Eutech Instruments HIGHLOW TDS TESTER User Manual

Open this leaflet for calibration/testing/maintenance instructions.
Changing Batteries
1. Open battery compartment lid (with attached lanyard loop).
2. Remove old batteries; replace with fresh ones. Note polarity as shown in battery compar tment. See Diagram 1.
3. Recalibrate after battery change.
INC
+
+
DEC
Diagram 1
Warranty
Eutech Instruments warrants its meter free from manufacturing defects for 2 years and electrode module for 6 months. If repair, adjustment or replacement is necessary and has not been the result of abuse or misuse within the time period, please return the tester – freight prepaid – and correction will be made without charge. Out of warranty products will be repaired on a charge basis.
Return Of Items
Authorization must be obtained from Eutech’s distributor before returning items for any reason. When applying for authorization, please include information regarding the reason the item(s) are to be returned.
Note: We reserve the right to make improvements in design, construction and appearance of products without notice. Prices are subject to change without notice.
Waterproof & Floats
Replaceable Electrode
Auto-off
Automatic Temperature
Waterproof ECScan High Instructions
Before You Begin
Remove electrode cap. Soak electrode for a few minutes in alcohol to remove oils.
Calibration
Tester is factory calibrated. However, to ensure accuracy, calibrate conductivity on a regular basis.
Select a calibration standard appropriate for the
ECScan High tester (between 2.00 mS and
19.90 mS). It is best to select a standard close
to the test solution value.
1. Open battery compartment lid (end with lanyard loop).
2. Orientate the battery compar tment as shown in Diagram 1.
3. Rinse electrode in deionized water, then rinse it in calibration standard, then dip it
into a container of calibration standard.
4. Switch unit on (ON/OFF key). Wait several minutes for display to stabilize.
5. Press the INC or DEC key to adjust reading to match the calibration
standard value.
6. After 3 seconds without a key press, the display flashes 3 times, then shows ‘ENT’. The tester accepts calibration value; returns to measurement mode.
7. Replace battery cap.
Conductivity Testing
1. Remove electrode cap. Switch unit on (ON/OFF key).
2. Dip electrode into test solution. Make sure sensor is fully covered.
3. Wait for reading to stabilize (Automatic Temperature Compensation corrects for temperature changes). Note reading.
4. Press ON/OFF to turn off tester. Replace electrode cap. Note: Tester automatically shuts off after
8.5 minutes of non-use.
HOLD Function
Press HOLD key to freeze display. Press HOLD again to release.
Tester Maintenance
• To improve performance, clean the electrode by rinsing them in alcohol for 10 to 15 minutes.
• Replace all batteries if low battery indicator
appears, or if readings are faint or unstable.
• If you experience drift, periodically let electrode
fully dry.
When you need a new electrode, see “Electrode Replacement” on insert in back of box.
Self-Diagnostic Messages
bAt Weak batteries–replace with fresh ones specified by manufacturer.
Err Wrong keypress.
Or, Ur Over range/Under range signal, or electrode
is not in contact with solution, or electrode is failing.
---- Calibration mode activated but not performed.
68X247755 11/00 Rev 0
Open this leaflet for calibration/testing/maintenance instructions.
Changing Batteries
1. Open battery compartment lid (with attached lanyard loop).
2. Remove old batteries; replace with fresh ones. Note polarity as shown in battery compartment. See Diagram 1.
3. Recalibrate after battery change.
INC
+
+
DEC
Diagram 1
Warranty
Eutech Instruments warrants its meter free from manufacturing defects for 2 years and electrode module for 6 months. If repair, adjustment or replacement is necessary and has not been the result of abuse or misuse within the time period, please return the tester – freight prepaid – and correction will be made without charge. Out of warranty products will be repaired on a charge basis.
Return Of Items
Authorization must be obtained from Eutech’s distributor before returning items for any reason. When applying for authorization, please include information regarding the reason the item(s) are to be returned.
Note: We reserve the right to make impro vements in design, construction and appearance of products without notice. Prices are subject to change without notice.
Waterproof & Floats
Replaceable Electrode
Auto-off
Automatic Temperature
Loading...
+ 6 hidden pages