Eutech instruments DO 700 User Manual

Instruction Manual
sy...
DO 700 Dissolved Oxygen/ºC/ºF
Bench Meter
Part of Thermo Fisher Scientific 68X541703 Rev. 2. Sep 2010
Tech nol o
adeEa
TABLE OF CONTENTS
1. INTRODUCTION.............................................................. 1
2. GETTING STARTED........................................................ 2
2.1 Keypad Functions ...................................................................2
2.2 LCD Annunciators...................................................................3
2.3 Meter Connections..................................................................3
3. DISSOLVED OXYGEN / ATC ELECTRODE................... 4
4. CALIBRATION................................................................. 5
4.1 Temperature Calibration .........................................................6
4.2 % Saturation Calibration .........................................................7
4.3 100% Calibration.....................................................................7
4.4 0% Calibration.........................................................................7
4.5 mg/L or ppm (Concentration) Calibration................................8
5. SETUP FUNCTIONS........................................................ 9
5.1 1.0 CAL (Calibration)...............................................................9
5.2 2.0 ELE (Electrode Information)..............................................10
5.3 3.0 ConF (Configuration).........................................................10
5.4 3.1 rdY (Ready Stability Indicator / Auto Hold)........................10
5.5 3.2 ºC ºF (Celsius or Fahrenheit)............................................10
5.6 3.3 dPr (Barometric Pressure Adjustment)..............................11
5.7 3.4 SALt (Salinity Adjustment) ................................................11
5.8 3.5 OFS (Offset Adjustment) from % mode.............................11
5.9 3.5 Unit (Unit Adjustment) from concentration mode...............12
5.10 4.0 rSt (Reset).........................................................................12
5.11 5.0 CLr (Clear Stored Memory)...............................................12
6. TAKING MEASUREMENTS............................................ 13
6.1 HOLD Function .......................................................................14
6.2 Storing and Recalling Data .....................................................14
7. ELECTRODE MAINTENANCE........................................ 15
8. DISSOLVED OXYGEN THEORY .................................... 18
9.
REPLACEMENTS AND ACCESSORIES........................ 22
10. TROUBLESHOOTING GUIDE......................................... 23
11. SPECIFICATIONS............................................................ 25
12. WARRANTY..................................................................... 26
13. RETURN OF ITEMS......................................................... 27

1. INTRODUCTION

Thank you for purchasing our DO 700 series benchtop meter. This microprocessor-based meter is economical and simple to use. The design incorporates a large LCD for clear viewing, yet offers a small footprint to conserve space.
The DO 700 measures Dissolved Oxygen in % saturation or
centration (mg/L or ppm) with temperature (ºC or ºF).
con Each meter includes a convenient slide-out card for quick
referen
ce. Also included is an electrode arm and metal bracket which can be easily attached to the left or right side of the meter according to your preference.
We take great pride in every instrument we manufacture and hope
this one serves you well.
1

2. GETTING STARTED

2.1 Keypad Functions

Powers the meter on and off. Upon power on, the meter automatic Calibration and memory values are retained even if meter is unplugged.
Toggle between available measurement modes—% saturatio switch to Temperature calibration during DO calibration modes.
Press and hold for 5 seconds to enter SETUP mode.
Toggles between measurement and calibration modes. In SET
Confirms calibration values in CAL mode. Confirms selections in SETUP mode. View recalled values in memory mode.
S MI (Memory Insert) stores values into memory.
ally begins in the mode that was last used.
n or concentration (mg/L or ppm). Also used to
UP mode, returns user to the measurement mode.
Increase value. Scroll up in SETUP & CAL modes.
MR (Memor T Decrease value. Scroll down in SETUP & CAL modes.
Freezes measured reading. Press again to resume live re
y Recall) recalls values from memory.
ading.
2

2.2 LCD Annunciators

2.3 Meter Connections
DC
ATC
DO
Power supply Temperature connection of Dissolved Oxygen electrode BNC connection of Dissolved Oxygen electrode
3

