This manual serves to explain the use of the hand-held series meters. The models
covered are the pH 11 and the pH 110 hand-held meters.
The manual functions in two ways, firstly as a step by step guide to help the user
operate the meter. Secondly, it serves as a handy reference guide.
This instruction manual is written to cover as many anticipated applications of the pH
meters as possible. If there are doubts in the use of the pH 11/110 meters, do not
hesitate to contact the nearest Authorised Distributor.
It is recommended that all operators should read this manual prior to working with
this instrument. Eutech Instruments / Oakton Instruments cannot accept any
responsibility for damage or malfunction to the meter caused by improper use of the
instrument.
The information presented in this manual is subject to change without notice as
improvements are made, and does not represent a commitment on the part of Eutech
Instruments Pte Ltd / Oakton Instruments.
5.3.1 Taking measurements with READY indicator selected on............................................................ 19
5.3.2 Taking measurements with the AUTO HOLD feature selected on................................................ 19
6 HOLD FUNCTION..................................................................................................................................... 20
9.6.2 User Reset..................................................................................................................................... 43
11.1.2 After Use ....................................................................................................................................... 57
19 RETURN OF ITEMS ............................................................................................................................. 71
Instruction Manual CyberScan pH 11 / 110
1 INTRODUCTION
Thank you for selecting the pH 11/110 meter. These meters are microprocessorbased instruments and are designed to be handy, capable of allowing one-hand
operation. Each has a large custom dual LCD for clear and easy reading. It is a
unique and intelligent instrument that has the capability to cater to the preferences of
the discerning individual. You have one of the two models:
• pH 11 meter
• pH 110 meter
Both meters have many user-friendly features – all of which are completely
accessible through the splash-proof membrane keypad. Your meter includes a
temperature probe, electrode holder, built-in meter stand and batteries. Eutech
Instruments/ Oakton Instruments offer a wide selection of pH and ORP electrodes.
Refer to Section 15 LIST OF ACCESSORIES on page 65 for more information.
The basic model is the pH 11 which is capable of measuring pH, Temperature, and
millivolt (mV).
The deluxe model is the pH 110 which measures pH, Temperature, millivolt (mV) and
relative millivolt (Rel mV). It has many advanced features and allows you to
customise the meter settings. It also has a RS232C port that allows the meter to be
connected to a computer or a printer via a cable for transferring data.
For power requirements, you can either use 4 AAA-sized batteries or an AC/DC
power adapter (sold separately).
Please read this manual thoroughly before operating your meter.
1
Instruction Manual CyberScan pH 11 / 110
2 DISPLAY & KEYPAD FUNCTIONS
2.1 Display
The LCD has a primary and secondary display.
• The primary display shows the measured pH or mV values.
• The secondary display shows the measured temperature.
The display also shows error messages, keypad functions and program functions.
10. ATC - Automatic
Temperature Compensation
indicator
11. MEM - Memory location
indicator
2
16. ON – READY/Auto HOLD set
up enable indicator.
OFF – READY set up disable
indicator
17. HOLD – Hold indicator
Instruction Manual CyberScan pH 11 / 110
2.2 Keypad
A large membrane keypadmakes the instrument easy to use. Each button, when
pressed, has a corresponding graphic indicator on the LCD (Figure 2 on previous
page). Some buttons have several functions depending on its mode of operation.
KEY FUNCTION
ON
OFF
CAL
MEAS
HOLD
ENTER
MI
MR
• ON/OFF - Powers on and shuts off the meter. The meter will start in a
measurement mode similar to the last measurement mode it is in before
being switched off.
• CAL - Activates the pH or Relative mV calibration mode and when used
with the MODE key, it activates the temperature calibration mode.
• MEAS - Allows return to measurement mode when canceling or terminating
any operation.
• HOLD - Activates/Deactivates freezing of the measured reading while in
measurement mode.
• ENTER - Confirms the calibration values in Calibration mode and the
selection in SETUP and Print Function mode.
Scroll through the memory location and the stored data during memory
recall.
• MI (Memory Input) - Captures the measured readings of the pH, Relative
mV or mV with its corresponding temperature values and stores them in
the memory.
• MR (Memory Recall) - Retrieves the stored data from memory.
• ▲▼ - Sets the calibration values during Relative mV or temperature
calibration.
Scrolls through each SETUP and its sub group menu.
Scroll through Print Function menu and its parameters.
(pH 110)
SETUP
MODE
• SETUP - Activates the parameter setting menu to allow you to customise
meter configuration, view calibration points and electrode offset data,
select auto power off, reset meter, and clear memory.
• MODE - Selects the measurement parameter option between pH with
temperature, Relative mV with temperature (only pH 110), and mV with
temperature.
• PRINT - Allows print of current measurement or stored data to either the
printer or the computer.
3
Instruction Manual CyberScan pH 11 / 110
CAL
MEAS
MIMR
SETUPMODE
HOLD
ENTER
ON
OFF
CAL
MEAS
SETUP
ON
OFF
HOLD
ENTER
MRMI
MODE
Keypad for pH 11 meterKeypad for pH 110 meter
Figure 2: pH 11 and pH 110 keypads
4
Instruction Manual CyberScan pH 11 / 110
3 PREPARATION
3.1 Inserting the Batteries
This meter is packaged with 4 “AAA” alkaline batteries required for operation. To
insert the batteries into the meter, follow the procedure outlined below.
1. To open the battery compartment, press down the catch of the battery cover.
See below.
2. Note the polarity and insert the batteries into the battery compartment
correctly
3. Replace the battery cover into its original position, ensuring the catch is
locked in position.
Your hand-held meter is now ready for operation.
LR03 'AAA' (AM4)
MADE IN S'PORE
Figure 4: Note battery's polarity
Figure 3: Open Battery Cover
5
Instruction Manual CyberScan pH 11 / 110
3.2 Connecting the pH Electrode Temperature Probe & Electrode
Holder
The pH 11/110 meter uses any standard pH, ORP, or Ion Selective Electrode (ISE)
with a BNC connector. For Automatic Temperature Compensation (ATC), this meter
requires a temperature probe with a phono-jack connector.
NOTE: It is important that water does not get into the BNC connector. Also avoid
touching the connector with soiled hands.
3.2.1 To connect pH, ORP or ISE electrode
1. Insert the BNC connector
from the electrode over the
BNC connector socket on the
meter.
2. Make sure the slots of the
connector are in line with the
posts of the socket.
3. Slide the BNC connector of
the electrode over (Figure 5).
4. Rotate the connector
clockwise until it locks.
5. To remove electrode, push
and rotate the connector
Figure 5: Insert the BNC connector from the
electrode cable into the BNC socket on the meter.
counterclockwise.
6. While holding onto the metal part of the connector, pull it away from the
meter.
7. Be careful not to use excessive force.
CAUTION: Do not pull on the probe cord or the probe wires might disconnect.
Refer to section 15 LIST OF ACCESSORIES section on page 65 for information on
temperature probe and other electrodes.
6
Instruction Manual CyberScan pH 11 / 110
3.2.2 To connect the temperature probe:
The temperature probe (provided) uses a 2.5 mm
phono jack to connect with the socket on the pH
11/110. Insert the jack fully into the socket (Figure
6).
Note: Calibrate your temperature probe when you
replace the probe and when using a “3-in-One”
combination pH and temperature probe. See
section 4.5 - Temperature Calibration on page 16
for instructions.
Figure 6: Insert temperature probe
3.2.3 Attaching the Electrode Holder to the
Meter
The pH meter is designed to allow you to do a one hand operation for any
measurement or calibration. For that purpose, two electrode holders are provided.
They are designed for easy use and installation. Care must be taken to avoid use of
excessive force in the process of attaching these components.
1. Locate the slot on the right-hand side of the meter.
2. Gently slide the flange of the holder into the slot on the meter. Make sure the
holder is secured properly into the slot (Figure 7).
3. You can attach the electrode holder in different positions (Figure 9).
Figure 7: Insert electrode holder
3.2.4 To attach a second electrode holder:
The electrode holder is designed such that you can attach one holder onto another.
Up to two electrodes (using the BNC connector and phono-jack) can be used with the
meter at any one time.
7
Instruction Manual CyberScan pH 11 / 110
1. Align the flange of the second electrode holder with the slot of the first holder
(Figure 8).
2. Slide the flange of the second holder into the slot of the first holder until the
tops of the holders are aligned and secure.
Figure 8: To attach a second electrode holder
3.2.5 Insert the electrode into the holder
1. Do not use excessive force
when inserting electrodes into
the holders.
2. Insert the pH electrode into the
opening of the first holder until
the top housing of the electrode
touches the top of the holder.
3. If you are using a separate
temperature probe, insert the
probe into the opening of the
second holder until the ridge on
the housing touches the top of
the holder.
Figure 9: Different positions for one hand operation
NOTE: The holder is designed for
probes 12 mm in diameter. Electrodes larger than 12 mm may not fit in the holder.
Forcing the electrode into the opening may damage the holder or your electrode.
The electrode holders can be attached in different positions for greater flexibility in
measurement and storage purposes. Simply slide out the electrode holders and
reorient into appropriate orientation before putting into position.
8
Instruction Manual CyberScan pH 11 / 110
3.3 Connecting the AC/DC Adapter
Besides using four AAA-sized batteries as a power source, the pH 11/110 meter can
also operate from the power mains using an AC/DC power adapter either at 120/220
VAC (sold separately). This is extremely useful if you have an A.C. power source
available (e.g. laboratory).
Before plugging in, switch off the meter and the power source of adapter. This is a
safety precaution that should be adhered to safeguard your meter.
1. Switch off the meter and power sources.
2. Select the correct AC/DC Adapter either at 120/220 VAC which matches your
input mains voltage.
3. Select the correct output voltage of the AC/DC adapter. (Output Voltage: 9 to
15 V DC, Current: >=50 mA).
4. Gently insert the power adapter D.C. jack into the meter power socket.
5. Switch on the power source of the adapter followed by the meter.
3.4 Connecting the RS232C Cable (Only for pH 110)
The pH 110 meter provides a RS232C output for you to
transmit your readings either to a printer or a computer via
a cable. This is useful in instances where the meter is used
for continuous monitoring of a certain process or
experiment. Data output to the printer or the computer can
then be evaluated.
