Eutech Instruments αlpha-CON560 Instruction Manual

gyM
sy...
/
µ
r
Instruction Manual
αlpha-CON560
Conductivity Controller
mS
S/ºC/ºF mete
Tec hno lo
adeEa
68X216878 Rev 3 Aug 07
Preface
This manual serves to explain the use of the αlpha-CON 560 Industrial Conductivity Controller. The manual functions in two ways, firstly as a step by step
guide to help the user operate the instrument. Secondly, it serves as a handy reference guide. This instruction manual is written to cover many anticipated applications of the αlpha-CON 560 controller. If you have doubts in the use of the instrument, please do not hesitate to contact the nearest Eutech Instruments Authorised Distributor.
The information presented in this manual is subject to change without notice as improvements are made, and does not represent a commitment on part of Eutech Instruments.
Eutech Instruments cannot accept any responsibility for damage or malfunction of the unit due to improper use of the instrument.
Copyright ©2006 All rights reserved.
Eutech Instruments Pte Ltd.
Table of Contents
1. Introduction ...........................................................................................................1
1.1 Before You Begin ................................................................................................................... 1
1.2 Intended Use .......................................................................................................................... 1
1.3 Safety Instructions.................................................................................................................. 2
1.4 Taking Out of Service / Correct Disposal of the Unit ............................................................. 2
2. Getting Started ......................................................................................................3
2.1 Description of Instrument ....................................................................................................... 3
2.2 Measurement and Control System......................................................................................... 4
2.3 Connecting Peripherals.......................................................................................................... 5
2.3.1 Connection Terminals.................................................................................................... 5
2.3.2 Switching Between Pt100 & Pt1000 Temperature Sensors.......................................... 6
2.3.3 Connecting Conductivity Probe ..................................................................................... 7
2.3.4 Connecting Temperature Probe .................................................................................... 7
2.4 Installation .............................................................................................................................. 8
2.4.1 Mechanical Dimensions................................................................................................. 8
2.4.2 Wall Mount..................................................................................................................... 8
2.4.3 Panel Mount................................................................................................................... 9
2.5 Display & Keypad ................................................................................................................. 10
2.5.1 Display Overview......................................................................................................... 10
2.5.2 Key Functions.............................................................................................................. 11
3. Operation .............................................................................................................13
3.1 Measurement mode ............................................................................................................. 13
3.2 Menu Overview .................................................................................................................... 14
4. Calibration Mode .................................................................................................15
4.1 About Calibration.................................................................................................................. 15
4.2 Calibration ............................................................................................................................ 16
5. Setup Mode ..........................................................................................................19
5.1 Enter Setup mode ................................................................................................................ 19
5.2 Temperature Coefficient Settings......................................................................................... 20
5.3 Temperature Settings........................................................................................................... 21
5.4 Relay 1/Relay 2 Settings...................................................................................................... 23
5.5 Controller Settings................................................................................................................ 25
5.6 Measuring Range Settings................................................................................................... 27
5.7 Configuration Settings.......................................................................................................... 28
5.8 Viewing Electrode Properties ............................................................................................... 30
6. Technical Specifications.....................................................................................31
7. List of Accessories .............................................................................................33
8. Troubleshooting ..................................................................................................34
9. General Information ............................................................................................35
9.1 Warranty............................................................................................................................... 35
9.2 Return of Goods ................................................................................................................... 35
9.3 Guidelines for Returning Unit for Repair .............................................................................. 35
10. Appendices .......................................................................................................36
10.1 Appendix 2 Conductivity of various aqueous solutions....................................................... 36
10.2 Appendix 2 Graphical Representation of the Function of Hysteresis ................................. 37
10.3 Appendix 3 General Instructions Concerning Controller Setting ........................................ 38
10.4 Appendix 4 - Abbreviations Used in LCD............................................................................. 39
Instruction Manual αlpha-CON 560
1. Introduction

1.1 Before You Begin

We thank you for purchasing the αlpha CON 560 controller.
The construction of the CON 560 controller employs leading edge technology and complies with safety regulations currently in force. Notwithstanding this, improper use could lead to hazards for the user or a third-party, and/or adverse effects on the plant or other equipment. Therefore, the operating instructions must be read and understood by the persons involved before working with the controller.
The instruction manual must always be stored close at hand, in a place accessible to all people working with the conductivity controller.
If you have questions, which are not or insufficiently answered in this instruction manual, please contact your authorized supplier. They will be glad to assist you.