3. DISSOLVED OXYGEN / ATC ELECTRODE

The DO 700 includes a galvanic DO electrode with built in temperature measurement.
The electrode has a dual cable; a BNC connector for DO measurement and a mini­phono plug for temperature. The electrode is pre-filled with electrolyte that is separated from the sample by an oxygen permeable membrane. The membrane is pre-assembled for you and fixed to a detachable cap. The pre-assembled cap design allows simple replacement and fast conditioning.
The electrode is light-weight with a 12-mm bod
y diameter and a 16-mm tip diameter from at the detachable Noryl cap. The compact sensing area reduces air entrapment resulting in quick, accurate, and stable readings.
Provide simple stirring for best results—
ally achieve a minimum water flow rate of
ide 2 inch/second on the membrane.
The probe is not recommended for samples be
yond 0 to 50 °C.
The membrane is thin and can be replaced if damaged. Use care to protect from scratches, abras results keep membrane clean by rinsing after daily use. See Section 7
– Electrode Maintenance.
ion, or contact with solids. For best
4

4. CALIBRATION

The DO 700 has three measurement modes; DO as % saturation, DO as mg/L (or ppm) concentration, and temperature.
Dissolved oxygen levels vary with temperature, barometric pressure,
linity, so calibration must be performed with consideration of
and sa these factors. It is necessary to set the proper temperature,
barometric pressure and salinity values prior to performing any DO calibration or measurement.
See Section 4.1 — Temperature Calibration
See Section 5.6 — Barometric Pressure Adjustment. See Section 5.7 — Salinity Adjustment.
The DO 700 will accept two % saturation calibration points; 100%
g saturated air or air-saturated water, and 0% using zero oxygen
usin solution. When 100% or 0% calibration is performed, the concentration value is adjusted accordingly. Therefore, it is not necessary to calibrate the concentration mode. If calibrating for 0% oxygen, note that the meter will take several minutes to reach 0% saturation value and constant stirring is not required.
The following table lists calibration values in % saturation calibration mode
with two different barometric pressures. Note that the saturation value (92.1%) has decreased due to the lower barometric pressure entered.
.
% Saturation
(of factory default value)
less than 10% 0% 0%
10.1% to 49.9% CAL ERR CAL ERR 50% to 150% 100% 92.1%
Calibration Value
(760mmHg)
Calibration Value
(700mmHg)
If calibration is attempted from 10.1% to 49.9%, CAL Err message is shown—calibration is rejected and the display will return to calibration mode.
5
The DO 700 will accept one calibration point in concentration mode. The minimum value is 2 mg/L (ppm), and the calibration window is +/­40% of the factory default value.
Temperature and % saturation calibration should take place be attempting to perform mg/L (ppm) concentration calibration.
Calibration of the concentration mode concentration calibration and does not affect the % saturation calibration.
To offset your % saturation reading to match
Section 5.8 – Offset Adjustment
Perform daily calibration for best results. Ne automatically override the existing data. To completely recalibrate the meter and when installing a replacement electrode, it is best to clear all calibration data. See Section 5.10 – Reset

4.1 Temperature Calibration

For best DO accuracy, ensure that the temperature is accurate. The thermistor sensor used for automatic temperature compensation and measurement is accurate and stable, so frequent calibration isn’t required. Temperature calibration is recommended upon probe replacement, whenever the temperature reading is suspect, or if matching against a certified thermometer is desired.
The DO 700 will default to 25 ºC / 77 ºF when the ATC plug is not
ected.
conn
1) Place the pr obe into a solution with a known accurate temperature such as a co
nstant temperature bath.
will only replace the previous
another instrument, see
.
w calibration values will
.
fore
2) Press measured temperature while the secondary displa y shows the factory default temperature.
4) Adjust the tem
º
5
C (or ± 9ºF) from the factory default temperature.
followed by . The primary display shows the
perature using
to cancel. The meter allows an adjustable maximum value of ±
or . Press to accept or
6

4.2 % Saturation Calibration

The DO700 can be easily calibrated in air. For best accuracy, ensure that the barom
The barometric pressure factory default is 760 mm Hg (1.013 bar),
hich results in a theoretical calibration value of 100% saturation in air.
w If the barometric pressure setting has been changed from 760 mm Hg, the meter will automatically adjust to a new % saturation calibration value instead of 100%. This new value is correct for the adjusted barometric pressure.