The data is output in the ASCII format. This format allows
the data to be imported by a wide variety of software that
read ASCII data (e.g. Microsoft’s Excel, Lotus, Quattro-pro
etc.). A complimentary Data Acquisition Software (DAS) is
provided and it captures data transmitted into an ASCII
file for later use.
1. Open the printer port cover located at the bottom
end of the meter. Do not use excessive force when doing this. See Figure 10.
2. Noting the orientation of the RS232C connector, plug the male connector into
the RS232C port of the meter.
Figure 10: Location of
RS232C
3. Fasten the RS232C connector by fastening the two screws at the side of the
male RS232C connector.
9
Instruction Manual CyberScan pH 11 / 110
3.4.1 RS232C Configuration
The pH 110 meter has a 9 pin female RS232C connector with the following pin out:
PIN NO. DESCRIPTION
1 -
2 Transmit Data
3 -
4 -
5 CTS (Clear to Send)
6 -
7 GND (Ground)
8 -
9 -
A one-to-one connection can be made with a 9 pin RS232C port of the computer.
5
4
3
8
9
7
1
2
6
In case pH 110 meter’s output has to be sent to a 25 pin RS232C connector, the
following cable configuration may be used:
pH 110 25 pin connector
2 (TxD) (RxD) 3
5 (CTS) (RTS) 4
7 (GND) (GND) 7
10
Instruction Manual CyberScan pH 11 / 110
4 CALIBRATION
4.1 Important Information on Meter Calibration
When you re-calibrate your meter, previous pH and Relative mV calibration points
are replaced on a point by point basis. For example, if you previously calibrated your
meter at pH 4.01, 7.00, and 10.01, and you have now re-calibrated at pH 7.00, the
meter retains the old calibration data at pH 4.01 and pH 10.01. To view current
calibration points, see section 9.3 - P2.0: Viewing Previous Calibration Data on page
38.
To completely re-calibrate your meter, or when you use a replacement probe, it is
best to set the meter to its factory defaults and re-calibrate the meter at all points. To
reset the meter to its factory defaults, see section 9.6.2 - User Reset on page 43.
4.2 Preparing the Meter for Calibration
Before starting calibration, make sure you are in the correct measurement mode.
When you switch on the meter, the meter starts up in the units last used. For
example, if you shut the meter off in “mV” units, the meter will read “mV” units when
you switch the meter on.
Be sure to remove the protective electrode storage bottle or rubber cap of the
electrode before calibration or measurement. If the electrode has been stored dry,
wet the electrode in tap water for 10 minutes before calibrating or taking readings to
saturate the pH electrode surface and minimise drift.
Wash your electrode in deionised water after use, and store in electrode storage
solution. If storage solution is not available, use pH 4.01 or 7.00 buffer solution.
Do not reuse buffer solutions after calibration. Contaminants in the solution can affect
the calibration, and eventually the accuracy of the measurements. See section 15.3 Calibration Solutions on page 67 for information on our high-quality pH buffer
solutions.
It is recommended that you perform at least a 2-Point Calibration using standard
buffers that adequately cover the expected measurement range prior to
measurement. 1-Point Calibration can also be used for quick measurements. Make
sure that the calibration point is close to the sample value to be measured.
11
Instruction Manual CyberScan pH 11 / 110
4.2.1 pH 11 meter calibration
The pH 11 meter is capable of up to 5-point pH calibration to ensure accuracy across
the entire pH range of the meter. You can select from the following buffer options:
• USA buffers --- pH 1.68, 4.01, 7.00, 10.01, and 12.45.
The pH 110 meter features four separate internationally recognised buffer standards.
Select the buffer standard you require in section 9.2.2 - P1.2: Select Calibration
Buffer Options on page 35.
This meter is capable of up to 6-point pH calibration, depending on the buffer
standard selection. You can select from the following buffer options:
• USA buffers --- pH 1.68, 4.01, 7.00, 10.01, and 12.45.
• DIN buffers --- pH 1.09, 3.06, 4.65, 6.79, 9.23, and 12.75.
• PWB (Low Ionic): pH 4.10 and 6.97
The meter automatically recognises and calibrates to these standard buffer values,
which makes pH calibration faster and easier.
12
Instruction Manual CyberScan pH 11 / 110
q
4.3 pH Calibration with ATC
NOTE: We recommend that you perform at least 2-point calibration using standard
buffers that bracket (one above and one below) the expected sample range.
1. If necessary, press the MODE key to select pH
mode. The pH indicator appears in the upper
right hand corner of the display.
MEAS
pH
2. Rinse the probe thoroughly with de-ionised
water or a rinse solution. Do not wipe the
probe; this causes a build-up of electrostatic
charge on the glass surface.
3. Dip the probe into the calibration buffer. The
end of the probe must be completely immersed
into the sample. Stir the probe gently to create a
homogeneous sample.
4. Press CAL/MEAS key to enter pH calibration
mode. The CAL indicator will be shown. The
primary display will show the measured reading
while the smaller secondary display will indicate
the pH standard buffer solution.
5. Wait for the measured pH value to stabilise
(The READY indicator will be shown only if it is
activated in the set up menu).
6. Press HOLD/ENTER key to confirm calibration.
The meter is now calibrated to the current
buffer.
READY
7.16
CAL
MEAS
7.16
7.16
22.3
CAL
7.00
CAL
7.00
°C
ATC
pH
pH
pH
pH
See figure on right.
13
HOLD
ENTER
CAL
READY
7.00
7.00
Figure 11: pH Calibration
se
uence
pH
pH
Instruction Manual CyberScan pH 11 / 110
p
7. Rinse the probe with de-ionised water or a rinse
solution, and place it in the next pH buffer.
8. Follow steps 5 and 6 for additional calibration
points. See Figure 12.
9. When all the calibration points set in the Unit
Configuration Setup (see P1.3 Select Number
of pH Calibration Points on page 36) are
completed, the meter returns to Measurement
mode automatically. However, if you wish to
terminate the calibration without completing the
number of points as set in the Unit
Configuration Setup menu, press CAL/MEAS to
return to pH measurement mode.
NOTES:
1. To exit from pH calibration mode without confirming
calibration, DO NOT press ENTER in step 6. Press
CAL/MEAS instead.
CAL
READY
4.15
4.01
HOLD
ENTER
CAL
READY
4.01
4.01
Figure 12: Proceed to next
calibration
oint
pH
pH
pH
pH
2. To limit the number of pH buffer values available during
calibration, see P1.3 Select Number of pH Calibration Points
on page 36.
14
Instruction Manual CyberScan pH 11 / 110
q
A
4.4 mV Calibration (Offset) (Only for pH 110)
With pH 110 meter, you can perform a calibration or an
offset of up to ±150 mV. Note: For Oxidation reduction
Potential (ORP) or Redox measurements, an ORP
MEAS
mV
electrode must be used.
133.4
1. While in the measurement function, press
MODE to enter the mV mode. The mV indicator
appears in the upper right hand corner. See
figure at right.
2. Press the CAL/MEAS key. The CAL indicator
appears above the primary display. The primary
display shows the relative mV reading and the
secondary display shows the absolute mV value.
24.5
CAL
MEAS
CAL
133.4
°C
ATC
mV
•If you have never calibrated relative mV or if the
meter has been reset, the value shown in the
primary display is the same as the absolute mV
value.
•“R.” annunciator will appear once mV calibration
is performed, an indication of mV offset.
3. Press the MI/▲ or MR/▼keys to enter the
relative mV value that matches your desired
reading.
4. Press the HOLD/ENTER key to confirm the
reading and to return to the measurement mode.
The primary display now shows the relative mV
reading.
NOTES:
133.4
MR
READY
MI
CAL
155.0
133.4
HOLD
ENTER
MEAS
READY
155.0
24.5
Figure 13: mV calibration
se
uence
mV
R.mV
°C
TC
1. To view the mV offset value, see section 9.4.2 - View mV
Offset on page 40.
2. To reset all calibration and offset values in memory to the
factory default settings, see section 9.6.1 - Calibration Reset
page 42.
15
Instruction Manual CyberScan pH 11 / 110
A
4.5 Temperature Calibration
The temperature sensor is factory calibrated. Calibrate
the temperature probe only if you suspect temperature
errors may have occurred over a long period of time or if
you have a replacement temperature probe. This
procedure offers offset adjustment of probe to ensure
more accurate temperature measurement.
1. Make sure the ATC probe (or temperature
connector of the electrode) is connected to the
phono-socket (see Figure 6).
MEAS
7.16
22.3
CAL
MEAS
CAL
pH
°C
ATC
pH
2. Switch the meter on. The ATC annunciator will
appear at the right-hand side of the LCD. Press
the MODE key to select pH mode.
3. Press the CAL/MEAS key to enter pH
calibration mode. The CAL indicator will appear
above the primary display.
4. While in pH calibration mode, press the MODE
key to enter temperature calibration mode. The
primary display shows the measured
temperature reading according to the last set
offset1 and the secondary display shows the
factory calibrated temperature measurement.
5. Dip the ATC probe (or electrode) into a solution
of known temperature (i.e. a temperature bath).
Allow time for the temperature probe to
stabilise.
6. Scroll with the MI/▲ or MR/▼ keys to set the
correct temperature value (i.e. the temperature
of the temperature bath). You can adjust the
reading in increments of 0.1°C.
7.16
MODE
HOLD
ENTER
7.00
CAL
22.3
22.3
MIMR
CAL
22.0
22.3
pH
°C
TC
°C
ATC
7. Once you have selected the correct
temperature, press the HOLD/ENTER key. The
meter automatically returns to pH measurement
mode. See Figure 14.
NOTES:
READY
MEAS
7.16
22.0
pH
°C
ATC
You can offset the temperature reading up to ±5°C from
original reading.
Figure 14: Temperature
calibration
To exit this program without confirming the temperature
calibration value, DO NOT press ENTER. Press CAL/MEAS instead.
1
It displays the value being measured currently, offset by the last calibration.
16
Instruction Manual CyberScan pH 11 / 110
5 MEASUREMENT
This meter is capable of taking measurements with automatic or manual temperature
compensation. Automatic temperature compensation only occurs when a
temperature sensor is plugged into the meter. If there is no temperature sensor
plugged into the meter, the default manual temperature setting is automatically 25.0
°C. You can manually set the temperature to match your working conditions using a
separate thermometer.
NOTE: Remove the protective rubber cap or soaker bottle of the electrode before
proceeding with measurement. Take care not to exert too much force as this may
cause damage to the electrode.
