Eutech Instruments ALPHA PH 100 User Manual

Page 1
p
Alpha pH 100 C/CX
pH/ORP
Controller / Transmitter
INSTRUCTION MANUAL
MODE
H/OR
7.0
o
SETU
C
Technology Made Easy ...
READY
ENTER
Page 2
Preface
This manual serves to explain the use of the Alpha pH 100 pH/ORP Controller/Transmitter. The units covered are Alpha C and CX pH/ORP Controller/ Transmitter.
The instruction manual functions in two ways: first, as a step by step guide to help you operate and understand the operation of the unit and second, as a handy reference guide.
The information presented in this manual is subject to change as improvements are made, and does not represent a commitment of Eutech Instruments Pte Ltd.
This instruction manual is written to cover as many anticipated applications of the Alpha pH 100 C/CX pH/ORP Controller/Transmitter as possible. If there are doubts in the use, please do not hesitate to contact the nearest Eutech Instruments’ Authorized Distributor or us at (65) 778-6876.
Eutech Instruments cannot accept any responsibility for damage or malfunction of the unit due to improper use of instrument.
Copyright 1994 Eutech Instruments Pte Ltd. Version 1.0
1
Page 3
TABLE OF CONTENTS
1. Introduction------------------------------------------------------------ 3
2. Getting Acquainted--------------------------------------------------4
2.1 Front Panel.............................................................4
2.2 Back Panel............................................................. 5
2.3 Selecting pH/ORP Measurement Mode..................6
2.4 Wiring.....................................................................6
3. Operating the Controller -------------------------------------------7
3.1 The Main Display.................................................... 7
4. Setting Up the Controller ------------------------------------------ 8
4.1 Setting and Changing the Password....................... 8
4.1.1 Setting New Password.........................................................8
4.1.2 Changing the Password.......................................................9
5. Calibrating the Controller---------------------------------------- 10
5.1 The Lower Level Menus........................................10
5.2 Calibrating for pH Measurement........................... 10
5.2.1 The CAL PH menu.............................................................10
5.3 Changing the Offset..............................................11
5.4 Calibrating for ORP Measurement........................12
5.5 Calibrating for Temperature Measurement ...........12
5.5.1 The CAL °C Menu..............................................................12
5.5.2 ATC ON/OFF.....................................................................13
5.6 Setting the Alarm Feature ...................................15
5.6.1 The Hi SET Menu ..............................................................15
5.6.2 The Hi HYS Menu..............................................................16
5.6.3 The Lo SEt Menu...............................................................17
5.6.4 The Lo HYS Menu .............................................................18
6. Using the Controller Current Loop for Datal oggi ng (for Transmitter Model Only)------------------------------------------------- 19
7. Additional Information-------------------------------------------- 20
8. Appendix 1 - Jumper Positions (I nside Controller) ------- 21
9. Appendix 2A - Jumper Positions (pH) ----------------------- 23
10. Appendix 2B - Jumper Positions, ORP --------------------- 24
11. Appendix 3 - pH Offset / Slope-------------------------------- 25
12. Technical Specifi cations ---------------------------------------- 26
2
Page 4
1. Introduction
The alpha pH 100 series pH/ORP Controller/ Transmitter is the latest innovation of process controllers from Eutech Instruments. Incorporated with the ASIC (Applic ation Specific Integrated Circuit) m i croprocessor technology, thi s panel­mounted on-line controller provi des many user-friendly features desirable in pH/ORP controll ers.
This versatile pH/ORP unit can be used either for measuring and monitoring pH or ORP values at one time. It cannot be used for measuring both parameters simultaneous l y. When shipped, it will be set internally to the pH mode by means of an internal jumper. This jumper can be reset by you to select between the pH and ORP mode.