4.3 100% Calibration

1) Rinse the probe well with clean water (do not dry).
etric pressure value is accurate.
2) Press
3) Hold the probe in the air with the tip facing downwards. Wait for
the read
4) Press
indicate the calibration point, 100.0.
5) Press
briefly before calibrating to 100.0%.
6) Press to return to measurement mode, or continue with 0%
calibration if desired.

4.4 0% Calibration

1) Rinse the probe well with clean water (do not dry).
2) Press
3) Submerse probe in zero oxygen calibration solution and stir
slo more than 10 minutes!
to select % saturation.
ing to stabilize.
. The CAL indicator will blink and secondary display will
to confirm the calibration. T he primary display will blink
to select % saturation.
wly. Wait for the reading to stabilize—this can often take
7
4) Press . The CAL indicator will blink and the secondary display
will indicate the calibration point, 0.0.
5) Press to confirm the calibration. T he primary display will blink
briefly before calibrating to 0.0%.
6) Press
calibration if desired (note 0% calibration followed by 100% calibration is not recommended---always perform 100% calibration first).

4.5 mg/L or ppm (Concentration) Calibration

Performing a saturation calibration at 100% or 0% will simultaneously calibr
ate the corresponding mg/L (ppm) concentration value. Therefore, additional mg/L (ppm) calibration isn’t required in most circumstances.
If desired, you can perform a calibration adjustment in mg/L (ppm)
ithout affecting your % saturation calibration value.
w
1) Rinse the probe well with clean water (do not dry).
2) Press
3) Dip the probe into a sample of known oxygen concentration (i.e.
determi to stabilize.
4) Press
briefly, before CAL flashes. The concentration value based on the factory default calibration is shown.
5) Press
oxygen concentration value.
to return to measurement mode, or continue with 100%
to select concentration mg/L (ppm).
ned by titration or another instrument). Wait for the reading
. The CAL indicator and current concentration display
or to adjust the reading to match the known
8
6) Press to confirm the calibration. T he primary display will blink
briefly before returning to measurement mode.
NOTE: calibration is limited to ±40% adjustment of the factory default value to prevent erroneous calibrations.
The minimum calibration value is 2 mg/L. In addition, the

5. SETUP FUNCTIONS

Use the setup feature to customize your instrument operation. There are two setup menus; one menu is derived from the %
saturation mode and the other from the mg/L (or ppm) concentration mode.
First, select the appropriate measurement mode you wish to adjust by pressi
ng
until the desired units are displayed. During
measurement, press and hold mode of the parameter being measured.
Press Press to select the program or confirm selection. Press to go back one level or return to measurement mode.

5.1 1.0 CAL (Calibration)

or to change programs or change options.
Press Note: % saturation mode will list two points, while concentration mode has only one point.
to view stored calibration points.
for 5 seconds to enter SETUP
9

5.2 2.0 ELE (Electrode Information)

Press TIP: The factor is an indication of the probe’s efficiency. It is the ratio of the actual mV produced by the probe to the theoretical mV value. The ratio displays from 0.5 to 1.999.
Press
Press Press

5.3 3.0 ConF (Configuration)

Press

5.4 3.1 rdY (Ready Stability Indicator / Auto Hold)

Press Press or to choose READY “On”, READY “OFF”, or
Auto HOLd. Press

5.5 3.2 ºC ºF (Celsius or Fahrenheit)

Press Press
to view calibration factor (CAL.F)
to view mV output for 100% (HI).
to view mV output for 0% (LO). to view Offset % (OFS)—% saturation mode only.
to access set-up programs 3.1 thru 3.5.
.
to confirm.
. or to select ºC or ºF.
Press to confirm.
10

5.6 3.3 dPr (Barometric Pressure Adjustment)

Press Press Press Press or to adjust the value, press to accept.

5.7 3.4 SALt (Salinity Adjustment)

For best accuracy, enter the actual salinity value if your samples are at least 0.1 ppt (100 ppm). Maximum adjustment is 50 ppt (50,000 ppm or 5% salt).
Press Press or to adjust the value (in parts per thousand).
Tip: 0.1 ppt = 100 ppm salinity Press

5.8 3.5 OFS (Offset Adjustment) from % mode

Use this feature to offset the meter’s value to another DO meter. Using the % saturation mode, observe the reading a sample solution after it has stabilized. Similarly, observe the reading of the same sample using another DO meter as a reference. The probe of the reference meter should be immersed in the same sample at the same depth.
The DO 700 allows +/- 10.0% offset adjustme user calibrations will reset the offset adjustment back to the factory default value of 0.0 %.
. or to select units; Hg (mm Mercury) or bAr. to confirm.
.
to confirm.
nt. Note that subsequent
Press Press or to adjust the % offset.
.
11
Press to confirm.

