EUTECH INSTRUMENTS PCSTestr35, PCTestr35, PTTestr35 Quick Manual

Page 1
QUICK GUIDE
MULTI-
PARAMETER
TESTER 35
pH/ Conductivity/ TDS/
Salinity/ Temperature
PCSTestr35, PCTestr35, PTTestr35
Technolo
gyM
adeEa
sy...
Part of Thermo Fisher Scientific
68X441601 Rev 0 Jan’08
Page 2
Thank you for purchasing the multi-parameter tester. Please use the following instructions to set up and calibrate your tester before taking measurements.
There are 3 stages to execute before using your meter:
STAGE 1 : Conditioning and temperature calibration
STAGE 2: Setup the system and the parameters
STAGE 3 : Calibrate the parameters
BEFORE YOU BEGIN
you must first load batteries, condition the tester and calibrate the temperature.
BATTERIES
Twist and unscrew the top cover and insert the 4 batteries (LR44 1.5V) as shown in this picture. Note the polarity.
2
Page 3
3
STAGE 1
CONDITIONING
Condition your tester by immersing it in tap water for at least 30 minutes before use.
DO NOT use de-ionized water.
CALIBRATING THE TEMPERATURE
Before taking measurements, it is important to calibrate your tester to read accurately. This procedure is only done once in a long while as the calibration for temperature will remain accurate unless used under severe conditions.
FOLLOW THESE STEPS:
1. Press the ON key
The tester is now in Measurement mode.
2. Place a standard thermometer in the sample for reference. If using a mercury thermometer, allow some time to reach correct reading.
3. Then place tester into the same sample. Allow some time for sensor to stabilize.
4. Press MODE to select the pH measuring mode. Then press the CAL key.
Page 4
5. Press the HOLD key for 5 seconds and wait till screen
changes.
6. Use the c and d keys to manually adjust until you reach the correct temperature of your sample according to the reference thermometer.
7. Press ENT to confirm and return to the PH measuring mode.
NOTE: Once calibrated, you don’t have to do this for any other parameter as temperature is common to all.
CHANGING THE SENSOR
Rot ate co ll ar
away from you
Ribbed collar
After twisting the ribbed collar off, hold the electrode/sensor module and pull straight out.
4
Page 5
Then replace the new electrode/sensor, ensuring the pins are lined up to the connector before inserting.
5
Insert Electrode Module
Electrode Module
Large O-Ring
Smal l O -Rin g S mal l Ta b Lar g e Tab
STAGE 2 : SET UP
1. Turn your tester off. Then
hold the ENT key down and press ON/OFF to enter the Setup mode.
2. The display will show
SEt UP
SEt
UP
Then it will automatically change to
PArA
3. Use the c and d keys to select
PArA
Or
4. Press ENT to confirm.
SySt
Page 6
Select SYST to enter the main menu for general settings.
OR
Select PARA to set up the various parameters – pH, COND, TDS, Salinity.
SAVING THE SET UP
After any manual changes have been done during set up, you are required to save the setting.
1. Press both c and d keys at the same time and release.
2. Press again until you see
SA YES
3. Use either c OR d key to toggle between YES and NO. Selecting YES will save your changes/ calibrations.
NOTE: If you did NOT make changes or recalibrate, this screen will NOT appear.
6
Page 7
SYSTEM SETUP
After selecting SYST, you will see the display
Unit rST
Use the c and d keys to switch to other settings and press ENT to select :
UNit
rST
Use this to reset the calibration or factory settings.
Use this to turn on or off the AUTO OFF function.
SET
A.off
Use this to set the temperature coefficient.
SET
TC
Set ATC
Use this to turn on or off the Automatic Temperature Compensation.
Use this to set the temperature unit in degrees Celsius or Fahrenheit.
Set °C°F
7
Page 8
Resetting previous calibrations or factory settings
1. When the above screen appears, press ENT to select.
2. Use the c and d keys to
toggle between
and
3. To reset previous calibrations in the pH parameter, select
by pressing ENT.
Then select by pressing ENT to reset your previous settings
OR
use the c and d keys to select to reset the factory settings.
Press ENT to confirm and return to
4. To reset calibrations for all other parameters, press ENT to select and follow
the same instructions as point number 3 above.
8
rst unit
EC rST
FCT rST
CAL rST
PH rST
EC rST
PH rST
UNit RST
Page 9
Turning the AUTO OFF function on or off
SET
A.off
1. Use the c and d keys from
UNit rst
and find
Set a. off
2. Press ENT to select it.
Use the c and d keys to switch between
9
and
NO a. off
YES a. off
Selecting YES means your tester will turn OFF automatically if it is not used after a 8.5 minutes.
Setting the Temperature Coefficient
SET
TC
1. Use the c and d keys from
UNIT
RST
and find
SET
TC
Page 10
2. Press ENT to select it. The standard coefficient used is
2.1. Press ENT if this is correct for your sample. If not, use the c and d keys to manually set the coefficient, then press ENT when done.
Turning on or off the Automatic Temperature Compensation
1. Use the c and d
keys from
and find
2. Press ENT to select it.
3. Use the c and d keys to
select either
OR
Then press ENT.
10
Selecting YES will activate the automatic temperature compensation function. Selecting
NO will activate manual
function.