Some of the features of this Controller are :
• Automatic temperature compensation (with PT 100)
• Manual temperature compensati on setting without the
temperature probe
• Push button calibration up to 3 points
• Offset adjustment
• High and low alarm triggering relays
• User-defined password feature to prevent unauthorized
entry to change the calibration data, hysteresis function, high and low setpoints information
• Alarm and operational message annunciators
• Hysteresis function t o prevent chattering of relays around
the setpoint
• High-end transmitter model with 4-20 mA current output
for datalogging purposes (pH CTP0100 only)
• Built-in memory backup to ensure that calibration and
other information will not be erased if power supply fails
• Switchable mains voltages of 110 VAC or 220 VAC via
user selectable internal j umper
3
Page 5
2. Getting Acquainted
SETU
Y
Y
2.1 Front Panel
The front panel consists of a 3 digit LED display together with 5 LED annunciators. There are also 4 k eys as shown below.
MODE
8 8 8
pH/ OR
The keys available are the ▲ (UP/INCREMENT), ▼ (DOWN/DECREMENT), MODE and ENTER keys.
The annunciators are pH/ORP, °C, SETUP, READY, and RELAY. The function of the annunciator is to provide additional information, for exampl e, the pH/ORP annunciator lights up i n pH/ORP measurement mode. The READY annunciator lights up when the pH/ORP stabilizes. RELAY lights up when any of the HI SET or LO SET relays i s activated.
The MODE key allows you to select between pH or ORP display, the temperature display or the SETUP menu di splay. While in one of the SETUP menus, it als o functions as an ESCAPE key. For example, while setting the Hi SET point, you can press MODE key if you change your mind about changing this information.
You can confirm changes or to enter into further levels of the lower menu by pressing the ENTER key. The ▲ (UP/INCREMENT) and ▼ (DOWN/DECREMENT) keys allow you to change information or t o select between different menus . Holding down the key increases the scrolling speed, i.e. changeover of 1st digit (ones) t o 2nd (tens) and then 3rd digit (hundreds).
o
REA D
P
RELA
4
Page 6
2.2 Back Panel
The back panel consists of t wo connectors . The f irs t one is a 4­way screw terminal and the second is a 12-way screw term inal. Refer to the label on top of the unit for diagram. The connection for the 4-way screw termi nals are (from the left to right):
1. PT 100 connection
2. PT 100 connection
3. No connection
4. No connection The connections for the 12-way screw term inals are (from left
to right),
5. Low Set Relay deactivated posit i o n
6. Low Set Relay center pole
7. Low Set Relay activated position
8. High Set Relay deactivat ed position
9. High Set Relay center pole
10. High Set Relay activat ed position
11. 4-20 mA - ve connection (for transmitter m odel s only)
12. 4-20 mA + ve connection (for transmitter m odel s only)
13. Protective earth
14. Protective earth
15. Neutral
16. Live
5
Page 7
2.3 Selecting pH/ORP Measurement Mode
You can set the appropriate pH/ORP measurement mode from the jumper. See Appendix 1 for details on the jumper positions for pH mode and Appendix 2 for jumper positions for ORP
mode.
2.4 Wiring
Connect the power supply to the GND (EARTH) - 13 or 14, NEUTRAL - 15 and LIVE - 16 screw terminals. Make sure that the power supply jumper setting matches the mains voltage (110 VAC or 220 VAC). See Appendix 2 for the jumper setting for the voltage selecti on.
Connect the pH electrode to the B NC connector at the back panel. Wire up the P T 100 t emperature probe to the PT 100 connections (see back panel connections).
Power on the controller and the display automatically shows the pH/ORP reading. The pH/ORP annunciat or l i ghts up. Once the reading is stable, the READY annunciator lights up. However, if the PT 100 temperature probe is not connected, automati c temperature compens ation does not function. You c an set the temperature at a select ed val ue. The temperature is set t o factory default at 25.0
NOTE : Eutech Instrum ent s will not be responsible for
incorrect application of the controller using improper voltage sources or wrong jumper set t ings.
o
C.