5.1 Automatic Temperature Compensation
For automatic temperature compensation (ATC) simply
plug the temperature probe into the meter. The ATC
indicator will light up on the LCD.
NOTE: If you are using a temperature probe, the probe
must be submersed in the liquid you are measuring.
MEAS
pH
7.05
°C
24.5
Figure 15: ATC indicator
ATC
17
Instruction Manual CyberScan pH 11 / 110
p
5.2 Manual Temperature Compensation
IMPORTANT: For manual compensation, you must disconnect the temperature
probe. The [ATC] annunciator will disappear from the LCD.
1. Switch the meter on. Press the MODE key to
select pH mode.
2. Press the CAL/MEAS key to enter pH
calibration mode. The CAL indicator will appear
above the primary display.
3. While in pH calibration mode, press the MODE
key to enter temperature calibration mode. The
primary display act as the adjustable
temperature setting and the secondary display
shows the default temperature value of 25°C or
indicates the last set temperature setting.
4. Check the temperature of your sample using an
accurate thermometer.
5. Press the MI/▲ or MR/▼ keys to set the
temperature to the measured value from step 4.
6. Press HOLD/ENTER key to confirm the
selected temperature and to return to the pH
measurement mode.
7. The meter will now compensate pH readings for
the manually set temperature.
READY
25.0
30.0
HOLD
ENTER
MEAS
7.05
CAL
°C
25.0
MIMR
CAL
°C
25.0
pH
°C
30.0
See Figure on right.
NOTES:
To exit this program without confirming the manual temperature
compensation value, DO NOT press ENTER in step 6. Press
CAL/MEAS instead.
18
Figure 16: Manual
tem
erature compensation
Instruction Manual CyberScan pH 11 / 110
5.3 Taking Measurements
Be sure to remove the electrode soaker bottle or protective rubber cap on the
electrode before measurement.
To take readings:
1. Rinse the probe with de-ionised or distilled
water before use to remove any impurities
adhering to the probe body. If the pH electrode
has dehydrated, soak it for 30 minutes in
electrode storage solution or 2M – 4M KCl
MEAS
8.23
21.3
pH
°C
ATC
solution (sold separately).
2. Press ON to switch on meter. The MEAS
annunciator appears on the top center of the
Figure 17: Measurement
mode
LCD. The ATC indicator appears in the lower
right-hand corner to indicator Automatic Temperature Compensation (section
5.2 on Manual Temperature Compensation).
3. Dip the probe into the sample.
When dipping the probe into the sample, the sensor or the glass bulb of the
electrode must be completely immersed into the sample. Stir the probe gently in the
sample to create a homogeneous sample.
4. Allow time for the reading to stabilise. Note the reading on the display.
5. To toggle between pH and mV (or Rel mV) readings, press the MODE key.
5.3.1 Taking measurements with READY indicator selected on
If the READY indicator has been activated, the READY annunciator lights when the
reading is stable2 *. You can switch the READY indicator on or off in SETUP
program. See 9.2.1 - P1.1: READY Indicator and Auto Hold function on page 34.
5.3.2 Taking measurements with the AUTO HOLD feature selected on
This feature is available on model pH 110 only.
When a reading is stable for more than 5 seconds, the AUTO HOLD feature will
automatically “HOLD” the reading. The “HOLD” indicator appears on the left side of
the display. Press the HOLD/ENTER key to release the reading. Switch the Auto
Hold feature on or off in SETUP program in section 9.2.1 - P1.1: READY Indicator
and Auto Hold function on page 34.
2
The READY indicator appears and the reading holds until the measured value exceeds the tolerance (±0.02 pH;
This feature lets you freeze the value of the pH, mV (or relative mV) and temperature
readings for a delayed observation. HOLD can be used any time when in MEAS
mode.
1. To hold a measurement, press the
HOLD/ENTER key while in measurement mode.
“HOLD” will appear on the display. See Figure
18.
2. To release the held value, press HOLD/ENTER
key again. Continue to take measurements.
HOLD
MEAS
6.89
20.7
pH
°C
ATC
Figure 18: Press HOLD to
freeze readin
NOTE:
a. This meter shuts off automatically 20 minutes after last key is
pressed. If the meter is shut off either automatically or
manually, the HOLD value will be lost. For longer storage,
use the Memory functions described in page 21.
b. For pH 110 model only: The pH 110 meter has an Auto Hold
feature. When this feature is switched on, the display will
automatically “HOLD” a reading that has been stable for
more than 5 seconds. When “HOLD” indicator appears,
press the HOLD/ENTER key to release the reading. To
switch on or off the Auto Hold feature, see section 9.2.1 P1.1: READY Indicator and Auto Hold function on page 34.
20
Instruction Manual CyberScan pH 11 / 110
y
7 MEMORY FUNCTION
The memory function is available in both pH 11 and pH
READY
MEAS
8.23
21.3
pH
°C
ATC
110 meters. These meters store data in sets:
• pH and temperature
• mV and temperature
• Relative mV and temperature (for pH 110)
The pH 11 meter can store up to 50 sets of data while
pH 110 meter can store up to 100 sets of data. You can
store data in any combination of values. For example,
you can store 17 pH and 19 mV values.
7.1 Memory Input
1. During any measurement function (MEAS),
press MI/▲ key to input any data into the
memory.
2. MEM, “Sto” and memory number will flash. The
meter then returns to measurement mode.
3. If the memory is full, the display will flash “MEM
– FUL” momentarily (Figure 20). The first value
stored will be erased to create space for the new
value.
MEM
FUL
MEM
READY
MI
MEM
MEAS
5.69
MI
MEM
MEM
1
22.4
MEM
2
pH
°C
ATC
Figure 19: Data storage in
memor
MEM
MEM
1
Figure 20: Memory full
21
Instruction Manual CyberScan pH 11 / 110
p
A
A
7.2 Memory Recall
This function recalls the previous readings stored in the
memory. You can only access MR from the
measurement mode. Memory recall is in “Last In First
Out” order. To recall readings:
MEAS
7.16
22.3
pH
°C
ATC
1. Press the MR/▼ key once to retrieve the last
reading stored. The memory location screen –
MEM, “Loc” and the memory number – will flash
on the display.
2. Press the HOLD/ENTER key to recall the
reading stored under that memory number.
3. To display next stored reading, press
HOLD/ENTER key. The display automatically
moves to the next lower memory location
screen.
4. To view a stored reading in a specific memory
location, press MI/▲ or MR/▼ keys to locate it,
and press HOLD/ENTER key to display the
reading.
NOTE: Readings stored in memory are retained even if
the unit is turned off. To erase all readings stored in
memory, use the SETUP in section 9.7 – P 6.0: Memory
Clear (CLr) on page 44.
MR
loc
HOLD
ENTER
5.69
HOLD
ENTER
MEM
2
MEM
22.4
MEM
pH
°C
TC
MEM
loc
26
MIMR
MEM
loc
25
Figure 22: Scroll up or down
for s
ecific memory location
loc
1
HOLD
ENTER
MEM
8.23
°C
21.3
Figure 21: Memory recall
TC
pH
22
Instruction Manual CyberScan pH 11 / 110
8 PRINT FUNCTION (Only for pH 110)
The pH 110 meter is equipped with an RS232 communication port and a PRINT ()
key that facilitate the printing of data from the meter to a printer or computer.
8.1 Using pH 110 Meter With Printer or Computer
1. To use the pH 110 directly with a printer, the printer should have either a 9
pin or a 25 pin RS 232C serial port.
2. Communication settings of the printer or computer’s data acquisition software
(CyberComm Portable DAS) MUST match the following parameters:
a. Baud Rate: 9600
b. Parity Bit: None
c. Stop Bit: One
d. Data Bit: 8
e. Flow Control: None
3. Please refer to your printer’s or computer’s instruction manual for more
information on its communication setup protocol.
4. Use 9-pin cable (Refer to section 15 - LIST OF ACCESSORIES on page 65)
to connect the meter to the printer. If the printer has a 25 pin connector, use
the 9 to 25 pin cable or make your own cable taking note of the connection
parameters as described in section 3.4 - Connecting the RS232C Cable
(Only for pH 110) on page 9.
5. Turn on pH 110 meter and note the following:
f. For printer, ensure it is powered on and paper is loaded.
g. For computer, ensure that Eutech Instruments’ CyberComm Portable
Data Acquisition Software is installed and activated. See section 10.3 Running CyberComm Portable on page 51.
6. To send data to the printer or computer, press the PRINT key.
8.2 Sending Data To Computer or Printer
When the PRINT key is pressed, there are two options:
• dAt / CUr: Print Data on Timed Interval basis. You can print any data
automatically based a pre-set timing from 5 to 30 seconds.
23
Instruction Manual CyberScan pH 11 / 110
• dAt / StO: Print Stored Data from Memory. This simply dumps all data stored
in memory. You cannot print data from selected memory locations.
See Figure below.
From pH or mV measurement mode.
HOLD
ENTER
MR
MI
HOLD
ENTER
Figure 23: Print data based on timed interval or from memory
24
Instruction Manual CyberScan pH 11 / 110
y
MEAS
8.2.1 Print Current Data Manually
READY
pH
6.55
To manually print any data that is currently being
measured:
23.4
°C
1. From measurement mode, press HOLD key to
freeze reading. See Figure 24.
2. Print PRINT () key.
3. The “Prn – CUr” annunciator will appear and
flash for about 0.5 seconds. The data is
transmitted to the printer or computer.
4. The meter returns to measurement mode.
5. Repeat steps 1 to 3 for next measurement.
8.2.2 Print Data on Timed Interval
This feature is useful to automate printing and
downloading ‘live’ measurement values into computer.
The time interval setting can range from 5 to 30
seconds.
READY
HOLD
HOLD
ENTER
MEAS
pH
6.55
°C
23.4
Flash for 0.5 seconds
1. From pH or mV measurement mode, press
PRINT key.
READY
MEAS
pH
6.55
2. Select “Dat/Cur” screen and press
HOLD/ENTER key.
3. The “Sec 5” screen appears. The default value
is 5 seconds time interval. See Figure 25.
Figure 24: Print current data
23.4
manuall
4. Press MI/▲ or MR/▼ key to set a value suitable for your application.
5. Press HOLD/ENTER key.
6. The meter goes into measurement mode with MEM annunciator flashing.
7. At a pre-set time interval, the screen flashes “Prn/CUr” momentarily (printing
data) and switches back to measurement mode with MEM annunciator
flashing.
