Eurotron MicroCal T100+, MicroCal T500+, MicroCal T1100+ Instruction Manual

MicroCal T100+ MicroCal T500+ MicroCal T1100+
Portable Temperature Calibrators
Instruction Manual MM850518 Ed.01
_________________________________________________________ _________________________________________________________
Instruction Manual MM850518 ed.01
______________________________________________2_______________________________________________
INTRODUCTORY NOTE
This manual has been with all the information you need to install, operate and maintain the dry block temperature calibrator MicroCal T100+/T500+/T1100+ and its accessories.
Eurotron has used the best care and efforts in preparing this book and believes the information in this publication are accurate. The Eurotron products are subjected to continuous improvement, in order to pursue the technological leadership; these improvements could require changes to the information of this book. Eurotron reserves the right to change such information without notice.
No part of this document may be stored in a retrieval system, or transmitted in any form, electronic or mechanical, without prior written permission of Eurotron Instruments SpA.
MicroCal T series calibrators use sophisticated analogic and digital technologies. Any maintenance operation must be carried out by qualified personnel
ONLY. Eurotron supplies instructions and operative procedures for any operation on
the instrument. We recommend to contact our technicians for any support requirements.
MicroCal T is fully tested in conformity with the directive n°89/336/CEE Electromagnetic Compatibility. Eurotron shall not be liable in any event, technical and publishing error or omissions, for any incidental and consequential damages, in connection with, or arising out of
the use of this book.
All right reserved Copyright © 1998, 2004
EUROPEAN Headquarters Eurotron Instruments SpA
Viale F.lli Casiraghi 409/413 20099 Sesto S. Giovanni (MI) Tel. : +39-02 24 88 201 FAX: +39-02 24 40 286 Mail: info@eurotron.com
USA Headquarters E-Instruments Group LLC
172 Middletown Blvd – Suite B201 Langhorne, PA 19047 Tel.: 215 750 1212 FAX: 215 750 1399 Mail: info@einstrumentsgroup.com
WARNING
Hazardous voltage are present in this electrical equipment during operation.
Non-observance of the safety instruction can result in severe personal injury or property damage.
Only qualified personnel should work on or around this equipment after becoming familiar with
all warnings, safety notices, and maintenance procedures contained herein.
Only qualified personnel or our personnel should work on this equipment for maintenance
operation.
The successful and safe operation of this equipment is dependant on proper handing, operation
and maintenance.
_________________________________________________________ _________________________________________________________
Instruction Manual MM850518 ed.01
______________________________________________3_______________________________________________
CONTENTS
1 SPECIFICATIONS.......................................................................................................5
1.1 MicroCal T100+ .................................................................................................................................... 5
1.2 MicroCal T500+ .................................................................................................................................... 6
1.3 MicroCal T1100+ .................................................................................................................................. 8
2 ORDERING CODE ......................................................................................................9
2.1 MicroCal T100+ .................................................................................................................................... 9
2.2 MicroCal T500+ .................................................................................................................................... 9
2.3 MicroCal T1100+ .................................................................................................................................. 9
3 RECOMMANDATIONS .............................................................................................11
3.1 Position of the probe:.......................................................................................................................... 11
4 SAFETY INSTRUCTIONS.........................................................................................13
5 PREPARATION.........................................................................................................14
5.1 Installation........................................................................................................................................... 14
5.1.1 Removal of packaging..................................................................................................................... 14
5.1.2 Positioning the calibrator ................................................................................................................. 14
5.1.3 Supply.............................................................................................................................................. 14
5.1.4 How to fit the equalising block in MicroCal T1100+ ........................................................................ 14
6 OPERATION PROCEDURE......................................................................................16
6.1 Operation description ......................................................................................................................... 16
6.2 Description of instrument.................................................................................................................... 16
6.2.1 Thermoregulator .............................................................................................................................. 16
6.2.2 Main switch...................................................................................................................................... 16
6.2.3 Overheating warning light................................................................................................................ 16
6.2.4 Ventilation holes .............................................................................................................................. 16
6.2.5 Heating resistance........................................................................................................................... 16
6.2.6 Equalising block .............................................................................................................................. 16
6.2.7 Temperature sensors ...................................................................................................................... 17
6.3 Start-up instructions............................................................................................................................ 17
6.4 Use of the function.............................................................................................................................. 19
6.4.1 Reading the external probes ........................................................................................................... 19
6.4.2 Switch test (SW. ON SW. OFF) ..................................................................................................... 19
6.4.3 Serial communication ...................................................................................................................... 19
6.5 Re-calibration methods....................................................................................................................... 19
7 TYPICAL FAULTS ....................................................................................................21
8 REGULATOR ............................................................................................................22
8.1 REGULATOR’S FUNCTIONS ................................................................................22
9 MICRO-PROCESSOR REGULATOR .......................................................................26
10 WARRANTY..............................................................................................................27
10.1 Warranty terms ................................................................................................................................ 27
_________________________________________________________ _________________________________________________________
Instruction Manual MM850518 ed.01
______________________________________________4_______________________________________________
10.2 Letter of conformity.......................................................................................................................... 27
APPENDIX ........................................................................................................................28
A1 EMC Conformity ................................................................................................................................. 29
_________________________________________________________ _________________________________________________________
Instruction Manual MM850518 ed.01
______________________________________________5_______________________________________________
1 SPECIFICATIONS
1.1 MicroCal T100+
Environmental range: temperature +10÷+35°C, R.H. max. 90%. (Standard models with ø26mm hole)
Operative range : -28 ÷ +150°C (ambient temperature = 18°C)
Stability : ±0.03°C **
Resolution : 0,01/0,1°C
Reading accuracy : ±0,15°C **
Regulation & reading probe : Pt 100 class A din43760
Auxiliary inputs : Pt100 and Tc J, K, N, R, S(only for Model 2I)
Reading : °C or °F
Serial communication : RS 232
Increase gradient : 20°C/1’ **
Decrease gradient : 25°C/1’ **
Standard block : ø26 x 135mm - Ø35 x 135mm
Temperature ramps : min. 0,1°C/1’
Thermostat test : 12 Vcc.
Voltage : 230V 50Hz (100/115V by required) 50/60Hz.
Power : 300VA.
Electric protection : 2,5A T. fuse (3A F for 100-115V)
Calibrator measurements : 160x370 x h. 330 mm
Measurements of case : 520x330 x h. 500 mm
Weight : 10 Kg only calibrator; 17Kg with travelling-case.
Structure in flanged plate with rotating handle.
Thermostatic well in aluminium with a holes ø26mm or 36mm, 135mm dept. Reducer inserts: ø25,7x135mm or
ø35,7x130mm.
Regulation of the temperature with PID µcontroller.
Switch test.
Internal cryostat with Peltier elements.
Electronic control components thermally insulated with forced air sistem.
Removable upper protection grid.
Total absence of environmentally harmful cooling liquids.
Socket with main cable and protection fuses.
Display back light control.
Electromagnetic compatibility : Emission EN50081-1 / Immunity EN50082-2
NOTE: The data marked with ** has been recorded at an ambient temperature of 20°C±3, power supply
230V±10%, with Pt100 ø6 inserted in the block. The above-mentioned data keep valid for one year after the issuing of the calibrating certificate; afterwards it is necessary to carry out the oven re­calibration.
_________________________________________________________ _________________________________________________________
Instruction Manual MM850518 ed.01
______________________________________________6_______________________________________________
1.2 MicroCal T500+
Environmental range: temperature +10 ÷ +35°C, R.H. max. 90%.
Standard model with aluminium block measuring 50mm diameter
Operative range: Environment ÷ +550°C**
Stability: ±0.05°C a 450°C **
Display resolution: 0,01/0,1°C
Reading accuracy: ±0,3°C a 450°C
Regulation & reading probe: Pt 100 class A din43760
Auxiliary input : Pt100 and Tc J, K, N, R, S(only for Model 2I)
Reading: °C or °F
Interface: RS 232
Maximum ascent rate: 20°C/1’ **
Maximum descent rate: 20÷25°C/1’ ** (depending on the starting temperature)
Test wells: 4 holes ø4,5-6,5-9,5-17 x 185mm
Temperature ramps: minimum 0,1°C/1’
Thermostat test: 12 Vcc.
Power supply: 230V 50/60Hz (110/115V by request)
Power: 800VA.
Size : 140x370 x 330 mm
Case size : 330x520 x H. 500 mm
Weight of calibrator : 10 Kg (17Kg including carrying case and accessories).
) Model with aluminium-bronze block measuring 50mm diameter
Operative range : Environment ÷ +600°C**
Stability : ±0.1°C a 450°C **
Maximum ascent rate : 20÷22°C/1’ **
Maximum descent rate : 8÷12°C/1’ ** (t = 300°C).
Weight of calibrator : 17Kg
Test wells with perforation standard or by request
) Model with aluminium block measuring 65mm diameter:
Operative range : Environment ÷ +550°C**
Stability : ±0.1°C a 450°C **
Maximum ascent rate : 15÷20°C/1’ **
Maximum descent rate : 3÷5°C/1’ ** (t = 300°C).
Weight of calibrator : 10Kg
Test wells with perforation by request
) Model with aluminium-bronze block measuring 65mm diameter:
Operative range : Environment ÷ +600°C**
Stability: ±0.1°C a 450°C **
Maximum ascent rate : 15÷20°C/1’ **
Maximum descent rate : 3÷5°C/1’ ** (t = 300°C).
Weight of calibrator : 17Kg
Test wells with perforation by request
Structure in flanged plate with rotating handle.
Micro-processor operated temperature regulator.
Automatic resetting safety thermostat.
Switch test.
Internal oven in stainless steel.
Electronic control components thermally insulated.
Forced air cooling system.
Removable upper protection grid.
Total absence of environmentally harmful cooling liquids.
Socket with main cable and protection fuses.
Display back light control.
Electromagnetic compatibility : Emission EN50081-2 / Immunity EN50082-2
NOTE: The data marked with ** has been recorded at an ambient temperature of 20°C±3, power supply
230V±10%, with Pt100 ø4.5mm inserted in the block.
The above-mentioned data keep valid for one year after the issuing of the calibrating certificate;
_________________________________________________________ _________________________________________________________
Instruction Manual MM850518 ed.01
______________________________________________7_______________________________________________
afterwards it is necessary to carry out the oven re-calibration.
_________________________________________________________ _________________________________________________________
Instruction Manual MM850518 ed.01
______________________________________________8_______________________________________________
1.3 MicroCal T1100+
Environmental range: temperature +10÷+35°C, R.H. max. 90%.
Operative range: 200÷+1100°C **.
Stability: ±0.3°C**.
Maximum ascent rate: 18°C/min.**(from 20 to 1000°C)
Maximum descent rate: 6°C/min.**( from 1000 to 600°C) - 3.8°C/min**( from 700 to 400°C)
Stability time: 20 min.
Display resolution: 1/0,1°C
Radial uniformity: ±0.4°C
Axial uniformity: ±0.4°C (on 100mm)
Reading precision: ±0.3% ±1 digit V.F.S.
Interface: RS 232
Power supply: 230V 50/60Hz (115/100V by request) 850 VA
Dimension of internal oven: ø45x200mm
Size: 170x330xh.450 mm
Weight: 12 Kg
Structure in flanged plate with handle.
Micro-processor operated temperature regulator.
Manual resetting safety thermostat.
The internal oven consist of a quartz tube hearted by a ceramic resistance, all mounted in an aluminium structure.
Standard insert Inconel made, ø44mmx175mm dept, fitted with 4 holes 7-9-11-13.5mm in diameter, 155 mm deep.
Possibility of special drilling.
Thermocouple for regulation and reading; thermocouple connected at the thermostat.
Forced air cooling system.
Upper protection grid.
Socket with main cable and protection fuses.
Electromagnetic compatibility : Emission EN50081-1 / Immunity EN50082-2
NOTE: The data marked with ** has been recorded at an ambient temperature of 20°C±3, power supply
230V±10%, thermocouples inserted in the block.
The above-mentioned data keep valid for one year after the issuing of the calibrating certificate;
afterwards it is necessary to carry out the oven re-calibration.
MICRO-PROCESSOR DATA
Display: 2 lines 20ch x line (3.2x5.5) back lighting.
Resolution: 0.01°C/0.1°C.
processor: 80C552
A/D converter: Σ- 24 bits
E2PROM memory for recording parameters.
RS232 Single serial output.
_________________________________________________________ _________________________________________________________
Instruction Manual MM850518 ed.01
______________________________________________9_______________________________________________
2 ORDERING CODE
2.1 MicroCal T100+
Cat. 8172 - A - B - C
Table A Options
0 None 1 Aluminium carrying case (EE880063) 3 Two channels TCs & Pt100 auxiliary input 4 Vinyl carrying case (BB880047) 9 Special drilling insert
Table B Power Supply
1 120 Vac with USA plug 2 230 Vac Schuko plug 3 240 Vac UK plug 4 230 Vac European plug 5 100 Vac with Japan/USA plug
Table C Calibration Certificate 1 Eurotron certificate
2.2 MicroCal T500+
Cat. 8174 - A - B - C
Table A Options
0 None 1 Aluminium carrying case (EE880063) 2 Extended temperature range to 600°C (MicroCal T500 only) 3 Two channels TCs & Pt100 auxiliary input 4 Vinyl carrying case (BB880047) 9 Special drilling insert
Table B Power Supply
1 120 Vac with USA plug 2 230 Vac Schuko plug 3 240 Vac UK plug 4 230 Vac European plug 5 100 Vac with Japan/USA plug
Table C Calibration Certificate 1 Eurotron certificate
2.3 MicroCal T1100+
Cat. 8176 - A - B - C
Table A Options
0 None 1 Aluminium carrying case (EE880064) 3 Two channels TCs & Pt100 auxiliary input 4 Vinyl carrying case (BB880047) 9 Special drilling insert
Table B Power Supply
1 120 Vac with USA plug