6
Page 8
3. Operating the Controller
R
o
R
R
3.1 The Main Display
Press MODE key to switch to three main displays - the pH/ORP display, the temperat ure display
7.4
pH/ORP
MODE
ENTE
and “SEt” display. Press MODE key once to get into
the temperature meas urement. The °C annunciator lights up when you are measuring temperature.
24.0
C
MODE
ENTE
The display shows current measured temperature (with A T C) or the temperature that was set i n MTC mode. Press MODE key again and the display toggles to
SEt
MODE
ENTE
SETUP
the “SEt” for SETUP menu.
Press ENTER key to go into the lower-level setup menus while the display shows "SEt". These lower-level menus allow calibration of pH/ORP, Temperature and other parameters including set password, Hi or Lo S et poi nts, and Hi or Lo Hysteresis values.
7
Page 9
4. Setting Up the Controller
R
R
R
R
R
4.1 Setting and Changing the Password
4.1.1 Setting New Password
To set a password, press MODE key
MODE
CAL
SETUP
PH
SETUP
SEt
SETUP
id
SETUP
123
SETUP
menu, press ENTER and set your desired password by following the above steps.
until the “SEt” is displayed.
ENTE
1. Press ENTER key and the display shows "CAL” and “PH” alternately. Press ▲ key once and the display shows “SEt”
MODE
ENTE
“id”.
2. Press ENTER to enter your desired password. Follow the steps below (e.g. using a
password “123”).
MODE
1. Initially, the display shows "000"
ENTE
with the first digit blinking.
2. Use ▲ (UP/INCREMENT) key
to enter “1” on the 1st digit of the password. Press ENTER
MODE
ENTE
once and the second digit will blink. Similarly, enter the second digit of the pass word “2” and press ENTER key again. Repeat with the third digit accordingly.
3. Press ENTER key and the
MODE
ENTE
display shows “CAL” and ”PH” alternately. Use ▲ or ▼ key to get into the “SEt” and “id”
1. Once completed, press ENTER to confirm and then press MODE key to return to the pH display.
8
Page 10
To calibrate the controller at any time, you may have to enter the password that you set, in order t o access the calibration mode. Once you have entered the pass word correctly, the display shows "CAL PH” indicating that you are i n one of the lower-level SETUP menus.
If you enter the wrong password, the display revert s back to the pH display. Alternatively, if you prefer no password protection, set the password to “000”, "CAL PH" immediately displays aft e r you press ENTER key while you are in the "SEt" menu.
NOTE : The user set password is a protec tion code. Thus, it is very important to keep thi s password strictly confi dential to authorized personnel. You are advised to remember the password that you have set, in order to prot ect the controller settings and prevent any unauthorized tampering to the system!
IMPORTANT : In case the password set is forgotten, use the master password “555”.
4.1.2 Changing the Password
Enter the Set id menu with already set password or “555”. Change a new password as per steps mentioned for setting the password.
9
Page 11
5. Calibrating the Controller
5.1 The Lower Level Menus
The "CAL PH/CAL OrP" display is the first menu seen upon entering the lower-level menus, depending on which mode is set. Press ▲ or ▼ key to display the various menus as shown in the figure below. Press ENTER key to go to the lower level menus.
START
CAL pH /
SEt id
Lo Hys
CAL OrP
CAL oC
Hi
SEt
Hi
Hys
Lo SEt
5.2 Calibrating for pH Measurement
5.2.1 The CAL PH menu
1. Enter this menu by pressing t he MODE key twice to the “SEt” display if you are in measurement mode.
2. Press ENTER key. If the password has been s et earlier, key in the password using the m ethod des cribed in s ect ion
4.1. Press ENTER once to confirm. After you set the correct password, you can see the “CAL” and “PH” displays, blink al ternately.
3. Press ENTER key to enter i nto the pH calibration mode. Dip both the pH electrode and temperature probe in the standard solution for a few mi nutes.
