Ethos 5020 Instructions Manual

1
ETHOS 5020
Auto Ranging Digital Multimeter
2
1. SAFETY INFORMATION
SAFETY SYMBOLS
Warning! Dangerous Voltage (Risk of electric shock). Caution! Refer to the user’s manual before using this Meter.
Double Insulation (Protection Class ).
Alternating Current (AC). Direct Current (DC). Either DC or AC.
Ground (maximum permitted voltage between terminal and ground).
The symbol indicating separate collection for electrical and electronic
equipment.
The RESPONSIBLE BODY shall be made aware that, if the instrument
is used in a manner not specified by the manufacturer, the protection provided by the instrument may be impaired.
The finger or any part of your body shall not be beyond the barrier of
the test probe when measuring.
Individual protective equipment must be used if HAZARDOUS LIVE
parts in the installation where measurement is to be carried out could be ACCESSIBLE.
The following safety information must be observed to insure maximum personal safety during the operation at this meter. Do not operate the meter if the body of meter or the test leads appear to be damaged. Check the main function dial and ensure it is at the correct position before each measurement. When making current measurements ensure that the circuit is not “live” before opening it in order to connect the test leads. Do not perform resistance, diode and continuity test on a live power system. Do not apply voltage between the test terminals and test terminal to ground that exceeds the maximum limit record in this manual. Exercise extreme caution when measuring live system with voltage greater than 60V DC or 30V AC.
Change the battery when the “
” symbol appears to avoid incorrect data.
Use the DMM indoor, altitude up to 2000m and temperature 5℃ to 40℃.
Maximum relative humidity 80% for temperatures up to 31℃, decreasing
3
linearly to 50% relative humidity at 40℃. Pollution Degree 2.
2. SPECIFICATIONS
2.1 GENERAL SPECIFICATIONS
Display: LCD with a max. reading of 1999. Display sign of the test lead connect to socket. Range control: Auto range & Manual range control Polarity: Automatic negative polarity indication. Zero adjustment: Automatic. Over range indication: The “OL” or “-OL” display.
Low battery indication: Display " " sign. Data hold: Display “DH” sign. Max hold: Display “MH” sign. Auto Power Off: When measurement exceeds 15 minutes without switching
mode and pressing key, the meter will switch to standby mode.
Press any key to exit standby mode. When restart the system, press and hold M.H key to disable auto power off.
Safety standards: EMC/LVD. CAT Ⅲ 600V.
The meter is up to the standards of IEC1010 Double InsulationPollution Degree 2, Overvoltage Category .
Operating environment: Temperature 32° to104°F (0°C to 40°C),
Humidity 80% RH.
Storage environment: Temperature -4° to140°F (-20°C to 60°C),
Humidity 90% RH.
Fuse: F0.5AL 250V 5 x 20mm, F10AL 250V 5 x 20mm. Power supply: 9V battery ( 6F22 ). Test lead probe: RE10, CAT 600V / CAT 1000V, 10A, L=90cm.
Dimension: 177(H) x 92(W) x 40(D) mm Weight: Approx. 320g (including battery).
2.2 ELECTRICAL SPECIFICATIONS Accuracies are ± (% of reading + number in last digit) At 23 ± 5°C ,75% RH.
2.2.1 DC Voltage
Range Accuracy Resolution 200mV
± ( 0.5%+2 )
0.1mV
2V 1mV
20V 10mV
200V 100mV 600V ± ( 0.8%+2 ) 1V
Overload protection: 600V DC or AC rms