8. To stop printing, press PRINT key. The MEM annunciator disappears and the
meter is on measurement mode.
°C
25
Instruction Manual CyberScan pH 11 / 110
"MEM" annunicator will keep
flashing during measurement
From measurement mode
mode & data will be printed on
pre-set timed interval.
To stop, press PRINT key
HOLD
ENTER
sec
MR
sec
5
10
MEAS
READY
MEM
pH
4.75
°C
21.0
Print on set timed
interval
MI
HOLD
ENTER
MEAS
READY
Flash for 0.5 seconds
MEM
pH
4.75
Set timed interval for
printing, eg, every 10
seconds
21.0
°C
Figure 25: Print data on timed interval
8.2.3 Print Data from Stored Memory
The pH 110 meter can print data that is stored in the memory. You can print out all
stored data or selected data based on its memory location.
To Print All Stored Data
1. From measurement mode, press PRINT key.
2. Use MI/▲ or MR/▼ key, select “dAt – StO” option and press HOLD/ENTER
key. See Figure 23.
3. At the “SEC – 0” option, use MI/▲ or MR/▼ key to select number of seconds
to print each set of stored data. This setting gives flexibility to match the
meter to a printer’s or computer’s ability (or buffer size) to accept data and
prevent overflow.
26
Instruction Manual CyberScan pH 11 / 110
4. Press HOLD/ENTER key. The “Prn” primary display will flash until all data is
printed.
5. To stop printing, press PRINT key and exit to measurement mode. See
Figure 26.
NOTE: All stored memory will remain in the meter after printing.
"Prn" display will keep flashing
until all stored data are printed. .
To stop, press PRINT key
HOLD
ENTER
MEAS
pH
°C
ATC
sec
0
READY
4.75
21.0
MR
MI
HOLD
ENTER
Flash until all stored
data are downloaded.
sec
Set timed interval to
download stored data
10
To Print Selected Stored Data:
1. From measurement mode, press MR/▼ key.
2. “MEM” indicator appears and the last memory input location is displayed.
3. Press MI/▲ or MR/▼ key to select a specific memory location, and press
HOLD/ENTER key.
4. When ready, press PRINT key.
to printer or computer
Figure 26: Print all stored data
5. The screen flashes “Prn/CUr” for about 0.5 seconds and switches back to the
stored data of the specific memory location.
6. Press HOLD/ENTER key. The meter is now ready to recall the next stored
data in the following memory location. See Figure 27.
27
Instruction Manual CyberScan pH 11 / 110
During measurement mode
MEAS
READY
7.03
pH
Repeat print procedure or
°C
press CAL to exit
23.4
MR
loc
20
MR
loc
10
HOLD
ENTER
MEM
Last memory input
location
MI
MEM
For example,select
memory location 10
MEM
pH
loc
HOLD
ENTER
5.69
Flash for 0.5 seconds
MEM
9
Moves to next
memory location
MEM
22.4
pH
°C
ATC
5.69
°C
22.4
Figure 27: Print based on selected stored data from memory
For more information about Eutech Instruments’ CyberComm Portable Data
Acquisition Software programme, see section 10 - CYBERCOMM PORTABLE DAS
(Only for pH 110) on page 45.
ATC
28
Instruction Manual CyberScan pH 11 / 110
9 ADVANCED SETUP FUNCTIONS
The advanced setup mode lets you customised your meter’s preferences and
defaults. The pH 11 and pH 110 meters feature different sub groups that organise all
setup parameters.
The sub-groups are:
For pH
For mV
P1.0 Unit Configuration P1.0 Unit Configuration
P2.0 Calibration Points P2.0 Not available
P3.0 Electrode Status P3.0 Electrode Status (pH 110)
P4.0 Auto Power Mode P4.0 Auto Power Mode
P5.0 Reset P5.0 Reset
P6.0 Memory Clear P6.0 Memory Clear
See Figure 28.
29
Instruction Manual CyberScan pH 11 / 110
From pH measurement modeFrom mV measurement mode
SETUPSETUP
SETUPSETUP
COFCOF
P 1.0P 1.0
SETUP
SETUP
SETUP
MR
MR
MR
MR
MI
CAL
P 2.0
MI
ELE
P 3.0
MI
A
P 4.0
MI
0
MR
pH
SETUP
MI
ELE
This program is
available
only in PH 110 but
not in PH 11.
P 3.0
MR
SETUP
A
MI
0
P 4.0
MR
SETUP
MI
P 5.0
MR
MI
SETUP
SETUP
P 5.0
MR
MI
MEM
SETUP
MEM
P 6.0
P 6.0
Figure 28: Setup overview for pH and mV operations
30
Instruction Manual CyberScan pH 11 / 110
9.1 Advanced SETUP Mode Overview
Press the SETUP key to enter Setup mode. Press the MI/▲ and MR/▼ keys to scroll
through the sub groups.
9.1.1 pH Setup mode
SETUP
SETUP
SETUP
SETUP
COF
P 1.0
CAL
P 2.0
ELE
P 3.0
A
0
P1.0: Unit Configuration
P1.1 Ready indicator (On, Off or Auto Hold)
P1.2 Select buffer option
P1.3 Number of calibration points: 2, 3, 4 or 5 (or 6 for DIN
pH
P1.4 Select °C or °F temperature units*** (pH 110 only)
P2.0: View Previous pH Calibration Data (View only)
P2.1 First calibration point
P2.2 Second calibration point
P2.3 Third calibration point
P2.4 Fourth calibration point
P2.5 Fifth calibration point
P2.6 Sixth calibration point (for DIN buffer option)
P3.0: View Electrode Data (View only)
P3.1 pH electrode offset
P3.2 pH electrode slope
P4.0: Select Auto Power On/Off
P4.0 Select auto power on or off
buffer option)
SETUP
SETUP
P 4.0
P 5.0
MEM
P 6.0
P5.0: Reset to Factory Default
P5.1 Reset pH calibrated data
P5.2 User reset back to factory default values
P6.0: Memory Clear
P6.0 Clear all stored readings
31
Instruction Manual CyberScan pH 11 / 110
9.1.2 mV Setup mode
SETUP
SETUP
SETUP
SETUP
COF
P 1.0
ELE
P 3.0
A
P 4.0
0
P1.0: Unit Configuration
P1.1 Ready indicator (On, Off or Auto Hold)
P1.4 Select °C or °F temperature units (pH 110 only)
P3.0: View Electrode Data (pH 110 only)
P3.1 Electrode offset in mV (pH 110 only)
P4.0: Select Auto Power On/Off
P4.0 Select auto power on or off
P5.0: Reset to Factory Default
P5.1 Reset Relative mV calibrated data
P5.2 User reset back to factory default values
SETUP
P 5.0
MEM
P 6.0
P6.0: Memory Clear
P6.0 Clear all stored readings
32
Instruction Manual CyberScan pH 11 / 110
g
9.2 P1.0 : Unit Configuration
This subgroup program allows customising the meter to
your specific needs. You can program the meter to:
1. Select READY function ON or OFF (pH 110
allows you an additional feature of switching the
AUTO HOLD function ON or OFF)
2. Select between 2 buffer options (4 buffer
options for pH 110)
3. Select the number of pH calibration points
4. Select between °C and °F units for temperature
readings (for pH 110 meter only)
Figure 29 on the right shows the setup sequence for
this program subgroup.
The following 3 pages explain each of the program in
detail.
SETUP
SETUP
READY
ON
SETUP
cof
HOLD
ENTER
HOLD
ENTER
usa
p 1.0
p 1.1
HOLD
ENTER
p 1.2
SETUP
3 p
p 1.3
HOLD
ENTER
SETUP
c
p 1.4
Figure 29: Setup sequence for
unit confi
uration
°C
33
Instruction Manual CyberScan pH 11 / 110
9.2.1 P1.1: READY Indicator and Auto Hold function
Program P1.1 lets you select “READY indicator on” to indicate when the reading is
stable, or select “READY indicator off” for faster meter response.
In the pH 110 meter, program P1.1 also lets you switch the auto hold function on or
off. Select auto hold on to “HOLD” the reading when it is stable for more than 5
seconds. The display automatically freezes, and the HOLD indicator appears on the
left side of the display. Press the HOLD/ENTER key to release the display and
access other functions. Select auto endpoint off to deactivate this feature.
From measurement mode
1. Press SETUP key to enter Set Up mode.
2. At COF – P1.0 parameter, press HOLD/ENTER
key to select parameter P1.1.
3. Press the MI/▲ or MR/▼ keys to select the
configuration you require.
• OFF switches the READY indicatory off.
• ON switches the READY3 indicator on.
• ON and HOLD together switches the auto
hold feature on (for pH 110 meter only).
4. Press the HOLD/ENTER key to confirm
selection and to proceed to step 2 of P1.2.
Press the CAL/MEAS key to return to
measurement mode.
SETUP
READY
ON
SETUP
READY
OFF
SETUP
READY
HOLD
ON
MR
MR
p 1.1
MI
p 1.1
MI
Figure 30: P1.1 - Configure
READY ON, OFF or HOLD
p 1.1
NOTE:
Meter default is set for READY indicator on and auto hold
function off.
3
The READY indicator appears and the reading holds until the measured value exceeds the tolerance (±0.02 pH;
The types of pH calibration buffer options available are:
Meter Meter Display pH Calibration Buffer Options
pH 11 USA
nSt
pH 110 USA
nSt
dn
PUr
USA buffers --- pH 1.68, 4.01, 7.00, 10.01, and 12.45.
NIST buffers --- pH 1.68, 4.01, 6.86, 9.18, and 12.45.
USA buffers --- pH 1.68, 4.01, 7.00, 10.01, and 12.45.
NIST buffers --- pH 1.68, 4.01, 6.86, 9.18, and 12.45.
DIN buffers --- pH 1.09, 3.06, 4.65, 6.79, 9.23, and 12.75.
PWB ( Low Ionic): pH 4.10 and 6.97
From measurement mode
Press SETUP key to enter Set Up mode.
1. Press the MI/▲ or MR/▼ keys to
scroll through subgroups until you
view parameter P1.0.
2. Press the HOLD/ENTER key two
times to select parameter 1.2.
