3
1.2 Connecting the Electrode and Temperature Probe
To connect the pH or ORP electrode:
The Eutech Instruments pH transmitter uses any
standard pH or ORP.
Remove BNC connector from cable. Two wires
will be exposed:
Strip insulation, just enough to expose bare wires.
IMPORTANT: For pH/ORP sensing cable, remember to strip the inner black
layer (screen) and expose the clear plastic sheath.
Insert pH sensing cable into pin 9 of connector. Tighten screw.
Insert pH reference cable into pin 7 of connector. Tighten screw.
To connect temperature probe:
For Automatic Temperature Compensation (ATC), this transmitter requires a
temperature probe, Pt 100, 3-wire.
3-wire probe:
Insert brown wire to pin 4 of connector. Tighten screw.
Insert red wire to pin 5 of connector. Tighten screw.
Insert green wire to pin 6 of connector. Tighten screw.
2-wire probe:
Take a small length of wire and short pins 4 and 5 of connector. Insert red wire into
pin 5 of connector and tighten screws.
Insert green wire to pin 6 of connector. Tighten screw.
IMPORTANT: Temperature probe should always be together with pH probe,
even during calibration. Make sure no bare wires are exposed and all screws
are tightened for proper contact.
To connect Potential Matching Pin (Liquid electrical interferences):
Connect the additional wire from the electrode to pin 3. Transmitter must be
configured to ‘Symmetrical mode of operation’ - Please refer to ‘Advanced Setup
Functions’, Section 6.6.2 - ‘Selecting Asymmetrical or Symmetrical Mode of
Operation.
Insulation
pH reference
Clear sheath
pH sensing
Screen