5.9 3.5 Unit (Unit Adjustment) from concentration mode

Press . Press or to select ppm or mg/L units. Press

5.10 4.0 rSt (Reset)

Press Press or to select “Yes” (Reset) or “No” (Cancel). If “Yes”, press or to select “Cal” (calibration reset
only) or “FCt” (complete reset to factory default settings). Press

5.11 5.0 CLr (Clear Stored Memory)

Press Press or to select “Yes” (Erase memory) or “No”. Press
to confirm.
.
to confirm.
.
to confirm.
12

6. TAKING MEASUREMENTS

During measurement the dissolved oxygen reading is automatically compensated for pressure and salinity. For best accuracy, ensure the pressure and salinity settings are adjusted accordingly from the setup menu. The factory default value for barometric pressure is 760 mm Hg or 101.3 bar (sea level). The factory default value for salinity is 0.0 ppt (no salinity).
See Section 5.6 — Barometric Pressure Adjustment See Section 5.7 — Salinity Adjustment
Press
to select the desired measurement mode:
Percentage Saturation (%) and Temp (
Concentration (mg/L or ppm) and Temp (
.
.
º
C or ºF)
º
C or ºF)
Follow these general rules when taking measurements:
Keep the membrane free from contact with solid objects.
Provide stirring of your solutionthis helps to overcome oxygen consumption of the probe and prevents air bubble entrapment.
Do not submerge the cable for extended periods.
Use with aqueous solutions only.
The pre-assembled cap must be completely submersed to obtain an accurate reading in solutions.
Rinse the probe well with DI water or rinse solution between measurements.
13

6.1 HOLD Function

For prolonged observation of a reading, press measurement mode to freeze the display. The “HOLD” indicator will display when the reading is held. To release the held value and
resume measurement, press memory by pressing .

6.2 Storing and Recalling Data

The DO 700 can retain up to 100 data points into memory for later retrieval.
1) In the measur ement mode, press
into memory. The stored memory location value (StO) is briefly displayed.
2) To recall data from memory, press . The location of the most
recent stored data is displayed first. Press location of the desired data, then press
3) Press to return to the stored data location. Press to return
to measurement mode. To erase stored data, see Section 5.11 —
Clear Stored Memory
again or insert the held value into
to insert the measured value
.
during
or to select the
to accept.
14

7. ELECTRODE MAINTENANCE

The DO 700 electrode produces an output proportional to the oxyge n present in the medium in which it is placed. The galvanic probe design allows for immediate use—unlike the typical 15 minute warm-up period required for polargraphic electrodes.
The probe consists of two parts. The upper p cathode, and dual cable. The lower part consists of a pre-assembled cap, and electrolyte solution.
Oxygen diffuses through the membrane onto the cathode, where it is consumed. This process produces an electrical current which flows through the cable to the meter. The electric current produced is proportional to the oxygen that passes through the membrane and the layer of electrolyte. This makes it possible to measure the partial pressure of oxygen in the sample at a given temperature.
Since the DO in the sample is consumed by the cathode it is essential to have flow past the membrane of the probe to prevent the occurrence of false readings. The probe uses very little oxygen for its measurement. This enables it to function correctly with liquid movement as low as 2 inch/sec across membrane.
The permeability of the membrane to oxygen varies greatly with temperature. Therefore compensation is needed for this variation. The DO 700 probe comes with built-in temperature compensation for the membrane variation.
Proper maintenance will help you maximize probe life and achieve most accurate
Deposits on the membrane surface act as a barrier to oxygen diffusing
h the membrane, so clean the membrane to assure maximum
throug reliability. After each use, rinse the probe with clean water to avoid any hardening of deposits or bacterial growth.
NOTE: Although the membrane is strong and not easily damaged, wipe it gently while cleaning it. If the membrane is punctured, damaged, or torn, the probe will not function properly.
readings.
art consists of the anode,
15
There are no special probe storage requirements.
Cap and Electrolyte Replacement
Replacement of the pre-assembled cap is required only when you cann
ot calibrate the probe, or if the membrane is damaged. Typical membrane damages are punctures or wrinkles caused during measurements or cleaning.
1) Unscrew the cap counter clockwise from the probe sensing tip.
2) Rinse the probe under running water.
3) Mount the nozzle tip onto the syringe provided. Fill the syringe
ith the refill solution through the tip of the plastic bottle.
w
4) Hold th e probe upside do wn. Insert the nozzle tip into one of the 4
hol
es surrounding the silver cathode. Inject the fill soluti on into the probe body until solution leaks out from the fill hole (approximatel y 5 mL).
5) Replace pre-assembled cap by tightening clockwise until hand ed.
tighten
6) Allow at least 1 hour for the electrode to equilibrate before
usage.
16
1
2
3
4
5
6
Electrolyte Solution
The electrolyte solution in your probe’s cap will deplete on usage and will need to be replaced periodically. The replacement electrolyte solution included with your probe comes premixed and ready to use. To order more electrolyte solution, see Section 9 – Replacements and
Accessories
.
17