3. Press the MI/▲ or MR/▼ keys to
select the buffer set you require:
4. Press the HOLD/ENTER key to
confirm selection and to return to the
subgroup menu. Press CAL/MEAS
key to return to measurement mode.
SETUP
SETUP
usa
p 1.2
MR
p 1.2
MR
MI
MI
See Figure 31.
SETUP
This screen is
available only
in the pH 110
meter
SETUP
This screen is
available only
in the pH 110
meter
p 1.2
MR
p 1.2
Figure 31: P1.2 - Select pH calibration
buffer standards
MI
35
Instruction Manual CyberScan pH 11 / 110
p
9.2.3 P1.3 Select Number of pH Calibration Points
Program P1.3 lets you select the number of calibration points that appear in pH
calibration mode: 2, 3, 4, or 5 (or 6 with DIN buffer option). This lets the meter scroll
through the calibration points more quickly if you regularly calibrate at less than 5
points.
From measurement mode:
1. Press SETUP key to enter Set Up mode.
2. Press the MI/▲ or MR/▼ keys to scroll through
subgroups until you view parameter P1.0.
3. Press the HOLD/ENTER key three times to
select parameter 1.3.
4. Press the MI/▲ or MR/▼ keys to select 2, 3, 4,
or 5 point pH calibration (up to 6 points for DIN
calibration set with pH 110 meter).
5. Press the HOLD/ENTER key to confirm
selection and to return to the subgroup menu.
Press CAL/MEAS key to return to measurement
mode.
See Figure 32.
SETUP
SETUP
SETUP
2 p
p 1.3
MR
3 p
p 1.3
MR
4 p
MI
MI
MR
p 1.3
MI
SETUP
5 p
p 1.3
Figure 32: P1.3 - Select
number of calibration
oints
36
Instruction Manual CyberScan pH 11 / 110
p
9.2.4 P1.4 Select °C or °F Temperature Units
Available in pH 110 meter only
The pH 110 meter lets you select between °C and °F
units for temperature readings.
From measurement mode
1. Press SETUP key to enter Set Up mode.
2. Press the MI/▲ or MR/▼ keys to scroll through
subgroups until you view parameter P1.0.
3. Press the HOLD/ENTER key four times to
select parameter 1.4.
4. Press the MI/▲ or MR/▼ keys to toggle
between °C and °F.
5. Press the HOLD/ENTER key to confirm
selection and to return to the subgroup
menu. Press the CAL/MEAS key to return to
measurement mode.
See Figure 33.
SETUP
p 1.4
MR
SETUP
p 1.4
Figure 33: P1.4 - Select
tem
erature units, °C or °F
C
°C
MI
f
°F
37
Instruction Manual CyberScan pH 11 / 110
p
9.3 P2.0: Viewing Previous Calibration Data
This mode lets you recall previous calibration data,
which helps you know when to re-calibrate your meter.
This is a “view only” mode.
SETUP
cal
pH
p 2.0
From measurement mode:
1. Press the SETUP key to enter Set up mode.
2. Press the MI/▲ or MR/▼ keys to scroll through
subgroups until you view parameter P2.0.
3. Press the HOLD/ENTER key repeatedly to
view previous calibration data. See Figure 34.
4. When you have scrolled through all calibration
data, you will automatically return to the
subgroup menu. Press CAL/MEAS key to
return to measurement mode.
HOLD
ENTER
SETUP
pH
---
p 2.1
HOLD
ENTER
SETUP
pH
7.00
p 2.3
HOLD
ENTER
SETUP
pH
10.01
p 2.4
NOTES:
If there is no previous calibration data at a particular point, the
primary display will show “ ----“.
38
Figure 34: P2.0 - View
revious calibration data
Instruction Manual CyberScan pH 11 / 110
p
9.4 P3.0: View Electrode Data
Program 3 has two “view only” options that let you check the electrode parameters
for diagnostic purposes.
•From pH measurement mode, you can view the electrode’s offset and slope
values.
• From mV measurement mode, you can view the electrode’s relative mV offset
value. (pH 110 only)
9.4.1 View pH Electrode Offset and Slope
From pH measurement mode:
1. Press the SETUP key to enter Set Up mode.
2. Press the MI/▲ or MR/▼ keys to scroll through
subgroups until you view parameter P3.0.
3. Press the HOLD/ENTER key.
4. The display shows the electrode offset value. It
is the mV offset at pH 7.00. If you have not
calibrated at any buffer, the primary display
shows ‘---‘.
5. Press the HOLD/ENTER key to proceed to
electrode slope display.
6. The display shows electrode slope in
percentage. Slope displayed will be the current
slope of the slope zone to which a measurement
is made or calibration is done.
7. At any point, you can press the CAL/MEAS key
to return to measurement mode.
SETUP
SETUP
SETUP
ele
p 3.0
HOLD
ENTER
5.3
p3.1
HOLD
ENTER
100.0
p 3.2
mV
pH
%
pH
See Figure 35.
39
Figure 35: P3.1/3.2 - View pH
electrode offset and slo
e
Instruction Manual CyberScan pH 11 / 110
9.4.2 View mV Offset (Only for pH 110)
If you are in mV measurement mode, the display shows the relative mV offset. You
can adjust the relative mV offset in relative mV calibration mode. See section 4.4 mV Calibration (Offset) on page 15.
From mV measurement mode:
1. If necessary, press the MODE key to select mV
measurement mode.
2. From mV measurement mode, press the SETUP
key to enter Set Up mode.
3. Press the MI/▲ or MR/▼ keys to scroll through
subgroups until you view parameter P3.0.
4. Press the HOLD/ENTER key.
5. The display shows the electrode offset value. It
is the relative mV offset. If you have not
calibrated at any mV standards, the primary
display shows 0.00 mV.
6. Press the HOLD/ENTER key to return to P3.0.
7. At any point, you can press the CAL/MEAS key
to return to measurement mode.
See Figure 36.
SETUP
ele
p 3.0
HOLD
ENTER
SETUP
20.0
p3.1
Figure 36: P3.0 - View mV
offset value
R.mV
40
Instruction Manual CyberScan pH 11 / 110
9.5 P4.0 : Auto-Off
In Program P4.0 you can activate the AUTO-OFF option to automatically power off
the instrument 20 minutes after the last key selection. This feature is useful for
conserving battery power. Under default conditions, the instrument has this option
activated.
1. Press the SETUP key to enter Set Up mode.
2. Press the MI/▲ or MR/▼ keys to scroll through
subgroups until you view parameter P4.0.
3. Press the HOLD/ENTER key.
4. Press MI/▲ or MR/▼ keys to select options
•‘YES” - Meter will automatically switch off
20 minutes after the last key operation.
•‘NO’ - Auto Off mode is switched off.
5. Press the HOLD/ENTER key to confirm and
return to P4.0.
6. At any point, you can press the CAL/MEAS
key to return to measurement mode.
SETUP
SETUP
SETUP
HOLD
ENTER
MR
P 4.0
yes
MI
Figure 37: P4.0 - Auto shut off
feature
41
Instruction Manual CyberScan pH 11 / 110
9.6 P5.0 : Reset to Factory Default
This mode lets you reset all parameters to factory
default settings. There are two levels of reset:
•Calibration Reset: Reset only the calibration
values. This clears all previous calibrated
values and reset to factory default. See page 68
and 69 for factory default settings.
SETUP
P 5.0
• User Reset: Clears all data including
calibration, memory and other customised
setup functions and reset it to factory default.
See page 68 and 69 for factory default settings.
9.6.1 Calibration Reset
From measurement mode:
1. Press SETUP key to enter Set Up mode.
2. Press the MI/▲ or MR/▼ keys to scroll
through subgroups until you view parameter
P5.0.
3. Press the HOLD/ENTER key.
4. Press the MI/▲ or MR/▼ keys to toggle
between NO and YES.
• ‘NO’ retains current settings
SETUP
SETUP
HOLD
ENTER
CAL
HOLD
ENTER
CAL
MR
P 5.1
MI
• ‘YES’ erase calibration settings.
5. Press the HOLD/ENTER key to confirm
selection.
• If ‘NO’ is selected, it will retains current
settings and proceed to next program
P5.2 – User Reset.
• If ‘YES’ is selected, all LCD segments will
lit up for one second, erase calibration
values, and the meter switches to the
measurement mode.
6. Otherwise press CAL/MEAS key to return to
measurement mode without resetting to factory default.
See Figure 38.
42
SETUP
CAL
yes
Select either "YES" or "nO",
press ENTER key to confirm.
Figure 38: P5.1 - Calibration
reset
Instruction Manual CyberScan pH 11 / 110
NOTE:
Calibration reset is independent for each measurement parameter.
That is:
• From pH mode, calibration reset will only erase pH calibration
values; and
• From mV mode, calibration reset will only erase mV calibration
values.
9.6.2 User Reset
This program clears all data including calibration,
memory and other customised setup functions and reset
it to factory default.
You can skip program P5.1 (Calibration Reset) and
proceed straight to Program P5.2:
1. From Program P5.0 - rSt, press HOLD/ENTER
key three times to P5.2.
2. Press the MI/▲ or MR/▼ keys to scroll through
subgroups until you view parameter P5.0.
• ‘NO’ retains current settings
• ‘YES’ resets to factory default settings.
3. Press the ENTER key to confirm selection.
•If ‘NO’ is selected, it will retains current
settings and goes back to program 5.0.
SETUP
SETUP
SETUP
HOLD
ENTER
MR
P 5.2
MI
•If ‘YES’ is selected, all LCD segments will lit
up for one second, resets to factory default
settings, and the meter will be in the
measurement mode.
4. Otherwise press CAL/MEAS key to return to
measurement mode without resetting to factory
default.
43
yes
Select either "YES" or "nO",
press ENTER key to confirm.
Figure 39: P5.2 - User reset
Instruction Manual CyberScan pH 11 / 110
y
9.7 P 6.0: Memory Clear (CLr)
Use this parameter to clear all memory values when you need to store a new series
of values. This lets you avoid confusing the old values with the new ones.
NO is the default setting.
NOTE: Selecting YES will wipe out all memory.
From measurement mode:
1. Press the SETUP key to enter Set Up mode.
2. Press the MI/▲ or MR/▼ keys to scroll through
subgroups until you view the parameter P6.0.
3. Press the HOLD/ENTER key.
4. Press the MI/▲ or MR/▼ keys to toggle
between NO and YES.
