G448-0002
ULTRA SLIMPAK
®
G448-0002
Bridge Input Field Configurable Isolator
Provides an Isolated DC Output
in Proportion to a Bridge/Strain Gauge Input
Adjustable Excitation 1 to 10V with up to 120mA drive
Field Configurable Inputs from 10mV to ±200mV
(0.5mV/V to >50mV/V)
Field Configurable Outputs:
0-5V, 0-10V, 0-1mA, 0-20mA and 4-20mA
Description
The G448 is a DIN rail mount, bridge or strain gauge input signal
conditioner with 1800VDC isolation between input, output and
power. The field configurable input and output offer flexible, wide
ranging capability for bridge or strain gauge input applications
from 0.5mV/V to over 50mV/V.
Wide ranging, precision zero and span pots allow 50% adjustablity
of offset and gain within each of the 11 switch selectable input
ranges. The output can be set for either 0-5V, 0-10V, 0-1mA, 020mA or 4-20mA.
Application
Three way isolation in the G448 completely eliminates ground
loops from any source. Isolation protects expensive SCADA
systems from ground faults and provides filtering for noise
reduction which can be a significant problem with small, millivolt,
bridge signals. Wide ranging flexibility allows the user to easily
zero out dead-loads in weighing systems or configure bipolar
input ranges for expansion-compression or vacuum-pressure
bridge applications.
Diagnostic LED
The G448 is equipped with a dual function LED signal monitor. The
green, front mounted LED indicates both DC power and input
signal status. Active DC power is indicated by an illuminated LED.
If the input signal is more than 110% of the full scale range, the LED
will flash at 8Hz. Below -10%, the flash rate is 4Hz.
Ultra Slim Housing for High Density Installations
Flexible Power Supply Accepts 9 to 30 VDC
ASIC Technology for Enhanced Reliability
RoHS Compliant
Configuration
The G448 has 11 input range switch settings. Trim potentiometers
allow 50% input zero and span adjustablity within each of the 11
full-scale, input ranges.
For example, the 200mV switch setting in Table 1 configures the
input for a 0 to 200mV range. Since the span can be contracted
by 50%, this enables an input span as narrow as 100mV of the
range, or 50%. This span can be positioned anywhere within the
0-200mV range with a zero off-set as large as 50% of the full scale
range (e.g. 100 to 200mV input).
Unless otherwise specified, the factory presets the Model G448 as
follows:
Input Setting: 0 to 50mV
Input Range:
Excitation: 10V
Operation: Direct
Output: 4 to 20mA
The DC power input accepts any DC source between 18 and 30V,
typically a 24VDC source is used (see Accessories).
For other I/O ranges refer to Tables 1 through 4 and reconfigure
switches SW1 and SW2 for the desired input range, function,
excitation and output range.
WARNING: Do not change switch settings with power applied.
Severe damage will result!
0 to 30mV (3mV/V)

Calibration
1. After configuring the DIP switches, connect the input to a
calibrated millivolt source. Connect the output to the device (or
a load equivalent to the device) and apply power. (see Wiring
Diagram, Figure 2 or 3).
Note: To maximize thermal stability, final calibration should be
performed in the operating installation, allowing approximately 1
to 2 hours for warm up and thermal equilibrium of the system.
Excitation (+)
Bridge (+)
2. Set the calibrator to the desired minimum and adjust the zero
potentiometer for the desired minimum output.
3. Set the calibrator to the desired maximum and adjust the span
potentiometer for the desired maximum output.
4. Repeat steps 2 and 3, if necessary for best accuracy.
Table 1: G448 Input Ranges
1WSrotceleS
:yeK desolCroNO=1=
tupnI
egnaRegnaR
egnaRegnaR
egnaR
12 34 5
Vm01ot0
Vm02ot0
Vm05ot0
Vm001ot0
Vm002ot0
Vm5ot5-
Vm01ot01-
Vm02ot02-
Vm05ot05-
Vm001ot001-
Vm002ot002-
Table 2: G448 Direct
or Reverse Operation
noitarepO
esreveR
:yeK desolCroNO=1=
Excitation (-)
Bridge (-)
1WS
6
tceriD
Table 3: G448 Bridge
Excitation Settings
egdirB
noitaticxEnoitaticxE
noitaticxEnoitaticxE
noitaticxE
V1.01ot8.9
V2.5ot8.4
V01ot0
V5.2ot0
:yeK desolCroNO=1=
1WS
78
Table 4: G448 Output Settings
tuptuO
12345678
V5+ot0
V01+ot0
Am1ot0
Am02ot4
Am02ot0
:yeK desolCroNO=1=
2WS

Figure 1: G448 Factory Cal: 0 to 30mV (0 -50mV Switch Settings)
10V Excitation, Direct Operation, 4-20mA Output
Figure 2: Wiring Diagram for G448-0001/0002
(-0001 for reference only, not an active product)
Figure 3: Wiring Diagram for G448-0000
(for reference only, not an active product)
Note: All Ultra SlimPak modules are designed to operate in ambient
temperatures from 0 to 55°C when mounted on a horizontal DIN rail. If
five or more modules are mounted on a vertical rail, circulating air or
model HS01 Heat Sink is recommended. Refer to HS01 Technical
Bulletin (#721-0549-00) or contact the factory for assistance.
Figure 4: Mounting Multiple Modules

Specifications
Input:
Voltage:
Full Scale Range: 10mV to ±200mV (Table 1).
Impedance: >1M Ohms
Overvoltage:
intermittent 400V, max.;
continuous 264V, max.
Common Mode (Input to Ground):
1800VDC, max.
Zero Turn-Up:
50% of full scale range
Span Turn-Down:
50% of full scale range
Operation:
direct or reverse acting
Output:
Voltage:
Output: 0-5V, 0-10V
Impedance: <10 Ohms
Drive: 10mA, max. (1K Ohms, min. @ 10V)
Current:
Output: 0-1mA, 0-20mA, 4-20mA
Impedance: >100K Ohms
Compliance:
0-1mA; 7.5V, max.(7.5K Ohms, max.)
0-20mA; 12V, max. (600 Ohms, max.)
4-20mA; 12V, max. (600 Ohms, max.)
Bridge Excitation:
1 to 10VDC, 120mA max.
Accuracy (Including Linearity, Hysteresis):
±0.1% typical, ±0.2% maximum of selected
input range at 25°C.
Stability:
±0.025%/°C typical, 0.05%/°C maximum, of
selected full scale input range.
Output Noise (maximum):
0.1% of span, rms, or 10mV whichever is
greater.
Response Time (10 to 90%):
<200mSec., typical.
Common ModeRejection:
DC to 60Hz: ³120dB, ³100dB for 0 -1mA range
Isolation:
1800VDC between input, output & power.
EMC Compliance (CE Mark):
EMC: EN61326-1:2013
Safety: EN61010-2:2013
LED Indication (green):
Input Range (approx.)
>110% input: 8Hz flash
<0% input: 4Hz flash
Humidity (Non-Condensing):
Operating: 15 to 95% @ 45°C
Soak: 90% for 24 hours @ 65°C
Temperature Range:
Operating: 0 to 55°C (32 to 131°F)
Storage: -25 to 70°C (-13 to 158°F)
Power:
Consumption:
2.5W typical (one 350 Ohm bridge),
4W max. (four 350 Ohm bridges).
Range: 18 to 30VDC
Weight:
0.54 lbs.
Wire Terminations:
Screw terminals for 12-22 AWG
Agency Approvals:
CSA certified per standard C22.2, No. 0-M91
and 142-M1987 (File No. LR42272).
UL recognized per standard UL508
(File No.E99775).
CE Conformance per EMC directive 2004/
108/EC and Low Voltage directive 2006/
95/EC.
RoHS Compliant
Ordering Information
Models & Accessories
Specify:
1. Model: G448-0002
2. Accessories: (see Accessories)
3. Optional Custom Factory Calibration; specify C620 with desired
input and output ranges.
Accessories
Slim Pak “G” series modules will mount on standard TS32 (model MD02) or
TS35 (model MD03) DIN rail. In addition, the following accessories are available:
HS01 Heat Sink
MD03 TS35 x 7.5 DIN rail
WV905 24VDC Power Supply (0.5 Amp)
H910 24VDC Power Supply (1 Amp)
H915 24VDC Power Supply (2.3 Amp)
MB03 End Bracket for MD03
C664 I/O Descriptive Tags
Printed on recycled pape
Dimensions
Action Instruments
Chessell
Eurotherm