postage when photometer is returned to customer. A maximum processing fee of $75 will be charged for repair or replacement of non-registered photometers and damages not covered by this warranty. The repair or replacement of the photometer will not extend or renew the period of guarantee. This warranty does not affect your statutory rights. The warranty is not transferable.
ITEM
INCLUDES PART NO.
Standard Carrying case
Blue carrying case with foam inserts (holds up to 7 bottles)
486111
XL Carrying case
Black carrying case with foam inserts (holds up to 14 bottles)
486001
Made of sturdy material lined with foam, the carrying case offers storage for a Well eXact® EZ photometer and eXact® Micro reagents.
CARRYING CASE FOR WELL EXACT® EZ
Store all your necessary reagents together with your Well eXact
®
EZ in our convenient carrying case!
TIP
ABOUT
Page 17
17
Sensafe.com
MSDS
Section 1 Chemical Identifi cation / Company / Undertaking
Name: EZ Open Reducer Powder Pillow For Total Iron Test Product code: 486601 used in kit 486651 Company Name: Industrial Test Systems, Inc.
1875 Langston Street, Rock Hill, SC 29730 USA
Phone: (800) 861-9712, (803) 329-9712, Fax: (803) 329-9743 eMail: [email protected]
Section 2 Composition / Information on Ingredients
Hazardous Ingredients: Sodium Hydrosulfi te (CAS 7775-14-6) 8-13% Sodium Thiosulfate (CAS 10102-17-7 or 7772-98-7) 16-22% Sodium Metabisulfate (CAS 7681-57-4) 13-21% Sodium Chloride (CAS 7757-82-6) 31-50% Sodium Dihydrogen Citrate (CAS 18996-35-5) 9-15%
Section 3 Hazards Identifi cation
Main Hazards: Harmful if swallowed. Irritating to eyes, respiratory system and skin. Contact with acids liberates toxic gases. Harmful to aquatic organisms
Section 4 First Aid Measures (Symptoms)
Skin Contact: There may be mild irritation at the site of contact. Eye Contact: There may be irritation and redness. Ingestion: There may be soreness and redness of the mouth and throat.
There may be diffi culty swallowing. Nausea and stomach pain may occur. There may be vomiting. Inhalation: Absorption through the lungs can occur causing symptoms similar to those of ingestion.
Section 4 First Aid Measures (Action)
Skin Contact: Wash immediately with plenty of soap and water. Eye Contact: Flush eyes with running water for 15 minutes. Call
physician. Ingestion: Wash out mouth with water. Do not induce vomiting. If conscious, drink 1-2 glasses of water immediately. Call physician immediately. Inhalation: Remove to fresh air, give artifi cial respiration if necessary. Call physician.
Section 5 Fire Fighting Measures
Extinguishing media: Suitable extinguishing media for the surrounding
fi re should be used.
Exposure hazards: In combustion emits toxic fumes. Protection of fi re-fi ghters: Wear self-contained breathing apparatus.
Wear protective clothing to prevent contact with skin and eyes.
Section 6 Accidental Release Measures
Personal precautions: Refer to section 8 of MSDS for personal protection details. Do not create dust. Mark the contaminated area with signs and prevent access to unauthorised personnel. If outside do not approach from downwind.
Environmental precautions: Do not discharge into drains or rivers. Environmental precautions: Transfer to a closable, labeled salvage
container for disposal by an appropriate method.
Page 18
18
©2017, Industrial Test Systems, Inc. Printed in USA
MSDS
Section 7 Handling and Storage
Handling requirements: Ensure there is exhaust ventilation in the area. Avoid the formation or spread of dust in the air. Avoid direct contact with the substance. Storage conditions: Store in cool, well ventilated area. Keep container tightly closed. Practices: Maintain general industrial hygiene practices when using this product.
Section 8 Exposure Controls / Protective Equipment
Hand protection: Protective gloves. Skin Protection: Protective clothing. Eye Protection: Safety glasses. Ensure eye wash station is available. Respiratory protection: Respiratory protective device with particle fi lter. Engineering measures: Ensure there is adequate ventilation or dust
mask.
Section 9 Physical and Chemical Properties
State: Powder Solubility in water: Soluble Odor: Unpleasant Color: Off-white Melting point/range°C: >400 pH: 5.3 (5% solution)
Section 10 Stability and Reactivity
Stability: Stable under normal conditions. Conditions to avoid: Heat. Materials to avoid: Oxidising agents and acids. Haz. decomp. products: In combustion emits toxic fumes.
Section 11 Toxicological Information
Engineering Measures: EZ Open Reducer Powder Pillow For Total Iron Test Respiratory protection: Refer to section 4 of MSDS for routes of exposure and corresponding symptoms.
