linearly to 50% relative humidity at 40℃. Pollution Degree 2.
2. SPECIFICATIONS
2.1 GENERAL SPECIFICATIONS
Display: LCD with a max. reading of 1999.
Display sign of the test lead connect to socket.
Range control: Auto range & Manual range control
Polarity: Automatic negative polarity indication.
Zero adjustment: Automatic.
Over range indication: The “OL” or “-OL” display.
Low battery indication: Display " " sign.
Data hold: Display “DH” sign.
Max hold: Display “MH” sign.
Auto Power Off: When measurement exceeds 15 minutes without switching
mode and pressing key, the meter will switch to standby mode.
Press any key to exit standby mode. When restart the system,
press and hold M.H key to disable auto power off.
Safety standards: EMC/LVD. CAT Ⅲ 600V.
The meter is up to the standards of IEC1010 Double
Insulation,Pollution Degree 2, Overvoltage Category Ⅲ.
Operating environment: Temperature 32° to104°F (0°C to 40°C),
Humidity ≤ 80% RH.
Storage environment: Temperature -4° to140°F (-20°C to 60°C),
Humidity ≤ 90% RH.
Fuse: F0.5AL 250V 5 x 20mm, F10AL 250V 5 x 20mm.
Power supply: 9V battery ( 6F22 ).
Test lead probe: RE10, CATⅢ 600V / CATⅡ 1000V, 10A, L=90cm.
Dimension: 177(H) x 92(W) x 40(D) mm
Weight: Approx. 320g (including battery).
2.2 ELECTRICAL SPECIFICATIONS
Accuracies are ± (% of reading + number in last digit)
At 23 ± 5°C ,≤75% RH.
2.2.1 DC Voltage
Range Accuracy Resolution
200mV
± ( 0.5%+2 )
0.1mV
2V 1mV
20V 10mV
200V 100mV
600V ± ( 0.8%+2 ) 1V
Overload protection: 600V DC or AC rms