4. When “CAL PH” displays, press ENTER key. The current measured value (7 ±0.3 pH) will be displayed. Once the reading stablizes, press ENTER to confirm.
5. Use ▼ keys to scroll to the other buf fer values “PH4” and “PH10”, in that sequence. Press ENTER to confirm and
10
Page 12
complete the pH cali brat i on, as fo pH 7.0. Press MODE
R
R
R
key to return to measurement mode.
Note : If the pH value is not near
CAL
SETUP
MODE
ENTE
to any buffer, pressing ENTER will not calibrate the controll er. If at any point, you decide not to calibrate, you can return to the previous mode by pressing the
MODE key.
Changing the Offset
5.3
If the controller cali bration has
PH
SETUP
MODE
ENTE
drifted, you can reset the of fset of controller. First, go to the
7.0
SETUP
MODE
ENTE
calibration mode. Press ENTER when pH7 is being displayed and the present pH value is shown
(blinking). Using the ▲ or ▼ key, change the current pH to 7.0. Press ENTER to confirm, and it resets the pH calibrati on curve completely.
NOTE : Gently stir the electrode in a container f illed with buffer solution to create a hom ogenous sample. Remem ber to rinse the electrode before calibrati ng t o the next solution. Minimum two points calibration is rec ommended. For full range accuracy, three points calibration i s advised.
This is useful for on-line operations. For example, the controller while in use shows a reading of pH 8.4. Using a pre-calibrated portable meter, a measurement is made and the reading shows pH 8.6. The offset of controller can be set to 8.6 by going into the CAL m ode. While the controll er displays “PH7”, press ENTER. The current reading will be displayed. Using the ▲ or ▼ key, change the reading to pH
8.6 to reset the offset. Press ENTER to confirm offset setting. Press MODE key returns to the controller back to the measurement m ode. This allows a one-point calibration to be made while the controller is still in operation to adjust f or slight drifts of t he controller. For details, prefer to Appendix 3.
11
Page 13
5.4 Calibrating for ORP Measurement
R
R
R
R
R
1. Make sure that the ORP mode is set before calibration.
2. Enter this menu by pressing MODE key twice to the “SEt”
SEt
SETUP
display if you are in meas urement mode.
3. Press ENTER key. If the password has been set earlier, key in t he password using the method described in section 4.1. Press
CAL
SETUP
ENTER once to confirm. After you set the correct password, you can see the “CAL” and “OrP” displays blink alternately.
4. Press ENTER key to enter into the ORP calibration mode. Di p both
OrP
SETUP
the ORP electrode and temperat ure probe i n the standard solution for a few minutes .
5. When “CAL OrP” displays, you can see the present mV value blinking, press ▲ or ▼ key to adjus t the mV value. To confirm press ENTER key.
6. Press MODE key to return the controller to its measurement m ode.
5.5 Calibrating for Temperature Measurement
5.5.1 The CAL °°°°C Menu
1. Enter this menu by pressing MODE key twice to the “SEt” display if you are in measurement mode.
1. Press ENTER key. If the password has been set earlier, key in t he password using the method described in section 4.1. Press ENTER once to confirm. After you set the correct password, you see
CAL
SETUP
o
C
SETUP
MODE
ENTE
MODE
ENTE
MODE
ENTE
MODE
ENTE
MODE
ENTE
12
Page 14
the “CAL” and “PH” display blinks alternatively. P ress ▼
R
A
R
R
key once, the display s hows “CAL”“
o
C” blinking.
2. Press ENTER and the display alternatively shows “Atc”
and “On” or “Atc” and “OFF”. Use ▲ and ▼ keys to choose between both ATC ON and ATC OFF.
5.5.2 ATC ON/OFF
tc
SETUP
On
SETUP
23.5
SETUP
MODE
ENTE
MODE
ENTE
MODE
ENTE
ATC ON : If a PT 100 is connected, use ▲ and ▼ keys to adjust the temperature offset of the PT 100 by ± 5 °C. Dip the temperature probe into the sample liqui d. Mak e sure that the display is s et to ATC ON, press ENTER and the display now shows the actual temperat ure reading (blinking).