Section 12 Ecological Information
Ecotoxicity operations: EZ Open Reducer Powder Pillow For Total Iron Test
Mobility: Soluble in water. Persistence and degradability: No data available. Bioaccumulative potential: No data available. Other adverse effects: Harmful to aquatic organisms.
Section 13 Disposal Considerations
Disposal operations: Dissolve or mix the material with a combustible solvent and burn in a chemical incinerator equipped with an afterburner and scrubber. Observe all federal, state and local regulations. Disposal of packaging: Burn in a chemical incinerator equipped with afterburners and scrubbers. NB: The user’s attention is drawn to the possible existence of regional or national regulations regarding disposal.
Section 14 Transportation Information
UN no: - Shipping name: “Not subject to ADR”
ADR / RID IMDG / IMO IATA / ICAO
UN no: - UN no: - UN no: -
Page 19
19
Sensafe.com
MSDS
Section 15 Regulatory Information
Hazard symbols: Harmful Risk phrases: R22: Harmful if swallowed R36/37/38: Irritating to eyes, respiratory system and skin R31: Contact with acids liberates toxic gas. R52: Harmful to aquatic organisms.
Safety phrases: S22: Do not breathe dust. S36/37/39: Wear suitable protective clothing, gloves and eye / face
protection. S26: In case of contact with eyes, rinse immediately with plenty of water and seek medical advice. S45: In case of accident or if you feel unwell, seek medical advice immediately (show the label where possible) S61: Avoid release to the environment. Refer to special instructions / safety data sheets. NOTE: The regulatory information given above only indicates the principal regulations specifi cally applicable to the product described in the material safety data sheet. The user’s attention is drawn to the possible existence of additional provisions which complete these regulations. Refer to all applicable national, international and local regulations or provisions.
Section 16 Other Information
Other Information: Amended by Scafell Organics Risk phrases used in s.2:
R22: Harmful in swallowed. R36/37/38: Irritating to eyes, respiratory system and skin. R31: Contact with acids liberates toxic gas. R52: Harmful to aquatic organisms.
Legal disclaimer: The above information is believed to be correct but does not purport to be all inclusive and shall be used only as a guide. This company shall not be held liable for any damage resulting from handling or from contact with the above product
Page 20
†
Value provided represents best possible accuracy under laboratory conditions, but may vary throughout the detection
range.
For a complete list of accuracies throughout all ranges, please visit sensafe.com/photometers/well-exact-ez/.
1
For detailed instructions on using this product, contact us (see below) or visit sensafe.com/photometers/well-exact-ez
For resellers and distributors - products sold in case quantities (12 units per case).
PATENT INFORMATION
US Patent #7,333,194; Euro Pat No. 1 725 864 DE FR UK; South Africa Pat No 2007/0628 by Industrial Test Systems, Inc., 1875 Langston Street, Rock Hill, SC USA. EXACT
®
is a registered trademark of Industrial Test Systems, Inc. Rock Hill, SC USA.
EXACT® MICRO TESTS & REAGENTS
R031017-WEZQG
Visit us online:
sensafe.com/photometers/well-exact-ez/
for up–to–date product information.
SENSAFE ITSSENSAFE
Industrial Test Systems, Inc. 1875 Langston Street, Rock Hill, SC 29730 1–800–861–9712—Inside the U.S.A. 1–803–329–9712—Outside the U.S.A. Fax:1–803–329–9743 Email: [email protected] WWW.SENSAFE.COM
ITS EUROPE, LTD The UK Centre for Homeland Security Building 7, Chilmark Salisbury, Wiltshire, SP3 5DU, UK Phone: +44 (0)1722 717911 Fax: +44 (0) 1722 717941 Email: [email protected] WWW.ITSEUROPE.CO.UK
MANUFACTURED BY
PARAMETER / TEST
PART #
RANGE
ppm
% BEST †
ACCURACY
# OF
TESTS
Alkalinity, Total 486641 10 - 200 10 100
Alkalinity, Total Range Extender
1486665 Each strip adds 130 ppm 100
Hardness, Total High 486656 60 - 600 12 50
Hardness, Total Conditioner 1486666
Use when Alkalinity >200 ppm
50
Iron, Total 486650 0.03 - 8 8 50
Nitrate 486655 0.25 - 32 ppm 15 50
pH 486639-II 6.4 - 8.4 pH 0.2 pH 100
EXACT REFILL BOX
PART #
PARAMETER / TESTS
eXact® Well Refi ll Box 486212
Total Alkalinity, Total Hardness High, Total Iron, Nitrate, and pH
ADVANCED PHOTOMETER SYSTEM
Well