Instruction Manual CyberScan CON 1500
9
4.2. Overview of Setup Pages in CyberScan CON 1500
Each mode selection (conductivity, resistivity or total dissolved solids)
has similar setup pages except for TDS mode which includes Setup
P6.0 for setting TDS factor. Refer to Overview of Setup Pages shown
below.
Mode Setup Pages Accessible
uS P1.0 to P15.0 except P6.0
kohm P1.0 to P15.0 except P6.0
TDS P1.0 to P 15.0
)
If you do not wish to make a selection or change the option, pressing
std key at any point of time allows you to exit the setup page/mode
and return to the measurement mode.
Overview of Setup Pages in CyberScan CON 1500
P 1.0: View calibration
data
- View the last calibrated
buffer value depending on the
display mode selected.
P 3.0: Set the
Temperature Coefficient
- Select temperature
coefficient value in the range
of 0.000 to 10.000%.
P 5.0: Set single- or
multi-point calibration
- Select ON to activate singlepoint calibration, OFF for
multi-point calibration.
P 4.0: Select the
Normalised temperature
- Set the normalised
temperature in the range of
15.0 to 30.0°C.
P 7.0: Set Temperature
Unit
-Set the temperature unit to
be °C or °F.
P 8.0: Set the Baud
Rate
- Set baud rate as
4800, 9600, 19200 or
38400 bps.
P 2.0: Set the cell
constant
- Adjust and select cell
constant to be 1.00, 10.00 or
0.10.
Print
Cell Constant
P 9.0: Set the Parity Bit
- Set parity bit as none(0),
odd(1) or even(2).
P 10.0: Set the Stop Bit
- Set stop bit as one(1) or
two (2).
P 11.0: Select the Print
Data Option
- Select to print current or
stored data.
COEFF
P 6.0: Set TDS Factor
- Set the TDS Factor from a
range of 0.40 to 1.00.
P 14.0: Clear the user
calibration
- Select to clear respective
user calibration in the
current mode.
View
select
Print
Print
Print
clear
clear
BUFFER
STABLE
%
TDS
P 5.0: Select the Auto or
Manual cal options.
(Only applicable for
Conductivity mode)
- Set Auto calibration or
Manual calibration for uS
P 8.0: Set the stability
indicator
- Select ON to activate the
stability indicator, OFF to
de-activate.
OFF
P 13.0: Clear the memory
- Select to clear stored
memory.
P 15.0: Replatinisation
process setting
- Select YES to replatinise
probe, NO to exit. Default
replatinisation duration is 5.00
minutes.