Ethicon NTLC55, NTLC75 Steps To Use

ETHICON Linear Cutter
Steps to Use
*
*Ethicon Endo-Surgery™ Linear Cutter
Alignment Slot
Steps to Use
Anvil Half
Gripping Sur face
Models available: (55mm, 75mm)
NTLC55, NTLC75
Reloads available:
SR55, SR75
Verify compatibility of all instruments and accessories prior to using the instrument. For complete product details, see Instructions for Use.
Staple Retaining Cap
Gripping Surface
1
Using sterile technique, remove the instrument from the package. To avoid damage, do not lip the Cartridge/Reload into the sterile ield.
Alignment Tab
Locking Rib
Cartridge/Reload Half
Firing Knob
Gripping Surface
Staple Height Selector
Cartridge/Reload Fork
Alignment/Locking Lever
2
If instrument is not separated into halves, separate instrument by completely disengaging the Alignment/Locking Lever.
3
Load the instrument by inserting the selectable Cartridge/Reload by placing the alignment tabs into the alignment slots and pivoting the selectable Cartridge/Reload onto the Cartridge/Reload Fork. Snap the Cartridge/Reload into position.
5
4
Remove Staple Retaining Cap by grasping the edge of the Staple Retaining Cap and lift straight up from the Cartridge/Reload. Discard the Staple Retaining Cap.
Note: The Staple Retaining Cap ensures proper staple orientation and protects the staple leg points during shipping and transportation.
Prior to placing instrument across the tissue, select the desired closed staple height. In order to select the desired staple height, push the Staple Height Selector so that it is aligned with desired color indicator, Blue 1.5 mm, Gold 1.8 mm or Green 2.0 mm. If tissue cannot be comfortably compressed to the closed staple height, or easily compresses to less than the closed staple height, either the setting or the tissue is contraindicated for the selected staple size or setting.
CAUTION: Careful consideration should be given to any pre-surgical treatment the patient may have undergone, which may require
alterations to surgical technique or alternative surgical procedures.
CAUTION: Tissue thickness should be carefully evaluated before iring any stapler.