NO ATC
YES ATC
Set ATC
UNit
RSt
Set ATC
Page 11
Setting up the temperature unit of measurement
Set °C°F
1. Use the c and d keys from
UNit
RST
and find
Set °C°F
2. Press ENT to select it.
3.Use the c and d keys to
select either
OR
11
°C
SET
Press ENT when done.
°F
SET
4. Save the changes as described on page 6.
-END OF SYSTEM SETUP –
PARAMETER SETUP
1. Return to
and press ENT.
SEt
UP
2. Then use the c and d keys to select
PArA
and Press ENT.
Page 12
You will see the
PH
displayed on the screen first.
3. Use the c and d keys to scroll to COND, TDS or SALT parameter.
4. Press ENT to select the parameter you want to set up.
12
Setting up the pH Buffer
PH
1. At the above display, press ENT.
2. Use the c and d keys to
select either
OR
Nist
BUFF
USA
BUFF
Then press ENT to confirm.
3. Again, use the c and d keys
to select either
OR
Five-point calibration or Three point calibration.
5-PT
CAL
3-pt
CAL
4. Then press ENT to confirm. This will bring you back to
PH
Page 13
CoND
Setting up the Conductivity Parameter
1. From press ENT.
PArA
Then use c and d keys to select
CoND
2. Press ENT.
3. Then use c and d keys to
select either
OR
13
YES
A.CAL
NO
A.CAL
Selecting YES means the automatic calibration will be activated. Press ENT to confirm.
4. Then use c and d keys to select either
OR
YES
SPC
NO
SPC
Selecting YES means the Single Point Calibration will be activated. Press ENT to confirm and return to the main menu.
Selecting NO means multipoint calibration will be activated.
Page 14
14
Setting up the TDS Factor
TDS
1. From press ENT.
PArA
Then use c and d keys to select
TDS
2. Press ENT.
3. Use the c and d keys to
change the TDS value. For example:
0.67 FACT
4. Press ENT to confirm and return to main menu.
Setting up the Salinity parameter
SALT
1. From press ENT,
PArA
then use c and d keys to select
SALT
2. Press ENT.
Page 15
3. Then use c and d keys to select either
OR
15
PPT UNIT
PER UNIT
to measure in Parts Per Thousand as PPT or Percentage as PER.
4. Press ENT to confirm and return to the main menu.
5. Save the changes as described on page 6.
-END OF PARAMETER SETUP-
STAGE 3: CALIBRATING THE PARAMETERS
Calibrating PH
1. In the measurement mode at press the CAL key.
PH
2. Place the tester in the buffer solution. The bottom reading will show the correct buffer value.
3. When the reading has stabilized, for instance: 3.87 for a 4.01 buffer, press ENT.
Page 16
4. To continue with another buffer calibration, simply remove tester and place in another buffer solution.
5. Press ENT again to confirm the buffer value when it has stabilized.
6. When calibration is done, press the CAL key to confirm.
Calibrating COND
COND
1. At the display press the CAL key.
NOTE: In your set up for COND, the tester must be in AUTO CAL mode. NOT manual CAL.
2. Place the tester in the buffer solution.
3. Follow the same instructions as above for PH calibration – Points 3 to 6. Press CAL key to confirm.
For manual calibration, place tester in the buffer solution then do the following:
16
Press c and d keys to scroll to the calibration standard value
Page 17
and press ENT. The tester will return to measurement mode.
Calibrating TDS
1. At the display press the CAL key.
2. Place the tester in the buffer
solution.
3. Press c and d keys to scroll to the calibration standard value and press ENT. The tester will return to measurement mode.
Calibrating SALT
1. At the display press the CAL key.
2. Place the tester in the buffer
solution.
3. Press c and d keys to scroll to the calibration standard value and press ENT. The tester will return to measurement mode.
END OF CALIBRATION SECTION
SALT
TDS
17
Page 18
18
HOLD FUNCTION
To stop measurements and lock on to a reading, press the HOLD key. The tester will not change the last reading on the screen until the HOLD key is pressed again.
This function is useful when taking measurements in a dim environment. It allows you to hold the reading until you are able to move to a well-lit area.
MEASUREMENT
1. To start measuring, press the ON/OFF key.
2. Press MODE/ ENT until you
get to your desired parameter.
3. Place tester in your sample and it will start measuring immediately.
Page 19
19
Warranty
The waterproof testers are warranted to be free from manufacturing defects for 1 year and the electrode module is warranted for 6 months, unless otherwise stated. If repair, adjustment or replacement is necessary and has not been the result of abuse or misuse within the time period specified, please return the tester – freight prepaid – and correction will be made without charge. Out of warranty products will be repaired on a charge basis.
Return of Items
Authorization must be obtained from your distributor before returning items for any reason.
When applying for authorization, please include information regarding the reason the item(s) are to be returned.
Note: We reserve the right to make improvements in design, construction and appearance of products without notice. Prices are subject to change without notice.
Page 20
20
For more information on Eutech Instruments’/ Oakton Instruments’ products, contact your nearest distributor or visit our website listed below:
Oakton Instruments
P.O Box 5136, Vernon Hills, IL 60061, USA Tel: (1) 888-462-5866 Fax: (1) 847-247-2984 E-mail:
Web-site: www.4oakton.com
Eutech Instruments Pte Ltd.
Blk 55, Ayer Rajah Crescent, #04-16/24 Singapore 139949 Tel: (65) 6778 6876 Fax: (65) 6773 0836 E-mail:
Web-site: www.eutechinst.com
Loading...