1.2 Intended Use

αlpha CON 560 controllers are intended solely for conductivity and temperature measurement, as described in this instruction manual.
Any other use, or use not mentioned here, that is incompatible with the technical specifications is deemed inappropriate. The operator is solely responsible for any damage arising from such use.
Other prerequisites for appropriate use include:
– Comply with the instructions, notes and requirements set out in this
instruction manual.
Comply with all local safety regulations concerning safety at work.
Comply with all information and warnings in the documentation dealing
with the products used together with the conductivity controller (housings, sensors, etc.).
– Comply with local environmental and operational conditions.
1
Instruction Manual αlpha-CON 560

1.3 Safety Instructions

- The αlpha CON 560 controller should be installed and operated only by personnel familiar with the instrument and who are qualified for such work.
- A defective CON controller must neither be installed nor put into service.
- The αlpha CON 560 must only be operated under the specified operating conditions (see section 6).
- The αlpha CON 560 must not be repaired by the customer.
- No modifications to the αlpha CON 560 are allowed. The manufacturer/supplier accepts no responsibility for damage caused by unauthorized modifications. The risk is borne entirely by the user.

1.4 Taking Out of Service / Correct Disposal of the Unit

Taking out of Service
First disconnect the unit from the power supply and then undo all electrical connections.
Remove the unit from the wall / panel.
Correct Disposal of the Instrument
When the αlpha CON 560 controller is permanently taken out of service, obey the local environmental regulations for correct disposal or send the instrument to your local distributor, they will take care of proper disposal.
2
Instruction Manual αlpha-CON 560

2. Getting Started

2.1 Description of Instrument

The Eutech Instruments’ αlpha CON 560 controller is used for measuring conductivity and temperature values. The controller can be used for applications such as water treatment and controlling, galvanic­decontamination, chemical processing, food processing, clean or wastewater control and neutralization processes.
The controller has many user-friendly and safety features which include:
Push-button keypad for calibration and setup
Built-in non-volatile memory to ensure that calibration and other
information are not erased if power supply fails
Menu-driven program that simplifies setup
Automatic temperature compensation (ATC)
Manual temperature compensation setting without the ATC probe, with
independent setting for calibration and process temperature
Large dual display LCD for easy reading with clear multiple annunciators,
operational mode indicators and error indicators
Two switching contacts as set-point relays
Limit (on/off) control mode
Separately adjustable high and low set-point hysteresis (dead bands) to
prevent chattering of relays around the set points
0 to 2000 seconds time delay adjustment on all relays – minimizes false
alarms
Backlight LCD display
3
Instruction Manual αlpha-CON 560
A

2.2 Measurement and Control System

A typical measurement system consists of:
A conductivity process controller
A conductivity sensor with integrated or separate Pt100/Pt1000
temperature sensor
An appropriate measurement cable
An immersion, flow or process assembly
A final control element such as pump or valve
lpha CON 560 Controller
Pump or Valve
lpha pH560
ESC
CAL
pH M onitor
ENT
Relay Output
Power Adaptor
Housing and Sensors
Measurement Cable
(+9 V DC)
4
Instruction Manual αlpha-CON 560