8. DISSOLVED OXYGEN THEORY

Dissolved Oxygen (DO) refers to the volume of oxygen that is contained in water. There are two main sources of DO in water; atmosphere and photosynthesis. Waves and tumbling water mix air into the water where oxygen readily dissolves until saturation occurs. Oxygen is also produced by aquatic plants and algae during photosynthesis.
The amount of DO that can be held b
1) TEMPER
ATURE:
y water depends on 3 factors:
DO increases with decreasing temperature (colder water holds more oxygen)
ALINITY:
2) S DO increases with decreasing salinity (freshwater holds more oxygen than saltwater does)
TMOSPHERIC PRESSURE:
3) A DO decreases with decreasing atmospheric pressure (amount of DO absorbed in water decreases as altitude increases)
Solubility of oxygen in water contact with water saturated air at
16
14
12
10
8
6
Solubility mg/L
4
2
0
0 5 10 15 20 25 30 35 40 45 50
standard atmospheric pressure
Tem perat ure ° C
DO Solubility in Water vs. Temperature
18
Measurement Units
One measure of DO in water is parts per million (ppm) which is the number of oxygen molecules (O
) per million total molecules in a sample.
2
The “mg/L” unit is equivalent to ppm (a liter of water weighs 1 million milligrams—one part in a million is similar to one milligram in a liter). Calculating the % saturation is another way to analyze DO levels. % saturation is the measured DO level divided by the greatest amount of oxygen that the water could hold under various temperature and atmospheric pressure conditions multiplied by 100.
What Is Being Measured?
DO probes respond to the partial pressure of oxygen in liquid or gas being
ed—they measure the “pressure” of oxygen rather than
measur concentration. All of the oxygen entering the probe is consumed at the cathode where it is electrochemically reduced to hydroxyl ions producing an electrical current within the probe:
+ 2 H2O + 4 e- Æ 4 OH –
O
2
Since all oxygen entering the probe is chemi
cally consumed, the partial pressure of oxygen in the electrolyte is zero. Therefore, a partial pressure gradient exists across the membrane and the rate at which oxygen enters the probe is a function of the partial pressure of oxygen in the gas or in liquid being measured.
When a probe is placed in air saturated
water, the current it produces will not be affected by the temperature or salinity of the water. The DO concentration in the water, however, will vary with temperature and salinity. Because it is convenient to report DO concentration in mg/L or ppm, it is necessary to adjust for temperature and salinity of the water to get correct readings in these units.
If DO were to be reported in terms of partial p
ressure or % Saturation, then temperature and/or salinity compensation for oxygen solubility would not be necessary. Most probes are temperature compensated—i.e. they convert the “partial pressure measurement” to mg/L of DO at whatever temperature the water happens to be at for a given salinity and barometric pressure.
19
Air Calibration
Understanding the principle of air calibration is easy, once you know that it is partial pressure that the probe is responding to. When the probe is in air, it is measuring the partial pressure of oxygen in air. If water is air saturated, then the partial pressure of oxygen in the water will be the same as it is in air. Therefore, all you need to know is the temperature of the air in which the probe is placed. By consulting solubility tables for oxygen at the particular barometric pressure and salinity of the water being measured, the corresponding concentration (mg/L or ppm) can be found for air saturated water at the air calibration temperature, and the meter can be set accordingly. Because most meters are temperature compensated, they will still give correct readings in mg/L even though the actual water temperature may be different to the air calibration temperature. Note: The closer the air calibration temperature is to the water temperature, the more accurate the calibration.
Applications
Oxygen is essential for fish, invertebrate, plant, and aerobic bacteria respirati organisms. Levels below 2 or 1 ppm will not support fish. Fish growth and activity usually require 5 to 6 ppm of DO, an important consideration for Aqua-culture industry.
Low DO indicates a demand on the oxygen of the system. Natural organic material s demand as it is decomposed. Organic materials from human activities also create an oxygen demand in the system. Micro-organisms consume oxygen as they decompose sewage, urban and agricultural run-off, and discharge from food-processing plants, meat-packing plants and diaries. There is an optimum DO level for this process and if the DO level falls too low, the micro-organisms die and the decomposition ceases. When DO level is too high, this indicates too much power is used than necessary for aeration resulting in an inefficient process.
In boiler water application, presence of oxygen in the water will increase corrosio such instance it is critical to keep DO concentration to a minimum.
on. DO levels below 3 ppm are stressful to most aquatic
uch as leaves accumulate in the stream and create an oxygen
n and helps build up boiler scale that inhibits heat transfer. In
20
Some pollutants such as acid mine drainage produce direct chemical demands on oxygen in the water. DO is consumed in the oxidation­reduction reactions of introduced chemical compounds such as nitrate
1-
) and ammonia (NH
(NO
3
ferrous (Fe
+
) and ferric (Fe
2
These are important consideration for w
1+
), sulfate (SO
4
+
) ions.
3
2-
), and sulfite (SO
4
2-
3
) and
ater and wastewater treatment
industry.
21