Use ▲ and ▼ keys to adjus t the reading to its actual temperature ­as measured by an external thermometer. Onc e done, press ENTER key and the display will flash “CAL” “PH” alternatively.
13
Page 15
R
A
tc
R
R
SETUP
OFF
SETUP
MODE
ENTE
MODE
ENTE
ATC OF F : If a PT 100 i s not used, then the ATC should be set to OFF. In step 3 above, choose by pressing ▲ and ▼ keys to select ATC OFF. Then press ENTER. The display will now show the default of 25
o
C or the last set value (blinking). Use ▲ and ▼ keys to set your desired value. Pres s ENTER to confirm and the display shows “CAL” “Con”.
mode.
25.0
SETUP
MODE
ENTE
Note : For ATC OFF, you can adjust the set temperature values from 0.0 to 99.9
o
C. This value will be used for its temperature compensation e.g. pH i n the MTC
14
Page 16
5.6 Setting the Alarm Feature
R
R
R
T
This menu allows you to change the High and Low Set poi nt and Hysteresis values. See Section 8 for hysteresis applications.
IMPORTANT : When SETUP mode is entered, the 4-20 mA output (only for transmi tt er CX model) freezes and t he relay de­activates (if it was in an al arm condition).
5.6.1 The Hi SET Menu
1. Enter this menu by pressing MODE key twice to the “SEt” display if you are in measurement m ode.
2. Press ENTER key. If the password has been set earlier, key in the password using the method described in sec tion
4.1. Press ENTER once to confirm.
3. After you set the correc t password, you see the “CAL” and “PH” display blinks alternately. Press ▼ key twice, the display shows “Hi” “SEt” blinking alternately.
1. Press ENTER key to access the Hi SET menu and the displ ay shows the last Hi SET value. Use ▲ and ▼ keys to change the value of the Hi SET point.
2. Press ENTER to confirm the value of the Hi SET point.
3. Press MODE key to exit to the measurement m ode. You can
E : The Hi SEt relay
activates if the c urrent pH/ORP reading exceeds Hi SET point + ½ Hi HYS value. The Hi SET value cannot be lower than the Low SET value.
press MODE key (as an ESCAPE key) to revert to the measurement value if ENTER is not pressed; the set value i s not stored into memory.
SEt
SETUP
Hi
SETUP
8.5
SETUP
MODE
ENTE
MODE
ENTE
MODE
ENTE
15
Page 17
5.6.2 The Hi HYS Menu
R
R
R
HYS
SETUP
Hi
SETUP
0.2
SETUP
MODE
ENTE
MODE
ENTE
MODE
ENTE
NOTE : The maximum value of Hysteresis is 0.2 pH to 2.0 pH or 20 mV to 200 mV. The HI HYS hysteresis is spread equally on either side of the Hi SET point.
1. Enter this menu by pressing the MODE key to the “SEt” display if you are in the measurem ent mode.
2. Press ENTER key. If the password has been set earlier, key in the password using the method described in sec tion
4.1. Press ENTER once to confirm. After you set the correct password, you see the “CAL” and “PH” display blinks alternatively. Press ▼ key thrice, the display shows “Hi” “HYS” blinking.
1. Press ENTER key to access the Hi HYS menu and the display shows the last Hi HY S
value. Use ▲ and ▼ keys to change the value of the Hi HYS point.
2. Press ENTER to confirm the value of the Hi HYS.
3. Press MODE key to exit to
measurement m ode. You can press MODE key (as an ESCAPE key) to revert to the measurement value if ENTER is not pressed; the set value i s not stored into mem ory.
16
Page 18
5.6.3 The Lo SEt Menu
R
R
R
1. Ent er this menu by pressi ng MODE key twice to the “SEt” display if you are in measurement mode.
2. Press ENTER key. If the password has been set earlier, key i n the password using the method described in section 4.1. Pr ess ENTER once to confirm.