2.3 Connecting Peripherals

2.3.1 Connection Terminals

Remove Back Cover:
Remove the screws from the four corners at the back of the conductivity controller. Remove the back cover. The connectors are exposed on the back PCBA as shown in the Figure 1 below.
Connectors:
J8 – 9V DC power connector (1.2mm diameter)
J9 – Relay connections
J10 - Conductivity electrode & Temperature probe connections (wiring has
to be done in the detachable connector
J10 Connections
1. CON Low Current
2. CON Low Voltage
3. CON High Voltage
4. CON High Current
5. Pt 100 Compensate
6. Pt 100 Sense
7. Pt 100 GND
8. (No connection)
J9 Connections
Relay 1 & 2
J8
9V DC Power
Screw
J9
REL 1
J10
REL 2
Figure 1: Outer Side of Back PCBA
5
Instruction Manual αlpha-CON 560

2.3.2 Switching Between Pt100 & Pt1000 Temperature Sensors

The controller supports both Pt100 & Pt1000 (2-wire or 3-wire) temperature sensors. The default factory setting is Pt100. If you need to use Pt1000 temperature sensor, you have to change the jumper setting (J7) as described below.
Remove Back Cover:
Remove screws from the four corners at the back of the conductivity controller. Remove the back cover.
Remove Back PCBA:
Remove the screw located center of the back PCBA (
Figure 1). Detach the back PCBA from the controller. Turn over the back PCBA. Locate J7 jumper on the inner side of the back PCBA as shown in Figure 2 below.
J7
Jumper Setting for
Pt100
(Factory Default)
Jumper Settings for
Pt1000
Figure 2: Inner Side of Back PCBA
Set Jumper J7:
Set the J7 jumper to required sensor (Pt100 or Pt1000) type.
6
Instruction Manual αlpha-CON 560

2.3.3 Connecting Conductivity Probe

Any industrial conductivity probe (2-cell or 4-cell) can be connected to the conductivity controller.
2-Cell Conductivity Probe:
1. Short Pin 1 & 2 of J10 connector using a jumper wire.
2. Connect CON low input wire (white) to Pin 2 of J10 connector.
3. Connect CON high input wire (black) to Pin 3 of J10 connector.
4. Short Pin 3 & 4 of J10 connector using a jumper wire.
4-Cell Conductivity Probe:
1. Connect CON low input wire (black) to Pin 1 of J10 connector
2. Connect CON low input wire (white) to Pin 2 of J10 connector.
3. Connect CON high input wire (orange) to Pin 3 of J10 connector.
4. Connect CON high input wire (blue) to Pin 4 of J10 connector.

2.3.4 Connecting Temperature Probe

For Automatic Temperature Compensated (ATC) conductivity readings, an in-built temperature sensor is usually integrated with industrial conductivity probes. Alternatively, a separate 100Ω Pt RTD temperature probe (2-wire or 3-wire) can be connected to the controller.
3-Wire Probe:
1. Connect PT100 compensate wire to Pin 5 of J10 connector.
2. Connect PT100 sense wire to Pin 6 of J10 connector.
3. Connect PT100 GND wire to Pin 7 of J10 connector.
2-Wire Probe:
4. Short Pin 5 & 6 of J10 connector using a jumper wire.
5. Connect PT100 sense wire to Pin 6 of J10 connector.
6. Connect PT100 GND wire to Pin 7 of J10 connector.
7
Instruction Manual αlpha-CON 560

2.4 Installation

2.4.1 Mechanical Dimensions

lpha CON560
ESC
CAL

2.4.2 Wall Mount

ENT
Conductivity Monitor
Cover the catch slots at
2
both sides with overlays
Pierce through holes at
11
both sides
8
Instruction Manual αlpha-CON 560

2.4.3 Panel Mount

Prepare panel cut-out of
1
1
92.0 mm X 92.0 mm
Remove back cover of
2
conductivity controller and slide it through panel cut-out
Gasket
Panel
Panel
Insert threaded rods through catch
4
until conductivity controller is held against panel
Attach catch to both sides
3
of conductivity controller
9
Loading...
+ 33 hidden pages