• NO retains current memory
• YES clears all memory.
See Figure 40.
5. Press the HOLD/ENTER key to confirm
selection and return to the subgroup menu.
6. Press CAL/MEAS key to return to measurement
mode.
SETUPMEM
p 6.0
HOLD
ENTER
SETUPMEM
MR
SETUPMEM
MI
7. Otherwise press CAL/MEAS key to return to
measurement mode without resetting to factory
default.
Figure 41: Insert Eutech Instruments' CD-ROM containing Data Acquisition Software (DAS) into your
CD-ROM drive. It will Auto Run and let you start loading the software by displaying the screen as in
Figure 44. Alternatively, you can also manually load the software by clicking the START button and RUN
command.
45
Instruction Manual CyberScan pH 11 / 110
Figure 42: Click on 'Browse' button and locate CD-ROM drive
Figure 43: Locate the CyberComm Portable Setup program in the CD-ROM under "Eutech DAS"
directory.
46
Instruction Manual CyberScan pH 11 / 110
Figure 44: Select ‘Portable Meters’ directory to go to CyberComm Portable.
Figure 45: Open ‘CyberComm Portable’ to enter to the Setup program.
47
Instruction Manual CyberScan pH 11 / 110
Figure 46: Select "SETUP.EXE" program and click the OPEN button
Figure 49: Key in your name and company name and click NEXT button
Figure 50: To select another Destination Directory to install the program, click on BROWSE button.
Otherwise, click NEXT button.
49
Instruction Manual CyberScan pH 11 / 110
Figure 51: Creating a new program folder. Click on NEXT button.
Figure 52: Click on NEXT button.
50
Instruction Manual CyberScan pH 11 / 110
Figure 53: The CyberComm Portable DAS program is fully installed. Click on FINISH button to end
installation.
10.3 Running CyberComm Portable
Before running the DAS program, please ensure that the RS232 cable is connected
between the computer’s serial port and the meter’s port. See section 3.4 Connecting the RS232C Cable (Only for pH 110) on page 9.
Figure 54: Run the CyberComm Software program
51
Instruction Manual CyberScan pH 11 / 110
Figure 55: The opening screen will appear as above
10.3.1 Buttons & Check-Box
•Enable Connection - Click this button to enable communication between
meter and computer.
• Clear Readings - To clear all data and start all over again.
• Save Readings - To save all data displayed in either *.dat or *.txt format.
• Time Stamp - To include Time and Date stamp when collecting the data.
Time and date information comes from the computer.
52
Instruction Manual CyberScan pH 11 / 110
Figure 56: Under File Menu setting, you can change various parameters. Under ABOUT menu, details of
Eutech Instruments' contact information, email address and updates are shown.
10.3.2 Menu
•Communication Settings - To set communication port number, baud rate
speed, parity and stop bits protocol.
• Open - To open previously saved data file.
• Save - To save current data captured.
• Save As - To save current data set in another format such as *.dat or
*.txt.
•Exit - To exit from CyberComm Data Acquisition Software program.
53
Instruction Manual CyberScan pH 11 / 110
Figure 57: Communication Settings for computer's Com port. It must match with COM port settings on
CyberScan pH 110 meter. Please refer to Section 8.1 Using pH 1 10 Meter With Printer or Computer on
page 23 for the settings.
10.3.3 Communication Settings
• Connect Using - For selecting communication port. (Select ‘1’)
• Baud Rate – For selecting baud rate. (Select ‘9600’ bps (bits per
second)).
• Parity – For selecting parity. (Select ‘None’)
• Stop Bits – For selecting stop bits. (Select ‘1’).
Figure 58: Under SAVE AS menu, you can save your data as *.dat or *.txt formats
54
Instruction Manual CyberScan pH 11 / 110
10.4 Capturing And Printing Data Into Computer Using Data Acquisition
After matching the Communication Settings between your computer using Data
Acquisition program and the pH 110 meter, you can now capture data into your
computer for analysis and storage purposes.
1. Ensure the 1-meter RS232 communication cable (Refer to Accessories
Section) is connected between the computer and the pH 110 meter’s Com
port. Refer to section 3.4 “Connecting the RS232C Cable (Only For pH 110)”
on page 9for connection procedure.
2. Switch on the pH 110 meter and run the Data Acquisition software.
3. Click “ENABLE CONNECTION” button.
4. With the pH 110 meter switched on, press the PRINT key to send data to the
computer. See Figure 57 below.
5. You can use MODE key on the meter and change to other parameter such
as mV or Rel mV and print data accordingly.
6. You can also check off the Time Stamp function, so as to print without the
Time and Date information.
7. You can click Clear Readings button to begin another set of measurements,
or click Save Readings to store readings for future retrieval.
Figure 59: Example of a set of data print in CyberComm Portable DAS
55
Instruction Manual CyberScan pH 11 / 110
10.5 Trouble-shooting Guide
a) Problem: Unable to PRINT
When press PRINT key on pH 110 meter, nothing is printed.
POSSIBLE CAUSES SOLUTIONS
You have not "ENABLE CONNECTION" in
the Data Acquisition program.
The "Communication Settings" in the Data
Acquisition program is different from
meter's setup.
The COM port number in the Data
Acquisition program is wrong.
Your computer's COM port setting may be
wrong.
You may have used the wrong
communication cable.
Click on "ENABLE CONNECTION" in the
Data Acquisition program.
Match the COM port number, baud rate,
parity and stop bits information between the
Data Acquisition program and the meter.
Change the COM port number (1 or 2) in the
Data Acquisition program.
Check your computer's hardware settings
(through Windows OS, BIOS, or any other
OS) and refer to computer's manual or
consult with the computer's manufacturer.
Make sure you use the RS232C cable
supplied together with the meter (Part No.
EC-CA01M09F09).
Check the RS232C configuration as
described in the meter's instruction manual.
To report any bugs, please e-mail to techsupport@eutechinst.com
56
Instruction Manual CyberScan pH 11 / 110
11 ELECTRODE CARE
11.1 Electrode Maintenance
pH electrodes are susceptible to dirt and contamination and need to be cleaned
regularly depending on the extent and condition of use.
11.1.1 Storage
The best results, always keep the pH bulb wet, preferably in electrode storage
solution or pH 4 buffers with 1/100 part of saturated KCl. Other pH buffers or tap
water are also acceptable storage media, but avoid storage in de-ionised water. The
protective electrode storage bottle or rubber cap filled with the buffer solution
provides ideal storage for long periods.
11.1.2 After Use
After measurement is complete, follow the sequence elaborated below for storage.
1. Wash the electrode and reference junction in deionised water.
2. Close the refilling hole by returning its rubber sleeve or stopper cap
(Necessary for only refillable electrodes).
3. Store the electrode as mentioned above (Storage section 11.1.1), or as
recommended by the manufacturer.
The reference electrolyte needs to be refilled when the electrode has been used for
an extended period, or when the internal electrolyte has dried up. To accomplish this,
follow the procedure detailed below.
1. Remove the protective rubber cap or sleeve
2. Remove the protective rubber sleeve to expose the filling port of the
electrode. Remove the old reference electrolyte with a syringe and a flexible
tube.
3. Fill the new reference electrolyte. Add in fresh electrolyte till the level of
refilling port. The reference electrolyte depends on your electrode. Use the
solution type recommended for your electrode. Replace the rubber sleeve.
4. Reuse the electrode. Rinse the liquid junction with deionised water and tap
dry.
NOTE: If these steps fail to restore normal electrode response, you may attempt to
57
Instruction Manual CyberScan pH 11 / 110
rejuvenate it. (See section 11.4 on Rejuvenation Procedure).
11.2 Electrode Cleaning
Electrodes that are mechanically intact can be restored to normal performance by
one or combination of the following procedures.
a) Salt deposits:
Dissolve the deposit by immersing the electrode in tap water for ten to fifteen
minutes. Then thoroughly rinse with deionised water.
b) Oil / Grease Films:
Wash electrode pH bulb in a little detergent and water. Rinse electrode tip with
deionised water.
c) Clogged Reference Junction:
Heat a dilute KCl solution to 60-80 °C. Place the sensing portion of the pH
electrode into the heated KCl solution for approximately 10 minutes. Allow the
electrode to cool while immersed in some unheated KCl solution.
d) Protein Deposits:
Prepare a 1% pepsin solution in 0.1M HCl. Allow the electrode to stand in this
solution for five to ten minutes. Rinse the electrode with deionised water.
11.3 Electrode Activation
Generally, if the procedure of storage and maintenance had been closely followed,
the electrode can be used immediately. However, should the electrode response
become sluggish, it may be possible that the bulb has dehydrated. The bulb can be
re-hydrated by immersing the electrode in an ideal storage solution (e.g. buffer pH 4
solution) for 10 to 30 minutes. If this fails, the electrode may require re-activation.
At no time should one touch or rub the glass bulb as this causes the build-up of
electrostatic charge.
If the above procedure does not re-activate the electrode to acceptable status, try
rejuvenating the electrode by following the procedure outlined below.
58
Instruction Manual CyberScan pH 11 / 110
11.4 Rejuvenation Procedure
1. Dip and stir the electrode in freon or alcohol for 5 minutes.
2. Leave the electrode in tap water for 15 minutes.
3. Dip and stir the electrode in concentrated acid (e.g. HCl, H2SO4) for 5
minutes.
4. Repeat Step 2.
5. Dip and stir in strong base (NaOH) for 5 minutes.
6. Leave for 15 minutes in tap water.
7. Test with standard calibration buffer solutions.
CAUTION: Proper eyewear and gloves must be worn when preparing these
chemicals and performing this operation.
Finally, test with standard calibration buffer solutions to see if the electrode yields
acceptable results. You may repeat steps ‘3’ to ‘6’ again for better response
(maximum 3 times). If the response does not improve, then the electrode has
completed its useful life. Replace with a new electrode.
59
Instruction Manual CyberScan pH 11 / 110
12 ERROR MESSAGES
The following table provides a guideline to enable diagnosis of possible problems
indicated by the messages generated by the meter. The table also provides possible
solutions to the problems encountered.
ERROR MESSAGE INDICATES POSSIBLE CAUSE CORRECTIVE
ACTION
Err. Annunciator Wrong keypad input Wrong input in
selected mode.
