prepare this solution from your own laboratory stock, half fill a 1,000 ml volumetric flask
with distilled water and add 264 grams of reagent-grade ammonium sulfate, (NH4)2SO4.
Swirl the flask gentle to dissolve the solid. Fill the flask to the mark with distilled water,
cap, and upend several times to mix the contents. ISA is added at the rate of 2 ml of ISA to
each 100 ml of standard or sample to adjust the ionic strength to about 0.12M.
3. Eutech Nitrate Standard, 0.1M NaNO
from your own laboratory stock, add 8.50 grams of reagent-grade sodium nitrate to a one
liter volumetric flask about half full of distilled water. Swirl the flask gently to dissolve the
solid. Fill to the mark with distilled water, cap and upend several times to mix the solution.
4. Eutech Nitrate Standard, 1,000 ppm NO
solution from your own laboratory stock, add 1.37 grams of reagent-grade sodium nitrate to
a one liter volumetric flask about half full with distilled water. Swirl the flask gently to
dissolve the solid. Fill to the mark with distilled water, cap and upend several times to mix
the solution.
, Code no. EC-SCS-NT1-BT. To prepare this solution
3
-1
, Code no. EC-SCS-NT2-BT. To prepare this
3
3
Nitrate Electrode Instruction Manual
5. Eutech Nitrate Standard, 100 ppm NO
-1
as N, Code No. EC-SCS-NT3-BT. To prepare this
3
solution from your own laboratory stock, add 0.61 grams of reagent-grade sodium nitrate to
a one liter volumetric flask about half full with distilled water. Swirl the flask gently to
dissolve the solid. Fill to the mark with distilled water, cap and upend several times to mix
the solution.
GENERAL PREPARATION
Electrode Preparation
Remove the rubber caps covering the electrode tips and any rubber inserts covering the filling holes of
the reference electrode. Fill the combination electrode or the reference electrode with the filling
solution shipped with the electrode to a level just below the fill hole. No preparation is required with a
sealed reference electrode. Gently shake the electrode downward in the same manner as a clinical
thermometer to remove any air bubbles which may be trapped behind the nitrate membrane. Prior to
first usage, or after long-term storage, immerse the nitrate membrane in nitrate standard for thirty
minutes. The electrode is now ready for use.
Connect the electrodes to the proper terminals of the meter as recommended by the meter manufacturer.
Electrode Slope Check (with pH/mV meter)
(check electrodes each day)
1. To a 150 ml beaker, add 100 ml of distilled water and two ml of ISA. Place the beaker on a
magnetic stirrer and begin stirring at a constant rate. After assuring that the meter is in the
millivolt mode, lower the electrode tips into the solution.
2. Using a pipet, add 1 ml of 0.1M or 1,000 ppm nitrate standard to the beaker. When the reading
has stabilized, record the mV reading.
3. Using a pipet, add 10 ml of the same nitrate standard used above to the beaker. When the
reading has stabilized, record the mV reading.
4. Determine the difference between the two readings. The electrode is operating correctly if the
millivolt potential has changed by 56 ± 2 mV, assuming the solution temperature is between
o
and 25oC. See the TROUBLESHOOTING sections if the potential change is not within this
20
range.
Slope
is defined as the change in potential observed when the concentration changes by a factor of 10.
Electrode Slope Check (with ion meter)
(check electrodes each day)
1. Prepare standard nitrate solutions whose concentrations vary by ten fold. Use either the 0.1M
or 1,000 ppm nitrate standard. Use the serial dilution method for this preparation.
2. To a 150 ml beaker, add 100 ml of the lower value standard and 2 ml of ISA. Place the beaker
on the magnetic stirrer and begin stirring at a constant rate. Lower the electrode tips into the
solution. Assure that the meter is in the concentration mode.
4
Instruction Manual Nitrate Electrode
3. Adjust the meter to the concentration of the standard and fix the value in the memory
according to the meter manufacturer's instructions.
