This manual serves to explain the use of the αlpha-CON 560 Industrial
Conductivity Controller. The manual functions in two ways, firstly as a step by step
guide to help the user operate the instrument. Secondly, it serves as a handy
reference guide. This instruction manual is written to cover many anticipated
applications of the αlpha-CON 560 controller. If you have doubts in the use of the
instrument, please do not hesitate to contact the nearest Eutech Instruments
Authorised Distributor.
The information presented in this manual is subject to change without notice as
improvements are made, and does not represent a commitment on part of Eutech
Instruments.
Eutech Instruments cannot accept any responsibility for damage or malfunction of the
unit due to improper use of the instrument.
10.1 Appendix 2 Conductivity of various aqueous solutions....................................................... 36
10.2 Appendix 2 Graphical Representation of the Function of Hysteresis ................................. 37
10.3 Appendix 3 General Instructions Concerning Controller Setting ........................................ 38
10.4 Appendix 4 - Abbreviations Used in LCD............................................................................. 39
Instruction Manual αlpha-CON 560
1. Introduction
1.1 Before You Begin
We thank you for purchasing the αlpha CON 560 controller.
The construction of the CON 560 controller employs leading edge technology
and complies with safety regulations currently in force. Notwithstanding this,
improper use could lead to hazards for the user or a third-party, and/or
adverse effects on the plant or other equipment. Therefore, the operating
instructions must be read and understood by the persons involved before
working with the controller.
The instruction manual must always be stored close at hand, in a place
accessible to all people working with the conductivity controller.
If you have questions, which are not or insufficiently answered in this
instruction manual, please contact your authorized supplier. They will be glad
to assist you.
1.2 Intended Use
αlpha CON 560 controllers are intended solely for conductivity and
temperature measurement, as described in this instruction manual.
Any other use, or use not mentioned here, that is incompatible with the
technical specifications is deemed inappropriate. The operator is solely
responsible for any damage arising from such use.
Other prerequisites for appropriate use include:
– Comply with the instructions, notes and requirements set out in this
instruction manual.
– Comply with all local safety regulations concerning safety at work.
– Comply with all information and warnings in the documentation dealing
with the products used together with the conductivity controller (housings,
sensors, etc.).
– Comply with local environmental and operational conditions.
1
Instruction Manual αlpha-CON 560
1.3 Safety Instructions
- The αlpha CON 560 controller should be installed and operated only
by personnel familiar with the instrument and who are qualified for
such work.
- A defective CON controller must neither be installed nor put into
service.
- The αlpha CON 560 must only be operated under the specified
operating conditions (see section 6).
- The αlpha CON 560 must not be repaired by the customer.
- No modifications to the αlpha CON 560 are allowed. The
manufacturer/supplier accepts no responsibility for damage
caused by unauthorized modifications. The risk is borne entirely
by the user.
1.4 Taking Out of Service / Correct Disposal of the Unit
Taking out of Service
• First disconnect the unit from the power supply and then undo all electrical
connections.
• Remove the unit from the wall / panel.
Correct Disposal of the Instrument
When the αlpha CON 560 controller is permanently taken out of service,
obey the local environmental regulations for correct disposal or send the
instrument to your local distributor, they will take care of proper disposal.
2
Instruction Manual αlpha-CON 560
2. Getting Started
2.1 Description of Instrument
The Eutech Instruments’ αlpha CON 560 controller is used for measuring
conductivity and temperature values. The controller can be used for
applications such as water treatment and controlling, galvanicdecontamination, chemical processing, food processing, clean or wastewater
control and neutralization processes.
The controller has many user-friendly and safety features which include:
• Push-button keypad for calibration and setup
• Built-in non-volatile memory to ensure that calibration and other
information are not erased if power supply fails
• Menu-driven program that simplifies setup
• Automatic temperature compensation (ATC)
• Manual temperature compensation setting without the ATC probe, with
independent setting for calibration and process temperature
• Large dual display LCD for easy reading with clear multiple annunciators,
operational mode indicators and error indicators
• Two switching contacts as set-point relays
• Limit (on/off) control mode
• Separately adjustable high and low set-point hysteresis (dead bands) to
prevent chattering of relays around the set points
• 0 to 2000 seconds time delay adjustment on all relays – minimizes false
alarms
• Backlight LCD display
3
Instruction Manual αlpha-CON 560
A
2.2 Measurement and Control System
A typical measurement system consists of:
• A conductivity process controller
• A conductivity sensor with integrated or separate Pt100/Pt1000
temperature sensor
• An appropriate measurement cable
• An immersion, flow or process assembly
• A final control element such as pump or valve
lpha CON 560 Controller
Pump or Valve
lpha pH560
ESC
CAL
pH M onitor
ENT
Relay Output
Power Adaptor
Housing and
Sensors
Measurement Cable
(+9 V DC)
4
Instruction Manual αlpha-CON 560
2.3 Connecting Peripherals
2.3.1 Connection Terminals
Remove Back Cover:
Remove the screws from the four corners at the back of the conductivity
controller. Remove the back cover. The connectors are exposed on the back
PCBA as shown in the Figure 1 below.
Connectors:
• J8 – 9V DC power connector (1.2mm diameter)
• J9 – Relay connections
• J10 - Conductivity electrode & Temperature probe connections (wiring has
to be done in the detachable connector
J10 Connections
1. CON Low Current
2. CON Low Voltage
3. CON High Voltage
4. CON High Current
5. Pt 100 Compensate
6. Pt 100 Sense
7. Pt 100 GND
8. (No connection)
J9 Connections
Relay 1 & 2
J8
9V DC Power
Screw
J9
REL 1
J10
REL 2
Figure 1: Outer Side of Back PCBA
5
Instruction Manual αlpha-CON 560
2.3.2 Switching Between Pt100 & Pt1000 Temperature Sensors
The controller supports both Pt100 & Pt1000 (2-wire or 3-wire) temperature
sensors. The default factory setting is Pt100. If you need to use Pt1000
temperature sensor, you have to change the jumper setting (J7) as described
below.
Remove Back Cover:
Remove screws from the four corners at the back of the conductivity
controller. Remove the back cover.
Remove Back PCBA:
Remove the screw located center of the back PCBA (
Figure 1). Detach the
back PCBA from the controller. Turn over the back PCBA. Locate J7 jumper
on the inner side of the back PCBA as shown in Figure 2 below.
J7
Jumper Setting for
Pt100
(Factory Default)
Jumper Settings for
Pt1000
Figure 2: Inner Side of Back PCBA
Set Jumper J7:
Set the J7 jumper to required sensor (Pt100 or Pt1000) type.
6
Instruction Manual αlpha-CON 560
2.3.3 Connecting Conductivity Probe
Any industrial conductivity probe (2-cell or 4-cell) can be connected to the
conductivity controller.
2-Cell Conductivity Probe:
1. Short Pin 1 & 2 of J10 connector using a jumper wire.
2. Connect CON low input wire (white) to Pin 2 of J10 connector.
3. Connect CON high input wire (black) to Pin 3 of J10 connector.
4. Short Pin 3 & 4 of J10 connector using a jumper wire.
4-Cell Conductivity Probe:
1. Connect CON low input wire (black) to Pin 1 of J10 connector
2. Connect CON low input wire (white) to Pin 2 of J10 connector.
3. Connect CON high input wire (orange) to Pin 3 of J10 connector.
4. Connect CON high input wire (blue) to Pin 4 of J10 connector.
2.3.4 Connecting Temperature Probe
For Automatic Temperature Compensated (ATC) conductivity readings, an
in-built temperature sensor is usually integrated with industrial conductivity
probes. Alternatively, a separate 100Ω Pt RTD temperature probe (2-wire or
3-wire) can be connected to the controller.
3-Wire Probe:
1. Connect PT100 compensate wire to Pin 5 of J10 connector.
2. Connect PT100 sense wire to Pin 6 of J10 connector.
3. Connect PT100 GND wire to Pin 7 of J10 connector.
2-Wire Probe:
4. Short Pin 5 & 6 of J10 connector using a jumper wire.
5. Connect PT100 sense wire to Pin 6 of J10 connector.
6. Connect PT100 GND wire to Pin 7 of J10 connector.
7
Instruction Manual αlpha-CON 560
2.4 Installation
2.4.1 Mechanical Dimensions
lpha CON560
ESC
CAL
2.4.2 Wall Mount
ENT
Conductivity Monitor
Cover the catch slots at
2
both sides with overlays
Pierce through holes at
11
both sides
8
Instruction Manual αlpha-CON 560
2.4.3 Panel Mount
Prepare panel cut-out of
1
1
92.0 mm X 92.0 mm
Remove back cover of
2
conductivity controller
and slide it through
panel cut-out
Gasket
Panel
Panel
Insert threaded rods through catch
4
until conductivity controller is held
against panel
Attach catch to both sides
3
of conductivity controller
9