3. Af ter you set the corr ect password, you see the “CAL” and “PH” display blinks alternately. Press ▼ key four times, the display shows “Lo” “SEt” blinking alternately.
1. Press ENTER key to access the Lo SET menu and the display shows the l ast Lo SET value. Use ▲ and ▼ keys to change the value of the Lo SET point.
2. Press ENTER to confirm the value of the Lo SET point.
3. Press MODE key to exit to measurement mode. You can press MODE key (as an ESCAPE key) to revert to the measurement value if ENTER is not pressed; the set value is not stored into memory.
SEt
SETUP
Lo
SETUP
5.0
SETUP
MODE
ENTE
MODE
ENTE
MODE
ENTE
NOTE : The Lo SEt relay activates if the current pH/ORP reading exceeds Lo SET point - ½ Lo HYS value. The Lo SET value cannot be higher than the Hi SET value.
17
Page 19
5.6.4 The Lo HYS Menu
R
R
R
HYS
SETUP
Lo
SETUP
0.2
SETUP
MODE
ENTE
MODE
ENTE
MODE
ENTE
NOTE : The maximum value of Hysteresis is
0.2 pH to 2.0 pH or 20 mV to 200 mV. The Lo HYS hysteresis is spread equally on either side of the Lo SET point.
1. Ent er this menu by pressi ng the MODE key to the “SEt” display if you are in the measurement mode.
2. Press ENTER key. If the password has been set earlier, key in the password using the method described in s ection
4.1.
3. Press ENTER once to confirm. After you set the correct password, you see the “CAL” and “PH” display blinks alternatively. Press ▼ key five times, the display shows “Lo” “HYS” blinking alternately.
1. Press ENTER key to access
the Lo HYS menu and the display shows the l ast Lo HYS value. Use ▲ and ▼ keys to change the value of the Lo HYS point.
2. Press ENTER to confirm the value of the Lo HYS. Press MODE key to exit to the measurement mode. You can press MODE key (as an ESCAPE key) to revert to the measurement value if ENTER is not pressed; the set value is not stored into memory.
IMPORTANT : The Lo SET + ½ Lo HYS should be less than Hi SET - ½ Hi HYS i.e. Hi SET and Lo SET value (in consideration of the hysteresis band) can never overlap. Lo HYS and Hi HYS can be set independent of each other; this allows non-symmetrical hysteresis option.
18
Page 20
6. Using the Controller Current Loop for Datalogging (for Transmitter Model Only)
The 4-20 mA Current Loop
A 4-20 mA current loop can be connected if a remote data logging is required. The current will be proportional to the pH/ORP displayed on the panel displ ay. The 4-20 m A current loop can drive a load resistance of no more than 200Ω.
If the pH is less than 0.0 pH, the current s ignal will be set to approximately 0 mA. If the pH is great er than 14.0 pH, it will be set to approximately 21 mA . I n thi s way, the remote data logger can detect when out of range conditions aris e. Pleas e note that the 4-20 mA loop is not galvani cally isolated from the s ource.
19
Page 21
7. Additional Information
A
A
The controller allows you to set High and Low alarm s that switch on or off relays, and ac tivating or deactivating devices linked to the controller. In cases where the pH/ORP values fluctuate close t o the high or low setpoints, the relays will continuously switch on/off very quickly and may cause problems to the linked devi ces. The hysteresis band all ows you to set an allowable range of fluctuations to prevent the relays from activating and deac tivating too quickly. S ee bel ow.
It is not possible to set the high set point lower than the l ow set point and vice versa (taking t he hysteresis band into consideration). For example, when the high set point is 8.50 pH and the high hysteresis band is set at 0.40 pH, then the low setpoint cannot be set at higher than 8.50 - ½ Hi HYS.
NOTE : The shaded area indicates the hys teresis band.
