Eutech instruments EcoTestr EC High User Manual

Operating Instructions
EcoTestr EC High Waterproof Pocket Tester
Before First Use: Soak sensor for 2 minutes in alcohol to remove stains/oils. Rinse in clean water.
Testing:
1. Remove cap and press
to turn on the tester
3. Stir once and let the reading stabilize. Note the measured value
4. To hold reading, press
. Screen fl ashes HO once, then displays measurement
with blinking unit (mS) to indicate that tester is in the hold mode. Press
again to
cancel hold mode (
HC
)
5. Press
to shut the tester off
Note: Tester automatically shuts off after 8.5 minutes of non-use to conserve batteries.
Calibration: The tester is factory calibrated. However, to ensure accuracy, regular calibration is recommended.
1. Soak sensor for 10 minutes in alcohol to remove stains/oils. Rinse in clean water and blot sensor dry
2. Press
to turn on the tester
3. Dip sensor fully into standard solution and wait for the value to stabilize
4. Press
to enter the calibration sequence. Display fl ashes
CAL
momentarily and
then shows a fl ashing default reading
5. Press
to scroll to the standard solution value
6. Release
and wait for the auto confi rmation
7. To abort calibration, press
to escape (
ESC
)
8. You have successfully calibrated the instrument
Maintenance: Clean the stainless steel sensor by periodically rinsing them in alcohol for 10 - 15 minutes. Rinse sensor in clean water after each use.
Useful Notes: To avoid cross contamination, rinse between samples and calibration standards with de-ionized water.
Eutech Order Code
Oakton Order Code
Description
ECCON2764BT 00653-20 2764 µS/cm KCl calibration solution ECCON1288BT 00606-10 12.88 mS/cm KCl calibration solution
LIFT LIFT
LIFT
Figure 1: Removing battery cover to change batteries
Battery
cover
Step 2
Lift and push up
Step 1
Lift and hold
Changing Batteries:
1. To remove batteries, lift up the front battery cover to release the front catch. Push up and hold in that position
2. Next, lift the pocket clip to release the back catch and slide upwards to remove the battery cover (Figure 1)
3. Replace old batteries with fresh ones. Note polarity as shown in battery compartment
4. No re-calibration is required after battery change. The tester’s non-volatile memory function retains calibration settings
Error Messages:
1.
bA
– Batteries are weak and need replacement
2.
Er
– Calibration is performed in solution value below 5 % of full-scale range. Use higher value calibration standard – Faulty temperature sensor
3.
Or
– Over range signal for measurements performed above the full-scale range
Reset Option: Reset option allows you to restore the calibration back to factory default settings.
1. Press
to turn off tester
2. Press and hold
, then press to turn the tester on. The display shows a
fl ashing
rS
(reset)
3. Press
again to confi rm reset (
En
) or press to escape (
ESC
) reset option
Warranty: This tester is supplied with a warranty against manufacturing defects for a period of ONE year from date of purchase.
Accessories:
Loading...