4. Rinse the electrodes with distilled water and blot dry.
5. To another 150 ml beaker, add 100 ml of the higher value standard and 2 ml of ISA. Place the
beaker on the magnetic stirrer and begin stirring at a constant rate. Lower the electrode tips
into the solution.
6. Adjust the meter to the concentration of the standard and fix the value in the memory.
7. Read the electrode slope according to the meter manufacturer's instructions. Correct electrode
operation is indicated by a slope of 90-100%. See the TROUBLESHOOTING sections if the
slope is not within this range.
MEASUREMENT
Measuring Hints
All samples and standards should be at the same temperature for precise measurement. A difference of
1oC in temperature will result in a 2% measurement error.
The sensing membrane is normally subject to water uptake and might appear milky. This has no effect
on performance.
Constant, but not violent, stirring is necessary for accurate measurement. Magnetic stirrers can
generate sufficient heat to change the solution temperature. To counteract this effect, place a piece of
insulating material, such as styrofoam sheet, between the stirrer and beaker.
Always rinse the electrodes with distilled water and blot dry between measurements. Use a clean, dry
tissue to prevent cross-contamination.
For samples with high ionic strength, prepare standards whose composition is similar to the sample.
Always check to see that the membrane is free from air bubbles after immersion into standard or
sample.
A slow responding electrode may be caused by interferences to the electrode. To restore proper
performance, soak the electrode in distilled water for about 5 minutes to clean the membrane, rinse, and
soak in standard solution for about 5 minutes.
Sample Requirements
All samples must be aqueous and not contain organics which can dissolve in the membrane or extract
out the liquid ion exchanger.
The temperature of the standard solutions and of the sample solutions should be the same and below
40oC. About a 2% error will be introduced for a 1oC difference in temperature.
5
Nitrate Electrode Instruction Manual
Interferences should be absent. If they are present, use the procedures found in the Interferences
section to remove them.
The pH range for the nitrate ion electrode is 2.5-11. Neutralize samples outside this range with acid or
base to bring them in range.
Units of Measurement
Nitrate concentrations are measured in units of ppm as sodium nitrate, ppm as nitrate, m oles per liter, or
any other convenient concentration unit. Table 1 indicates some of the concentration units.
TABLE 1: Concentration Unit Conversion Factors
ppm NaNO
ppm NO
3
850.0 620.0 1.0 X 10
-1
moles/liter
3
-2
85.0 62.0 1.0 X 10-3
8.5 6.2 1.0 X 10-4
MEASUREMENT PROCEDURE
Direct Measurement
Direct measurement is a simple procedure for measuring a large number of samples. A single meter
reading is all that is required for each sample. The ionic strength of samples and standards should be
made the same by adjustment with ISA for all nitrate solutions. The temperature of both sample
solution and standard solutions should be the same.
Direct Measurement of Nitrate (using a pH/mV meter)
1. By serial dilution, prepare three standard solutions from the 0.1M or 1,000 ppm stock
standard. The resultant concentrations should be 10-2, 10-3, and 10-4M or 1,000, 100 and 10
ppm nitrate standards. Add 2 ml of ISA to each 100 ml of standard. Prepare standards with a
composition similar to the samples if the samples have an ionic strength above 0.1M.
2. Place the most dilute solution (10-4M or 10 ppm) in a 150 ml beaker on the magnetic stirrer
and begin stirring at a constant rate. After assuring that the meter is in the mV mode, lower
the electrode tips into the solution. After the reading has stabilized, record the mV reading.
3. Place the mid-range solution (10-3M or 100 ppm) in a 150 ml beaker on the magnetic stirrer
and begin stirring. After rinsing the electrodes with distilled water, blot dry and immerse the
electrode tips in the solution. When the reading has stabilized, record the mV value.
4. Place the most concentrated solution (10-2M or 1,000 ppm) in a 150 ml beaker on the magnetic
stirrer and begin stirring. After rinsing the electrodes in distilled water, blot dry, and immerse
the electrode tips in the solution. When the reading has stabilized, record the mV reading.
6
Loading...
+ 12 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.