3) Amp meter should read 12mA; if not, calibration is necessary;
4) Locate “Offset adjustment” trimmer; Using a fine screwdriver,
slowly turn trimmer till Amp meter reads 12mA;
5) For greater accuracy, adjust slope with pH 4 or 10;
6) Rinse probe in de-ionised water and immerse in either pH 4.01 or pH
10.01 buffer solution;
7) In pH 4.01, the output is 8.54mA;
8) In pH 10.01, the output is 15.44mA;
9) Using a fine screwdriver, slowly turn trimmer, marked ‘slope’, till Amp
meter reads correct mA value with respective buffer solutions.
10) Remember to disconnect Amp meter and close cover.
NOTE: For other buffer values, expected current output is based on:
mA = {(16/14) x buffer pH value} + 4mA
8 ORP Transmitter Calibration
1) Prepare fresh Standard solutions, such as 255mV, 470mV.
2) Short ORP inputs and observe Amp meter display;
3) Amp meter should read 4mA; if not, calibration is necessary;
4) Locate “Offset adjustment” trimmer; Using a fine screwdriver,
slowly turn trimmer till Amp meter reads 4mA;
5) For greater accuracy, adjust the slope with Standard calibrating
solutions;
6) Rinse probe in de-ionised water and immerse in standard solution;
7) In 255mV solution, the output is 8.08mA;
8) In 470mV solution, the output is 11.52mA;
9) Using a fine screwdriver, slowly turn trimmer, marked ‘slope’, till Amp
meter reads correct mA value with respective Standard solutions.
10) Remember to disconnect Amp meter and close cover.
NOTE: For other Standard values, expected current output is based on:
mA = {(16/1000) x ORP value of Standard solution} + 4mA
9 Specifications
Model ECPHXTR ECORPXTR
Range 0.00 – 14.00 pH 0 to +1000mV
Accuracy
± 0.05 pH ± 3 mV
Calibration
Offset ±1pH
Slope 80 – 120%
Offset ±50mV
Slope 80 – 120%
Input Asymmetrical or Symmetrical (with Potential Matching
Pin)
Output 4-20 mA
Operating
Voltage
DC 12 V to 24 V
Load 100 Ohms max. for 12 V;
600 Ohms max. for 24V
Housing 77 x 77 x 28 mm, field mountable
10 Appendix 1
Connecting Amp Meter for Calibration
Figure 3 - Method One
Figure 4 - Method Two
11 Ordering Information
pH
Item Cat Code
pH Transmitter ECPHXTR
Combination pH Electrode, General, integral 5m cable ECARTSO05B
Combination pH Electrode, w/PMP, integral 5m cable ECARGTSO05B
Dissolved Oxygen
Item Cat Code
Dissolved Oxygen Transmitter ECDOXTR
General Purpose D.O. probe, 0.5 – 40 ppm, Industrial,
integral 5m cable
ECDOGEN
General Purpose D.O. probe, 0 – 10 ppm, Industrial,
integral 5m cable
ECDOTPll
ORP
Item Cat Code
ORP Transmitter ECORPXTR
Combination Platinum ORP Electrode, w/PMP,
integral 5m cable
ECHTPTTSO05B
Combination Gold ORP Electrode, w/PMP, integral
5m cable
ECHTAUTTSO05B
Conductivity
Item Cat Code
Conductivity Transmitter ECCONXTR
2-Pin SS,k=1.0, integral 3m cable ECCONSEN46
Temperature
Item Cat Code
Temperature Transmitter ECTEMPXTR
Pt 100 probe, SS316 ECPT10001M or
ECPT10005M
Power Supply
Item Cat Code
120V Power Supply EC-120-XTR
220V Power Supply EC-220-XTR