Secondary Display
Toggles Continuously
between Calibration
Buffer Values,
Electrode icon blinks &
Err annunciator lights
up
Battery icon lights up Low Battery Battery power is low Replace batteries
Calibration Incorrect buffer used
or contaminated buffer
solution
Release key.
Select valid
operations
depending on
mode.
Check if the right
buffer was
selected or use
fresh buffer
solution.
with fresh ones as
soon as possible.
60
Instruction Manual CyberScan pH 11 / 110
13 TROUBLE-SHOOTING
PROBLEM PROBABLE CAUSE REMEDIAL ACTION
Nothing is displayed when
the ON/OFF key is
selected.
2. Batteries (+ & - poles) not
in correct polarity.
3. Weak batteries.
electrolyte in electrode.
2. Broken electrode.
3. External “noises” or
induction (e.g. electrical
‘noise’ caused by a
nearby motor).
4. Dirty electrode.
1. HOLD mode in operation.
2. Internal program error.
1. Insert batteries.
2. Re-insert batteries in the
correct polarity.
3. Replace batteries or
attach AC adapter.
1. Fill electrode with
reference electrolyte.
2. Replace electrode.
3. Move or switch off
interfering device.
4. Clean the electrode.
Rejuvenate if necessary.
Recalibrate the meter
with the cleaned probe.
Rejuvenate if necessary.
1. Cancel HOLD mode.
2. Reset all internal
programs by re-inserting
battery.
61
Instruction Manual CyberScan pH 11 / 110
14 INFORMATION ON pH MEASUREMENT & ELECTRODE
The previous sections of this manual describe the various features of the instrument
and practical aspects of its operation. This section provides a general description of
pH measurement principles. It also provides tabulations of pH buffer changes with
temperature. Lastly, it includes a list of available accessories.
14.1 pH Measurements
14.1.1 Liquid Junction Potential
The liquid - junction potential is the difference in potential created at the liquid - liquid
pHase boundary at the electrode tip, due to dissimilar composition of the test solution
and the saturated KCl electrolyte. The sign and size of the liquid - junction potential
depends upon the composition of the two solutions, the temperature, and the
geometry of the type of junction used.
When two solutions of different compositions come into contact, ion diffusion occurs
at the point of contact.
When there is a high liquid junction potential, measured values differ greatly from the
true value.
14.1.2 Asymmetry Potential
Within the glass electrode is an internal electrode that contains a reference solution
of pH 7. When the electrode is placed in a pH 7 solution, the pH values within and
outside the electrode are the same, and no potential ought to be produced. In
practice, however, an electric potential called asymmetry potential, is generated.
It has been shown that asymmetry potential varies with the pH of the solution in
contact with the glass, with age, with temperature, with the shape of the membrane,
with the thickness of the membrane, with impurities in the reference solution, and any
imperfections in the glass.
In addition, when the glass membrane dries out, the asymmetry potential increases
and leads to measurement errors. That is why it is important to keep the glass
hydrated prior to use, by soaking the electrode in a buffered solution or tap water.
(Do not use deionised water). The purpose of soaking the glass electrode is to obtain
a swollen low resistance glass surface and to stabilise the asymmetry potential of the
electrode. This swollen surface is generally spoiled by drying, prolonged immersion
in dehydrating solutions or chemical attack, such as etching by alkalis or hydrofluoric
acid. Hence, it is necessary to work with a well-soaked glass electrode to ensure a
constant asymmetry potential.
62
Instruction Manual CyberScan pH 11 / 110
14.1.3 pH and Temperature
The electromotive force generated in the glass electrode varies with the temperature
of the solution. As a result the response of the pH electrode is subjected to variations
in temperature. Automatic Temperature Compensation compensates for this
variance.
A solution’s temperature dependence varies considerably. Some may show an
increase in pH while others a decrease for the same temperature variation. This is
why when you are measuring pH values at a particular temperature, even with a pH
meter having ATC, you must record the solution’s temperature along with the pH
value, or the measurement may be meaningless.
Temperature variation also causes variation in liquid junction potential, asymmetry
potential, and the pH value of the reference solution. To reduce these factors, it is
important to ensure that the temperature of the standard calibrating solution and the
sample solution are the same.
14.2 Use of Standard pH Buffers
Standard pH buffer solutions are used to calibrate or standardise a pH meter before
you measure the pH of a sample. They serve as reference standards for the basis of
comparison between measurements.
The more common standard buffers are the pH 4.01, pH 7.00 and pH 10.01. The
others include pH 1.68 and pH 12.45.
For 1-point calibration, you only need a general pH value. Use a standard buffer of
pH 7.00 or a standard buffer whose pH value is close to that of the sample.
2-point calibration is used when you know that the sample is acidic or basic.
• • For acidic sample: use standard buffers of pH 7.00 and pH 4.01
• For basic sample: use standard buffers of pH 7.00 and pH 10.01
3-point calibration is necessary when the sample’s pH is completely unknown. pH
7.00, pH 4.01 and pH 10.01 standard buffers should be used.
63
Instruction Manual CyberScan pH 11 / 110
14.3 Standard pH Buffers
The following table shows the various pH values at different temperature of the
solution during calibration.
Temperature
(ºC)
0 1.67 4.01 7.12 10.32 13.43
5 1.67 4.01 7.09 10.25 13.21
10 1.67 4.00 7.06 10.18 13.00
15 1.67 4.00 7.04 10.12 12.81
20 1.68 4.00 7.02 10.06 12.63
25 1.68 4.01 7.00 10.01 12.45
30 1.69 4.01 6.99 9.97 12.29
35 1.69 4.02 6.98 9.93 12.13
40 1.70 4.03 6.97 9.89 11.99
45 1.70 4.04 6.97 9.86 11.84
50 1.71 4.06 6.97 9.83 11.70
55 - 4.08 6.97 9.81 -
60 - 4.10 6.98 9.79 -
70 - 4.12 6.99 9.76 -
80 - 4.16 7.00 9.74 -
90 - 4.20 7.02 9.73 -
pH 1.68
(oxalate)
pH 4.01
(phthalate)
pH 7.00
(neutral
Phosphate)
pH 10.01
(carbonate)
pH 12.45
(Saturated
Calcium
Hydroxide
Solution)
64
Instruction Manual CyberScan pH 11 / 110
15 LIST OF ACCESSORIES
15.1 Replacement Meter and Meter accessories
Eutech Instruments
Item
CyberScan pH 11 Portable pH/mV Meter with 2 electrode holders
15X000700, temperature probe EC-PH5-TEM01P
CyberScan pH 110 Portable pH/mV/RS232 Meter with 2 electrode holders
15X000700, PC-based DAS, RS232C cable & temperature probe EC-PH5TEM01P
CyberScan pH 11 Portable pH/mV Meter with pH electrode EC-FC7252201B, 2 electrode holders 15X000700, temperature probe EC-PH5-TEM01P
& pH carrying kit set EC-PHWP-KIT
CyberScan pH 110 Portable pH/mV/RS232 Meter with pH electrode ECFC72522-01B, 2 electrode holders 15X000700, temperature probe EC-PH5TEM01P, PC-based DAS, RS232C cable & pH carrying kit set EC-PHWPKIT
Temperature probe for pH 11 and pH 110 meters. EC-PH5-TEM01P
Electrode Holder for probes and electrodes (12 mm diameter) 15X000700
General purpose epoxy-body pH combination electrode, single-junction,
12x110 mm, 1-m cable length.
General purpose epoxy-body pH combination electrode, double-junction,
12x110 mm, 1-m cable length.
Combination ORP electrode, epoxy-body, single-junction, 12x110 mm, 1-m
cable length
Combination ORP electrode, epoxy-body, double-junction, 12x110 mm, 1-m
cable length
RS232 Communication Cable: 9-pin male to 9-pin female connector, 1m
cable length
Serial Impact Micro-Printer: Dot-matrix, paper-roll portable printer with 25-pin
female connector, with a roll of paper and 110/120 VAC power adapter
provided
Eutech Instruments
Order Code
EC-PH11/01
EC-PH110/01
EC-PH11/02K
EC-PH110/02K
EC-FC72521-01B
EC-FC72522-01B
EC-FC73528-01B
EC-FC79601-01B
EC-FC79602-01B
EC-CA01M09F09
EC-
MICROPRNTR01
Serial Impact Micro-Printer: Dot-matrix, paper-roll portable printer with 25-pin
female connector, with a roll of paper and 220/230 VAC power adapter
provided
pH Carrying Kit Set - Plastic Carrying Case comprises 1 x pH 4.01, pH 7.00,
pH 10.01 buffer solutions (60 ml) and 1 x empty rinse water bottle (480 ml)
110/120 VAC Power Adapter (50/60 Hz) 2-flat pin US type, 9 VDC 500mA
(for standard portable meters)
220/230 VAC Power Adapter (50/60 Hz) 2-round pin Euro type, 9 VDC
500mA (for standard portable meters)
65
EC-MICROPRNTR02
EC-PH-WPKIT
EC-120-ADA
EC-220-ADA
Instruction Manual CyberScan pH 11 / 110
15.2 Oakton Instruments
Item
CyberScan pH 11 Portable pH/mV Meter with 2 electrode holders
15X000700, temperature probe EC-PH5-TEM01P
CyberScan pH 11 Portable pH/mV Meter with pH electrode EC-FC7252201B, 2 electrode holders 15X000700, temperature probe EC-PH5-TEM01P &
pH carrying kit set EC-PHWP-KIT
CyberScan pH 110 Portable pH/mV/RS232 Meter with 2 electrode holders
15X000700, PC-based DAS, RS232C cable & temperature probe EC-PH5TEM01P
CyberScan pH 110 Portable pH/mV/RS232 Meter with pH electrode ECFC72522-01B, 2 electrode holders 15X000700, temperature probe EC-PH5TEM01P, PC-based DAS, RS232C cable & pH carrying kit set EC-PHWPKIT
Temperature probe for pH 11 and pH 110 meters. 35613-05
Electrode Holder for probes and electrodes (12 mm diameter) 35615-06
General purpose epoxy-body pH combination electrode, single-junction,
12x110 mm, 1-m cable length.
General purpose epoxy-body pH combination electrode, double-junction,
12x110 mm, 1-m cable length.
Combination ORP electrode, epoxy-body, single-junction, 12x110 mm, 1-m
cable length
Combination ORP electrode, epoxy-body, double-junction, 12x110 mm, 1-m
cable length
RS232 Communication Cable: 9-pin male to 9-pin female connector, 1m
cable length.
