EUTECH INSTRUMENTS CyberScan PCD 650 Quick Manual

Use keypad buttons to access
the functions displayed above it.
The fi rst function group appears when you enter
“Measurement” mode. Use and to access the 2nd and 3rd function groups.
Use or to toggle selection and increase
or decrease values in Setup & Calibration modes.
calibrations during Setup and Calibration procedures.
System Set Up Mode
Select “System” to customise your parameters and system settings, including password confi guration.
Quick Guide
Waterproof CyberScan PCD 650
pH/mV/Ion/Conductivity/TDS/Resistivity/Salinity/ Dissolved Oxygen/ºC/ºF
Keys & Functions
Keys Functions
Activates/Deactivates freezing of measurement
Activates calibration mode
Selects parameter
Shuts off meter (hold for 3 seconds)
Switches between diff erent function groups
Activates setup mode
Recalls stored data from memory
Stores data into memory
Escapes to 1
st
function group
Shows calibration report
Sends data to PC or printer
Activates temperature calibration mode
1st Function Group
2nd Function Group
3rd Function Group
Please turn overleaf >>
System - Page 1
• ‘Stable’ Indicator: ENABLE
• Stability Criteria: FAST
• Auto Hold: DISABLE
• Tem. Display From: pH/Cond
• Display Setting:
- 1st Row: pH
- 2nd Row: Conductivity
- 3rd Row: DO (%)
System - Page 2
Year: 2006
Month: Jan
Day: 01
Hour: 00
Minute: 00
Second: 00
System - Page 2
Year: 2006
Month: Jan
Day: 01
Hour: 00
Minute: 00
Second: 00
System - Page 3
Auto-O : ENABLE
On Time: 10 Mins
Back Light Settings:
- Permanently On: DISABLE
- On With Key Press: DISABLE
- On Time With Key: 1 Min
System - Page 3
Auto-O : ENABLE
On Time: 10 Mins
Back Light Settings:
- Permanently On: DISABLE
- On With Key Press: DISABLE
- On Time With Key: 1 Min
System - Page 4
Print Mode: IrDA
Data Format: CyberComm
Current Data Set: TIMED
Interval: 9 Secs
Fixed Setting: 2400 8-N-1
System - Page 4
Print Mode: IrDA
Data Format: CyberComm
Current Data Set: TIMED
Interval: 9 Secs
Fixed Setting: 2400 8-N-1
System - Page 5
Password Protection: DISABLE
Set Password: 00000
Confi rm Password: NO
System - Page 5
Password
Protection: DISABLE
Set Password: 00000
Confi rm Password: NO
System - Page 6
Clear Logged Data Memory: NO
Factory Reset: NO
System - Page 6
Clear Logged Data
Memory: NO
Factory Reset: NO
Part of Thermo Fisher Scientifi c
68X068067 07/08 Rev 1
pH, Ion & Temperature Set Up Mode
Access “pH” to confi gure pH calibration-related settings, alarm set points and pH resolution.
Access “Ion” to select measurement units.
Access “Temperature” to select measurement units and temperature compensation mode.
pH - Page 1
Buff er: USA
Cal Points: 3
pH Alarm: DISABLE
Set Points:
- Hi pH: 20.00 pH
- Lo pH: 0.00 pH
pH - Page 1
Buff er: USA
Cal Points: 3
pH Alarm:
Set Points:
- Hi pH: 20.00 pH
- Lo pH: 0.00 pH
- Lo pH: 0.00 pH
pH - Page 2
• pH Calibration Due: 5 DAYS
• pH Resolution:
0.00
Ion - Page 1
Ion Unit: ppm
Ion - Page 1
Ion Unit: ppm
Temp - Page 1
• Unit: ºC
• Mode: ATC
Conductivity, TDS, Salinity, Resistivity Set Up Mode
Access “Conductivity”, “TDS”, “Salinity” and “Resistivity” to confi gure pure/linear temperature
coeffi cients, normalisation temperature, cell constant, calibration-related settings as well as alarm set points limits.
Conductivity - Page 1
• Pure Water Coeffi cient: ENABLE
• Linear Temp Coeffi cient: 2.100
• Normalisation Temp: 25.0
• Cell Constant: 1.000
• Calibration Due: DISABLE
• Calibration Points: SINGLE
• Calibration Std: AUTO
TDS - Page 1
• Pure Water Coeffi cient: ENABLE
• Linear Temp Coeffi cient: 2.100
• Normalisation Temp: 25.0
• Cell Constant: 1.000
• Calibration Due: DISABLE
• Calibration Points: SINGLE
• TDS Factor: 0.500
Salinity - Page 1
• Pure Water Coeffi cient: ENABLE
• Linear Temp Coeffi cient: 2.100
• Normalisation Temp: 25.0
• Cell Constant: 1.000
• Calibration Due: DISABLE
• Calibration Points: SINGLE
Resistivity - Page 1
• Pure Water Coeffi cient: ENABLE
• Linear Temp Coeffi cient: 2.100
• Normalisation Temp: 25.0
• Cell Constant: 1.000
• Calibration Due: DISABLE
• Calibration Points: SINGLE
Resistivity ­Page 2
• Resistivity Alarm: DISABLE
• Set Points:
- Hi: 2.00 Ω
- Lo: 1.00 Ω
Salinity ­Page 2
• Salinity Alarm: DISABLE
• Set Points:
- Hi: 2.000 ppm
- Lo: 1.000 ppm
Cond - Page 2
• Conductivity Alarm: DISABLE
• Set Points:
- Hi: 2.000 µS
- Lo: 1.000 µS
TDS - Page 2
• TDS Alarm: DISABLE
• Set Points:
- Hi: 2.00 ppm
- Lo: 1.00 ppt
Dissolved Oxygen % Saturation & Concentration Set Up Mode
Access “Setup” to confi gure calibration-related settings, unit of measure, pressure and salinity settings
as well as alarm set points limits.
DO (%) - Page 1
• Off set Cal: 0.0 %
• Calibration Due (%): 5 DAYS
• Select Pressure Unit: mmHg
• Measured Pressure: 760.0
• Adjusted Pressure: 760.0
• Pressure Comp: ENABLE
DO (mg/L) - Page 1
• Select DO Unit: mg/L
• Auto Salinity Comp: DISABLE
• Set Salinity:
0.0 ppt
DO (%) - Page 2
• DO Alarm: DISABLE
• Set Points:
- Hi DO: 200.0 %
- Lo DO: 10.0 %
DO (mg/L) - Page 2
• DO Alarm: DISABLE
• Set Points:
- Hi DO: 20.00 mg/L
- Lo DO: 1.00 mg/L