D
Hi
Hys
B
C
Hysteresis Band
E
ctual
pH
Hi
Set
Lo
Hys
Hysteresis Band
F
H
Lo
Set
G
20
Page 22
Explanation of the diagram in the previous page
F
A - Reading reaches (Hi SET - ½ HI HYS), the Hi-Set relay
remains inactivat ed.
B - Reading reaches above High Setpoint but below (Hi SET +
½ Hi HYS), Hi-Set relay remains inactivated.
C - Reading reaches between (Hi SET - ½ Hi HY S), the Hi-Set
relay remains inacti vated.
D - Reading reaches above (Hi SET + ½ Hi HYS), and t he Hi-
Set relay is activated.
E - The Hi-Set relay is inac tivated only when the reading falls
below (Hi SET - ½ Hi HYS).
F - Reading reaches between (Lo SET - ½ Lo HYS), the Lo-
Set relay remains inactivated.
G - Reading reaches below (Lo SET - ½ Lo HYS), the Lo-Set
relay is activated.
H - The Lo-Set relay remains activated until the reading goes
above the (Lo SET + ½ Lo HYS)
Appendix 1 - Jumper Positions (Inside Controller)
21
Page 23
JP1 Selects the input voltage between 110
VAC or 220 VAC
JP3 & JP4 Selects for pH or ORP mode. Please
refer to Appendix 2A and 2B for the diagrams
JP2
OPEN
SHORT
(DEFAULT)
Terminals 5 & 6,
8 & 9
6 & 7, 9 & 10
5 & 6,
8 & 9
6 & 7,
9 & 10 Power On OPEN CLOSE CLOSE OPEN pH within
OPEN CLOSE CLOSE OPEN
setting pH
CLOSE OPEN OPEN CLOSE exceeding setting
Fuse Note that there is a fuse internal to the
controller. Before opening the unit,
ENSURE that the power cable is
physically separ ated from the mains
supply. Replace the fuse with the one
recommended by the manufacturer.
22
Page 24
Appendix 2A - Jumper Positions, pH
Jumper Positions - inside the Controller
23
Page 25
Appendix 2B - Jumper Positions, ORP
Jumper Positions - inside the Controller
24
Page 26
Appendix 3 - pH Offset / Slope
A
mV Axis
Current pH curve
pH Axis
25
4.5
djusted pH curve
7.0
7.5
10.0
10.5
Page 27
8. Technical Specifications
pH Range 0.0 to 14.0 pH Resolution 0.1 pH Relative Accuracy +/- 0.1 pH mV Range -1250 to +1250 mV Resolution 10 mV Relative Accuracy +/- 10 mV Temperature Compensation Temperature 0 to 100 oC
Resolution 0.1 oC Relative Accuracy +/- 0.5 oC Sensor PT 100
Calibration Push-bush 1 to 3 points Calibration Options pH 4.0, 7.0 and 10.0 Output 4-20 mA, screw terminals (non-isolated);
Display LED, 3 digits Inputs BNC, screw terminals Input Impedance Input Bias Current 50 pA max. at 25 oC Recommended Input Cable Length Relays
No. Of Relays Maximum Voltage Maximum Current High Hys. Band Low Hys. Band
Power Requirements 110 VAC or 220 VAC (Jumper selectable) Environmental Requirements Operating
Storage Humidity Limits
Storage Temp. Range 0 to 50 oC Dimensions 1/8 DIN size; 96 (L) x 48 (H) mm (Front panel)
Automatic / Manual (0 to 100.0 oC)
Max. load 200 Ω
10 ^ 12 Ω
Less than 5 meters
2 - High set & Low set SPDT
240 VAC
0.2 to 2.0 pH or 20 to 200 mV
0.2 to 2.0 pH or 20 to 200 mV
10 to 95% RH (non condensing)
3A
0 to 50 oC
-10 to 60 oC
26
Page 28
NOTES
27
Loading...