Eutech Instruments DISSOLVED OXYGEN PROBE User Manual

Eutech Instruments
Dissolved Oxygen
Probe
USER MANUAL
1

1 DESCRIPTION

N

1.1 The Eutech Probe

The probe is a galvanic measuring element, which produces a millivolt output proportional to the oxygen present in the medium it is placed in. It consists of an upper part with cathode, anode and cable, and a cap with membrane and electrolyte.
Oxygen diffuses through the membrane onto the cathode, where it reacts chemically and then combines with the anode. This chemical process develops an electrical current, which flows through a built in resistor. The resistor converts the current (microamps) into millivolts. This millivolt signal is led to the transmitter via a two-core cable.
The probe is virtually maintenance free, requiring only that its membrane be kept reasonably clean. The anode remains fully active throughout its entire lifetime, which may be several years. The membrane is very robust and can be wiped clean with a cloth or paper. It is recommended to clean the membrane from time to time to ensure maximum reliability.
The probe uses very little oxygen for its measurement. This enables it to function correctly with liquid movement as low as 2-cm / sec (measured at 7 ppm and 13
o
C). The membrane material was specially chosen to ensure a negligible
influence of deposits on its surface, when compared with traditional systems.
Remember that in any biologically active system, deposits will be found on any surface, including the membrane of an oxygen probe. The deposit acts as barrier to the oxygen diffusing through the membrane. It must, therefore, be cleaned at regular intervals, the frequency depending on factors such as the biological loading of the system, the oxygen consumption of the probe and of course the accuracy expected.
In a sewage treatment works the percentage of solids and the wastewater composition are important factors. In a fish farm the species, kind of feed, population and water exchange frequency are important. This is valid for all types of oxygen meters.
A small hole in the upper part permits pressure equalizing and CO2 release. The hole also enables the probe to be filled completely with electrolyte. The probe can be serviced IN SITU without the use of tools by exchanging the cap with a new (spare). The membrane in the cap can also be exchanged, giving a very considerable reduction in maintenance cost compared with conventional systems.

1.2 Which kind of transmitter can be used with the Eutech Probe

As indicated above, the probe can be considered a low impedance millivolt generator. Its output of about 30mV with the probe in air is temperature compensated and linearly proportional to the oxygen concentration, either in mg/l or % saturation depending on the choice of membrane.
The output impedance is very low (<1 Kohm) so noise problems normally do not exist. However, it is important that the input impedance of the following transmitter is at least 2 Mohms in order not to disturb the temperature compensation. It is also extremely important that inputs from probes are galvanically isolated from each other if more than one probe is connected to the same electronics.
ote that the probe is available in two different versions.
Type 1 is particularly suited for long term measurements at relatively high DO levels – 2 to 50 mg/l. This type is also used for % sat measurements
Type 2 is specially designed for low DO applications -
0.0 to 10mg/1 and equivalent % sat measurements.
Membranes for mg/l and % sat measurements differ. The mg/l membrane sinks in water, the % sat membrane floats.
2

2 INSTALLATION

The Eutech probe is easy to install. It should be placed where there is some movement in the water-1 cm/sec is enough at 7 mg/1 and 13 C. Ensure that the probe does not strike against the tank wall, and do not mount it directly above diffusers etc.
Each probe is connected using ordinary 2-core cable, and it is fitted with 3 meters of PU cable unless otherwise specified. The brown wire is positive, the blue negative.

3 DIRECTIONS FOR USE

3.1 Calibration

Calibration is carried out by placing the probe in air or air saturated water, waiting for temperature equalization then adjusting the transmitter or electronics connected so that the output corresponds to the calibration value. If you are measuring % saturation on the calibration value is always 100%. If you are measuring mg/l (ppm) temperature, the barometric pressure and the water salinity all affect the calibration value.
The probe must have the same temperature as the water or air surrounding it, and must therefore be allowed to attain this temperature before being calibrated. Any robust long-life probe such as the Eutech can take up to an hour to respond to a 10-degree temperature change in air, which only takes 10 minutes in water. This must be taken into account when calibrating.
Take the probe up from the water, wipe the membrane dry, and hang it in free air, away from direct sunlight. When the probe output is stable adjust the transmitter or electronics connected to give the correct output.
Calibration value for systems measuring % saturation
The calibration value for systems measuring % saturation is always 100. For these systems, and only for these systems, temperature, barometric pressure and salinity have no influence.
Calibration value for systems measuring mg/l (ppm)
For these measurements adjustment until the output corresponds to the appropriate value from the calibration tables. Temperature, pressure and salinity affect the value. The first table at the end of this section shows values for various temperature and air pressures, the other values for various temperatures and salinity at 760 mbar.
It is essential to correct for temperature-use the air temperature near the probe. At high altitudes, or for greater accuracy, a correction must be made for pressure. The salinity should be corrected for in salt-water measurements. For a simple calibration at low altitudes assume a normal barometric pressure of 760 mmHg.
How often should calibration be performed?
Unfortunately, it is not possible to answer this question. A new probe needs more frequent calibration than an “old”. Calibration frequently also depends on the accuracy desired. If an accuracy of +/- 0.5 ppm is sufficient, calibration once a month is more than enough. It is, however, much better to perform one thorough calibration than two quick ones. Give the probe time to stabilize, check the barometer and decide if correction for air pressure is necessary. Check the salinity if your farm is a salt-water facility!
3
Loading...
+ 5 hidden pages