5. CALIBRATION
5.1 Drop a little oil on the 5 mm standard block 3-12.
5.2 Press the Calibration key 3-8, the ‘CAL’ be shown on the Display. ‘CAL is the short
for calibration.
5.3 Press the sensor 3-7 on the standard block. The coupling symbol ((
•
)) is on if
coupling well. ‘5.0’ mm (or ‘0.197’ inch) and ‘CAL’ will be shown on the Display in tum.
When steady, Press CAL key 3-8 to confirm and then the unit return to the state of
measurement.
5.4 The calibration result will be auto-saved to the unit once confirmation. It is
unnecessary to calibrate often unless you suspect the accuracy of measurement.
6. MEASURING PROCEDURE
6.1 Press the power key 3-4 to tum on the unit.
6.2 Press the mrnlinch key 3-3 to select the right measurement unit.
6.3 Press the Sensor 3-7 onto the material surface to measure on the premise that the
material code selected is right. Be sure that coupling is well and the symbol ((•)) is on.
The reading on display is the measurement value.
6.4 The reading is held till a new measurement value is coming. The last value is held
on the display till the power is off.
6.5 2 modes to tum off the power. Manual off at any time by pressing the power key or
Auto power off after about 1 minutes from last key operation.
7. MEASURING BY VELOCITY SETTING
7.1 Press the VEL key 3-13 and the display shows the velocity set last time.
7.2 How to measure its thickness by the velocity known? The velocity can be changed
by pressing the plus key or minus key to the value of known velocity. The increment is 1
Om/s every time when pressing the plus or minus key. And the increment is 1 OOm/s if
depressing the key form ore than about 4 seconds.
7.3 Drop a little oil onto the material to measure and press the Sensor 3-7 onto the
surface. The reading on the display is the thickness if coupling well. So if we have known
the velocity of a certain material, it is easy to measure the thickness by 7 .2.
7.4 How to measure the thickness by a sample of known thickness? Just get a sample
of known thickness. Then repeat 7.2 and 7.3 till the measurement value is totally same