9. REPLACEMENTS AND ACCESSORIES

Part number Ordering Code
Item Description
DO 700 meter with DO/ATC probe and 1-m cable, (2) Pre-assembled caps, electrolyte, integral stand, & 100/240 VAC adapter.
DO/ATC probe, 1-m cable, (2) Pre­assembled caps, electrolyte, & plastic syringe.
DO/ATC probe, 3-m cable, (2) Pre­assembled caps, electrolyte, & plastic syringe.
Pre-assembled cap 01X241608 — Pre-assembled cap, 10 mL
electrolyte, plastic syringe SMPS, 100/240 VAC, 9V, 6W power
adapter Zero Oxygen Solution (500 mL) 00653-00 DO probe electrolyte refilling solution
(60 mL)
Eutech Instruments
ECDO70042S
X543501
01
DO6HANDY
01
X233913
DO6HANDY3M
01
X233916
01X241609 35642-55
60X030130 35615-07
01X211226 —
Oakton Instruments
35415-00
35642-50
35462-52
22

10. TROUBLESHOOTING GUIDE

PROBLEM CAUSE SOLUTION
No display Main power not switched
“Ur” (Under range)
“Or” (Over range)
in primary or secondar display
Meter not responding to key press
Secondary display continually shows 0 and 100, during % calibration
on. AC Adapter socket not inserted properly.
DO or Temperature value is out of range.
Probe is not connected.
Electrode dirty or broken.
y
Meter not calibrated. Manual HOLD or Auto
HOLD is active.
Worn ke
Invalid key; Button is not functional in the current operation mod
The calibration value is too high to be accept 0% standard, and too low to be accepted as a 100% standard (10.1% to
49.9%), compared against factory default value.
ypad.
e.
ed as a
Switch on the power supply. Re-insert AC Adapter.
Confirm measurement condition, ensure temp is
Check electrode connections.
Check electrode-add electrolyte &/or replace me
Reset meter & recalibrate.
Press HOLD.
See Section 5.4 — Auto Hold disable Auto Hold feature if enabled.
Contact Technical Service. Press alternate key.
Select valid key depending on current mode.
Service probe. Check calibration solution. Check sample conditions (temp,
and pressure settings).
salinity Reset meter.
within range.
mbrane.
to deactivate manual
to
23
PROBLEM CAUSE SOLUTION CAL
CAL
Slow response or unstable readings
% calibration value is not acceptable. Electrode is disconnected or failing.
ppm or mg/L calibration is attempted calibrated absolute value is below 2.00
External noise pickup or induction caused b nearby electric motor
Insufficient electrolyte in probe
Air bubbles near sensor
Dirt Probe not deep enough in
sample Broken probe
when the factory
y
y or damaged probe
Check electrode connections. Check that corre selected & repeat cal procedure. Service electrode. Reset meter. Replace electrode.
Use concentration calibration solution higher than 2.00. Check electrode connections. Check that corre selected & repeat cal procedure. Service electrode. Reset meter. Replace electrode.
Remove or switch off interfering device
Fill probe with electrolyte & / or replace pre-
Stir or tap probe to remove bubbles, reduce
Clean the probe and re-calibrate Submerse sensor cap into sample
Replace probe o
ct DO unit is
ct DO unit is
assembled cap
stirring
r membrane
24

11. SPECIFICATIONS

DO % Saturation
Range Resolution Relative accuracy Offset adjustment Calibration Points
DO mg/L or ppm
Range Resolution Relative accuracy Calibration
Temperature
Range Resolution Relative accuracy Calibration
Salinity Correction Range Resolution Method
Pressure Correction Range Resolution Method
Probe (DO / Temp) Galvanic / Thermistor Avg. Response Time 60 seconds to achieve 95% of the reading Hold Function Yes Memory 100 data sets Input BNC for DO & 2.5 mm phono plug for temperature Power AC/DC 9V, 6W Adapter (100/240 VAC, 50-60Hz) Dimensions (mm) meter only = 175 (L) x 155 (W) x 69 (H)
(2 points) 100% in saturated air or air-saturated water,
(1 point) ± 40% from the factory default value.
Offset in 0.1 º increments; Offset range: ± 5 ºC / 9 ºF
Automatic correction after manual input
Automatic correction after manual input
± 10.0% of reading
0% in zero oxygen solution
Minimum reading allowed is 2.00.
0.0 to 50.0 ºC / 32.0 to 122.0 ºF
450 to 825 mmHg/ 0.6 to 1.1 bar
1 mm Hg / 0.1 bar
0 to 300%
0.1 %
± 0.5% of value
0.0 to 30.0
0.01
± 0.5% of value
0.1 ºC / 0.1 ºF
± 0.5ºC / ± 0.9ºF
0.0 to 50.0 ppt
0.1 ppt
25

12. WARRANTY

This meter is supplied with a warranty against significant deviations in material and workmanship for a period of THREE years from date of purchase whereas probe with a SIX month warranty.
If repair or adjustment is necessary a or misuse within the designated period, please return – freight prepaid – and correction will be made without charge. Eutech Instruments/Oakton Instruments will determine if the product problem is due to deviations or customer misuse.
Out of warranty products will be repaired on a charged basis. The warranty on your instrument shall not ap
nd has not been the result of abuse
ply to defects resulting from:
Improper or inadequate maintenance by customer
Unauthorized modification or misuse
Operation outside of the environment specifications of
the products
26

13. RETURN OF ITEMS

Authorization must be obtained from our Customer Service Department or authorized distributor before returning items for any reason. A “Return Material Authorization” (RMA) form is available through our authorized distributor. Please include data regarding the reason the items are to be returned. For your protection, items must be carefully packed to prevent damage in shipment and insured against possible damage or loss. Eutech Instruments will not be responsible for damage resulting from careless or insufficient packing. A restocking charge will be made on all unauthorized returns.
NOTE: Eutech Instruments Pte Ltd /Oakton Instruments res to make improvements in design, construction, and appearance of products without notice.
erve the right
27
NOTES
For more information on our products, please contact our channel partner or visit our websites listed below:
Oakton Instruments
625 E Bunker Court Vernon Hills, IL 60061 USA Tel: (1) 888-462-5866 Fax: (1) 847-247-2984

info@4oakton.com www.4oakton.com

Distributed by:
Eutech Instruments Pte Ltd
Blk 55, Ayer Rajah Crescent, #04-16/24 Singapore 139949 Tel: (65) 6778 6876 Fax: (65) 6773 0836
eutech@thermofisher.com www.eutechinst.com
Part of Thermo Fisher Scientific
Loading...