Electrical Test Instruments PI-250B
Section III – Operating Instructions
Copyright(C) 1994-2012 Electrical Test Instruments, LLC. All Rights Reserved
Page III - 10
1. Consult breaker manufacturer's literature to determine any necessary
precautions and expected test results. A test current of three times the
rating of the breaker should be used for this test.
2. Follow steps in pre-test installation and setup, as required, to set up the
breaker test set.
3. Connect appropriate output tap to one pole of breaker under test.
4. Setup for C.L. MOM
5. Set RANGE to lowest range that is greater than desired test current.
6. Repeatedly jog the INITIATE button, while adjusting the OUTPUT
CONTROL higher, until the desired test current is read on the ammeter.
7. Select MAINTAIN mode.
8. If desired, N.C. mode may be used; connect contacts leads to an unused
pole of the breaker.
9. Press and release INITIATE key; OUTPUT ON light should glow, timer
should run, and current display should read desired output current.
10. While test is running, observe current reading. If necessary, adjust to
correct value by using Vernier control.
11. When breaker trips, the test set output should de-energize, timer should
stop, and OUTPUT ON light should turn off.
12. Read and record trip current and trip time. Compare to manufacturer's
time/current curve. If necessary, adjust the trip unit and repeat the test.
13. Repeat the above tests for the other poles of the breaker, allowing time
for breaker to cool. Note: Results are acceptable if all poles of the breaker
trip within the manufacturer's time range for the value of test current
chosen. All poles of the breaker need not trip in exactly the same
amount of time.
14. Shut down test set, disconnect breaker, and prepare for subsequent
testing or relocation according to pre-test installation and setup
outlined above.
Timing Test, Short Time Delay
Note: The breaker's trip time for short time delay will always be less than 30
cycles and may be as low as 4-5 cycles. Therefore, setting test current will
require some practice. However, the magnitude of the test current is not
critical; for instance, if the S.T.D. is set for 1000 amperes, there will be little or no
difference in timing if a test current of 1500, 2000 or 2500 amperes is used.