Serial Impact Micro-Printer: Dot-matrix, paper-roll portable printer with 25-pin
female connector, with a roll of paper and 110/120 VAC power adapter
provided
Oakton Instruments
Order Code
35614-22
35614-20
35615-22
35615-20
59001-65
35641-51
35811-71
35811-72
35649-50
35649-51
35615-09
35622-00
Serial Impact Micro-Printer: Dot-matrix, paper-roll portable printer with 25-pin
female connector, with a roll of paper and 220/230 VAC power adapter
provided
Soft belt-loop carrying case 35615-75
pH kit for pH 11 & pH 110 meter – Hard plastic carrying case comprises of 1
pc deionised water bottle (480 ml, empty), 1 x pH 4.01, pH 7.00 & pH 10.01
buffer pouches (20-ml), 4 pcs calibration jars (30-ml, empty), & 1 pc
rinse/waste water bottle (480-ml, empty).
110/120 VAC Power Adapter (50/60 Hz) 2-flat pin US type, 9 VDC 500mA
(for standard portable meters)
220/230 VAC Power Adapter (50/60 Hz) 2-round pin Euro type, 9 VDC
500mA (for standard portable meters)
pH 4.01 buffer sachets, 20 ml x 20 pcs. EC-BU-4BS 35653-01
pH 7.00 buffer sachets, 20 ml x 20 pcs. EC-BU-7BS 35653-02
pH 10.01 buffer sachets, 20 ml x 20 pcs. EC-BU-10BS 35653-03
pH De-ionised water rinse sachets, 20 ml x 20 pcs EC-RIN-WT 35653-00
pH sachet assortment pack – 5 each of pH 4.01, pH
7.00, pH 10.01 and de-ionised water sachets per box.
Protein cleaning solution for pH electrode EC-DPC-BT 00653-06
Storage solution for pH electrode EC-RE-005 00653-04
Eutech
Instruments Order
Code
EC-AST-PK 35653-04
Oakton
Instruments Order
Code
Note: pH buffer solutions (480-ml bottle) have ±0.01 pH accuracy at 25 °C.
Sachets are individually sealed, single use pouch containing 20 ml of fresh,
contamination free calibration solution. pH buffer sachets have ±0.01 pH accuracy at
25°C.
67
Instruction Manual CyberScan pH 11 / 110
16 FACTORY DEFAULT SETTINGS
16.1 pH 11 Factory Default Settings
PROGRAM FUNCTION PARAMETER
OPTING KEYS
P 1.0 Unit Configuration Parameter Title
Only
P 1.1 Ready Indicator MI/▲; MR/▼ON ; OFF, ON
P 1.2 Select pH buffer option MI/▲; MR/▼USA, nSt, USA
P 1.3 Number of calibration
points
P 2.0 View previous calibration
data
P 2.1 First calibration point Viewing Only Viewing Only - - P 2.2 Second calibration point Viewing Only Viewing Only - - P 2.3 Third calibration point Viewing Only Viewing Only - - P 2.4 Fourth calibration point Viewing Only Viewing Only - - P 2.5 Fifth calibration point Viewing Only Viewing Only - - P 3.0 View electrode data Parameter Title
P 3.1 Electrode offset (pH mode) Viewing only Viewing only - - P 3.2 pH electrode slope Viewing only Viewing only - - -
P 4.0 Select auto power on/off MI/▲; MR/▼ YES, nO YES
P 5.0 Reset to factory default Parameter Title
P 5.1 Reset pH or ORP
calibrated data
P 5.2 User reset to factory
default
P 6.0 Memory clear MI/▲; MR/▼ nO, YES nO
MI/▲; MR/▼3 P, 4 P, 5 P, 2 P 3 P
Parameter Title
Only
Only
Only
MI/▲; MR/▼ nO, YES nO
MI/▲; MR/▼ nO, YES nO
PARAMETER
OPTIONS
Parameter Title
Only
Parameter Title
Only
Parameter Title
Only
Parameter Title
Only
DEFAULT
SETTING
NOTE: The HOLD/ENTER key is used to confirm every parameter opted. The
CAL/MEAS on the other hand will escape the set up operation to return to the
measurement mode.
68
Instruction Manual CyberScan pH 11 / 110
16.2 pH 110 Factory Default Settings
PROGRAM FUNCTION PARAMETER
OPTING KEYS
P 1.0 Unit Configuration Parameter Title
Only
P 1.1 Ready Indicator MI/▲; MR/▼ON ; OFF, HOLD ON
P 1.2 Select pH buffer option MI/▲; MR/▼USA, nSt, dn, PUr USA
P 1.3 Number of calibration
points
P 1.4 Select °C or °F MI/▲; MR/▼ C, F C
P 2.0 View previous calibration
data
P 2.1 First calibration point Viewing Only Viewing Only - - P 2.2 Second calibration point Viewing Only Viewing Only - - P 2.3 Third calibration point Viewing Only Viewing Only - - P 2.4 Fourth calibration point Viewing Only Viewing Only - - P 2.5 Fifth calibration point Viewing Only Viewing Only - - P 2.6 Sixth calibration point Viewing Only Viewing Only - - P 3.0 View electrode data Parameter Title
P 3.1 Electrode offset (pH mode) Viewing only Viewing only - - -
P 3.1 Electrode offset (mV
mode)
P 3.2 pH electrode slope Viewing only Viewing only - - -
P 4.0 Select auto power on/off MI/▲; MR/▼ YES, nO YES
P 5.0 Reset to factory default Parameter Title
P 5.1 Reset pH or ORP
calibrated data
P 5.2 User reset to factory
default
P 6.0 Memory clear MI/▲; MR/▼ nO, YES nO
MI/▲; MR/▼3 P, 4 P, 5 P, 2 P 3 P
Parameter Title
Only
Only
Viewing only Viewing only 0.0
Only
MI/▲; MR/▼ nO, YES nO
MI/▲; MR/▼ nO, YES nO
PARAMETER
OPTIONS
Parameter Title
Only
Parameter Title
Only
Parameter Title
Only
Parameter Title
Only
DEFAULT
SETTING
NOTE: The HOLD/ENTER key is used to confirm every parameter opted. The
CAL/MEAS on the other hand will escape the set up operation to return to the
measurement mode.
69
Instruction Manual CyberScan pH 11 / 110
17 SPECIFICATIONS
MODEL NO. pH 11 pH 110
pH Range -2.00 to 16.00 pH -2.00 to 16.00 pH
Resolution 0.01 pH 0.01 pH
Accuracy ± 0.01 pH ± 0.01 pH
Slope Range 80% to 120% 80% to 120%
Millivolt Range -1999 to +1999 mV -1999 to +1999 mV
Resolution 0.1 mV (between ±199.9 mV)
1 mV (beyond ±199.9 mV)
Accuracy ± 0.2 mV (between ±199.9 mV)
± 2.0 mV (beyond ±199.9 mV)
Relative mV Range - none - + 1999 to -1999 mV
Resolution - none - 0.1 mV (between ±199.9 mV)
Accuracy - none - 0.2 mV (between ±199.9 mV)
Temperature Range -10.0 to 110.0 ºC -10.0 to 110.0 ºC (14 to 230 °F)
Resolution 0.1 ºC 0.1 ºC (0.1 °F from 14.0 °F to 199.9 °F ) &
Accuracy ± 0.5 ºC ± 0.5 ºC (± 0.9 °F)
Temperature Compensation Automatic / Manual (0 to 100 ºC) Automatic / Manual (0 to 100 ºC)
No. Of Calibration Points 2 to 5 points 2 to 6 points
Environmental Requirements 0 to 50 °C (32 to 122 °F) 0 to 50 °C (32 to 122 °F)
eight: & Dimensions Meter: 18.7 x 9 x 4 cm; 290 g
Boxed: 24 x 23 x 7 cm; 650 g
Note: In strong electromagnetic environment, a deviation of up to 3 % of reading may be noticed.
0.1 mV (between ±199.9 mV)
1 mV (beyond ±199.9 mV)
± 0.2 mV (between ±199.9 mV)
± 2.0 mV (beyond ±199.9 mV)
1 mV (beyond ±199.9 mV)
2.0 mV (beyond ±199.9 mV)
(1 °F from 200 °F to 230 °F)
USA: pH 1.68, 4.01, 7.00, 10.01 & 12.45
NIST: pH 1.68, 4.01, 6.86, 9.18, & 12.45
DIN: pH 1.09, 3.06, 4.65, 6.79, 9.23 & 12.75
PWB: pH 4.10 and 6.97
4 ‘AAA’ Batteries;
AC/DC Adapter 9 V, 200 mA
Meter: 18.7 x 9 x 4 cm; 290 g
Boxed: 24 x 23 x 7 cm; 650 g
70
Instruction Manual CyberScan pH 11 / 110
18 WARRANTY
This meter is supplied with a three-year warranty, six-month warranty for probe
against significant deviations in material and workmanship.
If repair or adjustment is necessary and has not been the result of abuse or misuse
within the designated period, please return – freight pre-paid – and correction will be
made without charge. Eutech Instruments/ Oakton Instruments will determine if the
product problem is due to deviations or customer misuse.
Out of warranty products will be repaired on a charged basis.
Exclusions
The warranty on your instrument shall not apply to defects resulting from:
• Improper or inadequate maintenance by customer
• Unauthorised modification or misuse
• Operation outside of the environment specifications of the products
19 RETURN OF ITEMS
Authorisation must be obtained from our Customer Service Department or authorised
distributor before returning items for any reason. A “Return Goods Authorisation”
(RGA) form is available through our Authorised Distributor. Please include data
regarding the reason the items are to be returned. For your protection, items must be
carefully packed to prevent damage in shipment and insured against possible
damage or loss. Eutech Instruments/ Oakton Instruments will not be responsible for
damage resulting from careless or insufficient packing. A restocking charge will be
made on all unauthorised returns.
NOTE:
Eutech Instruments Pte Ltd/ Oakton Instruments reserve the
right to make improvements in design, construction, and
appearance of products without notice.
71
For more information on Eutech Instruments/ Oakton Instruments’ products, contact your
nearest distributor or visit our website listed below:
Oakton Instruments
P.O Box 5136,
Vernon Hills, IL60061, USA
Tel: (1) 888-462-5866
Fax: (1) 847-247-2984
E-mail: info@4oakton.com
Web-site: