Eutech Instruments CYBERSCAN PC 5000 User Manual

y..
.
Instruction Manual
CyberScan PC 5500/ 5000
Bench pH, Ion & Conductivity Meter/ Bench pH & Conductivity Meter
(Also Applicable for CyberScan PCD 5500)
Technolo
adeEas
68X292329
Rev.1 08/03
PREFACE
Thank you for selecting a Eutech Instruments CyberScan Series 5000 bench meter. This meter measures pH, conductivity and ion (PCD5500 & PC5500 only).
The instruction manual serves to explain the use of the CyberScan Series 5000 bench meters as a step-by-step operational guide to help you familiarize with the meter’s features and functions. It is structured sequentially with illustration of diagrams that explains the various functions and setup menus available.
This manual is written to cover as many anticipated applications and uses of the CyberScan 5000 series bench meters as possible. If there are doubts in the use of the meter, please do not hesitate to contact the nearest Eutech Instruments’ Authorized Distributors or call us at (65) 6778-6876 for Eutech Instruments’ Customer Service Dept. for assistance.
Kindly remember to complete the warranty card and mail it back to your Authorized Distributors or Eutech Instruments Pte Ltd.
Eutech Instruments reserve the rights to change, make improvement and modify specifications without prior notice and cannot accept any responsibility for damage or malfunction to the instrument caused by improper use.
Copyright © 2002 Eutech Instruments Pte Ltd All rights reserved. Rev.1 08/03.
TABLE OF CONTENTS
1 INTRODUCTION 1
1.1 Introducing the CyberScan Series 1
2 UNPACKING THE METER 2 3 GETTING STARTED 3
3.1 Connectors (PCD 5500 only) 3
3.2 pH Electrodes & Ion Specific Electrodes (PCD5500 & PC5500 only 6
3.3 Conductivity Probes 8
4 USING THE METER 10
4.1 Touch screen Operation 11
4.2 Choosing a channel (for PCD5500 & PC5500 only) 12
4.3 Button Functions 14
5 SYSTEM SETUP 16
5.1 Access system setup 17
5.2 Select Language 19
5.3 Set Date 21
5.4 Set Time 23
5.5 Set Beeper Status 25
5.6 Select procedural level 27
5.7 Set Print Configuration 29
5.8 Set Baud Rate 31
5.9 Set Number of Bits 32
5.10 Set Stop Bits 33
5.11 Set Parity 34
5.12 Set Operator 35
5.13 Set Display Contrast 37
5.14 Display Meter Information 39
5.15 Reset to Factory Defaults 40
6 pH SETUP 41
6.1 Access pH setup 42
6.2 Set Sample ID# 44
6.3 Select Buffer Group 48
6.4 Set pH Custom Buffer Group 50
6.5 Select Buffer Recognition 52
6.6 Select Auto Read Mode 53
6.7 Set pH Stability Criteria 54
6.8 Set Default Temperature 56
6.9 Set Isopotential Point 58
6.10 Set Alarm Limits 60
6.11 Set Print Criteria 62
6.12 Set Print Interval 64
6.13 Set Data Storage Criteria 66
6.14 Set Display Resolution 68
6.15 Set Display Configuration 70
6.16 View Stored Data 72
7 mV SETUP 76
7.1 Access mV Setup 77
7.2 Set Display Resolution 81
8 ION SETUP 83
8.1 Access Ion Setup 84
8.2 Select Ion Method 86
8.3 Select Measurement Units 88
8.4 Select Electrode Type 90
8.5 Set Significant Digits 92
9 CONDUCTIVITY SETUP 94
9.1 Access the Conductivity Setup 95
9.2 Select Cell Constant 99
9.3 Select Reference Temperature 101
9.4 Set Temperature Coefficient 103
9.5 Perform Replatinization 105
9.6 Set Significant Digits 109
10 pH OPERATION 111
10.1 Standardizing 113
10.2 pH Measurements 117
11 mV OPERATION 119
11.1 mV Measurement 120
12 ION OPERATION 121
12.1 Direct Reading Methods 123
12.2 Incremental Methods 129
12.3 Known Addition Method 131
12.4 Known Subtraction Method 135
12.5 Analate addition 137
12.6 Analate Subtraction 139
13 CONDUCTIVITY OPERATION 141
13.1 Conductivity Standardization 143
13.2 Measurement 146
14 CLEANING 148 15 TROUBLE SHOOTING 149 16 COMPLIANCE 150 17 APPENDIX 151
17.1 Determining Isopotential Points Experimentally 151
17.2 Data Management 152
17.3 Factory Default Settings 153
17.4 pH Theory 155
17.5 ISE Theory 160
17.6 Conductivity Theory 164
18 METER SPECIFICATIONS 169 19 ACCESSORIES 171
19.1 Replacement Meters and Adapters 171
19.2 General Purpose pH / ORP Electrodes SP 172
19.3 Specialty pH Electrode NP 173
19.4 pH Buffer Solutions & Single Use Sachets SP 174
19.5 ORP Solutions SP 174
19.6 Conductivity & TDS 442 Standard Solutions SP 174
20 WARRANTY 175
Instruction Manual CyberScan PC5500/ 5000
1 INTRODUCTION
1.1 Introducing the CyberScan Series
Thank you for selecting a Eutech Instruments CyberScan Bench meter. This manual describes the operation of the CyberScan Series 5000 bench meter. The state-of-art meter that you have purchased is easy to operate and will guide you through the various functions by displaying easy to understand prompts. This instruction manual should answer any questions that might arise in operating your meter, however, do not hesitate to call our Technical Support at (65) 6778 6876 (ext 839) or fax at (65) 6773 0836.
This bench meter provides microprocessor precision in a compact benchtop design that is easy to use. One touch screen controls all procedures, letting you:
Measure pH, absolute mV, relative mV or Conductivity  Select one of three sets of standard buffer groups  Implement automatic buffer recognition  Standardize with up to five standard or customer buffers  Customise your display screen and +  operating parameters  Assign operator and sample identification numbers  Store 250 data points (1000 for PCD5500) in the meter’s
memory or transfer data to a computer or printer.
Access extensive online help with just a touch a button
It all adds up to rapid, completely automatic, intuitive operation.
You will find this symbol appearing in this manual; it indicates useful tips that ease your meter operation.
1
Instruction Manual CyberScan PC5500/ 5000
2 UNPACKING THE METER
The following is a listing of what you should have received with your new CyberScan meter.
Meter package includes
Meter Power adapter (120 VAC/ 12VDC) OR (220 VAC/ 12 VDC)
depending on order code.
Electrode arm support bracket  Electrode arm ATC probe Instruction Manual
If any of these items are missing, please contact your nearest Authorized Distributor.
Accessory conductivity probes and ion selective electrodes are available and can be ordered by contacting any Authorized Distributor or Eutech Instruments Marketing (65) 67786876.
2
Instruction Manual CyberScan PC5500/ 5000
3 GETTING STARTED
3.1 Connectors (PCD 5500 only)
1. Review the layout and arrangement of the rear connector panel.
Back Panel of PCD 5500
3
Instruction Manual CyberScan PC5500/ 5000
Back Panel of PC 5500/ 5000
4
Instruction Manual CyberScan PC5500/ 5000
2. Connect the electrode arm to the base.
3. Connect the power cable to the connector cable to the rear connector panel power jack and to a power source.
To connect RS232, see Data Management on Section 17.2.
5
Instruction Manual CyberScan PC5500/ 5000
3.2 pH Electrodes & Ion Specific Electrodes (PCD5500 &
PC5500 only
This meter allows you to use two types of pH electrodes: the conventional glass pH electrode and the field effect transistor (FET) pH electrode. If both types of pH electrodes are connected, the meter will read the FET electrode.
1. Carefully remove the protective cover from the end of the electrode. Before first using your glass pH electrode, soak it 2-4 hours in an electrode storage solution, pH 4 buffer, or KCI solution.
2. Remove the shorting cap on BNC connector. Connect the combination pH electrode in to the BNC input connector.
If a combination electrode isn’t used, connect the indicating pH
electrode into the BNC input connector. Plug the reference electrode into the reference pin jack. Also, install the ATC probe
into the ATC jack.
Note: be sure to connect all probes to the appropriate channel connectors (for example: Input 1, Ref 1 and ATC1).
Option: Connect the optional FET electrode by plugging it into
the FET jack on the back meter panel. Allow the FET electrode to warm up five minutes before use.
Connect ion specific electrodes in the same manner as pH electrodes.
If both a conventional and a FET electrode are connected to the meter, do not put them in a solution together because you will get inaccurate measurements. Do not discard the BNC shorting cap.
6
Instruction Manual CyberScan PC5500/ 5000
3. Rinse and blot-dry (don’t wipe) electrodes between each measurement. Rinse electrodes with distilled or deionized water, or a portion of the next solution to be measured.
4. Between measurements, store conventional pH electrodes storage solution, pH 4 buffer, or KCI solution. Always leave the filling hole of liquid filled combination electrodes open. Refill when the level of solution gets below the manufacturer’s recommended level.
Proper electrode care is fundamental to obtaining reliable pH measurements. Improper care of the electrode may cause the meter reading to drift, respond slowly, or produce erroneous readings. For this reason, the electrode should always be conditional and used in accordance with manufacturer’s instructions.
7
Instruction Manual CyberScan PC5500/ 5000
3.3 Conductivity Probes
This meter allows you to use two types of conductivity cells: the 2-cell conductivity cell with dual pin connector and the 4-cell conductivity cell with DIN connector. If both a 2-cell and a 4-cell conductivity probe are connected, the meter will read the 4-cell conductivity probe.
1. Carefully remove the protective cover from the end of the conductivity cell. Before using your conductivity cell, soak it for 5 to 10 minutes in distilled or deionized water.
2. Connect the 2-cell conductivity cell to the 2-cell jack. Install the ATC probe in the ATC jack.
OR Connect the 4-cell conductivity cell to the 4-cell jack. The 4-cell
conductivity cells have built in ATC probes. Therefore, there is no need to install a separate ATC probe.
8
Instruction Manual CyberScan PC5500/ 5000
3. Rinse and blot-dry (don’t wipe) probe between each measurement. Rinse probe with distilled or deionized water, or a portion of the next solution to be measured.
4. Between measurements, store conductivity probes dry.
9
Instruction Manual CyberScan PC5500/ 5000
4 USING THE METER
E U T E C H
INSTRUMENTS
11:11 am
Touch anywhere to resume
10
Instruction Manual CyberScan PC5500/ 5000
4.1 Touch screen Operation
This new CyberScan research benchtop meter operates with a state of art touch screen. The touch screen makes this the easiest meter on the market to operate and care for. When this meter is first plugged in, the STANDBY screen will appear. Touch anywhere on this screen to access the functions of the meter.
The buttons on the right side of the screen control all of the functions of the meter. Alight touch on the screen is all you need to access the various functions. Once you touch a button you will get an audible tone; the screen will not change until you lift your finger.
This design prevents rapid uncontrolled scrolling through the various function screens. Easy to understand prompts guide you through the operation of the meter in the selected mode. If you are ever in doubt about what to do, just touch help on the bottom right corner of the screen for detailed information about the screen.
The touch screen is made of a durable polyester material that is chemically resistant. Maintenance is simple with this meter. To clean the screen you just need to wipe it with a damp cloth and dry it with a clean dry towel. For additional information, see cleaning and troubleshooting section of the manual (page 149).
11
Instruction Manual CyberScan PC5500/ 5000
J
J
n
4.2 Choosing a channel (for PCD5500 & PC5500 only)
Main Scree
Channel Screen
E U T E C H
INSTRUMENTS
pH
E U T E C H
1
INSTRUMENTS
CyberScan PC 5500
Select from the options to the right
January 15, 2002 11:11 am
mV
Ion
Cond Cond
setup
Measure
pH channel 1
9.295
ID#
00000
auto buffer
Buffers
anuary 15, 2002 11:11 am
µS/cm channel 1
ATC
ON
slope
000.0%
mV
0000.0
Last std: Jan 15 @ 11:11
CyberScan PC 5500
Select from the options to the right
anuary 1 5, 2002 11:11 am
1
2
25.0°C
am
2
dual
0.010
ID#
00000
ATC
26.2°C
Ref
25°C
Co
0.00%/°C
STND
Last std: Jan 15 @ 11:11
Dual Screen
am
12
Instruction Manual CyberScan PC5500/ 5000
The CyberScan PCD5500/ PC5500 are multi-channel meters. With this meter you can switch from channel 1 to channel 2. You can also view both channels at the same time by accessing the dual channel mode. If you set the meter to view both channels, you cannot change the parameters without choosing the screen that you want to modify. The setup parameters for each screen independently of one another.
Choosing a channel
1. Touch anywhere on the standby screen. OR Touch channel on the main screen THEN
2. Touch 1, 2 or dual to access the channel that you want to view.
To access System Setup from the standby screen
1. Touch anywhere on the standby screen.
2. Touch 1, 2 or dual to access the channel that you want to view.
3. Touch setup.
4. Touch system on the setup screen.
If you are in any measure made, touch mode until you access the main screen.
13
Instruction Manual CyberScan PC5500/ 5000
4.3 Button Functions
The touch screen of your CyberScan bench meter has “buttons” along the right side of the screen that are common to many of the screens. The following indicates the function of these common buttons.
This is the standby button and it allows you to access the standby mode. When in standby, the meter will not take measurements. It is in a state of rest. When you touch stdby the meter will return to the standby screen which says “Eutech Instruments” and displays the time.
This button allows you to access the channel screen to choose what channel you want to view. This button appears only on dual-channel meters such as PC 5500 and PCD 5500. This button also allows you to move from the main mode to the channel screen to access the standby mode.
The mode button allows you to switch between the various
mode
enter
save
close
more
back
operations of the meter. These operations, depending upon which meter you have, include four measuring modes (pH, mV, ion and conductivity) as well as a setup mode.
The enter button allows you to accept any changes that you have made on one setup screens or accept values that you have input with keypads. When touched, enter will save the changes and return you to the previous screen.
The save button allows you to save multiple changes that you have made on one screen as a group. It functions like the enter button does for an individual changes.
The exit button allows you to leave the screen you are currently viewing and return to the previous screen without making any changes.
The help button allows you to access helpful information on any screen. When you touch the help button, information about the current screen appears. This information will include step by step instructions for operating the meter from the current screen and possible applications information for that screen.
The close key appears on the bottom of all help screens and allows you to exit the help screen and return to the previous screen.
The more button appears on the help screens and allows you to advance to the next help screen for additional information.
The back button appears on the help screen and allows you to move back to a previous help screen.
14
Instruction Manual CyberScan PC5500/ 5000
This button accesses the standardization screen from the
std
meas
setup
print
S
edit
clear
BS
BS
delete
prev
next
various measurement modes and initiates standardization of the meter once the standardization screen is accessed.
This button is the measure button and directs the meter to measure your sample when in the Auto Read function of the pH mode.
This button will access the setup screens for the measuring mode that you are currently using. It can also be used to access the system setup screen that allows you to set parameters that are not related to measurements such as the times and the date.
The print button will send information to the output device that you have connected to your CyberScan meter. The output device can be a printer, data logger or a computer. In addition to this, touching the print button will also send data to the data storage center of the meter if a sample ID # has been assigned to your sample.
The arrow keys on he screen move the cursor up and down in order to highlight parameters that you would like to review or edit.
The edit button appears on the setup screens. After you have highlighted a parameter that you would like to change, the edit button allows you to access the available options for that parameter.
The clear button allows you to remove a setup parameter or standard buffer value from the meter’s memory that may have been entered at a previous time or by a previous user that is no longer of value to you. Touching the clear button erases the value so you may enter a new one. It can also erase the data from the memory of the meter.
The BS button is a backspace button. It appears on keypad screens and it allows you to back up and delete a character entered in error.
The delete button appears on the “View Stored Data” screens. This button allows you to erase the data from the memory of the meter.
The prev button appears on the Data Screens when the data stored in the meter’s memory has been accessed. It allows you to scroll through data points sorted and stored prior to the current data point displayed.
The next button appears on the Data Screens when the data stored in the meter’s memory has been accessed. It allows you to scroll through data points sorted and stored prior to the current data point displayed.
15
Instruction Manual CyberScan PC5500/ 5000
5 SYSTEM SETUP
System Setup
S
SYSTEM SETUP OPTIONS
-
Select
Language
- Set
Date
T
Time
- Set
- Set
Beeper Status
- Set
Print Configuration
edit
- Set
Operator
- Set
Display Contrast
- Display
- Reset to
Meter Information
Factory Defaults
16
Instruction Manual CyberScan PC5500/ 5000
J
5.1 Access system setup
E U T E C H
INSTRUMENTS
1
E U T E C H
INSTRUMENTS
pH
CyberScan PC 5500
Select from the options to the right
January 15, 2002 1 1:11 am
Setup
Select from the options to the right
2
dual
Cond Cond
pH
mV
Ion
CyberScan PC 5500
Select from the options to the right
anuary 15, 2002 11:11 am
System Setup
SYSTEM SETUP OPTIONS
- Selec t
Language
- Set
Date
Time
- Set
- Set
Beeper S tatus
- Set
Print Configuration
- Set
Operator
- Set
Display Contrast
- Disp lay
Mete r Infor matio n
- Rese t to
Factory Defaults
mV
Ion
setup
S
T
edit
Cond
setup
17
Instruction Manual CyberScan PC5500/ 5000
The system setup function allows you to customise the meter display options to meet your personal preference. Once set, these will rarely need to be changed.
To access System Setup
1. Touch anywhere on the standby screen
2. Touch setup on the main screen
3. Touch system on the setup screen The system setup options are now displayed on the screen.
To access a System Setup option
1. Use the arrow keys to scroll through the setup options and
highlight the option to be reviewed.
2. Touch edit to view the current status of the selected option.
The following is a detailed description of the system setup option screens
18
Instruction Manual CyberScan PC5500/ 5000
5.2 Select Language
Set
Language
Current LANGUAGE
English
French
German
Spanish
Use to highlight desired language
arrow keys
and then touch to accept
enter
English
S
T
enter
19
Instruction Manual CyberScan PC5500/ 5000
This option allows you to choose the language in which all prompts and directions will appear on the touch screen.
To Select Language
1. Access the select language screen from the system setup
screen. The current language is displayed on the screen.
2. Use the arrow keys to highlight the desired language
3. Touch enter to accept the language and return to the system
setup screen.
OR
Touch exit to return to the system setup screen, without making
any changes.
20
Instruction Manual CyberScan PC5500/ 5000
5.3 Set Date
Set
Date
enter
Current MM/DD/YY
1 2 3
4 5 6
7 8 9
BS 0 /
Touch clear to delete current date
Use numeric touchpad to input the new
date and then touch enter to accept
D/M/Y
clear
21
Instruction Manual CyberScan PC5500/ 5000
This screen can be used to set the present date which will be displayed on the measure screens. This date will also be printed on demand and stored in the data storage center of the meter when data is saved. There are two format options for the date: month/ day/ year
(M/ D/ Y) or the European format of day/ month/ year (D/ M/ Y)
To Set Date
1. Access the Set Date screen from the System Setup screen. The
current date and numeric keypad are displayed on the screen.
2. Touch clear to delete the current entry.
3. Touch D/ M /Y or M/ D/ Y to set the date format.
4. Use the numeric touch pad to enter the desired date, separating
the day, the month and the year by touching the / key on the keypad.
5. Touch enter to accept the date in the current format.
OR
Touch exit to return to the system setup screen, without making
any changes.
If you do not use /, the meter will not accept the date entry.
22
Instruction Manual CyberScan PC5500/ 5000
5.4 Set Time
Set
Time
enter
Current 12HR:
10:40 am
1 2 3
4 5 6
7 8 9
BS 0 /
Touch to delete current time
clear
Use to input the new
numeric touchpad
time and then touch to accept
enter
24hr
pm
clear
23
Instruction Manual CyberScan PC5500/ 5000
This screen can be used to set the present time which will be displayed on the measure screens. This time will also be printed on demand and stored in the data storage center of the meter when data is saved. There are two formats options for the time. The clock can be set as either a 12 hour clock or a 24 hour clock.
To Set Time
1. Access the set time screen from the System Setup screen. The
current time and numeric keypad are displayed on the screen.
2. Touch clear to delete the current entry.
3. Touch the 24hr or 12hr button to format the clock as either a 12
hour or a 24 hour clock.
4. Touch am or pm to set the appropriate time.
5. Use the numeric touch pad to enter desired time, separating the
hour and the minutes by touching the “
6. Touch enter to accept the time in the current format
OR
Touch exit to return to the System Setup screen, without making any changes.
Current 12HR
Current 24HR
5:15 pm
17:15
: key in the keypad.
If you do not use the “:, the meter will not accept the time.
24
Instruction Manual CyberScan PC5500/ 5000
5.5 Set Beeper Status
Set
Beeper Status
S
Current BEEPER STATUS
-
After STABLE Readin
OFF
- After KEY Touch
- On LIMIT Exceeded
Use to highlight beeper
arrow keys
option and then touch to change
Touch to accept changes
save
ON/OFF
OFF
OFF
T
ON
save
25
Loading...
+ 152 hidden pages